					Multimeter with Temperature   
					Instruction Sheet   
					P 
					Read First: Safety Information   
					To ensure that the meter is used safely, follow these instructions:   
					• 
					• 
					Do not use the meter if the meter or test leads appear damaged, or if   
					you suspect that the meter is not operating properly.   
					Disconnect the live test lead before disconnecting the common test   
					lead.   
					• 
					• 
					When using the probes, keep your fingers behind the finger guards on   
					the probes.   
					Do not use the V•Check mode to measure voltages in circuits that   
					could be damaged by this mode’s low input impedance   
					( 2 kΩ).   
					• 
					• 
					• 
					Turn off power to the circuit under test before cutting, desoldering, or   
					breaking the circuit. Small amounts of current can be dangerous.   
					Do not apply more than 600V rms between a meter terminal and earth   
					ground.   
					Use caution when working with voltages above 60V dc or 30V ac rms.   
					Such voltages pose a shock hazard.   
					PN 643657 June 1997 Rev.2, 3/98   
					1997, 1998 Fluke Corporation. All rights reserved. Printed in U.S.A.   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				AC and DC Voltage   
					Also refer to V•Check.   
					Volts AC   
					Input Impedance 5 MΩ   
					50 Hz to 400 Hz   
					Volts DC   
					Input Impedance 10 MΩ   
					MULTIMETER   
					MULTIMETER   
					16   
					16   
					AC   
					DC   
					4000 mV 4V 40V 400V 600V   
					OFF   
					OFF   
					V 
					AC DC   
					V 
					AC   
					/ 
					DC   
					/ 
					CAT   
					CAT   
					600V   
					600V   
					COM   
					+ 
					COM   
					+ 
					ip01i.eps   
					Resistance and Continuity eR   
					Turn off circuit power before testing. Also refer to V•Check.   
					Resistance   
					Continuity   
					16   
					MULTIMETER   
					MULTIMETER   
					MULTIMETER   
					16   
					16   
					Mk   
					OFF   
					OFF   
					OFF   
					V 
					•Check   
					V 
					•Check   
					V•Check   
					A 
					Low Impedance   
					V•Check   
					A 
					A 
					CAT   
					CAT   
					CAT   
					TEMPERATURE   
					TEMPERATURE   
					TEMPERATURE   
					600V   
					600V   
					600V   
					COM   
					+ 
					COM   
					+ 
					COM   
					+ 
					400Ω 4 kΩ 40 kΩ 400 kΩ 4 mΩ 40 mΩ   
					Short   
					Open   
					<25Ω   
					ip02i.eps   
					Detects shorts and opens ≥250 µS.   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				W V•Check   
					If a dc or ac voltage greater than about 4.5V is present across the inputs   
					when the meter is set to R, G, or e, the meter switches automatically   
					to dc or ac voltage mode (V•Check mode).   
					W Warning   
					Repetitive transients on a dc bus will cause V•Check to   
					select ac volts, even though a hazardous dc voltage may   
					be present. To avoid a misleading display and possible   
					electric shock, manually select the proper volts function   
					for measurements on these circuits.   
					When V•Check is activated, the meter has low input impedance (LoZ)   
					2 kΩ. This load can alter the voltages in electronic control circuits. Do   
					not use V•Check to measure voltage in circuits that could be damaged by   
					a 2 kΩ load.   
					Hint: V•Check can be effectively used to eliminate “ghost” voltages.   
					Disable and re-enable V•Check   
					For and Ω,   
					If AC detected   
					If DC detected   
					MULTIMETER   
					16   
					, 
					Disables V•Check and   
					locks meter in selected   
					mode.   
					Meter puts 2 kΩ   
					load on circuit.   
					2 sec   
					OFF   
					V 
					•Check   
					A 
					CAT   
					TEMPERATURE   
					600V   
					or   
					or   
					COM   
					+ 
					Re-enables   
					V•Check   
					or   
					ip09i.eps   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Temperature   
					MULTIMETER   
					16   
					°F   
					°C   
					OFF   
					V•Check   
					A 
					CAT   
					TEMPERATURE   
					600V   
					COM   
					+ 
					Vent   
					or   
					Pipe   
					80PK-1 Type K   
					Thermocouple   
					Probe   
					ip19f.eps   
					Note correct connector polarity.   
					To meet stated accuracy, the 80AK temperature adapter must be at the   
					same temperature as the meter. In addition, the 80AK and the 80PK-1   
					should be cleaned and lubricated regularly. Use any petroleum based   
					lubricant.   
