Invacare Oxygen Equipment IRC5LXAW User Manual

Operator's Manual  
Platinum™  
Oxygen Concentrators Standard,  
with/SensO2® and HomeFill®  
Model IRC5LXAW  
Model IRC5LXO2AW  
Model IRC5LXO2AWQ  
Model IRC9LXO2AWQ  
Dealer: This manual must be given to the  
end user.  
User: Before using this concentrator, read  
this manual and save for future reference.  
For more information regarding  
Invacare products, parts, and services,  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
TABLE OF CONTENTS  
TABLE OF CONTENTS  
SPECIAL NOTES ...............................................5  
RECYCLING INFORMATION .............................6  
SECTION 1—GENERAL GUIDELINES .....................7  
Maintenance.......................................................................................... 9  
Radio Frequency Interference.......................................................... 9  
To Reduce The Risk Of Burns, Electrocution, Fire Or Injury To  
Persons .................................................................................................. 9  
Oxygen Concentration....................................................................10  
SECTION 2—FEATURES ......................................11  
SECTION 3—SHIPPING AND HANDLING ............12  
Unpacking............................................................................................12  
Inspection............................................................................................12  
Storage.................................................................................................12  
SECTION 4—TYPICAL PRODUCT PARAMETERS .14  
SECTION 5—OPERATING INSTRUCTIONS ..........19  
Introduction........................................................................................19  
Select a Location ...............................................................................19  
Recommended Guidelines for Optimum Performance............20  
Plug in Power Cord ..........................................................................20  
Connect Humidifier (If So Prescribed).........................................21  
Power Switch .....................................................................................23  
Flowrate...............................................................................................24  
SensO2 Oxygen Purity Indicators (LX O2 Sensor Model).......25  
Indicators (LX Model)......................................................................26  
Main Power Loss Alarm System....................................................26  
Elapsed Time Meter (Hour Meter)...............................................26  
SECTION 6—MAINTENANCE .............................27  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
TABLE OF CONTENTS  
Routine Maintenance........................................................................27  
SECTION 7—TROUBLESHOOTING GUIDE ............31  
LIMITED WARRANTY .................................... 33  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
SPECIAL NOTES  
SPECIAL NOTES  
Signal words are used in this manual and apply to hazards or  
unsafe practices which could result in personal injury or  
property damage. Refer to the following table for definitions  
of the signal words.  
SIGNAL WORD  
MEANING  
Danger indicates an imminently hazardous situation  
which, if not avoided, will result in death or serious  
injury.  
DANGER  
Warning indicates a potentially hazardous situation  
which, if not avoided, could result in death or  
serious injury.  
WARNING  
Caution indicates a potentially hazardous situation  
which, if not avoided, may result in property  
damage or minor injury or both.  
CAUTION  
NOTICE  
The information contained in this document is subject to  
change without notice.  
DANGER  
DO NOT SMOKE while using this device. Keep all  
matches, lit cigarettes or other sources of ignition out of  
the room in which this product is located and away from  
where oxygen is being delivered.  
NO SMOKING signs should be prominently displayed.  
Textiles and other materials that normally would not  
burn are easily ignited and burn with great intensity in  
oxygen enriched air. Failure to observe this warning can  
result in severe fire, property damage and cause physical  
injury or death.  
5
Platinum™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RECYCLING INFORMATION  
RECYCLING INFORMATION  
This product has been supplied from an environmentally  
aware manufacturer that complies with the Waste Electrical  
and Electronic Equipment (WEEE) Directive 2002/96/CE.  
This product may contain substances that could be harmful to  
the environment if disposed of in places (landfills) that are not  
appropriate according to legislation.  
The ʹcrossed out wheelie binʹ symbol is placed on this product  
to encourage you to recycle wherever possible.  
Please be environmentally responsible and recycle this  
product through your recycling facility at its end of life.  
Platinum™  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 1—GENERAL GUIDELINES  
SECTION 1—GENERAL GUIDELINES  
WARNING  
SECTION 1 - GENERAL GUIDELINES contains  
important information for the safe operation and use of  
this product.  
In order to ensure the safe installation, assembly and  
operation of the Platinum concentrator these instructions  
MUST be followed.  
CAUTION  
“Caution: Federal law restricts this device to sale or rental by  
or on order of a physician, or any other practitioner licensed  
by the law of the governmental agency in which he/she  
practices to use or order the use of this device.”  
DANGER  
Risk of electric shock. DO NOT disassemble. Refer servicing  
to qualified service personnel.  
