Teledyne Microscope Magnifier GFC7000E User Manual

INSTRUCTION MANUAL  
MODEL GFC7000E  
CARBON DIOXIDE ANALYZER  
© TELEDYNE ANALYTICAL INSTRUMENTS  
16830 Chestnut St.  
City of Industry, Ca. 91748  
USA  
Phone: 626-961-9221  
Phone: 626-934-1500  
Fax: 626-961-2538  
Fax: 626-934-1651  
04584  
REV. A1  
02-August-2004  
Copyright 2003 Teledyne Instruments Incorporated  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
GFC7000E Documentation  
TABLE OF CONTENTS  
SAFETY MESSAGES I  
TABLE OF CONTENTS ...................................................................................................................... II  
LIST OF APPENDICES.................................................................................................................... VII  
LIST OF FIGURES  
VII  
LIST OF TABLES  
VIII  
1. MGFC7000E DOCUMENTATION..................................................................................................... 1  
1.1. Using This Manual ...................................................................................................................1  
2. SPECIFICATIONS, APPROVALS AND WARRANTY.......................................................................... 5  
2.1. Specifications..........................................................................................................................5  
2.2. CE Mark Compliance ................................................................................................................6  
2.3. Warranty................................................................................................................................6  
3. GETTING STARTED....................................................................................................................... 9  
3.1. Unpacking and Initial Set Up.....................................................................................................9  
3.1.1. Electrical Connections......................................................................................................11  
3.1.2. Pneumatic Connections:...................................................................................................15  
3.1.2.1. Basic Pneumatic Connections......................................................................................15  
3.1.2.2. Connections with Internal Valve Options Installed..........................................................19  
3.2. Initial Operation....................................................................................................................21  
3.2.1. Startup..........................................................................................................................22  
3.2.2. Warm Up .......................................................................................................................23  
3.2.3. Warning Messages ..........................................................................................................24  
3.2.4. Functional Check.............................................................................................................25  
3.3. Initial Calibration Procedure ....................................................................................................26  
4. FREQUENTLY ASKED QUESTIONS............................................................................................... 33  
4.1. FAQ’s...................................................................................................................................33  
4.2. Glossary...............................................................................................................................34  
5. OPTIONAL HARDWARE AND SOFTWARE .................................................................................... 37  
5.1. Rack Mount Kits (Options 20a, 20b & 21)..................................................................................37  
5.2. Current Loop Analog Outputs (Option 41) .................................................................................37  
5.3. Expendable Kits (Options 42C, 42D and 43) ..............................................................................38  
5.4. Calibration Valves Options ......................................................................................................38  
5.4.1. Zero/Span/Shutoff Valve (Option 50).................................................................................38  
5.4.2. Zero/Span/Shutoff with External CO2 Scrubber (Option 51)..................................................40  
5.4.3. Zero/Span Valve (Option 52) ............................................................................................40  
5.4.4. Zero/Span Valve with External CO2 Scrubber (Option 53)......................................................41  
5.5. Communication Options..........................................................................................................41  
5.5.1. RS232 Modem Cable (Option 60).......................................................................................41  
5.5.2. RS-232 Multidrop (Option 62)...........................................................................................42  
5.5.3. Ethernet (Option 63) .......................................................................................................42  
5.6. Additional Manuals.................................................................................................................42  
5.6.1. Printed Manuals (Option 70) .............................................................................................42  
5.6.2. Manual on CD (Part number 045840200)............................................................................43  
5.7. Extended Warranty (Options 92 & 93) ......................................................................................43  
5.8. Dilution Ratio Option (Option ??) .............................................................................................43  
5.9. Maintenance Mode Switch (Option ??) ......................................................................................43  
5.10. Second Language Switch (Option ??)......................................................................................44  
6. OPERATING INSTRUCTIONS ...................................................................................................... 45  
6.1. Overview of Operating modes .................................................................................................45  
6.2. Sample Mode........................................................................................................................46  
6.2.1. Test Functions ................................................................................................................46  
6.2.2. Warning Messages ..........................................................................................................48  
6.3. Calibration Mode ...................................................................................................................50  
6.3.1. SETUP – PASS: Calibration Password Security .....................................................................50  
6.4. SETUP Mode .........................................................................................................................52  
6.4.1. SETUP Mode Password Security.........................................................................................53  
6.5. SETUP – CFG: Viewing the Analyzer’s Configuration Information ..................................................53  
6.6. SETUP – CLK: Setting the Internal Time-of-Day Clock.................................................................54  
04584 Rev A1  
ii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
GFC7000E Documentation  
6.7. SETUP – RNGE: Analog Output Reporting Range Configuration.....................................................56  
6.7.1. Physical Range versus Analog Output Reporting Ranges........................................................56  
6.7.2. Reporting Range Modes ...................................................................................................57  
6.7.3. Single Range mode (SNGL) ..............................................................................................58  
6.7.4. Dual Range Mode (DUAL).................................................................................................60  
6.7.5. Auto Range Mode (AUTO).................................................................................................61  
6.7.6. Range Units ...................................................................................................................62  
6.7.7. Dilution Ratio .................................................................................................................63  
6.8. SETUP – VARS: Using the Internal Variables..............................................................................64  
6.9. SETUP – DIAG: Using the Diagnostics Functions.........................................................................66  
6.9.1. Accessing the Diagnostic Features .....................................................................................67  
6.9.2. Signal I/O......................................................................................................................67  
6.9.3. Analog Output Step Test ..................................................................................................68  
6.9.4. Analog I/O Configuration..................................................................................................69  
6.9.4.1. Analog Output Signal Type and Range Span Selection....................................................71  
6.9.4.2. Analog Output Calibration Mode..................................................................................71  
6.9.4.3. Manual Analog Output Calibration and Voltage Adjustment .............................................73  
6.9.4.4. Current Loop Output Adjustment.................................................................................75  
6.9.4.5. AIN Calibration.........................................................................................................77  
6.9.5. Electric Test ...................................................................................................................77  
6.9.6. Dark Calibration Test.......................................................................................................78  
6.9.7. Pressure Calibration ........................................................................................................78  
6.9.8. Flow Calibration..............................................................................................................80  
6.9.9. Test Channel Output........................................................................................................81  
6.10. SETUP – COMM: Using the Analyser’s Communication Ports.......................................................82  
6.10.1. Analyzer ID Code ..........................................................................................................82  
6.10.2. COMM Port Default Settings............................................................................................83  
6.10.3. COMM Port Cable Connections.........................................................................................84  
6.10.4. RS-485 Configuration of COM2........................................................................................84  
6.10.5. DTE and DCE Communication..........................................................................................85  
6.10.6. COMM Port Communication Modes ...................................................................................85  
6.10.7. COM Port Baud Rate ......................................................................................................88  
6.10.8. COM Port Testing ..........................................................................................................89  
6.10.9. Ethernet Card Configuration............................................................................................89  
6.10.9.1. Ethernet Card COM2 Communication Modes and Baud Rate...........................................90  
6.10.9.2. Configuring the Ethernet Interface Option using DHCP..................................................90  
6.10.9.3. Manually Configuring the Network IP Addresses...........................................................92  
6.10.9.4. Changing the Analyzer’s HOSTNAME..........................................................................94  
6.11. SETUP – ALRM: Using the Gas Concentration Alarms.................................................................95  
6.12. SETUP – DAS: Using the Data Acquisition System (iDAS)...........................................................96  
6.12.1. iDAS Structure..............................................................................................................96  
6.12.1.1. iDAS Channels........................................................................................................97  
6.12.1.2. iDAS Parameters.....................................................................................................98  
6.12.1.3. iDAS Triggering Events ............................................................................................98  
6.12.2. Default iDAS Channels ...................................................................................................99  
6.12.2.1. Viewing iDAS Data and Settings.............................................................................. 103  
6.12.2.2. Editing iDAS Data Channels.................................................................................... 104  
6.12.2.3. Trigger Events...................................................................................................... 105  
6.12.2.4. Editing iDAS Parameters ........................................................................................ 106  
6.12.2.5. Sample Period and Report Period............................................................................. 107  
6.12.2.6. Number of Records ............................................................................................... 109  
6.12.2.7. RS-232 Report Function......................................................................................... 110  
6.12.2.8. Compact Report.................................................................................................... 111  
6.12.2.9. Starting Date ....................................................................................................... 111  
6.12.2.10. Disabling/Enabling Data Channels.......................................................................... 111  
6.12.2.11. HOLDOFF Feature................................................................................................ 112  
6.12.3. Remote iDAS Configuration........................................................................................... 113  
6.13. Remote Operation of the Analyzer........................................................................................ 115  
6.13.1. Remote Operation Using the External Digital I/O.............................................................. 115  
6.13.1.1. Status Outputs ..................................................................................................... 115  
04584 Rev A1  
iii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
GFC7000E Documentation  
6.13.1.2. Control Inputs ...................................................................................................... 116  
6.13.2. Remote Operation Using the External Serial I/O............................................................... 117  
6.13.2.1. Terminal Operating Modes...................................................................................... 117  
6.13.2.2. Help Commands in Terminal Mode........................................................................... 118  
6.13.2.3. Command Syntax ................................................................................................. 118  
6.13.2.4. Data Types .......................................................................................................... 119  
6.13.2.5. Status Reporting................................................................................................... 120  
6.13.2.6. Remote Access by Modem...................................................................................... 120  
6.13.2.7. COM Port Password Security................................................................................... 122  
6.13.2.8. APICOM Remote Control Program............................................................................ 122  
6.13.3. Additional Communications Documentation..................................................................... 123  
6.13.4. Using the MGFC7000E with a Hessen Protocol Network ..................................................... 123  
6.13.4.1. General Overview of Hessen Protocol ....................................................................... 123  
6.13.4.2. Hessen COMM Port Configuration ............................................................................ 124  
6.13.4.3. Activating Hessen Protocol ..................................................................................... 124  
6.13.4.4. Selecting a Hessen Protocol Type ............................................................................ 125  
6.13.4.5. Setting The Hessen Protocol Response Mode............................................................. 126  
6.13.4.6. Hessen Protocol Gas ID.......................................................................................... 126  
6.13.4.7. Setting Hessen Protocol Status Flags ....................................................................... 127  
6.13.4.8. Instrument ID Code .............................................................................................. 128  
7. CALIBRATION PROCEDURES.................................................................................................... 129  
7.1. Before Calibration................................................................................................................ 129  
7.1.1. Zero Air and Span Gas...................................................................................................129  
7.1.2. Calibration Gas Traceability ............................................................................................130  
7.1.3. Data Recording Devices ................................................................................................. 130  
7.2. Manual Calibration without Zero/Span Valves .......................................................................... 130  
7.3. Manual Calibration Checks .................................................................................................... 133  
7.4. Manual Calibration with Zero/Span Valves............................................................................... 133  
7.5. Manual Calibration Checks with Zero/Span Valves .................................................................... 137  
7.5.1. Zero/Span Calibration on Auto Range or Dual Ranges......................................................... 137  
7.5.2. Use of Zero/Span Valves with Remote Contact Closure ....................................................... 138  
7.6. Automatic Zero/Span Cal/Check (AutoCal) .............................................................................. 139  
7.6.1. AutoCal with Auto or Dual Reporting Ranges Modes Selected............................................... 142  
7.7. Calibration Quality............................................................................................................... 142  
8. EPA PROTOCOL CALIBRATION ................................................................................................. 143  
9. MAINTENANCE SCHEDULE & PROCEDURES .............................................................................. 145  
9.1. Maintenance Schedule.......................................................................................................... 145  
9.2. Predicting Failures Using the Test Functions ............................................................................ 148  
9.3. Maintenance Procedures....................................................................................................... 148  
9.3.1. Replacing the Sample Particulate Filter............................................................................. 149  
9.3.2. Rebuilding the Sample Pump .......................................................................................... 150  
9.3.3. Performing Leak Checks................................................................................................. 150  
9.3.3.1. Vacuum Leak Check and Pump Check........................................................................ 150  
9.3.3.2. Pressure Leak Check ............................................................................................... 150  
9.3.4. Performing a Sample Flow Check..................................................................................... 151  
9.3.5. Cleaning the Optical Bench............................................................................................. 151  
9.3.6. Cleaning Exterior Surfaces of the MGFC7000E ................................................................... 151  
10. THEORY OF OPERATION......................................................................................................... 153  
10.1. Measurement Method......................................................................................................... 153  
10.1.1. Beer’s Law ................................................................................................................. 153  
10.1.2. Measurement Fundamentals ......................................................................................... 154  
10.1.3. Gas Filter Correlation................................................................................................... 154  
10.1.4. Ambient CO2 Interference Rejection ............................................................................... 158  
10.2. Pneumatic Operation.......................................................................................................... 158  
10.2.1. Sample Gas Flow......................................................................................................... 159  
10.2.1.1. Critical Flow Orifice ............................................................................................... 159  
10.2.1.2. Sample Pressure Sensor ........................................................................................ 160  
10.2.1.3. Sample Flow Sensor.............................................................................................. 160  
10.2.1.4. Valve Options....................................................................................................... 161  
10.2.2. Purge Gas Pressure and Flow Control ............................................................................. 161  
04584 Rev A1  
iv  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
GFC7000E Documentation  
10.3. Electronic Operation........................................................................................................... 161  
10.3.1. Overview ................................................................................................................... 161  
10.3.2. CPU........................................................................................................................... 163  
10.3.3. Optical Bench & GFC Wheel........................................................................................... 163  
10.3.3.1. Sample Gas and GFC Temperature Control ............................................................... 163  
10.3.3.2. IR Source ............................................................................................................ 164  
10.3.3.3. GFC Wheel........................................................................................................... 164  
10.3.3.4. IR Photo-Detector ................................................................................................. 165  
10.3.4. Synchronous Demodulator (Sync/Demod) Assembly......................................................... 165  
10.3.4.1. Overview.............................................................................................................165  
10.3.4.2. Signal Synchronization and Demodulation ................................................................ 166  
10.3.4.3. Phase Lock Warning .............................................................................................. 167  
10.3.4.4. Sync/Demod Status LED’s...................................................................................... 167  
10.3.4.5. Photo-Detector Temperature Control........................................................................ 168  
10.3.4.6. Dark Calibration Switch.......................................................................................... 168  
10.3.4.7. Electric Test Switch ............................................................................................... 168  
10.3.5. Relay Board................................................................................................................ 168  
10.3.5.1. Status LED’s.........................................................................................................169  
10.3.5.2. I2C Watch Dog Circuitry........................................................................................ 170  
10.3.6. Mother Board.............................................................................................................. 170  
10.3.6.1. A to D Conversion................................................................................................. 170  
10.3.6.2. Sensor Inputs....................................................................................................... 171  
10.3.6.3. Thermistor Interface.............................................................................................. 171  
10.3.6.4. Analog Outputs..................................................................................................... 172  
10.3.6.5. Internal Digital I/O................................................................................................ 172  
10.3.6.6. External Digital I/O ............................................................................................... 173  
10.3.7. I2C Data Bus............................................................................................................... 173  
10.3.8. Power Supply/ Circuit Breaker....................................................................................... 173  
10.4. Interface .......................................................................................................................... 174  
10.4.1. Front Panel Interface ................................................................................................... 175  
10.4.1.1. Analyzer Status LED’s............................................................................................ 175  
10.4.1.2. Keyboard............................................................................................................. 176  
10.4.1.3. Display................................................................................................................ 176  
10.4.1.4. Keyboard/Display Interface Electronics..................................................................... 177  
10.5. Software Operation............................................................................................................ 178  
10.5.1. Adaptive Filter ............................................................................................................ 179  
10.5.2. Calibration - Slope and Offset........................................................................................ 179  
10.5.3. Measurement Algorithm ............................................................................................... 180  
10.5.4. Temperature and Pressure Compensation ....................................................................... 180  
10.5.5. Internal Data Acquisition System (iDAS)......................................................................... 180  
11. TROUBLESHOOTING & REPAIR PROCEDURES ........................................................................ 183  
11.1. General Troubleshooting Hints............................................................................................. 183  
11.1.1. Interpreting WARNING Messages................................................................................... 184  
11.1.2. Fault Diagnosis with TEST Functions............................................................................... 186  
11.1.3. Using the Diagnostic Signal I/O Function......................................................................... 188  
11.1.4. Internal Electronic Status LED’s..................................................................................... 189  
11.1.4.1. CPU Status Indicator ............................................................................................. 189  
11.1.4.2. Sync Demodulator Status LED’s .............................................................................. 190  
11.1.4.3. Relay Board Status LED’s ....................................................................................... 191  
11.1.5. Gas Flow Problems ......................................................................................................192  
11.1.6. Typical Sample Gas Flow Problems................................................................................. 193  
11.1.6.1. Flow is Zero .........................................................................................................193  
11.1.6.2. Low Flow .............................................................................................................193  
11.1.6.3. High Flow ............................................................................................................ 194  
11.1.6.4. Displayed Flow = “XXXX” ....................................................................................... 194  
11.1.6.5. Actual Flow Does Not Match Displayed Flow .............................................................. 194  
11.1.6.6. Sample Pump....................................................................................................... 194  
11.1.7. Poor or Stopped Flow of Purge Gas ................................................................................ 194  
11.2. Calibration Problems .......................................................................................................... 195  
11.2.1. Miss-Calibrated ........................................................................................................... 195  
04584 Rev A1  
v
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
GFC7000E Documentation  
11.2.2. Non-Repeatable Zero and Span ..................................................................................... 195  
11.2.3. Inability to Span – No SPAN Key.................................................................................... 196  
11.2.4. Inability to Zero – No ZERO Key.................................................................................... 196  
11.3. Other Performance Problems............................................................................................... 196  
11.3.1. Temperature Problems................................................................................................. 197  
11.3.1.1. Box or Sample Temperature ................................................................................... 197  
11.3.1.2. Bench Temperature............................................................................................... 197  
11.3.1.3. GFC Wheel Temperature ........................................................................................ 198  
11.3.1.4. IR Photo-Detector TEC Temperature ........................................................................ 198  
11.3.2. Excessive Noise........................................................................................................... 198  
11.4. Subsystem Checkout.......................................................................................................... 199  
11.4.1. AC Mains Configuration ................................................................................................ 199  
11.4.2. DC Power Supply......................................................................................................... 200  
11.4.3. I2C Bus...................................................................................................................... 200  
11.4.4. Keyboard/Display Interface........................................................................................... 201  
11.4.5. Relay Board................................................................................................................ 201  
11.4.6. Sensor Assembly......................................................................................................... 202  
11.4.6.1. Sync/Demodulator Assembly .................................................................................. 202  
11.4.6.2. Opto Pickup Assembly ........................................................................................... 202  
11.4.6.3. GFC Wheel Drive................................................................................................... 203  
11.4.6.4. IR Source ............................................................................................................ 203  
11.4.6.5. Pressure/Flow Sensor Assembly .............................................................................. 203  
11.4.7. Motherboard............................................................................................................... 204  
11.4.7.1. A/D Functions....................................................................................................... 204  
11.4.7.2. Analog Outputs: Voltage ........................................................................................ 204  
11.4.7.3. Analog Outputs: Current Loop................................................................................. 204  
11.4.7.4. Status Outputs ..................................................................................................... 205  
11.4.7.5. Control Inputs – Remote Zero, Span........................................................................ 205  
11.4.8. CPU........................................................................................................................... 206  
11.4.9. RS-232 Communications .............................................................................................. 206  
11.4.9.1. General RS-232 Troubleshooting............................................................................. 206  
11.4.9.2. Troubleshooting Analyzer/Modem or Terminal Operation............................................. 207  
11.5. Repair Procedures.............................................................................................................. 207  
11.5.1. Repairing Sample Flow Control Assembly........................................................................ 207  
11.5.2. Removing/Replacing the GFC Wheel............................................................................... 209  
11.5.3. Disk-On-Chip Replacement Procedure............................................................................. 210  
12. A PRIMER ON ELECTRO-STATIC DISCHARGE.......................................................................... 213  
12.1. How Static Charges are Created........................................................................................... 213  
12.2. How Electro-Static Charges Cause Damage ........................................................................... 214  
12.3. Common Myths About ESD Damage ..................................................................................... 215  
12.4. Basic Principles of Static Control .......................................................................................... 216  
12.4.1. General Rules ............................................................................................................. 216  
12.4.2. Basic anti-ESD Procedures for Analyzer Repair and Maintenance ........................................ 218  
12.4.2.1. Working at the Instrument Rack.............................................................................. 218  
12.4.2.2. Working at a Anti-ESD Work Bench.......................................................................... 218  
12.4.2.3. Transferring Components from Rack To Bench and Back............................................. 219  
12.4.2.4. Opening Shipments from and Packing Components for Return to Teledyne Instruments  
Customer Service. ..............................................................................................................219  
04584 Rev A1  
vi  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
GFC7000E Documentation  
LIST OF APPENDICES  
APPENDIX A - VERSION SPECIFIC SOFTWARE DOCUMENTATION  
APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0  
APPENDIX A-2: Setup Variables For Serial I/O, Revision E.0  
APPENDIX A-3: Warnings and Test Functions, Revision E.0  
APPENDIX A-4: MGFC7000E Signal I/O Definitions, Revision E.0  
APPENDIX A-5: MGFC7000E iDAS Functions, Revision E.0  
APPENDIX A-6: Terminal Command Designators, Revision E.0  
APPENDIX B - GFC7000E SPARE PARTS LIST  
APPENDIX C - REPAIR QUESTIONNAIRE - MGFC7000E  
APPENDIX D - ELECTRONIC SCHEMATICS  
LIST OF FIGURES  
Figure 3-1:  
Figure 3-2:  
Figure 3-3:  
Figure 3-4:  
Figure 3-5:  
Figure 3-6  
Removal of Shipping Screws.......................................................................... 10  
Rear Panel Layout........................................................................................ 12  
Pneumatic Connections–Basic Configuration–Using Bottled Span Gas .................. 16  
Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator ........... 17  
Pneumatic Connections–MGFC7000E with Zero/Span/Shutoff Valves (OPT 50)...... 19  
Pneumatic Connections–MGFC7000E with Zero/Span/Shutoff Valves and External  
Zero Air Scrubber (OPT 51).......................................................................... 19  
Pneumatic Connections–MGFC7000E with Zero/Span Valves (OPT 52)................. 20  
Pneumatic Connections–MGFC7000E with Zero/Span Valves with External Zero air  
Scrubber (OPT 53)...................................................................................... 20  
Example of Pneumatic Set up for Multipoint Calibration of M360 ......................... 21  
Figure 3-7:  
Figure 3-8:  
Figure 3-9:  
Figure 3-10: Front Panel Layout ....................................................................................... 23  
Figure 3-11: Assembly Layout.......................................................................................... 30  
Figure 3-12: Optical Bench Layout.................................................................................... 31  
Figure 3-13: Internal Pneumatic Flow – Basic Configuration ................................................. 31  
Figure 5-1:  
Figure 5-2:  
Figure 5-3:  
Figure 5-4:  
Figure 6-1:  
Figure 6-2  
Figure 6-3  
Figure 6-4:  
Figure 6-5:  
Figure 6-6:  
Figure 6-7:  
Figure 6-8:  
Figure 6-9:  
Current Loop Option Installed on the Motherboard............................................ 37  
Internal Pneumatic Flow – Zero/Span/Shutoff Valves OPT 50 & 51...................... 39  
Internal Pneumatic Flow – Zero/Span OPT 52 & 53........................................... 41  
GFC7000E Ethernet Card and rear panel With Ethernet Installed......................... 42  
Front Panel Display ...................................................................................... 45  
Viewing MGFC7000E TEST Functions............................................................... 48  
Viewing and Clearing MGFC7000E WARNING Messages...................................... 50  
Analog Output Connector Pin Out................................................................... 56  
Setup for Calibrating Analog Voltage Outputs................................................... 74  
Setup for Calibrating Current Outputs............................................................. 75  
Default iDAS Channels Setup....................................................................... 101  
APICOM user interface for configuring the iDAS.............................................. 113  
iDAS Configuration Through a Terminal Emulation Program.............................. 114  
Figure 6-10: Status Output Connector............................................................................. 115  
Figure 6-11: Control Inputs ........................................................................................... 117  
Figure 6-12: APICOM Remote Control Program Interface.................................................... 123  
Figure 7-1:  
Figure 7-2:  
Figure 7-3:  
Figure 7-4  
Pneumatic Connections–Basic Configuration–Using Bottled Span Gas ................ 130  
Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator ......... 131  
Pneumatic Connections–MGFC7000E with Zero/Span/Shutoff Valves (OPT 50).... 134  
Pneumatic Connections–MGFC7000E with Zero/Span/Shutoff Valves and External  
Zero Air Scrubber (OPT 51)........................................................................ 134  
Pneumatic Connections–MGFC7000E with Zero/Span Valves (OPT 52)............... 135  
Figure 7-5:  
04584 Rev A1  
vii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
GFC7000E Documentation  
Figure 7-6:  
Figure 9-1:  
Pneumatic Connections–MGFC7000E with Zero/Span Valves with External Zero air  
Scrubber (OPT 53).................................................................................... 135  
Sample Particulate Filter Assembly ............................................................... 149  
Figure 10-1: Measurement Fundamentals ........................................................................ 154  
Figure 10-2: GFC Wheel................................................................................................ 155  
Figure 10-3: Measurement Fundamentals with GFC Wheel ................................................. 155  
Figure 10-4: Affect of CO2 in the Sample on CO2 MEAS & CO2 REF...................................... 156  
Figure 10-5: Effects of Interfering Gas on CO2 MEAS & CO2 REF......................................... 157  
Figure 10-6: Chopped IR Signal...................................................................................... 157  
Figure 10-7: Internal Pneumatic Flow – Basic Configuration ............................................... 159  
Figure 10-8: Flow Control Assembly & Critical Flow Orifice ................................................. 160  
Figure 10-9: GFC7000E Electronic Block Diagram ............................................................. 162  
Figure 10-10: GFC Light Mask.......................................................................................... 164  
Figure 10-11: Segment Sensor and M/R Sensor Output ....................................................... 165  
Figure 10-12: GFC7000E Sync / Demod Block Diagram ....................................................... 166  
Figure 10-13: Sample & Hold Timing................................................................................. 167  
Figure 10-14: Location of relay board Status LED’s ............................................................. 170  
Figure 10-15: Power Distribution Block Diagram................................................................. 174  
Figure 10-16: Interface Block Diagram.............................................................................. 175  
Figure 10-17: MGFC7000E Front Panel Layout.................................................................... 175  
Figure 10-18: Keyboard and Display Interface Block Diagram............................................... 177  
Figure 10-19: Basic Software Operation ............................................................................ 179  
Figure 11-1: Viewing and Clearing Warning Messages........................................................ 184  
Figure 11-2: Example of Signal I/O Function .................................................................... 189  
Figure 11-3: CPU Status Indicator .................................................................................. 190  
Figure 11-4: Sync/Demod Board Status LED Locations ...................................................... 191  
Figure 11-5: Relay Board Status LEDs............................................................................. 191  
Figure 11-6: Critical Flow Restrictor Assembly Disassembly................................................ 208  
Figure 11-7: Opening the GFC Wheel Housing .................................................................. 209  
Figure 11-8: Removing the GFC Wheel............................................................................. 210  
Figure 12-1: Triboelectric Charging................................................................................. 213  
Figure 12-2: Basic anti-ESD Work Station........................................................................ 216  
Figure A-1: Basic Sample Display Menu.......................................................................... 223  
Figure A-2: Sample Display Menu - Units with Z/S Valve or IZS Option installed................... 224  
Figure A-3: Primary Setup Menu (Except iDAS) ............................................................... 225  
Figure A-4: Primary Setup Menu (iDAS) ......................................................................... 226  
Figure A-5: Secondary Setup Menu (COMM & VARS) ........................................................ 227  
Figure A-6: Secondary Setup Menu (COMM Menu with Ethernet Card)................................. 228  
Figure A-7: Secondary Setup Menu (COMM Menu with HESSEN) ........................................ 229  
Figure A-8: Secondary Setup Menu (DIAG) ..................................................................... 230  
LIST OF TABLES  
Table 2-1:  
Table 3-1:  
Table 3-2:  
Table 3-3:  
Table 3-4:  
Table 3-5:  
Table 3-6:  
Table 5-1:  
Table 5-2:  
Table 6-1:  
Model GFC7000E Basic Unit Specifications ..........................................................5  
GFC7000E Analog Output Pin Outs .................................................................. 13  
GFC7000E Status Output Pin Outs................................................................... 14  
GFC7000E Control Input Pin Outs.................................................................... 15  
Model GFC7000E Rear Panel Pneumatic Connections.......................................... 16  
Front Panel Display During System Warm-Up.................................................... 23  
Possible Warning Messages at Start-Up............................................................ 24  
Zero/Span Valve Operating States for Options 50 & 51....................................... 38  
Zero/Span Valve Operating States for Options 52 & 53....................................... 40  
Analyzer Operating modes.............................................................................. 46  
04584 Rev A1  
viii  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
GFC7000E Documentation  
Table 6-2:  
Table 6-3:  
Table 6-4:  
Table 6-5:  
Table 6-6:  
Table 6-7:  
Table 6-8:  
Table 6-9:  
Test Functions Defined................................................................................... 46  
List of Warning Messages............................................................................... 49  
Primary Setup Mode Features and Functions ..................................................... 52  
Secondary Setup Mode Features and Functions ................................................. 52  
Variable Names (VARS) Revision B.3 ............................................................... 64  
GFC7000E Diagnostic (DIAG) Functions............................................................ 66  
DIAG - Analog I/O Functions........................................................................... 69  
Analog Output Voltage Ranges........................................................................ 69  
Table 6-10: Analog Output Current Loop Range.................................................................. 70  
Table 6-11: Analog Output Pin Assignments....................................................................... 70  
Table 6-12: Voltage Tolerances for Analog Output Calibration............................................... 73  
Table 6-13: Current Loop Output Calibration with Resistor ................................................... 76  
Table 6-14: Test Parameters Available for Analog Output A3 ................................................ 81  
Table 6-15: COM1 and COM2 DB-9 Pin Assignments............................................................ 85  
Table 6-16: COMM Port Communication modes................................................................... 86  
Table 6-17: Ethernet Status Indicators.............................................................................. 89  
Table 6-18: LAN/Internet Configuration Properties.............................................................. 90  
Table 6-19: Internet Configuration Keypad Functions .......................................................... 94  
Table 6-20: CO2 Concentration Alarm Default Settings......................................................... 95  
Table 6-21: Front Panel LED Status Indicators for iDAS ....................................................... 96  
Table 6-22: iDAS Data Channel Properties ......................................................................... 97  
Table 6-23: iDAS Data Parameter Functions....................................................................... 98  
Table 6-24: Status Output Pin Assignments ..................................................................... 116  
Table 6-25: Control Input Pin Assignments ...................................................................... 116  
Table 6-26: Terminal Mode Software Commands .............................................................. 118  
Table 6-27: Command Types ......................................................................................... 119  
Table 6-28: Serial Interface Documents........................................................................... 123  
Table 6-29: RS-232 Communication Parameters for Hessen Protocol ................................... 124  
Table 6-30: Teledyne Instruments Hessen Protocol Response Modes ................................... 126  
Table 6-31: Default Hessen Status Bit Assignments........................................................... 127  
Table 7-1:  
Table 7-2:  
Table 7-3:  
Table 9-1:  
Table 9-2:  
Table 9-3:  
AUTOCAL Modes ......................................................................................... 139  
AutoCal ATTRIBUTE Setup Parameters ........................................................... 139  
Calibration Data Quality Evaluation................................................................ 142  
GFC7000E Maintenance Schedule.................................................................. 146  
GFC7000E Test Function Record.................................................................... 147  
Predictive uses for Test Functions.................................................................. 148  
Table 10-1: Sync/Demod Status LED Activity ................................................................... 168  
Table 10-2: Relay Board Status LED’s ............................................................................. 169  
Table 10-3: Front Panel Status LED’s .............................................................................. 176  
Table 11-1: Warning Messages - Indicated Failures........................................................... 185  
Table 11-2: Test Functions - Indicated Failures................................................................. 187  
Table 11-3: Sync/Demod Board Status Failure Indications ................................................. 190  
Table 11-4: I2C Status LED Failure Indications ................................................................. 191  
Table 11-5: Relay Board Status LED Failure Indications ..................................................... 192  
Table 11-6: DC Power Test Point and Wiring Color Codes................................................... 200  
Table 11-7: DC Power Supply Acceptable Levels ............................................................... 200  
Table 11-8: Relay Board Control Devices ......................................................................... 201  
Table 11-9: Opto Pickup Board Nominal Output Frequencies............................................... 202  
Table 11-10: Analog Output Test Function - Nominal Values Voltage Outputs.......................... 204  
Table 11-11: Analog Output Test Function - Nominal Values Current Outputs ......................... 205  
Table 11-12: Status Outputs Check .................................................................................. 205  
Table 12-1: Static Generation Voltages for Typical Activities............................................... 214  
Table 12-2: Sensitivity of Electronic Devices to Damage by ESD ......................................... 214  
Table A-1:  
GFC7000E Setup Variables, Revision E.0 ........................................................ 231  
04584 Rev A1  
ix  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
GFC7000E Documentation  
Table A-2:  
Table A-3:  
Table A-4:  
Table A-5:  
Table A-6:  
Table A-7:  
Table D-1:  
GFC7000E Warning Messages, Revision E.0 .................................................... 238  
GFC7000E Test Functions, Revision E.0.......................................................... 239  
GFC7000E Signal I/O Definitions, Revision E.0 ................................................ 240  
GFC7000E DAS Trigger Events, Revision E.0................................................... 243  
GFC7000E iDAS Functions, Revision E.0......................................................... 244  
Terminal Command Designators, Revision E.0................................................. 245  
List of Included Electronic Schematics............................................................ 251  
User Notes  
04584 Rev A1  
x
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
MGFC7000E Documentation  
1. MGFC7000E DOCUMENTATION  
Thank you for purchasing the Model GFC7000E Gas Filter Correlation Carbon Dioxide Analyzer!  
The documentation for this instrument is available in several different formats:  
Printed format, part number 045840100  
Electronic format on a CD-ROM, part number 045840200  
The electronic manual is in Adobe® Systems Inc. “Portable Document Format”. The Adobe®  
Acrobat Reader® software, which is necessary to view these files, can be downloaded for free from  
the internet at http://www.adobe.com/.  
The electronic version of the manual has many advantages:  
Keyword and phrase search feature  
Figures, tables and internet addresses are linked so that clicking on the item will display  
the associated feature or open the website.  
A list of chapters and sections as well as thumbnails of each page are displayed to the left  
of the text.  
Entries in the table of contents are linked to the corresponding locations in the manual.  
Ability to print sections (or all) of the manual  
Additional documentation for the Model GFC7000E CO2 Analyzer is available from Teledyne  
Analytical Instruments’ website at http://www.teledyne-ai.com/manuals/  
APICOM software manual, part number 03945  
Multi-drop manual, part number 01842  
DAS Manual, part number 02837.  
1.1. Using This Manual  
This manual has the following data structures:  
1.0 Table of Contents:  
Outlines the contents of the manual in the order the information is presented. This is a good  
overview of the topics covered in the manual. There is also a list of tables, a list of figures and a  
list of appendices. In the electronic version of the manual, clicking on a any of these table entries  
automatically views that section.  
04584 Rev A1  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
MGFC7000E Documentation  
2.0 Specifications and Warranty  
This section contains a list of the analyzer’s performance specifications, a description of the  
conditions and configuration under which EPA equivalency was approved and Teledyne  
Instruments Incorporated’s warranty statement.  
3.0 Getting Started:  
A concise set of instructions for setting up, installing and running your analyzer for the first time.  
4.0 FAQ:  
Answers to the most frequently asked questions about operating the analyzer.  
5.0 Optional Hardware & Software  
A description of optional equipment to add functionality to your analyzer.  
6.0 Operation Instructions  
This section includes step by step instructions for operating the analyzer and using its various  
features and functions.  
7.0 Calibration Procedures  
General information and step by step instructions for calibrating your analyzer.  
8.0 EPA Protocol Calibration  
Because CO2 is not declared a criteria air pollutant by the US EPA, EPA equivalency is not required  
for this type of analyzer. Therefore no special calibration methods are needed to satisfy EPA  
requirements.  
9.0 Instrument Maintenance  
Description of certain preventative maintenance procedures that should be regularly performed on  
you instrument to keep it in good operating condition. This section also includes information on  
using the iDAS to record diagnostic functions useful in predicting possible component failures  
before they happen.  
10.0 Theory of Operation  
An in-depth look at the various principals by which your analyzer operates as well as a description  
of how the various electronic, mechanical and pneumatic components of the instrument work and  
interact with each other. A close reading of this section is invaluable for understanding the  
instrument’s operation.  
11.0 Troubleshooting Section:  
This section includes pointers and instructions for diagnosing problems with the instrument, such  
as excessive noise or drift, as well as instructions on performing repairs of the instrument’s major  
subsystems.  
04584 Rev A1  
2
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
MGFC7000E Documentation  
Appendices:  
For easier access and better updating, some information has been separated out of the manual  
and placed in a series of appendices at the end of this manual. These include: software menu  
trees, warning messages, definitions of iDAS & serial I/O variables, spare parts list, repair  
questionnaire, interconnect listing and drawings, and electronic schematics.  
NOTE  
Throughout this manual, words printed in capital, bold letters, such as SETUP or ENTR  
represent messages as they appear on the analyzer’s front panel display.  
NOTE  
The flowcharts in this manual contain typical representations of the analyzer’s display  
during the various operations being described. These representations are not intended  
to be exact and may differ slightly from the actual display of your instrument.  
User Notes  
04584 Rev A1  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Specifications, Approvals and Warranty  
2. SPECIFICATIONS, APPROVALS AND  
WARRANTY  
2.1. Specifications  
Table 2-1:  
Model GFC7000E Basic Unit Specifications  
Min/Max Range  
In 1ppb increments from 50ppb to 2 000ppm, dual ranges or auto  
ranging  
ppb, ppm, µg/m3, mg/m3, %(user selectable)  
(Physical Analog Output)  
Measurement Units  
Zero Noise  
< 0.1 ppm (RMS)  
Span Noise  
< 1% of reading (RMS)  
< 0.2 ppm1  
<0.25 ppm1  
<0.5 ppm1  
1% of reading above 50 PPM1  
Lower Detectable Limit1  
Zero Drift (24 hours)  
Zero Drift (7 days)  
Span Drift (7 Days)  
Linearity  
1% of full scale  
Precision  
0.5% of reading  
Temperature Coefficient  
Voltage Coefficient  
Lag Time  
< 0.1% of Full Scale per oC  
< 0.05% of Full Scale per V  
10 sec  
Rise/Fall Time  
95% in <60 sec  
Sample Flow Rate  
Temperature Range  
Humidity Range  
Dimensions H x W x D  
Weight, Analyzer  
AC Power Rating  
800cm3/min. ±10%  
5-40oC  
0 - 95% RH, non-condensing  
7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)  
38 lbs. (17 kg); add 1 lbs (0.5 kg) for IZS  
100 V, 50/60 Hz (3.25A);  
115 V, 60 Hz (3.0 A);  
220 – 240 V, 50/60 Hz (2.5 A)  
Environmental  
Installation category (over-voltage category) II; Pollution degree 2  
Three (3) Outputs  
Analog Outputs  
Analog Output Ranges  
0.1V, 1 V, 5 V, 10 V, 2-20 or 4-20 mA isolated current loop.  
All Ranges with 5% Under/Over Range  
Analog Output Resolution  
Status Outputs  
Control Inputs  
1 part in 4096 of selected full-scale voltage  
8 Status outputs - opto-isolated; including 2 alarm outputs  
6 Control Inputs, 3 defined, 3 spare  
Serial I/O  
One (1) RS-232; One (1) RS-485  
Baud Rate : 300 – 115200: Optional Ethernet Interface  
Certifications  
CE: EN61010-1:90 + A1:92 + A2:95, EN61326 - Class A  
1
At constant temperature and voltage.  
04584 Rev A1  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Specifications, Approvals and Warranty  
2.2. CE Mark Compliance  
Emissions Compliance  
The Teledyne Instruments Model GFC7000E Gas Filter Correlation CO2 Analyzer was tested and  
found to be fully compliant with:  
EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section 15.107 Class A, ICES-003  
Class A (ANSI C63.4 1992) & AS/NZS 3548 (w/A1 & A2; 97) Class A.  
Tested on 11-29-2001 at CKC Laboratories, Inc., Report Number CE01-249.  
Safety Compliance  
The Teledyne Instruments Model GFC7000E Gas Filter Correlation CO2 Analyzer was tested and  
found to be fully compliant with:  
IEC 61010-1:90 + A1:92 + A2:95,  
Tested on 02-06-2002 at Nemko, Report Number 2002-012219.  
2.3. Warranty  
Warranty Policy (02024)  
Prior to shipment, Teledyne Instruments Incorporated equipment is thoroughly inspected and  
tested. Should equipment failure occur, Teledyne Instruments Incorporated assures its customers  
that prompt service and support will be available.  
Coverage  
After the warranty period and throughout the equipment lifetime, Teledyne Instruments  
Incorporated stands ready to provide on-site or in-plant service at reasonable rates similar to  
those of other manufacturers in the industry. All maintenance and the first level of field  
troubleshooting is to be performed by the customer.  
Non-API Manufactured Equipment  
Equipment provided but not manufactured by Teledyne Instruments Incorporated is warranted  
and will be repaired to the extent and according to the current terms and conditions of the  
respective equipment manufacturers warranty.  
General  
Teledyne Instruments Incorporated warrants each product manufactured by Teledyne Instruments  
Incorporated to be free from defects in material and workmanship under normal use and service  
for a period of one year from the date of delivery. All replacement parts and repairs are warranted  
for 90 days after the purchase.  
If a product fails to conform to its specifications within the warranty period, Teledyne Instruments  
Incorporated shall correct such defect by, in Teledyne Instruments Incorporated’s discretion,  
repairing or replacing such defective product or refunding the purchase price of such product.  
04584 Rev A1  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Specifications, Approvals and Warranty  
The warranties set forth in this section shall be of no force or effect with respect to any product:  
(i) that has been altered or subjected to misuse, negligence or accident, or (ii) that has been used  
in any manner other than in accordance with the instruction provided by TELEDYNE ANALYTICAL  
INSTRUMENTS or (iii) not properly maintained.  
THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES THEREFORE ARE  
EXCLUSIVE AND IN LIEU OF ANY IMPLIED WARRANTIES OF MERCHANTABILITY,  
FITNESS FOR PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER  
EXPRESSED OR IMPLIED. THE REMEDIES SET FORTH IN THIS SECTION ARE THE  
EXCLUSIVE REMEDIES FOR BREACH OF ANY WARRANTY CONTAINED HEREIN. Teledyne  
Instruments Incorporated SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR  
CONSEQUENTIAL DAMAGES ARISING OUT OF OR RELATED TO THIS AGREEMENT OF  
TELEDYNE INSTRUMENTS INCORPORATED’S PERFORMANCE HEREUNDER, WHETHER  
FOR BREACH OF WARRANTY OR OTHERWISE.  
Terms and Conditions  
All units or components returned to Teledyne Amnalytical Instruments should be properly packed  
for handling and returned freight prepaid to the nearest designated Service Center. After the  
repair, the equipment will be returned, freight prepaid.  
User Notes  
04584 Rev A1  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
3. GETTING STARTED  
3.1. Unpacking and Initial Set Up  
CAUTION  
To avoid personal injury, always use two persons to lift and carry the  
Model GFC7000E.  
1. Verify that there is no apparent external shipping damage. If damage has occurred, please  
advise the shipper first, then Teledyne Analytical Instruments.  
2. Included with your analyzer is a printed record of the final performance characterization  
performed on your instrument at the factory. This record, titled Final Test and Validation Data  
Sheet (part number 04307) is an important quality assurance and calibration record for this  
instrument. It should be placed in the quality records file for this instrument.  
3. Carefully remove the top cover of the analyzer and check for internal shipping damage.  
Remove the set screw located in the top, center of the rear panel.  
Remove the four screws fastening the top cover to the unit (two per side).  
Lift the cover straight up. Do not slide backwards.  
NOTE  
Some versions of the GFC7000E CO2 Analyzer may have a spring loaded fastener at the  
top center of the rear panel and as many as eight screws (four per side) fastening the  
top cover to the chassis.  
NOTE  
Static sensitive parts are present on PCA (Printed Circuit Assemblies). Before touching  
PCAs, touch a bare metal part of the chassis to discharge any electrostatic potentials or  
connect a grounding strap to your wrist.  
CAUTION  
Never disconnect PCAs, wiring harnesses or electronic subassemblies  
while under power.  
4. Inspect the interior of the instrument to make sure all circuit boards and other components  
are in good shape and properly seated.  
04584 Rev A1  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
5. Check the connectors of the various internal wiring harnesses and pneumatic hoses to make  
sure they are firmly and properly seated.  
6. Verify that all of the optional hardware ordered with the unit has been installed. These are  
listed on the paperwork accompanying the analyzer.  
7. Once you have determined that no shipping damage exists, and the unit includes all expected  
hardware options, remove all red colored shipping screws from the bottom of the chassis as  
shown in Figure 3-1.  
Shipping  
Screws  
Figure 3-1:  
Removal of Shipping Screws  
NOTE  
Save these shipping screws and re-install them whenever the unit is shipped to another  
location.  
04584 Rev A1  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
8. VENTILATION CLEARANCE: Whether the analyzer is set up on a bench or installed into an  
instrument rack, be sure to leave sufficient ventilation clearance.  
AREA  
MINIMUM REQUIRED CLEARANCE  
Back of the instrument  
4 in.  
1 in.  
Sides of the instrument  
Above and below the instrument  
1 in.  
Various rack mount kits are available for this analyzer. See Section 5.1 of this manual  
for more information.  
3.1.1. Electrical Connections  
CAUTION  
Check the voltage and frequency label on the rear panel of the  
instrument (See Figure 3-2) for compatibility with the local  
power before plugging the MGFC7000E into line power.  
Do not plug in the power cord if the voltage or  
frequency is incorrect.  
CAUTION  
Power connection must have functioning ground connection.  
Do not defeat the ground wire on power plug.  
CAUTION  
Turn off analyzer power before disconnecting or  
connecting electrical subassemblies.  
Do not operate with cover off.  
04584 Rev A1  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
Figure 3-2:  
Rear Panel Layout  
9. Attach a strip chart recorder and/or data-logger to the appropriate analog output connections  
on the rear panel of the analyzer.  
ANALOG  
A1  
A2  
3 4  
A3  
A4  
7
1
2
5
6
8
When the instrument is in its default configuration, the A1 and A2 channels output a signal  
that is proportional to the CO2 concentration of the sample gas. Either can be used for  
connecting the analog output signal to a chart recorder or for interfacing with a datalogger.  
The third analog output, labeled A3 is special. It can be set by the user (see Section  
6.9.9) to output any one of the parameters accessible through the <TST TST> keys of the  
units sample display.  
The standard configuration for these outputs is mVDC. An optional current loop output is  
available for each.  
Output A4 is not used on the Model 306E.  
Pin-outs for the analog output connector at the rear panel of the instrument are:  
04584 Rev A1  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Table 3-1: GFC7000E Analog Output Pin Outs  
Getting Started  
Pin  
Analog Output  
VDC Signal  
mADC Signal  
1
2
3
4
5
6
7
8
V Out  
I Out +  
I Out -  
A1  
Ground  
V Out  
I Out +  
A2  
A3  
Ground  
V Out  
I Out -  
I Out +  
Ground  
V Out  
I Out -  
Not Available  
Not Available  
A4 (Spare)  
Ground  
The default analog output voltage setting of the GFC7000E CO2 Analyzer is 0 – 5 VDC  
with a range of 0 – 500 ppm.  
TO change these settings, see Sections 6.9.4 and 6.7 respectively.  
10.If you wish utilize the analyzer’s status outputs to interface with a device that accepts logic-  
level digital inputs, such as programmable logic controllers (PLC’s) they are accessed via a 12-  
pin connector on the analyzer’s rear panel labeled STATUS.  
STATUS  
1
2
3
4
5
6
7
8
D
+
NOTE  
Most PLC’s have internal provisions for limiting the current that the input will draw from  
an external device. When connecting to a unit that does not have this feature, an  
external dropping resistor must be used to limit the current through the transistor  
output to less than 50mA. At 50mA, the transistor will drop approximately 1.2V from its  
collector to emitter.  
04584 Rev A1  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
The pin assignments for the status outputs can be found in the table below:  
Table 3-2:  
GFC7000E Status Output Pin Outs  
Status  
Definition  
Output #  
Condition  
1
2
SYSTEM OK  
On if no faults are present.  
On if CO2 concentration measurement is valid.  
If the CO2 concentration measurement is invalid, this bit is OFF.  
CONC VALID  
3
4
5
6
HIGH RANGE  
ZERO CAL  
On if unit is in high range of DUAL or AUTO range modes.  
On whenever the instruments ZERO point is being calibrated.  
On whenever the instruments SPAN point is being calibrated.  
On whenever the instrument is in DIAGNOSTIC mode.  
SPAN CAL  
DIAG MODE  
On whenever the measured CO2 concentration is above the set  
point for ALM1  
7
8
ALARM1  
ALARM2  
On whenever the measured CO2 concentration is above the set  
point for ALM2  
D
+
EMITTER BUSS  
DC POWER  
The emitters of the transistors on pins 1-8 are bussed together.  
+ 5 VDC  
Digital Ground  
The ground level from the analyzer’s internal DC power supplies.  
11.If you wish to use the analyzer’s to remotely activate the zero and span calibration modes,  
several digital control inputs are provided via a 10-pin connector labeled CONTROL IN on the  
analyzer’s rear panel. Two methods for energizing the inputs are provided below; the first  
using the internal +5V available on the CONTROL IN connector and the second, if an external,  
isolated supply is employed.  
CONTROL IN  
CONTROL IN  
A
B
C
D
E
F
U
+
A
B
C
D
E
F
U
+
Z
E
R
S
P
A
N
Z
E
R
O
S
P
A
N
O
5 VDC Power  
Supply  
+
-
External Power Connections  
Local Power Connections  
04584 Rev A1  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
The pin assignments for the digital control inputs can be found in the table below:  
Table 3-3:  
GFC7000E Control Input Pin Outs  
Status  
Definition  
Input #  
On Condition  
REMOTE ZERO  
CAL  
The Analyzer is placed in Zero Calibration mode. The mode  
field of the display will read ZERO CAL R.  
A
B
REMOTE  
SPAN CAL  
The Analyzer is placed in Span Calibration mode. The mode  
field of the display will read SPAN CAL R.  
SPARE  
SPARE  
C
D
E
F
SPARE  
SPARE  
Digital Ground  
May be connected to the ground of the datalogger/recorder.  
Input pin for +5 VDC required to activate pins A – F. This  
can be from an external source or from the “+” pin of the  
instruments STATUS connector.  
Pullup supply  
for inputs  
U
Internal +5V  
Supply  
Internal source of +5V which can be used to actuate control  
inputs when connected to the U pin.  
+
12.If you wish to utilize either of the analyzer’s two serial interface COMM ports, refer to Section  
6.10 of this manual for instructions on their configuration and usage.  
13.If your unit has a Teledyne Instruments Ethernet card (Option 63), plug one end into the 7’  
CAT5 cable supplied with the option into the appropriate place on the back of the analyzer  
(see Figure 5-4 in Section 5.5.3) and the other end into any nearby Ethernet access port.  
3.1.2. Pneumatic Connections:  
3.1.2.1. Basic Pneumatic Connections  
Figures 3-3 and 3-4 illustrate the most common configurations for gas supply and exhaust lines to  
the Model GFC7000E Analyzer. Figure 3-13 illustrates the internal gas flow of the instrument in  
its basic configuration.  
Please refer to Figure 3-2 for pneumatic connections at the rear panel and Table 3-4 for  
nomenclature.  
NOTE  
Sample and calibration gases should only come into contact with PTFE (Teflon), FEP,  
glass, stainless steel or brass. CAUTION  
04584 Rev A1  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
CAUTION  
In order to prevent dust from getting into the gas flow channels of your analyzer, it was  
shipped with small plugs inserted into each of the pneumatic fittings on the back panel.  
Make sure that all of these dust plugs are removed before attaching  
exhaust and supply gas lines.  
Table 3-4:  
Rear Panel Label  
SAMPLE  
Model GFC7000E Rear Panel Pneumatic Connections  
Function  
Connect a gas line from the source of sample gas here.  
Calibration gasses are also inlet here on units without  
zero/span/shutoff valve or IZS options installed.  
EXHAUST  
Connect an exhaust gas line of not more than 10 meters long here.  
On units with zero/span/shutoff valve options installed, connect a  
gas line to the source of calibrated span gas here.  
PRESSURE SPAN  
Span gas vent outlet for units with zero/span/shutoff valve options  
installed.  
Connect an exhaust gas line of not more than 10 meters long here.  
VENT SPAN  
IZS  
Internal zero air scrubber.  
on units with zero/span/shutoff valve options installed but NO  
internal zero air scrubber, attach a gas line to the source of zero air  
here.  
This inlet supplies purge air to the GFC wheel housing (see section  
10.2.2)  
PURGE IN  
Connect a source of dried air that has been scrubbed of CO2.  
Calibrated CO2  
gas at desired  
span gas  
VENT  
Source of  
SAMPLE Gas  
Removed  
during  
Calibration  
concentration  
Needle  
valve to  
control  
flow  
Indicating  
soda-lime  
Sample  
Exhaust  
MODEL  
GFC7000E  
Valve  
Vent Span  
Pressure Span  
MODEL 701  
Zero Air  
IZS  
Generator  
Purge In  
Figure 3-3:  
Pneumatic Connections–Basic Configuration–Using Bottled Span Gas  
04584 Rev A1  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
Source of  
Calibrated  
CO2 Gas  
SAMPLE Gas  
Removed  
during  
MODEL 700  
Gas Dilution  
Calibrator  
Calibration  
Indicating  
soda-lime  
VENT  
Sample  
Exhaust  
Vent Span  
Pressure Span  
IZS  
MODEL  
GFC7000E  
MODEL 701  
Zero Air Generator  
Purge In  
Figure 3-4:  
Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator  
1. Attach a sample inlet line to the sample inlet port. The SAMPLE input line should not be more  
than 2 meters long.  
NOTE  
Ideally, the pressure of the sample gas should be at ambient pressure (0  
psig). Maximum pressure of sample gas should not exceed 1.5 in-Hg over  
ambient.  
In applications where the sample gas is received from a pressurized  
manifold, a vent must be placed as shown to equalize the sample gas with  
ambient atmospheric pressure before it enters the analyzer.  
This vent line must be:  
At least 0.2m long  
No more than 2m long and vented outside the shelter or immediate  
area surrounding the instrument.  
04584 Rev A1  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
2. Attach sources of zero air and span gas (see Figures 3-3 through 3-8 inclusive).  
Span Gas is a gas specifically mixed to match the chemical composition of the type of gas  
being measured at near full scale of the desired measurement range.  
In the case of CO2 measurements made with the Teledyne Analytical Instruments Model  
GFC7000E Analyzer it is recommended that you use a gas calibrated to have a CO2 content  
equaling 80% of the range of compositions being measured.  
EXAMPLE: If the application is to measure between 0 ppm and 500 ppm, an appropriate  
Span Gas would be 400 ppm. If the application is to measure between 0 ppm and 100  
ppm, an appropriate Span Gas would be 80 ppm.  
Span Gas can be purchased in pressurized canisters or created using Dynamic Dilution  
Calibrator such as the Teledyne Analytical Instruments Model 700 and a source of dried  
air scrubbed of CO2 such as a Teledyne Analytical Instruments Model 701 Zero Air  
Generator in combination with a canister of indicating soda-lime.  
Zero Air is similar in chemical composition to the earths atmosphere but scrubbed of all  
components that might affect the analyzer’s readings.  
In the case of CO2 measurements this means CO2 less than 0.1 ppm of CO2 and Water  
Vapor. Zero Air can be purchased in pressurized canisters or created using a Teledyne  
Instruments Model 701 Zero Air Generator in combination with a canister of indicating  
soda-lime.  
3. Attach an exhaust line to the exhaust outlet port.  
The exhaust from the pump and vent lines should be vented to atmospheric pressure using  
maximum of 10 meters of ¼” PTEF tubing.  
CAUTION  
Venting should be outside the shelter or immediate area surrounding the  
instrument.  
4. Attach a source of dried air scrubbed of CO2 to the purge inlet port  
NOTE  
The source of purge air should be at 20-25 psig and capable of maintaining a flow of at  
least 0.5 liters/min.  
Purge source air pressure should not exceed 35 pisg  
5. Once the appropriate pneumatic connections have been made, check all pneumatic fittings for  
leaks using a procedure similar to that defined in Section 9.3.3.  
04584 Rev A1  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
3.1.2.2. Connections with Internal Valve Options Installed  
Figures 3-5 through 3-8 show the proper pneumatic connections for MGFC7000E’s with various  
optional internal valve sets installed.  
Source of  
SAMPLE Gas  
VENT if input is pressurized  
Certified  
CO2 Gas  
Sample  
Needle  
Exhaust  
valve to  
control  
flow  
MODEL  
GFC7000E  
VENT  
Vent Span  
Pressure Span  
MODEL 701  
Zero Air  
Generator  
IZS  
VENT  
Purge In  
Indicating  
soda-lime  
Figure 3-5:  
Pneumatic Connections–MGFC7000E with Zero/Span/Shutoff Valves (OPT  
50)  
Source of  
SAMPLE Gas  
VENT if input is pressurized  
Certified  
CO2 Gas  
Sample  
Exhaust  
MODEL 701  
Zero Air  
Generator  
MODEL  
GFC7000E  
VENT  
Vent Span  
Pressure Span  
IZS  
External Zero  
Air Scrubber  
Purge In  
Indicating  
soda-lime  
Figure 3-6  
Pneumatic Connections–MGFC7000E with Zero/Span/Shutoff Valves and  
External Zero Air Scrubber (OPT 51)  
04584 Rev A1  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
Source of  
SAMPLE Gas  
VENT if input is pressurized  
MODEL 700  
Gas Dilution  
Calibrator  
Sample  
Exhaust  
Certified  
CO2 Gas  
Needle  
valve to  
control flow  
MODEL  
GFC7000E  
Vent Span  
VENT  
VENT  
Pressure Span  
Indicating  
soda-lime  
IZS  
Purge In  
MODEL 701  
Zero Air  
Generator  
Figure 3-7:  
Pneumatic Connections–MGFC7000E with Zero/Span Valves (OPT 52)  
Source of  
SAMPLE Gas  
VENT if input is pressurized  
Certified  
CO Gas  
MODEL 700  
Gas Dilution  
Calibrator  
Sample  
Exhaust  
Indicating  
soda-lime  
VENT  
MODEL  
GFC7000E  
Vent Span  
Pressure Span  
External Zero  
Air Scrubber  
IZS  
MODEL 701  
Zero Air Generator  
Purge In  
Figure 3-8:  
Pneumatic Connections–MGFC7000E with Zero/Span Valves with External  
Zero air Scrubber (OPT 53)  
04584 Rev A1  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
Some applications may require multipoint calibration checks where span gas of several different  
concentrations is needed. We recommend using high-concentration, certified, calibration gas  
supplied to the analyzer through a Gas Dilution Calibrator such as a Teledyne Instruments Model  
700. This type of calibrator precisely mixes Span Gas and Zero Air to produce any concentration  
level between 0 ppm and the concentration of the calibrated gas.  
