Agilent Technologies G1978B User Manual

Agilent G1978B  
Multimode Source for  
6100 Series Single Quad  
LC/MS  
Set-Up Guide  
Agilent Technologies  
In This Guide  
This guide explains how to install, maintain and troubleshoot your  
nanoelectospray ion source.  
This chapter tells you how to install the multimode ion source.  
This chapter tells you how to install the multimode ion source.  
This chapter describes basic operation and maintenance for the multimode  
ion source.  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
3
4
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
Contents  
6
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
Agilent G1978B Multimode Source for 6100 Series Single Quad LC/MS  
Set-Up Guide  
1
Installation  
This chapter contains instructions to install the multimode source on a 6100  
Series Single Quad LC/MS system, and also to remove and replace the source.  
Agilent Technologies  
7
     
1
Installation  
Installation  
Installation  
This section explains how to install the G1978B multimode source on a 6100  
Series Single Quad LC/MS instrument. The G1978A multimode source is not  
supported on these instruments.  
The 6100 Series Single Quad LC/MS with multimode source is supported on  
ChemStation B.03.01 or higher, or the ChemStation B.01.03 with a patch.  
Step 1. Prepare to install  
The Multimode Enablement Kit, G1978-60451, is shipped with the multimode  
source. This kit needs to be installed before the multimode source is used.  
Note that the multimode source and its accessories are to be installed by an  
Agilent Customer Engineer.  
1 Check that the Multimode Enablement Kit contains the following parts:  
Multimode Bd HV cable, p/n G1960-60858  
Multimode HV PCA, p/n G1960-61015  
Multimode Bd Power/Data Cable, p/n G1960-60873  
Figure 1  
From left to right: G1960-60858, G1960-61015 and G1960-60873  
2 Install the APCI Enablement Kit, G1947-60451, which is shipped with the  
multimode source.  
The APCI Enablement kit contains the following parts:  
Fast APCI HV Supply, p/n G1946-80058  
8
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
   
Installation  
1
Step 2. Install the HV control PCA and cables  
Valve BD-APCI supply, p/nG1960-60802  
Valve BD-APCI Needle Interlock, p/n G1960-60856  
Figure 2  
From left to right: G1946-80058, G1960-60802 and G1960-60856  
Step 2. Install the HV control PCA and cables  
1 Turn off the system power and remove the system power cord.  
Keep the power cord intact if the vacuum control switch box is used. The  
switch box is intended to keep the vacuum on while a CE works on the  
electronics. The switch box is for CE use only.  
2 Remove the CDS cover, top, side, front, and the Aux Module cover.  
3 Disconnect the ribbon cable that connects the valve PCA to the  
Vcap/Vchamber power supply. Then disconnect the Vcap and Vchamber  
cable from the power supply.  
Figure 3  
Disconnecting the Vcap/Vchamber power supply from the valve PCA (left) and the Vcap/Vchamber.  
4 Place the multimode HV power supply PCA in the slot between the valve  
PCA and the Vcap/Vchamber. Secure the board by pressing it down into its  
slot and then attach it with two screws.  
5 Connect the short gray cable from valve PCA to the multimode HV power  
supply.  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
9
 