7
Platinum™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 1—GENERAL GUIDELINES  
WARNINGS  
The use of oxygen therapy requires that special care be taken  
to reduce the risk of fire. Any materials that will burn in air,  
and some that will not, are easily ignited and burn rapidly in  
high concentrations of oxygen. For safety concerns, it is  
necessary that all sources of ignition be kept away from the  
product and preferably out of the room in which it is being  
used. NO SMOKING signs should be prominently displayed.  
A spontaneous and violent ignition may occur if oil, grease or  
greasy substances come in contact with oxygen under  
pressure. These substances MUST be kept away from the  
oxygen concentrator, tubing and connections, and all other  
oxygen equipment. DO NOT use any lubricants unless  
expressly recommended by Invacare.  
For optimum performance, Invacare recommends that each  
concentrator be on and running for a minimum of 30 minutes  
at a time. Shorter periods of operation may reduce maximum  
product life.  
If the concentrator has a damaged cord or plug, if it is not  
working properly, if it has been dropped or damaged, or  
dropped into water, call Qualified Service Personnel for  
examination and repair.  
Keep the oxygen tubing, cord, and unit away from heated or  
hot surfaces, including space heaters, blankets, stoves and  
similar electrical appliances.  
DO NOT move or relocate concentrator by pulling on  
the cord.  
NEVER drop or insert any object into any opening.  
NEVER block the air openings of the product or place it on a  
soft surface, such as a bed or couch, where the air opening may  
be blocked. Keep the openings free from lint, hair and the like.  
Platinum™  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 1—GENERAL GUIDELINES  
Fill humidifier with water to the level shown by the  
manufacturer. DO NOT overfill.  
Keep unit at least 7.5 cm (3 in) away from walls, draperies,  
furniture, and the like.  
Invacare recommends that Crushproof oxygen tubing  
(supplied by Invacare) be used with this product and NOT  
exceed 15.2 m (50 ft) in length. Recommended use for  
Platinum 9 is up to 15.2 m (50 ft) in length of high flow tubing.  
Maintenance  
The Invacare Oxygen Concentrator was specifically designed  
to minimize routine preventive maintenance at intervals of  
once per year. Only professionals of the healthcare field or  
persons fully conversant with this process such as authorized  
or factory trained personnel should perform preventive  
maintenance or performance adjustments on the oxygen  
concentrator.  
Radio Frequency Interference  
Most electronic equipment is influenced by Radio Frequency  
Interference (RFI). Caution should be exercised with regard to  
the use of portable communications equipment in the area  
around such equipment.  
To Reduce The Risk Of Burns,  
Electrocution, Fire Or Injury To Persons  
Avoid using while bathing. If continuous usage is required by  
the physicians prescription:  
The concentrator must be located in another room at least  
2.1 m (7 ft.) from the bath.  
DO NOT come in contact with the concentrator while wet.  
A product should NEVER be left unattended when plugged in.  
DO NOT place or store product where it can drop into water  
or other liquid.  
9
Platinum™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 1—GENERAL GUIDELINES  
DO NOT reach for product that has fallen into water. Unplug  
IMMEDIATELY.  
This device is to be used only in accordance with the  
prescription of a physician and this Owners Manual. If at any  
time the patient or attendant conclude that the patient is  
receiving an insufficient amount of oxygen, the supplier  
and/or physician should be contacted IMMEDIATELY. No  
adjustments should be made to the flowrate unless prescribed  
by a physician.  
Close supervision is necessary when this product is used near  
children or physicallychallenged individuals.  
Use this product for only intended use as described in this  
manual.  
DO NOT connect the concentrator in parallel or series with  
other oxygen concentrators or oxygen therapy devices.  
Use of some administration accessories or certain humidifiers  
not specified for use with oxygen concentrator may impair the  
performance.  
In certain circumstances oxygen therapy can be hazardous.  
Invacare recommends that you seek medical advice before  
using this machine.  
This equipment is to be used as an oxygen supplement and is  
not considered life supporting or life sustaining.  
Avoid creation of any spark near medical oxygen equipment.  
This includes sparks from static electricity created by any type  
of friction.  
Oxygen Concentration  
Take care not to set the flow above RED ring. An oxygen flow  
greater than 5 L/min. (IRC5LXO2AW, IRC5LXAW,  
IRC5LXO2AWQ) and 9 L/min. (IRC9LXO2AWQ) will  
decrease the oxygen concentration.  
Platinum™  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 2—FEATURES  
SECTION 2—FEATURES  
FRONT VIEW  
Oxygen Outlet  
Oxygen Purity and Fault Indicator  
Lights (SensO model)/Fault and  
2
Power Indicator Lights (LX model)  
Flowmeter  
Power Switch  
Circuit Breaker  
Elapsed Time Meter  
(Hour Meter)  
Cabinet Filter  
(one of two  
shown)  
*NOTE: This outlet fitting is to be used only for filling oxygen  
cylinders with the HomeFill home oxygen compressor. The outlet  
fitting does not affect concentrator performance. Refer to the  
HomeFill ownerʹs manual, part number 1145804, for connection  
and operating instructions. When not in use, the plug provided  
with the concentrator should be inserted into the outlet fitting.  