Figure 3-8 depicts the pneumatic set up in this sort of application of a Model GFC7000E CO2  
Analyzer with zero/span/shutoff valve option 50 installed (a common configuration for this type of  
application).  
Gas Pressure should  
be regulated at  
Source of  
30 – 35 PSIG  
SAMPLE Gas  
VENT if input is pressurized  
MODEL 700  
Gas Dilution  
Calibrator  
Sample  
VENT  
Exhaust  
MODEL  
GFC7000E  
VENT  
Calibrated  
CO2 Gas  
Vent Span  
Pressure Span  
IZS  
Indicating  
soda-lime  
VENT  
Purge In  
Needle  
valve to  
control flow  
MODEL 701  
Zero Air Generator  
Figure 3-9:  
Example of Pneumatic Set up for Multipoint Calibration of M360  
3.2. Initial Operation  
If you are unfamiliar with the MGFC7000E theory of operation, we recommend that you read  
Chapter 10.  
For information on navigating the analyzer’s software menus, see the menu trees described in  
Appendix A.1.  
NOTE  
The analyzer’s cover must be installed to ensure that the temperatures of the GFC  
wheel and absorption cell assemblies are properly controlled.  
04584 Rev A1  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
3.2.1. Startup  
After electrical and pneumatic connections are made, turn on the instrument and pump power.  
The exhaust and PMT cooler fans should start. The display should immediately display a single,  
horizontal dash in the upper left corner of the display. This will last approximately 30 seconds  
while the CPU loads the operating system.  
Once the CPU has completed this activity it will begin loading the analyzer firmware and  
configuration data. During this process, string of messages will appear on the analyzer’s front  
panel display:  
System waits 3 seconds  
then automatically begins  
its initialization routine.  
SELECT START OR REMOTE  
START  
:
3
.
No action required.  
CHECKING FLASH STATUS  
:
1
1
1
System is checking the format  
of the instrument’s flash  
memory chip.  
If at this point,  
STARTING INSTRUMENT CODE  
STARTING INSTRUMENT W/FLASH  
:
**FLASH FORMAT INVALID**  
appears, contact T–API customer service  
The instrument is loading  
configuration and calibration  
data from the flash chip  
:
The instrument is  
loading the analyzer  
firmware.  
M200E NOX ANALYZER  
BOOT PROGRESS [XXXXX 50%_ _ _ _ _]  
The revision level of the  
firmware installed in your  
analyzer is briefly displayed  
SOFTWARE REVISION D.6  
BOOT PROGRESS [XXXXXXXX 80% _ _]  
SAMPLE  
TEST  
SYSTEM RESET  
CO2=X.XXX  
CLR SETUP  
Firmware  
fully booted  
CAL  
Press CLR to clear initial  
warning messages.  
(see Section 3.2.3)  
04584 Rev A1  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
The analyzer should automatically switch to SAMPLE mode after completing the boot-up  
sequence and start monitoring CO2 gas. Warm-Up  
3.2.2. Warm Up  
The MGFC7000E requires about 30 minutes warm-up time before reliable CO2 measurements can  
be taken. During that time, various portions of the instrument’s front panel will behave as follows.  
See Figure 3-10 for locations.  
Table 3-5:  
Front Panel Display During System Warm-Up  
Name  
Color  
Behavior  
Significance  
Concentration  
Field  
N/A  
Displays current,  
compensated CO2  
Concentration  
N/A  
Mode Field  
N/A  
Displays blinking  
“SAMPLE”  
Instrument is in sample mode but is still in the  
process of warming up.  
STATUS LED’s  
Sample  
Green  
On  
Unit is operating in sample mode, front panel  
display is being updated.  
Flashes On/Off when adaptive filter is active  
The instrument’s calibration is not enabled.  
Cal  
Yellow  
Red  
Off  
Fault  
Blinking  
The analyzer is warming up and hence out of  
specification for a fault-free reading. various warning  
messages will appear.  
MODE FIELD MESSAGE FIELD  
CONCENTRATION FIELD STATUS LED’s  
FASTENER  
LOCKING SCREW  
FASTENER  
SAMPLE  
CAL  
CO2 = 400.0  
SETUP  
SAMPLE A  
<TST TST>  
RANGE = 500.0 PPM  
CAL  
FAULT  
POWER  
GAS FILTER CORRELATION CO2 ANALYZER- MODEL GFC7000E  
KEY DEFINITIONS KEYBOARD  
ON / OFF SWITCH  
Figure 3-10: Front Panel Layout  
04584 Rev A1  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
3.2.3. Warning Messages  
Because internal temperatures and other conditions may be outside be specified limits during the  
analyzer’s warm-up period, the software will suppress most warning conditions for 30 minutes  
after power up.  
If warning messages persist after the 30 minutes warm up period is over, investigate their cause  
using the troubleshooting guidelines in Chapter 11 of this manual. The following table includes a  
brief description of the various warning messages that may appear.  
Table 3-6:  
Possible Warning Messages at Start-Up  
MESSAGE  
MEANING  
The instruments A/D circuitry or one of its analog outputs is not calibrated.  
The Temperature of the optical bench is outside the specified limits.  
ANALOG CAL WARNING  
BENCH TEMP WARNING  
BOX TEMP WARNING  
Remote span calibration failed while the dynamic span feature was set to  
turned on  
CANNOT DYN SPAN  
CANNOT DYN ZERO  
Remote zero calibration failed while the dynamic zero feature was set to  
turned on  
Configuration was reset to factory defaults or was erased.  
Configuration storage was reset to factory configuration or erased.  
CONFIG INITIALIZED  
Concentration alarm 1 is enabled and the measured CO2 level is the set  
point.  
CONC ALRM1 WARNING  
Concentration alarm 2 is enabled and the measured CO2 level is the set  
point.  
CONC ALRM2 WARNING  
iDAS data storage was erased.  
DATA INITIALIZED  
FRONT PANEL WARN  
Firmware is unable to communicate with the front panel.  
The temperature of the IR photometer is outside the specified limits.  
PHOTO TEMP WARNING  
REAR BOARD NOT DET  
The CPU is unable to communicate with the motherboard.  
The firmware is unable to communicate with the relay board.  
The flow rate of the sample gas is outside the specified limits.  
Sample gas pressure outside of operational parameters.  
RELAY BOARD WARN  
SAMPLE FLOW WARN  
SAMPLE PRESS WARN  
The temperature of the sample gas is outside the specified limits.  
The IR source may be faulty.  
SAMPLE TEMP WARN  
SOURCE WARNING  
The instrument is not properly tracking the rotation of the Gas Filter  
Correlation wheel.  
SYNC WARNING  
The computer was rebooted.  
SYSTEM RESET  
The Gas Filter Correlation wheel temperature is outside the specified limits.  
WHEEL TEMP WARNING  
04584 Rev A1  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
To View and Clear the various warning messages press:  
SAMPLE  
HVPS WARNING  
CAL MSG  
CO2 = 0.00  
TEST deactivates warning  
TEST  
CLR SETUP  
messages  
MSG activates warning  
SAMPLE  
RANGE=500.000 PPM  
MSG  
CO2 = 0.00  
messages.  
<TST TST> keys replaced with  
< TST TST > CAL  
CLR SETUP  
TEST key  
SAMPLE  
HVPS WARNING  
CO2 = 0.00  
Press CLR to clear the current  
message.  
TEST  
CAL  
MSG  
CLR SETUP  
NOTE:  
If more than one warning is active, the  
next message will take its place  
If the warning message persists  
after several attempts to clear it,  
the message may indicate a  
real problem and not an artifact  
of the warm-up period  
Once the last warning has been  
cleared, the analyzer returns to  
SAMPLE mode  
Make sure warning messages are  
not due to real problems.  
3.2.4. Functional Check  
1. After the analyzer’s components has warmed up for at least 30 minutes, verify that the  
software properly supports any hardware options that were installed.  
2. Check to make sure that the analyzer is functioning within allowable operating parameters.  
Appendix C includes a list of test functions viewable from the analyzer’s front panel as well as  
their expected values. These functions are also useful tools for diagnosing performance  
problems with your analyzer (Section 11.1.2). The enclosed Final Test and Validation Data  
sheet (part number 04307) lists these values before the instrument left the factory.  
To view the current values of these parameters press the following key sequence on the  
analyzer’s front panel. Remember until the unit has completed its warm up these parameters  
may not have stabilized.  
SAMPLE  
RANGE = 500.000 PPM  
CO2 = XXX.X  
SETUP  
< TST TST > CAL  
RANGE  
RANGE11  
RANGE21  
STABIL  
Toggle <TST TST> keys to  
scroll through list of functions  
CO2 MEAS  
CO2 REF  
MR RATIO  
PRES  
SAMP FL  
SAMP TEMP  
BENCH TEMP  
WHEEL TEMP  
BOX TEMP  
PHT DRIVE  
SLOPE  
Refer to  
Section  
6.X.X for  
definitions  
of these  
test  
functions.  
1 Only appears instrument is set  
for DUAL or AUTO reporting  
range modes  
OFFSET  
TEST  
TIME  
04584 Rev A1  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
3. If your analyzer has a Ethernet card (Option 63) installed and your network is running a  
dynamic host configuration protocol (DHCP) software package, the Ethernet option will  
automatically configure its interface with your LAN. However, it is a good idea to check these  
settings to make sure that the DHCP has successfully downloaded the appropriate network  
settings from your network server (See Section 6.10.9.2).  
If your network is not running DHCP, you will have to configure the analyzer’s interface  
manually (See Section 6.10.9.3).  
3.3. Initial Calibration Procedure  
The next task is to calibrate the analyzer.  
To perform the following calibration you must have sources for zero air and span gas available for  
input into the sample port on the back of the analyzer. See Section 3.1.2 for instructions for  
connecting these gas sources.  
While it is possible to perform this procedure with any range setting we recommend that you  
perform this initial checkout using the 50 ppm range.  
NOTE  
The following procedure assumes that the instrument does not have any of  
the available Zero/Span Valve Options installed.  
See Section 7.4 for instructions for calibrating instruments possessing Z/S  
valve options.  
1. Set the Analog Output Range of the MGFC7000E  
04584 Rev A1  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
SETUP X.X  
CFG DAS RNGE PASS CLK MORE  
EXIT  
SETUP X.X  
RANGE CONTROL MENU  
MODE SET UNIT  
EXIT  
Press this button to select the  
concentration units of measure:  
Press this button to set  
the analyzer for SNGL  
DUAL or AUTO ranges  
PPB, PPM, UGM, MGM  
SETUP X.X  
RANGE: 500.000 CONC  
0
0
5
0
0
.0  
ENTR EXIT  
EXIT ignores the new setting and  
returns to the RANGE CONTROL  
MENU.  
To change the value of the  
reporting range span, enter the  
number by pressing the key under  
each digit until the expected value  
appears.  
ENTR accepts the new setting and  
returns to the  
SETUP X.X  
RANGE: 500.000 Conc  
.0  
RANGE CONTROL MENU.  
0
0
0
5
0
ENTR EXIT  
04584 Rev A1  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
2. Set the expected CO2 span gas concentration  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP  
< TST TST > CAL  
This sequence causes the  
analyzer to prompt for the  
expected CO2 span  
concentration.  
M-P CAL  
RANGE = 500.000 PPM  
CO2 =X.XXX  
EXIT  
< TST TST > ZERO  
CONC  
The CO2 span  
concentration values  
automatically default to  
400.0 Conc.  
EXIT ignores the new setting  
and returns to the previous  
display.  
M-P CAL  
CO2 SPAN CONC: 400.000 Conc  
To change this value to  
the actual concentration of  
the span gas, enter the  
number by pressing the  
key under each digit until  
the expected value  
ENTR accepts the new setting  
0
0
0
4
5
.0  
ENTR EXIT  
and returns to the  
previous display..  
appears.  
NOTE  
For this Initial Calibration it is important to independently verify the precise CO2  
Concentration Value of the SPAN gas.  
If the source of the Span Gas is from a Calibrated Bottle, use the exact concentration  
value printed on the bottle.  
04584 Rev A1  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
3. Perform the Zero/Span Calibration Procedure  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =XXX.X  
SETUP  
Set the Display to show the  
STABIL test function.  
This function calculates the  
stability of the COx  
< TST TST > CAL  
measurement  
SAMPLE  
STABIL=XXX.X PPM  
CO2 =XXX.X  
SETUP  
< TST TST > CAL  
ACTION:  
Allow zero gas to enter the sample port at the  
rear of the instrument.  
Wait until STABIL  
falls below 1.0 ppm.  
This may take several  
minutes.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
< TST TST > CAL  
SETUP  
M-P CAL  
STABIL=XXX.X PPM  
CONC  
CO2 =XXX.X  
EXIT  
< TST TST > ZERO  
Press ENTR to changes the  
OFFSET & SLOPE values for the  
CO2 measurements.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
< TST TST > ENTR  
CONC  
EXIT  
Press EXIT to leave the calibration  
unchanged and return to the  
previous menu.  
ACTION:  
Allow span gas to enter the sample port at the  
rear of the instrument.  
The value of  
STABIL may jump  
significantly.  
Wait until it falls back  
below 1.0 ppm  
The SPAN key now  
appears during the  
transition from zero to  
span.  
This may take several  
minutes.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
EXIT  
< TST TST >  
M-P CAL  
SPAN CONC  
You may see both keys.  
If either the ZERO or  
SPAN buttons fail to  
appear see Section 11  
for troubleshooting tips.  
Press ENTR to change the  
OFFSET & SLOPE values for the  
CO2 measurements.  
RANGE = 500.000 PPM CO2 =XXX.X  
EXIT  
Press EXIT to leave the calibration  
unchanged and return to the  
previous menu.  
< TST TST > ENTR SPAN CONC  
M-P CAL  
RANGE = 500.000 PPM CO2 =XXX.X  
CONC EXIT  
EXIT returns to the main  
SAMPLE display  
< TST TST > ENTR  
The Model GFC7000E Analyzer is now ready for operation  
04584 Rev A1  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
NOTE  
Once you have completed the above set-up procedures, please fill out the Quality  
Questionnaire that was shipped with your unit and return it to Teledyne Instruments.  
This information is vital to our efforts in continuously improving our service and our  
products.  
THANK YOU.  
Front Panel  
IR Source  
On/Off Switch &  
Circuit Breaker  
Particulate Filter  
GFC Wheel  
Housing Purge  
Gas Inlet & Flow  
Control Orifice  
Optical Bench  
Gas Inlet  
Optional  
Sample/Cal  
Valve  
GFC Motor  
GFC Wheel Housing  
& IR Source Heat  
Sample Gas  
Critical  
Flow Orifice  
Optional  
Zero/Span  
Valve  
Gas Flow  
Sensor Assy  
Flow Sensor  
Optional  
Shutoff  
Valve  
Purge Gas  
Pressure  
Control Assy  
Pump Assy  
Sample Gas  
Pressure Sensor  
Optical Bench  
Gas Outlet  
Sample Gas  
Temperature  
Sensor  
Optional  
Ethernet Card  
CPU Card  
Mother  
Board  
Power  
Receptacle  
Rear Panel  
Fan  
Figure 3-11: Assembly Layout  
04584 Rev A1  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
Sample Gas Outlet  
fitting  
Sample Gas Flow  
Sensor  
Sample Chamber  
Sync/Demod PCA  
Housing  
Pressure Sensor(s)  
Bench  
Temperature  
Thermistor  
Shock Absorbing  
Mounting Bracket  
Opto-Pickup  
PCA  
Purge Gas  
Pressure Regulator  
IR Source  
GFC Wheel  
Heat Sync  
GFC Wheel Motor  
GFC Temperature  
Sensor  
Purge Gas  
Inlet  
GFC Heater  
Figure 3-12: Optical Bench Layout  
INSTRUMENT CHASSIS  
SAMPLE GAS  
INLET  
PUMP  
GFC Wheel  
Motor  
Purge Gas  
Flow Rate  
Control  
EXHAUST GAS  
OUTLET  
Orifice  
GFC Motor  
Heat Sync  
Purge Gas  
Pressure  
Control Assy  
GFC Wheel  
Housing  
PURGE GAS  
INLET  
FLOW / PRESSURE  
SENSOR PCA  
SAMPLE  
PRESSURE  
SENSOR  
SAMPLE CHAMBER  
FLOW  
SENSOR  
VENT SPAN  
OUTLET  
PRESSURE  
SPAN INLET  
PARTICULATE  
FILTER  
IZS INLET  
Figure 3-13: Internal Pneumatic Flow – Basic Configuration  
04584 Rev A1  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Getting Started  
User Notes  
04584 Rev A1  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Frequently Asked Questions  
4. FREQUENTLY ASKED QUESTIONS  
4.1. FAQ’s  
The following is a list from the Teledyne Instruments’ Customer Service Department of the most  
commonly asked questions relating to the Model CO2 Analyzer.  
Q: How do I get the instrument to zero / Why is the zero key not displayed?  
A: See Section 11.2.4 Inability to zero.  
Q: How do I get the instrument to span / Why is the span key not displayed?  
A: See Section 11.2.3 Inability to span.  
Q: Why does the ENTR key sometimes disappear on the Front Panel Display?  
A: During certain types of adjustments or configuration operations, the ENTR key will disappear if  
you select a setting that is nonsensical (such as trying to set the 24-hour clock to 25:00:00) or  
out of the allowable range for that parameter (such as selecting an iDAS Holdoff period of more  
than 20 minutes).  
Once you adjust the setting in question to an allowable value, the ENTR key will re-appear.  
Q: Is there an optional midpoint calibration?  
A: There is an optional mid point linearity adjustment, however, midpoint adjustment is applicable  
only to applications where CO2 measurements are expected above 100 ppm. Call Teledyne  
Instruments’ Service Department for more information on this topic.  
Q: How do I make the display and datalogger analog input agree?  
A: This most commonly occurs when an independent metering device is used besides the  
datalogger/recorded to determine gas concentration levels while calibrating the analyzer. These  
disagreements result from the analyzer, the metering device and the datalogger having slightly  
different ground levels.  
If the only difference is a DC offset then it is possible to enter a compensating value in the analog  
outputs. This procedure is described in Section 6.9.4.3 of this manual.  
Alternately, use the datalogger itself as the metering device during calibrations procedures.  
Q: How do I perform a leak check?  
See Section 9.3.3.  
Q: How do I measure the sample flow?  
A: Sample flow is measured by attaching a calibrated rotameter, wet test meter, or other flow-  
measuring device to the sample inlet port when the instrument is operating. The sample flow  
should be 800 cm3/min ±10%. See Section 9.3.4.  
04584 Rev A1  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Frequently Asked Questions  
Q: How long does the IR source last?  
A: Typical lifetime is about 2-3 years.  
Q: Where is the sintered filter/sample flow control orifice?  
A: These components are located inside the flow control assembly that is attached to the inlet side  
of the sample pump, see Figure 3-13. See Section 11.5.1 for instructions on disassembly and  
replacement.  
Q: How do I set up a SEQUENCE to run a nightly calibration check?  
A: The setup of this option is located in Section 7.6.  
Q: How do I set the analog output signal range and offset?  
A: Instructions for this can be found in Section 6.9.4 which describes analog I/O configuration.  
Q: What is the averaging time for an GFC7000E?  
A: The default averaging time, optimized for ambient pollution monitoring, is 150 seconds for  
stable concentrations and 10 seconds for rapidly changing concentrations; see Section 10.5.1 for  
more information. However, it is adjustable over a range of 0.5 second to 200 seconds (please  
contact customer service for more information).  
4.2. Glossary  
ASSY - acronym for Assembly  
DAS - acronym for data acquisition system, the old acronym of iDAS.  
DIAG - acronym for diagnostics, the diagnostic settings of the analyzer  
DHCP: acronym for dynamic host configuration protocol. A protocol used by LAN or Internet  
servers to automatically set up the interface protocols between themselves and any other  
addressable device connected to the network.  
DOC - Disk On Chip, the analyzer’s central storage area for analyzer firmware, configuration  
settings and data. This is a solid state device without mechanically moving parts that acts as a  
computer hard disk drive under Æ DOS with disk label “C”. DOC chips come with 8 mb in the E-  
series analyzer standard configuration but are available in larger sizes.  
DOS - Disk Operating System. The E-series analyzers uses DR DOS  
EEPROM - also referred to as a FLASH chip.  
FLASH - flash memory is non-volatile, solid-state memory.  
GFC – Acronym for Gas Filter Correlation.  
I2C bus - a clocked, bi-directional, serial bus for communication between individual analyzer  
components  
04584 Rev A1  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Frequently Asked Questions  
iDAS - acronym for internal data acquisition system  
IP – acronym for internet protocol  
LAN - acronym for local area network  
LED - acronym for light emitting diode  
PCA - acronym for printed circuit assembly, the Æ PCB with electronic components, ready to use.  
PCB - acronym for printed circuit board, the bare board without electronic components  
RS-232 - an electronic communications type of a serial communications port  
RS-485 - an electronic communications type of a serial communications port  
TCP/IP - acronym for transfer control protocol / internet protocol, the standard communications  
protocol for Ethernet devices.  
VARS - acronym for variables, the variables settings of the analyzer  
User Notes  
04584 Rev A1  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Optional Hardware and Software  
5. OPTIONAL HARDWARE AND SOFTWARE  
This section includes a brief description of the hardware and software options available for the  
Model GFC7000E Gas Filter Correlation Carbon Dioxide Analyzer. For assistance with ordering  
these options please contact the Sales department of Teledyne – Advanced Pollution Instruments  
at:  
TEL: 626-961-9221  
TEL: 626-934-1500  
FAX: 626-961-2538  
WEB SITE: www.teledyne-ai.com  
5.1. Rack Mount Kits (Options 20a, 20b & 21)  
Option Number  
Description  
OPT 20A  
OPT 20B  
OPT 21  
Rack mount brackets with 26 in. chassis slides.  
Rack mount brackets with 24 in. chassis slides.  
Rack mount brackets only  
Each of these options, permits the Analyzer to be mounted in a standard 19" x 30" RETMA rack.  
5.2. Current Loop Analog Outputs (Option 41)  
This option adds isolated, voltage-to-current conversion circuitry to the analyzer’s analog outputs.  
This option may be ordered separately for any of the analog outputs, it can be installed at the  
factory or added later. Call Teledyne Instruments sales for pricing and availability.  
The current loop option can be configured for any output range between 0 and 20 mA.  
Information on calibrating or adjusting these outputs can be found in Section 6.9.4.9.  
Figure 5-1:  
Current Loop Option Installed on the Motherboard  
04584 Rev A1  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Optional Hardware and Software  
5.3. Expendable Kits (Options 42C, 42D and 43)  
OPTION NUMBER  
OPT 42C  
DESCRIPTION  
1 year’s supply of replacement of 47mm dia. particulate filters  
OPT 42D  
1 full replacement’s volume of indicating soda-lime for the external CO2  
scrubber included with options 51 & 53 ( approximate active lifetime: 1  
year)  
OPT 43  
Options 42 C & 42D  
5.4. Calibration Valves Options  
There are four available options involving Zero/Span/Shutoff valves. From an operational and  
software standpoint, all of the options are the same, only the source of the span and zero gases  
are different.  
5.4.1. Zero/Span/Shutoff Valve (Option 50)  
This option requires that both zero air and span gas be supplied from external sources. It is  
specifically designed for applications where span gas will be supplied from a pressurized bottle of  
calibrated CO2 gas. A critical flow control orifice, internal to the instrument ensures that the  
proper flow rate is maintained. An internal vent line, isolated by a shutoff valve ensures that the  
gas pressure of the span gas is reduced to ambient atmospheric pressure. normally zero air would  
be supplied from zero air module such as a Teledyne Instruments Model 701.  
In order to ensure that span gas does not migrate backwards through the vent line and alter the  
concentration of the span gas, a gas line not less than 2 meters in length should be attached to  
the vent span outlet on the rear panel of the analyzer. To prevent the buildup of back pressure,  
this vent line should not be greater than 10 meters in length.  
The following table describes the state of each valve during the analyzer’s various operational  
modes.  
Table 5-1:  
Zero/Span Valve Operating States for Options 50 & 51  
MODE  
VALVE  
CONDITION  
VALVE PORT CONNECTION  
(Fig. 5-2)  
Sample/Cal  
Zero/Span  
Open to SAMPLE inlet  
Open to ZERO AIR inlet  
Closed  
3 Æ 2  
3 Æ 2  
N/A  
SAMPLE  
(Normal State)  
Shutoff Valve  
Sample/Cal  
Zero/Span  
Open to zero/span inlet  
Open to ZERO AIR inlet  
Closed  
1 Æ 2  
3 Æ 2  
N/A  
ZERO CAL  
SPAN CAL  
Shutoff Valve  
Sample/Cal  
Zero/Span  
Open to ZERO/SPAN inlet  
Open to SPAN GAS inlet  
1 Æ 2  
1 Æ 2  
Shutoff Valve  
Open to PRESSURE SPAN  
Inlet  
1 Æ 2  
The minimum span gas flow rate required for this option is 800 cm3/min.  
04584 Rev A1  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Optional Hardware and Software  
The state of the zero/span valves can also be controlled:  
Manually from the analyzer’s front panel by using the SIGNAL I/O controls located under  
the DIAG Menu (Section 6.9.2),  
By activating the instrument’s AutoCal feature (Section 7.6),  
Remotely by using the external digital control inputs (Section 6.13.1.2 and Section 7.5.2),  
or  
Remotely through the RS-232/485 serial I/O ports (see Appendix A-6 for the appropriate  
commands).  
INSTRUMENT CHASSIS  
SAMPLE GAS  
INLET  
PUMP  
GFC Wheel  
Motor  
EXHAUST GAS  
Purge Gas  
OUTLET  
Flow Control  
Orifice  
GFC Motor  
Heat Sync  
Purge Gas  
Pressure  
Control Assy  
GFC Wheel  
Housing  
PURGE GAS  
INLET  
FLOW / PRESSURE  
SENSOR PCA  
SAMPLE  
PRESSURE  
SENSOR  
SAMPLE CHAMBER  
FLOW  
SENSOR  
VENT SPAN  
OUTLET  
PRESSURE  
SPAN INLET  
ZERO/SPAN  
VALVE  
SAMPLE /CAL  
VALVE  
SHUT OFF  
VALVE  
2
3
3
1
PARTICULATE  
FILTER  
1
2
1
2
IZS INLET  
Figure 5-2:  
Internal Pneumatic Flow – Zero/Span/Shutoff Valves OPT 50 & 51  
04584 Rev A1  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Optional Hardware and Software  
5.4.2. Zero/Span/Shutoff with External CO2 Scrubber  
(Option 51)  
Option 51 is operationally and pneumatically identical to Option 50 above (see section 5.4.1),  
except that the zero air is generated by an externally mounted zero air scrubber filled with  
indicating soda-lime that changes color from white to pink as it becomes saturated.  
5.4.3. Zero/Span Valve (Option 52)  
This valve option is intended for applications where zero air is supplied by a zero air generator like  
the Teledyne Instruments Model 701 and span gas are being supplied by Gas Dilution Calibrator  
Like the Teledyne Instruments Model 700 or 702. Internal zero/span and sample/cal valves  
control the flow of gas through the instrument, but because the calibrator limits the flow of span  
gas no shutoff valve is required.  
In order to ensure that span gas does not migrate backwards through the vent line and alter the  
concentration of the span gas, a gas line not less than 2 meters in length should be attached to  
the vent span outlet on the rear panel of the analyzer. To prevent the buildup of back pressure,  
this vent line should not be greater than 10 meters in length.  
The following table describes the state of each valve during the analyzer’s various operational  
modes.  
Table 5-2:  
Zero/Span Valve Operating States for Options 52 & 53  
Mode  
Valve  
Condition  
Valve Port Connection  
(Fig. 5-2)  
Sample/Cal  
Zero/Span  
Sample/Cal  
Zero/Span  
Sample/Cal  
Zero/Span  
Open to SAMPLE inlet  
Open to ZERO AIR inlet  
Open to zero/span inlet  
Open to ZERO AIR inlet  
Open to ZERO/SPAN inlet  
Open to SPAN GAS inlet  
3 Æ 2  
3 Æ 2  
1 Æ 2  
3 Æ 2  
1 Æ 2  
1 Æ 2  
SAMPLE  
(Normal State)  
ZERO CAL  
SPAN CAL  
The minimum span gas flow rate required for this option is 800 cm3/min.  
The state of the zero/span valves can also be controlled:  
Manually from the analyzer’s front panel by using the SIGNAL I/O controls located under  
the DIAG Menu (Section 6.9.2),  
By activating the instrument’s AutoCal feature (Section 7.6),  
Remotely by using the external digital control inputs (Section 6.13.1.2 and Section 7.5.2),  
or  
Remotely through the RS-232/485 serial I/O ports (see Appendix A-6 for the appropriate  
commands).  
04584 Rev A1  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Optional Hardware and Software  
INSTRUMENT CHASSIS  
SAMPLE GAS  
INLET  
PUMP  
GFC Wheel  
Motor  
EXHAUST GAS  
OUTLET  
Purge Gas  
Flow Control  
Orifice  
GFC Motor  
Heat Sync  
Purge Gas  
Pressure  
Control Assy  
GFC Wheel  
Housing  
PURGE GAS  
INLET  
FLOW / PRESSURE  
SENSOR PCA  
SAMPLE  
PRESSURE  
SENSOR  
SAMPLE CHAMBER  
FLOW  
SENSOR  
VENT SPAN  
OUTLET  
PRESSURE  
SPAN INLET  
ZERO/SPAN  
VALVE  
SAMPLE /CAL  
VALVE  
2
3
3
1
PARTICULATE  
FILTER  
1
2
IZS INLET  
Figure 5-3:  
Internal Pneumatic Flow – Zero/Span OPT 52 & 53  
5.4.4. Zero/Span Valve with External CO2 Scrubber (Option 53)  
Option 53 is operationally and pneumatically identical to Option 52 above (see Section 5.4.3),  
except that the zero air is generated by an externally mounted zero air scrubber filled with  
indicating soda-lime that changes color from white to pink as it becomes saturated.  
5.5. Communication Options  
5.5.1. RS232 Modem Cable (Option 60)  
This option consists of a cable to connect the analyzer’s COM1 port to a computer, a code  
activated switch or any other communications device that is equipped with a DB-9 male  
connector. The cable is terminated with two DB-9 female connectors, one of which fits the  
analyzer’s COM1 port.  
Some older computers or code activated switches with a DB-25 serial connector will need a  
different cable or an appropriate adapter.  
04584 Rev A1  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Optional Hardware and Software  
5.5.2. RS-232 Multidrop (Option 62)  
The multidrop option is used with any of the RS-232 serial ports to enable communications of  
several analyzers with the host computer over a chain of RS-232 cables. The option consists of a  
small box, which can be attached to the analyzer, with a termination switch, a power connector  
and two serial ports, one incoming from the analyzer (cable supplied) and one outgoing (requires  
additional cable) to the next analyzer’s multi-drop box. One Option 62 is required per analyzer.  
The first incoming port on the first box connects to the host computer and the outgoing port on  
the last multi-drop box needs to be terminated. Setup and user instructions are covered in the  
Teledyne Instruments’ multi-drop manual, part number 021790000.  
5.5.3. Ethernet (Option 63)  
The Ethernet option allows the analyzer to be connected to any Ethernet local area network (LAN)  
running TCP/IP. The local area network must have routers capable of operating at 10BaseT. If  
Internet access is available through the LAN, this option also allows communication with the  
instrument over the public Internet.  
When installed, this option is electronically connected to the instrument’s COM2 serial port making  
that port no longer available for RS-232/RS-485 communications through the COM2 connector on  
the rear panel. The option consists of a Teledyne Instruments designed Ethernet card (Figure 5-  
4), which is mechanically attached to the instrument’s rear panel. A 7-foot long, CAT-5 network  
cable terminated at both ends with standard RJ-45 connectors is included as well. Maximum  
communication speed is limited by the RS-232 port to 115.2 kBaud.  
Figure 5-4:  
MGFC7000E Ethernet Card and rear panel With Ethernet Installed  
5.6. Additional Manuals  
5.6.1. Printed Manuals (Option 70)  
Additional printed copies of this manual are available from Teledyne Instruments  
04584 Rev A1  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Optional Hardware and Software  
5.6.2. Manual on CD (Part number 045840200)  
This operators manual is also available on CD. The electronic document is stored in Adobe  
Systems Inc. Portable Document Format (PDF) and is viewable with Adobe Acrobat Reader®  
software, downloadable for free at http://www.adobe.com/  
The CD version of the manual has many advantages:  
Fully searchable text.  
Hypertext links for figures, tables, table of contents and embedded references for quick  
access of individual manual portions.  
A list of thumbnails, chapters and sections displayed at the left of the text.  
Internet links embedded in the manual will take you to the corresponding web site  
(requires an internet connection).  
5.7. Extended Warranty (Options 92 & 93)  
Two options are available for extending Teledyne Instruments’ standard warranty (Section 2.3).  
Both options have to be specified upon ordering the analyzer.  
Option Number  
OPT 92  
OPT 93  
Description  
Extends warranty to cover a two (2) year period from the date of  
purchase.  
Extends warranty to cover a five (5) year period from the date of  
purchase.  
5.8. Dilution Ratio Option  
The Dilution Ration Option is a software option that is designed for applications where the Sample  
gas is diluted before being analyzed by the Model GFC7000E. Typically this occurs in Continuous  
Emission Monitoring (CEM) applications where the quality of gas in a smoke stack is being tested  
and the sampling method used to remove the gas from the stack dilutes the gas.  
Once the degree of dilution is known, this feature allows the user to add an appropriate scaling  
factor to the analyzer’s CO2 concentration calculation so that the Measurement Range and  
concentration values displayed on the instrument’s Front Panel Display and reported via the  
Analog and Serial Outputs reflect the undiluted values.  
Instructions for using the dilution ratio option can be found in Section 6.7.7..  
5.9. Maintenance Mode Switch  
API’s instruments can be equipped with an switch that places the instrument in maintenance  
mode. When present, the switch accessed by opening the hinged front panel and is located on  
the rearward facing side of the display/keyboard driver PCA; on the left side; near the particulate  
filter.  
04584 Rev A1  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Optional Hardware and Software  
When in maintenance mode the instrument ignores all commands received via the COMM ports  
that alter the operation state of the instrument This includes all calibration commands, diagnostic  
menu commands and the reset instrument command. the instrument continues to measure  
concentration and send data when requested.  
This option is of particular use for instruments connected to multidrop or Hessen protocol  
networks.  
5.10. Second Language Switch  
API’s instruments can be equipped with switch that activates an alternate set of display message  
in a language other than the instruments default language. When present, the switch accessed by  
opening the hinged front panel and is located on the rearward facing side of the display/keyboard  
driver PCA; on the right side.  
To activate this feature, the instrument must also have a specially programmed Disk on Chip  
containing the second language.  
User Notes  
04584 Rev A1  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6. OPERATING INSTRUCTIONS  
To assist in navigating the analyzer’s software, a series of menu trees can be found in Appendix  
A-1 of this manual.  
NOTES  
The flow charts appearing in this section contain typical representations of the  
analyzer’s display during the various operations being described. These representations  
may differ slightly from the actual display of your instrument.  
The ENTR key may disappear if you select a setting that is invalid or out of the allowable  
range for that parameter, such as trying to set the 24-hour clock to 25:00:00. Once you  
adjust the setting to an allowable value, the ENTR key will re-appear.  
6.1. Overview of Operating modes  
The MGFC7000E software has a variety of operating modes. Most commonly, the analyzer will be  
operating in SAMPLE mode. In this mode, a continuous read-out of the CO2 concentration is  
displayed on the front panel and output as an analog voltage from rear panel terminals,  
calibrations can be performed, and TEST functions and WARNING messages can be examined.  
The second most important operating mode is SETUP mode. This mode is used for performing  
certain configuration operations, such as for the iDAS system, the reporting ranges, or the serial  
(RS-232/RS-485/Ethernet) communication channels. The SET UP mode is also used for  
performing various diagnostic tests during troubleshooting.  
Mode Field  
SAMPLE A  
RANGE = 500.00 PPM  
CO2 400.00  
SETUP  
<TST TST> CAL  
Figure 6-1:  
Front Panel Display  
The mode field of the front panel display indicates to the user which operating mode the unit is  
currently running.  
Besides SAMPLE and SETUP, other modes the analyzer can be operated in are:  
04584 Rev A1  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
Table 6-1:  
Analyzer Operating modes  
MEANING  
MODE  
DIAG  
One of the analyzer’s diagnostic modes is being utilized (see Section 6.9).  
M-P CAL  
This is the basic, multi-point calibration mode of the instrument and is activated  
by pressing the CAL key.  
SAMPLE  
SAMPLE A  
SETUP1  
Sampling normally, flashing indicates adaptive filter is on.  
Indicates that unit is in SAMPLE Mode and AUTOCAL feature is activated.  
SETUP mode is being used to configure the analyzer (CO2 sampling will continue  
during this process).  
SPAN CAL A  
Unit is performing span cal procedure initiated automatically by the analyzer’s  
AUTOCAL feature.  
SPAN CAL M  
SPAN CAL R  
Unit is performing span cal procedure initiated manually by the user.  
Unit is performing span cal procedure initiated remotely via the RS-232, RS-4485  
or digital i/o control inputs.  
ZERO CAL A  
Unit is performing zero cal procedure initiated automatically by the analyzer’s  
AUTOCAL feature.  
ZERO CAL M  
ZERO CAL R  
Unit is performing zero cal procedure initiated manually by the user.  
Unit is performing zero cal procedure initiated remotely via the RS-232, RS-4485  
or digital I/O control inputs.  
1
The revision of the Teledyne Instruments software installed in this analyzer will be displayed  
following the word SETUP. E.g. “SETUP E.0”  
Finally, the various CAL modes allow calibration of the analyzer. Because of its importance, this  
mode is described separately in Chapter 7.  
6.2. Sample Mode  
This is the analyzer’s standard operating mode. In this mode the instrument is analyzing the gas  
in the sample chamber, calculating CO2 concentration and reporting this information to the user  
via the front panel display, the analog outputs and, if set up properly, the RS-232/485/Ethernet  
ports.  
NOTE  
A value of “XXXX” displayed in the CO2 Concentration field means that the M/R ratio is  
invalid because CO2 REF is either too high(> 4950 mVDC) or too low (< 1250 VDC).  
6.2.1. Test Functions  
A series of test functions is available at the front panel while the analyzer is in SAMPLE mode.  
These parameters provide information about the present operating status of the instrument and  
are useful during troubleshooting (Section 11.1.2 ). They can also be recorded in one of the iDAS  
channels (Section 6.12) for data analysis. To view the test functions, press one of the <TST TST>  
keys repeatedly in either direction.  
Table 6-2:  
Units  
Test Functions Defined  
Meaning  
Parameter  
Display  
04584 Rev A1  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
Title  
The full scale limit at which the reporting range of the  
analyzer is currently set.  
THIS IS NOT the Physical Range of the instrument. See  
Section 6.7 for more information.  
Range  
RANGE  
PPB, PPM,  
UGM,  
RANGE11  
RANGE21  
MGM  
Standard deviation of CO2 concentration readings. Data  
points are recorded every ten second. The calculation uses  
the last 25 data points.  
Stability  
STABIL  
PPB, PPM  
UGM,  
MGM  
The demodulated, peak IR detector output during the  
measure portion of the CFG Wheel cycle.  
CO2 Measure  
CO2  
MEAS  
MV  
The demodulated, peak IR detector output during the  
reference portion of the CFG wheel cycle.  
CO2 Reference  
CO2 REF  
MV  
-
The result of CO2 MEAS divided by CO2 REF. This ratio is  
the primary value used to compute CO2 concentration. The  
value displayed is not linearized.  
Measurement / MR Ratio  
Reference  
Ratio  
The absolute pressure of the Sample gas as measured by a  
solid state pressure sensor located inside the sample  
chamber.  
Sample  
Pressure  
PRES  
In-Hg-A  
cc/min  
Sample mass flow rate. This is computed from the differential  
between the pressures measured up-stream and down-  
stream of the sample critical flow orifice pressures.  
Sample Flow  
SAMPLE  
FL  
The temperature of the gas inside the sample chamber.  
Sample  
Temperature  
SAMP  
TEMP  
°C  
°C  
Optical bench temperature.  
Bench  
Temperature  
BENCH  
TEMP  
Filter wheel temperature.  
Wheel  
Temperature  
WHEEL  
TEMP  
°C  
The temperature inside the analyzer chassis.  
Box  
BOX  
°C  
Temperature  
TEMP  
The drive voltage being supplied to the thermoelectric coolers  
of the IR photo-detector by the sync/demod Board.  
Photo-detector  
Temp. Control  
Voltage  
PHT  
DRIVE  
mV  
The sensitivity of the instrument as calculated during the last  
calibration activity. The SLOPE parameter is used to set the  
span calibration point of the analyzer.  
Slope  
SLOPE  
-
-
The overall offset of the instrument as calculated during the  
last calibration activity. The OFFSET parameter is used to  
set the zero point of the analyzer response.  
Offset  
OFFSET  
Displays the signal level of the TEST analog output channel.  
Only appears when the TEST channel has been activated.  
Test channel  
output signal  
TEST  
TIME  
mV, mA  
-
The current time. This is used to create a time stamp on  
iDAS readings, and by the AUTOCAL feature to trigger  
calibration events.  
Current Time  
1
Only appears when the instrument’s reporting range mode is set for DUAL or AUTO  
04584 Rev A1  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
To view the TEST Functions press the following Key sequence:  
SAMPLE  
RANGE = 500.000 PPM  
CO2 = XXX.X  
SETUP  
< TST TST > CAL  
RANGE  
RANGE11  
RANGE21  
STABIL  
Toggle <TST TST> keys to  
scroll through list of functions  
CO2 MEAS  
CO2 REF  
MR RATIO  
PRES  
SAMP FL  
SAMP TEMP  
BENCH TEMP  
WHEEL TEMP  
BOX TEMP  
PHT DRIVE  
SLOPE  
Refer to  
Table 6-2  
for  
definitions  
of these  
test  
functions.  
1 Only appears instrument is set  
for DUAL or AUTO reporting  
range modes  
OFFSET  
TEST  
TIME  
Figure 6-2  
Viewing MGFC7000E TEST Functions  
NOTE  
A value of “XXXX” displayed for any of the TEST functions indicates an out-of-range  
reading or the analyzer’s inability to calculate it.  
All pressure measurements are represented in terms of absolute pressure. Absolute,  
atmospheric pressure is 29.92 in-Hg-A at sea level. It decreases about 1 in-Hg per 300  
m gain in altitude. A variety of factors such as air conditioning and passing storms can  
cause changes in the absolute atmospheric pressure.  
6.2.2. Warning Messages  
The most common instrument failures will be reported as a warning on the analyzer’s front panel  
and through the COM ports. Section 11.1.1 explains how to use these messages to troubleshoot  
problems. Section 3.2.3 shows how to view and clear warning messages.  
Table 6-3 lists all warning messages for the current version of software.  
04584 Rev A1  
48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
Table 6-3:  
List of Warning Messages  
MESSAGE  
MEANING  
ANALOG CAL WARNING  
The instruments A/D circuitry or one of its analog outputs is not calibrated.  
The Temperature of the optical bench is outside the specified limits.  
BENCH TEMP WARNING  
BOX TEMP WARNING  
Remote span calibration failed while the dynamic span feature was set to  
turned on  
CANNOT DYN SPAN  
CANNOT DYN ZERO  
Remote zero calibration failed while the dynamic zero feature was set to  
turned on  
Configuration was reset to factory defaults or was erased.  
Concentration alarm 1 is enabled and the measured CO2 level is the set  
point.  
CONC ALRM1 WARNING  
CONC ALRM2 WARNING  
Concentration alarm 2 is enabled and the measured CO2 level is the set  
point.  
CONFIG INITIALIZED  
DATA INITIALIZED  
FRONT PANEL WARN  
Configuration storage was reset to factory configuration or erased.  
iDAS data storage was erased.  
Firmware is unable to communicate with the front panel.  
The temperature of the IR photometer is outside the specified limits.  
The CPU is unable to communicate with the motherboard.  
PHOTO TEMP WARNING  
REAR BOARD NOT DET  
RELAY BOARD WARN  
SAMPLE FLOW WARN  
SAMPLE PRESS WARN  
The firmware is unable to communicate with the relay board.  
The flow rate of the sample gas is outside the specified limits.  
Sample gas pressure outside of operational parameters.  
SAMPLE TEMP WARN  
SOURCE WARNING  
The temperature of the sample gas is outside the specified limits.  
The IR source may be faulty.  
SYNC WARNING  
The instrument is not properly tracking the rotation of the Gas Filter  
Correlation wheel.  
SYSTEM RESET  
The computer was rebooted.  
WHEEL TEMP WARNING  
The Gas Filter Correlation wheel temperature is outside the specified limits.  
04584 Rev A1  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
To view and clear warning messages  
SAMPLE  
HVPS WARNING  
CAL MSG  
CO2 = 0.00  
TEST deactivates warning  
TEST  
CLR SETUP  
messages  
MSG activates warning  
SAMPLE  
RANGE=500.000 PPM  
MSG  
CO2 = 0.00  
messages.  
<TST TST> keys replaced with  
< TST TST > CAL  
CLR SETUP  
TEST key  
SAMPLE  
HVPS WARNING  
CO2 = 0.00  
Press CLR to clear the current  
message.  
TEST  
CAL  
MSG  
CLR SETUP  
NOTE:  
If more than one warning is active, the  
next message will take its place  
If the warning message persists  
after several attempts to clear it,  
the message may indicate a  
real problem and not an artifact  
of the warm-up period  
Once the last warning has been  
cleared, the analyzer returns to  
SAMPLE mode  
Make sure warning messages are  
not due to real problems.  
Figure 6-3  
Viewing and Clearing MGFC7000E WARNING Messages  
6.3. Calibration Mode  
Pressing the CAL key switches the MGFC7000E into multi-point calibration mode. In this mode,  
the user can calibrate the instrument or check the instruments calibration with the use of  
calibrated zero or span gases.  
If the instrument includes either the zero/span valve option or IZS option, the display will also  
include CALZ and CALS keys. Pressing either of these keys also puts the instrument into  
multipoint calibration mode.  
The CALZ key is used to initiate a calibration of the zero point.  
The CALS key is used to calibrate the span point of the analyzer. It is recommended that  
this span calibration is performed at 90% of full scale of the analyzer’s currently selected  
reporting range.  
Because of their critical importance and complexity, calibration operations are described in detail  
in Chapter 7 of this manual. For more information concerning the zero/span, zero/span/shutoff  
and IZS valve options, see Section 5.4.  
6.3.1. SETUP – PASS: Calibration Password Security  
The MGFC7000E calibration functions may be password protected for to prevent inadvertent  
adjustments. When the calibration password has been enabled using the PASS menu item found  
under the Setup Menu (see below), the system will prompt the user for a password anytime CAL,  
CALZ, CALS activated.  
04584 Rev A1  
50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
The default status of the calibration password is OFF. To enable the calibration password press:  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
ENTR accepts  
displayed  
SAMPLE  
ENTER SETUP PASS : 818  
password value  
8
1
8
ENTR EXIT  
EXIT returns to  
SAMPLE display  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
SETUP X.X  
CAL. PASSWORD ENABLE: OFF  
CAL. PASSWORD  
default state is  
OFF  
Toggles  
password  
status On/Off  
OFF  
ENTR EXIT  
SETUP X.X  
PASSWORD ENABLE: ON  
ENTR EXIT  
ON  
ENTR accepts  
the change  
SETUP X.X  
ON  
PASSWORD ENABLE: ON  
ENTR EXIT  
EXIT ignores  
the change  
If the calibration password (101) is enabled, the following keypad sequence will be required to  
enter one of the calibration modes:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP  
< TST TST > CAL CALZ CALS  
SAMPLE  
ENTER SETUP PASS : 0  
0
Prompts  
password  
number  
0
0
0
ENTR EXIT  
ENTR EXIT  
SAMPLE  
ENTER SETUP PASS : 0  
1
Press  
individual  
keys to set  
1
101  
M-P CAL  
RANGE = 500.000 PPM  
CO2 =X.XXX  
EXIT  
< TST TST > ZERO  
CONC  
Continue calibration process …  
04584 Rev A1  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.4. SETUP Mode  
The SETUP mode contains a variety of choices that are used to configure the analyzer’s hardware  
and software features, perform diagnostic procedures, gather information on the instruments  
performance and configure or access data from the internal data acquisition system (iDAS). For a  
visual representation of the software menu trees, refer to Appendix A-1.  
The areas access under the Setup mode are:  
Table 6-4:  
Primary Setup Mode Features and Functions  
KEYPAD  
LABEL  
MANUAL  
SECTION  
MODE OR FEATURE  
DESCRIPTION  
Lists key hardware and software configuration  
information  
Analyzer Configuration  
Auto Cal Feature  
CFG  
6.5  
Used to set up an operate the AutoCal feature.  
Only appears if the analyzer has one of the internal  
valve options installed  
ACAL  
7.6  
Internal Data Acquisition  
(iDAS)  
DAS  
RNGE  
PASS  
CLK  
Used to set up the iDAS system and view recorded data  
6.12  
6.7  
Analog Output Reporting  
Range Configuration  
Used to configure the output signals generated by the  
instruments Analog outputs.  
Calibration Password  
Security  
Turns the calibration password feature ON/OFF  
6.3.1  
6.6  
Internal Clock  
Configuration  
Used to Set or adjust the instrument’s internal clock  
This button accesses the instruments secondary setup  
menu  
See  
Table 6-5  
Advanced SETUP features  
MORE  
Table 6-5:  
Secondary Setup Mode Features and Functions  
KEYPAD  
LABEL  
MANUAL  
SECTION  
MODE OR FEATURE  
DESCRIPTION  
Used to set up and operate the analyzer’s various  
external I/O channels including RS-232; RS 485,  
modem communication and/or Ethernet access.  
External Communication  
Channel Configuration  
6.10 &  
6.13  
COMM  
Used to view various variables related to the  
instruments current operational status  
System Status Variables  
VARS  
DIAG  
6.8  
6.9  
Used to access a variety of functions that are used to  
configure, test or diagnose problems with a variety of  
the analyzer’s basic systems  
System Diagnostic  
Features  
Used to activate the analyzer’s two gas concentration  
status alarms and set the alarm limits  
CO2 Concentration Alarms  
ALRM  
6.11  
NOTE  
Any changes made to a variable during one of the following procedures is not  
acknowledged by the instrument until the ENTR Key is pressed  
04584 Rev A1  
52  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
If the EXIT key is pressed before the ENTR key, the analyzer will beep alerting the user  
that the newly entered value has been lost.  
6.4.1. SETUP Mode Password Security  
Whenever the Model GFC7000E’s SETUP mode is activated the instrument will prompt the user to  
enter a security password. The default password is 818. This allows access to all of the  
instruments basic functions and operating modes as well as some of its more powerful diagnostic  
tools and variables.  
The analyzer will automatically insert 818 into the password prompt field. Simply press ENTR to  
proceed.  
Other password levels exist allowing access to special diagnostic tools and variables used only for  
specific and rarely needed troubleshooting and adjustment procedures. They may be made  
available as needed by Teledyne Instruments’ Customer Service department.  
6.5. SETUP – CFG: Viewing the Analyzer’s  
Configuration Information  
Pressing the CFG key displays the instrument configuration information. This display lists the  
analyzer model, serial number, firmware revision, software library revision, CPU type and other  
information. Use this information to identify the software and hardware when contacting customer  
service. Special instrument or software features or installed options may also be listed here.  
SAMPLE*  
RANGE = 500.000 PPB  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
Press NEXT of PREV to move back  
and forth through the following list  
of Configuration information:  
MODEL NAME  
SAMPLE  
PRIMARY SETUP MENU  
Press EXIT at  
any time to  
return to the  
SERIAL NUMBER  
SOFTWARE REVISION  
LIBRARY REVISION  
CFG DAS RNGE PASS CLK MORE  
EXIT  
iCHIP SOFTWARE REVISION1  
SAMPLE display  
HESSEN PROTOCOL REVISION1  
ACTIVE SPECIAL SOFTWARE  
SAMPLE  
GFC7000E CO2 ANALYZER  
OPTIONS1  
Press EXIT at  
any time to  
return to  
CPU TYPE  
DATE FACTORY CONFIGURATION  
SAVED  
NEXT PREV  
EXIT  
SETUP menu  
1Only appears if relevant option of Feature is active.  
04584 Rev A1  
53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.6. SETUP – CLK: Setting the Internal Time-of-Day  
Clock  
The MGFC7000E has a time of day clock that supports the AutoCal timer, time of day TEST  
function, and time stamps on most COM port messages. To set the time-of-day, press:  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
SETUP X.X  
TIME-OF-DAY CLOCK  
Enter Current  
Time-of-Day  
Enter Current  
Date-of-Year  
TIME DATE  
EXIT  
SETUP X.X  
DATE: 01-JAN-02  
SETUP X.X  
TIME: 12:00  
0
1
JAN  
0
2
ENTR EXIT  
1
2
: 0  
0
ENTR EXIT  
SETUP X.X  
JAN  
DATE: 01-JAN-02  
SETUP X.X3  
: 0  
TIME: 12:00  
0
1
0
2
ENTR EXIT  
1
2
0
ENTR EXIT  
SETUP X.X  
TIME-OF-DAY CLOCK  
TIME DATE  
SETUP X.X  
EXIT  
EXIT returns  
to the main  
SAMPLE display  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
04584 Rev A1  
54  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
In order to compensate for CPU clocks which run faster or slower, you can adjust a variable called  
CLOCK_ADJ to speed up or slow down the clock by a fixed amount every day. To change this  
variable, press:  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUPX.X  
1 ) CONC_PRECISION = 3  
EDIT PRNT EXIT  
< TST TST > CAL  
SETUP  
PREV NEXT JUMP  
SAMPLE  
ENTER SETUP PASS : 818  
Continue to press NEXT until …  
8
1
8
ENTR EXIT  
SETUP X.X  
4) CLOCK_ADJ=0 Sec/Day  
JUMP EDIT PRNT EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
PREV  
SETUP X.X  
CLOCK_ADJ:0 Sec/Day  
ENTR EXIT  
SETUP X.X  
SECONDARY SETUP MENU  
+
0
0
COMM VARS DIAG  
EXIT  
Enter sign and number of seconds per  
day the clock gains (-) or loses (+).  
SETUP X.X  
0 ) DAS_HOLD_OFF=15.0 Minutes  
SETUP X.X  
4) CLOCK_ADJ=0 Sec/Day  
NEXT JUMP  
EDIT PRNT EXIT  
PREV NEXT JUMP  
EDIT PRNT EXIT  
3x EXIT returns  
to the main SAMPLE display  
04584 Rev A1  
55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.7. SETUP – RNGE: Analog Output Reporting Range  
Configuration  
The analyzer has three active analog output signals, accessible through a connector on the rear  
panel.  
ANALOG OUT  
SO2 concentration  
Test Channel  
outputs  
Not Used  
A1  
A2  
A3  
A4  
+
-
+
-
+
-
+
-
LOW range when  
DUAL mode is selected  
HIGH range when  
DUAL mode is selected  
Figure 6-4:  
Analog Output Connector Pin Out  
All three outputs can be configured either at the factory or by the user for full scale outputs of 0.1  
VDC, 1VDC, 5VDC or 10VDC. Additionally A1 and A2 may be equipped with optional 0-20 mADC  
current loop drivers and configured for any current output within that range (e.g. 0-20, 2-20, 4-  
20, etc.). The user may also adjust the signal level and scaling of the actual output voltage or  
current to match the input requirements of the recorder or datalogger (see Section 6.9.4).  
The A1 and A2 channels output a signal that is proportional to the CO2 concentration of the  
sample gas. Several modes are available which allow them to operate independently or be slaved  
together (see Section 6.7). The user may also select between a variety of reporting range spans  
(see Sections 6.7.3, 6.7.4 and 6.7.5).  
EXAMPLE:  
A1 OUTPUT: Output Signal = 0-5 VDC representing 0-1000 ppm concentration values  
A2 OUTPUT: Output Signal = 0 – 10 VDC representing 0-500 ppm concentration values.  
The output, labeled A3 is special. It can be set by the user (see Section 6.9.9) to output several  
of the test functions accessible through the <TST TST> keys of the units sample display.  
Output A4 is not available on the Model GFC7000E analyzer.  
6.7.1. Physical Range versus Analog Output Reporting Ranges  
Functionally, the Model GFC7000E Gas Filter Correlation CO2 Analyzer has one hardware Physical  
Range that is capable of determining CO2 concentrations between 50 ppb and 2 000 ppm. This  
architecture improves reliability and accuracy by avoiding the need for extra, switchable, gain-  
amplification circuitry. Once properly calibrated, the analyzer’s front panel will accurately report  
concentrations along the entire span of its 50 ppb and 2 000 ppm physical range.  
Because, most applications use only a small part of the analyzer’s physical range, the width of the  
Model GFC7000E’s physical range can create data resolution problems for most analog recording  
devices. For example, in an application where the expected concentration of CO2 is typically less  
04584 Rev A1  
56  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
than 500 ppm, the full scale of expected values is only 25% of the instrument’s 2 000 ppm  
physical range. Unmodified, the corresponding output signal would also be recorded across only  
25% of the range of the recording device.  
The MGFC7000E solves this problem by allowing the user to select a scaled reporting range for the  
analog outputs that only includes that portion of the physical range relevant to the specific  
application. Only the reporting range of the analog outputs is scaled, the physical range of the  
analyzer and the readings displayed on the front panel remain unaltered.  
6.7.2. Reporting Range Modes  
The MGFC7000E provides three analog output range modes to choose from.  
Single range (SNGL) mode sets a single maximum range for the analog output. If single  
range is selected (see Section 6.7.3) both outputs are slaved together and will represent  
the same measurement span (e.g. 0-50 ppm), however their electronic signal levels may  
be configured for different ranges (e.g. 0-10 VDC vs. 0-.1 VDC – See Section 6.9.4.1).  
Dual range (DUAL) allows the A1 and A2 outputs to be configured with different  
measurement spans (see Section 6.7.4) as well as separate electronic signal levels (see  
Section 6.9.4.1).  
Auto range (AUTO) mode gives the analyzer to ability to output data via a low range and  
high range. When this mode is selected (see Section 6.7.5) the MGFC7000E will  
automatically switch between the two ranges dynamically as the concentration value  
fluctuates.  
Range status is also output via the External Digital I/O Status Bits (see Section 6.13.1.1).  
To select the Analog Output Range Type press:  
04584 Rev A1  
57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SAMPLE  
ENTER SETUP PASS : 818  
8 ENTR EXIT  
8
1
SETUP X.X  
CFG DAS RNGE PASS CLK MORE  
EXIT  
SETUP X.X  
RANGE CONTROL MENU  
MODE SET UNIT  
EXIT  
SETUP X.X  
RANGE MODE: SNGL  
EXIT Returns  
to the Main  
SNGL DUAL AUTO  
ENTR EXIT  
SAMPLE Display  
Only one of the  
range modes may  
be active at any  
time.  