1
Installation  
Step 2. Install the HV control PCA and cables  
Figure 4  
Connecting the valve PCA to the multimode HV power supply.  
6 Install the APCI HV power supply. The APCI HV power supply is located at  
the end of the AUX Module.  
7 Connect ribbon cable between valve PCA and Vcap/Vchamber power  
supply.  
Figure 5  
Connecting the valve PCA to the Vcap/Vchamber power supply.  
8 Connect the Vcap and Vchamber cables to the Vcap/Vchamber power  
supply.  
10  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
Installation  
1
Step 2. Install the HV control PCA and cables  
Figure 6  
Connecting the Vcap and Vchamber cables to the power supply.  
9 Connect the long ribbon cable, p/n G1960-60802, from the APCI HV power  
supply to the valve PCA.  
Figure 7  
Connecting the APCI HV power supply to the valve PCA.  
10 Insert one end of the APCI Needle Interlock cable, G1960-60856, through  
the slot at the front of the system and then plug it to the APCI HV  
connector. Attach the other end to the chassis with the o-ring and the nut.  
Figure 8  
Connecting the APCI HV to the chassis.  
11 Insert the cable, G1960-60858, to the top slot and attach it to the chassis.  
Plug the other two ends into the multimode HV PCA.  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
11  
1
Installation  
Step 2. Install the HV control PCA and cables  
Figure 9  
Connecting the HV PCA to the chassis.  
12 Close the AUX Module cover and reconnect all cables.  
13 Install the multimode source to the system and connect all connectors.  
Figure 10 Installing the multimode source (left) and connecting all connectors.  
14 Put back the side, top, front and CDS cover.  
15 Plug the system power cord back on and turn the front switch on.  
The pump down process will start.  
16 Start the ChemStation program.  
17 Click Method and Run Control View > MSD > Spray Chamber and verify  
that the source is an MM Source.  
18 Go to MSD Tune window and start the Auto tune with a multimode source.  
19 Verify that the multimode source is working correctly.  
12  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
Installation  
1
Changing Sources  
Changing Sources  
To remove the multimode source  
Do the following steps to remove the multimode source.  
1 Set the source temperatures for vaporizer heater and drying gas heater to  
minimum values to cool the source.  
Use the Tune > Instrument > Edit Spray Chamber menu item to display  
the Edit Spray Chamber dialog box. Set the drying gas flow, nebulizer gas  
flow, drying gas temperature and the vaporizer temperature to minimum  
values.  
Do not touch the multimode source or the capillary cap. They may be very hot. Let  
the parts cool before you handle them.  
WARNING  
WARNING  
Never touch the source surfaces, especially when you analyze toxic substances or  
when you use toxic solvents. The source has several sharp pieces which can pierce  
your skin including the APCI corona needle, vaporizer sensor and counter current  
electrode.  
Do not insert fingers or tools through the openings on the multimode chamber. When  
in use, the capillary and capillary cap are at high voltages up to 4 kV.  
WARNING  
2 Wait approximately 20 minutes or until the source is cool.  
3 Open the CDS door at the front of the MS to access the cables.  
4 Disconnect the ESI high voltage charging electrode cable.  
5 Disconnect the APCI Needle Interlock and multimode HV cable.  
6 Unscrew the nebulizer gas line from the nebulizer.  
7 Unscrew the LC sample tubing from the nebulizer.  
8 Open the latch on the source and open the source.  
9 Remove the multimode source from the spray chamber mount.  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
13  
   
1
Installation  
To convert from multimode to ESI, APCI or APPI  
10 Place the source shipping cover on the source.  
To convert from multimode to ESI, APCI or APPI  
Never touch the source surfaces, especially when you analyze toxic substances or  
WARNING  
when you use toxic solvents. The source has several sharp pieces which can pierce  
your skin including the APCI corona needle, vaporizer sensor and counter current  
electrode.  
1 If the source to be installed is an APPI source, disconnect the multimode  
high voltage PCA RS-232 serial cable from the serial port B connector of  
Smart Card.  
2 Unscrew and remove the multimode spray shield with the field shaping  
electrodes.  
3 Install the new source and the standard spray shield, making sure that the  
hole in the spray shield is in the 12 o'clock position.  
4 For APCI and APPI ion source, connect the vaporizer heater cable and the  
APCI high voltage cable. For the APPI source, connect the RS-232 cable to  
the serial port B connector of Smart Card.  
5 For all sources, reconnect the nebulizer gas line tubing and the LC/MS  
sample tubing.  
14  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
   
Installation  
1
To convert from ESI, APCI or APPI to the multimode source  
To convert from ESI, APCI or APPI to the multimode source  
If you are installing this source on this instrument for the first time, follow the steps in  
CAUTION  
1 Switch to the MSD Tune view.  
2 Select Instrument/Set Spray Chamber and set all gas flows and  
temperatures to 0.  
Drying Gas (L/min)  
Nebulizer Pressure (psig)  
Drying Gas Temperature (°C)  
Vaporizer Temperature (APCI source only)  
Lamp Off (APPI source only)  
3 Wait for the source to cool (until temperatures are at least below 100°C).  
4 Disconnect the nebulizer gas tubing from the currently installed ion source.  
5 Disconnect the LC/MS sample inlet tubing.  
6 If the APCI or APPI source is installed, remove the APCI vaporizer heater  
cable and APCI high voltage cable.  
7 If the APPI source is installed, remove the serial port B RS-232 cable.  
8 Remove the currently installed ion source.  
9 Unscrew and remove the spray shield. See Figure 11.  
Do not touch the multimode source or the capillary cap. They may be very hot. Let the  
parts cool before you handle them.  
WARNING  
WARNING  
Do not insert fingers or tools through the openings on the multimode chamber. When  
in use, the capillary and capillary cap are at high voltages up to 4 kV.  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
15  
   