For more information about the HomeFill, contact Invacare.  
Power Cord  
Outlet Fitting*  
ACCESSORIES  
(NOT SHOWN):  
HomeFill home oxygen  
compressor  
Contact your local Invacare  
representative for available  
accessory information.  
TOP VIEW  
11  
Platinum™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 3—SHIPPING AND HANDLING  
SECTION 3—SHIPPING AND  
HANDLING  
NOTE: For this procedure, refer to FIGURE 3.1 on page 13.  
Before you install and operate the concentrator, please refer to  
Platinum Series Ownerʹs Manual.  
The Platinum concentrator should ALWAYS be kept in the  
upright position to prevent cabinet damage while being  
transported. The shipping container has been designed to  
assure maximum protection of the concentrator.  
If the concentrator is to be reshipped by common carrier, it  
should be packed in a new carton. Additional cartons are  
available from Invacare.  
The air compressor suspension system has been engineered to  
withstand severe motion and orientation.  
Unpacking  
1. Check for any obvious damage to the carton or its contents.  
If damage is evident, notify the carrier, or Invacare.  
2. Remove all loose packing from the carton.  
3. Carefully remove all the components from the carton.  
NOTE: Retain carton and packing materials for storage of the  
Platinum.  
Inspection  
1. Examine exterior of the oxygen concentrator for nicks,  
dents, scratches or other damages. Inspect all components.  
Storage  
1. Store the repackaged oxygen concentrator in a dry area.  
2. DO NOT place other objects on top of the repackaged  
concentrator.  
Platinum™  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
SECTION 3—SHIPPING AND HANDLING  
FIGURE 3.1 Unpacking  
13  
Platinum™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
SECTION 4—TYPICAL PRODUCT PARAMETERS  
SECTION 4—TYPICAL PRODUCT  
PARAMETERS  
Regulatory EN/IEC 60601-1  
Listing:  
Alternating Current  
Type BF equipment  
Unit running  
Unit not running  
Attention - Consider Accompanying Documents  
DO NOT smoke  
Class II, Double Insulation  
Electrical Requirements:  
Rated Current Input:  
230 VAC ± 10% (253 VAC/207 VAC), 50 Hz  
1.4 A (IRC5LXO2AWQ)  
2.0 A (IRC5LXAW, IRC5LXO2AW)  
2.1 A (IRC9LXO2AWQ)  
Sound Level:  
Altitude:  
39.5 dB Average (IRC5LXO2AWQ)  
44 dB Average (IRC5LXAW, IRC5LXO2AW)  
50 dB Average (IRC9LXO2AWQ)  
Up to 1828 meters (6,000 ft.) above sea level  
without degradation of concentration levels.  
NOT RECOMMENDED FOR USE ABOVE  
1828 meters (6,000 ft)  
For IRC9LXO2AWQ - Up to 4,000 ft. (1230  
meters) above sea level without degradation  
of concentration levels. For 4,000 ft (1230  
meters) to 13,129 ft (4000 meters) below  
90% efficiency.  
Platinum™  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 4—TYPICAL PRODUCT PARAMETERS  
*Oxygen Output  
IRC5LXAW, IRC5LXO2AW and  
Concentration Levels:  
All 5LXO2AWQ/5LXAW/  
5LXO2AW/9LXO2AWQ  
models  
IRC5LXO2AWQ  
93% minimum at 1 to 3 L/min.  
91% minimum at 4 L/min.  
87% minimum at 5 L/min. (maximum  
recommended flow)  
* (Stated concentration levels  
achieved after initial warm-up IRC9LXO2AWQ  
period (approximately 30  
minutes).  
92% minimum at 1 to 2 L/min.  
91% minimum at 8 L/min.  
87% minimum at 9 L/min.  
93% minimum at 3 to 7 L/min.  
Maximum Outlet Pressure:  
IRC5LXAW, IRC5LXO2AW,  
IRC5LXO2AWQ  
34.5 ± 3.45 kPa (5 ± 0.5 psi)  
IRC9LXO2AWQ  
62.0 kPa ± 3.45 kPa (9 ± 0.5 psi)  
Flow Range:  
IRC5LXAW, IRC5LXO2AW,  
IRC5LXO2AWQ  
0.5 to 5 L/min. Flowrates less than 1 L/min.  
are not recommend  
IRC9LXO2AWQ  
1 L/min. to 9 L/min. Flowrates less than 1  
L/min. are not recommended.  
