Graham Field Personal Lift JB02008 User Manual

CONTENTS  
INTRODUCTION  
Caution: Indicates a potential hazard or unsafe  
practice that, if not avoided, could result in minor  
personal injury or product/property damage.  
Note: Provides application recommendations or other  
useful information to ensure that you get the most from  
your product.  
EXPLOSION HAZARD: Do not use the oximeter in  
the presence of flammable anesthetics, explosive  
substances, vapors or liquids.  
Dear User: Please read and understand all instructions  
before using the John Bunn DigiOx Finger Pulse  
Oximeter, JB02008.  
INTRODUCTION............................................................ 3  
INTENDED USE ....................................................... 3  
CONTRAINDICATIONS............................................ 3  
IMPORTANT SAFETY GUIDELINES - PLEASE  
READ BEFORE USE............................................ 3  
The oximeter is designed for real-time and  
rapid measurement. It is not suitable for long-time  
continuous patient monitoring. DO NOT apply the  
sensor to the same finger for more than two hours.  
If any readings are found to be abnormal, please  
change the position of the sensor.  
DigiOx  
Intended use  
Finger Pulse Oximeter  
The John Bunn DigiOx Finger Pulse Oximeter is used  
to measure oxygen saturation (SpO ) and pulse rate in  
home or clinical settings. It is design2 ed for adult and  
pediatric patients.  
OVERVIEW .................................................................... 7  
APPEARANCE .............................................................. 7  
FEATURES .................................................................... 7  
NAME AND MODEL ...................................................... 8  
BATTERY INSTALLATION ............................................ 8  
OPERATION INSTRUCTIONS ...................................... 8  
ACCESSORIES (INCLUDED) ..................................... 11  
REPAIR, MAINTENANCE, AND DISPOSAL............... 11  
MAINTENANCE...................................................... 11  
CLEANING AND DISINFECTION .......................... 12  
DISPOSAL.............................................................. 13  
TROUBLESHOOTING................................................. 13  
SYMBOL KEY.............................................................. 15  
TECHNICAL SPECIFICATIONS.................................. 16  
LIMITED WARRANTY ................................................. 18  
MANUAL DE OPERACIÓN (ESPAÑOL)..................... 21  
WARNINGS:  
Important! Read and understand this manual before  
using the oximeter. Do not use this device without  
proper instruction from a Healthcare Professional.  
If the oximeter is not properly operated, personal  
injury and/or damage to the oximeter could result.  
Contraindications  
The oximeter is intended only as an adjunct in  
patient assessment. It must be used in conjunction  
with clinical signs and symptoms. It is intended as a  
measuring, not a treatment, device.  
WARNING: The oximeter is not intended for  
continuous patient monitoring. DO NOT use the  
oximeter while patient is undergoing MRI or CT  
scanning.  
If components are damaged or missing, contact your  
dealer immediately. DO NOT use substitute parts.  
Magnetic and electrical fields are capable of interfer-  
ing with the proper performance of the oximeter. For  
this reason, ensure that all external devices operated  
in the vicinity of the oximeter comply with the relevant  
EMC requirements. Mobile phones, X-ray equipment,  
and/or MRI devices are a possible source of interfer-  
ence, as they may emit high levels of electromagnetic  
radiation. This device complies with IEC60601-1-2.  
Important Safety Guidelines - Please Read Before Use  
The safety statements presented in this chapter refer  
to the basic safety information that the operator of the  
John Bunn DigiOx Finger Pulse Oximeter shall pay  
attention to and abide by. There are additional safety  
statements in other chapters or sections, which may  
be the same as or similar to the following, or specific  
to the operations.  
Inspect the device before use to ensure that there is  
no visible damage that may affect patient or user’s  
safety or measurement performance. Thereafter,  
inspect the device at least once a week.  
Operation Manual  
Important: Do not operate the  
John Bunn DigiOx Finger Pulse  
Oximeter without first reading  
and understanding this manual!  
Save this manual for future use.  