1
Installation  
To convert from ESI, APCI or APPI to the multimode source  
Standard spray shield  
Capillary cap  
Figure 11 Standard spray shield and capillary cap for ESI or APCI  
10 Remove the capillary cap. If needed, moisten a clean cloth with isopropyl  
alcohol and wipe the capillary cap. See Figure 12.  
Capillary cap  
Figure 12 Spray shield removed.  
11 Place the capillary cap back on the capillary.  
12 Install the new spray shield with field shaping electrodes. See Figure 13.  
16  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
Installation  
1
To convert from ESI, APCI or APPI to the multimode source  
Figure 13 Multimode spray shield  
13 Screw the multimode spray shield into the holder for the spray shield. See  
Figure 14.  
Field shaping electrode  
9 o'clock position  
Field shaping electrode  
6 o'clock position  
Figure 14 Multimode spray shield installed  
The field shaping electrodes should be in the nine o’clock and the six o’clock position.  
Loosen the end plate screws on each side to adjust the field shaping electrodes position.  
NOTE  
14 Remove the shipping cover from the multimode source spray chamber.  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
17  
   
1
Installation  
To convert from ESI, APCI or APPI to the multimode source  
Figure 15 Multimode Spray Chamber  
15 Install the spray chamber on the spray chamber mount.  
I-Button  
Figure 16 Multimode source with I-Button  
16 Install the nebulizer on the multimode source spray chamber.  
18  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
Installation  
1
To convert from ESI, APCI or APPI to the multimode source  
Figure 17 No nebulizer on top of the multimode source  
17 Connect the 1/8-inch nebulizer gas tubing from the LC/MS mainframe to  
the nebulizer gas fitting. See Figure 18.  
Sample Tubing  
Nebulizer zero dead  
volume  
Nebulizer gas fitting  
Figure 18 Nebulizer with gas tubing connected  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
19  
 
1
Installation  
To convert from ESI, APCI or APPI to the multimode source  
18 Connect the LC/MS sample tubing to the LC/MS diverter valve inlet filter.  
See Figure 19 on page 20.  
The LC/MS Liquid Chromatograph diverter valve is an integral part of the G1978B  
safety system. The LC mobile phase flow must always be connected to the diverter  
valve inlet filter. Never bypass the diverter valve and connect directly to the  
nebulizer. If the diverter valve is used in a manner not specified by Agilent  
Technologies, the protections provided by the diverter valve may be impaired.  
WARNING  
LC/MS sample tubing  
Figure 19 LC/MS sample tubing connected to LC/MS inlet filter  
19 If you are installing the multimode source for the first time, follow the steps  
20  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
   
2
Verification  
To verify sensitivity for 6120 Quad - G1978B Multimode, Positive SIM Modes  
To verify sensitivity for 6120 Quad - G1978B Multimode, Positive  
SIM Modes  
The G6110A instrument does not support the multimode source without the Selection  
NOTE  
valve. Use the option kit G2735H to enable the G6110A to work with the multimode source.  
Preparation  
1 Transfer 1 mL of 5 ng/µL reserpine (Agilent G2423A) to a 50 mL volumetric  
flask. Use a clean graduated pipette.  
2 Dilute to the 50 mL mark with 75:25 methanol/water with 0.1% formic or  
acetic acid.  
3 Transfer 1 mL of the first dilution to a second 50 mL volumetric flask. Use a  
clean graduated pipette.  
4 Dilute to the 50 mL mark with 75:25 methanol/water with 0.1% formic or  
acetic acid.  
This provides the final 2 pg/µL reserpine concentration required for  
performance verification.  
5 Transfer approximately 1 mL of the second dilution to a vial for use in the  
LC autosampler.  
Table 1  
6110 and 6120 Quad Performance Verification Summary Table  
G1978B  
G1978B  
ES Positive SIM Mode  
APCI Positive SIM Mode  
Sample  
Reserpine, 5 ng/µL  
2 pg/µL  
Reserpine, 5 ng/µL  
2 pg/µL  
Concentration after dilution  
Injection volume  
5 µL  
5 µL  
Total sample amount injected 10 pg  
10 pg  
Sample order number  
Solvent  
G2423A  
G2423A  
75:25  
75:25  
methanol/water  
methanol/water  
with 0.1% formic or acetic acid with 0.1% formic or acetic acid  
22  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
 