Low Flow Alarm:  
0 L/min. to 0.5 L/min. Rapid Audible Alarm  
Beeping (No Accessories Connected).  
Average Power Consumption: (IRC5LXO2AWQ)  
300 W  
>
280 W @ 3 L/min.  
(IRC5LXO2AW)  
365 W  
>
340 W @ 3 L/min.  
380 W (IRC5LXAW)  
475 W (IRC9LXO2AWQ)  
Pressure Relief Mechanism  
Operational at:  
241 kPa ± 24.1 kPa (35 psi ± 3.5 psi)  
15  
Platinum™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 4—TYPICAL PRODUCT PARAMETERS  
Change in maximum  
recommended flow when  
back pressure of 7k Pa is  
applied:  
0.7 L/min.  
Filters:  
Cabinet (2), Outlet HEPA and Compressor  
Inlet.  
Safety System:  
Current overload or line surge shutdown.  
High temperature compressor shutdown.  
High Pressure Alarm w/compressor  
shutdown.  
Low Pressure Alarm w/compressor  
shutdown.  
Battery Free Power Loss Alarm.  
SensO2 Oxygen System (SensO2 Model) Low  
Flow Alarm.  
Width:  
Height:  
Depth:  
Weight:  
46.7 cm ± 1 cm (18 3/8 in ± 3/8 in)  
67.0 cm ± 1 cm (6 3/8 in. ± 3/8 in)  
36.5 cm ± 1 cm (14 3/8 in. ± 3/8 in)  
IRC5LXAW, IRC5LXO2AW,  
IRC5LXO2AWQ  
23.6 kg ± 1 kg (52 lbs ± 2 lbs)  
IRC9LXO2AWQ  
24.0 kg ± 1 kg (53 lbs ± 2 lbs)  
Shipping Weight:  
IRC5LXAW, IRC5LXO2AW,  
IRC5LXO2AWQ  
26.8 kg ± 1 kg (59 lbs ± 2 lbs)  
IRC9LXO2AWQ  
27.2 kg ± 1 kg (60 lbs ± 2 lbs)  
Operating Ambient  
Temperature:  
10°C - 35°C (50°F - 95°F) at 20-60% relative  
humidity.  
Exhaust:  
Less than Ambient +19°C (+ 35°F)  
Less than Ambient +2°C (+ 4°F).  
Oxygen Output:  
Cabinet:  
Impact Resistant flame-retardant plastic  
cabinet that conforms to UL 94-V0.  
Platinum™  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 4—TYPICAL PRODUCT PARAMETERS  
Regulatory Listing:  
IRC5LXAW/IRC5LXO2AW/  
IRC5LXO2AWQ/IRC9LXO2AWQ  
ETL tested as complying with:  
EN55011  
EN61000-3-2  
EN61000-3-3  
IEC 60601-1, A1, A2  
IEC 60601-1-2  
IRC5LXO2AW, IRC5LXO2AWQ,  
IRC9LXO2AWQ Models Only  
ISO8359  
MDD  
CE marked models:  
IRC5LXO2AW, IRC5LXO2AWQ,  
IRC9LXO2AWQ  
Electrical:  
No extension cords.  
Placement:  
No closer than 7.5 cm (3 inches) from any  
wall, furniture, draperies, or similar surfaces.  
Tubing:  
IRC5LXAW, IRC5LXO2AW,  
IRC5LXO2AWQ  
2.1 m (7 ft) cannula with a maximum 15.2 m  
(50 ft) of Crush-Proof Tubing (DO NOT  
pinch).  
IRC9LXO2AWQ  
Recommended use up to 15.2 M (50 ft) high  
flow tubing with high flow cannula at all flow  
rates.  
Relative Humidity:  
Time of Operation:  
20 to 60%  
Up to 24 hours per day.  
Recommended Storage and  
Shipping Temperature:  
-29°C to 65°C (-20°F to 150°F) at 15-95%  
relative humidity.  
17  
Platinum™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 4—TYPICAL PRODUCT PARAMETERS  
SENSO2 INDICATORS  
LABEL  
SYMBOL  
STATUS  
INDICATOR LIGHTS  
(LED)  
SYSTEM OKAY  
GREEN Indicator Light  
O2 over 85% (± 2%)  
O2  
O2 Between 73% (± 3%) YELLOW Indicator light  
A. YELLOW Solid  
B. YELLOW Flashing Sensor  
Failure  
to 85% (±2%)  
Call a qualified technician.  
