		 Campbell Scientific CS512   
					Oxyguard Type III   
					Dissolved Oxygen Probe   
					Revision: 12/06   
					C o p y r i g h t   
					© 
					2 0 0 1 - 2 0 0 6   
					C a m p b e l l S c i e n t i f i c , I n c .   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				CS512 Table of Contents   
					PDF viewers note: These page numbers refer to the printed version of this document. Use   
					the Adobe Acrobat® bookmarks tab for links to specific sections.   
					1. General Information.....................................................1   
					2. Specifications ..............................................................2   
					3. Optional Probe Accessories.......................................3   
					4. Optional Agitator .........................................................3   
					5. Application Information ..............................................4   
					6. Wiring............................................................................4   
					7. Programming ...............................................................5   
					8. Calibration....................................................................6   
					9. Maintenance.................................................................7   
					10. Agitator Control .........................................................7   
					Appendix   
					A. Dissolved Oxygen in Fresh Water Table............... A-1   
					Figures   
					1-1. CS512......................................................................................................1   
					4-1. Preventing Bio-fouling of the CS512......................................................3   
					4-2. CS512 with PT4 Agitator........................................................................4   
					10-1. Agitator Wiring .....................................................................................7   
					Table   
					6-1. Sensor Wiring..........................................................................................4   
					i 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				CS512 Table of Contents   
					Examples   
					1. Sample Program using P1.......................................................................... 5   
					2. Sample Program using P2.......................................................................... 5   
					3. Sample Program using VoltSE .................................................................. 5   
					4. Sample Program using VoltDiff ................................................................ 5   
					5. Sample Program using VoltSE .................................................................. 5   
					6. Sample Program using P36........................................................................ 6   
					7. Sample Program using Expression ............................................................ 6   
					8. Sample Program using P86 and P22 Instructions ...................................... 8   
					9. Sample Program using Portset ................................................................... 8   
					10. Sample Program using Portset ................................................................. 8   
					ii   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Campbell Scientific CS512 Oxyguard   
					Type III Dissolved Oxygen Probe   
					An accurate and reliable sensor is a critical element in any measurement system. The   
					Oxyguard Stationary Probe meets these criteria for the measurement of dissolved oxygen.   
					1. General Information   
					The Oxyguard Stationary Probe is a galvanic probe which produces a millivolt   
					signal proportional to the amount of oxygen present in the measured medium.   
					Oxygen diffuses through the membrane onto the cathode, reacts chemically, and   
					combines with the anode. An electrical current is produced by this chemical   
					reaction which is converted from microamps to millivolts by an in-line resistor.   
					An in-line thermistor also conditions the signal providing automatic temperature   
					compensation. With these features, the probe produces a millivolt output   
					proportional to the oxygen present in the medium in which it is placed. The probe   
					consists of two parts, an upper part with cathode, anode and cable, and a lower part   
					comprising a screw on membrane cap with fitted membrane. The cap is filled with   
					electrolyte and simply screwed onto the top component.   
					The probe is self-polarizing and requires no external power source. There are two   
					wires to connect.   
					Because the probe's output is linear, it is possible to connect it directly to a data   
					acquisition system capable of handling the small millivolt signal.   
					The probe's robust construction and simple design make maintenance and   
					servicing it straightforward. There is no need to send the probe back to the factory   
					for servicing. It utilizes a strong, easy-to-clean and easy-to-change membrane in a   
					screw-on membrane cap. Regular servicing is not required. When necessary the   
					probe can be fully overhauled in five minutes.   
					FIGURE 1-1. CS512   
					1 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe   
					2. Specifications   
					Principle of Measurement:   
					Membrane covered galvanic oxygen probe   
					2.5 to 5 mV per mg 1-1   
					Output Signal:   
					Repeatability:   
					Better than ±0.2 mg/l calibration temperature   
					equals measuring temperature ±5°C   
					Output Impedance:   
					Response Time:   
