SSI America Water Pump 90 2581 REV B User Manual

Binary Solvent Delivery Module  
Dual Series-III Pumps with Serial PC and Voltage Control  
Operator's Manual  
90-2581 rev B  
Scientific Systems, Inc. 349 N. Science Park Road State College, P  
16803  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1. INTRODUCTION  
This operator's manual contains information needed to install, operate, and  
perform minor maintenance on the Binary Solvent Delivery Module. The  
figures below are for reference of items described in the manual.  
Power  
Switch  
“T” With  
Outlet  
Left  
Right  
Drawer  
Drawer  
Control  
Keypads  
Prime Purge  
Valves  
Outlet  
Filters  
Cooling  
Fan  
Power  
Entry  
RS-232  
Serial Ports  
Auxiliary I/O  
Voltage Control, etc.  
1-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1.1  
Description of the Binary Solvent Delivery Module  
The system consists of two Series III high performance metering  
pumps, in a compact package with pulse dampener and an off-line  
pressure transducer. The pumps use a drawer system for easy access to  
internal components for troubleshooting & service. These features  
allow for:  
f Fast & easy setup:  
Two inlet connection (from solvent reservoirs)  
One outlet connection (to injector)  
All connections accessible from front panel  
f Modular:  
Can be added to any system  
f All Stainless steel or PEEK Fluid path, including pumps, valves  
and fittings  
f Modular pump bays for easy replacement and maintenance  
f Self flushing pump heads for extended seal life and reduced  
maintenance  
f Pulse Dampener for reduced pulsation.  
f Very high performance/price ratio  
f Easily user adjustable process set points (flow rate, pressure. etc.)  
via front key pads  
f Digital readout of process parameters  
f User settable upper/lower pressure and temperature limits  
f RS-232 serial PC interface  
f Compact size—Requires only 11 inches of bench space  
The low pulsation flow produced by the reciprocating, single-  
piston pump is achieved by using an advanced rapid-refill cam design,  
programmed stepper motor acceleration, and an internal pulse damper.  
1.1.1 Pump Features  
They include:  
Rapid refill mechanism to reduce pulsation  
Stainless Steel or PEEK™ pumps head  
LED front panel readout of flow rate, pressure and upper/lower  
pressure limits  
Flow adjustment in 0.001 ml increments, from 0.001 to 5.000  
ml/min with a precision of 0.2% RSD (5mL/min heads)  
1-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Flow adjustment in 0.01 ml increments, from 0.01 to 10.00  
ml/min with a precision of 0.2% RSD (10mL/min heads)  
Microprocessor advanced control  
Digital stepper motor design to prevent flow rate drift  
over time and temperature  
Back panel RS232 serial communications port for  
complete control and status monitoring  
Remote analog inputs (e.g. voltage) to control flow rate  
1.1.2 Wetted Pump Materials  
Pump heads, check valve bodies, and tubing are made out Stainless  
Steel or PEEK™. Other materials are synthetic ruby and sapphire  
(check valve internals and piston).  
1.1.3 Self-Flushing Pump Head  
Self-flushing pump heads provide continuous washing of the piston  
surface without the inconvenience of a manual flush or gravity feed  
arrangement. The self-flushing pump head uses a diaphragm and  
secondary set of check valves to create a continuous and positive flow  
in the area behind the high pressure pump seal. The flushing solution  
washes away any buffer salts that have precipitated onto the piston.  
If not removed, these precipitates can abrade the high pressure seal  
and cause premature seal failure, leakage, and can possibly damage the  
pump.  
SAPPHIRE  
PISTON  
MOBILE  
PHASE  
PISTON  
MOVEMENT  
FLOW OF  
FLUSH SOLUTION  
PRIMARY  
SECONDARY  
HIGH-PRESSURE SEAL  
SELF-FLUSH SEAL  
Figure 1-1. Self-Flushing Pump Head  
1-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1.1.4 Self-Flush and Seal Life  
It is recommended that the Self Flush feature be used to improve  
seal life in a number of applications. In particular, (as stated above) if  
pumping Buffers, Acids/Bases or any inorganic solution near  
saturation, the pump should utilize the Self Flush feature. With every  
piston stroke, an extremely thin film of solution is pulled back past the  
seal. If this zone is dry (without use of Self Flush), then crystals will  
form with continuous operation, which will ultimately damage the  
seal.  
