Omega Omegaette Hht 1500 User Manual

TABLE OF CONTENTS  
1. SPECIFICATIONS ...................................................................................... 1  
2. PANEL FUNCTIONS ................................................................................... 2  
3. OPERATING INSTRUCTION..................................................................... 4  
3.1. Normal Operation................................................................................... 5  
3.2. Using the Mechanical Adaptor (Option RM1502)................................ 6  
3.3. Measurement of Surface Speed(m/min, ft/min, or yd/min)................. 7  
3.4. Use as an Event Counter with External Light Source.......................... 8  
3.5. Use as an Event Counter with Internal Red LED................................. 8  
3.6. External Signal(TTL) Input (RM1501)................................................... 9  
3.7. Digital Pulse Signal Output (RM1501) ................................................. 9  
3.8. Measuring Slow Rotating Objects........................................................ 10  
3.9. Using the MAX/MIN/AVE Function...................................................... 10  
3.10. Turning On/Off Tachometer............................................................... 11  
3.11. Disable Auto-Power-Off Function...................................................... 11  
3.12. REPLACING THE BATTERIES ....................................................... 12  
4. Protocol of RS-232C Serial Interface (RM-1501)..................................... 13  
TM  
5. Installation of Window Application Software (RM-1501)........................ 14  
TM  
6. Description of Windows Application Program (RM-1501)..................... 15  
1. SPECIFICATIONS  
Speed (Non-contact):  
Range  
10.00 - 99999  
Resolution  
0.01/0.1/1  
Accuracy  
0.04% ±2 dgts  
RPM  
rps(Hz)  
0.200 - 2000.0 0.001/0.01/0.1 0.04% ±2 dgts  
Speed (Contact)  
Range  
20.00 - 29999  
Resolution  
0.01/0.1/1  
Accuracy  
0.04% ±2 dgts  
RPM / (/  
symoblizes  
"Contact"  
m/min  
2.000 - 2999.9 0.001/0.01/0.1 0.04% ±2 dgts  
ft/min  
yard/min  
6.00 - 10000  
4.00 - 3000  
0.01/0.1/1  
0.01/0.1/1  
0.04% ±2 dgts  
0.04% ±2 dgts  
Event Counter:  
Range  
Max. Input Frequency  
0 - 99999  
10KHz. with 5% duty cycle  
External TTL Input  
Display:  
Sampling Rate:  
High > 4.5V (RM1501)  
5 digits 99999 counts  
0.7 second (> 60 rpm)  
> 1 second (10 to 60 rpm)  
50 to 300 mm  
Measuring Distance:  
Time Base: 4.0 MHz Quartz Crystal  
Range Selection:  
Battery:  
Automatic  
Four 1.5V batteries (AA, UM-3)  
Power Consumption:  
1.5mA (Idle)  
5mA (1000.0 - 99999 rpm)  
25mA (10.00 - 999.99 rpm)  
25mA (Max, Min or Ave Enabled)  
25mA (Event Counter)  
Auto-Power-Off:  
Operating Temp.  
Size:  
Weight:  
Accessories:  
30 minutes.  
0 to 50°C (32 to 122 °F)  
172 mm x 63 mm x 36 mm (6.8" x 2.5" x 1.5")  
190 g(6.7 oz.) (including battery)  
Carrying case x 1, Reflective tape x 1,  
Instruction Manual x 1  
1
1
2. PANEL FUNCTIONS  
2
2
2.1. Reflecting tape  
Attach a reflecting tape to the surface of the unit to be measured  
2.2. Reflecting signal light beam  
It is recommenced that the reflecting surface is perpendicular to the  
emitting/receiving unit for best measurement.  
2.3. Emitting/Receiving Unit  
The red light emits from the left side of the tachometer. And a receiving  
sensor is installed in the right side to receive light signal from the reflecting  
surface.  
2.4. LCD  
The lower row of the LCD is a 5 digit number to display RPM, m/min,  
ft/min, yd/min or counts.  
The upper row of the LCD is a clock. If only the minutes is displayed,  
tachometer is in the auto-power off mode. Power will be turned off in 30  
minutes. To disable auto-power-off mode, press and hold the unit button  
(left button) for 2 seconds. Then the hours will be displayed.  
2.5. Function Button  
Press the button to select MAX, MIN or AVE function. To return to normal  
mode, press and hold the button for 2 seconds.  