SYSTEM FAILURE  
O2 Below 73% (±3%)  
RED Indicator Light  
LX INDICATORS  
LABEL  
SYMBOL  
STATUS  
INDICATOR LIGHTS  
(LED)  
SYSTEM OKAY  
GREEN Indicator Light  
I/O  
SYSTEM FAILURE  
RED Indicator light  
Continuous Audible Alarm  
Sieve-GARD™ Compressor  
Shutdown  
Call a qualified technician  
Platinum™  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 5—OPERATING INSTRUCTIONS  
SECTION 5—OPERATING  
INSTRUCTIONS  
Introduction  
Your oxygen concentrator is intended for individual use in the  
home. It is an electronically operated device that separates  
oxygen from room air. It provides high concentration of  
oxygen directly to you through a nasal cannula. Clinical  
studies have documented that oxygen concentrators are  
therapeutically equivalent to other types of oxygen delivery  
systems.  
Your provider will show you how to use your oxygen  
concentrator. He/She should be contacted with any questions  
or problems regarding your oxygen concentrator. This  
owners manual will tell you about your concentrator and will  
serve as a reference as you use your concentrator.  
Select a Location  
You may select a room in your house where using your  
oxygen concentrator would be most convenient. Your  
concentrator can be easily rolled from room to room on its  
casters.  
Your oxygen concentrator will perform best when operated  
under the following conditions. Usage in environments other  
than those described may result in the need for increased  
equipment maintenance. The air intake of the unit should be  
located in a well ventilated area to avoid airborne pollutants  
and/or fumes.  
19  
Platinum™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SECTION 5—OPERATING INSTRUCTIONS  
Recommended Guidelines for Optimum  
Performance  
Temperature:  
Electrical:  
(10°C - 35°C) 50°F - 95°F  
No extension cords.  
Placement:  
No closer than 7.5 cm (3 in) from the wall,  
furniture, draperies, or similar surfaces.  
Tubing and Cannula:  
IRC5LXAW, IRC5LXO2AW,  
IRC5LXO2AWQ  
2.1 m (7 ft) cannula with a maximum 15.2 m  
(50 ft) of Crush-Proof Tubing (DO NOT  
pinch).  
IRC9LXO2AWQ  
Recommended use up to 15.2 m (50 ft) high  
flow tubing with high flow cannula at all flow  
rates.  
Environment:  
Smoke and soot-free. No confined spaces  
(Example: No closets).  
Relative Humidity:  
Time of Operation:  
Flow Rate  
20 to 60%  
Up to 24 hours per day.  
IRC5LXAW, IRC5LXO2AW,  
IRC5LXO2AWQ  
0.5 L/min. to 5 L/min. Flow rates less than 1  
L/min., are not recommended.  
IRC9LXO2AWQ  
1 L/min. to 9L/min. Flow rates less than 1  
L/min. are not recommended.  
Minimum Operating Time:  
30 Minutes  
NOTE: Ensure that your concentrator is at least 7.5 cm (3 in) away  
from walls, draperies or furniture to assure sufficient air flow. Avoid  
deep pile carpets and heaters, radiators or hot air registers.  
Plug in Power Cord  
1. Plug in power cord to an electrical outlet.  
Platinum™  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
SECTION 5—OPERATING INSTRUCTIONS  
Connect Humidifier (If So Prescribed)  
NOTE: For this procedure, refer to FIGURE 5.1,FIGURE 5.2,  
1. Fill humidifier with distilled water to the level indicated  
by the manufacturer. (Please contact Invacare for  
recommendations).  
WARNING  
DO NOT overfill humidifier.  
Replace the humidifier cap and securely tighten.  
DO NOT reverse the oxygen input and output  
connections. Water from the humidifier bottle will travel  
through the cannula back to the patient.  
Humidifier Bottle  
FIGURE 5.1 Filling the Humidifier  
2. Place the humidifier bottle in the humidifier  
compartment.  
Humidifier Bottle  
FIGURE 5.2 Humidifier Compartment  
21  
Platinum™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
SECTION 5—OPERATING INSTRUCTIONS  
3. Remove the filter access door located on the side of the  
concentrator. Insert a flathead screwdriver in the plate  
groove on the top edge of the filter access door and gently  
pry the filter access door open. Refer to Detail A” in  
4. The humidifier adapter is next to the inlet filter. Pull up  
and remove the humidifier bottle adapter. Refer to Detail  
“B” in FIGURE 5.3.  
5. Attach it to the humidifier by turning the wing nut on the  
humidifier bottle counterclockwise until it is securely  
fastened. Refer to Detail “C” in FIGURE 5.3.  
DETAIL “A”  
Plate Groove (Insert Flat  
Head Screwdriver Here)  
Filter Access Door  
DETAIL “B”  
Humidifier Bottle  
Adapter  
DETAIL “C”  
Filter  
Access  
Door  
Humidifier Bottle  
Adapter  
FIGURE 5.3 Attaching the Humidifier Bottle Adapter  
Platinum™  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
SECTION 5—OPERATING INSTRUCTIONS  
6. Attach oxygen tubing from the humidifier bottle to  
oxygen outlet connector on the oxygen concentrator.  