Verification  
2
To verify sensitivity for 6120 Quad - G1978B Multimode, Positive SIM Modes  
Table 1  
6110 and 6120 Quad Performance Verification Summary Table  
G1978B  
G1978B  
ES Positive SIM Mode  
APCI Positive SIM Mode  
Method name  
6110SIMES_MM.M  
(6110 Quad)  
6110SIMCI_MM.M  
(6110 Quad)  
6120SIMES_MM.M  
(6120 Quad)  
6120SIMCI_MM.M  
(6120 Quad)  
Performance Specification  
20: 1 pk-pk  
100: 1 rms  
10: 1 pk-pk  
50: 1 rms  
Sensitivity Verification for Multimode in ESI Positive SIM Mode  
1 Start up the ChemStation software.  
2 Change the view to MSD Tune, and start an Autotune.  
After the autotune completes, you may need to wait up to 30 minutes to  
allow for the calibrant solution to be pumped out of the Quad instrument.  
This minimizes any background signal resulting from the calibrant.  
3 Change the view to Method and Run Control, and load the method  
6120SIMES_MM.M (6120 model).  
4 Edit the method to ensure that 75:25 methanol/water with 0.1% formic or  
acetic acid is selected as the LC solvent.  
5 Place the vials into the LC autosampler.  
Position #1: empty, uncapped vial  
Position #2: vial of the solvent used for dilution (solvent blank)  
Position #3: vial with the reserpine sample (2 pg/µL)  
6 Run the method.  
The method does an FIA run with one injection of the empty vial, five  
injections of the solvent blank, and five injections of the reserpine sample.  
You may need to further optimize the nebulizer pressure by running FIA  
experiments to achieve maximum instrument sensitivity.  
7 Review the results.  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
23  
2
Verification  
To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive SIM Modes  
When the method is finished, a report prints showing the signal-to-noise ratio  
for the five blank and five sample peaks, and a blank-subtracted average of the  
sample peaks. The blank-subtracted average signal-to-noise value must be  
greater than or equal to 20:1 peak-to-peak (100:1 RMS).  
To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode,  
Positive SIM Modes  
Preparation  
1 Transfer 1 mL of 5 ng/µL reserpine (Agilent G2423A) to a 50 mL volumetric  
flask. Use a clean graduated pipette.  
2 Dilute to the 50 mL mark with 75:25 methanol/water with 5 mM ammonium  
formate.  
3 Transfer 1mL of the first dilution to a 100 mL volumetric flask. Use a clean  
graduated pipette.  
4 Dilute to the 100 mL mark with 75:25 methanol/water with 5 mM  
ammonium formate.  
This provides the final 1 pg/µL reserpine concentration required for perfor-  
mance verification.  
5 Transfer approximately 1 mL of the second dilution to an autosampler vial.  
Table 2  
6130 and 6140 Quad Performance Verification Summary, SIM Mode  
G1978B  
G1978B  
ES Positive SIM Mode  
APCI Positive SIM Mode  
Sample  
Reserpine, 5 ng/µL  
1 pg/µL  
1 µL  
Reserpine, 5 ng/µL  
1 pg/µL  
1 µL  
Concentration after dilution  
Injection volume  
Total sample amount injected  
Sample order number  
1 pg  
1 pg  
G2423A  
G2423A  
24  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
 