2.6. Start/Hold/On/Off  
When the power is turned off, press this button to turn the power on. If the  
power is on, press this button to start measurement. Press again to hold  
the data and stop measurement. To turn the power off, press and hold  
button until "OFF" is displayed in the LCD (about 2 seconds). Release the  
button, then the power will be turned off.  
2.7. Unit Button  
Press this button to select different units, such as RPM, RPM / ( /  
Symbolizes "contact"), m/min, ft/min, yd/min, rps (Hz) or No. of counts. To  
measurement speed in RPM /, m/min, ft/min or yd/min, the option RM1502  
mechanical adaptor must be attached.  
When the No. symbol is displayed in the LCD, the unit becomes an event  
counter. The clock will be reset to zero, and start counting. To stop  
counting, press the HOLD (START) button once. To reset counts, press  
the function button(right button).  
2.8. MAX/MIN/AVE symbols  
Symbols to indicate which of MAX, MIN or AVE function is enabled.  
2.9. Low battery warning  
3
3
2.10. Start Measurement Symbol  
2.11. Hours:Minutes Clock  
4
4
3. OPERATING INSTRUCTION  
5
5
3.1. Normal Operation  
a. Install four 1.5V AA size batteries.  
b. Cut off a piece of reflective tape of size 1.0 cm x 1.4 cm (optimal  
size)  
c. Wipe off oil or stains from the surface where reflective tape will be  
adhered.  
d. Stick the self-adhesive reflective tape on the object whose  
rotational speed is to be measured. The reflective tape should be  
attached as close to the outer edge of the object to be measured as  
possible.  
e. Press the START button to turn on tachometer.  
f. Point the photoelectric probe at the object as Figure 2. Press the  
START button once and confirm that the red emitting LED lamp lights  
up. Then read the rotational speed on the LCD display.  
g. If the rotational speed is less than 60 rpm, it may take about 1 to 6  
seconds to measure (60 rpm takes 1 second, and 10 rpm takes 6  
seconds).  
h. The reading will be held for 6 seconds, if you aim the photoelectric  
probe at a 0 speed object right after one measurement.  
Note: If the speed is less than 999.99 rpm, the red LED is turned on all the  
time to detect speed. If the speed is great than 999.99, the unit goes into  
power-saving mode. The LED is turned on for a very short period of time,  
then turned off for most of the time to save power.  
6
6
3.2. Using the Mechanical Adaptor (Option RM1502)  
a. Select proper rubber head.  
b. Attach the mechanical adaptor to the unit.and fasten the screw.  
c. Press the unit button (left button) to select RPM / ( instead of RPM  
only). The auxiliary LED will be turned on while the main LED will be  
turned off.  
d. Press against the shaft of the object to be measured.  
e. Do not press against the shaft too hard to avoid hazard.  
WARNING: Do not exceed the speed limit of 29999 rpm when using  
mechanical adaptor.  
7
7
3.3. Measurement of Surface Speed(m/min, ft/min, or yd/min)  
a. Attach the mechanical adaptor (option RM-1502) to the unit.  
b. Press the unit button to select desired unit. The auxiliary LED will  
be turned on while the main LED will be turned off.  
c. Slightly touch against the surface to be measured.  
8
8
3.4. Use as an Event Counter with External Light Source  
a. Press the unit button several times to select the unit of "No. O"  
(counts). The "O" symbol is used to instruct users that external light  
source is required.  
b. When the symbol "No. O" appears, the tachometer reset the upper  
clock to 00:00 and start counting the pulses it detects. (Auto-power-  
off is automatically disabled)  
c. To stop counting, press the START button once.  
d. To reset counts to zero, press the function(right) button.  
3.5. Use as an Event Counter with Internal Red LED  
a. Press the unit button several times to select the unit of "No. "  
(counts).  
b. When the symbol "No." appears, the tachometer reset the upper  
clock to 00:00 and start counting the pulses it detects (Auto-power-  
off is automatically disabled).  
c. To stop counting, press the START button once.  
d. To reset counts to zero, press the function(right) button.  
9
9
3.6. External Signal(TTL) Input (RM1501)  
Instead of receiving signal from light source, users can input external  
TTL signal (High: > 4.5V, Low: 0V) through pin 8 of RS232C  
connector with pin 5 as signal ground.  