Another application where Self Flush is highly recommended is  
when pumping Tetrahydrofuran (a.k.a. THR, Diethylene Oxide) or  
other volatile solvents such as acetone (Note: THF and most solvents  
are compatible only with all-Stainless Steel systems. THF will attack  
PEEK). Volatile solvents will dry rapidly behind the seal (without the  
use of Self Flush), which will dry and degrade the seal.  
IPA, Methanol, 20% IPA/water mix or 20% Methanol/water mix  
are good choices for the flush solution. Consult the factory for  
specific recommendations.  
1-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
1.2  
Specifications for the Binary Solvent Delivery Modules  
Flow Rates ............0.000 to 5.000 mL/min for 5mL/min heads  
0.00 to 10.00 mL/min for 10mL/min heads  
Pressure ................0 to 6,000 p.s.i. for SS pump heads,  
0 to 5,000 p.s.i. for PEEK™ heads,  
Pressure Accuracy… .± 1% of full-scale pressure  
Pressure Zero Offset. .± 2 p.s.i.  
Flow Accuracy........± 2% for a flow rate of 0.20 mL/min and  
above.  
Flow Precision .......0.2% RSD  
Dimensions ............9" high x 10.5" wide x 18.5" deep  
Weight ...................34 lb  
Power ....................90-260 VAC, 50-60 Hz, 45W (The main  
voltage supply shall not exceed ±10%)  
Environmental……..Indoor use only  
Altitude……………..2000 M  
Temperature………10 to 30° C  
Humidity……………20 to 90% Relative humidity  
Remote Inputs . . . . RS-232, Voltage, Frequency  
1-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. INSTALLATION  
2.1  
Unpacking and Inspection  
Prior to opening the shipping container, inspect it for damage or  
evidence of mishandling. If it has been damaged or mishandled, notify  
the carrier before opening the container. Once the container is opened,  
inspect the contents for damage. Any damage should be reported to the  
carrier immediately. Save the shipping container. Check the contents  
against the packing list.  
2.2  
Location/Environment  
The preferred environment for the Binary Solvent Delivery  
Module is normal laboratory conditions. The area should be clean and  
have a stable temperature and humidity. The specific temperature and  
humidity conditions are 10 to 30 °C and 20% to 90% relative  
humidity. The instrument should be located on a stable flat surface  
with surrounding space for ventilation and the necessary electrical and  
fluid connections.  
2.3  
Fluid Connections & Priming  
There are only four connections to be made.  
1. Connect left drawer pump solvent line supplied  
2. Connect right drawer pump solvent line supplied  
3. Connect line to injection valve from outlet “T”  
4. Next, install the self flush. Connect the self flush inlet and  
outlet (opaque) tubing as shown on the the next page.  
Details on the proper installation of tubing and priming of the  
pump and self flush is shown on the next page.  
3-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SELF-FLUSH  
PUMP  
Connect wash solution inlet and  
outlet tubing  
Connect pump inlet tubing as  
shown.  
(opaque) to the flush housing as  
shown.  
Mak
posi
Make sure ferrule is in the  
correct  
p
Sc
Ins
int
Ensure inlet line filter is  
submersed into solvent.  
Attach syringe to Prime-Purge  
valve.  
Open Prime-Purge valve by  
t
t
D
D
b
3-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.4  
Electrical Connection  
The system utilizes Universal Switching Power Supply, and will  
accept voltages from 90 – 260 VAC, 50-60 Hz.  
WARNING: Do not bypass the safety ground connection as  
a serious shock hazard could result.  
2.5  
Solvent Preparation  
Proper solvent preparation will prevent a great number of pumping  
problems. The most common problem is bubble formation, which may  
affect the flow rate consistency. Aside from leaky fittings, the problem  
of bubble formation arises from two sources: solvent out-gassing and  
cavitation. Filtration of HPLC solvents is also required.  