GF Health Products, Inc. assumes no responsibility  
for any damage or injury caused by improper instal-  
lation or use of this product.  
DO NOT clip this device on edema or tender tissue.  
DO NOT open the oximeter housing. It contains  
no serviceable parts. Service must be performed  
by trained, authorized personnel only. Otherwise,  
device failure and health hazard may occur.  
WARNING: Indicates a potential hazard situation  
or unsafe practice that, if not avoided, could result in  
death or serious injury.  
Note: The most current version of this manual  
can be found online at www.grahamfield.com  
The infrared light emitted from the device (it is invis-  
ible) is harmful to the eyes. DO NOT stare into the light.  
Keep this device out of the reach of children.  
JB02008-INS-LAB-RevA09  
2
3
4
5
Cautions  
OVERVIEW  
from any angle. The oximeter features a low voltage  
indicator, powers off automatically in eight seconds  
when not in use, and provides approximately 30 hours  
of use with the use of alkaline batteries.  
Avoid placing the device on a limb which is wrapped  
with a cuff for blood pressure measurement or undergo-  
ing venous infusion.  
Ensure that no obstruction (contamination, scar tissue,  
fingernail polish, acrylic fingernail, etc.) exists at the  
site where the sensor is placed; such obstruction could  
affect the signal received by the sensor and result in an  
incorrect measurement.  
Vigorous exercise and electrosurgical device interfer-  
ence may affect the measuring accuracy.  
1. Open the clip as shown in  
Figure 4.  
a. Press the Power/Display Button  
once more; the screen will  
rotate 180° and display as  
To ensure patient safety, use only parts and ac-  
cessories specified in this manual.  
Keep the oximeter away from dust, vibration,  
corrosive substances, explosive materials, high  
temperature and moisture.  
If the oximeter gets wet, stop operation im-  
mediately and do not resume operation until it is  
dry. When it is carried from a cold environment  
to a warm and/or humid environment, allow it to  
warm to room temperature before using it.  
DO NOT use sharp objects to press the front  
panel button.  
DO NOT sterilize this device. Do not subject this  
device to irradiation, steam, ethylene oxide, high  
temperature, or immersion in liquid. Refer to the  
Repair and Maintenance/Cleaning and Disinfec-  
tion section of this manual for instructions for  
cleaning and disinfection.  
Appearance  
display screen  
power/display  
battery cover  
shown in Figure 6.  
Figure 6  
b. Press the Power/Display  
Button once more; the screen  
will rotate 90° and display as  
shown in Figure 7.  
button  
Name and Model  
Name: John Bunn DigiOx Finger Pulse Oximeter  
Model: JB02008  
BATTERY INSTALLATION  
1. Refer to Figure 3; insert  
two AAA batteries, oriented  
as shown, into the battery  
compartment.  
Figure 7  
c. Press the Power/Display Button  
rubber  
cushion  
once more; the screen will  
rotate 180° and display as  
shown in Figure 8.  
Figure 1  
Figure 2  
d. Press the Power/Display Button  
once more; the screen will  
2. Insert finger, nail up, into  
Figure 8  
2. Replace the cover.  
The DigiOx Finger Pulse Oximeter is shown above.  
the clip's rubber cushion  
(ensure that the finger is  
Figure 4  
Note: Ensure that the batteries  
Features  
operation  
rotate -90° and also show pulse  
as a filled wave-form bar graph  
along bottom of screen, as  
Figure 3  
are correctly installed. Incorrect  
installation will prevent the  
device from operating.  
The John Bunn DigiOx Finger Pulse Oximeter  
provides a simple way to spot-check patients by  
combining the sensor and monitor into one integrated,  
compact, easy-to-use device with single-button  
operation. The oximeter can accurately measure  
oxygen saturation and pulse rate. When a finger is  
inserted into the sensor’s rubber cushion, and the  
Power/Display Button is pressed to turn the oximeter  
on, then the SpO2 value and pulse rate value display  
on the screen. The oximeter features a bright-color  
LED screen with adjustable brightness and multi-  
directional displays that enable the user to read results  
battery installation  
correctly positioned), and then gently clip the finger.  