Go To  
Section  
6.7.3  
Go To  
Section  
6.7.4  
Go To  
Section  
6.7.5  
NOTE  
Upper span limit setting for the individual range modes are shared. Resetting the span  
limit in one mode also resets the span limit for the corresponding range in the other  
modes as follows:  
SNGL  
Range  
DUAL  
AUTO  
ÅÆ Range1 ÅÆ Low Range  
Range2 ÅÆ High Range  
6.7.3. Single Range mode (SNGL)  
This is the default reporting range mode for the analyzer. In single range mode both A1 and A2  
are set to the same reporting range. This reporting range can be any value between 50 ppb and 2  
000 ppm.  
While the two outputs always have the same reporting range, the span, signal offset and scaling  
of their electronic signals may be configured for differently (e.g., A1 = 0-10 V; A2 = 0-0.1 V).  
See Section 6.9.4 for instructions on adjusting these parameters.  
To select SNGLE range mode and to set the upper limit of the range, press:  
04584 Rev A1  
58  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 X.XXX  
< TST TST > CAL  
SETUP  
SETUP C.3  
RANGE MODE: SNGL  
SAMPLE  
ENTER SETUP PASS : 818  
SNGL DUAL AUTO  
ENTR EXIT  
8
1
8
ENTR EXIT  
SETUP C.3  
RANGE CONTROL MENU  
SETUP C.3  
PRIMARY SETUP MENU  
MODE SET UNIT  
EXIT  
CFG DAS RNGE PASS CLK MORE  
EXIT  
SETUP C.3  
RANGE CONTROL MENU  
SETUP C.3  
RANGE: 500.0 Conc  
.0  
MODE SET UNIT  
EXIT  
0
0
5
0
0
ENTR EXIT  
SETUP C.3  
RANGE MODE: SNGL  
SETUP C.3  
MODE SET UNIT  
RANGE CONTROL MENU  
EXIT x 2 returns  
to the main  
SAMPLE display  
SNGL DUAL AUTO  
ENTR EXIT  
EXIT  
04584 Rev A1  
59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.7.4. Dual Range Mode (DUAL)  
Selecting Dual Range mode allows the A1 and A2 outputs to be configured with different  
reporting ranges. The analyzer software calls these two ranges low and high. The low range  
setting corresponds with the analog output labeled A1 on the Rear Panel of the instrument. The  
high Range Setting corresponds with the A2 output. While the software names these two ranges  
low and high, they do not have to be configured that way. For example: The low range can be set  
for a span of 0-1000 ppm while the high range is set for 0-500 ppm.  
In DUAL range mode the RANGE test function displayed on the front panel will be replaced by  
two separate functions:  
RANGE1: The range setting for the A1 output.  
RANGE2: The range setting for the A2 output.  
To set the ranges press following keystroke sequence  
.
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP X.X  
RANGE MODE: DUAL  
< TST TST > CAL  
SETUP  
SNGL DUAL AUTO  
ENTR EXIT  
SAMPLE  
ENTER SETUP PASS : 818  
SETUP X.X  
RANGE CONTROL MENU  
8
1
8
ENTR EXIT  
MODE SET UNIT  
EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
SETUP X.X  
LOW RANGE: 500.0 Conc  
.0 ENTR EXIT  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
Toggle the  
Numeral Keys  
to set the upper  
limit of each  
range.  
0
0
1
5
0
0
SETUP X.X  
RANGE CONTROL MENU  
SETUP X.X  
HIGH RANGE: 500.0 Conc  
.0 ENTR EXIT  
MODE SET UNIT  
0
0
0
0
SETUP X.X  
RANGE MODE: SNGL  
SETUP X.X  
RANGE CONTROL MENU  
EXIT Returns  
to the Main  
SAMPLE Display  
SNGL DUAL AUTO  
ENTR EXIT  
MODE SET UNIT  
EXIT  
When the instrument’s range mode is set to DUAL the concentration field in the upper right hand  
corner of the display alternates between displaying the low range value and the high range value.  
The concentration currently being displayed is identified as follows: C1 = Low (or A1) and C2 =  
High (or A2).  
NOTE  
In DUAL range mode the LOW and HIGH ranges have separate slopes and offsets for  
computing CO2 concentration.  
The two ranges must be independently calibrated.  
04584 Rev A1  
60  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.7.5. Auto Range Mode (AUTO)  
In AUTO range mode, the analyzer automatically switches the reporting range between two user-  
defined ranges (low and high). The unit will switch from low range to high range when the CO2  
concentration exceeds 98% of the low range span. The unit will return from high range back to  
low range once both the CO2 concentration falls below 75% of the low range span.  
In AUTO Range mode the instrument reports the same data in the same range on both the A1  
and A2 outputs and automatically switches both outputs between ranges as described above.  
Also, the RANGE test function displayed on the front panel will automatically switch to show  
which range is in effect.  
The high/low range status is also reported through the external, digital status bits (Section  
6.13.1.1).  
To set individual ranges press the following keystroke sequence.  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP X.X  
RANGE MODE: AUTO  
< TST TST > CAL  
SETUP  
SNGL DUAL AUTO  
ENTR EXIT  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
SETUP X.X  
RANGE CONTROL MENU  
EXIT x 2 returns  
to the main  
SAMPLE display  
MODE SET UNIT  
EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP X.X  
LOW RANGE: 500.0 Conc  
.0 ENTR EXIT  
Toggle the numeral  
keys to set the  
LOW and HIGH  
range value.  
SETUP X.X  
RANGE CONTROL MENU  
0
0
5
5
0
0
MODE SET UNIT  
ENTR accepts the  
new setting, EXIT  
ignores the new  
setting.  
SETUP X.X  
RANGE MODE: SNGL  
SETUP X.X  
HIGH RANGE: 500.0 Conc  
.0 ENTR EXIT  
SNGL DUAL AUTO  
ENTR EXIT  
0
0
0
0
CAUTION  
In AUTO range mode the LOW and HIGH ranges have separate slopes and offsets for  
computing CO2 concentration.  
The two ranges must be independently calibrated.  
NOTE  
Avoid accidentally setting the low range of the instrument with a higher span limit than  
the high range. This will cause the unit to stay in the low reporting range perpetually  
and defeat the function of the AUTO range mode.  
04584 Rev A1  
61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.7.6. Range Units  
The MGFC7000E can display concentrations in parts per billion (109 mols per mol, PPB), parts per  
million (106 mols per mol, PPM), micrograms per cubic meter (µg/m3, UG), milligrams per cubic  
meter (mg/m3, MG) or percent (volume CO2/volume sample gas, %). Changing units affects all of  
the display, analog outputs, COM port and iDAS values for all reporting ranges regardless of the  
analyzer’s range mode.  
NOTE  
Concentrations displayed in mg/m3 and ug/m3 use 0°C, 760 mmHg for Standard  
Temperature and Pressure (STP). Consult your local regulations for the STP used by  
your agency.  
Conversion factors from volumetric to mass units are:  
CO2: ppb x 1.96 = µg/m3; ppm x 1.96 = mg/m3  
To change the concentration units:  
SAMPLE  
RANGE = 500.00 PPB  
CO2=X.XXX  
< TST TST > CAL  
SETUP  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
EXIT returns  
to the main menu.  
SETUP X.X  
RANGE CONTROL MENU  
MODE SET UNIT  
SETUP X.X  
CONC UNITS: PPM  
Select the preferred  
concentration unit.  
PPM PPB UGM MGM  
%
ENTER EXIT  
ENTER EXIT  
ENTR accepts  
the new unit,  
EXIT returns  
to the SETUP  
menu.  
SETUP X.X  
CONC UNITS: %  
PPM PPB UGM MGM  
%
NOTE  
Once the units of measurement have been changed the unit MUST be recalibrated, as  
the “expected span values” previously in effect will no longer be valid. Simply entering  
new expected span values without running the entire calibration routine is not  
sufficient.  
04584 Rev A1  
62  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.7.7. Dilution Ratio  
The dilution ratio is a software option that allows the user to compensate for any dilution of the  
sample gas before it enters the sample inlet. Using the dilution ratio option is a 4-step process:  
1. Select reporting range units: Follow the procedure in Section 6.7.6.  
2. Select the range: Use the procedures in Section 6.7.2 – 6.7.5. Make sure that the SPAN value  
entered is the maximum expected concentration of the undiluted calibration gas and that the  
span gas is either supplied through the same dilution inlet system as the sample gas or has an  
appropriately lower actual concentration. For example, with a dilution set to 100, a 10 ppm  
gas can be used to calibrate a 1000 ppm sample gas if the span gas is not routed through the  
dilution system. On the other hand, if a 1000 ppm span gas is used, it needs to pass through  
the same dilution steps as the sample gas.  
3. Set the dilution factor as a gain (e.g., a value of 20 means 20 parts diluting gas and 1 part of  
sample gas):  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
SETUP C.3  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
EXIT  
EXIT  
SETUP C.3  
RANGE CONTROL MENU  
DIL only appears  
if the dilution ratio  
option has been  
installed  
MODE SET UNIT DIL  
EXIT ignores the  
new setting.  
SETUP C.3  
DIL FACTOR: 1.0 GAIN  
.0 ENTR  
ENTR accepts the  
Toggle these keys to set the dilution  
factor.  
new setting.  
0
0
0
1
This is the number by which the  
analyzer will multiply the CO2  
concentrations of the gas passing  
through the reaction cell.  
SETUP C.3  
DIL FACTOR: 20.0 GAIN  
.0 ENTR  
0
0
2
0
The analyzer multiplies the measured gas concentrations with this dilution factor and displays the  
result.  
NOTE  
Once the above settings have been entered, the instrument needs to be recalibrated  
using one of the methods discussed in Chapter 7.  
04584 Rev A1  
63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.8. SETUP – VARS: Using the Internal Variables  
The MGFC7000E has several-user adjustable software variables, which define certain operational  
parameters. Usually, these variables are automatically set by the instrument’s firmware, but can  
be manually re-defined using the VARS menu. Table 6-6 lists all variables that are available within  
the 818 password protected level.  
Table 6-6:  
Variable Names (VARS) Revision B.3  
Description  
No.  
0
Variable  
Allowed Values  
Changes the internal data acquisition system (iDAS)  
hold-off time, which is the duration when data are  
not stored in the iDAS because the software  
considers the data to be questionable. That is the  
case during warm-up or just after the instrument  
returns from one of its calibration modes to SAMPLE  
mode. DAS_HOLD_OFF can be disabled entirely in  
each iDAS channel.  
Can be between  
0.5 and 20  
minutes  
DAS_HOLD_OFF  
Default=15 min.  
Allows the user to set the number of significant digits  
CONC_PRECISION to the right of the decimal point display of  
AUTO, 1, 2, 3, 4  
Default=AUTO  
1
2
concentration and stability values.  
Dynamic zero automatically adjusts offset and slope  
of the CO2 response when performing a zero point  
calibration during an AutoCal (Chapter 7).  
DYN_ZERO  
DYN_SPAN  
CLOCK_ADJ  
ON/OFF  
ON/OFF  
Dynamic span automatically adjusts slope and slope  
of the CO2 response when performing a zero point  
calibration during an AutoCal (Chapter 7).  
3
4
Note that the DYN_ZERO and DYN_SPAN features  
are not allowed for applications requiring EPA  
equivalency.  
Adjusts the speed of the analyzer’s clock. Choose the  
+ sign if the clock is too slow, choose the - sign if  
the clock is too fast.  
-60 to +60 s/day  
04584 Rev A1  
64  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
To access and navigate the VARS menu, use the following key sequence.  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
EXIT ignores the new setting.  
ENTR accepts the new setting.  
SETUP X.X  
SECONDARY SETUP MENU  
COMM VARS DIAG ALRM  
SETUP X.X  
0 ) DAS_HOLD_OFF=15.0 Minutes  
SETUP X.X  
.0  
DAS_HOLD_OFF=15.0 Minutes  
ENTR EXIT  
NEXT JUMP  
EDIT PRNT EXIT  
1
5
Toggle this keys to change setting  
SETUP X.X  
6) CONC_PRECUISION : 3  
SETUP X.X  
CONC_PRECUISION : 3  
PREV NEXT JUMP  
EDIT PRNT EXIT  
AUTO  
0
1
2
3
4
ENTR EXIT  
Toggle these keys to change setting  
SETUP X.X  
4 ) DYN_ZERO=ON  
SETUP X.X  
DYN_ZERO=ON  
PREV NEXT JUMP  
EDIT PRNT EXIT  
ON  
ENTR EXIT  
Toggle this keys to change setting  
SETUP X.X  
5) DYN_SPAN=ON  
PREV NEXT JUMP  
EDIT PRNT EXIT  
SETUP X.X  
DYN_SPAN=ON  
ENTR EXIT  
ON  
Toggle this keys to change setting  
SETUP X.X  
7) CLOCK_ADJ=0 Sec/Day  
SETUP X.X  
CLOCK_ADJ=0 Sec/Day  
ENTR EXIT  
PREV NEXT JUMP  
EDIT PRNT EXIT  
+
0
0
Toggle tese keys to change setting  
04584 Rev A1  
65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.9. SETUP – DIAG: Using the Diagnostics Functions  
A series of diagnostic tools is grouped together under the SETUPÆMOREÆDIAG menu. As these  
parameters are dependent on firmware revision (see Menu Tree A-6 in Appendix A). The individual  
parameters, however, are explained in more detail in the section indicated in Table 6-7. These  
tools can be used in a variety of troubleshooting and diagnostic procedures and are referred to in  
many places of the maintenance and trouble-shooting sections.  
Table 6-7:  
GFC7000E Diagnostic (DIAG) Functions  
Front Panel  
Mode  
DIAGNOSTIC FUNCTION AND MEANING  
SECTION  
Indicator  
SIGNAL I/O: Allows observation of all digital and analog  
signals in the instrument. Allows certain digital signals  
such as valves and heaters to be toggled ON and OFF.  
DIAG I/O  
6.9.2  
ANALOG OUTPUT: When entered, the analyzer performs  
an analog output step test. This can be used to calibrate  
a chart recorder or to test the analog output accuracy.  
DIAG AOUT  
DIAG AIO  
6.9.3  
6.9.4  
ANALOG I/O CONFIGURATION: the signal levels of the  
instruments analog outputs may be calibrated (either  
individually or as a group). Various electronic  
parameters such as signal span, and offset are available  
for viewing and configuration.  
ELECTRIC TEST: The analyzer is performing an electric  
test. This test simulates IR detector signal in a known  
manner so that the proper functioning of the  
sync/demod board can be verified.  
DIAG  
OPTIC  
6.9.5  
6.9.6  
6.9.7  
DARK CALIBRATION: The analyzer is performing a dark  
calibration procedure. This procedure measures and  
stores the inherent dc offset of the sync/demod board  
electronics.  
DIAG ELEC  
DIAG PCAL  
PRESSURE CALIBRATION: The analyzer records the  
current output of the sample gas pressure sensor. This  
value is used by the CPU to compensate the CO2  
concentration.  
FLOW CALIBRATION: This function is used to calibrate  
the gas flow output signals of sample gas and ozone  
supply. These settings are retained when exiting DIAG.  
DIAG FCAL  
DIAG TCHN  
6.9.8  
6.9.9  
TEST CHAN OUTPUT: Configures the A4 analog output  
channel.  
04584 Rev A1  
66  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.9.1. Accessing the Diagnostic Features  
To access the DIAG functions press the following keys:  
SAMPLE*  
RANGE = 500.00 PPM  
CO2 =X.XXX  
DIAG  
ANALOG I / O CONFIGURATION  
< TST TST > CAL  
SETUP  
PREV  
NEXT  
ENTR EXIT  
ENTR EXIT  
ENTR EXIT  
ENTR EXIT  
ENTR EXIT  
ENTR EXIT  
SAMPLE  
ENTER SETUP PASS : 818  
DIAG  
ELECTRICAL TEST  
DARK CALIBRATION  
8
1
8
ENTR EXIT  
PREV  
NEXT  
NEXT  
SETUP X.X  
PRIMARY SETUP MENU  
EXIT returns  
to the main  
SAMPLE  
display  
DIAG  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
PREV  
SETUP X.X  
SECONDARY SETUP MENU  
EXIT returns  
to the PRIMARY  
SETUP MENU  
DIAG  
PRESSURE CALIBRATION  
NEXT  
COMM VARS DIAG ALRM  
PREV  
DIAG  
SIGNAL I / O  
From this point  
forward, EXIT returns  
to the  
SECONDARY  
SETUP MENU  
DIAG  
FLOW CALIBRATION  
TEST CHAN OUTPUT  
NEXT  
NEXT  
ENTR EXIT  
ENTR EXIT  
PREV  
NEXT  
NEXT  
DIAG  
ANALOG OUTPUT  
DIAG  
PREV  
PREV  
6.9.2. Signal I/O  
The signal I/O diagnostic mode allows to review and change the digital and analog input/output  
functions of the analyzer. See Appendix A-4 for a complete list of the parameters available for  
review under this menu.  
NOTE  
Any changes of signal I/O settings will remain in effect only until the signal I/O menu is  
exited. Exceptions are the ozone generator override and the flow sensor calibration,  
which remain as entered when exiting.  
04584 Rev A1  
67  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
To enter the signal I/O test mode, press:  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
DIAG  
SIGNAL I / O  
Use the  
NEXT PREV  
&
keys to move between  
signal types.  
< TST TST > CAL  
SETUP  
PREV NEXT JUMP  
ENTR EXIT  
Use the JUMP key to  
go directly to a  
DIAG I / O  
Test Signals Displayed Here  
SAMPLE  
ENTER SETUP PASS : 818  
specific signal  
PREV NEXT JUMP  
PRNT EXIT  
8
1
8
ENTR EXIT  
See Appendix A-4 for  
a complete list of  
EXIT returns  
to the main  
SAMPLE display  
available SIGNALS  
EXAMPLE  
DIAG I / O  
JUMP TO: 12  
SETUP X.X  
PRIMARY SETUP MENU  
EXAMPLE:  
1
2
ENTR EXIT  
Enter 12 to Jump to  
12) ST_CONC_VALID  
CFG DAS RNGE PASS CLK MORE  
EXIT  
DIAG I / O  
ST_CONC_VALID = ON  
Exit to return  
to the  
SETUP X.X  
SECONDARY SETUP MENU  
DIAG menu  
PREV NEXT JUMP  
ON PRNT EXIT  
COMM VARS DIAG ALRM  
EXIT  
Pressing the PRNT key will send a formatted printout to the serial port and can be  
captured with a computer or other output device.  
6.9.3. Analog Output Step Test  
This test can be used to check the accuracy and proper operation of the analog outputs. The test  
forces all four analog output channels to produce signals ranging from 0% to 100% of the full  
scale range in 20% increments. This test is useful to verify the operation of the data  
logging/recording devices attached to the analyzer.  
To begin the Analog Output Step Test press:  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
DIAG  
SIGNAL I / O  
< TST TST > CAL  
SETUP  
NEXT  
ENTR EXIT  
SAMPLE  
ENTER SETUP PASS : 818  
DIAG  
ANALOG OUTPUT  
8
1
8
ENTR EXIT  
PREV  
NEXT  
ENTR EXIT  
Performs  
analog output  
step test.  
DIAG AOUT  
0%  
ANALOG OUTPUT  
ANALOG OUTPUT  
SETUP X.X  
PRIMARY SETUP MENU  
0% - 100%  
EXIT  
CFG DAS RNGE PASS CLK MORE  
EXIT  
Exit-Exit  
returns to the  
DIAG AOUT  
SETUP X.X  
SECONDARY SETUP MENU  
DIAG menu  
[0%]  
EXIT  
COMM VARS DIAG ALRM  
EXIT  
Pressing the key under “0%” while performing the test will  
pause the test at that level. Brackets will appear around  
the value: example: [20%] Pressing the same key again  
will resume the test.  
04584 Rev A1  
68  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.9.4. Analog I/O Configuration  
The analog I/O functions that are available in the MGFC7000E are:  
Table 6-8:  
DIAG - Analog I/O Functions  
Function  
Sub Menu  
AOUTS CALIBRATED: Shows the status of the analog output calibration (YES/NO) and initiates a  
calibration of all analog output channels.  
CONC_OUT_1  
Sets the basic electronic configuration of the A1 analog output (CO2 ) . There are  
three options:  
Range: Selects the signal type (voltage or current loop) and full scale level of the  
output.  
REC_OFS: Allows to set a voltage offset (not available when RANGE is set to  
CURRent loop.  
Auto_CAL: Performs the same calibration as AOUT CALIBRATED, but on this one  
channel only.  
NOTE: Any change to RANGE or REC_OFS requires recalibration of this output.  
CONC_OUT_2  
TEST OUTPUT  
Same as for CONC_OUT_1 but for analog channel 2 (CO2 )  
Same as for CONC_OUT_1 but for analog channel 4 (TEST)  
AIN CALIBRATED  
Shows the calibration status (YES/NO) and initiates a calibration of the analog to  
digital converter circuit on the motherboard.  
To configure the analyzer’s three analog outputs, set the electronic signal type of each channel  
and calibrate the outputs. This consists of:  
Selecting an output type (voltage or current, if an optional current output driver has been  
installed) and the signal level that matches the input requirements of the recording device  
attached to the channel, see Sections 6.9.4.1.  
Calibrating the output channel. This can be done automatically or manually for each  
channel, see Sections 6.9.4.2 and 6.9.4.3.  
Adding a bipolar recorder offset to the signal, if required (Section 6.9.4.2.)  
In its standard configuration, the analyzer’s outputs can be set for the following DC voltages. Each  
range is usable from -5% to + 5% of the nominal range.  
Table 6-9:  
Analog Output Voltage Ranges  
RANGE  
0-0.1 V  
0-1 V  
MINIMUM OUTPUT  
-5 mV  
MAXIMUM OUTPUT  
+105 mV  
-0.05 V  
+1.05 V  
0-5 V  
-0.25 V  
+5.25 V  
0-10 V  
-0.5 V  
+10.5 V  
The default offset for all ranges is 0 VDC.  
04584 Rev A1  
69  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
The following DC current output limits apply to the current loop modules:  
Table 6-10: Analog Output Current Loop Range  
RANGE  
MINIMUM OUTPUT  
MAXIMUM OUTPUT  
0-20 mA  
0 mA  
20 mA  
These are the physical limits of the current loop modules, typical  
applications use 2-20 or 4-20 mA for the lower and upper limits.  
Please specify desired range when ordering this option.  
The default offset for all ranges is 0 mA.  
Pin assignments for the output connector at the rear panel of the instrument are shown in Table  
6-11.  
ANALOG OUT  
A1  
A2  
A3  
A4  
+
-
+
-
+
-
+
-
Table 6-11: Analog Output Pin Assignments  
PIN  
ANALOG  
OUTPUT  
VOLTAGE  
SIGNAL  
CURRENT  
SIGNAL  
1
2
V Out  
I Out +  
A1  
A2  
Ground  
I Out -  
3
4
V Out  
I Out +  
I Out -  
Ground  
5
6
V Out  
I Out +  
I Out -  
A4  
A3  
Ground  
7 & 8  
Not Used  
Not Used  
See Figure 3-2 for a the location of the analog output connector on the instruments rear panel.  
04584 Rev A1  
70  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.9.4.1. Analog Output Signal Type and Range Span Selection  
To select an output signal type (DC Voltage or current) and level for one output channel, activate  
the ANALOG I/O CONFIGURATION MENU (see Section 6.9.1) then press:  
FROM ANALOG I/O CONFIGURATION MENU  
DIAG  
ANALOG I / O CONFIGURATION  
PREV  
NEXT  
ENTR  
EXIT  
EXIT  
DIAG AIO  
AOUTS CALIBRATED: NO  
Press SET> to select the  
analog output channel to be  
configured. Press EDIT to  
continue  
< SET SET> CAL  
DIAG AIO  
CONC_OUT_2:5V, CAL  
< SET SET> EDIT  
EXIT  
EXIT  
DIAG AIO  
CONC_OUT_2 RANGE: 5V  
SET> EDIT  
DIAG AIOOUTPUT RANGE: 5V  
These keys set  
the signal level  
and type of the  
selected channel  
0.1V 1V 5V 10V CURR  
ENTR EXIT  
Pressing ENTR records the new setting  
and returns to the previous menu.  
Pressing EXIT ignores the new setting and  
returns to the previous menu.  
DIAG AIOOUTPUT RANGE: 10V  
0.1V 1V 5V 10V CURR  
ENTR EXIT  
6.9.4.2. Analog Output Calibration Mode  
The analog outputs can be calibrated automatically or manually. In its default mode, the  
instrument is configured for automatic calibration of all channels. Manual calibration should be  
used for the 0.1V range or in cases where the outputs must be closely matched to the  
characteristics of the recording device. Outputs configured for automatic calibration can be  
calibrated as a group or individually. Calibration of the analog outputs needs to be carried out on  
first startup of the analyzer (performed in the factory as part of the configuration process) or  
whenever re-calibration is required.  
04584 Rev A1  
71  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
To calibrate the outputs as a group, activate the ANALOG I/O CONFIGURATION MENU (see  
Section 6.9.1), then press:  
STARTING FROM DIAGNOSTIC MENU  
(see Section 6.7.1)  
DIAG  
ANALOG I / O CONFIGURATION  
ENTR EXIT  
Exit at any time  
to return to the  
main DIAG  
menu  
PREV  
NEXT  
DIAG AIO  
AOUTS CALIBRATED: NO  
If AutoCal has been  
turned off for any  
channel, the message  
for that channel will be  
similar to:  
< SET SET> CAL  
EXIT  
DIAG AIO AUTO CALIBRATING CONC_OUT_1  
AUTO CALIBRATING CONC_OUT_2  
NOT AUTO CAL  
CONC_OUT_1  
AUTO CALIBRATING TEST_OUTPUT  
If any of the channels have  
not been calibrated this  
message will read NO.  
Exit to return to  
the I/O  
configuration  
menu  
DIAG AIO  
AOUTS CALIBRATED:  
YES  
< SET SET> CAL  
EXIT  
To automatically calibrate a single analog channel, activate the ANALOG I/O  
CONFIGURATION MENU (see Section 6.9.1), then press:  
DIAG  
ANALOG I / O CONFIGURATION  
NEXT ENTR EXIT  
EXIT to Return  
to the main  
Sample Display  
PREV  
DIAG AIO  
AOUTS CALIBRATED: NO  
<
SET> CAL  
EXIT  
Press SET> to select the  
Analog Output channel to  
be configured. Then Press  
EDIT to continue  
DIAG AIO  
CONC_OUT_2:5V, CAL  
< SET SET> EDIT  
EXIT  
EXIT  
DIAG AIO  
CONC_OUT_2 RANGE: 5V  
DIAG AIO  
CONC_OUT_2 CALIBRATED: NO  
CAL EXIT  
SET> EDIT  
<SET  
DIAG AIO  
CONC_OUT_2 REC OFS: 0 mV  
DIAG AIO  
AUTO CALIBRATING CONC_OUT_2  
< SET SET> EDIT  
EXIT  
DIAG AIO  
CONC_OUT_2 AUTO CAL: ON  
DIAG AIO  
<SET  
CONC_OUT_2 CALIBRATED: YES  
CAL EXIT  
< SET SET> EDIT  
EXIT  
04584 Rev A1  
72  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
To select manual output calibration for a particular channel, activate the ANALOG I/O  
CONFIGURATION MENU (see Section 6.9.1), then press:  
DIAG  
ANALOG I / O CONFIGURATION  
ENTR EXIT  
DIAG AIO  
CONC_OUT_2 REC OFS: 0 mV  
Exit to return to  
the main  
sample display  
PREV  
NEXT  
< SET SET> EDIT  
EXIT  
DIAG AIO  
AOUTS CALIBRATED: NO  
DIAG AIO  
CONC_OUT_2 AUTO CAL: ON  
< SET SET> CAL  
EXIT  
< SET SET> EDIT  
EXIT  
Press SET> to select the analog output channel to  
be configured. Then press EDIT to continue  
DIAG AIO  
CONC_OUT_2 AUTO CAL: ON  
DIAG AIO  
CONC_OUT_2:5V, CAL  
ON  
ENTR EXIT  
< SET SET> EDIT  
EXIT  
EXIT  
Toggles the  
auto cal mode  
ON/ OFF for  
this analog  
output channel  
only.  
ENTR accepts the new setting  
and returns to the previous menu.  
EXIT ignores the new setting and  
returns to the previous menu.  
DIAG AIO  
CONC_OUT_2 RANGE: 5V  
SET> EDIT  
Now the analog output channels should either be automatically calibrated or they should be set to  
manual calibration, which is described next.  
6.9.4.3. Manual Analog Output Calibration and Voltage Adjustment  
For highest accuracy, the voltages of the analog outputs can be manually calibrated. Calibration is  
done through the instrument software with a voltmeter connected across the output terminals  
(Figure 6-5). Adjustments are made using the front panel keys by setting the zero-point first and  
then the span-point (Table 6-12).  
The software allows this adjustment to be made in 100, 10 or 1 count increments.  
Table 6-12: Voltage Tolerances for Analog Output Calibration  
Full Scale  
0.1 VDC  
1 VDC  
Zero Tolerance  
±0.0005V  
±0.001V  
Span Voltage  
90 mV  
Span Tolerance  
±0.001V  
900 mV  
±0.001V  
5 VDC  
±0.002V  
4500 mV  
4500 mV  
±0.003V  
10 VDC  
±0.004V  
±0.006V  
NOTE  
Outputs configured for 0.1V full scale should always be calibrated manually  
04584 Rev A1  
73  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
See Table 3-1 for  
pin assignments  
of Analog Out  
connector on the  
rear panel  
V
+DC Gnd  
V OUT +  
V OUT -  
V IN +  
V IN -  
Recording  
Device  
ANALYZER  
Figure 6-5:  
Setup for Calibrating Analog Voltage Outputs  
To make these adjustments, the AOUT auto-calibration feature must be turned off (Section  
6.9.4.2). Activate the ANALOG I/O CONFIGURATION MENU (see Section 6.9.1), then press:  
FROM ANALOG I/O CONFIGURATION MENU  
DIAG AIO  
CONC_OUT_1 RANGE: 5V  
DIAG  
ANALOG I / O CONFIGURATION  
ENTR EXIT  
SET> EDIT  
EXIT  
PREV  
NEXT  
DIAG AIO  
CONC_OUT_1 REC OFS: 0 mV  
DIAG AIO  
AOUTS CALIBRATED: NO  
< SET SET> EDIT  
EXIT  
EXIT  
If AutoCal is ON, go to  
Section 6.7.3  
< SET SET> CAL  
EXIT  
DIAG AIO  
CONC_OUT_1 AUTO CAL: OFF  
Press SET> to select the analog output channel to be  
< SET SET> EDIT  
configured:  
DISPLAYED AS=  
CONC_OUT_1 =  
CONC_OUT_2 =  
TEST OUTPUT =  
CHANNEL  
A1  
A2  
A4  
DIAG AIO  
< SET  
CONC_OUT_2 CALIBRATED: NO  
CAL  
EXIT  
DIAG AIO  
CONC_OUT_1 :5V, NO CAL  
DIAG AIO CONC_OUT_1 VOLT–Z : 0 mV  
< SET SET> EDIT  
EXIT  
U100 UP10 UP DOWN DN10 D100 ENTR EXIT  
These keys increase / decrease the analog  
output by 100, 10 or 1 counts.  
EXIT ignores the  
new setting.  
ENTR accepts the  
Continue adjustments until the voltage measured  
at the output of the analyzer and/or the input of  
the recording device matches the value in the  
upper right hand corner of the display to the  
tolerance listed in Table 6-10.  
DIAG AIO CONC_OUT_1 VOLT–S : 4500 mV  
U100 UP10 UP DOWN DN10 D100 ENTR EXIT  
new setting.  
The concentration display will not change. Only  
the voltage reading of your voltmeter will change.  
DIAG AIO  
< SET  
CONC_OUT_1 CALIBRATED: YES  
CAL EXIT  
04584 Rev A1  
74  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.9.4.4. Current Loop Output Adjustment  
A current loop option is available and can be installed as a retrofit for each of the analog outputs  
of the analyzer (Sections 5.2). This option converts the DC voltage analog output to a current  
signal with 0-20 mA output current. The outputs can be scaled to any set of limits within that 0-20  
mA range. However, most current loop applications call for either 2-20 mA or 4-20 mA range. All  
current loop outputs have a +5% over-range. Ranges with the lower limit set to more than 1 mA  
(e.g., 2-20 or 4-20 mA) also have a -5% under-range.  
To switch an analog output from voltage to current loop after installing the current output printed  
circuit assembly, follow the instructions in Section 6.9.4.4 and select CURR from the list of  
options on the RANGE menu.  
Adjusting the signal zero and span values of the current loop output is done by raising or lowering  
the voltage of the respective analog output. This proportionally raises or lowers the current  
produced by the current loop option.  
Similar to the voltage calibration, the software allows this current adjustment to be made in 100,  
10 or 1 count increments. Since the exact current increment per voltage count varies from output  
to output and from instrument to instrument, you will need to measure the change in the current  
with a current meter placed in series with the output circuit (Figure 6-6).  
See Table 3-2 for  
pin assignments of  
mA  
the Analog Out  
connector on the  
rear panel.  
IN  
OUT  
V OUT +  
V OUT -  
I IN +  
I IN -  
Recording  
Device  
Analyzer  
Figure 6-6:  
Setup for Calibrating Current Outputs  
NOTE  
Do not exceed 60 V between current loop outputs and instrument ground.  
To adjust the zero and span values of the current outputs, activate the ANALOG I/O  
CONFIGURATION MENU (see Section 6.9.1), then press:  
04584 Rev A1  
75  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
FROM ANALOG I/O CONFIGURATION MENU  
The instrument attempt to automatically calibrate  
the channel … then beep.  
DIAG  
PREV  
ANALOG I / O CONFIGURATION  
NEXT ENTR EXIT  
DIAG AIO CONC_OUT_2 D/A/ CAL ERROR  
EXIT  
DIAG AIO  
AIN CALIBRATED: NO  
SET> EDIT  
EXIT  
DIAG AIO  
CONC_OUT_2 CURR-Z: 0 mV  
U100 UP10 UP DOWN DN10 D100 ENTR EXIT  
Press SET> to select the analog output channel  
to be configured:.  
DIAG AIO  
CONC_OUT_2 ZERO: 27 mV  
Increase or decrease the current  
output by 100, 10 or 1 counts.  
The resulting change in output  
voltage is displayed in the upper  
line.  
U100 UP10 UP DOWN DN10 D100 ENTR EXIT  
DIAG AIO CONC_OUT_2:CURR, NO CAL  
< SET SET> EDIT  
EXIT  
Continue adjustments until the  
correct current is measured with  
the current meter.  
DIAG AIO  
CONC_OUT_2 SPAN: 10000 mV  
DIAG AIO CONC_OUT_2 RANGE: CURR  
U100 UP10 UP DOWN DN10 D100 ENTR EXIT  
<SET SET> EDIT  
EXIT  
DIAG AIO  
CONC_OUT_2 ZERO: 9731 mV  
EXIT ignores the  
new setting, ENTR  
accepts the new  
setting.  
DIAG AIO CONC_OUT_2 CALIBRATED: NO  
< SET CAL EXIT  
U100 UP10 UP DOWN DN10 D100 ENTR EXIT  
DIAG AIO CONC_OUT_2 CALIBRATED: YES  
DIAG AIO AUTO CALIBRATING CONC_OUT_2  
< SET  
CAL  
EXIT  
If a current meter is not available, an alternative method for calibrating the current loop outputs  
is to connect a 250 Ω ± 1% resistor across the current loop output. Using a voltmeter connected  
across the resistor, follow the procedure above but adjust the output to the following values:  
Table 6-13: Current Loop Output Calibration with Resistor  
Voltage for 2-20 mA  
(measured across resistor)  
Voltage for 4-20 mA  
(measured across resistor)  
Full scale  
0%  
0.5 V  
5.0 V  
1.0 V  
5.0 V  
100%  
04584 Rev A1  
76  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.9.4.5. AIN Calibration  
This is the sub-menu to conduct the analog input calibration. This calibration should only be  
necessary after major repair such as a replacement of CPU, motherboard or power supplies.  
Activate the ANALOG I/O CONFIGURATION MENU (see Section 6.9.1), then press:  
STARTING FROM ANALOG I / O CONFIGURATION MENU  
Exit at any time to  
return to the main  
DIAG menu  
DIAG  
ANALOG I / O CONFIGURATION  
ENTR EXIT  
PREV  
NEXT  
Continue pressing SET? until …  
DIAG AIO  
AIN CALIBRATED: NO  
< SET SET> CAL  
EXIT  
DIAG AIO  
CALIBRATING A/D ZERO  
Instrument  
calibrates  
automatically  
CALIBRATING A/D SPAN  
Exit to return to the  
ANALOG I/O  
CONFIGURATION  
MENU  
DIAG AIO  
AIN CALIBRATED: YES  
< SET SET> CAL  
EXIT  
6.9.5. Electric Test  
The electric test function substitutes simulated signals for CO2 MEAS and CO2 REF, generated by  
circuitry on the sync/demod board, for the output of the IR photo-detector. While in this mode the  
user can also view the same test functions viewable from the main SAMPLE display. When the  
test is running, the concentration reported on the front panel display should be 40.0 ppm.  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
DIAG  
SIGNAL I / O  
< TST TST > CAL  
SETUP  
NEXT  
ENTR  
EXIT  
SAMPLE  
ENTER SETUP PASS : 818  
Repeat Pressing NEXT unti . . .  
8
1
8
ENTR EXIT  
DIAG  
ELECTRIC TEST  
SETUP X.X  
PRIMARY SETUP MENU  
PREV NEXT  
ENTR  
EXIT  
CFG DAS RNGE PASS CLK MORE  
EXIT  
DIAG ELEC  
RANGE=50.000 PPM  
CO2= 40.0  
EXIT  
Exit returns  
to the  
DIAG Menu  
SETUP X.X  
SECONDARY SETUP MENU  
<TST TST>  
COMM VARS DIAG ALRM  
EXIT  
Press <TST TST> to view Test Functions  
NOTE: CO MEAS and CO REF will be artificially altered to  
enforce a CO2 reading of 40.0 ppm.  
All other Test Functions will report the correct operational  
value  
04584 Rev A1  
77  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.9.6. Dark Calibration Test  
The dark calibration test interrupts the signal path between the IR photo-detector and the  
remainder of the sync/demod board circuitry. This allows the instrument to compensate for any  
voltage levels inherent in the sync/demod circuitry that might effect the calculation of CO2  
concentration. Performing this calibration returns two offset voltages, One for CO2 MEAS and on  
for CO2 REF that are automatically added to the CPU’s calculation routine. The two offset voltages  
from the last calibration procedure may be reviewed by the user via the front panel display.  
To activate the dark calibration procedure or review the results of a previous calibration, press:  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
DIAG  
SIGNAL I / O  
NEXT  
ENTR  
EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
Repeat Pressing NEXT until . . .  
SETUP X.X  
SECONDARY SETUP MENU  
DIAG  
DARK CALIBRATION  
COMM VARS DIAG ALRM  
EXIT  
PREV NEXT  
ENTR  
EXIT  
DIAG DARK  
CO DARK CALIBRATION  
Exit returns  
to the  
previous menu  
VIEW CAL  
EXIT  
Calibration runs automatically  
Electric offset for Reference signal  
Display  
tracks %  
complete  
DIAG DARK  
REF DARK OFFSET: 0.0 mV  
DIAG DARK  
DARK CAL 1% COMPLETE  
EXIT  
EXIT  
EXIT  
EXIT  
Electric offset for Measurement signal  
DIAG DARK  
MEAS DARK OFFSET: 0.0 mV  
DIAG DARK  
DARK CALIBRATION ABORTED  
6.9.7. Pressure Calibration  
A sensor at the exit of the sample chamber continuously measures the pressure of the sample  
gas. This data is used to compensate the final CO2 concentration calculation for changes in  
atmospheric pressure and is stored in the CPU’s memory as the test function PRES (also viewable  
via the front panel).  
NOTE  
04584 Rev A1  
78  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
This calibration must be performed when the pressure of the sample gas is equal to  
ambient atmospheric pressure.  
Before performing the following pressure calibration procedure, disconnect the sample  
gas pump and the sample gas-line vent from the sample gas inlet on the instrument’s  
rear panel.  
04584 Rev A1  
79  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
To cause the analyzer to measure and record a value for PRES, press.  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
DIAG  
SIGNAL I / O  
< TST TST > CAL  
SETUP  
NEXT  
NEXT  
ENTR  
ENTR  
EXIT  
EXIT  
SAMPLE  
ENTER SETUP PASS : 818  
DIAG  
SIGNAL I / O  
8
1
8
ENTR EXIT  
Exit at  
any time  
to return  
to main  
the  
SETUP X.X  
PRIMARY SETUP MENU  
Repeat Pressing NEXT until . . .  
SETUP  
menu  
CFG DAS RNGE PASS CLK MORE  
EXIT  
ENTR accepts the  
new value and  
returns to the  
previous menu  
EXIT ignores the  
new value and  
returns to the  
DIAG FCAL ACTUAL PRESS : 27.20 IN-HG-A  
SETUP X.X  
SECONDARY SETUP MENU  
2
7
.2  
0
ENTR EXIT  
COMM VARS DIAG ALRM  
EXIT  
Adjust these values until  
the displayed flow rate  
equals the flow rate being  
measured by the  
previous menu  
independent flow meter.  
6.9.8. Flow Calibration  
The flow calibration allows the user to adjust the values of the sample flow rates as they are  
displayed on the front panel and reported through COM ports to match the actual flow rate  
measured at the sample inlet. This does not change the hardware measurement of the flow  
sensors, only the software calculated values.  
To carry out this adjustment, connect an external, sufficiently accurate flow meter to the sample  
inlet (see Chapter 11 for more details). Once the flow meter is attached and is measuring actual  
gas flow, press:  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
DIAG  
SIGNAL I / O  
SAMPLE  
ENTER SETUP PASS : 818  
NEXT  
ENTR  
EXIT  
8
1
8
ENTR EXIT  
Exit at  
any time  
to return  
to main  
the  
SETUP X.X  
PRIMARY SETUP MENU  
Repeat Pressing NEXT until . . .  
CFG DAS RNGE PASS CLK MORE  
EXIT  
SETUP  
menu  
DIAG  
FLOW CALIBRATION  
SETUP X.X  
SECONDARY SETUP MENU  
PREV NEXT  
ENTR EXIT  
COMM VARS DIAG ALRM  
EXIT  
ENTR accepts the  
new value and  
returns to the  
previous menu  
EXIT ignores the  
new value and  
returns to the  
DIAG FCAL  
ACTUAL FLOW: 607 CC / M  
ENTR EXIT  
0
6
0
7
Adjust these values until  
the displayed flow rate  
equals the flow rate being  
measured by the  
independent flow meter.  
previous menu  
04584 Rev A1  
80  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.9.9. Test Channel Output  
When activated, output channel A3 can be used to report one of the test functions viewable from  
the SAMPLE mode display. To activate the A3 channel and select a test function, follow this key  
sequence:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
Continue to press NEXT until …  
< TST TST > CAL  
SETUP  
DIAG  
TEST CHAN OUTPUT  
SAMPLE  
ENTER SETUP PASS : 818  
EXIT returns  
to the main  
SAMPLE  
display  
PREV  
NEXT  
ENTR  
EXIT  
8
1
8
ENTR EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
DIAG TCHN  
TEST CHANNEL: NONE  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
NEXT  
ENTR  
EXIT  
SETUP X.X  
SECONDARY SETUP MENU  
DIAG TCHN TEST CHANNEL: CO2 MEASURE  
PREV NEXT ENTR  
COMM VARS DIAG ALRM  
EXIT  
DIAG  
SIGNAL I / O  
NEXT  
ENTR EXIT  
ENTR EXIT  
Press PREV or NEXT  
to move through the  
list of available  
parameters  
Press ENTR to select  
Press EXIT to  
return to the  
DIAG menu  
DIAG  
ANALOG OUTPUT  
the displayed  
parameter activating  
the test channel.  
PREV NEXT  
(Table 6-13)  
Table 6-14: Test Parameters Available for Analog Output A3  
Test Channel  
Zero  
Full Scale  
NONE  
Test Channel is turned off  
CO2 MEASURE  
CO2 REFERENCE  
SAMPLE PRESS  
SAMPLE FLOW  
SAMPLE TEMP  
BENCH TEMP  
WHEEL TEMP  
CHASSIS TEMP  
PHT DRIVE  
0 mV  
0 mV  
0 "Hg  
0 cc/m  
0°C  
5000 mV*  
5000 mV*  
40 "Hg  
1000 cc/m  
70°C  
0°C  
70°C  
0°C  
70°C  
0°C  
70°C  
0 mV  
5000 mV  
* This refers to the internal voltage level of the function NOT the output  
signal level of the Test channel itself.  
04584 Rev A1  
81  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.10. SETUP – COMM: Using the Analyser’s  
Communication Ports  
The MGFC7000E is equipped with two serial communication ports located on the rear panel  
(Figure 3-2). Both ports operate similarly and give the user the ability to communicate with, issue  
commands to, and receive data from the analyzer through an external computer system or  
terminal. By default, both ports operate on the RS-232 protocol. The COM2 port, however, can be  
configured for half-duplex RS-485 communication or can be used for the Teledyne Instruments  
Ethernet interface card (optional equipment, Section 5.5.3).  
Multidrop Communications  
There are two options to connect multiple analyzers to a single computer terminal or data logging  
device over a single serial communications line. Either port can be equipped with an optional RS-  
232 multidrop assembly (Section 5.5.2), or up to eight analyzers can be connected using COM2  
configured for RS-485 operation (contact the factory for further information). A third option is to  
use a code-activated switch (CAS), which can connect typically between 2 and 16 analyzers to one  
communications hub. Contact Teledyne Instruments sales for more information on CAS systems.  
Ethernet Communications  
When equipped with the optional Ethernet interface, the analyzer can be connected to any  
standard 10BaseT Ethernet network via low-cost network hubs, switches or routers. The interface  
operates as a standard TCP/IP device on port 3000. This allows a remote computer to connect  
through the internet to the analyzer using APIcom, terminal emulators or other programs.  
6.10.1. Analyzer ID Code  
The first entry in the COMM menu is for configuration of the analyzer ID code, a numerical value  
of up to 4 digits. Each type of Teledyne Instruments analyzer is configured with a default ID code.  
The MGFC7000E default ID code is 360. When more than one Teledyne Instruments analyzer of  
the same model type is connected to the same communications channel , such as two Model  
GFC7000E’s operating on the same Hessen Protocol network, each analyzer needs to be  
addressed with a unique ID number.  
To edit the instrument’s ID code, press:  
04584 Rev A1  
82  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SAMPLE  
ENTER SETUP PASS : 818  
8
8
1
ENTR EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
SETUP X.X SECONDARY SETUP MENU  
COMM VARS DIAG ALRM  
EXIT  
SETUP X.X  
COMMUNICATIONS MENU  
EXIT  
ID COM1 COM2  
Toggle these keys to  
cycle through the  
available character set:  
0-9  
ENTR key accepts the  
SETUP X.  
MACHINE ID: 360 ID  
new settings  
EXIT key ignores the  
0
3
6
0
ENTR EXIT  
new settings  
The ID number is only important if more than one analyzer is connected to the same  
communications channel (e.g., a multi-drop setup). Different models of Teledyne Instruments  
analyzers have different default ID numbers, but if two analyzers of the same model type are  
used on one channel (for example, two MGFC7000E’s), the ID of one instrument needs to be  
changed.  
The ID can also be used for to identify any one of several analyzers attached to the same network  
but situated in different physical locations.  
6.10.2. COMM Port Default Settings  
Received from the factory, the analyzer is set up to emulate a DCE or modem, with pin 3 of the  
DB-9 connector designated for receiving data and pin 2 designated for sending data.  
COM1: RS-232 (fixed), DB-9 male connector.  
o Baud rate: 19200 bits per second (baud).  
o Data Bits: 8 data bits with 1 stop bit.  
o Parity: None.  
COM2: RS-232 (configurable), DB-9 female connector.  
o Baud rate: 115000 bits per second (baud).  
o Data Bits: 8 data bits with 1 stop bit.  
o Parity: None.  
NOTE  
Cables that appear to be compatible because of matching connectors may incorporate  
internal wiring that make the link inoperable. Check cables acquired from sources other  
than Teledyne Instruments for pin assignments before using.  
04584 Rev A1  
83  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.10.3. COMM Port Cable Connections  
There are two DB-9 connectors on the MGFC7000E rear panel. COM1 is a male connector, COM2  
a female connector (Table 6-15 lists pin assignments). Teledyne Instruments offers two mating  
cables, one of which should be applicable for your use.  
Part number WR000077, a DB-9 female to DB-9 female cable, 6 feet long, allows connection of  
COM1 with the serial port of most personal computers. Also available as Option 60 (Section  
5.5.1).  
Part number WR0000024, a DB-9 female to DB-25 male cable. Allows connection to the most  
common styles of modems (e.g. Hayes-compatible) and code activated switches.  
Both cables are configured with straight-through wiring and should require no additional adapters.  
To assist in properly connecting the serial ports to either a computer or a modem, there are  
activity indicators just above each COM port. When power is applied to the analyzer, the red LED  
should be illuminated. If this LED is dark, it indicates a communications error between serial port  
and CPU.  
Once a cable is connected between the analyzer and a computer or modem, both the red and  
green LEDs should be on. If not, COM1 can be switched between DTE and DCE modes using a  
small switch on the rear panel to exchange the receive and transmit lines (emulating a cross-over  
or null-modem cable). If both LEDs are still not illuminated, check the cable for proper wiring. For  
COM2 it may be necessary to install a null-modem cable (contact customer service for  
information).  
6.10.4. RS-485 Configuration of COM2  
As delivered from the factory, COM2 is configured for RS-232 communications. This port can be  
re-configured for operation as a non-isolated, half-duplex RS-485 port with a 150 Ω termination  
resistor (Table 6-15 shows the pin assignments of the DB-9 connector).  
For RS-485 operation, jumper JP3 on the CPU board should be installed and switch 6 of SW1  
should be set to the ON position. For RS-232, remove the jumper and set the switch to OFF  
(default). JP3 is just to the right of the third connector from the left on the top of the CPU board  
(as seen from the inside of the analyzer). SW1 is in the middle of the CPU board between disk-on-  
chip and BIOS. For non-terminated RS-485 operation, remove the jumper on JP3 but leave switch  
6 in the ON position. Refer to Figure 3-11 to locate the CPU board.  
04584 Rev A1  
84  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
Table 6-15: COM1 and COM2 DB-9 Pin Assignments  
Pin #  
COM1 (RS-232)  
Not used  
COM2 (RS-232)  
Not used  
COM2 (RS-485)  
Not used  
1
2
3
4
5
6
7
8
Transmit Data*  
Receive Data*  
Not used  
Receive Data  
Transmit Data  
Not used  
DATA -  
Data +  
Not used  
Signal Ground  
Not used  
Signal Ground  
Not used  
Signal Ground  
Not used  
DATA SET READY*  
DATA SET READY  
REQUEST TO SEND  
Not used  
REQUEST TO SEND*  
(=DTE Ready)  
Not used  
9
Not used  
Not used  
Not used  
* Configurable for COM1 at rear panel using the DTE-DCE switch  
6.10.5. DTE and DCE Communication  
RS-232 was developed for allowing communications between data terminal equipment (DTE) and  
data communication equipment (DCE). Basic terminals always fall into the DTE category whereas  
modems are always considered DCE devices. The difference between the two is the pin  
assignment of the Data Receive and Data Transmit functions. DTE devices receive data on pin 2  
and transmit data on pin 3, DCE devices receive data on pin 3 and transmit data on pin 2.  
To allow the analyzer to be used with terminals (DTE), modems (DCE) and computers (which can  
be either), a switch mounted below the serial ports on the rear panel allows the user to switch  
between the two functions.  
6.10.6. COMM Port Communication Modes  
Each of the analyzer’s serial ports can be configured to operate in a number of different modes,  
listed in Table 6-16, which can be combined by adding the mode ID numbers. For example, quiet  
mode, computer mode and internet-enabled mode would carry a combined mode ID of 11, the  
standard configuration on the MGFC7000E COM2 port. Note that each COM port needs to be  
configured independently.  
04584 Rev A1  
85  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
Table 6-16: COMM Port Communication modes  
MODE1  
QUIET  
ID  
1
DESCRIPTION  
Quiet mode suppresses any feedback from the analyzer (iDAS reports, and  
warning messages) to the remote device and is typically used when the port  
is communicating with a computer program such as APICOM. Such feedback  
is still available but a command must be issued to receive them.  
COMPUTER  
SECURITY  
Computer mode inhibits echoing of typed characters and is used when the  
port is communicating with a computer program, such as APICOM.  
2
4
When enabled, the serial port requires a password before it will respond.  
The only command that is active is the help screen (? CR).  
HESSEN  
PROTOCOL  
The Hessen communications protocol is used in some European countries.  
Teledyne Instruments part number 02252 contains more information on this  
protocol.  
16  
E, 7, 1  
When turned on this mode switches the COMM port settings  
from  
No parity; 8 data bits; 1 stop bit  
to  
2048  
Even parity; 7 data bits; 1 stop bit  
RS-485  
Configures the COM2 Port for RS-485 communication. RS-485 mode has  
precedence over multidrop mode if both are enabled.  
1024  
32  
MULTIDROP  
PROTOCOL  
Multidrop protocol allows a multi-instrument configuration on a single  
communications channel. Multidrop requires the use of instrument IDs.  
ENABLE  
MODEM  
Enables to send a modem initialization string at power-up. Asserts certain  
lines in the RS-232 port to enable the modem to communicate.  
64  
ERROR  
Fixes certain types of parity errors at certain Hessen protocol installations.  
128  
256  
CHECKING2  
XON/XOFF  
Disables XON/XOFF data flow control also known as software handshaking.  
HANDSHAKE2  
HARDWARE  
HANDSHAKE  
Enables CTS/RTS style hardwired transmission handshaking. This style of  
data transmission handshaking is commonly used with modems or terminal  
emulation protocols as well as by Teledyne Instrument’s APICOM software.  
8
HARDWARE  
FIFO2  
Improves data transfer rate when on of the COMM ports.  
512  
COMMAND  
PROMPT  
Enables a command prompt when in terminal mode.  
4096  
1
Modes are listed in the order in which they appear in the  
SETUP Æ MORE Æ COMM Æ COM[1 OR 2] Æ MODE menu  
The default sting for this feature is ON. Do not disable unless instructed to by Teledyne Instruments  
2
Customer Service personnel.  
04584 Rev A1  
86  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
Press the following keys to select a communication mode for a one of the COMM Ports, such as the  
following example where HESSEN PROTOCOL mode is enabled:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
EXIT  
EXIT  
EXIT returns  
to the  
previous  
SETUP X.X  
SECONDARY SETUP MENU  
COMM VARS DIAG ALRM  
menu  
SETUP X.X  
COMMUNICATIONS MENU  
Select which COM  
port to configure  
ID COM1 COM2  
The sum of the mode  
IDs of the selected  
modes is displayed  
here  
SETUP X.X  
SET> EDIT  
COM1 MODE:0  
SETUP X.X  
COM1 QUIET MODE: OFF  
ENTR EXIT  
NEXT OFF  
Continue pressing next until …  
SETUP X.X COM1 HESSEN PROTOCOL : OFF  
Use PREV and NEXT  
keys to move between  
available modes.  
PREV NEXT OFF  
ENTR EXIT  
A mode is enabled by  
toggling the ON/OFF  
key.  
ENTR key accepts the  
SETUP X.X COM1 HESSEN PROTOCOL : ON  
new settings  
EXIT key ignores the  
PREV NEXT ON  
ENTR EXIT  
new settings  
Continue pressing the NEXT and PREV keys to select any other  
modes you which to enable or disable  
04584 Rev A1  
87  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.10.7. COM Port Baud Rate  
To select the baud rate of one of the COM Ports, press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
EXIT  
EXIT  
EXIT  
EXIT  
EXIT  
EXIT returns  
to the  
previous  
SETUP X.X  
SECONDARY SETUP MENU  
menu  
COMM VARS DIAG ALRM  
SETUP X.X  
COMMUNICATIONS MENU  
Select which COM  
port to configure.  
ID COM1 COM2  
SETUP X.X  
SET> EDIT  
COM1 MODE:0  
Press SET> until you  
reach COM1 BAUD  
RATE  
EXAMPLE  
SETUP X.X  
COM1 BAUD RATE:19200  
Use PREV and NEXT  
keys to move  
between available  
baud rates.  
EXIT key  
ignores  
the new  
setting  
<SET SET> EDIT  
300  
1200  
SETUP X.X  
COM1 BAUD RATE:19200  
ENTR  
4800  
9600  
19200  
38400  
57600  
115200  
ENTR key  
accepts  
the new  
setting  
PREV NEXT  
SETUP X.X  
COM1 BAUD RATE:9600  
ENTR  
NEXT ON  
04584 Rev A1  
88  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.10.8. COM Port Testing  
The serial ports can be tested for correct connection and output in the COMM menu. This test  
sends a string of 256 ‘w’ characters to the selected COM port. While the test is running, the red  
LED on the rear panel of the analyzer should flicker.  
To initiate the test press the following key sequence.  
SAMPLE  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP X.X  
COMMUNICATIONS MENU  
Select which  
COM port to  
test.  
< TST TST > CAL  
SETUP  
ID COM1 COM2  
EXIT  
EXIT  
SAMPLE  
ENTER SETUP PASS : 818  
SETUP X.X  
SET> EDIT  
COM1 MODE:0  
8
1
8
ENTR EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
SETUP X.X  
COM1 BAUD RATE:19200  
CFG DAS RNGE PASS CLK MORE  
EXIT  
<SET SET> EDIT  
EXIT  
SETUP X.X  
SECONDARY SETUP MENU  
SETUP X.X  
<SET  
COM1 : TEST PORT  
TEST  
COMM VARS DIAG ALRM  
EXIT  
EXIT  
SETUP X.X  
<SET  
TRANSMITTING TO COM1  
TEST  
Test runs  
automatically  
EXIT returns to  
COMM menu  
EXIT  
6.10.9. Ethernet Card Configuration  
The optional Ethernet card (Option 63) allows the analyzer to communicate via standard 10BaseT  
Ethernet protocol through the COM2 serial port. Refer to Figure and 5-4 for the physical location  
of this option.  
The card has four LEDs that are visible on the rear panel of the analyzer, indicating its current  
operating status.  
Table 6-17: Ethernet Status Indicators  
LED  
LNK (green)  
ACT (yellow)  
TxD (green)  
FUNCTION  
ON when connection to the LAN is valid.  
Flickers on any activity on the LAN.  
Flickers when the RS-232 port is transmitting  
data.  
RxD (yellow)  
Flickers when the RS-232 port is receiving data.  