Verification  
2
To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive SIM Modes  
Table 2  
6130 and 6140 Quad Performance Verification Summary, SIM Mode  
G1978B  
G1978B  
ES Positive SIM Mode  
APCI Positive SIM Mode  
Solvent  
75:25  
75:25  
methanol/water  
with 5mM  
methanol/water  
with 5mM  
ammonium formate  
ammonium formate  
Method name  
6130SIMES_MM.M  
(6130 model)  
6130SIMCI_MM.M  
(6140 model)  
6140SIMES_MM.M  
(6140 model)  
6140SIMCI_MM.M  
(6140 model)  
Performance Specification  
20: 1 pk-pk  
10: 1 pk-pk  
100: 1 rms  
50: 1 rms  
Sensitivity Verification for Multimode in ESI, Positive SIM Mode  
1 Start up the ChemStation software.  
2 Change the view to MSD Tune, and start an Autotune.  
After the autotune completes, you may need to wait up to 30 minutes to  
allow for the calibrant solution to be pumped out of the Quad instrument.  
This minimizes any background signal resulting from the calibrant.  
3 Change the view to Method and Run Control, and load the method  
6130SIMES_MM.M (6130 model) or 6140SIMES_MM.M (6140 model).  
4 Edit the method to ensure that 75:25 methanol/water with 5mM ammonium  
formate is selected as the LC solvent.  
5 Place the vials into the LC autosampler.  
Position #1: empty, uncapped vial  
Position #2: vial of the solvent used for dilution (solvent blank)  
Position #3: vial with the reserpine sample (1 pg/µL)  
6 Run the method.  
The method does an FIA run with one injection of the empty vial, five  
injections of the solvent blank, and five injections of the reserpine sample.  
You may need to further optimize the nebulizer pressure by running FIA  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
25  
2
Verification  
To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive SIM Modes  
experiments to achieve maximum instrument sensitivity  
7 Review the results.  
When the method is finished, a report prints showing the signal-to-noise ratio  
for the five blank and five sample peaks, and a blank-subtracted average of the  
sample peaks. The blank-subtracted average signal-to-noise value must be  
greater than or equal to 20:1 peak-to-peak (100:1 RMS).  
Sensitivity Verification for Multimode in APCI, Positive SIM Mode  
1 Start the ChemStation software.  
2 Change the view to MSD Tune, and start an Autotune.  
After the autotune completes, you may need to wait up to 30 minutes before  
you continue, to allow for the calibrant solution to be pumped out of the  
Quad instrument. This minimizes any background signal resulting from the  
calibrant.  
3 Change the view to Method and Run Control, and load the method  
6130SIMCI_MM.M (6130 model) or 6140SIMCI_MM.M (6140 model)  
4 Edit the method to ensure that 75:25 methanol/water with 5mM ammonium  
formate is selected as the LC solvent.  
5 Place the vials into the LC autosampler.  
Position #1: empty, uncapped vial  
Position #2: vial of the solvent used for dilution (solvent blank)  
Position #3: vial with the reserpine sample (1 pg/µL)  
6 Run the method.  
You may need to further optimize the nebulizer pressure by running FIA  
experiments to achieve maximum instrument sensitivity  
The method does an FIA run with one injection of the empty vial, five  
injections of the solvent blank, and five injections of the reserpine sample.  
7 Review the results.  
When the method is finished, a report prints showing the signal-to-noise ratio  
for the five blank and five sample peaks, and a blank-subtracted average of the  
sample peaks. The blank-subtracted average signal-to-noise value must be  
greater than or equal to 10:1 peak-to-peak (50:1 RMS).  
26  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
Verification  
2
To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive Scan Modes  
To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode,  
Positive Scan Modes  
Preparation  
1 Transfer 1 mL of 5 ng/µL reserpine (Agilent G2423A) to a 50 mL volumetric  
flask. Use a clean graduated pipette.  
2 Dilute to the 50 mL mark with 75:25 methanol/water with 5 mM ammonium  
formate.  
3 Transfer 5 mL of the first dilution to a 50 mL volumetric flask. Use a clean  
graduated pipette.  
4 Dilute to the 50 mL mark with 75:25 methanol/water with 5 mM ammonium  
formate.  
This provides the final 10 pg/µL reserpine concentration required for  
performance verification.  
5 Transfer approximately 1 mL of the second dilution to an autosampler vial.  
Table 3  
6130 and 6140 Quad Performance Verification Summary, Scan Mode  
G1978B  
G1978B  
ES Positive Scan Mode  
APCI Positive Scan Mode  
Sample  
Reserpine, 5 ng/µL  
10 pg/µL  
Reserpine, 5 ng/µL  
10 pg/µL  
5 µL  
Concentration after dilution  
Injection volume  
5 µL  
Total sample amount injected 50 pg  
50 pg  
Sample order number  
Solvent  
G2423A  
G2423A  
75:25  
75:25  
methanol/water  
with 5mM  
methanol/water  
with 5mM  
ammonium formate  
ammonium formate  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
27  
 