2.5.1 Solvent Out-gassing and Sparging  
Solvent out-gassing occurs because the mobile phase contains  
dissolved atmospheric gases, primarily N and O . These dissolved  
2
2
gases may lead to bubble formation and should be removed by  
degassing the mobile phase before or during use. The best practical  
technique for degassing is to sparge the solvent with standard  
laboratory grade (99.9+%) helium. Helium is only sparingly soluble  
in HPLC solvents, so other gases dissolved in the solvent diffuse into  
the helium bubbles and are swept from the system. Solvent filtration is  
not an effective alternative to helium degassing.  
It is recommended that you sparge the solvent vigorously for 10 to  
15 minutes before using it. Then maintain a trickle sparge during use  
to keep atmospheric gases from dissolving back into the mobile phase.  
The sparged solvent must be continually blanketed with helium at 2 to  
3 psi. Non- blanketed, sparged solvents will allow atmospheric gases  
to dissolve back into the mobile phase within four hours.  
Solvent mixtures using water and organic solvents (like methanol or  
acetonitrile) hold less dissolved gas than pure solvents. Sparging to  
reduce the amount of dissolved gas is therefore particularly important  
when utilizing solvent mixture.  
Even with sparging some out-gassing may be occur. A back  
pressure regulator installed after the detector flow cell will help  
prevent bubbles from forming and thus limit baseline noise.  
WARNING: Always release pressure from the pump slowly. A rapid  
pressure release could cause the pulse damper diaphragm to rupture.  
3-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.5.2 Cavitation  
Cavitation occurs when inlet conditions restrict the flow of solvent  
and vapor bubbles are formed during the inlet stroke. The key to  
preventing cavitation is to reduce inlet restrictions. The most common  
causes of inlet restrictions are crimped inlet lines and plugged inlet  
filters. Inlet lines with tubing longer than 48" (120 cm) or with tubing  
of less than 0.085" (2 mm) ID may also cause cavitation.  
Placing the solvent reservoirs below the pump level also promotes  
cavitation. The optimal location of the reservoirs is slightly above the  
pump level, but it is adequate to have them on the same level as the  
pump.  
2.5.3 Filtration  
Solvent filtration is good practice for the reliability of the Binary  
Solvent Delivery Module and other components in a HPLC system.  
Solvents should always be filtered with a 0.5 micron filter prior to use.  
This ensures that no particles will interfere with the reliable operation of  
the piston seals and check valves. Solvents in which buffers or other salts  
readily precipitate out will need to be filtered more often. After filtration,  
the solvents should be stored in a closed, particulate-free bottle.  
2.5.4 Initial system pressurization (Daily)  
IMPORTANT: To maximize accuracy at all pressures and flows  
this pump contains a pulse damper for each solvent used. It is  
important that the pump be brought to pressure with both pumps  
running to insure that each pulse damper has the correct solvent in it.  
It is recommended that the flow be set to the operational flow with an  
equal mixture of the two solvents. Once the system has reached a  
constant pressure the composition should be set to the initial  
conditions.  
2.6  
Instrument Installation  
2.6.1 Mobile Phase Reservoirs  
The mobile phase reservoir should be placed at the same level or  
slightly higher than the pump, never below the pump, and the inlet  
tubing should be as short as practical. These steps minimize pressure  
losses on the inlet side of the pump during refill and help to avoid  
bubble formation. These steps are particularly important when using  
high vapor pressure solvents (hexane, methylene chloride, etc.).  
Mobile phases should be degassed, filtered and covered. (See Section  
2.4.)  
3-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.6.2 Self-Flush Solution  
Self-flush heads require 250-500 mL of flushing solution. See  
section 1.1.4 for self-flush solution recommendations. A pH indicator  
that will indicate the concentration of salts in the solution is  
recommended as a reminder to change the solution. This flush solution  
should be replaced with a fresh solution weekly to avoid frequent  
pump maintenance.  
2.6.3 Inlet Tubing and Filters  
The table below shows the inlet tubing and filter used in the Binary  
Solvent Delivery Module. All inlet lines are supplied in a 30" (76 cm)  
length and are made of a fluoropolymer material.  