NOTE: It is strongly recommended that the emiting/receiving unit be  
covered to avoid undesired light signal coming in.  
3.7. Digital Pulse Signal Output (RM1501)  
Users can output digital pulse signal through pin 8 of RS232C  
connector with pin 5 as signal ground to oscilloscope.  
10  
3.8. Measuring Slow Rotating Objects  
If the rotating speed of the object which you are measuring is very  
slow, it is recommended that you use a tripod to hold the tachometer  
in place and use multiple pieces of tape for more accurate result.  
If multiple pieces of tape are used, each piece of tape should be  
located at a equal distance to each other on the object. Any number  
of pieces of tape are acceptable. But you need to divide the reading  
by the number of pieces to obtain the correct result.  
Correct RPM = Reading / (number of pieces of tape)  
Example: 4 pieces of tape and reading is 12 RPM.  
Correct Speed = 12 / 4 = 3 rpm  
3.9. Using the MAX/MIN/AVE Functions  
These functions allow users to analyze the stability of rotational  
speed. These value are actually peak (MAX), valley(MIN), and true  
average(AVE) values.  
MAX: Press the function button to select maximum function. The  
MAX function can be used to find the peak value of speeds  
measured.  
MIN: Press the function button to select minimum function. The MIN  
function can be used to find the valley value of speeds measured.  
AVE: Press the function button to select average function. The  
average function is a true average value.  
1
n
RPMAVE= å rpm)n ; å rpm) < 232 , n < 65535  
If summation of average value overflow, or n > 65535, the last  
average RPM will be used as initial value for successive calculation.  
And n will be reset to 1.  
When users press the HOLD button (same as START button), the  
max/min/ave values are still stored in memory. Users can press the  
function button (right button) to review the values. To return to  
normal mode, press and hold the function (right button) for 2  
11  
11  
seconds,  
and the max/min/ave symbols will disappear in LCD.  
3.10. Turning On/Off Tachometer  
Press any button will turn on the power of tachometer. To turn off the  
power of tachometer, press and hold the START button until word "OFF"  
is displayed in LCD. Then release the START button.  
3.11. Disable Auto-Power-Off Function  
If tachometer is in auto-power-off mode, only the minutes will be  
displayed. And tachometer will be turned off in 30 minutes. To disable  
the auto-power-off function, press and hold the UNIT button (left button)  
for 2 seconds until the hours is displayed.  
12  
12  
3.12.  
a. When  
REPLACING THE BATTERIES  
the symbol (low battery) appears in the LCD, it is  
time to replace the batteries.  
b. Remove the screw of the battery cover and remove the battery cover.  
c
Replace the old batteries with 4 new batteries. Do not mix different  
type of batteries together.  
d. Replace the battery cover and fasten the screw.  
13  
13  
4. Protocol of RS-232C Serial Interface (RM-1501)  
Ten bytes are sent out through RS-232C connector to PC. The definitions of  
each byte are as followings:  
Byte 1:  
Byte 2:  
Leading byte 0x0D  
Decimal point of LCD display  
bit0: dp1 (0000.0), if 1  
bit1: dp2 (000.00), if 1  
bit2: dp3 (00.000), if 1  
Flags for current status  
bit0: low battery, if 1  
bit1: Max. value overflow, if 1  
bit2: Counter overflow, if 1  
bit3: Auto-power-off disable, if 1  
bit4: Min. value overflow, if 1  
bit5: Ave. value overflow, if 1  
Unit  
Byte 3:  
Byte 4:  
bit0: rpm, if 1  
bit1: m/min, if 1  
bit2: ft/min, if 1  
bit3: yd/min, if 1  
bit4: rps, if 1  
bit5: counter with external light source, if 1  
bit6: counter without external light source, if 1  
Function  
bit0: normal, if 1  
bit1: max, if 1  
bit2: min, if 1  
bit3: average, if 1  
Flags for current status  
bit4: LCD reading over load (OL)  
bit7: Hold  
Byte 5:  
Byte 6:  
Byte 7:  
Byte 8:  
Byte 9:  
Byte 10:  
Least significant byte of reading in binary format  
2nd byte of reading in binary format  
3rd byte of reading in binary format  
Most significant byte of reading in binary format  
14  
14  
TM  
5. Installation of Window Application Software (RM-1501)  
5.1. For Windows 3.1  
TM  
TM  
A. Start Microsoft Windows  
B. Insert disk in drive A (or B)  
C. From Program Manager, select File menu and choose Run  
D. Type a:\setup (or b:\setup) and press Enter key  
5.2. For Windows 95  
TM  
A. Start Windows 95  
B. Insert disk in drive A (or B)  
C. Press START button, then select Run  
D. Type in a:\setup (or b:\setup) and press Enter key  
15  
15  
TM  
6. Description of Windows Application Program (RM-1501)  
Main Window:  
When the program is executed, the program automatically search for connected  
tachometer or available serial port. If no serial port is available, then a message  
of "No communication port" will be displayed, and the program exits . Once  
communication port is setup, a main window will be displayed on the screen as  
below:  
Sample: Sampling time.  