2
Verification  
To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive Scan Modes  
Table 3  
6130 and 6140 Quad Performance Verification Summary, Scan Mode  
G1978B  
G1978B  
ES Positive Scan Mode  
APCI Positive Scan Mode  
Method name  
6130SCNES_MM.M  
(6130 Quad)  
6130SCNCI_MM.M  
(6130 Quad)  
6140SCNES_MM.M  
(6140 Quad)  
6140SCNCI_MM.M  
(6140 Quad)  
Performance Specification  
20: 1 pk-pk  
100: 1 rms  
10: 1 pk-pk  
50: 1 rms  
Sensitivity Verification for Multimode in ESI, Positive Scan Mode  
1 Start the ChemStation software.  
2 Change the view to MSD Tune, and start an Autotune.  
After the autotune completes, you may need to wait up to 30 minutes to  
allow for the calibrant solution to be pumped out of the Quad instrument.  
This minimizes any background signal resulting from the calibrant.  
3 Change the view to Method and Run Control, and load the method  
6130SCNES_MM.M (for 6130) or 6140SCNES_MM.M (for 6140).  
4 Edit the method to ensure that 75:25 methanol/water with 5mM ammonium  
formate is selected as the LC solvent.  
5 Place the vials into the LC autosampler.  
Position #1: empty, uncapped vial  
Position #2: vial of the solvent used for dilution (solvent blank)  
Position #3: vial with the reserpine sample (10 pg/µL)  
6 Run the method.  
You may need to further optimize the nebulizer pressure by running FIA  
experiments to achieve maximum instrument sensitivity.  
The method does an FIA run with one injection of the empty vial, five  
injections of the solvent blank, and five injections of the reserpine sample.  
7 Review the results.  
28  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
Verification  
2
To verify sensitivity for 6130 and 6140 Quads - G1978B Multimode, Positive Scan Modes  
When the method is finished, a report prints showing the signal-to-noise  
ratio for the five blank and five sample peaks, and a blank-subtracted  
average of the sample peaks. The blank-subtracted average signal-to-noise  
value must be greater than or equal to 20:1 peak-to-peak (100:1 RMS).  
Sensitivity Verification for Multimode in APCI, Positive Scan Mode  
1 Start the ChemStation software.  
2 Change the view to MSD Tune, and start an Autotune.  
After the autotune completes, you may need to wait up to 30 minutes to  
allow for the calibrant solution to be pumped out of the Quad instrument.  
This minimizes any background signal resulting from the calibrant.  
3 Change the view to Method and Run Control, and load the method  
6130SCNCI_MM.M (6130 model) or 6140SCNCI_MM.M (6140 model).  
4 Edit the method to ensure that 75:25 methanol/water with 5mM ammonium  
formate is selected as the LC solvent.  
5 Place the vials into the LC autosampler.  
Position #1: empty, uncapped vial  
Position #2: vial of the solvent used for dilution (solvent blank)  
Position #3: vial with the reserpine sample (10 pg/µL)  
6 Run the method.  
The method does an FIA run with one injection of the empty vial, five  
injections of the solvent blank, and five injections of the reserpine sample.  
You may need to further optimize the nebulizer pressure by running FIA  
experiments to achieve maximum instrument sensitivity  
7 Review the results.  
When the method is finished, a report prints showing the signal-to-noise ratio  
for the five blank and five sample peaks, and a blank-subtracted average of the  
sample peaks. The blank-subtracted average signal-to-noise value must be  
greater than or equal to 10:1 peak-to-peak (50:1 RMS).  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
29  
2
Verification  
To verify sensitivity for the G1978B Multimode using Multiple FIA method in Mixed Mode operation  
To verify sensitivity for the G1978B Multimode using Multiple FIA  
method in Mixed Mode operation  
Preparation  
1 Do an autotune if necessary.  
After the autotune has completed, you may need to wait up to 30 minutes  
before continuing to allow the calibrant solution to be pumped out of the  
Quad instrument. This minimizes any background signal from the calibrant.  
You may need to further optimize the nebulizer pressure to achieve  
maximum instrument sensitivity.  
2 Load the method MMCheckSL_ES.M and:  
a Edit the method to ensure that 65:35 methanol/water with .2% acetic acid  
is selected as the LC solvent.  
b Save the method.  
3 Repeat for MMCheckSL_CI.M and MMCheckSL_MX.M.  
4 Place the vial into the LC autosampler.  
Position 21: vial with the ESI + APCI LC Demo Sample (p/n  
G1978-85000)  
Create and run a Multiple FIA Method Sequence  
1 Select RunControl > Run Multiple FIA Methods.  
2 In the Run Multiple FIA Methods dialog box, click Group > Add Group.  
3 Type in a unique directory name for the data files to be stored.  
4 In the Run Multiple FIA Methods dialog box, click Methods > Add Method,  
select MMCheckSL_ES.M and click OK.  
5 Repeat to add MMCheckSL_CI.M.  
6 Repeat to add MMCheckSL_MX.M.  
7 Click Data File for the MMCheckSL_ES.M data file, then:  
a Change the Subdirectory value to a unique subdirectory name in which  
to store the data files.  
b Change Data File to Multi_ES.  
c Change Operator to the user’s name or identification code.  
30  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
 