04584 Rev A1  
89  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.10.9.1. Ethernet Card COM2 Communication Modes and Baud Rate  
The firmware on board the Ethernet card automatically sets the communication modes and baud  
rate (115 200 kBaud ) for the COM2 port. Once the Ethernet option is installed and activated,  
the COM2 submenu is replaced by a new submenu, INET. This submenu is used to manage and  
configure the Ethernet interface with your LAN or Internet Server(s).  
6.10.9.2. Configuring the Ethernet Interface Option using DHCP  
The Ethernet option for you MGFC7000E uses Dynamic Host Configuration Protocol (DHCP) to  
automatically configure its interface with your LAN. This requires your network servers also be  
running DHCP. The analyzer will do this the first time you turn the instrument on after it has been  
physically connected to your network. Once the instrument is connected and turned on it will  
appear as an active device on your network without any extra set up steps or lengthy procedures.  
Should you need to, the following Ethernet configuration properties are viewable via the analyzer’s  
front panel.  
Table 6-18: LAN/Internet Configuration Properties  
PROPERTY  
DEFAULT STATE  
DESCRIPTION  
This displays whether the DHCP is turned ON or  
OFF.  
DHCP  
STATUS  
On  
Editable  
EDIT key  
disabled  
when DHCP  
This string of four packets of 1 to 3 numbers each  
(e.g. 192.168.76.55.) is the address of the  
analyzer itself.  
INSTRUMENT  
IP ADDRESS  
Configured  
by DHCP  
is ON  
EDIT key  
disabled  
when DHCP  
is ON  
A string of numbers very similar to the Instrument  
IP address (e.g. 192.168.76.1.)that is the address  
of the computer used by your LAN to access the  
Internet.  
GATEWAY IP  
ADDRESS  
Configured  
by DHCP  
Also a string of four packets of 1 to 3 numbers  
each (e.g. 255.255.252.0) that defines that  
identifies the LAN the device is connected to.  
All addressable devices and computers on a LAN  
must have the same subnet mask. Any  
transmissions sent devices with different assumed  
to be outside of the LAN and are routed through  
gateway computer onto the Internet.  
EDIT key  
disabled  
when DHCP  
is ON  
SUBNET  
MASK  
Configured  
by DHCP  
This number defines the terminal control port by  
which the instrument is addressed by terminal  
emulation software, such as Internet or Teledyne  
Instruments’ APICOM.  
TCP PORT  
3000  
Editable  
Editable  
The name by which your analyzer will appear  
when addressed from other computers on the LAN  
or via the Internet. While the default setting for  
all Teledyne Instruments MGFC7000E analyzers is  
“mGFC7000E” the host name may be changed to  
fit customer needs.  
HOST NAME  
MGFC7000E  
1 Do not change the setting for this property unless instructed to by Teledyne Instruments  
04584 Rev A1  
90  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
Customer Service personnel.  
NOTE  
It is a good idea to check these settings the first time you power up your analyzer after  
it has been physically connected to the LAN/Internet to make sure that the DHCP has  
successfully downloaded the appropriate information from you network server(s).  
If the gateway IP, instrument IP and the subnet mask are all zeroes (e.g. “0.0.0.0”),  
the DCHP was not successful.  
You may have to manually configure the analyzer’s Ethernet properties.  
See your network administrator.  
To view the above properties, press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP X.X  
DHCP: ON  
< TST TST > CAL  
SETUP  
SET> EDIT  
EXIT  
EXIT  
EXIT  
EXIT  
EXIT  
EXIT  
SAMPLE  
ENTER SETUP PASS : 818  
SETUP X.X  
INST IP: 0.0.0.0  
8
1
8
ENTR EXIT  
<SET SET>  
SETUP X.X  
PRIMARY SETUP MENU  
SETUP X.X GATEWAY IP: 0.0.0.0  
EDIT Key  
Disabled  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
EXIT  
<SET SET>  
SETUP X.X SECONDARY SETUP MENU  
SETUP X.X  
SUBNET MASK: 0.0.0.0  
COMM VARS DIAG ALRM  
<SET SET>  
SETUP X.X  
ID  
INET  
COMMUNICATIONS MENU  
SETUP X.X  
TCP PORT: 3000  
COM1  
<SET SET> EDIT  
From this point on,  
EXIT returns to  
COMMUNICATIONS  
MENU  
SETUP X.X HOSTNAME: M360E  
<SET  
EDIT  
Don not alter unless  
directed to by Teledyne  
Instruments Customer  
Service personnel  
04584 Rev A1  
91  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.10.9.3. Manually Configuring the Network IP Addresses  
There are several circumstances when you may need to manually configure the interface settings  
of the analyzer’s Ethernet card. The INET sub-menu may also be used to edit the Ethernet card’s  
configuration properties  
Your LAN is not running a DHCP software package,  
The DHCP software is unable to initialize the analyzer’s interface;  
You wish to program the interface with a specific set of IP addresses that may not be the  
ones automatically chosen by DHCP.  
Editing the Ethernet Interface properties is a two step process.  
STEP 1: Turn DHCP OFF: While DHCP is turned ON, the ability to manually set INSTRUMENT IP,  
GATEWAY IP and SUBNET MASK is disabled  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP X.X  
ID  
INET  
COMMUNICATIONS MENU  
< TST TST > CAL  
SETUP  
COM1  
EXIT  
EXIT  
SAMPLE  
ENTER SETUP PASS : 818  
SETUP X.X  
DHCP: ON  
8
1
8
ENTR EXIT  
<SET SET> EDIT  
SETUP X.X  
DHCP: ON  
SETUP X.X  
PRIMARY SETUP MENU  
ON  
ENTR EXIT  
ENTR EXIT  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP X.X  
DHCP: ON  
SETUP X.X SECONDARY SETUP MENU  
OFF  
COMM VARS DIAG ALRM  
Continue with editing of Ethernet interface  
properties (see Step 2, below).  
ENTR accept  
new settings  
EXIT ignores  
new settings  
04584 Rev A1  
92  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
STEP 2: Configure the INSTRUMENT IP, GATEWAY IP and SUBNET MASK addresses by  
pressing:  
Internet Configuration Keypad Functions  
From Step 1 above)  
KEY  
[0]  
FUNCTION  
Press this key to cycle through the range of  
numerals and available characters (“0 – 9” & “ . ”)  
<CH CH>  
DEL  
Moves the cursor one character left or right.  
Deletes a character at the cursor location.  
SETUP X.X  
DHCP: OFF  
Accepts the new setting and returns to the previous  
menu.  
ENTR  
EXIT  
SET> EDIT  
EXIT  
EXIT  
Ignores the new setting and returns to the previous  
menu.  
Some keys only appear as needed.  
SETUP X.X INST IP: 000.000.000.000  
<SET SET> EDIT  
Cursor  
location is  
indicated by  
brackets  
SETUP X.X INST IP: [0] 00.000.000  
<CH CH>  
DEL [0]  
ENTR EXIT  
SETUP X.X GATEWAY IP: 000.000.000.000  
<SET SET> EDIT  
EXIT  
SETUP X.X GATEWAY IP: [0] 00.000.000  
<CH CH> DEL [?] ENTR EXIT  
SETUP X.X SUBNET MASK:255.255.255.0  
<SET SET> EDIT  
EXIT  
SETUP X.X SUBNET MASK:[2]55.255.255.0  
<CH CH> DEL [?] ENTR EXIT  
SETUP X.X TCP PORT 3000  
<SET  
EDIT  
EXIT  
The PORT number needs to remain at 3000.  
Do not change this setting unless instructed to by  
Teledyne Instruments Customer Service personnel.  
Pressing EXIT from  
any of the above  
display menus  
causes the Ethernet  
option to reinitialize  
its internal interface  
firmware  
SETUP X.X  
INITIALIZING INET 0%  
INITIALIZING INET 100%  
SETUP X.X  
INITIALIZATI0N SUCCEEDED  
SETUP X.X  
INITIALIZATION FAILED  
Contact your IT  
Network Administrator  
SETUP X.X  
COMMUNICATIONS MENU  
ID  
INET  
COM1  
EXIT  
04584 Rev A1  
93  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.10.9.4. Changing the Analyzer’s HOSTNAME  
The HOSTNAME is the name by which the analyzer appears on your network. The default name  
for all Teledyne Instruments Model GFC7000E analyzers is MGFC7000E. To change this name  
(particularly if you have more than one Model GFC7000E analyzer on your network), press.  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP X.X  
DHCP: ON  
< TST TST > CAL  
SETUP  
SET> EDIT  
EXIT  
Continue pressing SET> UNTIL …  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
SETUP X.X HOSTNAME: M360E  
<SET EDIT  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
EXIT  
EXIT  
SETUP X.X SECONDARY SETUP MENU  
SETUP X.X HOSTNAME: [M]360E  
COMM VARS DIAG ALRM  
<CH CH> INS DEL [?]  
ENTR EXIT  
SETUP X.X  
ID INET  
COMMUNICATIONS MENU  
SETUP X.X HOSTNAME: M360E-FIELD1  
COM1  
<SET  
EDIT  
EXIT  
SETUP X.X  
INITIALIZING INET 0%  
INITIALIZING INET 100%  
SETUP X.X  
INITIALIZATI0N SUCCEEDED  
SETUP X.X  
INITIALIZATION FAILED  
SETUP X.X  
COMMUNICATIONS MENU  
Contact your IT  
Network Administrator  
ID  
INET  
COM1  
EXIT  
Table 6-19: Internet Configuration Keypad Functions  
FUNCTION  
KEY  
<CH  
CH>  
INS  
DEL  
[?]  
Moves the cursor one character to the left.  
Moves the cursor one character to the right.  
Inserts a character before the cursor location.  
Deletes a character at the cursor location.  
Press this key to cycle through the range of numerals and characters available  
for insertion. 0-9, A-Z, space ’ ~ ! © # $ % ^ & * ( ) - _ = +[ ] { } <  
>\ | ; : , . / ?  
ENTR  
EXIT  
Accepts the new setting and returns to the previous menu.  
Ignores the new setting and returns to the previous menu.  
04584 Rev A1  
94  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
Some keys only appear as needed.  
6.11. SETUP – ALRM: Using the Gas Concentration  
Alarms  
The Model GFC7000E includes two CO2 concentration alarms Each alarm has a user settable  
limit, and is associated with an opto-isolated TTL relay accessible via the status output connector  
on the instrument’s back panel (see Section 6.13.1.1). If the CO2 concentration measured by the  
instrument rises above that limit, the alarm‘s status output relay is closed.  
The default settings for ALM1 and ALM2 are:  
Table 6-20: CO2 Concentration Alarm Default Settings  
1
ALARM  
STATUS  
LIMIT SET POINT  
100 ppm  
ALM1  
ALM2  
Disabled  
Disabled  
300 ppm  
1
Set points listed are for PPM. Should the reporting range units of measure be changed (see Section  
6.7.6) the analyzer will automatically scale the set points to match the new range unit setting.  
Note  
To prevent the concentration alarms from activating during span calibration operations  
make sure to press CAL or CALS button prior to introducing span gas into the analyzer.  
To enable either of the CO2 concentration alarms and set the Limit points, press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
SETUP X.X  
ALARM MENU  
SETUP X.X  
PRIMARY SETUP MENU  
ALM1 ALM2  
EXIT  
CFG DAS RNGE PASS CLK MORE  
EXIT  
SETUP X.  
ALARM 1 LIMIT: OFF  
SETUP X.X SECONDARY SETUP MENU  
OFF  
ENTR EXIT  
COMM VARS DIAG ALRM  
EXIT  
SETUP X.  
ALARM 1 LIMIT: ON  
ON  
ENTR EXIT  
Toggle these keys to  
cycle through the  
available character set:  
0-9  
SETUP X.  
ALARM 1 LIMIT: 200,00 PPM  
.0 ENTR EXIT  
ENTR key accepts the  
new settings  
0
1
0
0
0
EXIT key ignores the  
new settings  
04584 Rev A1  
95  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.12. SETUP – DAS: Using the Data Acquisition System  
(iDAS)  
The MGFC7000E analyzer contains a flexible and powerful, internal data acquisition system (iDAS)  
that enables the analyzer to store concentration and calibration data as well as a host of  
diagnostic parameters. The iDAS of the MGFC7000E can store up to about one million data points,  
which can, depending on individual configurations, cover days, weeks or months of valuable  
measurements. The data are stored in non-volatile memory and are retained even when the  
instrument is powered off. Data are stored in plain text format for easy retrieval and use in  
common data analysis programs (such as spreadsheet-type programs).  
The iDAS is designed to be flexible, users have full control over the type, length and reporting  
time of the data. The iDAS permits users to access stored data through the instrument’s front  
panel or its communication ports. Using APICOM, data can even be retrieved automatically to a  
remote computer for further processing.  
The principal use of the iDAS is logging data for trend analysis and predictive diagnostics, which  
can assist in identifying possible problems before they affect the functionality of the analyzer. The  
secondary use is for data analysis, documentation and archival in electronic format.  
To support the iDAS functionality, Teledyne Instruments offers APICOM, a program that provides  
a visual interface for remote or local setup, configuration and data retrieval of the iDAS (Section  
6.12). The APICOM manual, which is included with the program, contains a more detailed  
description of the iDAS structure and configuration, which is briefly described in this section.  
The MGFC7000E is configured with a basic iDAS configuration, which is enabled by default. New  
data channels are also enabled by default but each channel may be turned off for later or  
occasional use. Note that iDAS operation is suspended while its configuration is edited through the  
front panel. To prevent such data loss, it is recommended to use the APICOM graphical user  
interface for iDAS changes.  
The green SAMPLE LED on the instrument front panel, which indicates the analyzer status, also  
indicates certain aspects of the iDAS status:  
Table 6-21: Front Panel LED Status Indicators for iDAS  
LED STATE  
OFF  
iDAS Status  
System is in calibration mode. Data logging can be enabled or disabled for this  
mode. Calibration data are typically stored at the end of calibration periods,  
concentration data are typically not sampled, diagnostic data should be collected.  
BLINKING  
ON  
Instrument is in hold-off mode, a short period after the system exits calibrations.  
IDAS channels can be enabled or disabled for this period. Concentration data are  
typically disabled whereas diagnostic should be collected.  
Sampling normally.  
The iDAS can be disabled only by disabling or deleting its individual data channels.  
6.12.1. iDAS Structure  
The iDAS is designed around the feature of a “record”. A record is a single data point of one  
parameter, stored in one (or more) data channels and generated by one of several triggering  
event. The entire iDAS configuration is stored in a script, which can be edited from the front panel  
04584 Rev A1  
96  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
or downloaded, edited and uploaded to the instrument in form of a string of plain-text lines  
through the communication ports.  
iDAS data are defined by the PARAMETER type and are stored through different triggering EVENTS  
in data CHANNELS, which relate triggering events to data parameters and define certain  
operational functions related to the recording and reporting of the data.  
6.12.1.1. iDAS Channels  
The key to the flexibility of the iDAS is its ability to store a large number of combinations of  
triggering events and data parameters in the form of data channels. Users may create up to 20  
data channels and each channel can contain one or more parameters. For each channel one  
triggering event is selected and up to 50 data parameters, which can be the same or different  
between channels. Each data channel has several properties that define the structure of the  
channel and allow the user to make operational decisions regarding the channel (Table 6-22).  
Table 6-22: iDAS Data Channel Properties  
PROPERTY  
NAME  
DESCRIPTION  
DEFAULT  
“NONE”  
SETTING RANGE  
The name of the data channel.  
Up to 6 letters and  
digits (more with  
APICOM, but only the  
first six are displayed  
on the front panel).  
TRIGGERIN  
G EVENT  
The event that triggers the data channel to  
measure and store its data parameters. See  
APPENDIX A-5 for a list of available triggering  
events.  
ATIMER  
See Appendix A-5 For  
a complete list.  
NUMBER &  
A User-configurable list of data types to be  
1 – DETMES See Appendix A-5 For  
PARAMETER recorded in any given channel. See APPENDIX  
a complete list.  
LIST  
A-5 for a list of available parameters  
STARTING  
DATE  
The starting date when a channel starts  
collecting data  
01-JAN-03  
000:01:00  
Any actual date in the  
past or future.  
SAMPLE  
PERIOD  
The amount of time between each data point  
that is averaged into one mean reported every  
REPORT PERIOD.  
000:00:01 to  
366:23:59  
(Days:Hours:Minutes)  
REPORT  
PERIOD  
The amount of time between each channel  
data point.  
000:01:00  
100  
000:00:01 to  
366:23:59  
(Days:Hours:Minutes)  
NUMBER OF The number of reports that will be stored in  
RECORDS  
1 to 1 million, limited  
by available storage  
space.  
the data file. Once the specified limit has been  
exceeded, the oldest data are over-written to  
make space for new data.  
RS-232  
REPORT  
Enables the analyzer to automatically report  
channel values to the RS-232 ports.  
OFF  
ON  
OFF or ON  
OFF or ON  
CHANNEL  
ENABLED  
Enables or disables the channel. Provides a  
convenient means to temporarily disable a  
data channel.  
CAL HOLD  
OFF  
Disables sampling of data parameters while  
instrument is in calibration mode.  
OFF  
OFF or ON  
(Section 6.12.2.11.)  
When enabled here – there is also a length of  
the DAS HOLD OFF after calibration mode,  
which is set in the VARS menu  
04584 Rev A1  
97  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.12.1.2. iDAS Parameters  
Data parameters are types of data that may be measured and stored by the iDAS. For each  
Teledyne Instruments analyzer model, the list of available data parameters is different, fully  
defined and not customizable. Appendix A-5 lists firmware specific data parameters for the  
MGFC7000E. iDAS parameters include things like CO2 concentration measurements, temperatures  
of the various heater placed around the analyzer, pressures and flows of the pneumatic  
subsystem and other diagnostic measurements as well as calibration data such as slope and  
offset.  
Most data parameters have associated measurement units, such as mV, ppb, cm³/min, etc.,  
although some parameters have no units. With the exception of concentration readings, none of  
these units of measure can be changed. To change the units of measure for concentration  
readings see Section 6.7.6.  
Note  
iDAS does not keep track of the unit of each concentration value and iDAS data files  
may contain concentrations in multiple units if the unit was changed during data  
acquisition.  
Each data parameter has user-configurable functions that define how the data are recorded:  
Table 6-23: iDAS Data Parameter Functions  
FUNCTION  
EFFECT  
PARAMETER  
Instrument-specific parameter name.  
SAMPLE MODE INST: Records instantaneous reading.  
AVG: Records average reading during reporting interval.  
MIN: Records minimum (instantaneous) reading during reporting interval.  
MAX: Records maximum (instantaneous) reading during reporting interval.  
PRECISION  
Decimal precision of parameter value(0-4).  
STORE NUM.  
SAMPLES  
OFF: stores only the average (default).  
ON: stores the average and the number of samples in each average for a  
parameter. This property is only useful when the AVG sample mode is used. Note  
that the number of samples is the same for all parameters in one channel and  
needs to be specified only for one of the parameters in that channel.  
Users can specify up to 50 parameters per data channel (the MGFC7000E provides about 30  
parameters). However, the number of parameters and channels is ultimately limited by available  
memory.  
6.12.1.3. iDAS Triggering Events  
Triggering events define when and how the iDAS records a measurement of any given data  
channel. Triggering events are firmware-specific and a complete list of Triggers for this model  
analyzer can be found in Appendix A-5. The most commonly used triggering events are:  
ATIMER: Sampling at regular intervals specified by an automatic timer. Most  
trending information is usually stored at such regular intervals, which can be  
instantaneous or averaged.  
04584 Rev A1  
98  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
EXITZR, EXITSP, SLPCHG (exit zero, exit span, slope change): Sampling at the  
end of (irregularly occurring) calibrations or when the response slope changes.  
These triggering events create instantaneous data points, e.g., for the new slope  
and offset (concentration response) values at the end of a calibration. Zero and  
slope values are valuable to monitor response drift and to document when the  
instrument was calibrated.  
WARNINGS: Some data may be useful when stored if one of several warning  
messages appears such as WTEMPW (GFC wheel temperature warning) or  
PPRESW (purge pressure warning). This is helpful for trouble-shooting by  
monitoring when a particular warning occurred.  
6.12.2. Default iDAS Channels  
A set of default Data Channels has been included in the analyzer’s software for logging CO2  
concentration and certain predictive diagnostic data. These default channels include but are not  
limited to:  
CONC: Samples CO2 concentration at one minute intervals and stores an average every hour  
with a time and date stamp. Readings during calibration and calibration hold off are not included  
in the data. By default, the last 800 hourly averages are stored.  
PNUMTC: Collects sample flow and sample pressure data at five minute intervals and stores an  
average once a day with a time and date stamp. This data is useful for monitoring the condition of  
the pump and critical flow orifice (sample flow) and the sample filter (clogging indicated by a drop  
in sample pressure) over time to predict when maintenance will be required. The last 360 daily  
averages (about 1 year) are stored.  
CALDAT: Logs new slope and offset every time a zero or span calibration is performed. This Data  
Channel also records the instrument readings just prior to performing a calibration. This  
information is useful for performing predictive diagnostics as part of a regular maintenance  
schedule (see Section 9.1).  
STBZRO: Logs the concentration stability, the electronic output of the IR detector of the most  
recent measure phase and the measure/reference ratio every time the instrument exits zero  
calibration mode. Data from the last 200 zero calibrations is stored. A time and date stamp is  
recorded for every data point logged. This information is useful for performing predictive  
diagnostics as part of a regular maintenance schedule (see Section 9.1).  
STBSPN: Logs the electronic output of the IR detector of the most recent measure phase and the  
measure/reference ratio every time the instrument exits span calibration mode. Data from the  
last 200 zero calibrations is stored. A time and date stamp is recorded for every data point logged.  
This information is useful for performing predictive diagnostics as part of a regular maintenance  
schedule (see Section 9.1).  
TEMP: Samples the analyzer’s bench temperature, box temperature and PHT cooler drive voltage  
every five minutes and records a average once every six hours. Data from the last 400  
averaging periods is recorded. A time and date stamp is recorded for every data point logged.  
This information is useful for performing predictive diagnostics as part of a regular maintenance  
schedule (see Section 9.1).  
Note  
04584 Rev A1  
99  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
The CALDAT, STBZRO and STBSPN channels collect data based on events (e.g. a  
calibration operation) rather than a timed interval. This does not represent any specific  
length of time since it is dependent on how often calibrations are performed.  
04584 Rev A1  
100  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
Triggering Events and Data Parameters/Functions for these default channels are:  
LIST OF CHANNELS  
LIST OF PARAMETERS  
NAME: CONC  
EVENT: ATIMER  
PARAMETER:  
CONC1  
REPORT PERIOD: 000:01:00  
MODE:  
AVG  
NO. OF RECORDS: 800  
RS-232 REPORT: OFF  
CHANNEL ENABLED: ON  
CAL HOLD OFF: ON  
PRECISION:  
1
STORE NUM SAMPLES  
OFF  
PARAMETER:  
SMPLFLW  
MODE:  
AVG  
NAME: PNUMTC  
PRECISION:  
1
EVENT: ATIMER  
STORE NUM SAMPLES  
OFF  
REPORT PERIOD: 001:00:00  
NO. OF RECORDS: 360  
RS-232 REPORT: OFF  
CHANNEL ENABLED: ON  
CAL HOLD OFF: OFF  
PARAMETER:  
SMPLPRS  
MODE:  
AVG  
PRECISION:  
1
STORE NUM SAMPLES  
OFF  
OFF  
PARAMETER:  
SLOPE1  
MODE:  
INST  
PRECISION:  
3
STORE NUM SAMPLES  
NAME: CALDAT  
EVENT: SLPCHG  
PARAMETER:  
OFSET1  
REPORT PERIOD: N/A  
NO. OF RECORDS:200  
RS-232 REPORT: OFF  
CHANNEL ENABLED: ON  
CAL HOLD OFF: OFF  
MODE:  
INST  
PRECISION:  
1
STORE NUM SAMPLES  
OFF  
OFF  
OFF  
PARAMETER:  
ZSCNC1  
MODE:  
INST  
PRECISION:  
1
STORE NUM SAMPLES  
PARAMETER:  
STABIL  
MODE:  
INST  
PRECISION:  
2
STORE NUM SAMPLES  
NAME: STBZRO  
EVENT: EXITZR  
PARAMETER:  
DETMES  
REPORT PERIOD: N/A  
NO. OF RECORDS:200  
RS-232 REPORT: OFF  
CHANNEL ENABLED: ON  
CAL HOLD OFF: OFF  
MODE:  
INST  
PRECISION:  
1
STORE NUM SAMPLES  
OFF  
OFF  
OFF  
PARAMETER:  
RATIO  
MODE:  
INST  
PRECISION:  
3
STORE NUM SAMPLES  
PARAMETER:  
DETMES  
MODE:  
INST  
NAME: STBSPN  
PRECISION:  
1
EVENT: EXITSP  
STORE NUM SAMPLES  
REPORT PERIOD: N/A  
NO. OF RECORDS:200  
RS-232 REPORT: OFF  
CHANNEL ENABLED: ON  
CAL HOLD OFF: OFF  
PARAMETER:  
RATIO  
MODE:  
INST  
PRECISION:  
3
STORE NUM SAMPLES  
OFF  
OFF  
PARAMETER:  
BNTEMP  
MODE:  
AVG  
PRECISION:  
1
STORE NUM SAMPLES  
NAME: TEMP  
EVENT: ATIMER  
PARAMETER:  
BOXTMP  
REPORT PERIOD: 000:06:00  
NO. OF RECORDS:400  
RS-232 REPORT: OFF  
CHANNEL ENABLED: ON  
CAL HOLD OFF: OFF  
MODE:  
AVG  
PRECISION:  
1
STORE NUM SAMPLES  
OFF  
PARAMETER:  
PHTDRV  
MODE:  
AVG  
PRECISION:  
1
STORE NUM SAMPLES  
OFF  
Figure 6-7:  
Default iDAS Channels Setup  
04584 Rev A1  
101  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
These default Data Channels can be used as they are, or they can be customized from the front  
panel to fit a specific application. They can also be deleted to make room for custom user-  
programmed Data Channels.  
Appendix A-5 lists the firmware-specific iDAS configuration in plain-text format. This text file can  
either be loaded into APICOM and then modified and uploaded to the instrument or can be copied  
and pasted into a terminal program to be sent to the analyzer.  
NOTE  
Sending an iDAS configuration to the analyzer through its COM ports will replace the  
existing configuration and will delete all stored data. Back up any existing data and the  
iDAS configuration before uploading new settings.  
These default Data Channels can be used as they are, or they can be customized from the front  
panel to fit a specific application. They can also be deleted to make room for custom user-  
programmed Data Channels.  
Appendix A-5 lists the firmware-specific iDAS configuration in plain-text format. This text file can  
either be loaded into APICOM and then modified and uploaded to the instrument or can be copied  
and pasted into a terminal program to be sent to the analyzer.  
NOTE  
Sending an iDAS configuration to the analyzer through its COM ports will replace the  
existing configuration and will delete all stored data. Back up any existing data and the  
iDAS configuration before uploading new settings.  
04584 Rev A1  
102  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.12.2.1. Viewing iDAS Data and Settings  
iDAS data and settings can be viewed on the front panel through the following keystroke  
sequence.  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
VIEW KEYPAD FUNCTIONS  
KEY  
FUNCTION  
SAMPLE  
ENTER SETUP PASS : 818  
<PRM  
PRM>  
Moves to the next Parameter  
8
1
8
ENTR EXIT  
Moves to the previous  
Parameter  
SETUP X.X  
PRIMARY SETUP MENU  
NX10  
NEXT  
PREV  
PV10  
Moves the view forward 10  
data points/channels  
CFG DAS RNGE PASS CLK MORE  
EXIT  
Moves to the next data  
point/channel  
Moves to the previous data  
point/channel  
SETUP X.X  
DATA ACQUISITION  
Moves the view back 10 data  
points/channels  
VIEW EDIT  
EXIT  
Keys only appear as needed  
SETUP X.X  
CONC : DATA AVAILABLE  
NEXT VIEW  
EXIT  
SETUP X.X  
00:00:00 NXCNC1=0.0 PPM  
PV10 PREV NEXT NX10 <PRM PRM>  
EXIT  
SETUP X.X  
PNUMTC: DATA AVAILABLE  
PREV NEXT VIEW  
EXIT  
SETUP X.X  
00:00:00 SMPFLW=000.0 cc / m  
<PRM  
PRM>  
EXIT  
SETUP X.X  
CALDAT: DATA AVAILABLE  
PREV NEXT VIEW  
EXIT  
SETUP X.X  
00:00:00 NXSLP1=0.000  
<PRM  
PRM>  
PV10 PREV  
EXIT  
SETUP X.X  
STBZRO: DATA AVAILABLE  
PREV NEXT VIEW  
EXIT  
SETUP X.X  
00:00:00 STABIL=0.000  
<PRM PRM>  
PV10 PREV  
EXIT  
EXIT  
EXIT  
SETUP X.X  
STBSPN: DATA AVAILABLE  
PREV NEXT VIEW  
EXIT  
SETUP X.X  
00:00:00 DETMES=0.000  
<PRM  
PRM>  
PV10 PREV  
SETUP X.X  
TEMP: DATA AVAILABLE  
PREV NEXT VIEW  
EXIT  
SETUP X.X  
00:00:00 BOXTMP=0.000  
<PRM  
PRM>  
PV10 PREV  
04584 Rev A1  
103  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.12.2.2. Editing iDAS Data Channels  
IDAS configuration is most conveniently done through the APICOM remote control program. The  
following list of key strokes shows how to edit using the front panel.  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SAMPLE  
ENTER SETUP PASS : 818  
EXIT will return to the  
previous SAMPLE  
display.  
8
1
8
ENTR EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
Main Data Acquisition Menu  
SETUP X.X  
DATA ACQUISITION  
VIEW EDIT  
EXIT  
Edit Data Channel Menu  
Moves the  
display up &  
down the list of  
Data Channels  
SETUP X.X  
0) CONC: ATIMER, 1,  
900  
Exits to the Main  
Data Acquisition  
Menu  
PREV NEXT  
INS DEL EDIT PRNT EXIT  
Exports the  
Inserts a new Data  
Channel into the list  
BEFORE the Channel  
configuration of all  
data channels to  
RS-232 interface.  
currently being displayed  
Deletes The Data  
Channel currently  
being displayed  
Moves the display  
between the  
SETUP X.X  
NAME:CONC  
Exits returns to the  
previous Menu  
PROPERTIES for this  
data channel.  
<SET SET> EDIT PRNT  
EXIT  
Reports the configuration of current  
data channels to the RS-232 ports.  
Allows to edit the channel name, see next key sequence.  
When editing the data channels, the top line of the display indicates some of the configuration  
parameters. For example, the display line:  
0) CONC : ATIMER, 4, 800  
translates to the following configuration:  
Channel No.: 0  
NAME: CONC  
TRIGGER EVENT: ATIMER  
04584 Rev A1  
104  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
PARAMETERS: Four parameters are included in this channel  
EVENT: This channel is set up to record 800 data points.  
To edit the name of a data channel, follow the above key sequence and then press:  
From the end of the previous key sequence …  
SETUP X.X  
NAME:CONC  
<SET SET> EDIT PRINT  
EXIT  
ENTR accepts the new string  
and returns to the previous  
menu.  
SETUP X.X  
NAME:CONC  
C
O
N
C
-
-
ENTR  
EXIT  
EXIT ignores the new string  
and returns to the previous  
menu.  
Press each key repeatedly to cycle through the  
available character set:  
0-9, A-Z, space ’ ~ ! © # $ % ^ & * ( ) - _ = +[ ]  
{ } < >\ | ; : , . / ?  
6.12.2.3. Trigger Events  
To edit the list of data parameters associated with a specific data channel, press:  
From the DATA ACQUISITION menu  
(see Section 6.12.2.2)  
Edit Data Channel Menu  
SETUP X.X  
0) CONC: ATIMER, 1,  
900  
Exits to the Main  
Data Acquisition  
menu  
PREV NEXT  
INS DEL EDIT PRNT EXIT  
SETUP X.X  
NAME:CONC  
<SET SET> EDIT PRINT  
EXIT  
SETUP X.X  
EVENT:ATIMER  
<SET SET> EDIT PRINT  
EXIT  
ENTR accepts the new string  
and returns to the previous  
menu.  
EXIT ignores the new string  
and returns to the previous  
menu.  
SETUP X.X  
EVENT:ATIMER  
<PREV NEXT>  
ENTR  
EXIT  
Press each key repeatedly to cycle through the  
list of available trigger events.  
04584 Rev A1  
105  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.12.2.4. Editing iDAS Parameters  
Data channels can be edited individually from the front panel without affecting other data  
channels. However, when editing a data channel, such as during adding, deleting or editing  
parameters, all data for that particular channel will be lost, because the iDAS can store only data  
of one format (number of parameter columns etc.) for any given channel. In addition, an iDAS  
configuration can only be uploaded remotely as an entire set of channels. Hence, remote update  
of the iDAS will always delete all current channels and stored data.  
To modify, add or delete a parameter, follow the instruction shown in section 6.12.2.2 then press:  
From the DATA ACQUISITION menu  
(see Section 6.12.2.2)  
Edit Data Channel Menu  
SETUP X.X  
0) CONC: ATIMER, 1,  
900  
Exits to the main  
Data Acquisition  
menu  
PREV NEXT  
INS DEL EDIT PRNT EXIT  
SETUP X.X  
NAME:CONC  
<SET SET> EDIT PRINT  
EXIT  
Press SET> key until…  
SETUP X.X  
PARAMETERS:1  
<SET SET> EDIT PRINT  
EXIT  
SETUP X.X  
EDIT PARAMS (DELETE DATA)  
YES will delete  
all data in that  
entire channel.  
NO returns to  
the previous  
menu and  
YES NO  
retains all data.  
Edit Data Parameter Menu  
Moves the  
display between  
existing  
SETUP X.X 0) PARAM=CONC1, MODE=AVG  
PREV NEXT INS DEL EDIT  
Exits to the main  
Data Acquisition  
menu  
EXIT  
Parameters  
Inserts a new Parameter  
before the currently  
displayed Parameter  
Use to configure  
the functions for  
this Parameter.  
Deletes the Parameter  
currently displayed.  
04584 Rev A1  
106  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
To configure a specific data parameter, press:  
FROM THE EDIT DATA PARAMETER MENU  
(see previous section)  
SETUP X.X 0) PARAM=CONC1, MODE=AVG  
PREV NEXT  
INS DEL EDIT  
EXIT  
SETUP X.X PARAMETERS:CONC!  
SET> EDIT  
EXIT  
SETUP X.X PARAMETERS: PMTDET  
PREV NEXT ENTR  
EXIT  
If more than on parameter is active for  
this channel, these cycle through list of  
existing Parameters.  
SETUP X.X SAMPLE MODE:AVG  
<SET SET> EDIT  
EXIT  
SETUP X.X SAMPLE MODE: INST  
INST AVG MIN MAX  
EXIT  
Press the key for the desired mode  
ENTR accepts the new  
setting and returns to the  
previous menu.  
SETUP X.X PRECISION: 1  
EXIT ignores the new setting  
and returns to the previous  
<SET SET> EDIT  
EXIT  
SETUP X.X PRECISION: 1  
1
EXIT  
Set for 0-4  
<SET Returns to  
previous  
SETUP X.X STORE NUM. SAMPLES: OFF  
Functions  
<SET  
EDIT  
EXIT  
SETUP X.X STORE NUM. SAMPLES: OFF  
OFF  
ENTR EXIT  
Turn ON or OFF  
6.12.2.5. Sample Period and Report Period  
The iDAS defines two principal time periods by which sample readings are taken and permanently  
recorded:  
SAMPLE PERIOD: Determines how often iDAS temporarily records a sample reading of  
the parameter in volatile memory. The SAMPLE PERIOD is set to one minute by default  
and generally cannot be accessed from the standard iDAS front panel menu, but is  
available via the instruments communication ports by using APICOM or the analyzer’s  
standard serial data protocol.  
SAMPLE PERIOD is only used when the iDAS parameter’s sample mode is set for AVG,  
MIN or MAX.  
04584 Rev A1  
107  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
REPORT PERIOD: Sets how often the sample readings stored in volatile memory are  
processed, (e.g. average, minimum or maximum are calculated) and the results stored  
permanently in the instruments Disk-on-Chip as well as transmitted via the analyzer’s  
communication ports. The REPORT PERIOD may be set from the front panel.  
If the INST sample mode is selected the instrument stores and reports an instantaneous  
reading of the selected parameter at the end of the chosen REPORT PERIOD  
In AVG, MIN or MAX sample modes, the settings for the SAMPLE PERIOD and the REPORT  
PERIOD determine the number of data points used each time the average, minimum or  
maximum is calculated, stored and reported to the COMM ports. The actual sample readings are  
not stored past the end of the of the chosen REPORT PERIOD.  
Also, the SAMPLE PERIOD and REPORT PERIOD intervals are synchronized to the beginning  
and end of the appropriate interval of the instruments internal clock.  
If SAMPLE PERIOD were set for one minute the first reading would occur at the beginning  
of the next full minute according to the instrument’s internal clock.  
If the REPORT PERIOD were set for of one hour the first report activity would occur at  
the beginning of the next full hour according to the instrument’s internal clock.  
EXAMPLE: Given the above settings, if iDAS were activated at 7:57:35 the first sample  
would occur at 7:58 and the first report would be calculated at 8:00 consisting of data  
points for 7:58. 7:59 and 8:00.  
During the next hour (from 8:01 to 9:00) the instrument will take a sample reading every  
minute and include 60 sample readings.  
When the STORE NUM. SAMPLES feature is turned on the instrument will also store how many  
sample readings were used for the AVG, MIN or MAX calculation but not the readings  
themselves.  
REPORT PERIODS IN PROGRESS WHEN INSTRUMENT IS POWERED OFF  
If the instrument is powered off in the middle of a REPORT PERIOD, the samples accumulated so  
far during that period are lost. Once the instrument is turned back on, the iDAS restarts taking  
samples and temporarily them in volatile memory as part of the REPORT PERIOD currently  
active at the time of restart. At the end of this REPORT PERIOD only the sample readings taken  
since the instrument was turned back on will be included in any AVG, MIN or MAX calculation.  
Also, the STORE NUM. SAMPLES feature will report the number of sample readings taken since  
the instrument was restarted.  
04584 Rev A1  
108  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
To define the REPORT PERIOD, follow the instruction shown in section 6.12.2.2 then press:  
From the DATA ACQUISITION menu  
(see Section 6.12.2.2)  
Edit Data Channel Menu  
Use the PREV and NEXT  
keys to scroll to the data  
channel to be edited.  
SETUP X.X  
0) CONC: ATIMER, 1,  
900  
Exits to the main  
Data Acquisition  
menu.  
PREV NEXT  
INS DEL EDIT PRNT EXIT  
SETUP X.X  
NAME:CONC  
<SET SET> EDIT PRINT  
EXIT  
Press SET> key until you reach REPORT PERIOD …  
SETUP X.X  
REPORT PERIOD:000:01:00  
<SET SET> EDIT PRINT  
EXIT  
SETUP X.X  
REPORT PERIODD:DAYS:0  
Set the number of days  
between reports (0-366).  
0
0
0
ENTR EXIT  
Press keys to set hours  
between reports in the format :  
HH:MM (max: 23:59). This is a  
24 hour clock . PM hours are 13  
thru 23, midnight is 00:00.  
SETUP X.X  
REPORT PERIODD:TIME:01:01  
ENTR EXIT  
ENTR accepts the new string and  
returns to the previous menu.  
EXIT ignores the new string and  
returns to the previous menu.  
0
1
0
0
IIf at any time an illegal entry is selected (e.g., days > 366)  
the ENTR key will disappear from the display.  
Example 2:15 PM = 14:15  
6.12.2.6. Number of Records  
The number of data records in the MGFC7000E is limited to about a cumulative one million data  
points in all channels (one megabyte of space on the disk-on-chip). However, the actual number  
of records is also limited by the total number of parameters and channels and other settings in the  
iDAS configuration. Every additional data channel, parameter, number of samples setting etc. will  
reduce the maximum amount of data points somewhat. In general, however, the maximum data  
capacity is divided amongst all channels (max: 20) and parameters (max: 50 per channel).  
The iDAS will check the amount of available data space and prevent the user from specifying too  
many records at any given point. If, for example, the iDAS memory space can accommodate 375  
more data records, the ENTR key will disappear when trying to specify more than that number of  
records. This check for memory space may also make an upload of an iDAS configuration with  
APICOM or a Terminal program fail, if the combined number of records would be exceeded. In this  
case, it is suggested to either try from the front panel what the maximum number of records can  
be or use trial-and-error in designing the iDAS script or calculate the number of records using the  
DAS or AICOM manuals. To set the number of records for one channel from the front panel, press  
SETUP-DAS-EDIT-ENTR and the following key sequence.  
04584 Rev A1  
109  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
From the DATA ACQUISITION menu  
(see Section 6.12.2.2)  
Edit Data Channel Menu  
SETUP X.X  
0) CONC: ATIMER, 1 2,  
900  
Exits to the main  
Data Acquisition  
menu  
PREV NEXT  
INS DEL EDIT PRNT EXIT  
SETUP X.X  
NAME:CONC  
<SET SET> EDIT PRINT  
EXIT  
Press SET> key until…  
SETUP X.X  
NUMBER OF RECORDS:000  
<SET SET> EDIT PRINT  
EXIT  
SETUP X.X  
EDIT RECOPRDS (DELET DATA)  
NO returns to the  
previous menu.  
YES will delete all data  
in this channel.  
YES  
NO  
ENTR accepts the new  
setting and returns to the  
previous menu.  
EXIT ignores the new setting  
and returns to the previous  
menu.  
Toggle keys to set  
number of records  
(1-99999)  
SETUP X.X  
REPORT PERIODD:DAYS:0  
ENTR EXIT  
0
0
0
0
0
6.12.2.7. RS-232 Report Function  
The MGFC7000E iDAS can automatically report data to the communications ports, where they can  
be captured with a terminal emulation program or simply viewed by the user.  
04584 Rev A1  
110  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
To enable automatic COM port reporting, follow the instruction shown in section 6.12.2.2 then  
press:  
From the DATA ACQUISITION menu  
(see Section 6.12.2.2)  
Edit Data Channel Menu  
SETUP X.X  
0) CONC: ATIMER, 1,  
900  
Exits to the main  
Data Acquisition  
menu  
PREV NEXT  
INS DEL EDIT PRNT EXIT  
SETUP X.X  
NAME:CONC  
<SET SET> EDIT PRINT  
EXIT  
Press SET> key until…  
SETUP X.X  
RS-232 REPORT: OFF  
<SET SET> EDIT PRINT  
EXIT  
ENTR accepts the new  
setting and returns to the  
previous menu.  
EXIT ignores the new setting  
and returns to the previous  
menu.  
SETUP X.X  
RS-232 REPORT: OFF  
Toggle key to turn  
reporting ON or OFF  
OFF  
ENTR EXIT  
6.12.2.8. Compact Report  
When enabled, this option avoids unnecessary line breaks on all RS-232 reports. Instead of  
reporting each parameter in one channel on a separate line, up to five parameters are reported in  
one line.  
6.12.2.9. Starting Date  
This option allows to specify a starting date for any given channel in case the user wants to start  
data acquisition only after a certain time and date. If the Starting Date is in the past, the iDAS  
ignores this setting.  
6.12.2.10. Disabling/Enabling Data Channels  
Data channels can be temporarily disabled, which can reduce the read/write wear on the disk-on-  
chip. The ALL_01 channel of the MGFC7000E, for example, is disabled by default.  
04584 Rev A1  
111  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
To disable a data channel, follow the instruction shown in section 6.12.2.2 then press:  
From the DATA ACQUISITION menu  
(see Section 6.12.2.2)  
Edit Data Channel Menu  
SETUP X.X  
0) CONC: ATIMER, 1,  
900  
Exits to the main  
Data Acquisition  
menu  
PREV NEXT  
INS DEL EDIT PRNT EXIT  
SETUP X.X  
NAME:CONC  
<SET SET> EDIT PRINT  
EXIT  
Press SET> key until…  
SETUP X.X  
CHANNEL ENABLE:ON  
<SET SET> EDIT PRINT  
EXIT  
ENTR accepts the new  
setting and returns to the  
previous menu.  
EXIT ignores the new setting  
and returns to the previous  
menu.  
SETUP X.X  
CHANNEL ENABLE:ON  
Toggle key to turn  
channel ON or OFF  
OFF  
ENTR EXIT  
6.12.2.11. HOLDOFF Feature  
The iDAS HOLDOFF feature allows to prevent data collection during calibrations and during the  
DAS_HOLDOFF period enabled and specified in the VARS (Section 6.8). To enable or disable the  
HOLDOFF, follow the instruction shown in section 6.12.2.2 then press:  
From the DATA ACQUISITION menu  
(see Section 6.12.2.2)  
Edit Data Channel Menu  
SETUP X.X  
0) CONC: ATIMER, 1,  
900  
Exits to the main  
Data Acquisition  
menu  
PREV NEXT  
INS DEL EDIT PRNT EXIT  
SETUP X.X  
NAME:CONC  
<SET SET> EDIT PRINT  
EXIT  
Press SET> key until…  
SETUP X.X  
CAL HOLD OFF:ON  
SET> EDIT PRINT  
EXIT  
ENTR accepts the new  
setting and returns to the  
previous menu.  
EXIT ignores the new setting  
and returns to the previous  
menu.  
SETUP X.X  
CAL HOLD OFF:ON  
Toggle key to turn  
HOLDOFF ON or OFF  
ON  
ENTR EXIT  
04584 Rev A1  
112  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.12.3. Remote iDAS Configuration  
Editing channels, parameters and triggering events as described in this section can be performed  
via the APICOM remote control program using the graphic interface shown in Figure 6-8. Refer to  
Section 6.13 for details on remote access to the MGFC7000E analyzer.  
Figure 6-8:  
APICOM user interface for configuring the iDAS.  
Once an iDAS configuration is edited (which can be done offline and without interrupting DAS data  
collection), it is conveniently uploaded to the instrument and can be stored on a computer for  
later review, alteration or documentation and archival. Refer to the APICOM manual for details on  
these procedures. The APICOM user manual (Teledyne Instruments part number 039450000) is  
included in the APICOM installation file, which can be downloaded at http://www.teledyne-  
api.com/software/apicom/.  
Although Teledyne Analytical Instruments recommends the use of APICOM, the iDAS can also be  
accessed and configured through a terminal emulation program such as HyperTerminal (Figure  
6-9). However, all configuration commands must be created following a strict syntax or be pasted  
in from of a text file, which was edited offline and then uploaded through a specific transfer  
procedure.  
04584 Rev A1  
113  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
Figure 6-9:  
iDAS Configuration Through a Terminal Emulation Program.  
Both procedures are best started by downloading the default iDAS configuration, getting familiar  
with its command structure and syntax conventions, and then altering a copy of the original file  
offline before uploading the new configuration.  
CAUTION  
Whereas the editing, adding and deleting of iDAS channels and parameters of one  
channel through the front-panel keyboard can be done without affecting the other  
channels, uploading an iDAS configuration script to the analyzer through its  
communication ports will erase all data, parameters and channels by replacing them  
with the new iDAS configuration. Backup of data and the original iDAS configuration is  
advised before attempting any iDAS changes.  
04584 Rev A1  
114  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.13. Remote Operation of the Analyzer  
6.13.1. Remote Operation Using the External Digital I/O  
6.13.1.1. Status Outputs  
The status outputs report analyzer conditions via optically isolated NPN transistors, which sink up  
to 50 mA of DC current. These outputs can be used interface with devices that accept logic-level  
digital inputs, such as programmable logic controllers (PLC’s). Each Status bit is an open collector  
output that can withstand up to 40 VDC. All of the emitters of these transistors are tied together  
and available at D.  
NOTE  
Most PLC’s have internal provisions for limiting the current that the input will draw from  
an external device. When connecting to a unit that does not have this feature, an  
external dropping resistor must be used to limit the current through the transistor  
output to less than 50 mA. At 50 mA, the transistor will drop approximately 1.2V from  
its collector to emitter.  
The status outputs are accessed via a 12-pin connector on the analyzer’s rear panel labeled  
STATUS. The function of each pin is defined in Table 6–24.  
STATUS  
1
2
3
4
5
6
7
8
D
+
Figure 6-10: Status Output Connector  
04584 Rev A1  
115  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
The pin assignments for the Status Outputs are:  
Table 6-24: Status Output Pin Assignments  
Status  
Definition  
Output #  
Condition  
1
2
SYSTEM OK  
On if no faults are present.  
On if CO2 concentration measurement is valid.  
If the CO2 concentration measurement is invalid, this bit is OFF.  
CONC VALID  
3
4
5
6
HIGH RANGE  
ZERO CAL  
On if unit is in high range of DUAL or AUTO range modes.  
On whenever the instruments zero point is being calibrated.  
On whenever the instruments span point is being calibrated.  
On whenever the instrument is in diagnostic mode.  
SPAN CAL  
DIAG MODE  
On whenever the measured CO2 concentration is above the set  
point for ALM1  
7
8
ALARM1  
ALARM2  
On whenever the measured CO2 concentration is above the set  
point for ALM2  
D
+
EMITTER BUSS  
DC POWER  
The emitters of the transistors on pins 1-8 are bussed together.  
+ 5 VDC  
Digital Ground  
The ground level from the analyzer’s internal DC power supplies.  
6.13.1.2. Control Inputs  
These inputs allow the user to remotely initiate Zero and Span calibrations. Two methods for  
energizing the inputs is provided below; the first using the internal +5V available on the CONTROL  
IN connector and the second, if an external, isolated supply is employed.  
Table 6-25: Control Input Pin Assignments  
Input  
Status  
Condition when enabled  
EXTERNAL ZERO  
CAL  
Zero calibration mode is activated. The mode field of the  
display will read ZERO CAL R.  
A
B
EXTERNAL SPAN  
CAL  
Span calibration mode is activated. The mode field of the  
display will read SPAN CAL R.  
C
D
E
F
Unused  
Unused  
Unused  
Unused  
DIGITAL GROUND Provided to ground an external device (e.g., recorder).  
DC power for  
Input pull ups  
Input for +5 VDC required to activate inputs A - F. This voltage  
can be taken from an external source or from the “+” pin.  
U
Internal +5V  
Supply  
Internal source of +5V which can be used to activate inputs  
when connected to pin U.  
+
04584 Rev A1  
116  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
There are two methods to activate control inputs. The internal +5V available from the “+” pin is  
the most convenient method (Figure 6.11). However, to ensure that these inputs are truly  
isolated, a separate, external 5 VDC power supply should be used (Figure 6.11).  
CONTROL IN  
CONTROL IN  
A
B
C
D
E
F
U +  
A
B
C
D
E
F
U
+
Z
E
R
S
P
A
N
Z
E
R
O
S
P
A
N
O
5 VDC Power  
Supply  
+
-
External Power Connections  
Local Power Connections  
Figure 6-11: Control Inputs  
6.13.2. Remote Operation Using the External Serial I/O  
6.13.2.1. Terminal Operating Modes  
The Model GFC7000E can be remotely configured, calibrated or queried for stored data through  
the serial ports. As terminals and computers use different communication schemes, the analyzer  
supports two communicate modes specifically designed to interface with these two types of  
devices.  
Computer mode is used when the analyzer is connected to a computer with a dedicated  
interface program such as APICOM. More information regarding APICOM can be found in  
later in this section or on the Teledyne Instruments website at http://www.teledyne-  
api.com/software/apicom/.  
Interactive mode is used with a terminal emulation programs such as HyperTerminal or a  
“dumb” computer terminal. The commands that are used to operate the analyzer in this  
mode are listed in Table 6-26.  
04584 Rev A1  
117  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.13.2.2. Help Commands in Terminal Mode  
Table 6-26: Terminal Mode Software Commands  
COMMAND  
Control-T  
Function  
Switches the analyzer to terminal mode (echo, edit). If mode flags 1 & 2 are OFF,  
the interface can be used in interactive mode with a terminal emulation program.  
Control-C  
Switches the analyzer to computer mode (no echo, no edit).  
CR  
A carriage return is required after each command line is typed into the  
terminal/computer. The command will not be sent to the analyzer to be executed  
until this is done. On personal computers, this is achieved by pressing the ENTER  
key.  
(carriage return)  
BS  
Erases one character to the left of the cursor location.  
(backspace)  
ESC  
Erases the entire command line.  
(escape)  
? [ID] CR  
This command prints a complete list of available commands along with the  
definitions of their functionality to the display device of the terminal or computer  
being used. The ID number of the analyzer is only necessary if multiple analyzers  
are on the same communications line, such as the multi-drop setup.  
Control-C  
Control-P  
Pauses the listing of commands.  
Restarts the listing of commands.  
6.13.2.3. Command Syntax  
Commands are not case-sensitive and all arguments within one command (i.e. ID numbers,  
keywords, data values, etc.) must be separated with a space character.  
All Commands follow the syntax:  
X [ID] COMMAND <CR>  
Where  
X
is the command type (one letter) that defines the type of command. Allowed  
designators are listed in Table 6-27 and Appendix A-6.  
[ID]  
is the analyzer identification number (Section 6.10.1.). Example: the Command  
“? 200” followed by a carriage return would print the list of available commands  
for the revision of software currently installed in the instrument assigned ID  
Number 200.  
COMMAND is the command designator: This string is the name of the command being  
issued (LIST, ABORT, NAME, EXIT, etc.). Some commands may have additional  
arguments that define how the command is to be executed. Press ? <CR> or  
refer to Appendix A-6 for a list of available command designators.  
<CR>  
is a carriage return. All commands must be terminated by a carriage return  
(usually achieved by pressing the ENTER key on a computer).  
04584 Rev A1  
118  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Table 6-27: Command Types  
Operating Instructions  
COMMAND  
COMMAND TYPE  
Calibration  
Diagnostic  
C
D
L
Logon  
T
Test measurement  
Variable  
V
W
Warning  
6.13.2.4. Data Types  
Data types consist of integers, hexadecimal integers, floating-point numbers, Boolean expressions  
and text strings.  
Integer data are used to indicate integral quantities such as a number of records, a filter  
length, etc. They consist of an optional plus or minus sign, followed by one or more digits.  
For example, +1, -12, 123 are all valid integers.  
Hexadecimal integer data are used for the same purposes as integers. They consist of the  
two characters “0x,” followed by one or more hexadecimal digits (0-9, A-F, a-f), which is  
the ‘C’ programming language convention. No plus or minus sign is permitted. For  
example, 0x1, 0x12, 0x1234abcd are all valid hexadecimal integers.  
Floating-point numbers are used to specify continuously variable values such as  
temperature set points, time intervals, warning limits, voltages, etc. They consist of an  
optional plus or minus sign, followed by zero or more digits, an optional decimal point, and  
zero or more digits. (At least one digit must appear before or after the decimal point.)  
Scientific notation is not permitted. For example, +1.0, 1234.5678, -0.1, 1 are all valid  
floating-point numbers.  
Boolean expressions are used to specify the value of variables or I/O signals that may  
assume only two values. They are denoted by the keywords ON and OFF.  
Text strings are used to represent data that cannot be easily represented by other data  
types, such as data channel names, which may contain letters and numbers. They consist  
of a quotation mark, followed by one or more printable characters, including spaces,  
letters, numbers, and symbols, and a final quotation mark. For example, “a”, “1”,  
“123abc”, and “()[]<>” are all valid text strings. It is not possible to include a quotation  
mark character within a text string.  
Some commands allow you to access variables, messages, and other items, such as iDAS  
data channels, by name. When using these commands, you must type the entire name of  
the item; you cannot abbreviate any names.  
04584 Rev A1  
119  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.13.2.5. Status Reporting  
Reporting of status messages as an audit trail is one of the three principal uses for the RS-232  
interface (the other two being the command line interface for controlling the instrument and the  
download of data in electronic format). You can effectively disable the reporting feature by setting  
the interface to quiet mode (Section 6.10.6., Table 6-16).  
Status reports include iDAS data (when reporting is enabled), warning messages, calibration and  
diagnostic status messages. Refer to Appendix A-3 for a list of the possible messages, and this  
section for information on controlling the instrument through the RS-232 interface.  
General Message Format  
All messages from the instrument (including those in response to a command line request) are in  
the format:  
X DDD:HH:MM [Id] MESSAGE<CRLF>  
Where:  
X
is a command type designator, a single character indicating the message  
type, as shown in the Table 6-27.  
DDD:HH:MM is the time stamp, the date and time when the message was issued. It  
consists of the Day-of-year (DDD) as a number from 1 to 366, the hour of  
the day (HH) as a number from 00 to 23, and the minute (MM) as a number  
from 00 to 59.  
[ID]  
is the analyzer ID, a number with 1 to 4 digits.  
MESSAGE  
is the message content that may contain warning messages, test  
measurements, iDAS reports, variable values, etc.  
<CRLF>  
is a carriage return / line feed pair, which terminates the message.  
The uniform nature of the output messages makes it easy for a host computer to parse them into  
an easy structure. Keep in mind that the front panel display does not give any information on the  
time a message was issued, hence it is useful to log such messages for trouble-shooting and  
reference purposes. Terminal emulation programs such as HyperTerminal can capture these  
messages to text files for later review.  
6.13.2.6. Remote Access by Modem  
The MGFC7000E can be connected to a modem for remote access. This requires a cable between  
the analyzer’s COM port and the modem, typically a DB-9F to DB-25M cable (available from  
Teledyne Instruments with part number WR0000024).  
Once the cable has been connected, check to make sure the DTE-DCE is in the correct position.  
Also make sure the MGFC7000E COM port is set for a baud rate that is compatible with the  
modem, which needs to operate with an 8-bit word length with one stop bit.  
The first step is to turn on the MODEM ENABLE communication mode (Mode 64, Section 6.10.6).  
Once this is completed, the appropriate setup command line for your modem can be entered into  
the analyzer. The default setting for this feature is  
AT Y0 &D0 &H0 &I0 S0=2 &B0 &N6 &M0 E0 Q1 &W0  
04584 Rev A1  
120  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
This string can be altered to match your modem’s initialization and can be up to 100 characters  
long.  
To change this setting press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
SET> EDIT  
COM1 MODE:0  
EXIT  
EXIT  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
SETUP X.X  
COM1 BAUD RATE:19200  
<SET SET> EDIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
SETUP X.X  
COM1 MODEM INIT:AT Y&D&H  
EXIT  
EXIT returns  
to the  
<SET SET> EDIT  
SETUP X.X  
SECONDARY SETUP MENU  
previous  
menu  
ENTR accepts the  
new string and returns  
to the previous menu.  
EXIT ignores the new  
string and returns to  
the previous menu.  
COMM VARS DIAG ALRM  
EXIT  
EXIT  
SETUP X.X  
COM1 MODEM INIT:[A]T Y&D&H  
ENTR EXIT  
<CH CH> INS DEL [A]  
SETUP X.X  
COMMUNICATIONS MENU  
Select which  
COM Port is  
tested  
ID COM1 COM2  
Press the [?]  
key repeatedly to cycle through the  
available character set:  
0-9  
The INS key  
inserts a character  
before the cursor  
location.  