					∼1 kohm   
					After equilibration, 1 minute for 95% of final   
					value   
					Materials of Construction:   
					Probe body:   
					Delrin   
					O-rings:   
					Membrane O-ring = Buna N   
					Body Seal O-ring = Viton   
					0.05 mm (2 mil) Teflon   
					Membrane:   
					Dimensions and Weight:   
					2.28” diameter x 2.3” height   
					15.9 oz.   
					Cable:   
					standard length 7 m (23 ft); other lengths   
					available upon request   
					2-wire 22 awg shielded, PVC jacketed   
					Operating Conditions:   
					Temperature   
					0° to 40°C (32° to 122°F)   
					0 to 29 psig   
					Pressure:   
					Minimum Submersion Depth:   
					Minimum Water Flow:   
					60 mm (2 ½ in)   
					1 cm/s (.39 in/sec) across membrane   
					Calibration:   
					In air or in air saturated water   
					Automatic from 4° - 40°C (40° - 104°F)   
					0-20 mg/l, 0-200% Sat   
					Temperature Compensation:   
					Range of Dissolved Oxygen:   
					Ships with:   
					5 spare membranes with O-rings   
					50 ml electrolyte   
					cathode cleaning pad   
					2 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe   
					3. Optional Probe Accessories   
					PT4-L — Agitator for stagnant conditions   
					18026 — DO electrolyte, 1 liter   
					18027 — Membrane kit, 10 membranes   
					4. Optional Agitator   
					The PT4 Agitator is a reliable and robust agitator for use in conjunction with   
					probes subjected to bio-fouling in ponds and stagnant water conditions.   
					O2 probes require a minimum water velocity across their membranes to function   
					properly. Therefore, to measure DO in stagnant water conditions, it is necessary to   
					move the water past the membrane to get accurate and reliable DO measurements.   
					In many instances the water also has a high bio-loading and the probes become   
					fouled resulting in inaccurate DO measurements.   
					FIGURE 4-1. Preventing Bio-fouling of the CS512   
					The PT4 Agitator overcomes these problems. The device is designed so that a soft   
					bristle brush sweeps across the probe membrane or sensor tip. This sweeping   
					action of the brush provides the required water velocity as well as prevents the   
					membrane from becoming bio-fouled.   
					The optimum sweeping frequency depends upon the design of probe and type of   
					membrane used and water conditions. An ON-time of 0.25 seconds and OFF-time   
					of 5 seconds is suitable in most circumstances.   
					Agitator’s overall size: 7-1/8" x 3-1/4" diameter, 1-1/4 lb. (180mm x 83mm, 0.6 kg)   
					Supplied with 10 ft. (3 meters) cable.   
					Power required: 10.5 to 18 VDC at the agitator, 1.1 amps. Maximum ON-time is   
					3 seconds.   
					3 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe   
					FIGURE 4-2. CS512 with PT4 Agitator   
					Optional Repeat Cycle Timer for Agitators: Reciprocating action may be   
					controlled by the optional solid state Repeat Cycle Timer. It sends 12 VDC pulses   
					to the agitator coil. The ON-time is 0.25 sec. The OFF-time is adjustable from 3   
					to 12 sec.; requires supply voltage 10 to 17 VDC; housed in watertight cylinder   
					6.6" x 2" diameter (170 mm X 50 mm).   
					5. Application Information   
					SCWin users: This manual was written primarily for those   
					whose needs are not met by SCWin. Your procedure is much   
					simpler: just add the CS512 probe (it’s in the water folder), save   
					your program, and follow the wiring shown in Step 2 of SCWIN.   
					NOTE   
					6. Wiring   
					The CS512 probe can use one differential channel or one single-ended channel.   
					Differential wiring is better at rejecting electrical noise and ground loop error.   
					TABLE 6-1. Sensor Wiring   
					CR510, CR10X, CR800, CR850,   
					Colo   
					r 
					Functio   
					n 
					CR23X, CR1000, CR3000   
					CR200 Series   
					Whit   
					e 
					Signal +   
					Differential High,   
					or Single-Ended Channel   
					Single-Ended   
					Channel   
					Blue   
					Signal -   
					Differential Low or AG   
					Ground   
					4 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe   
					7. Programming   
					In the CR510, CR10X, and CR23X datalogger, Instruction P1 or P2 can be used.   
					Example 1 uses a single-ended measurement for the CS512 probe; example 2 uses   
					a differential measurement. The example measurement instructions that follow do   
					not store data to final storage. Additional instructions (typically P92, P77, and   
					output processing instructions such as P70) are required to store data permanently.   
					EXAMPLE 1. Sample Program using P1   
					1: Volt (SE) (P1)   
					1:   
					1 
					Reps   
					2: 24   
					250 mV 60 Hz Rejection Range ; code 23 used for CR23X   
					3:   
					4:   
					5: 1.0   
					6: 0.0   
					1 
					1 
					SE Channel   
					Loc [ DOmV   
					Multiplier   
					Offset   
					] 
					*See Calibration*   
					EXAMPLE 2. Sample Program using P2   
					1: Volt (Diff) (P2)   
					1:   
					1 
					Reps   
					2: 24   
					250 mV 60 Hz Rejection Range ; code 23 used for CR23X   
					3:   
					4:   
					5: 1.0   
					6: 0.0   
					1 
					1 
					DIFF Channel   
					Loc [ DOmV   
					Multiplier   
					Offset   
					] 
					*See Calibration*   
					In the CR1000, CR800, CR850, and CR3000 dataloggers, instruction VoltDiff   
					or VoltSE is used to measure the CS512 probe.   
					EXAMPLE 3. Sample Program using VoltSE   
					VoltSE (DOmV,1,mV250,1,1,0,_60Hz,1.0,0)   
					EXAMPLE 4. Sample Program using VoltDiff   