3. Press the Power/Display Button to turn the oxim-  
eter on.  
4. The screen will then display the  
data as shown in Figure 5. After  
the values become stable, the  
shown in Figure 9.  
Figure 9  
OPERATION INSTRUCTIONS  
e. Press the Power/Display Button  
once more; the screen will now  
show pulse as a hollow wave-  
form bar graph along bottom of  
Notes:  
Clean the test finger with isopropyl alcohol before and  
after each measurement. Ensure that the finger is dry  
before being inserted into the oximeter.  
Do not shake the finger; keep finger motionless during  
oximeter use.  
user can read the information  
on the display screen.  
Figure 5  
Figure 10  
screen, as shown in Figure 10.  
f. Press the Power/Display Button once more; the  
screen will return to the original screen as shown  
in Figure 5 on previous page.  
5. When the screen displays as shown in Figure 5,  
pressing the Power/Display Button will change the  
screen's orientation:  
6
7
8
9
10  
6. To change the oximeter’s display screen bright-  
ness, press the Power/Display Button for more  
than one second. There are 10 brightness levels,  
and the level, from “BR 1” to “BR 10”, will show  
on the screen during brightness adjustment. The  
default is level four. Keep the Power/Display Button  
depressed for as long as it takes to cycle to the  
desired brightness level.  
7. The oximeter will automatically turn off if no finger  
is sensed for 8 seconds. If the finger is removed  
and the Power/Display Button is pressed, the  
oximeter will remain on and change screens, but  
will display “FINGER OUT”.  
Disposal  
SYMBOL KEY  
Change the batteries when the low-voltage indicator displays.  
Troubleshooting continued  
Do not treat this device or its batteries as household  
waste. Separate collection is required.  
Clean the surface of the device before using: Wipe the device  
with a clean, soft cloth saturated with a solution of 70% isopropyl  
alcohol, then let it air dry or wipe it dry. Repeat as necessary.  
Symbol  
Description  
Problem  
Possible Reason  
Solution  
Type BF  
TROUBLESHOOTING  
The device  
does not  
turn on  
Batteries are  
low  
Replace batteries  
Take out the batteries if the oximeter will not be used for a  
long period of time.  
Problem  
Possible Reason  
Solution  
Warning: See operation manual  
The pulse oxygen saturation  
Pulse rate (beats per minute)  
Batteries  
incorrectly  
installed  
Reinstall batteries  
correctly  
SpO2 and  
Pulse Rate  
display  
unstable or  
hard to see  
Finger  
Reposition finger  
correctly and try again  
incorrectly  
positioned in  
device  
Environment (Note: Operation: 41°F - 104°F  
Humidity may shorten (5°C - 40°C), 15% - 80% RH  
the life of the device,  
The device has  
malfunctioned  
Contact your GF Health  
Products, Inc. distributor  
Storage: -4°F - 158°F  
(-20°C - 70°C), < 93% RH  
or even damage it)  
Finger is  
shaking or  
patient is  
moving  
Ensure patient remains  
immobile  
The device  
turns off  
suddenly  
The device  
powers off  
automatically  
8 seconds  
after finger is  
removed  
Normal  
Low battery voltage  
ACCESSORIES (INCLUDED)  
Cleaning and Disinfection  
Serial number  
1
2
3
One lanyard  
Two AAA alkaline batteries  
One operation manual  
1. Do not immerse the device in liquid. Clean the de-  
vice only by using a surface-clean technique: Wipe  
with a clean, soft cloth saturated with a solution of  
70% isopropyl alcohol.  
SN  
Patient’s SpO  
value is too lo2w  
to measure  
Remeasure; if  
Not for continuous monitoring  
unsuccessful, yet device  
is functioning correctly,  
contact a medical  
Batteries are  
almost drained  
Replace batteries  
REPAIR, MAINTENANCE, AND DISPOSAL  
professional immediately  
2. Wipe with a clean, soft cloth saturated with clean  
water and air dry, or dry with a clean, soft cloth.  