2.6.4 Priming the Pump and the Flushing Lines  
Connect a syringe to the outlet tubing. Run the pump at a flowrate  
of 3 to 5 ml/min. Prime the pump by pulling mobile phase and any air  
bubbles through the system and into the syringe (a minimum of 20  
ml).  
To prime the flush lines for a self-flush head, simply place the inlet  
line in the flush solution and connect a syringe to the outlet line and  
apply suction until the line is filled with flush solution. Place the  
outlet line in the flush solution. Secure both flush lines in the flush  
solution container so they stay immersed during pump operation.  
2.6.6 Long Term Pressure Calibration Accuracy  
This note applies if your pump is equipped with an electronic  
pressure transducer. The transducer has been zeroed and calibrated at  
the factory. Over the life of the pump, some drift may occur. For  
example, it is typical for the zero to drift < 10 p.s.i. after about 1 year  
of operation (i.e., with no back pressure on the pump a reading of 1-9  
p.s.i. may be displayed). A similar drift may also occur at higher  
pressures, and are typically less than 1% (e.g. <50 p.s.i. at 6,000 p.s.i.  
back pressure).  
If pressure calibration and/or drift is a concern, consult the factory.  
The pump can be shipped back to SSI for recalibration. Alternatively,  
written calibration and zero-reset procedures are available. Consult the  
factory to receive these instructions.  
3-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.7  
Preparation for Storage or Shipping  
2.7.1 Isopropanol Flush  
Disconnect the outlet tubing from the pump. Place the inlet filter in  
isopropanol. Use a syringe to draw a minimum of 50 ml through the  
pump. Pump a minimum of 5 ml of isopropanol to exit. Leave the inlet  
tubing connected to the pump. Place the inlet filter in a small plastic  
bag and attach it to the tubing with a rubber band. Plug the outlet port  
with the shipping plug or leave a length of outlet tubing on the pump  
or cover the outlet port with plastic film.  
2.7.2 Packaging for Shipping  
CAUTION: Reship in the original carton, if possible. If the original carton is  
not available, wrap the pump in several layers of bubble wrap and cushion the  
bottom, top, and all four sides with 2" of packaging foam. Although heavy, an  
HPLC pump is a delicate instrument and must be carefully packaged to  
withstand the shocks and vibration of shipment.  
3-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. OPERATION  
3.1  
Pump Front Panel Controls and Indicators  
3.1.1 Control Panel  
3.1.1.1 Digital Display  
The 3-digit display shows the pump flow rate (mL/min), system  
pressure (psi), or the set upper or lower pressure limit (psi) when  
operating. Choice of display is selected with the MODE key.  
3.1.1.2 Digital Display Pump Keypad  
When pressed, this button increases the flow rate.  
When pressed, this button decreases the flow rate.  
Use this button to cycle through the four display modes: flow rate,  
pressure, upper pressure limit, or lower pressure limit. A status LED  
to the right of the digital display indicates which mode is active.  
MODE  
RUN  
STOP  
When pressed, this button alternately starts and stops the pump.  
Fast And Slow Button Repeat On The Up And Down Arrow Buttons:  
If the UP-ARROW or DOWN-ARROW button is held down for more  
than approximately one half of a second, the button press will repeat at  
a slow rate of approximately 10 times a second. Once slow button  
repeat has begun, fast button repeat can be initiated by using a second  
finger to press down the second arrow button. During fast button  
repeat, the button press will repeat at a rate of approximately 100  
times a second. Switching back and forth between repeat speeds can  
be accomplished by pressing and releasing the second arrow button  
while keeping the first arrow button held down.  
3-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3.1.1.3 Status LEDs  
Q ml/min  
When lit, the digital display shows flow rate in mL/min.  
Q PSI  
When lit, the digital display shows system pressure in psi.  
When lit, the display shows the user-set upper pressure limit in psi.  
When lit, the display shows the user-set lower pressure limit in psi.  
Lights to indicate that the pump is running.  