Option:  
Unit:  
Value:  
Range:  
Display option of anemometer. Normal, MAX, MIN, AVE  
rpm, m/min, ft/min, yard/min, counter 0, counter  
The value under VALUE is the reading from the tachometer  
The text displayed under RANGE is the range of the unit selected at  
the tachometer  
Minimum: The minimum value ever recorded by PC.  
Reset: Clear minimum and maximum value recorded.  
Maximum: The maximum value ever recorded by PC.  
Lower:  
Upper:  
The minimum value of the range specified.  
The maximum value of the range specified.  
File: If you select File, a pull down menu will show six options: Name,  
Start Recording, End Recording, VIEW, Plot Data from File, and Exit.  
Name:  
Enter file name to store data dispalyed in LCD at specified interval.  
Start Recording:  
End Recording:  
Start recording when selected  
End recroding when selected.  
16  
16  
VIEW FILE:  
If the View option under FILE in the Main Window is selected, a view window will  
be shown as above, and the users can review your ASCII data file. If a printer is  
connected to the PC, users can print out content selectively.  
File:Open users data file by selecting this menu. Users will be asked to enter file  
name. After the name being entered, program will read in one block  
of data. Number of records in one block depends on the memory  
size of PC. The bigger memory, the more records in one block.  
Blocks Read:  
Indicate how many blocks of records have been read.  
Records Read:  
Indicate how many records have been read.  
Chars./Record:  
Indicate how many characters in one record.  
Current Block:  
Indicate the current block number being reviewed.  
17  
17  
Selected:  
Indicate the current record number being selected.  
DISPLAY: DISPLAY menu has four options: DIGITAL, ANALOG, LIST, and  
GRAPHIC.  
DIGITAL:  
If this option is selected or CTRL+D is pressed, a window, which emulates  
multimeter's LCD display, shall appears on the screen.  
ANALOG:  
18  
18  
If this option is selected or CTRL+A is pressed, a window, which emulates an  
analog meter, shall appear on the screen.  
LIST:  
If this option is selected or CTRL+L is pressed, a window, which lists the date,  
function, range, and value every sampling, shall appear on the screen.  
GRAPHIC:  
If this option is selected or CTRL+G is pressed, a window, which emulates strip  
chart recorder, shall appear on the screen. The graphic window has two menu,  
PRINT and SCALE. Select the PRINT menu, the graphics will be printed  
through any printer connected to the PC. The SCALE menu allows user to set  
19  
19  
the minimum and maximum value for the Y (vertical) axis.  
Option:  
If you select Option, a pull down menu will show three options:  
Sample Rate, Baud Rate.  
Upper Limit: Enter upper limit. If upper limit is exceeded, a OVER message  
will be displayed in screen.  
Lower Limit: Enter lower limit. If displayed value is less than lower limit, a  
UNDER message will be displayed in screen.  
Graphic Mode:  
Sample Rate:  
Baud Rate:  
Select way of display in the graphic windows. Dot or  
BAR.  
Select this option to enter PC sampling time, or click at  
the SAMPLE on screen to enter sampling time.  
9600.  
20  
20  
21  
21  

Mitsubishi Electronics Muz Ge15na User Manual
Mitsubishi Motors Automobile Parts F9q User Manual
Panasonic Cu E12gke User Manual
Panasonic Rod End Bearings Hr E User Manual
Savin 3695 User Manual
Schumacher Automatic Swi 1555a User Manual
Toshiba Air Conditioner Rav Sm806bt Tr User Manual
Trane Central Air Conditioning User Manual
Trane Trani 22 5150 03 497 User Manual
York Millennium Nd600 User Manual