					Warning   
					To avoid possible electric shock, DO NOT apply   
					thermocouple tip to any conductor that is greater than   
					30V AC, 42.4V pk, or 60V DC to earth.   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Capacitance E   
					Turn off circuit power; then disconnect and discharge the capacitor before   
					measuring capacitance.   
					MULTIMETER   
					16   
					If the capacitor requires more   
					discharging, diSC is displayed   
					while the capacitor discharges.   
					OFF   
					V 
					•Check   
					A 
					CAT   
					TEMPERATURE   
					600V   
					COM   
					+ 
					1 µF 10 µF   
					100 µF   
					10,000 µF   
					+ 
					ip05i.eps   
					Note correct probe polarity for polarized capacitors.   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Microamps µA   
					Range 0 to 200 µA   
					MULTIMETER   
					16   
					µA AC   
					µA DC   
					OFF   
					V 
					•Check   
					A 
					CAT   
					TEMPERATURE   
					600V   
					COM   
					+ 
					Flame   
					sensor   
					probe   
					Control   
					module   
					ip20i.eps   
					Example: Flame rectification circuit.   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Diode Test G   
					Turn off circuit power before testing. For best results diodes   
					should be measured out of circuit.   
					Also refer to V•Check.   
					Good Diode   
					MULTIMETER   
					Typical voltage   
					for silicon diode   
					MULTIMETER   
					16   
					16   
					Low Impedance   
					V•Check   
					OFF   
					OFF   
					V 
					•Check   
					V•Check   
					A 
					A 
					CAT   
					CAT   
					TEMPERATURE   
					TEMPERATURE   
					600V   
					600V   
					COM   
					+ 
					COM   
					+ 
					Single beep   
					ip03i.eps   
					Forward Bias   
					Reverse Bias   
					Bad Diode   
					MULTIMETER   
					16   
					MULTIMETER   
					16   
					OFF   
					OFF   
					V•Check   
					V•Check   
					A 
					A 
					CAT   
					TEMPERATURE   
					CAT   
					TEMPERATURE   
					600V   
					600V   
					COM   
					+ 
					COM   
					+ 
					or   
					ip04i.eps   
					Shorted   
					Open   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				µF   
					MIN MAX   
					(Records the lowest and highest measurements)   
					V•Check, autoranging, and Battery Saver are disabled. Put the meter   
					in the proper range before entering MIN MAX.   
					When the reading changes more than about 50 digits, the meter gives a   
					short beep. When a new minimum or maximum is recorded, the meter   
					gives a long beep.   
					Enter   
					MIN MAX   
					Present Reading with   
					Input Change AlertTM   
					Maximum Reading   
					2 sec   
					To Exit   
					Minimum Reading   
					ip10i.eps   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				µF   
					MIN MAX with Elapsed Time   
					Records the hours and minutes between when MIN MAX was entered   
					and the last high and low was recorded. OL is displayed for times longer   
					than 99:59.   
					To enable the MIN MAX timer, hold down M while turning the rotary   
					switch from OFF to either measurement mode.   
					Enter MIN MAX. View Maximum Reading   
					Present Reading   
					Maximum Reading   
					Elapsed Time   
					2 sec   
					Minimum Reading   
					Elapsed Time   
					Minimum Reading   
					Exit MIN MAX   
					ip11i.eps   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Disabling the Beeper   
					To disable the beeper for all modes, hold down r for 2 seconds while   
					turning the meter on.   
					Continuity Capture   
					To set up the meter to capture intermittent shorts and opens, turn the   
					switch to GR, connect the leads to the circuit; then press M.   
					ip12i.eps   
					Captures transitions longer than 250 µs (1/4000th of a second).   
					Transitions after the first transition cause the meter to beep, but the   
					display does not change.   
					To reset the display to the current condition, press M.   
					To exit, press M for 2 seconds, or change the measurement mode.   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Maintenance   
					Clean the case with a damp cloth and detergent. Do not use abrasives or   
					solvents.   
					Battery Replacement   
					Remove the test leads before disassembling the case.   
					COM   
					600V   
					+ 
					+ 
					ip13i.eps   
					Replacement Parts   
					Fluke TL-75 (Double-insulated leads)   
					80AK-8001 (Adapter, Thermocouple)   
					80PK-1 (Probe, Thermocouple, Beaded K-Type)   
					PN 855705   
					PN 919683   
					PN 750422   
					Service and Parts   
					This meter should be serviced only by a qualified service technician. For   
					service information in the USA and Canada call 1-800-526-4731. To   
					locate an authorized service center, visit us on the World Wide Web:   
					
					below.   