					VoltDiff (DOmV,1,mV250,1,True ,0,_60Hz,1.0,0)   
					In the CR200 series dataloggers, only the instruction VoltSE is used to   
					measure the CS512 probe.   
					EXAMPLE 5. Sample Program using VoltSE.   
					VoltSE (DOmV,1,1,1.0,0)   
					5 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe   
					8. Calibration   
					The multiplier is used to calibrate the CS512 probe. To calculate the multiplier:   
					1) Program the datalogger using a multiplier of one.   
					2) Place the OxyGuard probe in the air, shaded from the sun. Wait for readings   
					to stabilize. This may take 15 minutes or more.   
					3) Determine the air temperature and barometric pressure.   
					4) Using a calibration chart such as that provided in the probe’s manual,   
					determine the oxygen concentration of the air.   
					5) Use the following equation to calculate the multiplier:   
					M = P/R   
					M = Multiplier   
					P = Concentration in PPM of the air (from the calibration chart in Appendix A)   
					R = The signal output of the OxyGuard probe when using a multiplier of one   
					6) Change the multiplier in the datalogger program from one to the calculated   
					number.   
					A more common way to enter the multiplier is to insert a separate instruction in   
					the program. This will allow a new multiplier to be added to the program   
					without rewriting, compiling, and downloading the program to the data logger.   
					In the CR510, CR10X, and CR23X, use Instruction P36. The multiplier is   
					entered into an input location called DOmult using the numeric display in   
					PC200W, PC208W, LoggerNet, PC400, PConnect, PConnectCE, or the   
					CR10KD.   
					EXAMPLE 6. Sample Program using P36   
					57: Z=X*Y (P36)   
					1:   
					2:   
					3:   
					1 
					2 
					3 
					X Loc [ DOmV   
					Y Loc [ DOmult   
					Z Loc [ DOppm   
					] 
					] 
					] 
					In all dataloggers that use CRBasic, an expression is written. The multiplier   
					value is entered into the expression through the Public Table using the numeric   
					display in PC200W, LoggerNet, PC400, PConnect, and PConnectCE.   
					EXAMPLE 7. Sample Program using Expression   
					DOppm = DOMult * DOmV   
					6 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe   
					9. Maintenance   
					The CS512 probe needs little maintenance. Regular cleaning of the membrane is   
					all that is required. The membrane is very durable and can be cleaned with a cloth   
					or soft paper. Do not scratch it clean with your fingernail.   
					10. Agitator Control   
					In low flow conditions (less than about 5 cm/sec), it may be necessary to add   
					an agitator to the CS512 probe. Campbell Scientific ships the agitator with a   
					repeat cycle timer. Using the repeat cycle timer requires no datalogger   
					programming. However, some users choose to use a solid state relay and have   
					the datalogger agitate the water on the probe face either periodically   
					throughout the day or just before measurement. Agitating just before the   
					measurement saves on power and causes less wear and tear on the agitator and   
					probe membrane.   
					The wiring for the agitator as controlled by this example program would be as   
					follows:   
					BLACK   
					1 
					Ground/G   
					Datalogger   
					4 
					3 
					Agitator   
					#7321 Relay   
					Crydom D1D07   
					Port C1   
					WHITE   
					2 
					+ 
					- 
					12 V Battery   
					FIGURE 10-1. Agitator Wiring   
					The following instructions would trigger the agitator as discussed in the   
					agitator manual.   
					In the CR510, CR10X, and CR23X, use instruction P86 and P22.   
					7 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Campbell Scientific CS512 Oxyguard Type III Dissolved Oxygen Probe   
					EXAMPLE 8. Sample Program using P86 and P22 Instructions   
					45: Do (P86)   
					1: 41   
					Set Port 1 High   
					46: Excitation with Delay (P22)   
					1:   
					1 
					Ex Channel   
					2: 20   
					Delay W/Ex (units = 0.01 sec)   
					Delay After Ex (units = 0.01 sec)   
					mV Excitation   
					3:   
					4:   
					0 
					0 
					47: Do (P86)   
					1: 51   
					Set Port 1 Low   
					48: End (P95)   
					In the CR200 series dataloggers, use the Portset instruction.   
					EXAMPLE 9. Sample Program using Portset   
					Portset (1,1)   
					Delay(500,msec)   
					Portset(1,0)   
					In the CR1000, CR800, CR850, and CR3000 dataloggers, use the Portset   
					instruction.   
					EXAMPLE 10. Sample Program using Portset   
					Portset (1,1)   
					Delay (1,500,msec)   
					Portset (1,0)   
					The above example is not as power efficient as possible and would require AC   
					power to maintain a sufficient battery charge. If it is necessary to operate an   
					agitator without AC power available, write the program so that the agitator is   
					only operated for a short period of time just before the measurement is to be   
					taken.   
					8 
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Appendix A. Dissolved Oxygen in   
					Fresh Water Table   
					A-1   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Appendix A. Dissolved Oxygen in Fresh Water Table   
					A-2   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				This is a blank page.   
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				Campbell Scientific Companies   
					Campbell Scientific, Inc. (CSI)   
					815 West 1800 North   
					Logan, Utah 84321   
					UNITED STATES   
					