The DEL key  
deletes a character  
at the cursor  
location.  
A-Z  
The <CH and CH> keys move  
the [ ] cursor left and right  
along the text string  
space ’ ~ ! © # $ % ^ & * ( ) - _ =  
+[ ] { } < >\ | ; : , . / ?  
To Initialize the modem press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP  
< TST TST > CAL  
SETUP X.X  
SET> EDIT  
COM1 MODE:0  
EXIT  
EXIT  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
SETUP X.X  
COM1 BAUD RATE:19200  
<SET SET> EDIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT returns  
to the  
previous  
SETUP X.X  
COM1 MODEM INIT:AT Y&D&H  
SETUP X.X  
SECONDARY SETUP MENU  
<SET SET> EDIT  
EXIT  
menu  
COMM VARS DIAG ALRM  
EXIT  
SETUP X.X  
COM1 INITIALIZE MODEM  
EXIT  
SETUP X.X  
COMMUNICATIONS MENU  
Select which  
COM Port is  
tested  
<SET SET> INIT  
ID COM1 COM2  
EXIT  
SETUP X.X  
INITIALIZING MODEM  
<SET SET> INIT  
EXIT  
EXIT returns to the  
Communications Menu.  
04584 Rev A1  
121  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.13.2.7. COM Port Password Security  
In order to provide security for remote access of the MGFC7000E, a LOGON feature can be  
enabled to require a password before the instrument will accept commands. This is done by  
turning on the SECURITY MODE (Mode 4, Section 6.10.6). Once the SECURITY MODE is  
enabled, the following items apply.  
A password is required before the port will respond or pass on commands.  
If the port is inactive for one hour, it will automatically logoff, which can also be achieved  
with the LOGOFF command.  
Three unsuccessful attempts to log on with an incorrect password will cause subsequent  
logins to be disabled for 1 hour, even if the correct password is used.  
If not logged on, the only active command is the '?' request for the help screen.  
The following messages will be returned at logon:  
LOGON SUCCESSFUL - Correct password given  
LOGON FAILED - Password not given or incorrect  
LOGOFF SUCCESSFUL - Connection terminated successfully  
To log on to the model GFC7000E analyzer with SECURITY MODE feature enabled, type:  
LOGON 940331  
940331 is the default password. To change the default password, use the variable RS232_PASS  
issued as follows:  
V RS232_PASS=NNNNNN  
Where N is any numeral between 0 and 9.  
6.13.2.8. APICOM Remote Control Program  
APICOM is an easy-to-use, yet powerful interface program that allows to access and control any of  
Teledyne Instruments’ main line of ambient and stack-gas instruments from a remote connection  
through direct cable, modem or Ethernet. Running APICOM, a user can:  
Establish a link from a remote location to the MGFC7000E through direct cable connection via RS-  
232 modem or Ethernet.  
View the instrument’s front panel and remotely access all functions that could be accessed  
when standing in front of the instrument.  
Remotely edit system parameters and set points.  
Download, view, graph and save data for predictive diagnostics or data analysis.  
Retrieve, view, edit, save and upload iDAS configurations.  
04584 Rev A1  
122  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
Check on system parameters for trouble-shooting and quality control.  
APICOM is very helpful for initial setup, data analysis, maintenance and trouble-shooting. Figure  
6-8 shows an example of APICOM being used to remotely configuration the instruments iDAS  
feature. Figure 6-12 shows examples of APICOM’s main interface, which emulates the look and  
functionality of the instruments actual front panel  
Figure 6-12: APICOM Remote Control Program Interface  
APICOM is included free of cost with the analyzer and the latest versions can also be downloaded  
for free at http://www.teledyne-api.com/software/apicom/.  
6.13.3. Additional Communications Documentation  
Table 6-28: Serial Interface Documents  
INTERFACE / TOOL  
DOCUMENT TITLE  
PART  
NUMBER  
AVAILABLE  
ONLINE*  
APICOM  
Multi-drop  
DAS Manual  
APICOM User Manual  
039450000  
021790000  
028370000  
YES  
YES  
YES  
RS-232 Multi-drop Documentation  
Detailed description of the iDAS.  
* These documents can be downloaded at http://www.teledyne-api.com/manuals/  
6.13.4. Using the GFC7000E with a Hessen Protocol Network  
6.13.4.1. General Overview of Hessen Protocol  
The Hessen protocol is a multidrop protocol, in which several remote instruments are connected  
via a common communications channel to a host computer. The remote instruments are regarded  
as slaves of the host computer. The remote instruments are unaware that they are connected to  
a multidrop bus and never initiate Hessen protocol messages. They only respond to commands  
from the host computer and only when they receive a command containing their own unique ID  
number.  
04584 Rev A1  
123  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
The Hessen protocol is designed to accomplish two things: to obtain the status of remote  
instruments, including the concentrations of all the gases measured; and to place remote  
instruments into zero or span calibration or measure mode. API’s implementation supports both of  
these principal features.  
The Hessen protocol is not well defined, therefore while API’s application is completely compatible  
with the protocol itself, it may be different from implementations by other companies.  
The following subsections describe the basics for setting up your instrument to operate over a  
Hessen Protocol network. for more detailed information as well as a list of host computer  
commands and examples of command and response message syntax, download the Manual  
Addendum for Hessen Protocol from the Teledyne Instruments web site: http://www.teledyne-  
ai.com/manuals/index.asp .  
6.13.4.2. Hessen COMM Port Configuration  
Hessen protocol requires the communication parameters of the MGFC7000E’s COMM ports to be  
set differently than the standard configuration as shown in the table below.  
Table 6-29: RS-232 Communication Parameters for Hessen Protocol  
Parameter  
Data Bits  
Stop Bits  
Parity  
Standard  
Hessen  
8
1
7
2
None  
Full  
Even  
Half  
Duplex  
To change the rest of the COMM port parameters. see Section 6.10.6.  
To change the baud rate of the GFC7000E’s COMM ports, see Section 6.10.7.  
NOTE  
Make sure that the communication parameters of the host computer are also properly  
set.  
Also, the instrument software has a 200 ms. latency before it responds to commands  
issued by the host computer. This latency should present no problems, but you should  
be aware of it and not issue commands to the instrument too frequently.  
6.13.4.3. Activating Hessen Protocol  
The first step in configuring the GFC7000E to operate over a Hessen protocol network is to  
activate the Hessen mode for COMM ports and configure the communication parameters for the  
port(s) appropriately. Press:  
04584 Rev A1  
124  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP X.X  
COM1 QUIET MODE: OFF  
ENTR EXIT  
Repeat the  
entire process to  
set up the  
< TST TST > CAL  
SETUP  
NEXT OFF  
COM2 port  
SAMPLE  
ENTER SETUP PASS : 818  
Continue pressing next until …  
8
1
8
ENTR EXIT  
SETUP X.X COM1 HESSEN PROTOCOL : OFF  
SETUP X.X  
PRIMARY SETUP MENU  
PREV NEXT OFF  
ENTR EXIT  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
EXIT  
EXIT  
Toggle OFF/ON keys  
to change  
SETUP X.X COM1 HESSEN PROTOCOL : ON  
SETUP X.X SECONDARY SETUP MENU  
activate/deactivate  
selected mode.  
PREV NEXT ON  
ENTR EXIT  
COMM VARS DIAG  
ALRM  
SETUP X.X  
COMMUNICATIONS MENU  
SETUP X.X  
COM1 E,7,1 MODE: OFF  
Select which COMM  
port to configure  
ID COM1 COM2  
PREV NEXT OFF  
ENTR EXIT  
The sum of the mode  
IDs of the selected  
modes is displayed  
here  
SETUP X.X  
SET> EDIT  
COM1 MODE:0  
SETUP X.X  
COM1 E,7,1 MODE: ON  
ENTR key accepts the  
new settings  
PREV NEXT ON  
ENTR EXIT  
EXIT key ignores the  
new settings  
6.13.4.4. Selecting a Hessen Protocol Type  
Currently there are two versions of Hessen Protocol in use. The original implementation, referred  
to as TYPE 1, and a more recently released version, TYPE 2 that has more flexibility when  
operating with instruments that can measure more than one type of gas. For more specific  
information about the difference between TYPE 1and TYPE 2 download the Manual Addendum for  
Hessen Protocol from the Teledyne Instruments web site: http://www.teledyne-  
ai.com/manuals/index.asp .  
To select a Hessen Protocol Type press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SETUP X.X  
COMMUNICATIONS MENU  
SAMPLE  
ENTER SETUP PASS : 818  
ID HESN COM1 COM2  
EXIT  
8
1
8
ENTR EXIT  
SETUP X.  
HESSEN VARIATION: TYPE 1  
SETUP X.X  
PRIMARY SETUP MENU  
SET> EDIT  
EXIT  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
ENTR key accepts the  
new settings  
SETUP X.X HESSEN VARIATION: TYPE 1  
TYE1 TYPE 2 ENTR EXIT  
EXIT key ignores the  
SETUP X.X  
SECONDARY SETUP MENU  
new settings  
COMM VARS DIAG ALRM  
SETUP X.X HESSEN VARIATION: TYPE 2  
PREV NEXT OFF ENTR EXIT  
Press to change  
protocol type.  
04584 Rev A1  
125  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
NOTE  
While Hessen Protocol Mode can be activated independently for COM1 and COM2, The  
TYPE selection affects both Ports.  
6.13.4.5. Setting The Hessen Protocol Response Mode  
The Teledyne Instruments implementation of Hessen Protocol allows the user to choose one of  
several different modes of response for the analyzer.  
Table 6-30: Teledyne Instruments Hessen Protocol Response Modes  
MODE ID  
CMD  
MODE DESCRIPTION  
This is the Default Setting. Reponses from the instrument are encoded as the traditional  
command format. Style and format of responses depend on exact coding of the initiating  
command.  
Responses from the instrument are always delimited with <STX> (at the beginning of the  
response, <ETX> (at the end of the response followed by a 2 digit Block Check Code  
(checksum), regardless of the command encoding.  
BCC  
Responses from the instrument are always delimited with <CR> at the beginning and the end  
of the string, regardless of the command encoding.  
TEXT  
To Select a Hessen response mode, press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
SETUP X.X  
COMMUNICATIONS MENU  
ID HESN COM1 COM2  
EXIT  
EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
SETUP X.X  
HESSEN VARIATION: TYPE 1  
SET> EDIT  
SETUP X.X  
SECONDARY SETUP MENU  
ENTR key accepts the  
new settings  
COMM VARS DIAG ALRM  
EXIT key ignores the  
SETUP X.X  
HESSEN RESPONSE MODE :CMD  
new settings  
<SET SET> EDIT  
EXIT  
Press to  
change  
response  
mode.  
SETUP X.X  
HESSEN RESPONSE MODE :CMD  
BCC TEXT EDIT  
ENTR EXIT  
6.13.4.6. Hessen Protocol Gas ID  
The Model GFC7000E Analyzer is a single gas instrument that measures CO2. As such it’s default  
gas ID has already been set to 310. There is no need to change this setting.  
04584 Rev A1  
126  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
6.13.4.7. Setting Hessen Protocol Status Flags  
Teledyne Instruments’ implementation of Hessen protocols includes a set of status bits that the  
instrument includes in responses to inform the host computer of its condition. Each bit can be  
assigned to one operational and warning message flag. The default settings for these bit/flags  
are:  
Table 6-31: Default Hessen Status Bit Assignments  
STATUS FLAG NAME  
DEFAULT BIT  
ASSIGNMENT  
WARNING FLAGS  
SAMPLE FLOW WARNING  
BENCH TEMP WARNING  
SOURCE WARNING  
0001  
0002  
0004  
0008  
0010  
0020  
0040  
0080  
BOX TEMP WARNING  
WHEEL TEMP WARNING  
SAMPLE TEMP WARNING  
SAMPLE PRESSURE WARNING  
INVALID CONC  
(The Instrument’s Front Panel Display Will Show  
The Concentration As “XXXX”)  
OPERATIONAL FLAGS  
Instrument Off  
0100  
0200  
0400  
0800  
In Manual Calibration Mode  
In Zero Calibration Mode  
In Span Calibration Mode  
UNITS OF MEASURE FLAGS  
UGM  
0000  
2000  
MGM  
PPB  
4000  
PPM  
6000  
SPARE/UNUSED BITS  
UNASSIGNED FLAGS (0000)  
100, 1000, 8000  
Sync Warning  
Relay Board Warning  
Front Panel Warning  
Conc Alarm 1  
Conc Alarm 2  
Analog Cal Warning  
Cannot Dyn Zero  
Cannot Dyn Span  
Invalid Conc  
Photo Temp Warning  
System Reset  
Rear Board Not Detected  
NOTES:  
It is possible to assign more than one flag to the same Hessen status bit. This allows  
the grouping of similar flags, such as all temperature warnings, under the same status  
bit.  
Be careful not to assign conflicting flags to the same bit as each status bit will be  
triggered if any of the assigned flags is active.  
04584 Rev A1  
127  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Operating Instructions  
To assign or reset the status flag bit assignments, press:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
SETUP X.X  
SECONDARY SETUP MENU  
COMM VARS DIAG ALRM  
EXIT  
EXIT  
SETUP X.X  
COMMUNICATIONS MENU  
ID HESN COM1 COM2  
Repeat pressing SET> until …  
SETUP X.  
HESSEN STATUS FLAGS  
<SET SET> EDIT  
EXIT  
SETUP X.  
SYNC WARNING: 0000  
PREV NEXT  
EDIT PRNT EXIT  
Repeat pressing NEXT or PREV until the desired  
message flag is displayed. See Table 6-27.  
For xxample …  
SETUP X.  
SYSTEM RESET: 0000  
EDIT PRNT EXIT  
PREV NEXT  
The <CH and  
CH> keys move  
the [ ] cursor left  
and right along  
the bit string.  
SETUP X.  
SYSTEM RESET: [0]000  
[0]  
ENTR key accepts the  
new settings  
<CH CH>  
ENTR EXIT  
EXIT key ignores the  
new settings  
Press the [?] key repeatedly to cycle through the available character set: 0-9  
Note: Values of A-F can also be set but are meaningless.  
6.13.4.8. Instrument ID Code  
Each instrument on a Hessen Protocol network must have a unique ID code. The MGFC7000E is  
programmed with a default ID code of 360. To change this code see Section 6.10.1  
User Notes  
04584 Rev A1  
128  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Calibration Procedures  
7. CALIBRATION PROCEDURES  
This section contains a variety of information regarding the various methods for calibrating a  
Model GFC7000E CO2 Analyzer as well as other supporting information.  
NOTE  
The procedures in this section assume that the calibration password feature is disabled  
(the instruments default state). If it is enabled a password prompt screen (see Section  
6.3.1) will appear after the CAL, CALZ or CALS buttons are pushed but before the  
instrument enters the associated calibration mode.  
7.1. Before Calibration  
The calibration procedures in this section assume that the Range Type, Range Span and units of  
measure have already been selected for the analyzer. If this has not been done, please do so  
before continuing (see Section 6.7 for instructions).  
All Gas lines should be PTFE (Teflon), FEP, glass, stainless steel or brass.  
NOTE  
If any problems occur while performing the following calibration procedures, refer to  
Chapter 11 of this manual for troubleshooting tips.  
7.1.1. Zero Air and Span Gas  
To perform the following calibration you must have sources for zero air and span gas available.  
Zero Air is similar in chemical composition to the Earth’s atmosphere but scrubbed of all  
components that might affect the analyzer’s readings. Zero air should contain less than 25 ppb of  
CO2 and other major interfering gases such as CO and Water Vapor. It should have a dew point  
of -5°C or less  
Span Gas is a gas specifically mixed to match the chemical composition of the type of gas being  
measured at near full scale of the desired measurement range. It is recommended that the span  
gas used have a concentration equal to 80% of the full measurement range.  
If Span Gas is sourced directly from a calibrated, pressurized tank, the gas mixture should be CO2  
mixed with Zero Air or N2 at the required ratio.  
Zero air generators that condition ambient air by drying and removal of pollutants are available on  
the commercial market such as the Teledyne Instruments Model 701 Zero Air Generator. We  
recommend this type of device, in conjunction with a CO2 scrubber such as soda-lime, for  
generating zero air.  
04584 Rev A1  
129  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Calibration Procedures  
7.1.2. Calibration Gas Traceability  
All equipment used to produce calibration gases should be verified against standards of the  
National Institute for Standards and Technology (NIST). To ensure NIST traceability, we  
recommend to acquire cylinders of working gas that are certified to be traceable to NIST Standard  
Reference Materials (SRM). These are available from a variety of commercial sources.  
7.1.3. Data Recording Devices  
A strip chart recorder, data acquisition system or digital data acquisition system should be used to  
record data from the MGFC7000E’s serial or analog outputs. If analog readings are used, the  
response of the recording system should be checked against a NIST traceable voltage source or  
meter. Data recording device should be capable of bi-polar operation so that negative readings  
can be recorded. For electronic data recording, the MGFC7000E provides an internal data  
acquisition system (iDAS), which is described in detail in Section 6.12.  
7.2. Manual Calibration without Zero/Span Valves  
This is the basic method for manually calibrating the Model GFC7000E CO2 Analyzer without  
functioning zero/span valve options. It is identical to the method described in the GETTING  
STARTED (Chapter 3) of this manual and is repeated her for you convenience.  
STEP ONE: Connect the Sources of Zero Air and Span Gas as shown below.  
Calibrated CO2  
gas at desired  
VENT  
Source of  
SAMPLE Gas  
Removed  
during  
Calibration  
span gas  
concentration  
Needle  
valve to  
control  
flow  
Indicating  
soda-lime  
Sample  
Exhaust  
MODEL  
GFC7000E  
Valve  
Vent Span  
Pressure Span  
MODEL 701  
Zero Air  
IZS  
Generator  
Purge In  
Figure 7-1:  
Pneumatic Connections–Basic Configuration–Using Bottled Span Gas  
04584 Rev A1  
130  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Calibration Procedures  
Source of  
SAMPLE Gas  
Removed  
during  
Calibrated  
CO2 Gas  
MODEL 700  
Gas Dilution  
Calibrator  
Calibration  
Indicating  
soda-lime  
VENT  
Sample  
Exhaust  
Vent Span  
Pressure Span  
IZS  
MODEL  
GFC7000E  
MODEL 701  
Zero Air Generator  
Purge In  
Figure 7-2:  
Pneumatic Connections–Basic Configuration–Using Gas Dilution Calibrator  
STEP TWO: Set the expected CO2 Span Gas concentration:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP  
< TST TST > CAL  
This sequence causes the  
analyzer to prompt for the  
expected COx span  
concentration.  
M-P CAL  
RANGE = 500.000 PPM  
CO2 =X.XXX  
EXIT  
< TST TST > ZERO  
CONC  
The SO2 span  
concentration values  
automatically default to  
400.0 Conc.  
EXIT ignores the new setting  
and returns to the previous  
display.  
M-P CAL  
SO2 SPAN CONC: 400.000 Conc  
To change this value to  
the actual concentration of  
the span gas, enter the  
number by pressing the  
key under each digit until  
the expected value  
ENTR accepts the new setting  
0
0
0
4
5
.0  
ENTR EXIT  
and returns to the  
previous display..  
appears.  
NOTE  
For this Initial Calibration it is important to independently verify the PRECISE CO2  
Concentration Value of the SPAN gas.  
If the source of the Span Gas is from a Calibrated Bottle, use the exact concentration  
value printed on the bottle.  
04584 Rev A1  
131  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Calibration Procedures  
STEP THREE: Perform the Zero/Span Calibration Procedure:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =XXX.X  
SETUP  
Set the Display to show the  
STABIL test function.  
This function calculates the  
stability of the COx  
< TST TST > CAL  
measurement  
SAMPLE  
STABIL=XXX.X PPM  
CO2 =XXX.X  
SETUP  
< TST TST > CAL  
ACTION:  
Allow zero gas to enter the sample port at the  
rear of the instrument.  
Wait until STABIL  
falls below 1.0 ppb.  
This may take several  
minutes.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
< TST TST > CAL  
SETUP  
M-P CAL  
STABIL=XXX.X PPM  
CONC  
CO2 =XXX.X  
EXIT  
< TST TST > ZERO  
Press ENTR to changes the  
OFFSET & SLOPE values for the  
CO2 measurements.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
< TST TST > ENTR  
CONC  
EXIT  
Press EXIT to leave the calibration  
unchanged and return to the  
previous menu.  
ACTION:  
Allow span gas to enter the sample port at the  
rear of the instrument.  
The value of  
STABIL may jump  
significantly.  
Wait until it falls back  
below 1.0 ppb.  
The SPAN key now  
appears during the  
transition from zero to  
span.  
This may take several  
minutes.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
EXIT  
< TST TST >  
M-P CAL  
SPAN CONC  
You may see both keys.  
If either the ZERO or  
SPAN buttons fail to  
appear see Section 11  
for troubleshooting tips.  
Press ENTR to change the  
OFFSET & SLOPE values for the  
CO2 measurements.  
RANGE = 500.000 PPM CO2 =XXX.X  
EXIT  
Press EXIT to leave the calibration  
unchanged and return to the  
previous menu.  
< TST TST > ENTR SPAN CONC  
M-P CAL  
RANGE = 500.000 PPM CO2 =XXX.X  
CONC  
EXIT  
EXIT returns to the main  
SAMPLE display  
< TST TST > ENTR  
If the ZERO or SPAN keys are not displayed, this means that the measurement made during that  
part of the procedure is too far out of the allowable range to do allow a reliable calibration. The  
reason for this must be determined before the analyzer can be calibrated. See Chapter 11 for  
troubleshooting tips.  
04584 Rev A1  
132  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Calibration Procedures  
7.3. Manual Calibration Checks  
Informal calibration checks, which only evaluate but do not alter the analyzer’s response curve,  
are recommended as a regular maintenance item and in order to monitor the analyzer’s  
performance. To carry out a calibration check rather than a full calibration, follow these steps.  
STEP ONE: Connect the sources of zero air and span gas as shown in Figures 7.1 or 7.2.  
STEP TWO: Perform the zero/span calibration check procedure:  
ACTION:  
Supply the instrument with zero gas.  
SAMPLE  
RANGE = 500.0 PPM  
CO2=X.XXX  
SETUP  
Scroll the display to the  
STABIL test function.  
< TST TST > CAL  
SAMPLE  
STABIL=XXX.X PPM  
CO2=X.XXX  
SETUP  
< TST TST > CAL  
Wait until  
STABIL is  
below 1.0 ppb.  
This may take  
several minutes.  
ACTION:  
Record the CO2  
concentration  
reading.  
SAMPLE  
STABIL=XXX.X PPM  
CO2=X.XXX  
< TST TST > CAL  
SETUP  
The value of  
STABIL may jump  
significantly.  
ACTION:  
Supply span gas to the instrument  
Wait until it falls  
below 1.0 ppb. This  
may take several  
minutes.  
ACTION:  
Record the CO2  
concentration  
reading.  
SAMPLE  
STABIL=XXX.X PPM  
CO2=X.XXX  
SETUP  
< TST TST > CAL  
The SPAN key appears during the transition from zero to  
span. You may see both keys.  
7.4. Manual Calibration with Zero/Span Valves  
There are four different zero/span valve option configurations (see Section 5.4). They all operate  
identically, differing only in the method used to supply calibration gas to the Analyzer.  
STEP ONE: Connect the sources of Zero Air and Span Gas as shown below.  
Figures 7-3 through 7-6 show the proper pneumatic connections for GFC7000E’s with various  
optional internal valve sets installed.  
04584 Rev A1  
133  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Calibration Procedures  
Source of  
SAMPLE Gas  
VENT if input is pressurized  
Certified  
CO2 Gas  
Sample  
Exhaust  
Needle  
valve to  
control  
MODEL  
GFC7000E  
VENT  
Vent Span  
flow  
MODEL 701  
Zero Air  
Pressure Span  
Generator  
IZS  
VENT  
Purge In  
Indicating  
soda-lime  
Figure 7-3:  
Pneumatic Connections–MGFC7000E with Zero/Span/Shutoff Valves (OPT  
50)  
Source of  
SAMPLE Gas  
VENT if input is pressurized  
Certified  
CO2 Gas  
Sample  
Exhaust  
MODEL 701  
Zero Air  
Generator  
MODEL  
GFC7000E  
VENT  
Vent Span  
Pressure Span  
IZS  
External Zero  
Air Scrubber  
Purge In  
Indicating  
soda-lime  
Figure 7-4  
Pneumatic Connections–MGFC7000E with Zero/Span/Shutoff Valves and  
External Zero Air Scrubber (OPT 51)  
Source of  
SAMPLE Gas  
VENT if input is pressurized  
MODEL 700  
Gas Dilution  
Calibrator  
Sample  
Certified  
CO2 Gas  
Needle  
valve to  
Exhaust  
MODEL  
GFC7000E  
control flow  
Vent Span  
VENT  
Pressure Span  
Indicating  
soda-lime  
IZS  
VENT  
Purge In  
MODEL 701  
Zero Air  
Generator  
04584 Rev A1  
134  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Calibration Procedures  
Figure 7-5:  
Pneumatic Connections–MGFC7000E with Zero/Span Valves (OPT 52)  
Source of  
SAMPLE Gas  
VENT if input is pressurized  
Certified  
CO Gas  
MODEL 700  
Gas Dilution  
Calibrator  
Sample  
Exhaust  
Indicating  
soda-lime  
VENT  
MODEL  
GFC7000E  
Vent Span  
Pressure Span  
External Zero  
Air Scrubber  
IZS  
MODEL 701  
Zero Air Generator  
Purge In  
Figure 7-6:  
Pneumatic Connections–MGFC7000E with Zero/Span Valves with External  
Zero air Scrubber (OPT 53)  
STEP TWO: Set the expected CO2 Span Gas concentration:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP  
< TST TST > CAL CALZ CALS  
This sequence causes the  
analyzer to prompt for the  
expected COx span  
concentration.  
M-P CAL  
RANGE = 500.000 PPM  
CONC  
CO2 =X.XXX  
EXIT  
< TST TST > ZERO  
The SO2 span  
concentration values  
automatically default to  
400.0 Conc.  
EXIT ignores the new setting  
and returns to the previous  
display.  
M-P CAL  
SO2 SPAN CONC: 450.000 Conc  
To change this value to  
the actual concentration of  
the span gas, enter the  
number by pressing the  
key under each digit until  
the expected value  
ENTR accepts the new setting  
0
0
0
4
5
.0  
ENTR EXIT  
and returns to the  
previous display..  
appears.  
NOTE  
For this Initial Calibration it is important to independently verify the PRECISE CO2  
Concentration Value of the SPAN gas.  
If the source of the Span Gas is from a Calibrated Bottle, use the exact concentration  
value printed on the bottle.  
04584 Rev A1  
135  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Calibration Procedures  
STEP THREE: Perform the zero/span calibration. Zero and span checks using the zero/span  
valve option are similar to that described in Section 3.3, except that zero air and span gas is  
supplied to the analyzer through the zero/span valves rather than through the sample inlet port.  
The zero and cal operations are initiated directly and independently with dedicated keys (CALZ &  
CALS).  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =XXX.X  
Set the Display to show the  
STABIL test function.  
This function calculates the  
stability of the COx  
<TST TST> CAL CALZ CALS SETUP  
measurement  
SAMPLE  
STABIL=XXX.X PPM  
CO2 =XXX.X  
< TST TST > CAL CALZ CALS SETUP  
ACTION:  
Allow zero gas to enter the sample port at the  
rear of the instrument.  
Wait until STABIL  
falls below1.0 ppb.  
This may take several  
minutes.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
SETUP  
< TST TST > CAL CALZ CALS  
M-P CAL  
STABIL=XXX.X PPM  
CONC  
CO2 =XXX.X  
EXIT  
< TST TST > ZERO  
Press ENTR to changes the  
OFFSET & SLOPE values for the  
CO2 measurements.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
< TST TST > ENTR  
CONC  
EXIT  
Press EXIT to leave the calibration  
unchanged and return to the  
previous menu.  
ACTION:  
Allow span gas to enter the sample port at the  
rear of the instrument.  
The value of  
STABIL may jump  
significantly.  
Wait until it falls back  
below 1.0 ppb.  
The SPAN key now  
appears during the  
transition from zero to  
span.  
This may take several  
minutes.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
EXIT  
< TST TST >  
M-P CAL  
SPAN CONC  
You may see both keys.  
If either the ZERO or  
SPAN buttons fail to  
appear see Section 11  
for troubleshooting tips.  
Press ENTR to change the  
OFFSET & SLOPE values for the  
CO2 measurements.  
RANGE = 500.000 PPM CO2 =XXX.X  
EXIT  
Press EXIT to leave the calibration  
unchanged and return to the  
previous menu.  
< TST TST > ENTR SPAN CONC  
M-P CAL  
RANGE = 500.000 PPM CO2 =XXX.X  
CONC  
EXIT  
EXIT returns to the main  
SAMPLE display  
< TST TST > ENTR  
04584 Rev A1  
136  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Calibration Procedures  
7.5. Manual Calibration Checks with Zero/Span Valves  
Zero and span checks using the VARIOUS zero/span valve options available for the MGFC7000E  
are similar to that described in Section 7.3, except that the zero and calibration operations are  
initiated directly and independently with dedicated keys CALZ and CALS.  
To perform a manual calibration check of an analyzer with a valve option installed, use the  
following method.  
STEP ONE: Connect the sources of Zero Air and Span Gas as shown in figures 7-3 through 7-6.  
STEP TWO: Perform the zero/span check.  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
Scroll to the STABIL test  
< TST TST > CAL CALZ CALS  
SETUP  
function.  
SAMPLE  
STABIL=XXX.X PPM  
CO2 =X.XXX  
SETUP  
Wait until STABIL  
falls below 1.0  
ppb.  
< TST TST > CAL CALZ CALS  
ACTION:  
Record the  
CO2 readings  
presented in the  
upper right corner of  
the display.  
This may take  
several minutes.  
ZERO CAL M  
STABIL=XXX.X PPM CO2 =X.XXX  
CONC EXIT  
< TST TST > ZERO  
SAMPLE  
STABIL=XXX.X PPM CO2 =X.XXX  
ACTION:  
Record the  
CO2 readings  
presented in the  
upper right corner of  
the display.  
The value of STABIL  
may jump  
significantly. Wait  
until STABIL falls  
below 1.0 ppb. This  
may take several  
minutes.  
< TST TST > CAL CALZ CALS  
SETUP  
SPAN CAL M  
STABIL=XXX.X PPM  
CO2 =X.XXX  
EXIT  
EXIT returns to the main  
< TST TST > ZERO SPAN CONC  
SAMPLE display  
7.5.1. Zero/Span Calibration on Auto Range or Dual Ranges  
If the analyzer is being operated in dual range mode or auto range mode, then the high and low  
ranges must be independently calibrated.  
When the analyzer is in either dual or auto range modes the user must run a separate calibration  
procedure for each range. After pressing the CAL, CALZ or CALS keys the user is prompted for  
the range that is to be calibrated as seen in the CALZ example below:  
04584 Rev A1  
137  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Calibration Procedures  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =XXX.X  
SETUP  
Set the Display to show the  
STABIL test function.  
This function calculates the  
stability of the COx  
<TST TST> CAL CALZ CALS  
measurement  
SAMPLE  
STABIL=XXX.X PPM  
CO2 =XXX.X  
SETUP  
< TST TST > CAL CALZ CALS  
ACTION:  
Allow zero gas to enter the sample port at the  
rear of the instrument.  
Wait until STABIL  
falls below 1.0 ppb.  
This may take several  
minutes.  
M-P CAL  
STABIL=XXX.X PPM  
CO2 =XXX.X  
< TST TST > CAL CALZ CALS  
SETUP  
SAMPLE  
RANGE TO CAL: LOW  
ENTR  
LOW HIGH  
SETUP  
SETUP  
SAMPLE  
RANGE TO CAL: HIGH  
ENTR  
LOW HIGH  
ANALYZER ENTERS  
ZERO CAL MODE  
ZERO CAL M  
RANGE = 500.000 PPB SO2 =X.XXX  
< TST TST > ZERO SPAN CONC  
EXIT  
Continue Calibration as per  
Standard Procedure  
Once this selection is made, the calibration procedure continues as previously described in Section  
7.2. The other range may be calibrated by starting over from the main SAMPLE display.  
7.5.2. Use of Zero/Span Valves with Remote Contact Closure  
Contact closures for controlling calibration are located on the rear panel CONTROL IN connector.  
Instructions for setup and use of these contacts are found in Section 6.11.1.2. When the contacts  
are closed for at least 5 seconds, the instrument switches into zero or span mode. The remote  
calibration contact closures may be activated in any order. It is recommended that contact  
closures remain closed for at least 10 minutes to establish a reliable reading.  
The instrument will stay in the selected mode for as long as the contacts remain closed. If  
calibration is enabled, the MGFC7000E will re-calibrate when the contact is opened, then go into  
SAMPLE mode. If calibration is disabled, the instrument will return to SAMPLE mode, leaving the  
calibration unchanged.  
04584 Rev A1  
138  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Calibration Procedures  
7.6. Automatic Zero/Span Cal/Check (AutoCal)  
The AutoCal system allows unattended periodic operation of the ZERO/SPAN valve options by  
using the MGFC7000E’s internal time of day clock. AutoCal operates by executing SEQUENCES  
programmed by the user to initiate the various calibration modes of the analyzer and open and  
close valves appropriately. It is possible to program and run up to 3 separate sequences (SEQ1,  
SEQ2 and SEQ3). Each sequence can operate in one of 3 Modes, or be disabled.  
Table 7-1:  
AUTOCAL Modes  
MODE NAME  
DISABLED  
ZERO  
ACTION  
Disables the Sequence  
Causes the Sequence to perform a zero calibration/check  
ZERO-SPAN  
Causes the Sequence to perform a zero and span  
concentration calibration/check  
SPAN  
Causes the Sequence to perform a span concentration  
calibration/check  
For each mode there are seven parameters that control operational details of the SEQUENCE.  
They are:  
Table 7-2:  
AutoCal ATTRIBUTE Setup Parameters  
ATTRIBUTE NAME  
TIMER ENABLED  
STARTING DATE  
STARTING TIME  
DELTA DAYS  
ACTION  
Turns on the Sequence timer  
Sequence will operate after Starting Date  
Time of day sequence will run  
Number of days to skip between each Seq. execution  
Number of hours later each “Delta Days” Seq is to be run  
Number of minutes the sequence operates  
Enable to do a calibration – Disable to do a cal check only  
DELTA TIME  
DURATION  
CALIBRATE  
04584 Rev A1  
139  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Calibration Procedures  
The following example sets Sequence #2 to do a Zero-Span Calibration every other day starting  
at 1 am on September 4, 2001, lasting 15 minutes, without calibration. This will start ½ hour later  
each iteration.  
MODE AND  
ATTRIBUTE  
VALUE  
COMMENT  
Define Sequence #2  
Sequence  
2
ZERO-SPAN  
ON  
Mode  
Select Zero and Span Mode  
Enable the timer  
Timer Enable  
Starting Date  
Starting Time  
Delta Days  
Delta Time  
Duration  
Sept. 4, 2001  
01:00  
2
Start after Sept 4, 2001  
First Span starts at 1:00AM  
Do Sequence #2 every other day  
Do Sequence #2 ½ hr later each day  
Operate Span valve for 15 min  
Do not calibrate at end of Sequence  
00:30  
15.0  
Calibrate  
NO  
NOTE  
The programmed STARTING_TIME must be a minimum of 5 minutes later than the real  
time clock (See Section 6.6 for setting real time clock).  
NOTE  
Avoid setting two or more sequences at the same time of the day. Any new sequence  
which is initiated whether from a timer, the COM ports, or the contact closure inputs  
will override any sequence which is in progress.  
NOTE  
If at any time an illegal entry is selected (Example: Delta Days > 367) the ENTR key will  
disappear from the display.  
04584 Rev A1  
140  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Calibration Procedures  
To program the Sequence:  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
SETUP  
SETUP X.X STARTING TIME:14:15  
< TST TST > CAL CALZ CALS  
<SET SET> EDIT  
EXIT  
EXIT  
SAMPLE  
ENTER SETUP PASS : 818  
SETUP X.X  
DELTA DAYS: 1  
8
1
8
ENTR EXIT  
<SET SET> EDIT  
Toggle keys  
to set  
SETUP X.X  
PRIMARY SETUP MENU  
number of  
days  
between  
procedures  
(1-367)  
SETUP X.X DELTA DAYS: 1  
CFG ACAL DAS RNGE PASS CLK MORE  
EXIT  
0
0
2
ENTR EXIT  
SETUP X.X SEQ 1) DISABLED  
SETUP X.X DELTA DAYS:2  
NEXT MODE  
EXIT  
<SET SET> EDIT  
EXIT  
SETUP X.X SEQ 2) DISABLED  
PREV NEXT MODE  
EXIT  
SETUP X.X DELTA TIME00:00  
<SET SET> EDIT  
EXIT  
SETUP X.X MODE: DISABLED  
Toggle keys  
to set  
delay time for  
each iteration  
of the  
sequence:  
HH:MM  
NEXT  
ENTR EXIT  
ENTR EXIT  
ENTR EXIT  
EXIT  
SETUP X.X DELTA TIME: 00:00  
0
0
:3  
0
ENTR EXIT  
SETUP X.X MODE: ZERO  
(0 – 24:00)  
PREV NEXT  
SETUP X.X DELTA TIEM:00:30  
<SET SET> EDIT  
EXIT  
SETUP X.X MODE: ZERO–SPAN  
PREV NEXT  
SETUP X.X DURATION:15.0 MINUTES  
Toggle keys  
to set  
duration for  
each  
iteration of  
the  
sequence:  
Set in  
<SET SET> EDIT  
EXIT  
ENTR EXIT  
EXIT  
SETUP X.X SEQ 2) ZERO–SPAN, 1:00:00  
PREV NEXT MODE SET  
SETUP X.X DURATION 15.0MINUTES  
3
0
.0  
Default  
value is  
ON  
SETUP X.X TIMER ENABLE: ON  
Decimal  
minutes  
from  
SET> EDIT  
EXIT  
0.1 – 60.0  
SETUP X.X DURATION:30.0 MINUTES  
<SET SET> EDIT  
SETUP X.X STARTING DATE: 01–JAN–02  
<SET SET> EDIT  
EXIT  
Toggle keys  
to set  
day, month &  
year:  
SETUP X.X  
CALIBRATE: OFF  
SETUP X.X STARTING DATE: 01–JAN–02  
<SET SET> EDIT  
EXIT  
ENTR EXIT  
EXIT  
0
4
SEP  
0
3
ENTR EXIT  
Format :  
DD-MON-YY  
Toggle key  
between  
Off and  
ON  
SETUP X.X  
ON  
CALIBRATE: OFF  
SETUP X.X STARTING DATE: 04–SEP–03  
<SET SET> EDIT  
EXIT  
SETUP X.X  
CALIBRATE: ON  
SETUP X.X STARTING DATE: 04–SEP–03  
<SET SET> EDIT  
<SET SET> EDIT  
EXIT  
Toggle keys to  
set time:  
SETUP X.X SEQ 2) ZERO–SPAN, 2:00:30  
EXIT returns  
to the SETUP  
Menu  
SETUP X.X STARTING TIME:00:00  
Format : HH:MM  
PREV NEXT MODE SET  
EXIT  
<SET SET> EDIT  
EXIT  
This is a 24 hr  
clock .  
PM hours are  
13 – 24.  
Sequence  
Delta Time  
MODE  
Delta Days  
SETUP X.X STARTING TIME:00:00  
Example  
2:15 PM = 14:15  
1
4
: 1  
5
ENTR EXIT/  
04584 Rev A1  
141  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Calibration Procedures  
7.6.1. AutoCal with Auto or Dual Reporting Ranges Modes  
Selected  
SETUP C.4  
<SET  
RANGE TO CAL: LOW  
EDIT  
EXIT  
SETUP C.4  
RANGE TO CAL: LOW  
RANGE TO CAL: HIGH  
LOW HIGH  
ENTR SETUP  
SETUP C.4  
<SET  
EDIT  
EXIT  
EXIT  
SETUP C.4 SEQ 2) ZERO–SPAN, 2:00:30  
EXIT returns to the  
PRIMARY SETUP  
Menu  
PREV NEXT MODE SET  
NOTE  
In order to automatically calibrate both the HIGH and LOW ranges, you must set up a  
separate sequence for each.  
7.7. Calibration Quality  
After completing one of the calibration procedures described above, it is important to evaluate the  
analyzer’s calibration SLOPE and OFFSET parameters. These values describe the linear response  
curve of the analyzer. The values for these terms, both individually and relative to each other,  
indicate the quality of the calibration. To perform this quality evaluation, you will need to record  
the values of both test functions (Section 6.2.1 or Appendix A-3), all of which are automatically  
stored in the iDAS channel CALDAT for data analysis, documentation and archival.  
Make sure that these parameters are within the limits listed in Table 7-3.  
Table 7-3:  
Calibration Data Quality Evaluation  
FUNCTION  
SLOPE  
MINIMUM VALUE  
OPTIMUM VALUE  
1.000  
MAXIMUM VALUE  
0.700  
1.300  
0.500  
OFFS  
-0.500  
0.000  
These values should not be significantly different from the values recorded on the Teledyne  
Instruments Final Test and Validation Data sheet that was shipped with your instrument. If  
they are, refer to the troubleshooting Chapter 11.  
User Notes  
04584 Rev A1  
142  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
EPA Protocol Calibration  
8. EPA PROTOCOL CALIBRATION  
At the writing of this manual there is no EPA requirements for the monitoring of CO2 or published  
calibration protocols.  
User Notes  
04584 Rev A1  
143  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
MAINTENANCE SCHEDULE & PROCEDURES  
9. MAINTENANCE SCHEDULE & PROCEDURES  
Predictive diagnostic functions including failure warnings and alarms built into the analyzer’s  
firmware allow the user to determine when repairs are necessary without performing painstaking  
preventative maintenance procedures. There are, however, a minimal number of simple  
procedures that when performed regularly will ensure that the analyzer continues to operate  
accurately and reliably over its the lifetime. Repairs and troubleshooting are covered in Chapter 11  
of this manual.  
9.1. Maintenance Schedule  
Table 9-1 shows a typical maintenance schedule for the analyzer. Please note that in certain  
environments (i.e. dusty, very high ambient pollutant levels) some maintenance procedures may  
need to be performed more often than shown.  
NOTE  
A Span and Zero Calibration Check (see CAL CHECK REQ’D Column of Table 9-1) must be  
performed following certain of the maintenance procedure listed below.  
See Sections 7.3, 7.5 and 7.6 for instructions on performing checks.  
CAUTION  
Risk of electrical shock. Disconnect power before performing any of the  
following operations that require entry into the interior of the analyzer.  
NOTE  
The operations outlined in this chapter are to be performed by qualified  
maintenance personnel only.  
04584 Rev A1  
145  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
MAINTENANCE SCHEDULE & PROCEDURES  
Table 9-1:  
Cal  
Check  
Req’d.  
GFC7000E Maintenance Schedule  
Date Performed  
Manual  
Section  
Item  
Action  
Freq  
Particulate  
Filter  
Weekly or as  
needed  
Replace  
Yes  
No  
Weekly or  
after any  
Maintenance  
or Repair  
Verify Test Record and  
Functions  
analyze  
Pump  
Diaphragm  
Every 2  
years  
Replace  
Yes  
No  
Perform  
Flow Check  
Every 6  
Months  
Check Flow  
Annually or  
after any  
Maintenance  
Perform  
Leak Check  
Verify Leak  
Tight  
Yes  
or Repair  
Pneumatic  
lines  
Examine  
and clean  
Yes if  
cleaned  
As needed  
As needed  
Only if  
cover  
Cleaning  
Clean  
remv’d  
04584 Rev A1  
146  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
MAINTENANCE SCHEDULE & PROCEDURES  
Table 9-2:  
GFC7000E Test Function Record  
Date Recorded  
Operating  
Function  
Mode*  
ZERO CAL  
STABILITY  
ZERO CAL  
CO2 MEAS  
Zero CAL  
MR RATIO  
SPAN CAL  
SAMPLE  
PRES  
SAMPLE After  
Warn-up  
PHT DRIVE  
SLOPE  
SPAN CAL  
ZERO CAL  
OFFSET  
04584 Rev A1  
147  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
MAINTENANCE SCHEDULE & PROCEDURES  
9.2. Predicting Failures Using the Test Functions  
The Test Functions can be used to predict failures by looking at how their values change over  
time. Initially it may be useful to compare the state of these Test Functions to the values recorded  
on the printed record of the final calibration performed on your instrument at the factory, p/n  
04307. Table 9-3 can be used as a basis for taking action as these values change with time. The  
internal data acquisition system (iDAS) is a convenient way to record and track these changes.  
Use APIcom to download and review this data from a remote location.  
Table 9-3:  
Predictive uses for Test Functions  
Function  
Stability  
Condition  
Behavior  
Interpretation  
ƒ Pneumatic Leaks – instrument & sample  
system  
ƒ Detector deteriorating  
ƒ Source Aging  
ƒ Detector deteriorating  
ƒ Optics getting dirty or contaminated  
ƒ Source Aging  
ƒ Detector deteriorating  
ƒ Contaminated zero gas (H2O)  
ƒ Source Aging  
ƒ Detector deteriorating  
ƒ GFC Wheel Leaking  
ƒ Pneumatic Leaks  
Zero Cal  
Zero Cal  
Increasing  
CO2 MEAS  
Decreasing  
Increasing  
Zero Cal  
Decreasing  
Increasing  
MR Ratio  
ƒ Contaminated zero gas (CO2)  
ƒ Source Aging  
ƒ Pneumatic Leaks – instrument & sample  
system  
Span Cal  
Sample  
ƒ Calibration system deteriorating  
ƒ Source Aging  
ƒ GFC Wheel Leaking  
Decreasing  
ƒ Calibration system deteriorating  
ƒ Pneumatic Leak between sample inlet  
and Sample Cell  
ƒ Change in sampling manifold  
ƒ Dirty particulate filter  
ƒ Pneumatic obstruction between sample  
inlet and Sample Cell  
ƒ Obstruction in sampling manifold  
ƒ Mechanical Connection between IR-  
Detector and Sample Cell deteriorating  
ƒ IR-Photodetector deteriorating  
ƒ See MR Ratio - Zero Cal Decreasing  
above  
ƒ See MR Ratio - Zero Cal Increasing above  
ƒ See MR Ratio - Span Cal Decreasing  
above  
Increasing > 1”  
Pres  
Decreasing > 1”  
Increasing  
Any, but with  
Bench Temp  
at 48°C  
PHT Drive  
Offset  
Increasing  
Decreasing  
Increasing  
Zero Cal  
Slope  
Span Cal  
ƒ See MR Ratio – Span Cal Increasing  
above  
Decreasing  
9.3. Maintenance Procedures  
The following procedures are to be performed periodically as part of the standard maintenance of  
the Model GFC7000E.  
04584 Rev A1  
148  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
MAINTENANCE SCHEDULE & PROCEDURES  
9.3.1. Replacing the Sample Particulate Filter  
The particulate filter should be inspected often for signs of plugging or contamination. We  
recommend that when you change the filter, handle it and the wetted surfaces of the filter  
housing as little as possible. Do not touch any part of the housing, filter element, PTFE retaining  
ring, glass cover and the o-ring.  
To change the filter:  
1. Turn OFF the analyzer to prevent drawing debris into the instrument.  
2. Open the MGFC7000E’s hinged front panel and unscrew the knurled retaining ring on the filter  
assembly.  
Figure 9-1:  
Sample Particulate Filter Assembly  
3. Carefully remove the retaining ring, PTFE o-ring, glass filter cover and filter element.  
4. Replace the filter, being careful that the element is fully seated and centered in the bottom of  
the holder.  
5. Re-install the PTFE o-ring with the notches up, the glass cover, then screw on the retaining  
ring and hand tighten. Inspect the seal between the edge of filter and the o-ring to assure a  
proper seal.  
6. Re-start the Analyzer.  
04584 Rev A1  
149  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
MAINTENANCE SCHEDULE & PROCEDURES  
9.3.2. Rebuilding the Sample Pump  
The diaphragm in the sample pump periodically wears out and must be replaced. A sample  
rebuild kit is available – see Appendix B of this manual for the part number of the pump rebuild  
kit. Instructions and diagrams are included with the kit.  
Always perform a Flow and Leak Check after rebuilding the Sample Pump.  
9.3.3. Performing Leak Checks  
Leaks are the most common cause of analyzer malfunction; Section 9.3.3.1 presents a simple leak  
check procedure. Section 9.3.3.2 details a more thorough procedure.  
9.3.3.1. Vacuum Leak Check and Pump Check  
This method is easy and fast. It detects, but does not locate most leaks, it also verifies that the  
sample pump is in good condition.  
1. Turn the analyzer ON, and allow enough time for flows to stabilize.  
2. Cap the sample inlet port.  
3. After several minutes, when the pressures have stabilized, note the following. In the TEST  
menu, note the SAMPLE PRESSURE reading.  
4. If the reading is < 10 in-Hg, the pump is in good condition and there are no large leaks.  
5. If both readings are equal within 10%, the instrument is free of large leaks.  
9.3.3.2. Pressure Leak Check  
If you can’t locate the leak by the above procedure, use the following procedure. Obtain a leak  
checker similar to the Teledyne Instruments part number 01960, which contains a small pump,  
shut-off valve, and pressure gauge. Alternatively, a convenient source of low-pressure gas is a  
tank of span gas, with the two-stage regulator adjusted to less than 15 psi with a shutoff valve  
and pressure gauge.  
CAUTION  
Do not use bubble solution with vacuum applied to the analyzer. The  
solution may contaminate the instrument. Do not exceed 15 PSIG  
pressure.  
1. Turn OFF power to the instrument.  
2. Install a leak checker or tank of gas as described above on the sample inlet at the rear panel.  
3. Remove the instrument cover and locate the inlet side of the sample pump. Remove the flow  
assembly from the pump and plug it with the appropriate gas-tight fitting.  
04584 Rev A1  
150  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
MAINTENANCE SCHEDULE & PROCEDURES  
4. Pressurize the instrument with the leak checker, allowing enough time to fully pressurize the  
instrument through the critical flow orifice. Check each fitting with soap bubble solution,  
looking for bubbles. Once the fittings have been wetted with soap solution, do not re-apply  
vacuum, as it will suck soap solution into the instrument and contaminate it. Do not exceed  
15 psi pressure.  
5. If the instrument has one of the zero and span valve options, the normally closed ports on  
each valve should also be separately checked. Connect the leak checker to the normally closed  
ports and check with soap bubble solution.  
6. Once the leak has been located and repaired, the leak-down rate should be < 1 in-Hg (0.4 psi)  
in 5 minutes after the pressure is shut off.  
9.3.4. Performing a Sample Flow Check  
CAUTION  
Always use a separate calibrated flow meter capable of measuring flows  
in the 0 – 1000 cc/min range to measure the gas flow rate though the  
analyzer.  
DO NOT use the built in flow measurement viewable from the Front Panel  
of the instrument. This measurement is only for detecting major flow  
interruptions such as clogged or plugged gas lines.  
See Figure 3-2 for sample port location.  
1. Turn off power.  
2. Attach the Flow Meter to the sample inlet port on the rear panel. Ensure that the inlet to the  
Flow Meter is at atmospheric f.  
3. Turn on instrument power.  
4. Sample flow should be 800 cc/min ± 10%.  
5. Once an accurate measurement has been recorded by the method described above, adjust the  
analyzer’s internal flow sensors (see Section 6.9.8)  
Low flows indicate blockage somewhere in the pneumatic pathway. High flows indicate leaks  
downstream of the Flow Control Assembly.  
9.3.5. Cleaning the Optical Bench  
The MGFC7000E sensor assembly and optical bench is complex and delicate. Disassembly and  
cleaning is not recommended. Please check with the factory before disassembling the optical  
bench.  
9.3.6. Cleaning Exterior Surfaces of the MGFC7000E  
If necessary, the exterior surfaces of the MGFC7000E can be cleaned with a clean damp cloth. Do  
not submerge any part of the instrument in water or cleaning solution.  
04584 Rev A1  
151  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
MAINTENANCE SCHEDULE & PROCEDURES  
User Notes  
04584 Rev A1  
152  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
10. THEORY OF OPERATION  
The Model GFC7000E Gas Filter Correlation Carbon Dioxide Analyzer is a microprocessor-  
controlled analyzer that determines the concentration of carbon dioxide (CO2) in a sample gas  
drawn through the instrument. It requires that sample and calibration gasses be supplied at  
ambient atmospheric pressure in order to establish a stable gas flow through the sample chamber  
where the gases ability to absorb infrared radiation is measured.  
Calibration of the instrument is performed in software and does not require physical adjustments  
to the instrument. During calibration the microprocessor measures the current state of the IR  
Sensor output and various other physical parameters of the instrument and stores them in  
memory.  
The microprocessor uses these calibration values, the ir absorption measurements made on the  
sample gas along with data regarding the current temperature and pressure of the gas to  
calculate a final co2 concentration.  
This concentration value and the original information from which it was calculated are stored in  
one of the unit’s internal data acquisition system (iDAS - see Sections 6.12) as well as reported to  
the user via a vacuum florescent display or a variety of digital and analog signal outputs.  
10.1. Measurement Method  
10.1.1. Beer’s Law  
The basic principle by which the analyzer works is called Beer’s Law. It defines the how light of a  
specific wavelength is absorbed by a particular gas molecule over a certain distance. The  
mathematical relationship between these three parameters is:  
-αLc  
I = Io e  
Where:  
Io  
I
is the intensity of the light if there was no absorption.  
is the intensity with absorption.  
L
is the absorption path, or the distance the light travels as it is being absorbed.  
C
is the concentration of the absorbing gas. In the case of the Model GFC7000E,  
carbon dioxide (CO2).  
α
is the absorption coefficient that tells how well CO2 absorbs light at the specific  
wavelength of interest.  
04584 Rev A1  
153  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
10.1.2. Measurement Fundamentals  
In the most basic terms, the Model GFC7000E uses a high energy heated element to generate a  
beam of broad-band IR light with a known intensity (measured during Instrument calibration.  
This beam is directed through multi-pass cell filled with sample gas. The sample cell uses mirrors  
at each end to reflect the IR beam back and forth through the sample gas to generate a 2.5 meter  
absorption path (see Figure 10–1). This length was chosen to give the analyzer maximum  
sensitivity to fluctuations in CO2 density.  
Band-Pass  
Filter  
Sample Chamber  
IR  
Source  
Photo-Detector  
IR Beam  
Figure 10-1: Measurement Fundamentals  
Upon exiting the sample cell, the beam shines through a band-pass filter that allows only light at  
a wavelength of 4.3 µm to pass. Finally, the beam strikes a solid-state photo-detector that  
converts the light signal into a modulated voltage signal representing the attenuated intensity of  
the beam.  
10.1.3. Gas Filter Correlation  
Unfortunately, water vapor absorbs light at 4.3 µm too. To overcome the interfering effects of  
water vapor the Model GFC7000E adds another component to the IR light path called a gas filter  
correlation (GFC) wheel (see Figure 10-2).  
04584 Rev A1  
154  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
Measurement  
Cell  
(Pure N2)  
Reference Cell  
(N2 with CO2)  
Figure 10-2: GFC Wheel  
A GFC wheel is a metallic wheel into which two chambers are carved. The chambers are sealed on  
both sides with material transparent to 4.3 µm IR radiation creating two airtight cavities. Each  
cavity is filled with specially composed gases. One cell is filled with pure N2 (the measure cell).  
The other is filled with a combination of N2 and a high concentration of CO2 (the reference cell).  
IR unaffected by N2 in Measurement Cell  
Δ H  
IR IS affected by CO in Reference Cell  
M
IR  
Source  
Photo-Detector  
R
GFC Wheel  
Figure 10-3: Measurement Fundamentals with GFC Wheel  
As the GFC wheel spins, the IR light alternately passes through the two cavities. When the beam  
is exposed to the reference cell, the CO2 in the gas filter wheel strips the beam of most of the IR  
at 4.3μm. When the light beam is exposed to the measurement cell, the N2 in the filter wheel  
04584 Rev A1  
155  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
does not absorb IR light. This results in a fluctuation in the intensity of the IR light striking the  
photo-detector (See Figure 10-3) that results in the output of the detector resembling a square  
wave.  
The Model GFC7000E determines the amount of CO2 in the sample chamber by computing the  
ratio between the peak of the measurement pulse (CO2 MEAS) and the peak of the reference  
pulse (CO2 REF).  
IR unaffected by N2 in Measurement  
Cell of the GDC Wheel and no  
additional CO2 in the Sample Chamber  
CO2 MEAS  
CO2 REF  
IR affected by CO2 in Reference  
Cell with no interfering gas in the  
Sample Chamber  
IR shinning through Measurement Cell  
of the GDC Wheel is reduced by  
additional CO2 in the Sample Chamber  
M/R  
is reduced  
IR shining through Reference Cell  
is also reduced by additional CO2  
in the Sample Chamber, but to a  
lesser extent  
Figure 10-4: Affect of CO2 in the Sample on CO2 MEAS & CO2 REF  
If no gases exist in the sample chamber that absorb light at 4.3μm, the high concentration of CO2  
in the gas mixture of the reference cell will attenuate the intensity of the IR beam by 60% giving  
a M/R ratio of approximately 2.4:1.  
Adding CO2 to the sample chamber causes the peaks corresponding to both cells to be attenuated  
by a further percentage. Since the intensity of the light passing through the measurement cell is  
greater, the effect of this additional attenuation is greater. This causes CO2 MEAS to be more  
sensitive to the presence of CO2 in the sample chamber than CO2 REF and the ratio between  
them (M/R) to move closer to 1:1 as the concentration of CO2 in the sample chamber increases.  
Once the Model GFC7000E has computed this ratio, a look-up table is used, with interpolation, to  
linearize the response of the instrument. This linearized concentration value is combined with  
04584 Rev A1  
156  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
calibration SLOPE and OFFSET values to produce the CO2 concentration which is then normalized  
for changes in sample pressure.  
M/R  
is Shifted  
IR shining through both cells is  
effected equally by interfering gas  
in the Sample Chamber  
Figure 10-5: Effects of Interfering Gas on CO2 MEAS & CO2 REF  
If an interfering gas, such as H2O vapor is introduced into the sample chamber, the spectrum of  
the IR beam is changed in a way that is identical for both the reference and the measurement  
cells, but without changing the ratio between the peak heights of CO2 MEAS and CO2 REF. In  
effect, the difference between the peak heights remains the same.  
Thus, the difference in the peak heights and the resulting M/R ratio is only due to CO2 and not to  
interfering gases. In this way, Gas filter correlation rejects the effects of interfering gases and so  
that the analyzer responds only to the presence of CO2.  
To improve the signal-to-noise performance of the IR photo-detector, the GFC wheel also  
incorporates an optical mask that chops the IR beam into alternating pulses of light and dark at  
six times the frequency of the measure/reference signal. This limits the detection bandwidth  
helping to reject interfering signals from outside this bandwidth improving the signal to noise  
ration.  