7. Attach the cannula/patient supply tubing to the  
humidifier bottle outlet.  
8. After assembly, ensure that oxygen is flowing through the  
cannula.  
9. Replace the filter access door on the side of concentrator  
by snapping it back into the plate groove.  
Oxygen Outlet  
Connector  
Oxygen  
Tubing  
Humidifier Outlet  
FIGURE 5.4 Attaching the Oxygen Tubing  
WARNING  
For optimum performance, Invacare recommends that  
each concentrator be on and running for a minimum of 30  
minutes at a time. Shorter periods of operation may  
reduce maximum product life.  
Power Switch  
NOTE: For this procedure, refer to FIGURE 5.5 on page 24.  
1. Press power switch to on position. All the panel lights and  
the audible alarm will come on for one second, indicating  
that the unit is functioning properly.  
23  
Platinum™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
SECTION 5—OPERATING INSTRUCTIONS  
On/Off (I/O)  
Power Switch  
FIGURE 5.5 Power Switch  
Flowrate  
NOTE: For this procedure, refer to FIGURE 5.6 on page 25.  
1. Turn the flowrate knob to the setting prescribed by your  
physician or therapist.  
WARNING  
DO NOT change the L/min. setting on the flowmeter  
unless a change has been prescribed by your physician or  
therapist.  
NOTE: To properly read the flowmeter, locate the prescribed  
flowrate line on the flowmeter. Next, turn the flow knob until the  
ball rises to the line. Now, center the ball on the L/min. line  
prescribed.  
2. If the flowrate on the flowmeter ever falls below 0.5 L/min.  
for more than about one minute, the LOW FLOW alarm  
will be triggered. This is a rapid beeping of the audible  
alarm. Check your tubing or accessories for blocked or  
kinked tubing or a defective humidifier bottle. After rated  
flow is restored, more than 0.5 L/min., the LOW FLOW  
audible alarm will go off.  
Platinum™  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
SECTION 5—OPERATING INSTRUCTIONS  
NOTE: The use of some accessories such as the pediatric flowstand  
and the HomeFill compressor will deactivate the Low Flow Alarm.  
Flow Knob  
Ball  
FIGURE 5.6 Flowrate  
SensO2 Oxygen Purity Indicators (LX O2  
Sensor Model)  
This feature monitors the purity of oxygen generated by the  
oxygen concentrator. If purity falls below factory preset  
standards, indicator lights on the control panel will illuminate.  
Initial Startup of the Concentrator  
NOTE: Concentrator may be used during the initial start warmup  
time (approximately 30 min.) while waiting for the O2 purity to  
reach maximum.  
When the unit is turned on, the GREEN light will come on  
(SYSTEM OK/O2 greater than 85%). After five minutes, the  
oxygen sensor will be operating normally and will control the  
indicator lights depending on oxygen concentration values.  
Explanation of Oxygen Purity Indicator Lights  
NOTE: For this procedure, refer to FIGURE 5.7.  
25  
Platinum™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
SECTION 5—OPERATING INSTRUCTIONS  
OXYGEN PURITY  
Auto Shut Down  
Use Backup  
RED  
Call Supplier  
YELLOW  
Below  
Normal  
Normal  
GREEN  
O2  
FIGURE 5.7 SensO2 Oxygen Purity Indicators (LX O2 Sensor  
GREEN light (O2) Normal Operation.  
YELLOW light ( ) Call supplier IMMEDIATELY. You may  
continue to use the concentrator unless instructed otherwise  
by your supplier. Be certain that backup oxygen is nearby.  
RED light ( )Total unit shutdown. Switch IMMEDIATELY  
to backup oxygen supply. Call supplier IMMEDIATELY  
GREEN light with YELLOW light flashing Call supplier  
IMMEDIATELY. Oxygen sensor malfunctioning; you may  
continue to use the concentrator (FIGURE 5.7).  
Indicators (LX Model)  
NOTE: For this procedure, refer to FIGURE 5.8.  
Explanation of LX models Indicator Lights  
RED light ( ) Total Unit ShutDown. Switch  
IMMEDIATELY to a backup oxygen supply. Call supplier  
IMMEDIATELY.  
GREEN light (I/O) On/Off. System okay.  
Platinum™  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
   
SECTION 5—OPERATING INSTRUCTIONS  
Auto Shut Down  
RED  
Call Supplier  
Use Backup  
Normal  
GREEN  
I/O  
Main Power Loss Alarm System  
The alarm system contains no battery and is thereby  
maintenance free. The alarm system is powered by a capacitor  
that is continuously charged and ready to trigger the alarm in  
the event of main power loss.  