Verification  
2
To verify sensitivity for the G1978B Multimode using Multiple FIA method in Mixed Mode operation  
d Click OK.  
8 Repeat for the MMCheckSL_CI.M data file. Use the same subdirectory and  
change Data File to Multi_CI.  
9 Repeat for the MMCheckSL_MX.M data file. Use the same subdirectory and  
change Data File to Multi_MX as the data file name.  
10 Click Run to start the sequence.  
11 Review the results.  
When the last method has finished, the report “Multimode Verification Report”  
prints.  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
31  
2
Verification  
To do an autotune  
To do an autotune  
Tuning the multimode source is done in MM-APCI mode only. Autotune is  
done from the same menu as with all sources.  
From the MSD Tune view, select Instrument > Autotune menu item.  
The tune report will have a header with either the title MM-APCI Positive  
Mode - Standard Scan or MM-APCI Negative Mode - Standard Scan. You can  
run the check tune after an autotune to validate that the instrument passes  
check tune criteria. Do the autotune after the system has had at least 8 hours  
to equilibrate vacuum and temperatures.  
Figure 20 Autotune report  
32  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
   
Verification  
2
Example of multimode verification report  
Example of multimode verification report  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
33  
 
2
Verification  
Example of multimode verification report  
34  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
Agilent G1978B Multimode Source for 6100 Series Single Quad LC/MS  
Set-Up Guide  
3
Methods  
This chapter describes the tasks that you need to set up methods for the  
multimode source. See “Guidelines” in the Agilent G1978A/B Multimode  
Source Maintenance Guide for suggested source conditions for the multimode  
source for the different ionization modes.  
Agilent Technologies  
35  
       
3
Methods  
To setup a method to use the multimode source  
To setup a method to use the multimode source  
To have your method use a multimode source, follow these steps:  
1 Open the MSD Spray Chamber dialog box by clicking Instrument > MSD  
Spray Chamber in the Method and Run Control view.  
2 Set Method Spray Chamber to MM-ES+APCI.  
3 Verify that Installed Spray Chamber is set to MM-ES+APCI.  
4 Make any other changes that are necessary for your method.  
5 Click the OK button.  
6 Open The Set up MSD Signals dialog box by clicking Instrument > More >  
Set up MSD Signals in the Method and Run Control view.  
7 Choose the desired ionization mode from the Ionization list. This list is  
only visible if the Method Spray Chamber was set to MM-ES+APCI. You  
may set the ionization mode to one of the following:  
MM-ES  
MM-APCI  
MM-ES+APCI  
8 Make any other changes that are necessary for your method.  
9 Click the OK button.  
The 6100 Series Single Quad LC/MS Liquid Chromatograph diverter valve is an  
WARNING  
integral part of the G1978B safety system. The LC mobile phase flow must always be  
connected to the diverter valve inlet filter. Never bypass the diverter valve and  
connect directly to the nebulizer. If the diverter valve is used in a manner not  
specified by Agilent Technologies, the protections provided by the diverter valve may  
be impaired and the system may catch fire.  
36  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
   