The IR Signal as the Photo-Detector  
sees it after being chopped by the GFC  
Wheel Screen  
CO MEAS  
CO REF  
Figure 10-6: Chopped IR Signal  
04584 Rev A1  
157  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
10.1.4. Ambient CO2 Interference Rejection  
CO2 absorbs IR light very well. So well that even the narrow volume of ambient air between the  
IR source and the sample chamber is enough to alter the analyzer’s measured concentration of  
CO2. Also, ambient air, which averages around 350 ppm to 400 ppm, will vary significantly over  
the course of the day. The ambient CO2 concentration can rise as high as 1 000 ppm during the  
time of the day when people are present. It can fluctuate ± 300 ppm as the photosynthesis of  
plant life in the nearby area increases during the day and decreases at night.  
The basic design of the GFC7000E rejects most of this interference at a 100:1 ratio, however this  
still can allow small fluctuations in CO2 concentration during the course of the day. To completely  
remove all effects of ambient CO2 from the analyzer’s measurement of CO2, dried air, scrubbed of  
all CO2 is pumped into the GFC wheel housing to purge all ambient CO2 (see Figure 10-7)  
10.2. Pneumatic Operation  
Caution  
It is important that the sample airflow system is both leak tight and  
not pressurized over ambient pressure.  
Regular leak checks should be performed on the analyzer as described  
in the maintenance schedule, Table 9-1.  
Procedures for correctly performing leak checks can be found in  
Section 9.3.3.  
An internal pump evacuates the sample chamber creating a small vacuum that draws sample gas  
into the analyzer. Normally the analyzer is operated with its inlet near ambient pressure either  
because the sample is directly drawn at the inlet or a small vent is installed at the inlet. There are  
several advantages to this “pull through” configuration.  
By placing the pump down stream from the sample chamber several problems are avoided.  
First the pumping process heats and compresses the sample air complicating the  
measurement process.  
Additionally, certain physical parts of the pump itself are made of materials that might  
chemically react with the sample gas.  
Finally, in certain applications where the concentration of the target gas might be high  
enough to be hazardous, maintaining a negative gas pressure relative to ambient means  
that should a minor leak occur, no sample gas will be pumped into the atmosphere  
surrounding analyzer.  
04584 Rev A1  
158  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
10.2.1. Sample Gas Flow  
INSTRUMENT CHASSIS  
SAMPLE GAS  
INLET  
PUMP  
GFC Wheel  
Motor  
Purge Gas  
EXHAUST GAS  
OUTLET  
Flow Rate  
Control  
Orifice  
GFC Motor  
Heat Sync  
Purge Gas  
Pressure  
Control Assy  
GFC Wheel  
Housing  
PURGE GAS  
INLET  
FLOW / PRESSURE  
SENSOR PCA  
SAMPLE  
PRESSURE  
SENSOR  
SAMPLE CHAMBER  
FLOW  
SENSOR  
VENT SPAN  
OUTLET  
PRESSURE  
SPAN INLET  
PARTICULATE  
FILTER  
IZS INLET  
Figure 10-7: Internal Pneumatic Flow – Basic Configuration  
10.2.1.1. Critical Flow Orifice  
The most important component of this flow control assembly is the critical flow orifice.  
Critical flow orifices are a remarkably simple way to regulate stable gas flow rates. They operate  
without moving parts by taking advantage of the laws of fluid dynamics. By restricting the flow of  
gas though the orifice, a pressure differential is created. This pressure differential combined with  
the action of the analyzer’s external pump draws the gas through the orifice.  
As the pressure on the downstream side of the orifice (the pump side) continues to drop, the  
speed that the gas flows though the orifice continues to rise. Once the ratio of upstream pressure  
to downstream pressure is greater than 2:1, the velocity of the gas through the orifice reaches  
the speed of sound. As long as that ratio stays at least 2:1 the gas flow rate is unaffected by any  
fluctuations, surges, or changes in downstream pressure because such variations only travel at  
the speed of sound themselves and are therefore cancelled out by the sonic shockwave at the  
downstream exit of the critical flow orifice.  
04584 Rev A1  
159  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
CRITICAL  
FLOW  
ORIFICE  
AREA OF  
LOW  
AREA OF  
HIGH  
PRESSURE  
PRESSURE  
Sonic  
Shockwave  
O-RINGS  
SPRING  
FILTER  
Figure 10-8: Flow Control Assembly & Critical Flow Orifice  
The actual flow rate of gas through the orifice (volume of gas per unit of time), depends on the  
size and shape of the aperture in the orifice. The larger the hole, the more gas molecules, moving  
at the speed of sound, pass through the orifice. Also, because the flow rate of gas through the  
orifice is only related to the minimum 2:1 pressure differential and not absolute pressure:  
Pressure wave created by the pump’s action are filtered out.  
The flow rate of gas through the sample chamber will be the same regardless of whether  
the analyzer is at the bottom of Death Valley or on top of Pikes Peak.  
The flow rate of the gas is also unaffected by degradations in pump efficiency due to age.  
The critical flow orifice used in the Model GFC7000E is designed to provide a flow rate of 800  
cm3/min.  
10.2.1.2. Sample Pressure Sensor  
An absolute value pressure transducer plumbed to the outlet of the sample chamber is used to  
measure sample pressure. The output of the sensor is used to compensate the concentration  
measurement for changes in air pressure. This sensor is mounted to a printed circuit board with  
the sample flow sensor on the sample chamber; see following section and Figure 3-11.  
10.2.1.3. Sample Flow Sensor  
A thermal-mass flow sensor is used to measure the sample flow through the analyzer. The sensor  
is calibrated at the factory with ambient air or N2, but can calibrated to operate with samples  
consisting of other gases such as CO2, see Section 9.3.4. This sensor is mounted to a printed  
circuit board with the Sample Pressure sensor on the sample chamber; see previous section and  
Figure 3-11.  
04584 Rev A1  
160  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
Particulate Filter  
The Model GFC7000E Analyzer comes equipped with a 47 mm diameter, Teflon, particulate filter  
with a 5 micron pore size. The filter is accessible through the front panel, which folds down to  
allow access, and should be changed according to the suggested maintenance schedule described  
in Table 9-1.  
10.2.1.4. Valve Options  
A variety of optional valve sets can be purchased for the analyzer which allow the user to more  
easily supply and manipulate various calibration gases, such as zero air and span gas during  
various calibration procedures. For more information of these options see Section 5.4.  
10.2.2. Purge Gas Pressure and Flow Control  
In order to ensure that all of the ambient CO2 is purged from the GFC Wheel housing a adequate  
supply of dried air, scrubbed of CO2 must be supplied to the PURGE AIR inlet at the back of the  
instrument. The source of purge air must be capable of maintaining a pressure of 20-25 psig at a  
flow rate of at least 0.5 liters/min. Purge source air pressure should not exceed 35 pisg.  
In order to maintain the proper pressure differential between the inside of the GFC wheel housing  
and ambient air, the M360 design includes a manually settable pressure regulator that maintains  
the pressure of the purge air feed at 7.5 psig and a flow control orifice that ensures a 0.5 liter/min  
flow though the GFC wheel housing.  
10.3. Electronic Operation  
10.3.1. Overview  
Figure 10-9 shows a block diagram of the major electronic components of the Model GFC7000E.  
At its heart the analyzer is a microcomputer (CPU) that controls various internal processes,  
interprets data, makes calculations, and reports results using specialized firmware developed by  
Teledyne Instruments. It communicates with the user as well as receives data from and issues  
commands to a variety of peripheral devices via a separate printed circuit assembly called the  
Mother Board.  
The mother board collects data, performs signal conditioning duties and routs incoming and  
outgoing signals between the cpu and the analyzer’s other major components.  
Data is generated by a gas-filter-correlation optical bench which outputs an analog signal  
corresponding to the concentration of CO2 in the sample gas. This analog signal is transformed  
into two, pre-amplified, DC voltages (CO2 MEAS and CO2 REF) by a synchronous demodulator  
printed circuit assembly. CO2 MEAS and CO2 REF are converted into digital data by a unipolar,  
analog-to-digital converter, located on the mother board.  
A variety of sensors report the physical and operational status of the analyzer’s major  
components, again through the signal processing capabilities of the mother board. These status  
reports are used as data for the CO2 concentration calculation and as trigger events for certain  
control commands issued by the CPU. They are stored in memory by the CPU and in most cases  
can be viewed but the user via the front panel display.  
The CPU communicates with the user and the outside world in a variety of manners:  
04584 Rev A1  
161  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
Through the analyzer’s keyboard and vacuum florescent display over a clocked, digital,  
serial I/O bus (using a protocol called I2C);  
RS 232 & RS485 Serial I/O channels;  
Via an optional Ethernet communications card:  
Various DCV and DCA analog outputs, and  
Several sets of Digital I/O channels.  
Finally, the CPU issues commands via a series of relays and switches (also over the I2C bus)  
located on a separate printed circuit assembly to control the function of key electromechanical  
devices such as heaters, motors and valves.  
Back Panel  
Connectors  
Analog Outputs  
A1  
A2  
A3  
Optional  
4-20 mA  
Control Inputs:  
1 – 8  
COM-A COM-B  
Optional  
Ethernet  
Interface  
Status Outputs:  
1 – 8  
Analog  
Outputs  
(D/A)  
External  
Digital I/O)  
PC 104  
CPU Card  
A/D  
Converter  
(V/F)  
RS–232  
or RS-485  
Power-Up  
Circuit  
Disk On  
Chip  
RS – 232  
MOTHER  
BOARD  
Flash Chip  
Box  
Temp  
PC 104 Bus  
Zero/Span  
Valve  
Options  
PUMP  
Thermistor  
Interface  
I2C  
Bus  
Internal  
Digital I/O  
Sensor Inputs  
SAMPLE  
TEMP  
C
O2  
C
O2  
RELAY  
BOARD  
Sample  
Flow &  
Pressure  
Keyboard &  
Display  
CPU Status  
Sensors  
M
E
A
R
E
F
LED  
BENCH  
TEMP  
S
Sensor  
Status  
&
TEC Control  
IR  
Source  
Control  
PHT  
WHEEL  
TEMP  
Photo-  
detector  
SYNC  
DEMOD  
Drive  
Detector  
Output  
GFC  
Motor  
GFC  
Wheel  
Optical  
Bench  
Schmidt  
Trigger  
Wheel  
Heater  
Segment Sensor  
Bench Heater  
M / R Sensor  
Figure 10-9: GFC7000E Electronic Block Diagram  
04584 Rev A1  
162  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
10.3.2. CPU  
The Model GFC7000E’s CPU is a, low power (5 VDC, 0.8A max), high performance, 386-based  
microcomputer running MS-DOS. Its operation and assembly conform to the PC/104 Specification  
version 2.3 for embedded PC and PC/AT applications. It has 2 MB of DRAM on board and operates  
at 40MHz over an internal 32-bit data and address bus. Chip to chip data handling is performed  
by two 4-channel DMA devices over data busses of either 8-bit or 16-bit configuration. The CPU  
supports both RS-232 and RS-485 serial I/O.  
The CPU includes two types of non-volatile data storage.  
Disk On Chip  
While technically an EEPROM, the Disk –on-Chip (DOC), this device appears to the CPU as,  
behaves as, and performs the same function in the system as an 8MB disk drive. It is used to  
store the operating system for the computer, the Teledyne Instruments Firmware, and most of  
the operational data generated by the analyzer’s internal data acquisition system (iDAS - see  
Section 6.12).  
Flash Chip  
Another, smaller EEPROM used to store critical calibration and configuration data. Segregating  
this data on a separate, less heavily accessed chip significantly decreases the chance of this key  
data being corrupted.  
10.3.3. Optical Bench & GFC Wheel  
Electronically, the Model GFC7000E’s optical bench, GFC wheel and associated components do  
more than simply measure the amount of CO2 present in the sample chamber. A variety of other  
critical functions are performed here as well.  
10.3.3.1. Sample Gas and GFC Temperature Control  
Because the temperature of a gas affects its density and therefore the amount of light absorbed  
by that gas it is important to reduce the effect of fluctuations in ambient temperature on the  
Model GFC7000E’s measurement of CO2. To accomplish this both the temperature of the sample  
chamber and the GFC Wheel are maintained at constant temperatures above their normal  
operating ranges.  
Bench Temperature : To minimize the effects of ambient temperature variations on the sample  
measurement, the sample chamber is heated to 48°C (8 degrees above the maximum suggested  
ambient operating temperature for the analyzer). A strip heater attached to the underside of the  
chamber housing is the heat source. The temperature of the sample chamber is sensed by a  
thermistor, also attached to the sample chamber housing.  
Wheel Temperature: To minimize the effects of temperature variations caused by the near  
proximity of the IR Source to the GFC wheel on the gases contained in the wheel, it is also raised  
to a high temperature level. Because the IR Source itself is very hot, the set point for this heat  
circuit is 68°C. A cartridge heater is implanted into the heat sync on the motor is the heat source.  
The temperature of the wheel/motor assembly is sensed by a thermistor also inserted into the  
heat sync.  
Both heaters operate off of the AC line voltage supplied to the instrument.  
04584 Rev A1  
163  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
10.3.3.2. IR Source  
The light used to detect CO2 in the sample chamber is generated by a element heated to  
approximately 1100oC producing infrared radiation across a broad band. This radiation is optically  
filtered after it has passed through the GFC Wheel and the sample chamber and just before it  
reaches the photo-detector to eliminate all black body radiation and other extraneous IR emitted  
by the various components of those components.  
10.3.3.3. GFC Wheel  
A synchronous AC motor turns the GFC wheel motor. For analyzers operating on 60Hz line power  
this motor turns at 1800 rpm. For those operating on 50Hz line power the spin rate is 1500 rpm.  
The actual spin rate is unimportant within a large rate since a phase lock loop circuit is used to  
generate timing pulses for signal processing.  
In order to accurately interpret the fluctuations of the IR beam after it has passed through the  
sample gas, the GFC wheel several other timing signals are produced by other photo  
emitters/detectors. These devices consist of a combination LED and detector mounted so that the  
light emitted by the LED shines through the same mask on the GFC wheel that chops the IR  
beam.  
KEY:  
Detection Beam shining  
through MEASUREMENT  
side of GFC Wheel  
Detection Beam shining  
through REFERENCE  
side of GFC Wheel  
IR Detection Ring  
Segment Sensor Ring  
M/R Sensor Ring  
Figure 10-10: GFC Light Mask  
M/R Sensor  
The emitter/detector assembly that produces this signal shines through a portion of the mask that  
allows light to pass for half of a full revolution of the wheel. The resulting light signal tells the  
analyzer whether the IR beam is shining through the measurement or the reference side of the  
GFC wheel.  
Segment Sensor  
This emitter/detector shines through a portion of the mask that is divided into the same number  
of segments as the portion of the mask through which the IR beam passes. It is used by the  
04584 Rev A1  
164  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
synchronous / demodulation circuitry of the analyzer to latch onto the most stable part of each  
measurement and reference IR pulse.  
Measurement  
Pulses  
Reference  
Pulses  
IR Beam  
Pulses  
Segment Sensor  
Pulses  
MR Sensor  
Pulses  
Figure 10-11: Segment Sensor and M/R Sensor Output  
Schmidt Triggers  
To ensure that the waveforms produced by the Segment Sensor and the M/R Sensor are properly  
shaped and clean, these signals are passed through a set of Schmidt Triggers circuits.  
10.3.3.4. IR Photo-Detector  
The IR beam is converted into an electrical signal by a cooled solid-state photo-conductive  
detector The detector is composed of a narrow-band optical filter, a piece of lead-salt crystal  
whose electrical resistance changes with temperature, and a two-stage thermo-electric cooler.  
When the analyzer is on, a constant electrical current is directed through the detector, The IR  
beam is focused onto the detector surface, raising its temperature and lowering its electrical  
resistance that results in a change in the voltage drop across the detector.  
During those times that the IR beam is bright, the temperature of the detector is high; the  
resistance of the detector is correspondingly low and the its output voltage output is low. During  
those times when the IR beam intensity is low or completely blocked by the GFC Wheel mask, the  
temperature of the detector is lowered by the two-stage thermo-electric cooler, increasing the  
detectors resistance and raising the output voltage.  
10.3.4. Synchronous Demodulator (Sync/Demod) Assembly  
10.3.4.1. Overview  
While the photo-detector converts fluctuations of the IR beam into electronic signals, the Sync /  
Demod Board amplifies these signals and converts them into usable information. Initially the  
output by the photo-detector is a complex and continuously changing waveform made up of  
Measure and Reference pulses. The sync/demod board demodulates this waveform and outputs  
two analog DC voltage signals, corresponding to the peak values of these pulses. CO2 MEAS and  
CO2 REF are converted into digital signals by circuitry on the Motherboard then used by the CPU  
to calculate the CO2 concentration of the sample gas.  
Additionally the synch/demod board contains circuitry that controls the photo-detector’s  
thermoelectric cooler as well as circuitry for performing certain diagnostic tests on the analyzer.  
04584 Rev A1  
165  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
56V  
Bias  
CO2 MEAS  
Variable  
Gain Amp  
Dark  
Switch  
Sample  
&
Pre  
Amp  
Photo-  
detector  
Signal  
Conditioner  
Hold  
Circuits  
TEC Control  
PHT DRIVE  
E-Test  
Generator  
CO2 Reference  
Signal  
Conditioner  
(x4)  
Thermo-Electric  
Cooler  
Control Circuit  
E Test A Gate  
E Test B Gate  
Dark Test Gate  
Measure Gate  
Compact  
Measure Dark Gate  
Reference Gate  
Programmable  
Logic Device  
Reference Dark Gate  
Phase Lock Warning  
M/R Sensor  
From GFC  
Wheel  
Segment  
Sensor  
Segment Clock  
X1 Reference  
E Test Control  
Phase  
Lock  
Loop  
x10  
From CPU  
via Mother  
Board  
÷10  
Dark Switch  
Control  
X10 Clock  
M/R  
Segment  
Status LED  
Status LED  
Phase Lock  
Figure 10-12: GFC7000E Sync / Demod Block Diagram  
10.3.4.2. Signal Synchronization and Demodulation  
The signal emitted by the IR photo-detector goes through several stages of amplification before it  
can be accurately demodulated. The first is a pre-amplification stage that raises the signal to  
levels readable by the rest of the synch/demod board circuitry. The second is a variable  
amplification stage that is adjusted at the factory to compensate for performance variations of  
mirrors, detectors, and other components of the optical bench from instrument to instrument.  
The workhorses of the sync/demod board are the four sample-and-hold circuits that capture  
various voltage levels found in the amplified detector signal needed to determine the value of CO2  
MEAS and CO2 REF. They are activated by logic signals under the control of a compact  
programmable logic device (PLD), which in turn responds to the output of the Segment Sensor  
and M/R Sensor described in Figure 10–11.  
04584 Rev A1  
166  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
The four sample and hold circuits are:  
DESIGNATION  
ACTIVE WHEN:  
IR BEAM PASSING THROUGH  
SEGMENT SENSOR PULSE IS:  
Measure Gate  
Measure Dark Gate  
Reference Gate  
MEASUREMENT cell of GFC Wheel  
MEASUREMENT Cell of GFC Wheel  
REFERENCE cell of GFC Wheel  
REFERENCE cell of GFC Wheel  
HIGH  
LOW  
HIGH  
LOW  
Reference Dark Gate  
Timing for activating the Sample and Hold circuits is provided by a phase lock loop circuit (PLL).  
Using the segment sensor output as a reference signal the PLL generates clock signal at ten times  
that frequency. This faster clock signal is used by the PLD to make the sample and hold circuits  
capture the signal during the center portions of the detected waveform, ignore the rising and  
falling edges of the detector signal.  
Sample & Hold  
Active  
Detector  
Output  
Sample & Hold  
Inactive  
Figure 10-13: Sample & Hold Timing  
10.3.4.3. Phase Lock Warning  
In order to detect critical fault conditions such as a failure of either the segment sensor or the GFC  
wheel motor, the synch/demod board also performs a simple check of the above signal  
synchronization to make sure everything is operation. The PLD divides the X10 clock signal by ten  
and sends this signal back to the PLL circuit which compares it to the original segment sensor  
reference signal.  
If these two signals match, the PLL sends a status level to the PLD that the phase lock is OK. If  
for some reason the two signals do not match, the PLL alerts the PLD that phase lock has been  
lost and the PLD issues a phase lock warning to the CPU. Should this occur, A SYNC warning will  
appear on the analyzer’s front panel display (see Section 11.1.1 for more information).  
10.3.4.4. Sync/Demod Status LED’s  
The following two status LED’s located on the synch/demod board provide additional diagnostic  
tools for checking the GFC wheel rotation.  
04584 Rev A1  
167  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
Table 10-1: Sync/Demod Status LED Activity  
LED  
D1  
FUNCTION  
STATUS OK  
FAULT STATUS  
M/R Sensor Status  
LED flashes approximately  
2/second  
LED is stuck  
ON or OFF  
D2  
Segment Sensor  
Status  
LED flashes approximately  
6/second  
LED is stuck  
ON or OFF  
See Section 11.1.4 for more information.  
10.3.4.5. Photo-Detector Temperature Control  
The synch/demod board also contains circuitry that controls the IR photo-detector’s thermoelectric  
coolers. A drive voltage, PHT DRIVE, is supplied to the coolers by the synch/demod board which  
is adjusted by the synch/demod board based on a return signal called TEC control which alerts  
informs the synch/demod board of the detector’s temperature. The warmer the detector, the  
harder the coolers are driven.  
PHT DRIVE is one of the Test Functions viewable by the user via the form panel. Press <TST or  
TST> until it appears on the display.  
10.3.4.6. Dark Calibration Switch  
This switch initiates the Dark Calibration procedure. When initiated by the user (see Section 6.9.6  
for more details), the dark calibration process opens this switch, interrupting the signal from the  
IR photo-detector. This allows the analyzer to measure any offset caused by the synch/demod  
board circuitry.  
10.3.4.7. Electric Test Switch  
When active this circuit generates a specific waveform intended to simulate the function of the IR  
photo-detector but with a known set of value which is substituted for the detector’s actual signal  
via the dark switch. It may also be initiated by the user (see Section 6.9.5 for more details).  
10.3.5. Relay Board  
By actuating various switches and relays located on this board, the CPU controls the status of  
other key components. The relay board receives instructions in the form of digital signals over  
the I2C bus, interprets these digital instructions and activates its various switches and relays  
appropriately.  
Heater Control  
The two heaters attached to the sample chamber housing and the GFC wheel motor are controlled  
by solid state relays located on the relay board.  
The GFC wheel heater is simply turned on or off, however control of the bench heater also  
includes circuitry that selects which one of its two separate heating elements is activated  
depending on whether the instrument is running on 100 VAC, 115 VAC or 230 VAC line power.  
04584 Rev A1  
168  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
GFC Wheel Motor Control:  
The GFC wheel operates from a AC voltage supplied by a multi-input transformer located on the  
relay board. The step-down ratio of this transformer is controlled by factory-installed jumpers to  
adjust for 100 VAC, 115 VAC or 230 VAC line power. Other circuitry slightly alters the phase of the  
AC power supplied to the motor during start up based on whether line power is 50Hz or 60 Hz.  
Normally, the GFC Wheel Motor is always turning while the analyzer is on. A physical switch  
located on the relay board can be used to turn the motor off for certain diagnostic procedures.  
Zero/Span Valve Options  
Any zero/span/shutoff valve options installed in the analyzer are controlled by a set of electronic  
switches located on the relay board. These switches, under CPU control, supply the +12VDC  
needed to activate each valve’s solenoid.  
IR Source  
The Relay board supplies a constant 11.5VDC to the IR Source. Under normal operation the IR  
source is always on.  
10.3.5.1. Status LED’s  
Eight LED’s are located on the analyzer’s relay board to show the current status on the various  
control functions performed by the relay board (see Figure 10-14). They are:  
Table 10-2: Relay Board Status LED’s  
LED  
COLOR  
FUNCTION  
STATUS WHEN LIT  
STATUS WHEN UNLIT  
D1  
RED  
Watchdog Circuit Cycles On/Off Every 3 Seconds under direct control of the  
analyzer’s CPU.  
D2  
D3  
D4  
D5  
YELLOW  
YELLOW  
YELLOW  
GREEN  
Wheel Heater  
Bench Heater  
Spare  
HEATING  
HEATING  
N/A  
NOT HEATING  
NOT HEATING  
N/A  
Sample/Cal Gas  
Valve Option  
Valve Open to CAL GAS  
FLOW  
Valve Open to SAMPLE GAS  
FLOW  
D6  
D7  
D8  
GREEN  
GREEN  
GREEN  
Zero/Span Gas  
Valve Option  
Valve Open to SPAN GAS  
FLOW  
Valve Open to ZERO GAS  
FLOW  
Shutoff Valve  
Option  
Valve Open to CAL GAS  
FLOW  
Valve CLOSED to CAL GAS  
FLOW  
IR SOURCE  
Source ON  
Source OFF  
04584 Rev A1  
169  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
DC VOLTAGE TEST  
POINTS  
STATUS LED’s  
RELAY PCA  
PN 04135  
Figure 10-14: Location of relay board Status LED’s  
10.3.5.2. I2C Watch Dog Circuitry  
Special circuitry on the relay board monitors the activity on the I2C bus and drives LED D1.  
Should this LED ever stay ON or OFF for 30 seconds, the watchdog circuit will automatically shut  
of all valves as well as turn off the IR Source and all heaters. The GFC wheel motor will still be  
running as will the Sample Pump, which is not controlled by the relay board.  
10.3.6. Mother Board  
This printed circuit assembly provides a multitude of functions including, A/D conversion, digital  
input/output, PC-104 to I2C translation, temperature sensor signal processing and is a pass  
through for the RS-232 and RS-485 signals.  
10.3.6.1. A to D Conversion  
Analog signals, such as the voltages received from the analyzer’s various sensors, are converted  
into digital signals that the CPU can understand and manipulate by the analog to digital converter  
(A/D). Under the control of the CPU, this functional block selects a particular signal input (e.g.  
BOX TEMP, CO2 MEAS, CO2 REF, etc.) and then coverts the selected voltage into a digital word.  
The A/D consists of a voltage-to-frequency (V-F) converter, a programmable logic device (PLD),  
three multiplexers, several amplifiers and some other associated devices. The V-F converter  
produces a frequency proportional to its input voltage. The PLD counts the output of the V-F  
during a specified time period, and sends the result of that count, in the form of a binary number,  
to the CPU.  
The A/D can be configured for several different input modes and ranges but in the MGFC7000E is  
used in uni-polar mode with a +5 V full scale. The converter includes a 1% over and under-  
range. This allows signals from –0.05 V to +5.05 V to be fully converted.  
04584 Rev A1  
170  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
For calibration purposes, two reference voltages are supplied to the A/D converter: Reference  
Ground and +4.096 VDC. During calibration, the device measures these two voltages, outputs  
their digital equivalent to the CPU. The CPU uses these values to compute the converter’s offset  
and slope and uses these factors for subsequent conversions.  
See Section 6.9.4 for instructions on performing this calibration.  
10.3.6.2. Sensor Inputs  
The key analog sensor signals are coupled to the A/D through the master multiplexer from two  
connectors on the motherboard. 100K terminating resistors on each of the inputs prevent cross  
talk from appearing on the sensor signals.  
Co2 Measure And Reference  
These are the primary signals that are used in the computation of the CO2 concentration. They  
are the demodulated IR-sensor signals from the sync demodulator board.  
Sample Pressure And Flow  
These are analog signals from two sensors that measure the pressure and flow rate of the gas  
stream at the outlet of the sample chamber. This information is used in two ways. First, the  
sample pressure is used by the CPU to calculate CO2 Concentration. Second, the pressure and  
flow rate are monitored as a test function to assist the user in predicting and troubleshooting  
failures.  
10.3.6.3. Thermistor Interface  
This circuit provides excitation, termination and signal selection for several negative-coefficient,  
thermistor temperature sensors located inside the analyzer. They are:  
Sample Temperature Sensor  
The source of this signal is a thermistor located inside the sample chamber of the Optical Bench.  
It measures the temperature of the sample gas in the chamber. This data is used to during the  
calculation of the CO2 concentration value.  
Bench Temperature Sensor  
This thermistor, attached to the sample chamber housing, reports the current temperature of the  
chamber housing to the CPU as part of the bench heater control loop.  
Wheel Temperature Sensor  
This thermistor attached a the heat-sync on the GFC wheel motor assembly reports the current  
temperature of the wheel/motor assembly to the cpu as part of the Wheel Heater control loop.  
Box Temperature Sensor  
A thermistor is attached to the mother board. It measures the analyzer’s inside temperature.  
This information is stored by the CPU and can be viewed by the user for troubleshooting purposes  
via the front panel display (see Section 11.1.2).  
04584 Rev A1  
171  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
10.3.6.4. Analog Outputs  
The analyzer comes equipped with four analog outputs: A1, A2, A3 and a fourth that is a spare.  
A2 and A1 Output  
The first two, A2 and A1 are normally set up to operate in parallel so that the same data can be  
sent to two different recording devices. While the names imply that one should be used for  
sending data to a chart recorder and the other for interfacing with a datalogger, either can be  
used for both applications.  
Both of these channels output a signal that is proportional to the CO2 concentration of the sample  
gas. The A1 and A2 outputs can be slaved together or set up to operated independently. A variety  
of scaling factors are available, See Section 6.9.4 for information on setting the range type and  
scaling factors for these output channels.  
A3 Output  
The third analog output, labeled A3 is special. It can be set by the user (see Section 6.9.9) to  
carry the current signal level of any one of the parameters accessible through the A3 menu of the  
unit’s software.  
Spare Output  
The fourth analog output, termed SPARE is not used in the MGFC7000E.  
In its standard configuration, the analyzer comes with all four of these channels set up to output a  
DC voltage. However, 4-20mA current loop drivers can be purchased for the first three of these  
outputs: A2, A2 & A3.  
Output Loop-back  
All four analog outputs are connected back to the A/D converter through a Loop-back circuit. This  
permits the voltage outputs to be calibrated by the CPU without need for any additional tools or  
fixtures.  
10.3.6.5. Internal Digital I/O  
This channel is used to communicate digital status and control signals about the operation of key  
components of the Optical Bench. The CPU sends signals to the synch/demod board that initiate  
the ELECTRICAL TEST and DARK CALIBRATION procedures. Likewise, the synch/demod board  
uses this interface to send the SYNC warning signal to the CPU (see Sections 6.9.5, 6.9.6 and  
11.1.1).  
04584 Rev A1  
172  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
10.3.6.6. External Digital I/O  
This External Digital I/O performs two functions.  
Status Outputs  
Logic-Level voltages are output through an optically isolated 8-pin connector located on the rear  
panel of the analyzer. These outputs convey good/bad and on/off information about certain  
analyzer conditions. They can be used to interface with certain types of programmable devices  
(see Section 6.13.1.1).  
Control Inputs  
By applying +5VDC power supplied from an external source such as a PLC or Datalogger (see  
Section 6.13.1.2), Zero and Span calibrations can be initiated by contact closures on the rear  
panel.  
10.3.7. I2C Data Bus  
An I2C data bus is used to communicate data and commands between the cpu and the  
keyboard/display interface and the relay board. I2C is a two-wire, clocked, digital serial i/o bus  
that is used widely in commercial and consumer electronic systems. A transceiver on the  
motherboard converts data and control signals from the PC-104 bus to I2C. The data is then fed  
to the keyboard/display interface and finally onto the relay board.  
Interface circuits on the keyboard/display interface and relay boards convert the i2c data to  
parallel inputs and outputs. An additional, interrupt line from the keyboard to the motherboard  
allows the CPU to recognize and service key presses on the keyboard.  
Power up Circuit  
This circuit monitors the +5V power supply during start-up and sets the Analog outputs, external  
digital I/O ports, and I2C circuitry to specific values until the CPU boots and the instrument  
software can establish control.  
10.3.8. Power Supply/ Circuit Breaker  
The analyzer operates on 100 VAC, 115 VAC or 230 VAC power at either 50Hz or 60Hz. Individual  
units are set up at the factory to accept any combination of these five attributes. As illustrated in  
Figure 10-15, power enters the analyzer through a standard IEC 320 power receptacle located on  
the rear panel of the instrument. From there it is routed through the On/Off switch located in the  
lower right corner of the Front Panel. A 6.75 Amp circuit breaker is built into the ON/OFF Switch.  
AC power is distributed directly to the sample gas pump. The bench and GFC wheel heaters as  
well as the GFC wheel receive AC power via the relay board.  
AC Line power is converted stepped down and converted to DC power by two DC power supplies.  
One supplies +12 VDC, for valves and the IR source, while a second supply provides +5 VDC and  
±15 VDC for logic and analog circuitry. All DC voltages are distributed via the relay board.  
04584 Rev A1  
173  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
CAUTION  
Should the AC power circuit breaker trip, investigate and correct the  
condition causing this situation before turning the analyzer back on.  
ON/OFF  
SWITCH  
AC POWER  
Pressure  
Sensors  
ENTRANCE  
Display  
Keypad  
CPU  
PS 1 (+5 VDC; ±15 VDC)  
RELAY  
BOARD  
KEY  
AC POWER  
DC POWER  
Mother  
Board  
PS 2 (+12 VDC)  
Sync/Demod  
IR Source  
Cooling Fan  
Pump  
M/R &  
Segment  
Sensors  
GFC  
Wheel  
Motor  
Valve  
Options  
Heaters  
Figure 10-15: Power Distribution Block Diagram  
10.4. Interface  
The analyzer has several ways to communicate the outside world, see Figure 10-16. Users can  
input data and receive information directly via the Front panel keypad and display. Direct  
communication with the CPU is also available by way of the analyzer’s RS232 & RS485 I/O ports  
or an optional Ethernet port. The analyzer can also send and receive different kinds of information  
via its external digital I/O connectors and the three analog outputs located on the rear panel.  
04584 Rev A1  
174  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
COMM A  
Male  
RS–232 ONLY  
RS-232 or RS–485  
COMM B  
Female  
CPU  
Mother  
Board  
Control Inputs:  
ETHERNET  
OPTION  
1 – 6  
Status Outputs:  
1 – 8  
PC/104 BUS  
Analog Outputs  
KEYBOARD  
A1  
A2  
Optional  
4-20 mA  
I2C BUS  
A3  
I2C BUS  
A4ST  
FRONT PANEL DISPLAY  
RELAY  
BOARD  
Figure 10-16: Interface Block Diagram  
10.4.1. Front Panel Interface  
MODE FIELD MESSAGE FIELD  
CONCENTRATION FIELD STATUS LED’s  
LOCKING SCREW  
FASTENER  
FASTENER  
SAMPLE  
CAL  
CO2 = 400.0  
SETUP  
RANGE = 500.0 PPM  
SAMPLE A  
<TST TST> CAL  
FAULT  
POWER  
GAS FILTER CORRELATION CO2 ANALYZER- MODEL GFC7000E  
KEY DEFINITIONS KEYBOARD  
ON / OFF SWITCH  
Figure 10-17: GFC7000E Front Panel Layout  
The most commonly used method for communicating with the MGFC7000E Analyzer is via the  
instrument’s front panel which includes a set of three status LEDs, a vacuum florescent display  
and a keyboard with 8 context sensitive keys.  
10.4.1.1. Analyzer Status LED’s  
Three LEDS are used to inform the user of the instruments basic operating status  
04584 Rev A1  
175  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
Table 10-3: Front Panel Status LED’s  
NAME  
COLOR STATE  
DEFINITION  
SAMPLE  
Green  
Off  
On  
Unit is not operating in sample mode, iDAS is disabled.  
Sample Mode active; Front Panel Display being updated, iDAS data being stored.  
Unit is operating in sample mode, front panel display being updated, iDAS hold-off  
mode is ON, iDAS disabled  
Blinking  
CAL  
Yellow  
Red  
Off  
On  
Auto Cal disabled  
Auto Cal enabled  
Blinking  
Unit is in calibration mode  
FAULT  
Off  
CO2 warnings exist  
Warnings exist  
Blinking  
10.4.1.2. Keyboard  
A row of eight keys just below the vacuum florescent display (see Figure 10-17) is the main  
method by which the user interacts with the analyzer. As the software is operated, labels appear  
on the bottom row of the display directly above each active key, defining the function of that key  
as it is relevant for the operation being performed. Pressing a key causes the associated  
instruction to be performed by the analyzer.  
Note that the keys do not auto-repeat. In circumstances where the same key must be activated  
for two consecutive operations, it must be released and re-pressed.  
10.4.1.3. Display  
The main display of the analyzer is a vacuum florescent display with two lines of 40 text  
characters each. Information is organized in the following manner (see Figure 10-17):  
Mode Field: Displays the name of the analyzer’s current operating mode.  
Message Field: Displays a variety of informational messages such as warning messages, operation  
data and response messages during interactive tasks.  
Concentration Field: Displays the actual concentration of the sample gas currently being measured  
by the analyzer  
Keypad Definition Field: Displays the definitions for the row of keys just below the display. These  
definitions dynamic, context sensitive and software driven.  
04584 Rev A1  
176  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
10.4.1.4. Keyboard/Display Interface Electronics  
I2C to Relay Board  
Key Press  
Detect  
Display Data  
Decoder  
Display  
Controller  
Display Power  
Watchdog  
Keypad  
Decoder  
I2C Interface  
Serial  
Data  
From 5 VDC  
Power Supply  
Optional  
Maintenance  
LED  
Sample LED  
(Green)  
Maint.  
Switch  
2nd Lang.  
Switch  
Cal LED  
(Yellow)  
2 x 40 CHAR. VACUUM  
FLUORESCENT DISPLAY  
Fault LED  
(Red)  
KEYBOARD  
Beeper  
FRONT PANEL  
Figure 10-18: Keyboard and Display Interface Block Diagram  
The keyboard/display interface electronics of the MGFC7000E Analyzer watches the status of the  
eight front panel keys, alerts the CPU when keys are depressed, translates data from parallel to  
serial and back and manages communications between the keyboard, the CPU and the front panel  
display. Except for the Keyboard interrupt status bit, all communication between the CPU and the  
keyboard/display is handle by way of the instrument’s I2C buss. The CPU controls the clock signal  
and determines when the various devices on the bus are allowed to talk or required to listen. Data  
packets are labeled with addresses that identify for which device the information is intended.  
Keypad Decoder  
Each key on the front panel communicates with a decoder IC via a separate analog line. When a  
key is depressed the decoder chip notices the change of state of the associated signal; latches and  
holds the state of all eight lines (in effect creating an 8-bit data word); alerts the key-depress-  
detect circuit (a flip-flop IC); translates the 8-bit word into serial data and; sends this to the I2C  
interface chip.  
Key-Depress-Detect Circuit  
This circuit flips the state of one of the inputs to the I2C interface chip causing it to send an  
interrupt signal to the CPU  
I2C Interface Chip  
04584 Rev A1  
177  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
This IC performs several functions:  
Using a dedicated digital status bit, it sends an interrupt signal alerting the CPU that new  
data from the keyboard is ready to send.  
Upon acknowledgement by the CPU that it has received the new keyboard data, the I2C  
interface chip resets the key-depress-detect flip-flop.  
In response to commands from the CPU, it turns the front panel status LEDs on and off and  
activates the beeper.  
Informs the CPU when the optional maintenance and second language switches have been  
opened or closed (see Chapter 5 for information on these options).  
Display Data Decoder  
This decoder the serial translates the data sent by the CPU (in TTY format) into a bitmapped  
image which is sent over a parallel data bus to the display.  
Display Controller  
This circuit manages the interactions between the display data decoder and the display itself. It  
generates a clock pulse that keeps the two devices synchronized. It can also, in response to  
commands from the CPU turn off and/or reset the display.  
Additionally, for analyzers with the optional maintenance switch is installed (See Chapter 5), the  
display controller turns on an LED located on the back of the keyboard interface PCA whenever  
the instrument is placed in maintenance mode.  
Display Power Watchdog  
The Model GFC7000E’s display can begin to show garbled information or lock-up if the DC voltage  
supplied to it falls too low, even momentarily. To alleviate this, a brown-out watchdog circuit  
monitors the level of the power supply and in the event that the voltage level falls below a certain  
level, turns the display off, then on resetting it  
I2C Link To The Relay PCA  
While the CPU’s I2C communication with the relay board is also routed through the  
keyboard/display interface, information passed to and from the relay board via this channel is not  
recognized by, acted upon or affected by the circuitry of the keyboard/display interface.  
10.5. Software Operation  
The Model GFC7000E Gas Filter Correlation Carbon Dioxide Analyzer is at its heart a high  
performance, 386-based microcomputer running MS-DOS. Inside the DOS shell, special software  
developed by Teledyne Instruments interprets user commands via the various interfaces,  
performs procedures and tasks, stores data in the CPU’s various memory devices and calculates  
the concentration of the sample gas.  
04584 Rev A1  
178  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
DOS Shell  
AI FIRMWARE  
Analyzer Operations  
Calibration Procedures  
Configuration Procedures  
Autonomic Systems  
Memory Handling  
IDAS Records  
Calibration Data  
System Status Data  
PC/104 BUS  
Diagnostic Routines  
ANALYZER  
HARDWARE  
Interface Handling  
Sensor input Data  
Display Messages  
Keypad  
Measurement  
Algorithm  
Analog Output Data  
RS232 & RS485  
External Digital I/O  
PC/104 BUS  
Linearization Table  
Figure 10-19: Basic Software Operation  
10.5.1. Adaptive Filter  
The MGFC7000E software processes the CO2 MEAS and CO2 REF signals, after they are  
digitized by the motherboard, through an adaptive filter built into the software. Unlike other  
analyzers that average the output signal over a fixed time period, the MGFC7000E averages over  
a set number of samples, where each sample is 0.2 seconds. This is technique is known as  
boxcar averaging. During operation, the software automatically switches between two different  
length filters based on the conditions at hand. Once triggered, the short filter remains engaged  
for a fixed time period to prevent chattering.  
During conditions of constant or nearly constant concentration the software, by default, computes  
an average of the last 750 samples, or approximately 150 seconds. This provides the calculation  
portion of the software with smooth stable readings. If a rapid change in concentration is detected  
the filter includes, by default, the last 48 samples, approximately 10 seconds of data, to allow the  
analyzer to more quickly respond. If necessary, these boxcar lengths can be changed between 1  
and 1000 samples but with corresponding tradeoffs in rise time and signal-to-noise ratio (contact  
customer service for more information).  
Two conditions must be simultaneously met to switch to the short filter. First the instantaneous  
concentration must exceed the average in the long filter by a fixed amount. Second the  
instantaneous concentration must exceed the average in the long filter by a portion, or  
percentage, of the average in the long filter.  
10.5.2. Calibration - Slope and Offset  
Calibration of the analyzer is performed exclusively in software.  
04584 Rev A1  
179  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
During instrument calibration (see Chapter 7) the user enters expected values for zero and span  
via the front panel keypad and commands the instrument to make readings of calibrated sample  
gases for both levels. The readings taken are adjusted, linearized, and compared to the expected  
values, With this information the software computes values for instrument slope and offset and  
stores these values in memory for use in calculating the CO2 concentration of the sample gas.  
The instrument slope and offset values recorded during the last calibration can be viewed by  
pressing the following keystroke sequence:  
SAMPLE  
RANGE = 50.0 MGM  
CO =XX.XX  
SETUP  
< TST TST > CAL  
SAMPLE  
OFFSET = 0.000  
CO =XX.XX  
SETUP  
< TST TST > CAL  
SAMPLE  
TIME = 16:23:34  
CO =XX.XX  
SETUP  
< TST TST > CAL  
SAMPLE  
SLOPE = 1.000  
CO =XX.XX  
SETUP  
< TST TST > CAL  
10.5.3. Measurement Algorithm  
Once the IR photo-detector is signal is demodulated into CO2 MEAS and CO2 REF by the  
sync/demod board and converted to digital data by the mother board the MGFC7000E analytical  
software calculates the ratio between CO2 MEAS and CO2 REF. this value is compared to a look-  
up table is used, with interpolation, to linearize the response of the instrument. The linearized  
concentration value is combined with calibration slope and offset values, then normalized for  
changes in sample gas pressure to produce the final CO2 concentration. This is the value that is  
displayed on the instrument front panel display and is stored in memory by the analyzer’s iDAS  
system.  
10.5.4. Temperature and Pressure Compensation  
Changes in pressure can have a noticeable, effect on the CO2 concentration calculation. To  
account for this, the Model GFC7000E software includes a feature which allows the instrument to  
compensation of the CO2 calculations based on changes in ambient pressure.  
The TPC feature multiplies the analyzer’s CO2 concentration by a factor which is based on the  
difference between the ambient pressure of the sample gas normalized to standard atmospheric  
pressure. As ambient pressure increases, the compensated CO2 concentration is increased.  
10.5.5. Internal Data Acquisition System (iDAS)  
The iDAS is designed to implement predictive diagnostics that stores trending data for users to  
anticipate when an instrument will require service. Large amounts of data can be stored in non-  
volatile memory and retrieved in plain text format for further processing with common data  
analysis programs. The iDAS has a consistent user interface in all Teledyne Instruments  
analyzers. New data parameters and triggering events can be added to the instrument as needed.  
04584 Rev A1  
180  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
THEORY OF OPERATION  
Depending on the sampling frequency and the number of data parameters the iDAS can store  
several months of data, which are retained even when the instrument is powered off or a new  
firmware is installed. The iDAS permits users to access the data through the instrument’s front  
panel or the remote interface. The latter can automatically download stored data for further  
processing. For information on using the iDAS, refer to Sections 6.12.  
User Notes  
04584 Rev A1  
181  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
11. TROUBLESHOOTING & REPAIR PROCEDURES  
This section contains a variety of methods for identifying the source of performance problems with  
the analyzer. Also included in this section are procedures that are used in repairing the  
instrument.  
CAUTION  
The operations outlined in this chapter are to be performed by qualified  
maintenance personnel only.  
CAUTION  
Risk of electrical shock. Disconnect power before performing the  
following operations.  
11.1. General Troubleshooting Hints  
The analyzer has been designed so that problems can be rapidly detected, evaluated and  
repaired. During operation, the analyzer continuously performs self-check diagnostics and  
provides the ability to monitor the key operating parameters of the instrument without disturbing  
monitoring operations.  
A systematic approach to troubleshooting will generally consist of the following four steps:  
1. Note any WARNING MESSAGES and take corrective action as required.  
2. Examine the values of all TEST functions and compare to factory values. Note any major  
deviations from the factory values and take correction action as required.  
3. Use the internal electronic status LED’s to determine whether the CPU and I2C buses are  
running, and if the sync/demodulator and relay board are operating properly. Verify that the  
DC power supplies are operating properly by checking the voltage test points on the relay  
board. Please note that the analyzer’s DC power wiring is color-coded and these colors match  
the color of the corresponding test points on the relay board.  
4. SUSPECT A LEAK FIRST! Data from Teledyne Instruments’ service department indicates  
that 50% of all problems are eventually traced to leaks in the pneumatic connections and gas  
lines of the analyzer itself, the source of zero air, span gases or sample gas delivery system.  
Check for gas flow problems such as clogged or blocked internal/external gas lines, damaged  
seals, punctured gas lines, a damaged pump diaphragm, etc.  
5. Follow the procedures defined in Section 11.4 for confirming that the analyzer’s basic  
components are working (power supplies, CPU, relay board, sync/demod board, keypad, GFC  
wheel motor, etc.). See Figure 3-11 for general layout of components and sub-assemblies in  
the analyzer. See the wiring Interconnect Drawing and Interconnect List, documents 04216  
and 04217.  
04584 Rev A1  
183  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
11.1.1. Interpreting WARNING Messages  
The most common and/or serious instrument failures will result in a warning message being  
displayed on the front panel. Table 11-1 lists warning messages, along with their meaning and  
recommended corrective action.  
It should be noted that if more than two or three warning messages occur at the same time, it is  
often an indication that some fundamental analyzer sub-system (power supply, relay board,  
motherboard) has failed rather than indication of the of the specific failures referenced by the  
warnings. In this case, it is recommended that proper operation of power supplies (see Section  
11.4.2), the relay board (see Section 11.4.5), and the A/D Board (see Section11.4.7.1) be  
confirmed before addressing the specific warning messages.  
The analyzer will alert the user that a Warning Message is active by displaying the keypad label  
MSG on the Front Panel. In this case the Front panel display will look something like the  
following:  
SAMPLE  
RANGE=500.0 PPM  
CO2 = 00.00  
CLR SETUP  
<TST TST> CAL  
MSG  
The analyzer will also alert the user via the Serial I/O COM port(s) and cause the FAULT LED on  
the front panel to blink.  
To view or clear the various warning messages press:  
SAMPLE  
WHEEL TEMP WARNING  
CAL MSG  
CO2 = XX.XX  
CLR SETUP  
TEST deactivates Warning  
Messages until New warning(s)  
are activated  
TEST  
MSG activates Warning  
SAMPLE  
RANGE=500.00 PPM  
MSG  
CO2 = XX.XX  
CLR SETUP  
Messages.  
<TST TST> keys replaced with  
< TST TST > CAL  
TEST key  
SAMPLE  
WHEEL TEMP WARNING  
CO2 = XX.XX  
Press CLR to clear the  
message currently being  
Displayed.  
< TST TST > CAL  
MSG  
CLR SETUP  
If more than one warning is  
active the next message will  
take its place  
Once the last warning has been  
cleared, the analyzer returns to  
SAMPLE Mode  
Make sure warning messages  
are not due to  
legitimate problems..  
Figure 11-1: Viewing and Clearing Warning Messages  
NOTE: A failure of the analyzer’s CPU or mother board can result in any or all of the following  
messages.  
04584 Rev A1  
184  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
Table 11-1: Warning Messages - Indicated Failures  
WARNING  
MESSAGE  
FAULT CONDITION  
The optical bench temp Bad bench heater  
POSSIBLE CAUSES  
BENCH TEMP  
WARNING  
is controlled at 48 ± 5  
Bad bench temperature sensor  
°C.  
Bad relay controlling the bench heater  
Entire relay board is malfunctioning  
I2C buss malfunction  
BOX TEMP  
WARNING  
Box Temp is < 5 °C or  
> 48 °C.  
NOTE: Box temperature typically runs ~7oc warmer than  
ambient temperature.  
Poor/blocked ventilation to the analyzer.  
Stopped exhaust-fan  
Ambient temperature outside of specified range  
Measured concentration value is too high or low.  
Concentration slope value to high or too low  
Measured concentration value is too high.  
Concentration offset value to high.  
Failed disk on chip  
CANNOT DYN  
SPAN  
CANNOT DYN  
ZERO  
CONFIG  
INITIALIZED  
Dynamic Span  
operation failed  
Dynamic Zero  
operation failed  
Configuration and  
Calibration data reset  
to original Factory  
state.  
User erased data  
Concentration alarm 1  
is enabled and the  
measured CO2 level is  
the set point.  
CONC ALRM1  
WARNING  
Concentration alarm 2  
is enabled and the  
measured CO2 level is  
the set point.  
CONC ALRM2  
WARNING  
DATA  
Data Storage in iDAS  
was erased  
The CPU is unable to  
Communicate with the  
Front Panel Display  
/Keyboard  
Failed disk on chip  
User cleared data  
INITIALIZED  
FRONT PANEL  
WARN  
Warning only appears on serial I/O com port(s)  
Front panel display will be frozen, blank or will not respond.  
Failed keyboard  
I2C buss failure  
Loose connector/wiring  
PHOTO TEMP  
WARNING  
PHT DRIVE is >2500  
mVDC  
Failed IR photo-detector  
Failed sync/demod board  
IR photo-detector improperly attached to the sample  
chamber  
Bench temp too high.  
REAR BOARD  
NOT DET  
Mother Board not  
detected on power up.  
Warning only appears on serial i/o com port(s)  
Front panel display will be frozen, blank or will not respond.  
Massive failure of mother board  
I2C buss failure  
RELAY BOARD  
WARN  
The CPU cannot  
communicate with the  
Relay Board.  
Failed relay board  
Loose connectors/wiring  
SAMPLE FLOW  
WARN  
Sample flow rate is <  
500 cc/min or > 1000  
cc/min.  
Failed sample pump  
Blocked sample inlet/gas line  
Dirty particulate filter  
Leak downstream of critical flow orifice  
Failed flow sensor/circuitry  
04584 Rev A1  
185  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
Table 11-1: Warning Messages – Indicated Failures (cont.)  
WARNING  
MESSAGE  
FAULT CONDITION  
POSSIBLE CAUSES  
SAMPLE PRES  
WARN  
Sample Pressure is  
<10 in-Hg or  
> 35 in-Hg  
Normally 29.92 in-Hg  
at sea level decreasing  
at 1 in-Hg per 1000 ft  
of altitude (with no  
flow – pump  
If sample pressure is < 10 in-hg:  
o Blocked particulate filter  
o Blocked sample inlet/gas line  
o Failed pressure sensor/circuitry  
If sample pressure is > 35 in-hg:  
o Blocked vent line on pressurized sample/zero/span gas  
supply  
o Bad pressure sensor/circuitry  
disconnected).  
SAMPLE TEMP  
WARN  
Sample temperature is Ambient temperature outside of specified range  
< 10oC or > 100oC.  
Failed bench heater  
Failed bench temperature sensor  
Relay controlling the bench heater  
Failed relay board  
I2C buss  
SOURCE  
WARNING  
Occurs when CO2 Ref  
is <1250 mVDC or  
>4950 mVDC.  
GFC wheel stopped  
Failed sync/demod board  
If status LED’s on the sync/demod board ARE flashing the  
cause is most likely a failed:  
IR source  
Either of these  
conditions will result in Relay board  
I2C buss  
IR photo-detector  
an invalid M/R ratio.  
SYNC WARNING Phase Lock Loop (PLL)  
has lost lock on wheel  
GFC wheel stopped  
GFC wheel rotation very slow/dragging  
Failed sync/demod board  
rotation.  
If status LED’s on the sync/demod board ARE flashing the  
cause is most likely an a problem with the analyzer’s main  
power:  
Intermittent loss of the power supply too short to cause a  
SYSTEM RESET warning;  
Frequency problem with the AC mains.  
This message occurs at power on.  
If it is confirmed that power has not been interrupted:  
Failed +5 VDC power,  
SYSTEM RESET  
The computer has  
rebooted.  
Fatal error caused software to restart  
Loose connector/wiring  
WHEEL TEMP  
WARNING  
The filter wheel  
Blocked Cooling Vents below GFC Assembly  
Analyzer’s top cover removed  
Wheel heater  
temperature is  
controlled at 68 ± 5 °C  
Wheel temperature sensor  
Relay controlling the wheel heater  
Entire relay board  
I2C buss  
11.1.2. Fault Diagnosis with TEST Functions  
Besides being useful as predictive diagnostic tools, the test functions viewable from the front  
panel can be used to isolate and identify many operational problems when combined with a  
thorough understanding of the analyzer’s theory of operation (see Chapter 10).  
The acceptable ranges for these test functions are listed in the “Nominal Range” column of the  
analyzer Final Test and Validation Data Sheet (p/n 04307) shipped with the instrument. Values  
outside these acceptable ranges indicate a failure of one or more of the analyzer’s subsystems.  
04584 Rev A1  
186  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
Functions whose values are still within the acceptable range but have significantly changed from  
the measurement recorded on the factory data sheet may also indicate a failure. A worksheet has  
been provided in Appendix C to assist in recording the value of these test functions.  
The following table contains some of the more common causes for these values to be out of  
range.  
Table 11-2: Test Functions - Indicated Failures  
TEST  
FUNCTIONS  
(As Displayed)  
INDICATED FAILURE(S)  
Time of day clock is too fast or slow  
To adjust see Section 6.6.  
TIME  
Battery in clock chip on CPU board may be dead.  
Incorrectly configured measurement range(s) could cause response problems with a  
Datalogger or chart recorder attached to one of the analog output.  
If the Range selected is too small, the recording device will over range.  
If the Range is too big, the device will show minimal or no apparent change in readings.  
Indicates noise level of instrument or CO2 concentration of sample gas (see Section 11.3.2  
for causes).  
RANGE  
STABIL  
If the value displayed is too high the IR Source has become brighter. Adjust the variable  
gain potentiometer on the sync/demod board  
If the value displayed is too low or constantly changing and the CO2 REF is OK:  
Failed multiplexer on the mother board  
Failed sync/demod board  
Loose connector or wiring on sync/demod board  
Flow of purge gas to the GFC wheel housing may have stopped  
If the value displayed is too low or constantly changing and the CO2 REF is BAD:  
GFC wheel stopped or rotation is too slow  
Failed sync/demod board IR source  
CO2 MEAS  
&
CO2 REF  
Failed IR source  
Failed relay board  
Failed I2C buss  
Failed IR photo-detector  
When the analyzer is sampling zero air and the ratio is too low:  
The reference cell of the GFC wheel is contaminated or leaking.  
The alignment between the GFC wheel and the segment sensor, the M/R sensor or both  
is incorrect.  
MR RATIO  
Failed sync/demod board  
Flow of purge gas to the GFC wheel housing may have stopped  
When the analyzer is sampling zero air and the ratio is too high:  
Zero air is contaminated  
Failed IR photo-detector  
See Table 11-1 for SAMPLE PRES WARN  
PRES  
Check for gas flow problems. see Section 11.1.6  
SAMPLE FL  
SAMPLE TEMP should be close to BENCH TEMP. Temperatures outside of the specified  
range or oscillating temperatures are cause for concern  
SAMPLE  
TEMP  
Bench temp control improves instrument noise, stability and drift. Temperatures outside of  
the specified range or oscillating temperatures are cause for concern. See Table 11-1 for  
BENCH TEMP WARNING  
BENCH  
TEMP  
Wheel temp control improves instrument noise, stability and drift. Outside of set point or  
oscillating temperatures are cause for concern. See Table 11-1 for WHEEL TEMP  
WARNING  
WHEEL  
TEMP  
If the box temperature is out of range, check fan in the power supply module. Areas to the  
side and rear of instrument should allow adequate ventilation. See Table 11-1 for BOX  
TEMP WARNING.  
BOX TEMP  
04584 Rev A1  
187  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
Table 11-2: Test Functions - Indicated Failures (cont.)  
TEST  
FUNCTIONS  
(As Displayed)  
INDICATED FAILURE(S)  
If this drive voltage is out of range it may indicate one of several problems:  
- A poor mechanical connection between the various components in inside the detector  
housing  
- An electronic failure of the IR Photo-Detector’s built-in cooling circuitry, or;  
- A temperature problem inside the analyzer chassis. In this case other temperature  
warnings would also be active such as BENCH TEMP WARNING or BOX TEMP  
WARNING.  
PHT DRIVE  
Values outside range indicate  
Contamination of the zero air or span gas supply  
Instrument is miss-calibrated  
Blocked gas flow  
Contaminated or leaking GFC wheel (either chamber)  
Faulty IR photo-detector  
SLOPE  
Faulty sample faulty IR photo-detector pressure sensor (P1) or circuitry  
Invalid M/R ratio (see above)  
Bad/incorrect span gas concentration due.  