Elapsed Time Meter (Hour Meter)  
The hour meter displays the cumulative number of hours the  
unit has operated.  
27  
Platinum™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
SECTION 6—MAINTENANCE  
SECTION 6—MAINTENANCE  
NOTE: The Invacare concentrators are specifically designed to  
minimize routine preventive maintenance. In places with high dust  
or soot levels, maintenance may need to be performed more often.  
Only authorized or factorytrained personnel should perform  
preventive maintenance on the concentrator. Power should be  
disconnected before beginning preventive maintenance on the  
Invacare concentrator.  
NOTE: At a minimum, preventive maintenance MUST be  
performed according to the maintenance record guidelines. Refer to  
Routine Maintenance  
NOTE: The following routine maintenance should be performed  
more frequently. See recommended intervals in each section and  
preventive maintenance record.  
Cleaning the Cabinet Filter  
NOTE: For this procedure, refer to FIGURE 6.1 on page 29.  
CAUTION  
DO NOT operate the concentrator without the filters  
installed.  
NOTE: There are two cabinet filters. One filter located on each side  
of the cabinet.  
1. Remove each filter and clean at least once a week  
depending on environmental conditions.  
NOTE: Environmental conditions that may require more frequent  
cleaning of the filters include but are not limited to: high dust,  
smoking, air pollutants, etc.  
2. Clean the cabinet filters with a vacuum cleaner or wash in  
warm soapy water and rinse thoroughly.  
Platinum™  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
SECTION 6—MAINTENANCE  
3. Dry the filters thoroughly before reinstallation.  
Cabinet Filter  
(located on both  
sides of  
concentrator)  
Cabinet  
WARNING  
Unplug the concentrator when cleaning. To avoid  
electrical shock, DO NOT remove cabinet.  
1. Clean the cabinet with a mild household cleaner and  
nonabrasive cloth or sponge.  
Humidifier  
NOTE: To clean the oxygen humidifier, follow the instructions  
provided by the manufacturer. If none are provided, follow these  
steps:  
1. Clean the humidifier every day.  
2. Wash it in soapy water and rinse it with a solution of ten  
parts water and one part vinegar.  
3. Rinse thoroughly with hot water and refill with distilled  
water to the level shown on the humidifier.  
WARNING  
DO NOT overfill humidifier bottle.  
29  
Platinum™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
SECTION 6—MAINTENANCE  
MODEL NO. IRC  
SERIAL NO.  
FIGURE 6.2 Preventive Maintenance Record  
Platinum™  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
SECTION 7—TROUBLESHOOTING GUIDE  
SECTION 7—TROUBLESHOOTING  
GUIDE  
SYMPTOM:  
PROBABLE CAUSE:  
SOLUTION:  
Alarm:  
Main Power Loss:  
Short beeps, long  
pause  
1. Power cord NOT  
plugged in.  
1. Insert plug into outlet.  
Concentrator not  
operating,  
power switch On.  
2. No power at outlet.  
2. Inspect house circuit  
breakers or fuses. If  
problem recurs, use a  
different outlet.  
Beep......Beep......  
3. Tripped circuit  
breaker.  
3. Call Service Provider.  
Alarm:  
System Failure:  
1a. Remove and clean  
Continuous  
1. Unit overheating due cabinet filters.  
to blocked air intake.  
Concentrator not  
operating, Power  
Switch On.  
1b. Move oxygen  
concentrator at least 7.5 cm  
(3 in) away from walls,  
draperies or furniture.  
Beeeeeeeep....  
2. Insufficient power at  
outlet.  
2. DO NOT use extension  
cords. Move to another  
electrical outlet or circuit.  
3. Internal repairs  
required  
3. Call Service Provider.  
31  
Platinum™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 
SECTION 7—TROUBLESHOOTING GUIDE  
SYMPTOM:  
PROBABLE CAUSE:  
SOLUTION:  
YELLOW or RED  
light illuminated.  
1. Low oxygen purity.*  
1. Clean or Replace filters.  
2. Kinked or blocked  
tubing, cannula or  
humidifier.*  
2. Inspect for kinks or  
blockages. Correct, clean or  
replace item. Once  
Alarm:  
Continuous  
On red light only.  
corrected, turn power OFF  
for 60 seconds and then  
turn power back ON.  
3. Flowmeter set at 0.5  
L/min. or less.*  
3. Check flowmeter is set to  
to Recommended  
Guidelines for Optimum  
* Only applies to  
SensO2 models.  