Methods  
3
To create a method for positive/negative mixed mode operation  
To create a method for positive/negative mixed mode operation  
1 Open the MSD Spray Chamber dialog box by clicking the Instrument > Set  
Up MSD Signals in the Method and Run Control view.  
2 Select MM-ES_APCI from the Method Spray Chamber drop-down list.  
3 Check that Installed Spray Chamber is also set to MM-ES+APCI.  
4 Make any other changes that are necessary for your method.  
5 Click the OK button.  
6 Open the Set up MSD Signals dialog box by clicking the Instrument > MSD  
Spray chamber in the Method and Run Control view.  
7 Modify the settings so that Signal 1 has Positive polarity and Signal 2 has  
Negative polarity as shown in Figure 21.  
8 Make any other changes that are necessary for your method.  
9 Click the OK button.  
Fast positive/negative polarity switching is a very useful technique but it  
requires time for the ion chemistry to be established and the optics path to  
refill with ions. The gas density plays a role in the speed of refilling the ion  
path. The gas density is affected by source temperature. For a method running  
positive/negative switching, use a lower vaporizer temperature (150 to 200°C)  
and a lower Vcap (approximately 1000 V). These will greatly affect the quality  
of the results in positive/negative switching experiments.  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
37  
   
3
Methods  
To create a method for positive/negative mixed mode operation  
Polarity options for  
Signal 1 and Signal 2  
Figure 21 Positive/Negative polarity switching using Set Up MSD Signals dialog box  
38  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
 
Methods  
3
To create a method for alternating ESI and APCI operation  
To create a method for alternating ESI and APCI operation  
1 Open The MSD Spray Chamber dialog box by clicking Instrument > MSD  
Spray Chamber in the Method and Run Control view.  
2 Set Method Spray Chamber to MM-ES+APCI.  
Figure 22 Method Spray Chamber set to MM-ES+APCI.  
3 Check that Installed Spray Chamber is also set to MM-ES+APCI.  
4 Make any other changes that are necessary for your method.  
5 Click the OK button.  
6 Open the Set up MSD Signals dialog box by clicking Instrument > Set Up  
MSD Signals in the Method and Run Control view.  
7 Modify the settings so that Signal 1 has Ionization value of MM-ES and  
Signal 2 has Ionization value of MM-APCI as shown in Figure 23.  
8 Make any other changes that are necessary for your method.  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
39  
   
3
Methods  
To create a method for alternating ESI and APCI operation  
9 Click the OK button.  
In general, use the mixed mode operation (MM-ES+APCI setting with Signal 1) instead of  
NOTE  
alternating MM-ES (Signal 1) and MM-APCI (Signal 2) mode switching. Two times as many  
scans are obtained during elution of a chromatographic peak, and no delay is needed  
between scans, resulting in better data. It is rarely necessary to know whether a compound  
responds purely in ESI or APCI modes on a chromatographic time scale.  
Ionization mode for  
Signal 1 and Signal 2  
Figure 23 MM-ES and MM-APCI switching using Set Up MSD Signals dialog box  
40  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
Index  
Index  
A
autotune, 32  
C
converting from ESI, APCI or APPI, 15  
converting to ESI or APCI, 14  
D
diverter valve inlet filter, 20  
E
ESI  
convert from, 15  
convert to, 14  
I
installation, 7, 21  
L
LC/MSD sample tubing, 20  
M
method  
alternating ESI and APCI, 39  
basic setup, 36  
positive/negative mixed mode, 37  
multimode  
nebulizer, 19  
source image, 35  
P
parts  
multimode spray shield, 17  
S
spray shield for multimode source, 17  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
41  
Index  
42  
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide  
In This Book  
This book contains  
installation, operation,  
maintenance and  
troubleshootinginstruction  
for the Multimode Source  
for 6100 Series Single Quad  
LC/MS.  
© Agilent Technologies, Inc. 2008  
Printed in USA  
First Edition, December 2008  
*G1978-90070*  
G1978-90070  
Agilent Technologies  

Kompernass KH 2254 User Manual
iLive ISPK2806 User Manual
Hitachi Travelstar HTS547550A9E384 User Manual
Giant Coffeemaker GP7150A User Manual
Fujitsu MHZ2120BJ User Manual
Fluke 1654B User Manual
Emerson CK5250 User Manual
Bunn 37614 041410 User Manual
BT 1010 User Manual
Antec Quiet Computing 0761345 03270 4 User Manual