					
					Campbell Scientific Africa Pty. Ltd. (CSAf)   
					PO Box 2450   
					Somerset West 7129   
					SOUTH AFRICA   
					
					
					Campbell Scientific Australia Pty. Ltd. (CSA)   
					PO Box 444   
					Thuringowa Central   
					QLD 4812 AUSTRALIA   
					
					
					Campbell Scientific do Brazil Ltda. (CSB)   
					Rua Luisa Crapsi Orsi, 15 Butantã   
					CEP: 005543-000 São Paulo SP BRAZIL   
					
					
					Campbell Scientific Canada Corp. (CSC)   
					11564 - 149th Street NW   
					Edmonton, Alberta T5M 1W7   
					CANADA   
					
					
					Campbell Scientific Ltd. (CSL)   
					Campbell Park   
					80 Hathern Road   
					Shepshed, Loughborough LE12 9GX   
					UNITED KINGDOM   
					
					
					Campbell Scientific Ltd. (France)   
					Miniparc du Verger - Bat. H   
					1, rue de Terre Neuve - Les Ulis   
					91967 COURTABOEUF CEDEX   
					FRANCE   
					
					
					Campbell Scientific Spain, S. L.   
					Psg. Font 14, local 8   
					08013 Barcelona   
					SPAIN   
					
					
					
					Download from Www.Somanuals.com. All Manuals Search And Download.   
				 |