					USA and Canada: 1-800-44-FLUKE (1-800-443-5853)   
					Europe: +31 402-678-200   
					Japan: +81-3-3434-0181   
					Singapore: +65- -276-6196   
					* 
					Anywhere in the world: +1-425-356-5500   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Specifications   
					Accuracy is specified for a period of one year after calibration, at 18°C to   
					28°C (64°F to 82°F) with relative humidity to 90%. AC conversions are   
					ac-coupled, average responding, and calibrated to the rms value of a sine   
					wave input. Accuracy specifications are given as follows:   
					±([% of reading] + [number of least significant digits])   
					Function   
					Range   
					Resolution   
					Accuracy   
					-10°C to 400°C   
					14°F to 752°F   
					0.1°C or   
					0.2°F   
					±(1.0% + 0.8°C)   
					typical   
					Temperature   
					±(1.0% + 1.5°F) typical   
					±(5.0% + 1.5°C)   
					typical   
					(Type K Thermocouple) -40°C to -10°C   
					-40°F to 14°F   
					0.1°C or   
					0.2°F   
					±(5.0% + 3.3°F) typical   
					Error does not include Type K Thermocouple errors.   
					Function   
					Range   
					Resolution   
					Accuracy   
					4000 mV1   
					4.000V   
					1 mV   
					±(1.9% + 3)   
					±(1.9% + 3)   
					±(1.9% + 3)   
					±(1.9% + 3)   
					±(1.9% + 3)   
					0.001V   
					00.01V   
					000.1V   
					1V   
					40.00V   
					400.0V   
					(50 to 400 Hz)   
					600V   
					4000 mV1   
					4.000V   
					1 mV   
					±(0.9% + 2)   
					±(0.9% + 2)   
					±(0.9% + 1)   
					±(0.9% + 1)   
					0.001V   
					00.01V   
					000.1V   
					1V   
					40.00V   
					400.0V   
					600V   
					±(0.9% + 1)   
					400.0Ω   
					0.1Ω   
					±(0.9% + 2)   
					4.000 kΩ   
					40.00 kΩ   
					400.0 kΩ   
					4.000 MΩ   
					40.00 MΩ   
					1.000 µF   
					10.00 µF   
					100.0 µF   
					10000 µF   
					0.001 kΩ   
					0.01 kΩ   
					0.1 kΩ   
					0.001 MΩ   
					0.01 MΩ   
					0.001 µF   
					0.01 µF   
					0.1 µF   
					1 µF   
					±(0.9% + 1)   
					±(0.9% + 1)   
					±(0.9% + 1)   
					±(0.9% + 1)   
					Ω 
					±(1.5% + 3)   
					±(1.9% + 2)   
					±(1.9% + 2)   
					±(1.9% + 2)   
					≤1000 µF ±(1.9% + 2)   
					>1000 µF ±(10% + 90) typical   
					±(1.9% + 2)2   
					2.000V   
					0.001V   
					1.The 4000 mV range can be entered only in manual range mode. Use the   
					4000 mV range with accessories.   
					2.The beeper is guaranteed to come on at <25Ω and turn off at >250Ω. The   
					meter detects opens or shorts ≥250 µs.   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Function   
					Range Resolution   
					Accuracy   
					Burden Voltage   
					0 to 200 µA   
					0.1 µA   
					±(2% + 3 counts)   
					<5 mV/µA   
					(50 Hz to 400   
					Hz)   
					0 to 200 µA   
					0.1 µA   
					±(1% + 2 counts)   
					<5 mV/µA   
					Overload   
					Function   
					Input Impedance (Nominal)   
					Protection1   
					>5 MΩ <100 pF   
					600V rms   
					V•Check and LoZ = >2 kΩ <200 pF (ac coupled)2   
					>10MΩ <100 pF   
					600V rms   
					V•Check and LoZ = >2 kΩ <200 pF2   
					Common Mode Rejection   
					Normal Mode Rejection   
					Ratio (1 kΩ Unbalanced)   
					600V rms >60 dB at dc 50 or 60 Hz   
					600V rms >100 dB at dc, 50 or 60 Hz >50 dB at 50 Hz or 60 Hz   
					Full Scale Voltage   
					Open Circuit Test Voltage   
					To 4.0 MΩ   
					40 MΩ   
					600V rms   
					600V rms   
					<1.5V dc   
					<450 mV dc <1.5V dc   
					2.400V dc   
					2.4-3.0V dc   
					Short Circuit Current   
					600V rms   
					600V rms   
					<500 µA   
					0.95 mA (typical)   
					1. 3 x 106 V Hz maximum   
					2. 2 kΩ input impedance up to 50V. Impedance increases with input voltage to   
					>300 kΩ at 600V.   