4. Unit overheating due 4.a. Remove and clean  
to blocked air intake.  
cabinet filters.  
4b. Move concentrator at  
least 7.5 cm (3 in) from  
walls, draperies, and  
furniture.  
5. Insufficient power at  
outlet.  
5. DO NOT use extension  
cords. Move to another  
electrical outlet or circuit.  
6. Internal repairs  
required.  
6. Call Service Provider.  
1. Call Service Provider.  
GREEN light  
YELLOW Light  
Flashing.  
1. Internal repairs  
required.  
Alarm:  
RAPID  
Low Flow Alarm  
1a. Kinked or blocked  
tubing, cannula or  
humidifier.  
1a. Inspect for kinks or  
blockages. Correct, clean or  
replace item. Once  
corrected, turn power OFF  
for 60 seconds and then  
turn power back ON.  
Beep....Beep...  
Beep...Beep  
1b. Flowmeter set at 0.5 1b. Reset flowmeter to  
L/min. or less.  
prescribed flowrate.  
Flowrates less than1L/min.  
are not recommended.  
Platinum™  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
LIMITED WARRANTY  
LIMITED WARRANTY  
Available At:  
Belgium & Luxemburg:  
Invacare nv, Autobaan 22, B8210 Loppem  
Tel: (32) (0)50 83 10 10, Fax: (32) (0)50 83 10 11  
Danmark:  
Deutschland:  
info@invacareaquatec.com  
European Distributor Organisation:  
Invacare, Kleiststraße 49, D32457 Porta Westfalica  
Tel: (49) (0)57 31 754 540, Fax: (49) (0)57 31 754 541  
España:  
Invacare SA, c/Areny s/n, Polígon Industrial de Celrà, E17460  
Tel: (34) (0)972 49 32 00, Fax: (34) (0)972 49 32 20  
France:  
Ireland:  
Invacare Ireland Ltd, Unit 5 Seatown Business Campus,  
Seatown Road, Swords, County Dublin Ireland  
Tel: (353) 1 810 7084, fax: (353) 1 810 7085  
33  
Platinum™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
LIMITED WARRANTY  
Italia:  
Nederland:  
Invacare BV, Celsiusstraat 46, NL6716 BZ Ede  
Norge:  
Invacare AS, Grensesvingen 9, Postboks 6230, Etterstad,  
Tel: (47) (0)22 57 95 00, Fax: (47) (0)22 57 95 01  
Österreich:  
Tel.: (43) 6232 5535 0, Fax.: (43) 6232 5535 4  
Portugal:  
Invacare Lda, Rua Estrada Velha, 949, P4465784 Leca do  
Tel: (351) (0)225 1059 46/47, Fax: (351) (0)225 1057 39  
Sverige & Suomi:  
Platinum™  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
LIMITED WARRANTY  
Switzerland:  
office@mobitecrehab.ch  
United Kingdom:  
Invacare Ltd, South Road, Bridgend Industrial Estate,  
Tel: (44) (0)1656 664 321, Fax: (44) (0)1656 667 532  
35  
Platinum™  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Invacare Corporation  
USA  
Manufacturer:  
One Invacare Way  
Elyria, Ohio USA  
44036-2125  
Invacare Medical  
Equipment (Suzhou) Co.,  
Ltd.  
440-329-6000  
800-333-6900  
Technical Services  
440-329-6593  
800-832-4707  
Unit 2B, Export  
Processing Zone  
No. 200 Suhong Road SIP  
(Suzhou Industrial Park)  
Suzhou, Jiangsu, PRC  
215021  
Tel: 86-512-62586192  
Fax: 86-512-62586072  
REP  
EC  
All rights reserved. Trademarks are  
identified by ™ and ®.  
All trademarks are owned by or  
licensed to Invacare Corporation or  
its subsidiaries unless otherwise  
noted.  
EU Representative  
Invacare International Sarl  
Route de Cité Ouest 2  
1196 Gland  
Switzerland  
Tel: +41 22 354 60 10  
Fax: +41 22 354 60 11  
© 2009 Invacare Corporation  
Rev E - 04/09  
Part No 1134866  
Download from Www.Somanuals.com. All Manuals Search And Download.  
 

Indesit Washer WIL113UK User Manual
Jensen CD Player JMC 326 User Manual
JL Audio Home Theater System IWS SYS 1 User Manual
JVC Cassette Player KS F315EE User Manual
JVC Stereo Receiver RX 730RBK User Manual
Kalorik Cookware NM 39527 User Manual
Kenwood Two Way Radio KSGCP401 User Manual
KiSS Networked Entertainment DVR KiSS VR 558 User Manual
KitchenAid Blender KHBC208 User Manual
Kompernass Microwave Oven KH 1166 User Manual