Values outside range indicate  
Contamination of the zero air supply  
Contaminated or leaking GFC wheel (either chamber)  
Faulty IR photo-detector  
OFFSET  
11.1.3. Using the Diagnostic Signal I/O Function  
The Signal I/O parameters found under the DIAG Menu (see Section 6.9.2 and Appendix A)  
combined with a thorough understanding of the instruments theory of operation (found in Chapter  
10) are useful for troubleshooting in three ways:  
The technician can view the raw, unprocessed signal level of the analyzer’s critical inputs  
and outputs.  
All of the components and functions that are normally under algorithmic control of the CPU  
can be manually exercised.  
The technician can directly control the signal level Analog and Digital Output signals.  
This allows the technician to systematically observe the effect of directly controlling these signals  
on the operation of the analyzer. Below in Figure 11-2 is an example of how to use the signal I/O  
menu to view the raw voltage of an input signal or to control the state of an output voltage or  
control signal. The specific parameter will vary depending on the situation.  
04584 Rev A1  
188  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
SAMPLE*  
RANGE = 500.000 PPM  
CO2 =X.XXX  
< TST TST > CAL  
SETUP  
SAMPLE  
ENTER SETUP PASS : 818  
8
1
8
ENTR EXIT  
SETUP X.X  
PRIMARY SETUP MENU  
CFG DAS RNGE PASS CLK MORE  
EXIT  
EXIT  
EXIT  
SETUP X.X  
SECONDARY SETUP MENU  
COMM VARS DIAG  
DIAG  
SIGNAL I/O  
PREV NEXT  
ENTR  
DIAG I/O  
0 ) EXT_ZERO_CAL=ON  
PREV NEXT JUMP  
PRNT EXIT  
If parameter is an  
input signal  
If parameter is an output  
signal or control  
DIAG I/O  
28) SAMPLE_PRESSURE=2540 MV  
DIAG I/O  
22) WHEEL_HTR=ON  
PREV NEXT JUMP  
PRNT EXIT  
PREV NEXT JUMP  
ON PRNT EXIT  
Toggles parameter  
ON/OFF  
DIAG I/O  
22 ) WHEEL_HTR=OFF  
PREV NEXT JUMP  
OFF PRNT EXIT  
Exit returns to  
DIAG display & all values  
return to software control  
Figure 11-2: Example of Signal I/O Function  
11.1.4. Internal Electronic Status LED’s  
Several LED’s are located inside the instrument to assist in determining if the analyzer’s CPU, I2C  
buss and relay board, GFC wheel and the sync/demodulator board are functioning properly.  
11.1.4.1. CPU Status Indicator  
DS5, a red LED, that is located on upper portion of the motherboard, just to the right of the CPU  
board, flashes when the CPU is running the main program loop. After power-up, approximately  
30 to 60 seconds, DS5 should flash on and off. If characters are written to the front panel display  
but DS5 does not flash then the program files have become corrupted, contact customer service  
04584 Rev A1  
189  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
because it may be possible to recover operation of the analyzer. If after 30 – 60 seconds neither  
the DS5 is flashing or no characters have been written to the front panel display then the CPU is  
bad and must be replaced.  
Mother Board  
P/N 04069  
CPU Status LED  
Figure 11-3: CPU Status Indicator  
11.1.4.2. Sync Demodulator Status LED’s  
Two LED’s located on the Sync/Demod Board and are there to make it obvious that the GFC  
Wheel is spinning and the synchronization signals are present:  
Table 11-3: Sync/Demod Board Status Failure Indications  
LED  
D1  
Function  
M/R Sensor  
Status  
Fault Status  
LED is stuck  
ON or OFF  
Indicated Failure(s)  
GFC Wheel is not turning  
M/R Sensor on Opto-Pickup Board failed  
Sync/Demod Board failed  
JP 4 Connector/Wiring faulty  
Failed/Faulty +5 VDC Power Supply (PS1)  
D2  
Segment  
Sensor Status  
LED is stuck  
ON or OFF  
GFC Wheel is not turning  
Segment Sensor on Opto-Pickup Board failed  
Sync/Demod Board failed  
JP 4 Connector/Wiring faulty  
Failed/Faulty +5 VDC Power Supply (PS1)  
04584 Rev A1  
190  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
D1 – M/R Sensor Status  
JP4 Connector to Opto-Pickup  
Board  
D2 – Segment Sensor Status  
Figure 11-4: Sync/Demod Board Status LED Locations  
11.1.4.3. Relay Board Status LED’s  
There are eight LED’s located on the Relay Board. The most important of which is D1, which  
indicates the health of the I2C buss. If D1 is blinking the other faults following LED’s can be used  
in conjunction with DIAG menu signal I/O to identify hardware failures of the relays and switches  
on the relay (see Section 6.9.2 and Appendix D).  
Table 11-4: I2C Status LED Failure Indications  
LED  
FUNCTION  
FAULT STATUS  
INDICATED FAILURE(S)  
Failed/Halted CPU  
Faulty Mother Board, Keyboard or Relay  
Board  
Faulty Connectors/Wiring between Mother  
Board, Keyboard or Relay Board  
Failed/Faulty +5 VDC Power Supply (PS1)  
I2C buss  
Health  
(Watchdog  
Continuously ON  
or  
Continuously OFF  
D1  
(Red)  
Circuit)  
DC VOLTAGE TEST  
POINTS  
STATUS LED’s  
RELAY PCA  
PN 04135  
Figure 11-5: Relay Board Status LEDs  
04584 Rev A1  
191  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
Table 11-5: Relay Board Status LED Failure Indications  
SIGNAL I/O PARAMETER  
LED  
FUNCTION  
DIAGNOSTIC TECHNIQUE  
ACTIVATED BY  
VIEW RESULT  
Voltage displayed should change. If not:  
Failed Heater  
WHEEL_TEMP Faulty Temperature Sensor  
D2  
Yellow  
WHEEL  
HEATER  
WHEEL_HEATER  
Failed AC Relay  
Faulty Connectors/Wiring  
Voltage displayed should change. If not:  
Failed Heater  
D3  
Yellow  
BENCH  
HEATER  
BENCH_HEATER  
N/A  
BENCH_TEMP Faulty Temperature Sensor  
Failed AC Relay  
Faulty Connectors/Wiring  
N/A  
D4  
Yellow  
SPARE  
N/A  
N/A  
Sample/Cal Valve should audibly change  
states. If not:  
Failed Valve  
Failed Relay Drive IC on Relay Board  
Failed Relay Board  
Faulty +12 VDC Supply (PS2)  
Faulty Connectors/Wiring  
Zero/Span Valve should audibly change  
states. If not:  
Failed Valve  
Failed Relay Drive IC on Relay Board  
Failed Relay Board  
Faulty +12 VDC Supply (PS2)  
Faulty Connectors/Wiring  
Shutoff Valve should audibly change  
states. If not:  
Failed Valve  
Failed Relay Drive IC on Relay Board  
Failed Relay Board  
Faulty +12 VDC Supply (PS2)  
Faulty Connectors/Wiring  
Voltage displayed should change. If not:  
Failed IR Source  
SAMPLE/CAL  
GAS VALVE  
OPTION  
D5  
Green  
CAL_VALVE  
SPAN_VALVE  
SHUTOFF_VALVE  
IR_SOURCE  
ZERO/SPAN  
GAS VALVE  
OPTION  
D6  
Green  
N/A  
N/A  
D7  
Green  
SHUTOFF  
VALVE OPTION  
Faulty +12 VDC Supply (PS2)  
D8  
Green  
IR SOURCE  
CO2_MEASURE Failed Relay Board  
Failed IR Photo-Detector  
Failed Sync/Demod Board  
Faulty Connectors/Wiring  
11.1.5. Gas Flow Problems  
In general, flow problems can be divided into three categories:  
1. Flow is too high  
2. Flow is greater than zero, but is too low, and/or unstable  
3. Flow is zero (no flow)  
When troubleshooting flow problems, it is a good idea to first confirm that the actual flow and not  
the analyzer’s flow sensor and software are in error, or the flow meter is in error. Use an  
04584 Rev A1  
192  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
independent flow meter to perform a flow check as described in Section 9.3.4. If this test shows  
the flow to be correct, check the pressure sensors as described in Section 11.4.6.5.  
11.1.6. Typical Sample Gas Flow Problems  
11.1.6.1. Flow is Zero  
The unit displays a SAMPLE FLOW warning message on the front panel display or the SAMPLE  
FLOW test function reports a zero or very low flow rate.  
Confirm that the sample pump is operating (turning). If not, use an AC voltmeter to make sure  
that power is being supplied to the pump. If no power is present at the electrical leads of the  
pump.  
1. If AC power is being supplied to the pump, but it is not turning, replace the pump.  
2. If the pump is operating but the unit reports no gas flow, perform a flow check as described in  
Section 9.3.4.  
3. If no independent flow meter is available:  
Disconnect the gas lines from both the sample inlet and the exhaust outlet on the rear  
panel of the instrument.  
Make sure that the unit is in basic SAMPLE Mode.  
Place a finger over an Exhaust outlet on the rear panel of the instrument.  
If gas is flowing through the analyzer, you will feel pulses of air being expelled from the  
Exhaust outlet.  
4. If gas flows through the instrument when it is disconnected from its sources of zero air, span  
gas or sample gas, the flow problem is most likely not internal to the analyzer. Check to make  
sure that:  
All calibrators/generators are turned on and working correctly.  
Gas bottles are not empty or low.  
Valves, regulators, and gas lines are not clogged or dirty.  
11.1.6.2. Low Flow  
1. Check if the pump diaphragm is in good condition. If not, rebuild the pump (see Section  
9.3.2). Check the Spare Parts List for information of pump rebuild kits.  
2. Check for leaks as described in Section 9.3.3. Repair the leaking fitting, line or valve and re-  
check.  
3. Check for the sample filter and the orifice filter for dirt. Replace filters (see Sections 9.3.1 and  
11.5.1 respectively).  
4. Check for partially plugged pneumatic lines, orifices, or valves. Clean or replace them.  
04584 Rev A1  
193  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
5. If an IZS option is installed in the instrument, press CALZ and CALS. If the flow increases  
then suspect a bad sample/cal valve.  
11.1.6.3. High Flow  
The most common cause of high flow is a leak in the sample flow control assembly or between  
there and the pump. If no leaks or loose connections are found in the fittings or the gas line  
between the orifice and the pump, rebuild/clean the sample flow control assembly as described in  
Section 11.5.1.  
11.1.6.4. Displayed Flow = “XXXX”  
This warning means that there is inadequate gas flow. There are four conditions that might cause  
this:  
1. A leak upstream or downstream of the flow sensor  
2. A flow obstruction upstream or downstream of the flow sensor  
3. Bad Flow Sensor Board  
4. Bad pump  
To determine which is the case, view the sample pressure and sample flow functions on the front  
panel. If the sample pressure is reading abnormally low, then the cause is likely a flow obstruction  
upstream of the flow sensor. First, check the sample filter and make sure it is not plugged and  
then systematically check all the other components upstream of the orifice to ensure that they are  
not obstructed.  
If the sample pressure is reading normal but the sample flow is reading low then it is likely that  
the pump diaphragm is worn or there is an obstruction downstream of the flow sensor.  
11.1.6.5. Actual Flow Does Not Match Displayed Flow  
If the actual flow measured does not match the displayed flow, but is within the limits of 720-880  
cc/min, adjust the calibration of the flow measurement as described in Section 6.9.8.  
11.1.6.6. Sample Pump  
The sample pump should start immediately after the front panel power switch is turned ON. With  
the Sample Inlet plugged, the test function PRES should read about 10”-Hg for a pump in good  
condition. Readings above 15”-Hg indicate that the pump needs rebuilding. If the test function  
SAMP FL is greater than 10 cc the instrument’s there is a leak in the pneumatic lines.  
11.1.7. Poor or Stopped Flow of Purge Gas  
If sufficient purge gas is not supplied to the GFC wheel housing, cyclical fluctuations in readings  
at zero or low CO2 concentrations, such as < 100 ppm, may occur. These fluctuations are the  
result of changes in the CO2 concentration of the ambient atmosphere throughout the course of  
the day and night. In isolated areas with relatively few people working nearby the ambient CO2  
concentration will fall during the day and rise during the night as rate of photosynthesis of the  
plants in the surrounding area decreases and increases. In a lab environment with a relatively  
high human occupancy the ambient CO2 concentration will increase during those parts of the day  
when the highest number of workers are present. If the GFC wheel housing is allowed to fill with  
04584 Rev A1  
194  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
ambient air, these natural, diurnal fluctuations might be detected by the instrument and cause its  
in its calculation of the CO2 concentration of the sample gas to drift.  
Another possible symptom of poor or stopped purge gas flow would be the inability to measure  
zero concentrations accurately at the end of a work day on a system that was calibrated at the  
beginning of a workday. Again this could be due to local fluctuations in ambient CO2  
concentration between the time of the day when the calibration was performed and other times  
during the day.  
11.2. Calibration Problems  
11.2.1. Miss-Calibrated  
There are several symptoms that can be caused by the analyzer being miss-calibrated. This  
condition is indicated by out of range Slopes and Offsets as displayed through the test functions  
and is frequently caused by the following:  
1. BAD SPAN GAS. This can cause a large error in the slope and a small error in the offset.  
Delivered from the factory, the MGFC7000E’s slope is within ±15% of nominal. Bad span gas  
will cause the analyzer to be calibrated to the wrong value. If in doubt have the span gas  
checked by and independent lab.  
2. CONTAMINATED ZERO GAS. Excess H2O can cause a positive or negative offset and will  
indirectly affect the slope.  
3. Dilution calibrator not set up correctly or is malfunctioning. This will also cause the slope, but  
not the zero to be incorrect. Again the analyzer is being calibrated to the wrong value.  
4. Too many analyzers on the manifold. This can cause either a slope or offset error because  
ambient gas with its pollutants will dilute the zero or span gas.  
11.2.2. Non-Repeatable Zero and Span  
As stated earlier, leaks both in the MGFC7000E and in the external system are a common source  
of unstable and non-repeatable readings.  
1. Check for leaks in the pneumatic systems as described in Section 9.3.3. Don’t forget to  
consider pneumatic components in the gas delivery system outside the MGFC7000E. Such as:  
A change in zero air source such as ambient air leaking into zero air line, or;  
A change in the span gas concentration due to zero air or ambient air leaking into the  
span gas line.  
2. Once the instrument passes a leak check, do a flow check (see Section 9.3.4) to make sure  
adequate sample is being delivered to the sensor assembly.  
3. A failing IR photo-detector may be at fault. Check the CO2 MEAS and CO2 REF test  
functions via the front panel display to make sure the signal levels are in the normal range  
(See Appendix A) and are quiet.  
04584 Rev A1  
195  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
4. Confirm the sample pressure, wheel temperature, bench temperature, and sample flow  
readings are correct and have steady readings.  
5. Disconnect the exhaust line from the optical bench near the rear of the instrument and plug  
this line into the SAMPLE inlet creating a pneumatic loop. The CO2 concentration (either zero  
or span) now must be constant. If readings become quiet, the problem is in the external  
pneumatics supplies for sample gas, span gas or zero air.  
6. If pressurized span gas is being used with a zero/span valve option, make sure that the  
venting is adequate (See Section 3.1.2 and 5.4)  
7. If it is the zero point that is non-repeatable, and if that non-repeatability seems to only occur  
at a certain time of day, such as when worker occupancy is highest or lowest, make sure the  
flow of purge gas to the GFC wheel housing has not stopped (see Sections 10.2.2 and 11.1.7.  
for more information).  
11.2.3. Inability to Span – No SPAN Key  
1. Confirm that the carbon dioxide span gas source is accurate; this can be done by switching  
between two span-gas tanks. If the CO2 concentration is different, there is a problem with  
one of the tanks.  
2. Check for leaks in the pneumatic systems as described in Section 9.3.3.  
3. Make sure that the expected span gas concentration entered into the instrument during  
calibration is the correct span gas concentration and not too different from expected span  
value. This can be viewed via the RNG Menu (see Section 6.7).  
4. Check to make sure that there is no ambient air or zero air leaking into span gas line.  
11.2.4. Inability to Zero – No ZERO Key  
1. Confirm that there is a good source of zero air. Dilute a tank of span gas with the same  
amount of zero air from two different sources. If the CO2 Concentration of the two  
measurements is different, there is a problem with one of the sources of zero air.  
2. Check for leaks in the pneumatic systems as described in Section 9.3.3.  
3. The Internal zero air scrubber may need maintenance. This device is only present if the  
analyzer has had zero/span valve options 51 or 53 installed this.  
4. Check to make sure that there is no ambient air leaking into zero air line.  
11.3. Other Performance Problems  
Dynamic problems (i.e. problems which only manifest themselves when the analyzer is monitoring  
sample gas) can be the most difficult and time consuming to isolate and resolve. The following  
section provides an itemized list of the most common dynamic problems with recommended  
troubleshooting checks and corrective actions.  
04584 Rev A1  
196  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
11.3.1. Temperature Problems  
Individual control loops are used to maintain the set point of the absorption bench, filter wheel,  
and IR photo-detector temperatures. If any of these temperatures are out of range or are poorly  
controlled, the MGFC7000E will perform poorly.  
11.3.1.1. Box or Sample Temperature  
Box Temperature  
The box temperature sensor is mounted to the motherboard and cannot be disconnected to check  
its resistance. Rather check the BOX TEMP signal using the SIGNAL I/O function under the  
DIAG Menu (see Section 11.1.3). This parameter will vary with ambient temperature, but at  
~30oC (6-7° above room temperature) the signal should be ~1450 mV.  
Sample Temperature  
The Sample Temperature should closely track the bench temperature. If it does not, locate the  
sensor, which is located at the midpoint of the optical bench in a brass fitting. Unplug the  
connector labeled “Sample”, and measure the resistance of the thermistor; at room temperature  
(25°C) it should be ~30K Ohms, at operating temperature, 48°C, it should be ~ 12K Ohms  
11.3.1.2. Bench Temperature  
There are three possible failures that could cause the Bench temperature to be incorrect.  
1. The heater mounted to the bottom of the Absorption bench is electrically shorted or open.  
Check the resistance of the two heater elements by measuring between pin 2 and 4 (~76  
Ohms), and pin 3 and 4 (~330 Ohms), of the white five-pin connector just below the sample  
temperature sensor on the Bench (pin 1 is the pointed end).  
2. Assuming that the I2C buss is working and that there is no other failure with the relay board,  
the solid-state relay (K2) on the relay board may have failed. Using the BENCH_HEATER  
parameter under the signal I/O function, as described above, turn on and off K2 (D3 on the  
relay board should illuminate as the heater is turned on). Check the AC voltage present  
between pin 2 and 4, for a 100 or 115 VAC model, and pins 3 and 4, for a 220-240 VAC  
model.  
WARNING:  
HAZARDOUS VOLTAGES ARE PRESENT DURING THIS TEST  
3. If the relay has failed there should be no change in the voltage across pins 2 and 4 or 3 and 4.  
Note, K2 is in a socket for easy replacement.  
4. If K2 checks out OK, the thermistor temperature sensor located on the optical bench near the  
front of the instrument could be at fault. Unplug the connector labeled “Bench”, and measure  
the resistance of the thermistor. At room temperature it should have approximately 30K  
Ohms resistance, near the 48oC set point it should have ~11K ohms.  
04584 Rev A1  
197  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
11.3.1.3. GFC Wheel Temperature  
Like the bench heater above there are three possible causes for the GFC wheel temperature to  
have failed.  
1. The wheel heater has failed. Check the resistance between pins 1 and 4 on the white five-pin  
connector just below the sample temperature sensor on the bench (pin 1 is the pointed end).  
It should be approximately 275 ohms.  
2. Assuming that the I2C buss is working and that there is no other failure with the relay board,  
the solid-state relay (K1) on the relay board may have failed. Using the WHEEL_HEATER  
parameter under the signal I/O function, as described above, turn on and off K1 (D2 on the  
relay board should illuminate as the heater is turned on). Check the AC voltage present  
between pin 1 and 4.  
WARNING:  
HAZARDOUS VOLTAGES ARE PRESENT DURING THIS TEST  
3. If the relay has failed there should be no change in the voltage across pins 1 and 4. Note, K1  
is socketed for easy replacement.  
4. If K1 checks out OK, the thermistor temperature sensor located at the front of the filter wheel  
assembly may have failed. Unplug the connector labeled “Wheel”, and measure the resistance  
of the thermistor. The resistance near the 68oC set point is ~5.7k ohms.  
11.3.1.4. IR Photo-Detector TEC Temperature  
If the PHT DRIVE test parameter described above in Table 11-2 is out of range there are two  
four possible causes of failure.  
1. The screws retaining the IR photo detector to the absorption bench have become loose.  
Carefully tighten the screws, hand-tight and note whether, after the analyzer has come up to  
operating temperature, whether the PHT DRIVE voltage has returned to an acceptable level.  
2. The two large transistor-type devices mounted to the side of the Absorption Bench have come  
loose from the bench. Tighten the retaining screws and note whether there is an improvement  
in the PHT DRIVE voltage.  
3. The photo-detector has failed. Contact the factory for instructions.  
4. The sync demodulator circuit board has failed. Contact the factor for instructions.  
11.3.2. Excessive Noise  
Noise is continuously monitored in the TEST functions as the STABIL reading and only becomes  
meaningful after sampling a constant gas concentration for at least 10 minutes. Compare the  
current STABIL reading with that recorded at the time of manufacture (included in the  
04584 Rev A1  
198  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
MGFC7000E Final Test And Validation Data Sheet-p/n 04271 shipped with the unit from Teledyne  
Instruments).  
1. The most common cause of excessive noise is leaks. Leak check and flow check the  
instrument described in Section 9.3.  
2. Detector failure – caused by failure of the hermetic seal or over-temperature due to poor heat  
sinking of the detector can to the optical bench. In addition to increased noise due to poor  
signal-to-noise ratio, another indicator of detector failure is a drop in the signal levels of the  
CO2 MEASURE signal and CO2 REFERENCE signal.  
3. Sync/Demod Board failure. There are many delicate, high impedance parts on this board.  
Check the CO2 MEAS and CO2 REF Test Functions via the Front Panel Display.  
4. The detector cooler control circuit can fail for reasons similar to the detector itself failing.  
Symptoms would be a change in MR RATIO Test Function when zero air is being sampled.  
Also check the SIGNAL I/O parameter PHT DRIVE. After warm-up, and at 25oC ambient, if  
PHT DRIVE < 2500 mV, the cooler is working properly. If PHT DRIVE is > 2500 mV there is a  
malfunction.  
5. The +5 and ±15 VDC voltages in the MGFC7000E are provided by switching power supplies.  
Switch mode supplies create DC outputs by switching the input AC waveform at high  
frequencies. As the components in the switcher age and degrade, the main problem observed  
is increased noise on the DC outputs. If a noisy switcher power supply is suspected, attach an  
oscilloscope to the DC output test points located on the top right hand edge of the Relay  
board. Look for short period spikes > 100 mV p-p on the DC output.  
11.4. Subsystem Checkout  
The preceding sections of this manual discussed a variety of methods for identifying possible  
sources of failures or performance problems within the analyzer. In most cases this included a list  
of possible causes. This section describes how to determine individually determine if a certain  
component or subsystem is actually the cause of the problem being investigated.  
11.4.1. AC Mains Configuration  
The analyzer is correctly configured for the AC mains voltage in use if:  
1. The Sample Pump is running.  
2. The GFC wheel motor is spinning (a slight vibration should be apparent to the touch).  
3. If incorrect power is suspected, check that the correct voltage and frequency is present at the  
line input on the rear panel.  
If the unit is set for 230 VAC and is plugged into 115VAC, or 100VAC the sample pump  
will not start, and the heaters will not come up to temperature.  
If the unit is set for 115 or 100 VAC and is plugged into a 230 VAC circuit, the circuit  
breaker built into the ON/OFF Switch on the Front Panel will trip to the OFF position  
immediately after power is switched on.  
04584 Rev A1  
199  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
11.4.2. DC Power Supply  
If you have determined that the analyzer’s AC mains power is working, but the unit is still not  
operating properly, there may be a problem with one of the instrument’s switching power  
supplies. The supplies can have two faults, namely no DC output, and noisy output.  
To assist tracing DC Power Supply problems, the wiring used to connect the various printed circuit  
assemblies and DC Powered components and the associated test points on the relay board follow  
a standard color-coding scheme as defined in Table.  
Table 11-6: DC Power Test Point and Wiring Color Codes  
NAME  
Dgnd  
+5V  
TEST POINT#  
TP AND WIRE COLOR  
1
2
3
4
5
6
7
Black  
Red  
Agnd  
+15V  
-15V  
Green  
Blue  
Yellow  
Purple  
Orange  
+12V  
+12R  
A voltmeter should be used to verify that the DC voltages are correct per the values in Table  
below, and an oscilloscope, in AC mode, with band limiting turned on, can be used to evaluate if  
the supplies are producing excessive noise (> 100 mV p-p).  
Table 11-7: DC Power Supply Acceptable Levels  
CHECK RELAY BOARD TEST POINTS  
POWER  
VOLTAG  
E
FROM TEST  
POINT  
SUPPLY  
ASSY  
MIN V  
MAX V  
TO TEST POINT  
NAME  
Dgnd  
#
NAME  
+5  
#
2
PS1  
PS1  
PS1  
PS1  
PS1  
PS2  
PS2  
+5  
+15  
1
3
3
3
1
6
6
4.8  
5.25  
16V  
Agnd  
+15  
4
13.5  
-15  
Agnd  
-15V  
Dgnd  
Chassis  
+12V  
Dgnd  
5
-14V  
-0.05  
-0.05  
11.75  
-0.05  
-16V  
0.05  
0.05  
12.5  
0.05  
Agnd  
Chassis  
+12  
Agnd  
1
Dgnd  
N/A  
7
+12V Ret  
+12V Ret  
Dgnd  
1
11.4.3. I2C Bus  
Operation of the I2C buss can be verified by observing the behavior of D1 on the Relay Board in  
conjunction with the performance of the front panel display. Assuming that the DC power  
supplies are operating properly and the wiring from the Motherboard to the Keyboard, and the  
wiring from the keyboard to the Relay board, is intact, the I2C buss is operating properly if:  
04584 Rev A1  
200  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
D1 on the relay board is flashing, or  
D1 is not flashing but pressing a key on the front panel results in a change to the display.  
11.4.4. Keyboard/Display Interface  
The front panel keyboard, display and Keyboard Display Interface PCA (03975 or 04258) can be  
verified by observing the operation of the display when power is applied to the instrument and  
when a key is pressed on the front panel. Assuming that there are no wiring problems and that  
the DC power supplies are operating properly:  
1. The vacuum fluorescent display is good if on power-up a “-“ character is visible on the upper  
left hand corner of the display.  
2. The CPU Status LED, DS5, is flashing, see Section 11.1.4.1.  
3. If there is a “-“ character on the display at power-up and D1 on the relay board is flashing  
then the keyboard/display interface PCA is bad.  
4. If the analyzer starts operation with a normal display but pressing a key on the front panel  
does not change the display, then there are three possible problems:  
One or more of the keys is bad,  
The interrupt signal between the Keyboard Display interface and the motherboard is  
broken, or  
The Keyboard Display Interface PCA is bad.  
11.4.5. Relay Board  
The relay board PCA (04135) can be most easily checked by observing the condition of the its  
status LEDs on the relay board, as described in Section 11.1.4.3, and the associated output when  
toggled on and off through signal I/O function in the diagnostic menu, see Section 11.1.3.  
1. If the front panel display responds to key presses and D1 on the relay board is NOT flashing  
then either the wiring between the Keyboard and the relay board is bad, or the relay board is  
bad.  
2. If D1 on the relay board is flashing and the status indicator for the output in question (heater  
power, valve drive, etc.) toggles properly using the signal I/O function, then the associated  
control device on the relay board is bad. Several of the control devices are in sockets and can  
be easily replaced. The table below lists the control device associated with a particular  
function:  
Table 11-8: Relay Board Control Devices  
CONTROL  
DEVICE  
K1  
FUNCTION  
IN SOCKET  
Wheel Heater  
Bench Heater  
Spare AC Control  
IZS Valves  
Yes  
Yes  
Yes  
Yes  
K2  
K3  
U4  
04584 Rev A1  
201  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
IR Source Drive  
U5  
No  
The IR source drive output can be verified by measuring the voltage at J16 with the IR source  
disconnected. It should be 11.5± 0.5 VDC.  
11.4.6. Sensor Assembly  
11.4.6.1. Sync/Demodulator Assembly  
To verify that the Sync/Demodulator Assembly is working follow the procedure below:  
1. Verify that D1 and D2 are flashing.  
If not check the opto pickup assembly, Section 11.4.6.2 and the GFC wheel drive, Section  
11.4.6.3  
If the wheel drive and opto pickup are working properly then verify that there is 2.4 ±0.1  
VAC and 2.5 ±0.15 VDC between digital ground and TP 5 on the sync demod board. If not  
then check the wiring between the sync/demod and opto pickup assembly (see  
interconnect drawing 04216). If good then the sync/demod board is bad.  
2. Verify that the IR source is operating, Section 11.4.6.4.  
3. With the analyzer connected to zero air, measure between TP11 (measure) and analog  
ground, and TP12 (reference) and analog ground.  
If they are similar to values recorded on the Factory Data sheet then there is likely a  
problem with the wiring or the A/D converter.  
If they are not then either the sync demodulator board or the IR-photodetector are bad.  
See also section 11.3.1.4 for problems with the IR-Photodetector TEC drive.  
11.4.6.2. Opto Pickup Assembly  
Operation of the opto pickup PCA (04088) can be verified with a voltmeter. Measure the AC and  
DC voltage between digital ground on the Relay board, or Keyboard and TP1 and TP2 on the sync  
pickup PCA. For a working board, with the GFC motor spinning, they should read 2.4 ±0.1 VAC  
and 2.5 ±0.15 VDC.  
Further confirmation that the pickups and motor are operating properly can be obtained by  
measuring the frequency at TP1 and TP2 using a frequency counter, a digital volt meter with a  
frequency counter, or an oscilloscope per the table below.  
Table 11-9: Opto Pickup Board Nominal Output Frequencies  
NOMINAL MEASURED FREQUENCY  
AC MAINS FREQ.  
50 Hz  
TP1  
25  
TP2  
300  
360  
60 Hz  
30  
04584 Rev A1  
202  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
11.4.6.3. GFC Wheel Drive  
If the D1 and D2 on the sync demodulator board are not flashing then:  
1. Check for power to the motor by measuring between pins 1 and 3 on the connector feeding  
the motor. For instruments configured for 120 or 220-240VAC there should be approximately  
88 VAC for instruments configured for 100VAC, it should be the voltage of the AC mains,  
approximately 100VAC.  
2. Verify that the frequency select jumper, JP4, is properly set on the Relay Board. For 50 Hz  
operation it should be installed. For 60 Hz operation may either be missing or installed in a  
vertical orientation.  
3. If there is power to the motor and the frequency select jumper is properly set then the motor  
is likely bad. See Section 11.5.2 for instructions on removing and replacing the GFC assembly  
that the motor is bolted to.  
11.4.6.4. IR Source  
The IR source can be checked using the following procedure:  
1. Disconnect the source and check its resistance when cold. When new, the source should have  
a cold resistance of more than 1.5 Ohms but less than 3.5 Ohms. If not, then the source is  
bad.  
2. With the source disconnected, energize the analyzer and wait for it to start operating.  
Measure the drive Voltage between pins 1 and 2 on the jack that the source is normally  
connected to, it should be 11.5 ± 0.25 VDC. If not, then there is a problem with either the  
wiring, the Relay Board, or the +12V power supply.  
3. If the drive voltage is correct in step 2, then remove the source from the heat sink assembly  
(2 screws on top) and connect to its mating connector. Observe the light being emitted from  
the source. It should be centered at the bottom of the U-shaped element. If there is either  
no emission or a badly centered emission then the source is bad.  
11.4.6.5. Pressure/Flow Sensor Assembly  
The pressure/flow sensor PCA, located on the top of the absorption bench, can be checked with a  
Voltmeter using the following procedure which, assumes that the wiring is intact, and that the  
Motherboard and the power supplies are operating properly:  
1. For Pressure related problems:  
Measure the voltage across C1 it should be 5 ± 0.25 VDC. If not then the board is bad.  
Measure the voltage across TP4 and TP1. With the sample pump disabled it should be  
4500 mV ±250 mV. With the pump energized it should be approximately 200 mV less. If  
not, then S1, the pressure transducer is bad, the board is bad, or there is a pneumatic  
failure preventing the pressure transducer from sensing the absorption cell pressure  
properly.  
2. For flow related problems:  
Measure the voltage across TP2 and TP1 it should be 10 ±0.25 VDC. If not then the board  
is bad.  
04584 Rev A1  
203  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
Measure the voltage across TP3 and TP1. With proper flow (800 sccm at the sample inlet)  
this should be approximately 4.5V (this voltage will vary with altitude). With flow stopped  
(sample inlet blocked) the voltage should be approximately 1V. If the voltage is incorrect,  
the flow sensor is bad, the board is bad or there is a leak upstream of the sensor.  
11.4.7. Motherboard  
11.4.7.1. A/D Functions  
The simplest method to check the operation of the A-to-D converter on the motherboard is to use  
the Signal I/O function under the DIAG Menu to check the two A/D reference Voltages and input  
signals that can be easily measured with a voltmeter.  
1. Use the Signal I/O function (see Section 11.1.3 and Appendix A) to view the value of  
REF_4096_MV and REF_GND. If both are within 3 mV of nominal (4096 and 0), and are  
stable, ±0.5 mV then the basic A/D is functioning properly. If not then the Motherboard is  
Bad  
2. Choose a parameter in the Signal I/O function such as SAMPLE_PRESSURE,  
SAMPLE_FLOW, CO2_MEASURE or CO2_REFERENCE. Compare these Voltages at their  
origin (see Interconnect drawing 04215 and Interconnect list 04216) with the voltage  
displayed through the Signal I/O function. If the wiring is intact but there is a large difference  
between the measured and displayed voltage (±10 mV) then the Motherboard is bad.  
11.4.7.2. Analog Outputs: Voltage  
To verify that the analog outputs are working properly, connect a voltmeter to the output in  
question and perform an analog output step test as described in Section 6.9.3.  
For each of the steps, taking into account any offset that may have been programmed into  
channel (see Section 6.9.4), the output should be within 1% of the nominal value listed in the  
table below except for the 0% step, which should be within 2 to 3 mV. If one or more of the steps  
fails to be within this range then it is likely that there has been a failure of the either or both of  
the DACs and their associated circuitry on the Motherboard.  
Table 11-10: Analog Output Test Function - Nominal Values Voltage Outputs  
FULL SCALE OUTPUT VOLTAGE  
100MV  
1V  
5V  
10V  
STEP  
%
0
NOMINAL OUTPUT VOLTAGE  
1
2
3
4
5
6
0
0
0
1
2
3
4
5
0
2
20  
40  
60  
80  
100  
20 mV  
40 mV  
60 mV  
80 mV  
100 mV  
0.2  
0.4  
0.6  
0.8  
1.0  
4
6
8
10  
11.4.7.3. Analog Outputs: Current Loop  
To verify that the analog outputs with the optional current mode output are working properly,  
connect a 250 ohm resistor across the outputs and use a voltmeter to measure the output as  
04584 Rev A1  
204  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
described in Section 6.9.4.2 and then perform an Analog Output Step Test as described in Section  
6.9.3.  
For each step the output should be within 1% of the nominal value listed in the table below.  
Table 11-11: Analog Output Test Function - Nominal Values Current Outputs  
OUTPUT RANGE  
2 -20  
4 -20  
NOMINAL OUTPUT VALUES  
STEP  
%
0
CURRENT  
2 mA  
5.6  
V(250 OHMS)  
CURRENT  
4
V(250 OHMS)  
1
2
3
4
5
6
0.5V  
1.4  
2.3  
3.2  
4.1  
5
1
20  
40  
60  
80  
100  
7.2  
1.8  
2.6  
3.4  
4.2  
5
9.2  
10.4  
13.6  
16.8  
20  
12.8  
16.4  
20  
11.4.7.4. Status Outputs  
The procedure below can be used to test the Status outputs:  
Connect a jumper between the “D“ pin and the “V” pin on the status output connector.  
Connect a 1000 ohm resistor between the “+” pin and the pin for the status output that is being  
tested.  
Connect a voltmeter between the “V” pin and the pin of the output being tested (see table  
below).  
Under the DIAGÆ SIGNAL I/O menu (see Section 11.1.3), scroll through the inputs and outputs  
until you get to the output in question. Alternately turn on and off the output noting the voltage  
on the Voltmeter, it should vary between 0 volts for ON and 5 volts for OFF.  
Table 11-12: Status Outputs Check  
PIN (LEFT TO RIGHT)  
STATUS  
SYSTEM OK  
CONC VALID  
HIGH RANGE  
ZERO CAL  
SPAN CAL  
DIAG MODE  
ALRM1  
1
2
3
4
5
6
7
8
SALRM2  
11.4.7.5. Control Inputs – Remote Zero, Span  
The control input bits can be tested by the following procedure:  
04584 Rev A1  
205  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
Connect a jumper from the +5 pin on the Status connector to the x5V on the Control In  
connector.  
Connect a second jumper from the ‘-‘ pin on the Status connector to the A pin on the Control In  
connector. The instrument should switch from SAMPLE mode to ZERO CAL R mode.  
Connect a second jumper from the ‘-‘ pin on the Status connector to the B pin on the Control In  
connector. The instrument should switch from SAMPLE mode to SPAN CAL R mode.  
In each case, the MGFC7000E should return to SAMPLE mode when the jumper is removed.  
11.4.8. CPU  
There are two major types of failures associated with the CPU board: complete failure and a  
failure associated with the Disk On Chip on the CPU board. If either of these failures occur,  
contact the factory.  
1. For complete failures, assuming that the power supplies are operating properly and the wiring  
is intact, the CPU is bad if on powering the instrument:  
The vacuum fluorescent display shows a dash in the upper left hand corner.  
The CPU Status LED, DS5, is not flashing. (see Section 11.1.4.1.)  
There is no activity from the primary RS-232 port (COM-A) on the rear panel even if “?  
<ret>” is pressed.  
In some rare circumstances this failure may be caused by a bad IC on the Motherboard,  
specifically U57 the large, 44 pin device on the lower right hand side of the board. If this  
is true, removing U57 from its socket will allow the instrument to startup but the  
measurements will be incorrect.  
2. If the analyzer stops part way through initialization (there are words on the vacuum  
fluorescent display) then it is likely that the DOC has been corrupted.  
11.4.9. RS-232 Communications  
11.4.9.1. General RS-232 Troubleshooting  
Teledyne Instruments analyzers use the RS-232 communications protocol to allow the instrument  
to be connected to a variety of computer-based equipment. RS-232 has been used for many  
years and as equipment has become more advanced, connections between various types of  
hardware have become increasingly difficult. Generally, every manufacturer observes the signal  
and timing requirements of the protocol very carefully.  
04584 Rev A1  
206  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
Problems with RS-232 connections usually center around 4 general areas:  
1. Incorrect cabling and connectors. See Table 6-15 for connector and pin-out information.  
2. The BAUD rate and protocol are incorrectly configured. See Section 6.10.7.  
3. If a modem is being used, additional configuration and wiring rules must be observed. See  
Section 6.13.2.6  
4. Incorrect setting of the DTE – DCE Switch is set correctly See Section 6.10.5  
5. Verify that cable (03596) that connects the serial COM ports of the CPU to J12 of the  
Motherboard is properly seated  
11.4.9.2. Troubleshooting Analyzer/Modem or Terminal Operation  
These are the general steps for troubleshooting problems with a modem connected to a Teledyne  
Instruments analyzer.  
1. Check Cables for proper connection to the modem, terminal or computer.  
2. Check to make sure the DTE-DCE is in the correct position as described in Section 6.10.5.  
3. Check to make sure the set up command is correct (See Section 6.13.2.7)  
4. Verify that the Ready to Send (RTS) signal is at logic high. The MGFC7000E sets pin 7 (RTS) to  
greater than 3 volts to enable modem transmission.  
5. Make sure the BAUD rate, word length, and stop bit settings between modem and analyzer  
match, see Section 6.10.7.  
6. Use the RS-232 test function to send “w” characters to the modem, terminal or computer; See  
Section 6.10.8.  
7. Get your terminal, modem or computer to transmit data to the analyzer (holding down the  
space bar is one way); the green LED should flicker as the instrument is receiving data.  
8. Make sure that the communications software or Terminal emulation software is functioning  
properly.  
Further help with serial communications is available in a separate manual “RS-232 Programming  
Notes” Teledyne Instruments part number 013500000.  
11.5. Repair Procedures  
This section contains procedures that might need to be performed on rare occasions when a major  
component of the analyzer requires repair or replacement.  
11.5.1. Repairing Sample Flow Control Assembly  
The Critical Flow Orifice is housed in the Flow Control Assembly (Teledyne Instruments part  
number: 001760400) located on the top of the optical bench. A sintered filter protects the jewel  
orifice so it is unusual for the orifice to need replacing, but if does, or the filter needs replacement  
04584 Rev A1  
207  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
please use the following procedure (see the Spare Parts list in Appendix B for part numbers and  
kits):  
1. Turn off power to the analyzer.  
2. Locate the assembly attached to the sample pump, see Figure 3–11.  
3. Disconnect the pneumatic connection from the flow assembly and the assembly from the  
pump.  
4. Remove the fitting and the components as shown in the exploded view in Figure 11.6.  
5. Replace the o-rings (p/n:OR_01) and the sintered filter (p/n:FL_01).  
6. If replacing the critical flow orifice itself (p/n:00094100), make sure that the side with the  
colored window (usually red) is facing upstream to the flow gas flow.  
7. Re-assemble in reverse order.  
8. After reconnecting the power and pneumatic lines, flow check the instrument as described in  
the Section 9.3.4.  
Pneumatic Connector, Male 1/8”  
(P/N FT_70  
Spring  
(P/N HW_20)  
Sintered Filter  
(P/N FL_01)  
Critical Flow Orifice  
(P/N 00094100)  
O-Ring  
(P/N OR_01)  
Purge Housing  
(P/N 000850000)  
Figure 11-6: Critical Flow Restrictor Assembly Disassembly  
04584 Rev A1  
208  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
11.5.2. Removing/Replacing the GFC Wheel  
When Removing or replacing the GFC Wheel it is important to perform the disassembly in the  
following order to avoid damaging the components:  
1. Turn off the analyzer.  
2. Remove the top cover as described in “Getting Started” Section 3.1.  
3. Open the Instrument’s hinged front panel.  
4. Locate the GFC wheel/motor assembly (see Figure 3-11).  
5. unplug the following electronic components:  
The GFC wheel housing temperature sensor;  
GFC wheel heater  
GFC wheel motor power supply  
IR source  
6. Unscrew the purge gas line hex nut and remove the 1/8 inch FEP purge gas line.  
10  
7
8
9
Figure 11-7: Opening the GFC Wheel Housing  
7. Remove the two (2) screws holding the opto-pickup printed circuit assembly to the GFC wheel  
housing.  
8. Carefully remove the opto-pickup printed circuit assembly.  
9. Remove the four (4) screws holding the GFC wheel motor/heat sink assembly to the GFC  
wheel housing.  
10.Carefully remove the GFC Wheel motor/heat sink assembly from the GFC wheel housing.  
04584 Rev A1  
209  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
11  
12  
Figure 11-8: Removing the GFC Wheel  
11.Remove the ONE (1) screw fastening the GFC wheel/mask assembly to the gfc motor hub.  
12.Remove the GFC wheel/mask assembly.  
13.Follow the previous steps in reverse order to put the GFC wheel/motor assembly back  
together.  
11.5.3. Disk-On-Chip Replacement Procedure  
Replacing the Disk-on-Chip, may be necessary in certain rare circumstances or to load new  
instrument software. This will cause all of the instrument configuration parameters and iDAS data  
to be lost. However a backup copy of the operating parameters are stored in a second non-  
volatile memory and will be loaded into the new the Disk-on-Chip on power-up. To change the  
Disk-on-Chip follow this procedure.  
1. Turn off power to the instrument.  
2. Fold down the rear panel by loosening the thumbscrews on each side  
3. Locate the Disk-on-Chip in the rightmost socket near the right hand side of the CPU assembly.  
Remove the IC by gently prying it up from the socket.  
4. Reinstall the new Disk-on-Chip, making sure the notch in the end of the chip is facing upward.  
5. Close the rear panel and turn on power to the machine.  
04584 Rev A1  
210  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
TROUBLESHOOTING & REPAIR PROCEDURES  
User Notes  
04584 Rev A1  
211  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
A Primer on Electro-Static Discharge  
12. A PRIMER ON ELECTRO-STATIC DISCHARGE  
Teledyne Instruments considers the prevention of damage caused by the discharge of static  
electricity to be extremely important part of making sure that your analyzer continues to provide  
reliable service for a long time. This section describes how static electricity occurs, why it is so  
dangerous to electronic components and assemblies as well as how to prevent that damage from  
occurring.  
12.1. How Static Charges are Created  
Modern electronic devices such as the types used in the various electronic assemblies of your  
analyzer, are very small, require very little power and operate very quickly. Unfortunately the  
same characteristics that allow them to do these things also makes them very susceptible to  
damage from the discharge of static electricity. Controlling electrostatic discharge begins with  
understanding how electro-static charges occur in the first place.  
Static electricity is the result of something called triboelectric charging which happens whenever  
the atoms of the surface layers of two materials rub against each other. As the atoms of the two  
surfaces move together and separate, some electrons from one surface are retained by the other.  
Materials  
Makes  
Contact  
Materials  
Separate  
+
+
+
+
+
+
+
+
PROTONS = 3  
ELECTRONS = 2  
PROTONS = 3  
ELECTRONS = 4  
PROTONS = 3  
ELECTRONS = 3  
PROTONS = 3  
ELECTRONS = 3  
NET CHARGE = -1  
NET CHARGE = +1  
NET CHARGE = 0  
NET CHARGE = 0  
Figure 12-1: Triboelectric Charging  
If one of the surfaces is a poor conductor or even a good conductor that is not grounded, the  
resulting positive or negative charge can not bleed off and becomes trapped in place, or static.  
The most common example of triboelectric charging happens when someone wearing leather or  
rubber soled shoes walks across a nylon carpet or linoleum tiled floor. With each step electrons  
change places and the resulting electro-static charge builds up, quickly reaching significant levels.  
Pushing an epoxy printed circuit board across a workbench, using a plastic handled screwdriver or  
even the constant jostling of Styrofoam pellets during shipment can also build hefty static charges  
04584 Rev A1  
213  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
A Primer on Electro-Static Discharge  
Table 12-1: Static Generation Voltages for Typical Activities  
65-90%  
RH  
10-25%  
RH  
MEANS OF GENERATION  
Walking across nylon carpet  
Walking across vinyl tile  
Worker at bench  
1,500V  
250V  
35,000V  
12,000V  
6,000V  
100V  
Poly bag picked up from  
bench  
1,200V  
1,500V  
20,000V  
18,000V  
Moving around in a chair  
padded with urethane foam  
12.2. How Electro-Static Charges Cause Damage  
Damage to components occurs when these static charges come in contact with an electronic  
device. Current flows as the charge moves along the conductive circuitry of the device and the  
typically very high voltage levels of the charge overheat the delicate traces of the integrated  
circuits, melting them or even vaporizing parts of them. When examined by microscope the  
damage caused by electro-static discharge looks a lot like tiny bomb craters littered across the  
landscape of the component’s circuitry.  
A quick comparison of the values in Table 12-1 with the those shown in the Table 12-2, listing  
device susceptibility levels, shows why Semiconductor Reliability News estimates that  
approximately 60% of device failures are the result of damage due to electro-static discharge.  
Table 12-2: Sensitivity of Electronic Devices to Damage by ESD  
DAMAGE SUSCEPTIBILITY  
VOLTAGE RANGE  
DEVICE  
DAMAGE BEGINS  
OCCURRING AT  
CATASTROPHIC  
DAMAGE AT  
MOSFET  
VMOS  
10  
100  
30  
1800  
100  
NMOS  
GaAsFET  
EPROM  
JFET  
60  
60  
2000  
100  
100  
140  
150  
190  
200  
7000  
500  
SAW  
Op-AMP  
CMOS  
2500  
3000  
2500  
3000  
Schottky Diodes 300  
Film Resistors  
300  
This Film  
Resistors  
300  
7000  
ECL  
500  
500  
500  
500  
SCR  
1000  
2500  
Schottky TTL  
04584 Rev A1  
214  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
A Primer on Electro-Static Discharge  
Potentially damaging electro-static discharges can occur:  
Any time a charged surface (including the human body) discharges to a device. Even  
simple contact of a finger to the leads of an sensitive device or assembly can allow enough  
discharge to cause damage. A similar discharge can occur from a charged conductive  
object, such as a metallic tool or fixture.  
When static charges accumulated on a sensitive device discharges from the device to  
another surface such as packaging materials, work surfaces, machine surfaces or other  
device. In some cases, charged device discharges can be the most destructive.  
A typical example of this is the simple act of installing an electronic assembly into the  
connector or wiring harness of the equipment in which it is to function. If the assembly is  
carrying a static charge, as it is connected to ground a discharge will occur.  
Whenever a sensitive device is moved into the field of an existing electro-static field, a  
charge may be induced on the device in effect discharging the field onto the device. If the  
device is then momentarily grounded while within the electrostatic field or removed from  
the region of the electrostatic field and grounded somewhere else, a second discharge will  
occur as the charge is transferred from the device to ground.  
12.3. Common Myths About ESD Damage  
I didn’t feel a shock so there was no electro-static discharge: The human nervous  
system isn’t able to feel a static discharge of less than 3500 volts. Most devices are  
damaged by discharge levels much lower than that.  
I didn’t touch it so there was no electro-static discharge: Electro Static charges are  
fields whose lines of force can extend several inches or sometimes even feet away from the  
surface bearing the charge.  
It still works so there was no damage: Sometimes the damaged caused by electro-  
static discharge can completely sever a circuit trace causing the device to fail immediately.  
More likely, the trace will be only partially occluded by the damage causing degraded  
performance of the device or worse, weakening the trace. This weakened circuit may  
seem to function fine for a short time, but even the very low voltage and current levels of  
the device’s normal operating levels will eat away at the defect over time causing the  
device to fail well before its designed lifetime is reached.  
These latent failures are often the most costly since the failure of the equipment in which  
the damaged device is installed causes down time, lost data, lost productivity, as well as  
possible failure and damage to other pieces of equipment or property.  
Static Charges can’t build up on a conductive surface: There are two errors in this  
statement.  
Conductive devices can build static charges if they are not grounded. The charge will be  
equalized across the entire device, but without access to earth ground, they are still  
trapped and can still build to high enough levels to cause damage when they are  
discharged.  
A charge can be induced onto the conductive surface and/or discharge triggered in the  
presence of a charged field such as a large static charge clinging to the surface of a nylon  
jacket of someone walking up to a workbench.  
04584 Rev A1  
215  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
A Primer on Electro-Static Discharge  
As long as my analyzer is properly installed it is safe from damage caused by  
static discharges: It is true that when properly installed the chassis ground of your  
analyzer is tied to earth ground and its electronic components are prevented from building  
static electric charges themselves. This does not, however, prevent discharges from static  
fields built up on other things, like you and your clothing, from discharging through the  
instrument and damaging it.  
12.4. Basic Principles of Static Control  
It is impossible to stop the creation of instantaneous static electric charges. It is not, however  
difficult to prevent those charges from building to dangerous levels or prevent damage due to  
electro-static discharge from occurring.  
12.4.1. General Rules  
Only handle or work on all electronic assemblies at a properly set up ESD station.  
Setting up an ESD safe work station need not be complicated. A protective mat properly tied to  
ground and a wrist strap are all that is needed to create a basic anti-ESD workstation (see figure  
12-2).  
Protective  
Wrist Strap  
Mat  
Ground Point  
Figure 12-2: Basic anti-ESD Work Station  
For technicians that work in the field, special lightweight and portable anti-ESD kits are available  
from most suppliers of ESD protection gear. These include everything needed to create a  
temporary anti-ESD work area anywhere.  
Always wear an Anti-ESD wrist strap when working on the electronic assemblies  
of your analyzer. An anti-ESD wrist strap keeps the person wearing it at or near the  
same potential as other grounded objects in the work area and allows static charges to  
dissipate before they can build to dangerous levels. Anti-ESD wrist straps terminated with  
04584 Rev A1  
216  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
A Primer on Electro-Static Discharge  
alligator clips are available for use in work areas where there is no available grounded  
plug.  
Also, anti-ESD wrist straps include a current limiting resistor (usually around one meg-  
ohm) that protects you should you accidentally short yourself to the instrument’s power  
supply.  
Simply touching a grounded piece of metal is insufficient. While this may  
temporarily bleed off static charges present at the time, once you stop touching the  
grounded metal new static charges will immediately begin to re-build. In some conditions  
a charge large enough to damage a component can rebuild in just a few seconds.  
Always store sensitive components and assemblies in anti-ESD storage bags or  
bins: Even when you are not working on them, store all devices and assemblies in a  
closed anti-Static bag or bin. This will prevent induced charges from building up on the  
device or assembly and nearby static fields from discharging through the it.  
Use metallic anti-ESD bags for storing and shipping ESD sensitive components  
and assemblies rather than pink-poly bags. The famous, pink-poly bags are made of  
a plastic that is impregnated with a liquid (similar to liquid laundry detergent) which very  
slowly sweats onto the surface of the plastic creating a slightly conductive layer over the  
surface of the bag.  
While this layer may equalizes any charges that occur across the whole bag, it does not  
prevent the build up of static charges. If laying on a conductive grounded surface, these  
bags will allow charges to bleed away but the very charges that build up on the surface of  
the bag itself can be transferred through the bag by induction onto the circuits of your ESD  
sensitive device. Also, the liquid impregnating the plastic is eventually used up after which  
the bag is as useless for preventing damage from ESD as any ordinary plastic bag.  
Anti-Static bags made of plastic impregnated with metal (usually silvery in color) provide  
all of the charge equalizing abilities of the pink-poly bags but also, when properly sealed,  
create a Faraday cage that completely isolates the contents from discharges and the  
inductive transfer of static charges.  
Storage bins made of plastic impregnated with carbon (usually black in color) are also  
excellent at dissipating static charges and isolating their contents from field effects and  
discharges.  
Never use ordinary plastic adhesive tape near an ESD sensitive device or to close  
an anti-ESD bag. The act of pulling a piece of standard plastic adhesive tape, such as  
Scotch® tape, from its roll will generate a static charge of several thousand or even tens of  
thousands of volts on the tape itself and an associated field effect that can discharge  
through or be induced upon items up to a foot away.  
04584 Rev A1  
217  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
A Primer on Electro-Static Discharge  
12.4.2. Basic anti-ESD Procedures for Analyzer Repair and  
Maintenance  
12.4.2.1. Working at the Instrument Rack  
When working on the analyzer while it is in the instrument rack and plugged into a properly  
grounded power supply  
1. Attach you anti-ESD wrist strap to ground before doing anything else.  
Use a wrist strap terminated with an alligator clip and attach it to a bare metal portion of  
the instrument chassis. This will safely connect you to the same ground level to which the  
instrument and all of its components are connected.  
2. Pause for a second or two to allow any static charges to bleed away.  
3. Open the casing of the analyzer and begin work. Up to this point the closed metal casing of  
your analyzer has isolated the components and assemblies inside from any conducted or  
induces static charges.  
4. If you must remove a component from the instrument, do not lay it down on a non-ESD  
preventative surface where static charges may lie in wait.  
5. Only disconnect your wrist strap after you have finished work and closed the case of the  
analyzer.  
12.4.2.2. Working at a Anti-ESD Work Bench.  
When working on an instrument of an electronic assembly while it is resting on a anti-ESD work  
bench  
1. Plug you anti-ESD wrist strap into the grounded receptacle of the work station before touching  
any items on the work station and while standing at least a foot or so away, This will allow  
any charges you are carrying to bleed away through the ground connection of the work station  
and prevent discharges due to field effects and induction from occurring.  
2. Pause for a second or two to allow any static charges to bleed away.  
3. Only open any anti-ESD storage bins or bags containing sensitive devices or assemblies after  
you have plugged your wrist strap into the work station.  
Lay the bag or bin on the workbench surface.  
Before opening the container, wait several seconds for any static charges on the outside  
surface of the container to be bled away by the work station’s grounded protective mat.  
4. Do not pick up tools that may be carrying static charges while also touching or holding an ESD  
Sensitive Device.  
Only lay tools or ESD-sensitive devices and assemblies on the conductive surface of your  
workstation. Never lay them down on a non-ESD preventative surfaces.  
5. Place any static sensitive devices or assemblies in anti-static storage bags or bins and close  
the bag or bin before unplugging your wrist strap.  
04584 Rev A1  
218  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
A Primer on Electro-Static Discharge  
6. Disconnecting your wrist strap is always the last action taken before leaving the work bench.  
12.4.2.3. Transferring Components from Rack To Bench and Back  
When transferring a sensitive device from an installed Teledyne Instruments analyzer to a Anti-  
ESD workbench or back:  
1. Follow the instructions listed above for working at the instrument rack and work station.  
2. Never carry the component or assembly without placing it in a anti-ESD bag or bin.  
3. Before using the bag or container allow any surface charges on it to dissipate:  
If you are at the instrument rack hold the bag in one hand while your wrist strap is  
connected to a ground point.  
If you are at a anti-ESD work bench, lay the container down on the conductive work  
surface.  
In either case wait several seconds.  
4. Place the item in the container.  
5. Seal the container. If using a bag, fold the end over and fastening it with anti-ESD tape.  
Never use standard plastic adhesive tape as a sealer.  
Folding the open end over isolates the component(s) inside from the effects of static fields.  
Leaving the bag open or simply stapling it shut without folding it closed prevents the bag  
from forming a complete protective envelope around the device.  
6. Once you have arrived at your destination, allow any surface charges that may have built up  
on the bag or bin during travel to dissipate:  
Connect your wrist strap to ground.  
If you are at the instrument rack hold the bag in one hand while your wrist strap is  
connected to a ground point.  
If you are at a anti-ESD work bench, lay the container down on the conductive work  
surface  
In either case wait several seconds  
7. Open the container.  
12.4.2.4. Opening Shipments from and Packing Components for Return to  
Teledyne Instruments Customer Service.  
Packing materials such as bubble pack and Styrofoam pellets are extremely efficient generators of  
static electric charges. To prevent damage from ESD, Teledyne Instruments ships all electronic  
components and assemblies in properly sealed ant-ESD containers.  
04584 Rev A1  
219  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
A Primer on Electro-Static Discharge  
Static charges will build up on the outer surface of the anti-ESD container during shipping as the  
packing materials vibrate and rub against each other. To prevent these static charges from  
damaging the components or assemblies being shipped make sure that you:  
Always unpack shipments from Teledyne Instruments Customer Service by:  
Opening the outer shipping box away from the anti-ESD work area  
Carry the still sealed ant-ESD bag, tube or bin to the anti-ESD work area  
Follow steps 6 and 7 of Section 12.4.2.3 above when opening the anti-ESD container at  
the work station  
Reserve the anti-ESD container or bag to use when packing electronic components or  
assemblies to be returned to Teledyne Instruments  
Always pack electronic components and assemblies to be sent to Teledyne Instruments  
Customer Service in anti-ESD bins, tubes or bags.  