					MIN MAX Recording Accuracy and Response Time   
					Specified accuracy of the measurement function ±12 digits in dc for changes   
					>200 ms in duration (±40 digits in ac). Typical 100 ms response to 80%.   
					Example 1: This would mean ±1.2° when recording temperature.   
					Example 2: This would mean ±12 µA when recording µA or ±12A if used with a   
					dc amp probe (with a mV input).   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				MIN MAX Recording with Elapsed Time   
					Elapsed Time   
					Resolution   
					Accuracy   
					0 to 100 hours (99:59)   
					1 minute   
					0.3% typical   
					Maximum Voltage   
					Between any Terminal   
					and Earth Ground:   
					600V rms   
					Display:   
					3 3/4-digits, 4000 counts, updates 4/sec   
					Operating Temperature: -10°C to 50°C (14°F to 122°F)   
					Storage Temperature:   
					-30°C to 60°C (-22°F to 140°F)   
					indefinitely (to -40°C (-40°F) for 100 hrs)   
					Temperature   
					Coefficient:   
					(.1 x specified accuracy)/°C (<18°C or   
					>28°C)   
					Relative Humidity:   
					0% to 90% (-10°C to 35°C; 14°F to 95°F)   
					0% to 70% (35°C to 50°C; 95°F to 122°F)   
					9V, NEDA 1604 or IEC 6F22   
					650 continuous hours with alkaline   
					450 continuous hours with carbon-zinc   
					3 meter drops.   
					Battery Type:   
					Battery Life:   
					Shock, Vibration:   
					Size (H x W x L):   
					3.46 cm x 7.05 cm x 14.23 cm   
					(1.35 in x 2.75 in x 5.55 in)   
					Weight:   
					Safety:   
					286g (10 oz)   
					Designed to Protection Class II   
					requirement of UL1244, ANSI/ISA-S82,   
					CSA C22.2 No 231, and VDE 0411, and   
					IEC 1010 overvoltage Category III (CAT   
					III, 600 Volts).   
					EMI Regulations:   
					Certifications:   
					Complies with FCC Part 15, Class B, and   
					VDE 0871B. Trademark of TÜV Product   
					Services. Complies with EN 61010-1:   
					1993.   
					TUV, UL and VDE pending   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				LIMITED WARRANTY & LIMITATION OF LIABILITY   
					Each Fluke product is warranted to be free from defects in material and workmanship under   
					normal use and service. The warranty period is three years and begins on the date of shipment.   
					Parts, product repairs and services are warranted for 90 days. This warranty extends only to the   
					original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses,   
					disposable batteries or to any product which, in Fluke’s opinion, has been misused, altered,   
					neglected or damaged by accident or abnormal conditions of operation or handling. Fluke   
					warrants that software will operate substantially in accordance with its functional specifications for   
					90 days and that it has been properly recorded on non-defective media. Fluke does not warrant   
					that software will be error free or operate without interruption.   
					Fluke authorized resellers shall extend this warranty on new and unused products to end-user   
					customers only but have no authority to extend a greater or different warranty on behalf of Fluke.   
					Warranty support is available if product is purchased through a Fluke authorized sales outlet or   
					Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for   
					importation costs of repair/replacement parts when product purchased in one country is submitted   
					for repair in another country.   
					Fluke’s warranty obligation is limited, at Fluke’s option, to refund of the purchase price, free of   
					charge repair, or replacement of a defective product which is returned to a Fluke authorized   
					service center within the warranty period.   
					To obtain warranty service, contact your nearest Fluke authorized service center or send the   
					product, with a description of the difficulty, postage and insurance prepaid (FOB Destination), to   
					the nearest Fluke authorized service center. Fluke assumes no risk for damage in transit.   
					Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB   
					Destination). If Fluke determines that the failure was caused by misuse, alteration, accident or   
					abnormal condition of operation or handling, Fluke will provide an estimate of repair costs and   
					obtain authorization before commencing the work. Following repair, the product will be returned to   
					the Buyer transportation prepaid and the Buyer will be billed for the repair and return   
					transportation charges (FOB Shipping Point).   
					THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL   
					OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY   
					IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.   
					FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR   
					CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER   
					ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE OR   
					ANY OTHER THEORY.   
					Since some countries or states do not allow limitation of the term of an implied warranty, or   
					exclusion or limitation of incidental or consequential damages, the limitations and exclusions of   
					this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or   
					unenforceable by a court of competent jurisdiction, such holding will not affect the validity or   
					enforceability of any other provision.   
					Fluke Corporation   
					P.O. Box 9090   
					Everett WA   
					Fluke Europe B.V.   
					P.O. Box 1186   
					5602 B.D. Eindhoven   
					The Netherlands   
					98206-9090   