Caution: Do not sterilize this device. Do  
not subject this device to irradiation, steam,  
ethylene oxide, high temperature, or immersion  
in liquid.  
WARNING: DO NOT open the oximeter housing.  
It contains no serviceable parts. Service must be  
performed by trained, authorized personnel only. Oth-  
erwise, device failure and health hazard may occur.  
Excessive  
Reduce ambient light  
Screen  
Device  
malfunction  
Contact your GF Health  
Products, Inc. distributor  
ambient light  
displays  
“Error3”,  
“Error4”,  
“Error6” or  
“Error7”  
Maintenance  
In order to ensure the oximeter's longevity and  
continued dependable operation, please observe the  
following maintenance recommendations.  
Caution: Do not use the sensor if it is dam-  
aged.  
11  
12  
13  
14  
15  
TECHNICAL SPECIFICATIONS  
LIMITED WARRANTY  
TECHNICAL SPECIFICATIONS continued  
GF Health Products, Inc. warrants the John Bunn  
DigiOx Finger Pulse Oximeter, Model JB02008, for a  
period of one year against manufacturer's defects. If  
a product is deemed to be under warranty, GF Health  
Products, Inc. shall provide, at its option, (1) replace-  
ment of any defective part or product or (2) a credit of  
the original selling price made to GF Health Products,  
Inc.'s initial customer. The warranty does not include  
any labor charges incurred in replacement part(s) in-  
stallation or any associated freight or shipping charges  
to GF Health Products, Inc.  
Display screen  
Organic LED Display  
Classification  
The type of protection against electric  
shock: Internally powered equipment  
Power supply  
requirement  
Two 1.5V AAA alkaline batteries  
The degree of protection against  
electric shock: Type BF applied part  
SpO2 Parameter  
Measuring range: 0%-100%  
Specifications  
The degree of protection against  
harmful ingress of liquids: Ordinary  
equipment without protection against  
ingress of water  
Accuracy: 70%-100%, ±3% < 70%,  
no definition  
USA Corporate Headquarters:  
Pulse Rate  
Parameter  
Specifications  
Measuring range: 30-235 bpm  
GF Health Products, Inc.  
2935 Northeast Parkway  
Atlanta, Georgia 30360  
Accuracy: 30-99bpm, ±2 bpm; 100-  
235 bpm, ±2%  
Electro-Magnetic Compatibility: Group  
I, Class B  
Dimensions (L x  
W x H)  
2.28" x 1.26" x 1.34" (58 mm x 32 mm  
x 34 mm)  
© 2009 GF Health Products, Inc.  
Made in China  
telephone: 800-347-5678, 770-447-1609  
fax: 800-726-0601, 678-291-3232  
Environment  
Operation: 41°F - 104°F (5°C - 40°C),  
15% - 80% RH  
Storage: -4°F - 158°F (-20°C - 70°C),  
< 93% RH  
Graham-Field and John Bunn are registered  
trademarks of GF Health Products, Inc.  
Net weight with  
batteries  
2 oz. (50g)  
Packaging, warranties, products, and specifications  
are subject to change without notice.  
GF Health Products, Inc. is not responsible for  
typographical errors.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
16  
17  
18  
19  

Graco Inc Pressure Washer 802 776 User Manual
Grizzly Sander H8120 User Manual
Haier Washer HK1007TME User Manual
Hamilton Beach Slow Cooker 840105900 User Manual
HannsG Computer Monitor HH 181APB User Manual
Harman Kardon Stereo Amplifier PA5800 User Manual
HP Hewlett Packard All in One Printer 3392 All in One User Manual
HP Hewlett Packard Server T1428 90026 User Manual
Iiyama Car Video System L323W User Manual
iLive Speaker System ITP100B User Manual