Do not use pink-poly bags.  
If you do not already have an adequate supply of anti-ESD bags or containers  
available, Teledyne Instruments’ Customer Service department) will supply them (see  
Section 11.7 for contact information.  
Always follow steps 1 through 5 of Section 12.4.1.3  
User Notes  
04584 Rev A1  
220  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction ManualAPPENDIX A - Version Specific Software Documentation  
APPENDIX A - Version Specific Software Documentation  
APPENDIX A-1: Model GFC7000E Software Menu Trees  
APPENDIX A-2: Model GFC7000E Setup Variables Available Via Serial I/O  
APPENDIX A-3: Model GFC7000E Warnings and Test Measurements Via Serial I/O  
APPENDIX A-4: Model GFC7000E Signal I/O Definitions  
APPENDIX A-5: Model GFC7000E iDAS Functions  
045150110 Rev A1  
221  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0  
APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0  
SAMPLE  
TEST1  
CAL  
MSG1,2  
CLR1,3  
SETUP  
Only appear if  
reporting range  
is set for  
AUTO range  
mode.  
ENTER SETUP PASS: 818  
LOW  
HIGH  
<TST TST>  
(Primary Setup Menu)  
CFG  
DAS  
RANG PASS  
CLK  
MORE  
ZERO  
SPAN CONC  
RANGE  
(Secondary Setup Menu)  
STABIL  
PRES  
SAMP FL  
PMT  
NORM PMT  
UV LAMP  
LAMP RATIO  
STR. LGT  
DARK PMT  
DARK LAMP  
SLOPE  
COMM VARS  
DIAG  
OFFSET  
HVPS  
RCELL TEMP  
BOX TEMP  
PMT TEMP  
IZS TEMP1  
TEST2  
TEST FUNCTIONS  
Viewable by user while  
instrument is in  
SAMPLE Mode  
(see Section 6.2.1)  
1
2
3
Only appears when warning messages are activated  
(see Section 6.2.2)  
Press this key to cycle through list of active warning  
messages.  
Press this key to clear/erase the warning message  
currently displayed  
TIME  
Figure A-1: Basic Sample Display Menu  
045840110 Rev B.3  
223  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0  
SAMPLE  
TEST1  
CAL  
CALS  
MSG1,2  
CLR1,3  
CALZ  
SETUP  
Only appear if  
reporting range  
is set for  
<TST TST>  
LOW  
HIGH  
LOW  
HIGH  
LOW  
HIGH  
AUTO range  
mode.  
RANGE  
RANGE1*  
RANGE2*  
STABIL  
ZERO  
SPAN CONC  
ZERO  
SPAN CONC  
ENTER SETUP PASS: 818  
CO2 MEAS  
CO2 REF  
MR RATIO  
PRES  
SAMP FL  
(Primary Setup Menu)  
SAMP TEMP  
BENCH TEMP  
WHEEL TEMP  
BOX TEMP  
PHT DRIVE  
SLOPE  
CFG  
DAS  
RANG PASS  
CLK  
MORE  
OFFSET  
TEST  
TIME  
(Secondary Setup Menu)  
TEST FUNCTIONS  
Viewable by user while  
instrument is in SAMPLE Mode  
(see Section 6.2.1)  
1
2
3
Only appears when warning messages are activated  
(see Section 6.2.2)  
Press this key to cycle through list of active warning  
messages.  
Press this key to clear/erase the warning message  
currently displayed  
COMM VARS  
DIAG  
*Only appears instrument is set for  
DUAL or AUTO reporting range  
modes  
Figure A-2: Sample Display Menu - Units with Z/S Valve or IZS Option installed  
045840110 Rev B.3  
224  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0  
SETUP  
ENTER SETUP PASS: 818  
ACAL1  
NEXT  
CFG  
DAS  
RNGE  
PASS  
CLK  
MORE  
ON  
PREV  
NEXT  
Go To iDAS  
MENU TREE  
(Fig. A-5)  
PREV  
MODE  
SET2  
OFF  
MODEL NAME  
SERIAL NUMBER  
TIME  
DATE  
SEQ 1)  
SEQ 2)  
SOFTWARE  
REVISION  
SEQ 3  
)
LIBRARY REVISION  
MODE  
SET  
UNIT  
iCHIP SOFTWARE  
REVISION1  
HESSEN PROTOCOL  
REVISION1  
ACTIVE  
PREV  
NEXT  
SPECIAL SOFTWARE  
OPTIONS1  
CPU TYPE  
SNGL  
DUAL  
AUTO  
DATE FACTORY  
CONFIGURATION  
SAVED  
DISABLED  
ZERO  
ZERO/SPAN  
SPAN  
PPB  
PPM  
UGM  
MGM  
%
ENTR  
ENTR  
TIMER ENABLE  
STARTING DATE  
STARTING TIME  
DELTA DAYS  
DELTA TIME  
Go To  
1
2
Only appears if a applicable  
option/feature is installed  
and activated.  
Appears whenever the  
currently displayed  
sequence is not set for  
DISABLED.  
Only appears when  
reporting range is set to  
AUTO range mode.  
SECONDARY SETUP MENU  
<SET SET>  
LOW3 HIGH3  
EDIT  
(Fig. A-4)  
DURATION  
CALIBRATE  
3
RANGE TO CAL3  
Figure A-3: Primary Setup Menu (Except iDAS)  
045840110 Rev B.3  
225  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0  
SETUP  
ENTER SETUP PASS: 818  
ACAL1  
CFG  
DAS  
RNGE  
PASS  
CLK  
MORE  
VIEW  
EDIT  
PREV NEXT  
CONC  
PNUMTC  
CALDAT  
ZTBZRO  
STBSPN  
TEMP  
PREV NEXT  
INS  
DEL  
EDIT  
PRNT  
YES  
NO  
CONC  
VIEW  
PREV NEXT  
PNUMTC  
CALDAT  
ZTBZRO  
STBSPN  
TEMP  
<SET  
SET>  
EDIT  
PRNT  
<PRM PRM> PV10  
NX10  
Selects data point to view.  
Creates/changes name  
(see Section 6.12.2).  
Cycles through  
lists of  
parameters  
chosen for this  
iDAS channel  
NAME  
EVENT  
PARAMETERS  
REPORT PERIOD  
NUMBER OF RECORDS  
RS-232 REPORT  
CHANNEL ENABLE  
CAL. HOLD  
Sets the  
amount of time  
between each  
report.  
YES  
NO  
PREV NEXT  
PREV NEXT  
INS  
DEL  
EDIT  
Cycles through  
available trigger  
events  
(see Section 6.12.3).  
YES  
NO  
ON  
YES  
NO  
OFF  
<SET  
SET>  
EDIT  
Cycles through  
already active  
parameters  
(see Section 6.12.4).  
Selects max  
no. of records  
for this channel  
PARAMETER  
PREV NEXT  
SAMPLE MODE  
PRECISION  
INST  
AVG  
MIN  
MAX  
Cycles through available/active parameters  
(see Section 6.12.4).  
1
Only appears if a valve option is installed.  
Figure A-4: Primary Setup Menu (iDAS)  
045840110 Rev B.3  
226  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0  
SETUP  
ENTER SETUP PASS: 818  
ACAL1  
CFG  
DAS  
RNGE  
PASS  
CLK  
MORE  
COMM  
VARS  
DIAG  
ID  
COM1  
COM2  
PREV NEXT JUMP  
EDIT PRINT  
DAS_HOLD_OFF  
CONC PRECISION  
DYN_ZERO OFF  
DYN_SPAN OFF  
CLOCK_ADJ  
<SET  
SET>  
EDIT  
MODE  
BAUD RATE TEST PORT  
PREV NEXT  
PREV NEXT  
TEST  
QUIET  
COMPUTER  
SECURITY  
HESSEN PROTOCOL  
COMx E,7,1  
RS-485  
MULTIDROP  
ENABLE MODEM  
ENABLE INTERNET  
IGNORE ERRORS  
DISABLE XON/XOFF  
COMMAND PROMPT  
300  
1200  
2400  
4800  
9600  
Go To  
DIAG MENU TREE  
(Fig A-6)  
19200  
38400  
57760  
115200  
ON  
1
Only appears if a valve is installed.  
OFF  
Figure A-5: Secondary Setup Menu (COMM & VARS)  
045840110 Rev B.3  
227  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0  
SETUP  
ENTER SETUP PASS: 818  
ACAL1  
CFG  
DAS  
RNGE  
PASS  
CLK  
MORE  
COMM  
ID  
VARS  
DIAG  
COM1  
PREV NEXT JUMP  
EDIT PRINT  
INET2  
DAS_HOLD_OFF  
COMM - MENU TREE  
CONC PRECISION  
DYN_ZERO OFF  
DYN_SPAN OFF  
CLOCK_ADJ  
(Fig A-5)  
<SET  
SET>  
EDIT  
DHCP  
INSTRUMENT IP  
GATEWAY IP  
SUBNET MASK  
TCP PORT3  
HOSTNAME4  
Go To  
DIAG MENU TREE  
(Fig A-6)  
ON  
INSTRUMENT IP5  
GATEWAY IP5  
SUBNET MASK5  
TCP PORT3  
OFF  
EDIT  
1
Only appears if a valve option is installed.  
Only appears when the Ethernet card (option 63) is installed.  
Although TCP PORT is editable regardless of the DHCP state, do not change the setting for this property unless  
instructed to by Teledyne Instruments Customer Service personnel.  
HOST NAME is only editable when DHCP is ON.  
2
3
4
5
INSTRUMENT IP, GATEWAY IP & SUBNET MASK are only editable when DHCP is OFF.  
Figure A-6: Secondary Setup Menu (COMM Menu with Ethernet Card)  
045840110 Rev B.3  
228  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0  
SETUP  
.
ENTER SETUP PASS: 818  
ACAL1  
CFG  
DAS  
RNGE  
PASS  
CLK  
MORE  
COMM  
ID  
VARS  
DIAG  
HESSEN2  
COM1 COM2  
See  
See  
Fig A-5  
Fig A-6  
<SET  
SET> EDIT  
<SET  
SET> EDIT  
VARIATION RESPONSE MODE GAS LIST STATUS FLAGS  
TYPE 1  
TYPE 2  
BCC  
TEXT  
CMD  
MODE  
BAUD RATE TEST PORT  
PREV NEXT TEST  
CO2, 310, REPORTED  
PREV NEXT  
PREV NEXT  
INS  
DEL  
EDIT  
PRNT  
300  
1200  
2400  
4800  
SAMPLE FLOW WARNING  
BENCH TEMP WARNING  
SOURCE WARNING  
QUIET  
COMPUTER  
SECURITY  
HESSEN PROTOCOL  
COM[1,2] E,7,1  
RS-485  
BOX TEMP WARNING  
WHEEL TEMP WARNING  
SAMPLE TEMP WARNING  
SAMPLE PRESSURE WARNING  
INVALID CONC  
9600  
19200  
38400  
57760  
115200  
MULTIDROP  
INSTRUMENT OFF  
ENABLE MODEM  
ENABLE INTERNET  
IGNORE ERRORS  
DISABLE XON/XOFF  
COMMAND PROMPT  
IN MANUAL CALIBRATION MODE  
IN ZERO CALIBRATION MODE  
IN SPAN CALIBRATION MODE  
UGM  
1
2
Only appears if a valve is installed.  
Only appears when the HESSEN  
mode is enabled for either COM1  
or COM2.  
MGM  
See Table 6-27 for  
Flag Assignments  
PPB  
PPM  
ON  
OFF  
Figure A-7: Secondary Setup Menu (COMM Menu with HESSEN)  
045840110 Rev B.3  
229  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX A-1: MGFC7000E Software Menu Trees, Revision E.0  
SETUP  
ENTER SETUP PASS: 818  
ACAL1  
DAS  
RNGE  
PASS  
CLK  
MORE  
CFG  
DIAG  
COMM  
VARS  
PREV NEXT  
SIGNAL  
I/O  
ANALOG  
OUTPUT  
ANALOG I/O  
CONFIGURATION  
ELECTRICAL  
DARK  
CALIBRATION  
PRESSURE  
CALIBRATION  
FLOW  
CALIBRATION  
TEST  
CHANNEL  
TEST  
OUTPUT  
ENTR  
ENTR  
ENTR  
ENTR  
PREV NEXT  
CAL  
EDIT  
EXIT  
Start step Test  
Starts Test  
Starts Test  
VIEW  
CAL  
0) EXT ZERO CAL  
1) EXT SPAN CAL  
2) SYNC OK  
ENTR  
3) MAINT MODE  
4) LANG2 SELECT  
NONE  
CO2 MEASURE  
CO2 REFERENCE  
AOUTS CALIBRATED  
CAL  
<SET  
SET>  
5) SAMPLE LED  
6) CAL LED  
7) FAULT LED  
8) AUDIBLE BEEPER  
9) ELEC TEST  
SAMPLE PRESSURE  
SAMPLE FLOW  
SAMPLE TEMP  
BENCH TEMP  
SAMPLE LOW = XXX.X MV  
SAMPLE LOW = X.X IN-HG-A  
SAMPLE HIGH = XXX.X MV  
SAMPLE HIGH = X.X IN-HG-A  
CONC OUT 1  
CONC OUT 2  
TEST OUTPUT  
10) DARK CAL  
WHEEL TEMP2  
CHASSIS TEMP  
PHT DRIVE  
11) ST SYSTEM OK  
12) ST CONC VALID  
13) ST HIGH RANGE  
14) ST ZERO CAL  
15) ST SPAN CAL  
16) ST DIAG MODE  
17) ST SYSTEM OK2  
18) ST CONC ALARM 1  
19) ST CONC ALARM 2  
20) RELAY WATCHDOG  
21) WHEEL HEATER  
22) BENCH HEATER  
23) CAL VALVE  
ON  
EDIT  
<SET SET>  
OFF  
24) SPAN VALVE  
25) SHUTOFF VALVE  
RANGE  
REC OFFSET  
AUTO CAL  
CALIBRATED  
CAL  
26  
INTERNAL ANALOG  
VOLTAGE SIGNALS  
(see Appendix A)  
45  
ON  
0.1V 1V  
5V 10V CURR  
46) CONC OUT 1  
47) CONC OUT 2  
48) TEST OUTPUT  
OFF  
MV  
1
2
Only relevant to on M200EH  
Only relevant to analyzers with IZS options installed  
Figure A-8: Secondary Setup Menu (DIAG)  
045840110 Rev B.3  
230  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX A-2: Setup Variables For Serial I/O, Revision  
E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0  
APPENDIX A-2: Setup Variables For Serial I/O, Revision E.0  
Table A-1:  
MGFC7000E Setup Variables, Revision E.0  
SETUP VARIABLE  
NUMERIC  
UNITS  
DEFAULT VALUE  
VALUE RANGE  
DESCRIPTION  
STANDARD SETUP VARIABLES  
DAS_HOLD_OFF  
Minutes  
15  
3
0.5–20  
Duration of DAS hold off  
period.  
CONC_PRECISION  
AUTO, 0, 1, 2, 3,  
4
Number of digits to display to  
the right of the decimal point  
for concentrations on the  
display.  
DYN_ZERO  
DYN_SPAN  
CLOCK_ADJ  
OFF  
OFF  
0
ON, OFF  
ON, OFF  
-60–60  
ON enables remote dynamic  
zero calibration; OFF disables  
it.  
ON enables remote dynamic  
span calibration; OFF disables  
it.  
Sec./Day  
Time-of-day clock speed  
adjustment.  
MEDIUM ACCESS LEVEL SETUP VARIABLES  
0
LANGUAGE_SELECT  
MAINT_TIMEOUT  
ENGL  
ENGL,  
SECD,  
EXTN  
Selects the language to use  
for the user interface.  
Hours  
2
0.1–100  
Time until automatically  
switching out of software-  
controlled maintenance  
mode.  
0
CONV_TIME  
33 MS  
33 MS, 66 MS,  
133 MS,  
266 MS,  
Conversion time for  
measure/reference detector  
channel.  
533 MS, 1 SEC, 2  
SEC  
CO_DWELL  
CO_SAMPLE  
Seconds  
Samples  
0.2  
1
0.1–30  
Dwell time before taking  
measure or reference sample.  
1–30  
Number of samples to take in  
measure or reference mode.  
3, 8  
FILT_SIZE  
Samples  
Samples  
750, 200  
1–1000  
1–1000  
Moving average filter size.  
3, 8  
FILT_ASIZE  
48, 20  
4, 15  
10  
Moving average filter size in  
adaptive mode.  
3, 8  
FILT_DELTA  
FILT_PCT  
PPM  
%
1–1000  
1–100  
Absolute change to trigger  
adaptive filter.  
Percent change to trigger  
adaptive filter.  
FILT_DELAY  
FILT_ADAPT  
Seconds  
90  
0–180  
Delay before leaving adaptive  
filter mode.  
ON  
ON  
5
ON, OFF  
ON, OFF  
1–300  
ON enables adaptive filter;  
OFF disables it.  
10  
CO2_FILT_ADAPT  
ON enables CO2 adaptive  
filter; OFF disables it.  
10  
CO2_FILT_SIZE  
Samples  
Samples  
%
CO2 moving average filter  
size.  
10  
CO2_FILT_ASIZE  
1
1–300  
CO2 moving average filter  
size in adaptive mode.  
10  
CO2_FILT_DELTA  
0.5  
5
0.01–10  
0.1–100  
0–300  
Absolute CO2 conc. change to  
trigger adaptive filter.  
10  
CO2_FILT_PCT  
%
Percent CO2 conc. change to  
trigger adaptive filter.  
10  
CO2_FILT_DELAY  
Seconds  
5
Delay before leaving CO2  
adaptive filter mode.  
045840110 Rev B.3  
231  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX A-2: Setup Variables For Serial I/O, Revision  
E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0  
SETUP VARIABLE  
NUMERIC  
UNITS  
DEFAULT VALUE  
VALUE RANGE  
DESCRIPTION  
0
3, 8  
USER_UNITS  
DIL_FACTOR  
PPM  
PPB, PPM,  
,
Concentration units for user  
interface.  
3, 8  
UGM, MGM  
,
1
0.1–1000  
Dilution factor. Used only if is  
dilution enabled with  
FACTORY_OPT variable.  
DARK_CAL_DURATION  
Seconds  
180,  
60  
10–600  
Duration of dark cal. First  
two-thirds is stabilization  
period; final third is measure  
period.  
4
DARK_MEAS_MV  
DARK_REF_MV  
mV  
mV  
0
0
-1000–1000  
-1000–1000  
1–10000  
Dark offset for measure  
reading.  
Dark offset for reference  
reading.  
LIN_TARGET_CONC1  
Conc  
300  
Target concentration during  
linearity adjustment for range  
1.  
LIN_NORM_CONC1  
LIN_RATIO1  
PPM  
300  
1
0.01–10000  
0.01–100  
Target concentration during  
linearity adjustment  
normalized for T/P for range  
1.  
Measure/reference ratio  
measured during linearity  
adjustment for range 1.  
LIN_CORRECT1  
1
0.001–999.999  
1–10000  
Linearity correction factor for  
range 1.  
LIN_TARGET_CONC2  
Conc  
300  
Target concentration during  
linearity adjustment for range  
2.  
LIN_NORM_CONC2  
LIN_RATIO2  
PPM  
300  
1
0.01–10000  
0.01–100  
Target concentration during  
linearity adjustment  
normalized for T/P for range  
2.  
Measure/reference ratio  
measured during linearity  
adjustment for range 2.  
LIN_CORRECT2  
LIN_TARGET_CONC  
LIN_NORM_CONC  
1
0.001–999.999  
1–10000  
Linearity correction factor for  
range 2.  
Conc  
PPM  
300  
300  
Target concentration during  
linearity adjustment.  
0.01–10000  
Target concentration during  
linearity adjustment  
normalized for T/P.  
LIN_RATIO  
1
0.01–100  
Measure/reference ratio  
measured during linearity  
adjustment.  
LIN_CORRECT  
1
0.001–999.999  
ON, OFF  
Linearity correction factor.  
CO2_COMP_ENABLE  
OFF  
ON enables CO2  
compensation; OFF disables  
it.  
CO2_COMP_CONC  
%
0
0–20  
CO2 concentration to  
compensate for.  
3, 8  
CO_CONST1  
CO_CONST2  
700, 500 4, 40000  
100–50000  
0–10  
CO calculation constant.  
CO calculation constant.  
0.13, 0.198 8 1.448  
4,1.458 12, 0.187  
3
ET_MEAS_GAIN  
ET_REF_GAIN  
1
0.0001–9.9999  
0.0001–9.9999  
0–5000  
Electrical test gain factor for  
measure reading.  
1
Electrical test gain factor for  
reference reading.  
ET_TARGET_DET  
mV  
4375  
Target detector reading  
during electrical test.  
045840110 Rev B.3  
232  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX A-2: Setup Variables For Serial I/O, Revision  
E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0  
SETUP VARIABLE  
NUMERIC  
UNITS  
DEFAULT VALUE  
VALUE RANGE  
DESCRIPTION  
3, 8  
ET_TARGET_CONC  
ET_CONC_RANGE  
STD_TEMP  
PPM  
Conc.  
ºK  
40, 400  
1–10000  
0.1–50000  
1–500  
Target concentration during  
electrical test.  
3, 8  
50, 5000  
D/A concentration range  
during electrical test.  
321  
Standard temperature for  
temperature compensation.  
12  
STD_PRESS  
"Hg  
ºC  
28.5, 28.7 8, 28.8  
,
1–50  
Standard pressure for  
pressure compensation.  
4
28.1  
BENCH_SET  
48  
0–100  
Optical bench temperature  
set point and warning limits.  
Warnings: 43–53  
WHEEL_SET  
ºC  
68  
0–100  
Wheel temperature set point  
and warning limits.  
Warnings: 63–73  
3, 5, 8  
8
ZERO_ENABLE  
ON, OFF  
OFF, ON  
ON enables auto-zero  
calibration using scrubber;  
OFF disables it.  
3, 5, 8  
5
ZERO_FREQ  
Minutes  
5, 20  
0.1–1440  
Auto-zero calibration  
frequency.  
3, 5, 8  
5
5
ZERO_DWELL  
Seconds,  
Minutes  
7, 5  
1–60, 1–30  
Dwell time after closing or  
opening zero scrubber valve.  
5
3, 5, 8  
ZERO_SAMPLES  
Samples  
Samples  
Ratio  
15  
5
1–1000  
1–100  
0–5  
Number of zero samples to  
average.  
3, 5, 8  
ZERO_FILT_SIZE  
Auto-zero offset moving  
average filter size.  
3, 5, 8  
3, 8  
ZERO_LIMIT  
1.2, 1.15  
Minimum auto-zero ratio  
allowed; must be greater  
than this value to be valid.  
3, 5, 8  
ZERO_CAL  
Ratio  
1.18  
0.5–5  
Calibrated auto-zero ratio.  
3, 8  
CO_SPAN1  
Conc.  
40, 400  
1–10000,  
1–20000  
Target CO concentration  
during span calibration of  
range 1.  
3, 8  
CO_SLOPE1  
CO_OFFSET1  
CO_SPAN2  
1
0
0.001–999.999  
-10–10  
CO slope for range 1.  
CO offset for range 1.  
3, 8  
Conc.  
40, 400  
1–10000,  
Target CO concentration  
during span calibration of  
range 2.  
3, 8  
1–20000  
CO_SLOPE2  
CO_OFFSET2  
RANGE_MODE  
1
0
0.001–999.999  
-10–10  
CO slope for range 2.  
CO offset for range 2.  
Range control mode.  
0
SNGL  
SNGL, DUAL,  
AUTO  
3, 8  
CONC_RANGE1  
Conc.  
Conc.  
%
50, 200 6, 500  
50, 200 6, 500  
0.1–50000  
0.1–50000  
1–15  
D/A concentration range 1.  
D/A concentration range 2.  
1
3, 8  
CONC_RANGE2  
10  
CONC_RANGE3  
15  
D/A concentration range 3  
(CO2).  
045840110 Rev B.3  
233  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX A-2: Setup Variables For Serial I/O, Revision  
E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0  
SETUP VARIABLE  
RS232_MODE  
NUMERIC  
UNITS  
DEFAULT VALUE  
VALUE RANGE  
DESCRIPTION  
BitFlag  
0
0–65535  
RS-232 COM1 mode flags.  
Add values to combine flags.  
1 = quiet mode  
2 = computer mode  
4 = enable security  
8 = enable hardware  
handshaking  
16 = enable Hessen protocol  
11  
32 = enable multi-drop  
64 = enable modem  
128 = ignore RS-232 line  
errors  
256 = disable XON / XOFF  
support  
512 = disable hardware  
FIFOs  
1024 = enable RS-485 mode  
2048 = even parity, 7 data  
bits, 1 stop bit  
4096 = enable command  
prompt  
0
BAUD_RATE  
19200  
300, 1200, 2400,  
4800,  
RS-232 COM1 baud rate.  
9600, 19200,  
38400, 57600,  
115200  
MODEM_INIT  
“AT Y0 &D0 &H0 &I0  
S0=2 &B0 &N6 &M0  
Any character in  
the allowed  
character set. Up  
to 100 characters  
long.  
RS-232 COM1 modem  
initialization string. Sent  
verbatim plus carriage return  
to modem on power up or  
manually.  
0
E0 Q1 &W0”  
RS232_MODE2  
BAUD_RATE2  
BitFlag  
0
0–65535  
RS-232 COM2 mode flags.  
(Same settings as  
RS232_MODE.)  
0
19200  
300, 1200, 2400,  
4800, 9600,  
RS-232 COM2 baud rate.  
19200, 38400,  
57600, 115200  
MODEM_INIT2  
“AT Y0 &D0 &H0 &I0  
S0=2 &B0 &N6 &M0  
Any character in  
the allowed  
character set. Up  
to 100 characters  
long.  
RS-232 COM2 modem  
initialization string. Sent  
verbatim plus carriage return  
to modem on power up or  
manually.  
0
E0 Q1 &W0”  
RS232_PASS  
MACHINE_ID  
Password  
ID  
940331  
0–999999  
0–9999  
RS-232 log on password.  
4
300, 320  
Unique ID number for  
instrument.  
0
COMMAND_PROMPT  
“Cmd> ”  
Any character in  
the allowed  
character set. Up  
to 100 characters  
long.  
RS-232 interface command  
prompt. Displayed only if  
enabled with RS232_MODE  
variable.  
045840110 Rev B.3  
234  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX A-2: Setup Variables For Serial I/O, Revision  
E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0  
SETUP VARIABLE  
TEST_CHAN_ID  
NUMERIC  
UNITS  
DEFAULT VALUE  
VALUE RANGE  
DESCRIPTION  
0
NONE  
NONE,  
Diagnostic analog output ID.  
CO MEASURE,  
CO REFERENCE,  
VACUUM  
PRESSURE,  
SAMPLE  
PRESSURE,  
SAMPLE FLOW,  
SAMPLE TEMP,  
BENCH TEMP,  
WHEEL TEMP,  
CHASSIS TEMP,  
PHT DRIVE  
0
REMOTE_CAL_MODE  
LOW  
LOW,  
HIGH  
Range to calibrate during  
contact closure or Hessen  
calibration.  
PASS_ENABLE  
STABIL_GAS  
OFF  
ON, OFF  
ON enables passwords; OFF  
disables them.  
0
CO  
CO,  
CO2  
Selects gas for stability  
measurement.  
STABIL_FREQ  
STABIL_SAMPLES  
Seconds  
Samples  
10  
25  
1–300  
Stability measurement  
sampling frequency.  
2–40  
Number of samples in  
concentration stability  
reading.  
PHOTO_TEMP_SET  
mV  
2000  
0–5000  
Photometer temperature  
warning limits. Set point is  
not used.  
Warnings: 500–3000  
SAMP_PRESS_SET  
PURGE_PRESS_SET  
SAMP_FLOW_SET  
In-Hg  
PSIG  
cc/m  
29.92  
Warnings: 15–35  
0–100  
0–100  
0–5000  
Sample pressure warning  
limits. Set point is not used.  
7.5  
Purge pressure warning  
limits. Set point is not used.  
Warnings: 2.5–12.5  
750, 800 3, 8, 3000 4,  
Sample flow warning limits.  
Set point is not used.  
13  
2000  
Warnings:  
500–1000, 500–  
12003,8  
,
1000–50004,  
8,9  
400–1000  
1500–2500  
,
13  
SAMP_FLOW_SLOPE  
VAC_SAMP_RATIO  
1
0.001–100  
0.1–2  
Slope term to correct sample  
flow rate.  
0.53,  
0.61  
Maximum vacuum pressure /  
sample pressure ratio for  
valid sample flow calculation.  
13  
SAMP_TEMP_SET  
BOX_SET  
ºC  
ºC  
30  
0–100  
0–100  
Sample temperature warning  
limits. Set point is not used.  
Warnings: 10.1–100  
30  
Internal box temperature  
warning limits. Set point is  
not used.  
Warnings: 5–48  
BENCH_CYCLE  
BENCH_PROP  
Seconds  
1/ºC  
2
5
0.5–30  
0–100  
Optical bench temperature  
control cycle period.  
Optical bench temperature  
PID proportional coefficient.  
Proportional band is the  
reciprocal of this setting.  
BENCH_INTEG  
0.5  
0–100  
Optical bench temperature  
PID integral coefficient.  
045840110 Rev B.3  
235  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX A-2: Setup Variables For Serial I/O, Revision  
E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0  
SETUP VARIABLE  
NUMERIC  
UNITS  
DEFAULT VALUE  
VALUE RANGE  
DESCRIPTION  
BENCH_DERIV  
WHEEL_CYCLE  
WHEEL_PROP  
2
0–100  
0.5–30  
0–100  
Optical bench temperature  
PID derivative coefficient.  
4
Seconds  
1/ºC  
4, 2  
Wheel temperature control  
cycle period.  
1
Wheel temperature PID  
proportional coefficient.  
Proportional band is the  
reciprocal of this setting.  
4
WHEEL_INTEG  
WHEEL_DERIV  
TPC_ENABLE  
0.135, 0.035  
0–100  
Wheel temperature PID  
integral coefficient.  
2
0–100  
Wheel temperature PID  
derivative coefficient.  
ON  
OFF, ON  
ON enables temperature/  
pressure compensation; OFF  
disables it.  
CONC_LIN_ENABLE  
SERIAL_NUMBER  
ON  
OFF, ON  
ON enables concentration  
linearization; OFF disables it.  
0
“00000000 ”  
Any character in  
the allowed  
Unique serial number for  
instrument.  
character set. Up  
to 100 characters  
long.  
0
DISP_INTENSITY  
HIGH  
HIGH,  
MED,  
LOW,  
DIM  
Front panel display intensity.  
I2C_RESET_ENABLE  
FACTORY_OPT  
ON  
OFF, ON  
ON enables automatic reset  
of the I2C bus in the event of  
communication failures; OFF  
disables automatic reset.  
9
BitFlag  
0, 4  
0–65535  
Factory option flags. Add  
values to combine flags.  
1 = enable dilution factor  
2 = zero/span valves  
installed  
4 = conc. alarms routed to  
relays  
8 = enable linearity  
adjustment factor  
16 = display units in  
concentration field  
32 = enable software-  
controlled maintenance mode  
3, 5  
64  
= span valve installed  
128 = enable switch-  
controlled maintenance mode  
2048 = enable Internet  
7
option  
4096 = use “old” style  
numeric data entry menus  
when editing conc. table  
045840110 Rev B.3  
236  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX A-2: Setup Variables For Serial I/O, Revision  
E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0  
SETUP VARIABLE  
CLOCK_FORMAT  
NUMERIC  
UNITS  
DEFAULT VALUE  
VALUE RANGE  
DESCRIPTION  
“TIME=%H:%M:%S”  
Any character in  
the allowed  
character set. Up  
to 100 characters  
long.  
Time-of-day clock format  
flags. Enclose value in double  
quotes (") when setting from  
the RS-232 interface.  
“%a” = Abbreviated weekday  
name.  
“%b” = Abbreviated month  
name.  
“%d” = Day of month as  
decimal number (01 – 31).  
“%H” = Hour in 24-hour  
format (00 – 23).  
“%I” = Hour in 12-hour  
format (01 – 12).  
“%j” = Day of year as  
decimal number (001 – 366).  
“%m” = Month as decimal  
number (01 – 12).  
“%M” = Minute as decimal  
number (00 – 59).  
“%p” = A.M./P.M. indicator  
for 12-hour clock.  
“%S” = Second as decimal  
number (00 – 59).  
“%w” = Weekday as decimal  
number (0 – 6; Sunday is 0).  
“%y” = Year without century,  
as decimal number (00 – 99).  
“%Y” = Year with century, as  
decimal number.  
“%%” = Percent sign.  
ALARM_TRIGGER3,4  
REF_SDEV_LIMIT  
Cycles  
mV  
10  
50  
1–100  
Concentration alarm trigger  
sensitivity adjustment.  
0.1–500  
Reference detector standard  
deviation must be below this  
limit to switch out of startup  
mode.  
0
Enclose value in double quotes (") when setting from the RS-232 interface.  
Multi-range modes.  
Hessen protocol.  
M300EH.  
GFC7000E.  
M300ES.  
Fixed range special.  
iCHIP option.  
1
2
3
4
5
6
7
8
M300EM.  
FC7000E.  
M306E.  
9
10  
11  
12  
Must power-cycle instrument for these options to take effect.  
GFC7000EU.  
045840110 Rev B.3  
237  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX A-3: Warnings and Test Functions, Revision  
E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0  
APPENDIX A-3: Warnings and Test Functions, Revision E.0  
Table A-2:  
GFC7000E Warning Messages, Revision E.0  
NAME  
WSYSRES  
MESSAGE TEXT  
DESCRIPTION  
Instrument was power-cycled or the CPU was reset.  
Data storage was erased.  
SYSTEM RESET  
DATA INITIALIZED  
CONFIG INITIALIZED  
WDATAINIT  
WCONFIGINIT  
Configuration storage was reset to factory configuration  
or erased.  
WSOURCE  
SOURCE WARNING  
AZERO WARN 1.001  
Reference reading below 1250 mV (50 mV in N2O, CO2,  
and high level instruments) or above 4950 mV in normal  
sample mode.  
4, 5  
WAUTOZERO  
WBENCHTEMP  
WWHEELTEMP  
Auto-reference ratio below limit specified by ZERO_LIMIT  
variable.  
BENCH TEMP  
WARNING  
Bench temperature outside of warning limits specified by  
BENCH_SET variable.  
WHEEL TEMP  
WARNING  
Wheel temperature outside of warning limits specified by  
WHEEL_SET variable.  
6
WSAMPFLOW  
WSAMPPRESS  
WSAMPTEMP  
WPURGEPRESS  
WBOXTEMP  
SAMPLE FLOW WARN  
SAMPLE PRESS WARN  
SAMPLE TEMP WARN  
PURGE PRESS WARN  
BOX TEMP WARNING  
SYNC WARNING  
Sample flow outside of warning limits specified by  
SAMP_FLOW_SET variable.  
Sample pressure outside of warning limits specified by  
SAMP_PRESS_SET variable.  
Sample temperature outside of warning limits specified  
by SAMP_TEMP_SET variable.  
9
Purge pressure outside of warning limits specified by  
PURGE_PRESS_SET variable.  
Chassis temperature outside of warning limits specified  
by BOX_SET variable.  
WSYNC  
SYNC_OK digital input is off.  
WPHOTOTEMP  
PHOTO TEMP  
WARNING  
Photometer temperature outside of warning limits  
specified by PHOTO_TEMP_SET variable.  
WDYNZERO  
WDYNSPAN  
CANNOT DYN ZERO  
Contact closure zero calibration failed while DYN_ZERO  
was set to ON.  
CANNOT DYN SPAN  
Contact closure span calibration failed while DYN_SPAN  
was set to ON.  
WREARBOARD  
WRELAYBOARD  
WFRONTPANEL  
WANALOGCAL  
REAR BOARD NOT DET  
RELAY BOARD WARN  
FRONT PANEL WARN  
Rear board was not detected during power up.  
Firmware is unable to communicate with the relay board.  
Firmware is unable to communicate with the front panel.  
ANALOG CAL WARNING The A/D or at least one D/A channel has not been  
calibrated.  
045840110 Rev B.3  
238  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Table A-3:  
GFC7000E Test Functions, Revision  
E.0APPENDIX A-3: Warnings and Test Functions, Revision E.0  
Table A-3:  
GFC7000E Test Functions, Revision E.0  
TEST FUNCTION  
NAME1  
MESSAGE TEXT  
DESCRIPTION  
RANGE  
RANGE=50.0 PPM 3  
CO RANGE=50.0 PPM 3, 7  
D/A range in single or auto-range modes.  
RANGE1  
RANGE2  
RANGE1=50.0 PPM 3  
D/A #1 range in dual range mode.  
D/A #2 range in dual range mode.  
D/A #3 range.  
CO RANGE1=50.0 PPM 3, 7  
RANGE2=50.0 PPM 3  
CO RANGE2=50.0 PPM 3, 7  
CO2 RANGE=50.0 PPM 3, 7  
RANGE3  
STABILITY  
STABIL=0.0 PPM 3  
CO/CO2 STB=0.0 PPM 3,7  
Concentration stability (standard deviation based on setting of  
STABIL_FREQ and STABIL_SAMPLES).  
RESPONSE 2  
RSP=0.20(0.00) SEC  
Instrument response. Length of each signal processing loop.  
Time in parenthesis is standard deviation.  
COMEAS  
COREF  
CO MEAS=4125.0 MV  
CO REF=3750.0 MV  
MR RATIO=1.100  
Detector measure reading.  
Detector reference reading.  
Measure/reference ratio.  
Measure/reference ratio during auto-reference.  
Sample pressure.  
MRRATIO  
4, 5  
AUTOZERO  
AZERO RATIO=1.234  
PRES=29.9 IN-HG-A  
PURGE=7.5 PSIG  
SAMPPRESS  
9
PURGEPRESS  
Purge pressure  
8
VACUUM  
VAC=6.8 IN-HG-A  
SAMP FL=751 CC/M  
SAMPLE TEMP=26.8 C  
BENCH TEMP=48.1 C  
WHEEL TEMP=68.1 C  
BOX TEMP=26.8 C  
PHT DRIVE=2500.0 MV  
SLOPE=1.000  
Vacuum pressure.  
6
SAMPFLOW  
Sample flow rate.  
SAMPTEMP  
BENCHTEMP  
WHEELTEMP  
BOXTEMP  
Sample temperature.  
Bench temperature.  
Wheel temperature.  
Internal chassis temperature.  
Photometer temperature.  
PHOTOTEMP  
COSLOPE  
CO slope for current range, computed during zero/span  
calibration.  
7
CO SLOPE=1.000  
COOFFSET  
OFFSET=0.000  
CO offset for current range, computed during zero/span  
calibration.  
7
CO OFFSET=0.000  
7
CO2SLOPE  
CO2 SLOPE=1.000  
CO2 slope for current range, computed during zero/span  
calibration.  
7
CO2OFFSET  
CO2 OFFSET=0.000  
CO2 offset for current range, computed during zero/span  
calibration.  
CO  
CO2  
CO=17.7 PPM 3  
CO2=15.0 % 7  
CO concentration for current range.  
CO2 concentration for current range.  
TESTCHAN  
TEST=1751.4 MV  
Value output to TEST_OUTPUT analog output, selected with  
TEST_CHAN_ID variable.  
CLOCKTIME  
TIME=09:52:20  
Current instrument time of day clock.  
1
The name is used to request a message via the RS-232 interface, as in “T BOXTEMP”.  
2
3
4
5
6
7
8
9
Engineering software.  
Current instrument units.  
M300ES.  
M300EH.  
Except GFC7000EU (APR version).  
M306E.  
Sample pressure or differential pressure flow measurement option.  
GFC7000E.  
045840110 Rev B.3  
239  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A-4: MGFC7000E Signal I/O Definitions, Revision E.0  
APPENDIX A-4: MGFC7000E Signal I/O Definitions, Revision E.0  
Table A-4:  
GFC7000E Signal I/O Definitions, Revision E.0  
SIGNAL NAME  
BIT OR CHANNEL  
NUMBER  
DESCRIPTION  
1 = sync. OK  
SYNC_OK  
0
0 = sync. error  
1–7  
Spare  
Internal outputs, U8, J108, pins 18 = bits 07, default I/O  
ELEC_TEST  
DARK_CAL  
0
1 = electrical test on  
0 = off  
1
1 = dark calibration on  
0 = off  
2–5  
6
Spare  
I2C_RESET  
1 = reset I2C peripherals  
0 = normal  
I2C_DRV_RST  
7
0 = hardware reset 8584 chip  
1 = normal  
Control inputs, U11, J1004, pins 1–6 = bits 0–5, default I/O address 321 hex  
EXT_ZERO_CAL  
0
1
2
0 = go into zero calibration  
1 = exit zero calibration  
EXT_SPAN_CAL  
0 = go into span calibration  
1 = exit span calibration  
REMOTE_RANGE_HI  
0 = remote select high range  
1 = default range  
3–5  
6–7  
Spare  
Always 1  
Control inputs, U14, J1006, pins 16 = bits 05, default I/O  
0–5  
6–7  
Spare  
Always 1  
Control outputs, U17, J1008, pins 18 = bits 07, default I/  
0–7 Spare  
Control outputs, U21, J1008, pins 912 = bits 03, default I/  
0–3 Spare  
Alarm outputs, U21, J1009, pins 112 = bits 47, default I/O  
ST_SYSTEM_OK2  
4
5
6
7
1 = system OK  
0 = any alarm condition or in diagnostics mo  
ST_CONC_ALARM_1  
1 = conc. limit 1 exceeded  
0 = conc. OK  
2 + 8  
ST_CONC_ALARM_2  
1 = conc. limit 2 exceeded  
0 = conc. OK  
2 + 8  
Spare  
A status outputs, U24, J1017, pins 18 = bits 07, default I/  
ST_SYSTEM_OK  
ST_CONC_VALID  
ST_HIGH_RANGE  
ST_ZERO_CAL  
0
1
2
3
0 = system OK  
1 = any alarm condition  
0 = conc. valid  
1 = hold off or other conditions  
0 = high auto-range in use  
1 = low auto-range  
0 = in zero calibration  
1 = not in zero  
045840110 Rev B.3  
240  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A-4: MGFC7000E Signal I/O Definitions, Revision E.0  
SIGNAL NAME  
BIT OR CHANNEL  
NUMBER  
DESCRIPTION  
ST_SPAN_CAL  
4
5
6
7
0 = in span calibration  
1 = not in span  
ST_DIAG_MODE  
0 = in diagnostic mode  
1 = not in diagnostic mode  
ST_CONC_ALARM_1  
0 = conc. limit 1 exceeded  
1 = conc. OK  
2
ST_CONC_ALARM_2  
0 = conc. limit 2 exceeded  
1 = conc. OK  
2
B status outputs, U27, J1018, pins 18 = bits 07, default I/  
2, 3  
ST_AUTO_REF  
0
0 = in auto-reference mode  
1 = not in auto-reference mode  
1–7  
Spare  
Front panel I2C keyboard, default I2C address 4  
MAINT_MODE  
LANG2_SELECT  
SAMPLE_LED  
CAL_LED  
5 (input)  
6 (input)  
0 = maintenance mode  
1 = normal mode  
0 = select second language  
1 = select first language (English)  
8 (output)  
9 (output)  
0 = sample LED on  
1 = off  
0 = cal. LED on  
1 = off  
FAULT_LED  
10 (output)  
14 (output)  
0 = fault LED on  
1 = off  
AUDIBLE_BEEPER  
0 = beeper on (for diagnostic testing only)  
1 = off  
Relay board digital output (PCF8574), default I2C add  
RELAY_WATCHDOG  
WHEEL_HTR  
0
1
2
3
4
5
6
6
7
Alternate between 0 and 1 at least every 5 se  
relay board active  
0 = wheel heater on  
1 = off  
BENCH_HTR  
0 = optical bench heater on  
1 = off  
5
O2_CELL_HEATER  
0 = O2 sensor cell heater on  
1 = off  
CAL_VALVE  
SPAN_VALVE  
0 = let cal. gas in  
1 = let sample gas in  
0 = let span gas in  
1 = let zero gas in  
ZERO_SCRUB_VALV  
0 = open zero scrubber valve  
1 = close  
2
E
SHUTOFF_VALVE  
IR_SOURCE_ON  
0 = energize shutoff valve  
1 = de-energize  
0 = IR source on  
1 = off  
Rear board primary MUX analog inputs  
SAMPLE_PRESSURE  
0
Sample pressure  
Vacuum pressure  
1
1
Purge pressure  
045840110 Rev B.3  
241  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A-4: MGFC7000E Signal I/O Definitions, Revision E.0  
SIGNAL NAME  
BIT OR CHANNEL  
NUMBER  
DESCRIPTION  
CO_MEASURE  
2
3
Detector measure reading  
Detector reference reading  
Temperature MUX  
CO_REFERENCE  
4
SAMPLE_FLOW  
PHOTO_TEMP  
TEST_INPUT_7  
TEST_INPUT_8  
REF_4096_MV  
5
Sample flow  
6
Photometer detector temperature  
Diagnostic test input  
7
8
Diagnostic test input  
9
4.096V reference from MAX6241  
O2 concentration sensor  
Spare  
5
O2_SENSOR  
10  
11  
12  
13  
14  
15  
7
CO2_SENSOR  
CO2 concentration sensor  
Spare  
DAC loopback MUX  
REF_GND  
Ground reference  
Rear board temperature MUX analog input  
Internal box temperature  
Sample temperature  
BOX_TEMP  
SAMPLE_TEMP  
BENCH_TEMP  
WHEEL_TEMP  
TEMP_INPUT_4  
TEMP_INPUT_5  
O2_CELL_TEMP  
0
1
2
3
4
5
6
7
Optical bench temperature  
Wheel temperature  
Diagnostic temperature input  
Diagnostic temperature input  
O2 sensor cell temperature  
Spare  
Rear board DAC MUX analog inputs  
DAC channel 0 loopback  
DAC channel 1 loopback  
DAC channel 2 loopback  
DAC channel 3 loopback  
Rear board analog outputs  
Concentration output #1  
Concentration output #2  
Test measurement output  
Concentration output #3 (CO2)  
DAC_CHAN_0  
DAC_CHAN_1  
DAC_CHAN_2  
DAC_CHAN_3  
0
1
2
3
CONC_OUT_1  
CONC_OUT_2  
TEST_OUTPUT  
0
1
2
3
7
CONC_OUT_3  
1
Hessen protocol.  
M300EH.  
2
3
M300ES.  
M320E.  
O2 option.  
4
5
6
Sample pressure or differential pressure flow measurement option.  
7
M306E.  
8
Factory option enables concentration alarms on relay outputs.  
9
GFC7000E.  
GFC7000E.  
10  
045840110 Rev B.3  
242  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A-5: GFC7000E iDAS Functions, Revision E.0  
APPENDIX A-5: GFC7000E iDAS Functions, Revision E.0  
Table A-5:  
MGFC7000E DAS Trigger Events, Revision E.0  
NAME  
ATIMER  
EXITZR  
EXITSP  
EXITMP  
EXITC2 5  
SLPCHG  
DESCRIPTION  
Automatic timer expired  
Exit zero calibration mode  
Exit span calibration mode  
Exit multi-point calibration mode  
Exit CO2 calibration mode  
Slope and offset recalculated  
CO2 slope and offset recalculated  
Exit diagnostic mode  
5
CO2SLP  
EXITDG  
SOURCW  
Source warning  
1, 2  
AZEROW  
Auto-zero warning  
CONCW1 1, 3, 4  
Concentration limit 1 exceeded  
Concentration limit 2 exceeded  
Sync warning  
1, 3, 4  
CONCW2  
SYNCW  
BNTMPW  
WTEMPW  
STEMPW  
Bench temperature warning  
Wheel temperature warning  
Sample temperature warning  
Sample flow warning  
6
SFLOWW  
SPRESW  
Sample pressure warning  
Purge pressure warning  
4
PPRESW  
BTEMPW  
PTEMPW  
Box temperature warning  
Photometer detector temperature warning  
1
2
3
4
5
6
M300EH.  
M300ES.  
M320E.  
GFC7000E.  
M306E.  
Except GFC7000EU (APR version).  
045840110 Rev B.3  
243  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A-5: GFC7000E iDAS Functions, Revision E.0  
Table A-6:  
NAME  
MGFC7000E iDAS Functions, Revision E.0  
DESCRIPTION  
UNITS  
mV  
DETMES  
DETREF  
RATIO  
Detector measure reading  
Detector reference reading  
mV  
M/R ratio.  
None  
None  
None  
None  
None  
None  
%
SLOPE1  
SLOPE2  
Slope for range #1  
Slope for range #2  
CO2 slope  
5
CO2SLP  
OFSET1  
OFSET2  
CO2OFS  
Offset for range #1  
Offset for range #2  
CO2 offset  
5
1,2  
AZERO  
Auto-zero reading  
M/R  
ZSCNC1  
Concentration for range #1 during zero/span calibration, just before  
computing new slope and offset  
PPM  
ZSCNC2  
Concentration for range #2 during zero/span calibration, just before  
computing new slope and offset  
PPM  
%
5
CO2ZSC  
CO2 concentration during zero/span calibration, just before  
computing new slope and offset  
CONC1  
CONC2  
Concentration for range #1  
Concentration for range #2  
CO2 concentration  
PPM  
PPM  
%
5
CO2CNC  
STABIL  
Concentration stability  
PPM  
°C  
BNTEMP  
WTEMP  
SMPTMP  
Bench temperature  
Wheel temperature  
°C  
Sample temperature  
°C  
6
SMPFLW  
SMPPRS  
Sample flow  
cc/m  
"Hg  
"Hg  
PSIG  
°C  
Sample pressure  
1, 3, 6  
VACUUM  
PRGPRS  
Vacuum pressure  
4
Purge pressure  
BOXTMP  
PHTDRV  
TEST7  
Internal box temperature  
Photometer detector temperature drive  
Diagnostic test input (TEST_INPUT_7)  
Diagnostic test input (TEST_INPUT_8)  
Diagnostic temperature input (TEMP_INPUT_4)  
Diagnostic temperature input (TEMP_INPUT_5)  
Ground reference (REF_GND)  
4096 mV reference (REF_4096_MV)  
Bench temperature control duty cycle  
mV  
mV  
mV  
°C  
TEST8  
TEMP4  
TEMP5  
°C  
REFGND  
RF4096  
BNCDTY  
mV  
mV  
Fraction  
0 = off,  
1 = on  
WHLDTY  
Wheel temperature control duty cycle  
Fraction  
1
2
3
4
5
6
7
M300EH.  
M300ES.  
M320E.  
GFC7000E.  
M306E.  
Except GFC7000EU (APR version).  
The units, including the concentration units, are always fixed, regardless of the current instrument  
units.  
045840110 Rev B.3  
244  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A-6: Terminal Command Designators, Revision E.0  
APPENDIX A-6: Terminal Command Designators, Revision E.0  
Table A-7: Terminal Command Designators, Revision E.0  
COMMAND  
? [ID]  
ADDITIONAL COMMAND SYNTAX  
DESCRIPTION  
Display help screen and commands list  
Establish connection to instrument  
Terminate connection to instrument  
Display test(s)  
LOGON [ID]  
LOGOFF [ID]  
password  
SET ALL|name|hexmask  
LIST [ALL|name|hexmask] [NAMES|HEX]  
name  
Print test(s) to screen  
Print single test  
T [ID]  
CLEAR ALL|name|hexmask  
SET ALL|name|hexmask  
LIST [ALL|name|hexmask] [NAMES|HEX]  
name  
Disable test(s)  
Display warning(s)  
Print warning(s)  
W [ID]  
Clear single warning  
CLEAR ALL|name|hexmask  
ZERO|LOWSPAN|SPAN [1|2]  
ASEQ number  
Clear warning(s)  
Enter calibration mode  
Execute automatic sequence  
Compute new slope/offset  
Exit calibration mode  
C [ID]  
COMPUTE ZERO|SPAN  
EXIT  
ABORT  
Abort calibration sequence  
Print all I/O signals  
LIST  
name[=value]  
Examine or set I/O signal  
Print names of all diagnostic tests  
Execute diagnostic test  
Exit diagnostic test  
LIST NAMES  
ENTER name  
EXIT  
RESET [DATA] [CONFIG] [exitcode]  
PRINT ["name"] [SCRIPT]  
RECORDS ["name"]  
Reset instrument  
D [ID]  
Print iDAS configuration  
Print number of iDAS records  
REPORT ["name"] [RECORDS=number]  
[FROM=<start date>][TO=<end  
date>][VERBOSE|COMPACT|HEX] (Print DAS  
records)(date format: MM/DD/YYYY(or YY)  
[HH:MM:SS]  
Print iDAS records  
CANCEL  
Halt printing iDAS records  
Print setup variables  
LIST  
name[=value [warn_low [warn_high]]]  
name="value"  
Modify variable  
Modify enumerated variable  
Print instrument configuration  
Enter/exit maintenance mode  
Print current instrument mode  
Upload iDAS configuration  
Upload single iDAS channel  
Delete iDAS channels  
V [ID]  
CONFIG  
MAINT ON|OFF  
MODE  
DASBEGIN [<data channel definitions>] DASEND  
CHANNELBEGIN propertylist CHANNELEND  
CHANNELDELETE ["name"]  
The command syntax follows the command type, separated by a space character. Strings in [brackets] are  
optional designators. The following key assignments also apply.  
045840110 Rev B.3  
245  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A-6: Terminal Command Designators, Revision E.0  
Table A-8:  
Terminal Key Assignments, Revision E.0  
TERMINAL KEY ASSIGNMENTS  
ESC  
Abort line  
CR (ENTER)  
Execute command  
Ctrl-C  
Switch to computer mode  
COMPUTER MODE KEY ASSIGNMENTS  
LF (line feed)  
Ctrl-T  
Execute command  
Switch to terminal mode  
045840110 Rev B.3  
246  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model 360E Instruction Manual  
APPENDIX A-6: Terminal Command Designators, Revision E.0  
045840110 Rev B.3  
247  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX B - GFC7000E Spare Parts List  
APPENDIX B - GFC7000E Spare Parts List  
NOTE  
The spare parts list for the Model GFC7000E is currently being compiled.  
For information is spare parts please contact Teledyne Analytical Instruments customer  
service department at  
16830 Chestnut St., City of Industry, Ca. 91748  
Phone: +1 626 961-9221 or 1-626-934-1500.  
Fax: +1 626-961-2538  
045840120 Rev B.3  
249  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
Appendix C - Repair Questionnaire - GFC7000E  
Appendix C - Repair Questionnaire - GFC7000E  
Company: _________________________ Contact Name: _____________________________  
Phone Number: ___________ Fax Number: _____________ Email: ____________________  
Site Address: __________________________________________________________________  
Can we connect to the instrument? If so, provide IP address or modem #:___________________  
Model GFC7000E Serial Number:_________________ Firmware revision: _________________  
The serial number can be found on the back of the instrument, the firmware revision is displayed in the upper left corner of the  
display when pressing SETUP on the front panel (Example: C.3).  
1. List all front panel error/warning messages:_________________________________________  
______________________________________________________________________________  
2. Please complete the following table: (Depending on options installed, not all test parameters  
shown below may be available in your instrument)  
RECORDED  
VALUE  
ACCEPTABLE  
VALUE  
RECORDED  
VALUE  
ACCEPTABLE  
VALUE  
PARAMETER  
PARAMETER  
RANGE  
STABIL  
ppb/ppm 50 ppb - 20 ppm  
SLOPE  
1.0 ± 0.3  
< 250  
ppb 1 ppb with zero  
OFFSET  
mV  
air  
SAMP FL  
cm³/min  
mV  
500 ± 50  
HVPS  
V
5500-900  
PMT SIGNAL  
WITH ZERO  
AIR  
-20 to 150  
ETEST  
mV  
2000 ± 1000  
PMT SIGNAL  
AT  
SPAN GAS  
CONC  
mV  
ppb/ppm  
0-5000  
0-20 000 ppb  
OTEST  
mV  
°C  
2000 ± 1000  
50 ± 1  
NORM PMT  
AT  
mV  
ppb/ppm  
0-5000  
0-20 000 ppb  
RCELL TEMP  
SPAN GAS  
CONC  
UV LAMP  
STR. LGT  
mV 2 000 to 4 000  
BOX TEMP  
PMT TEMP  
°C Ambient + ~5  
ppm 100 ppb/ zero  
°C  
7 ± 2  
air  
DARK PMT  
mV  
mV  
-50 to 200  
-50 to 200  
IZS TEMP  
°C  
50 ± 3  
DARK LAMP  
3. Has the analyzer been checked for leaks? Yes  
No  
For proper flows? Yes  
No  
4. What are the failure symptoms? _________________________________________________  
______________________________________________________________________________  
______________________________________________________ Continue on back if necessary  
5. Which tests have you done trying to solve the problem? _______________________________  
________________________________________________________ Continue on back if necessary  
6. If possible, fax a portion of a strip chart or email a data file to customer service.  
CUSTOMER SERVICE CONTACT INFORMATION: 16830 Chestnut St., City of Industry, Ca. 91748  
PHONE: +1 626-961-9221 or 1-626-934-1500. FAX: +1 626-961-2538.  
You can access and submit an online version of this form at http://www.teledyne-api.com/forms/csformGFC7000E.asp  
048450130 Rev A1  
251  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Notes and further information: _____________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
______________________________________________________________________________  
045150130 Rev A1  
252  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
APPENDIX D - ELECTRONIC SCHEMATICS  
APPENDIX D - ELECTRONIC SCHEMATICS  
Table D-1:  
List of Included Electronic Schematics  
DOCUMENT # DOCUMENT TITLE  
04217  
04216  
04136  
03297  
04070  
03632  
03976  
04259  
04003  
04089  
04468  
Interconnect List - M300E SNs >=100  
Interconnect Drawing - M300E SNs >=100  
PCA, 04135 Rev A, M300E Relay  
PCA, 03296, IR Photodetector Preamp and Sync Demodulator  
PCA, 04069, Motherboard, E-series  
PCA, 03631, 0-20mA driver  
PCA, 03975, Keyboard & Display Driver  
PCA, 04258, Keyboard & Display Driver  
PCA, 04003, Pressure/Flow Transducer Interface  
PCA, 04088, Opto Pickup Interface  
PCA, 04467, Analog Output Series Res  
045840140 Rev A1  
253  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Model GFC7000E Instruction Manual  
User Notes  
04584 Rev A1  
254  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Taylor Blender custard machine User Manual
Tech Craft TV Video Accessories BEL41S User Manual
Technicolor Thomson DVD Player DTH220E User Manual
Technicolor Thomson Network Router 605S User Manual
Toshiba Marine Radio u300 User Manual
Triarch Indoor Furnishings 75161 User Manual
Univex Water Heater FGR 100 User Manual
Velodyne Acoustics Speaker VA 806 907 User Manual
Vertical Communications Telephone 8112N User Manual
ViewSonic Car Video System VS12843 User Manual