Q HI PR  
Q LO PR  
Q RUN  
Q FAULT  
Lights when a fault occurs and stops the pump.  
3.1.1.2 Power-up Configuration  
Non-volatile Memory Reset: If the pump is operating erratically,  
there is the possibility that the memory has been corrupted. To reset  
the memory and restore the pump to it's default parameters, press and  
hold the UP-ARROW button when the power is switched on. Release  
the button when the display reads "rES". The parameters stored in  
non-volatile memory, i.e., the flowrate, the pressure compensation, the  
voltage/frequency select, the lower pressure limit, and the upper  
pressure limit will be set to the factory default values. The head type  
setting is the only parameter not changed by the non-volatile memory  
reset function. If the firmware is upgraded to a newer version, a non-  
volatile memory reset will automatically occur the first time the power  
is switched on.  
3.1.1.5 Power-Up Tests  
Display Software Version Mode: The software version can be  
displayed during power-up by pressing and holding the RUN/STOP  
and the UP-ARROW buttons when the power is switched on. Release  
the buttons when the display reads "UEr". The decimal point number  
displayed on the display is the software version. To exit this mode,  
press the RUN/STOP button.  
3-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Align Refill Switch Mode: The signal that initiates the refill phase  
can be displayed during power-up by pressing and holding the PRIME  
and the UP-ARROW buttons when the power is switched on. Release  
the buttons when the display displays "rFL". When the slotted disk  
allows the light beam to pass from the emitter to the detector on the  
slotted optical switch a pulse will be generated which signals the  
beginning of refill. When this pulse occurs the three horizontal  
segments displayed at the top of the display will turn off and the three  
horizontal segments at the bottom of the display will turn on. To exit  
this mode, press the RUN/STOP button.  
Serial Port Loopback Test Mode: If an external device will not  
communicate to the pump via the serial port, the serial port loopback  
test can be used to verify that the serial port is functioning properly.  
During power-up press and hold the UP-ARROW and the DOWN-  
ARROW buttons when the power is switched on and then release the  
buttons. The display must display "C00" for the first half of the test to  
pass. Plug in the serial port loop back plug (A modular plug with pins  
2 & 5 jumpered together and pins 3 & 4 jumpered together.). The  
display must read "C11" for the second half of the test to pass. To exit  
this mode, press the RUN/STOP button.  
3.5  
Rear Panel Remote Input  
An RS-232 modular jacks are provided on the back panel. A  
computer, with appropriate software, can be used as a remote  
controlling device for pump operation via this connection.  
See Appendix A for details on connection and operation.  
3.6  
Symbols  
The following symbols may appear on back panel of the unit:  
Caution: To avoid chemical or electrical hazards, always observe safe  
laboratory practices while operating this equipment.  
Caution: To avoid electrical shock and possible injury, remove the power  
cord from the back panel of this equipment before performing any  
type of service procedures.  
Note: The user shall be made aware that, if equipment is used in a  
manner not specified by the manufacturer, the protection provided by  
the equipment may be impaired.  
3-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. THEORY OF OPERATION  
4.1  
Pump Mechanical Operation  
4.1.1 Liquid System Flow Path  
The flow path of the Binary Solvent Delivery Module starts at the  
inlet reservoir filter passes through the inlet check valve, then through  
the pump head, and finally exits through the outlet check valve.  
4.1.2 Pump Cycle  
The pump cycle consists of two phases, the pumping phase, when  
fluid is metered out of the pump at high pressure, and the refill phase,  
when fluid is rapidly drawn into the pump.  
During the pumping phase, the pump piston moves forward at a  
programmed speed; this results in a stable flow from the pump. The piston  
is driven by a linear rapid refill cam which is belt driven by the motor.  
At the end of the pumping phase, the pump enters the refill phase.  
The piston quickly retracts, refilling the pump head with solvent, and  
the piston begins to move forward again as the pumping phase begins.  
The motor speed is increased during refill to reduce refill time and  
to pre-compress the solvent at the beginning of the pumping phase.  
Since the output flow completely stops during refill, an optional,  
external pulse damper is necessary for applications requiring  
extremely low pulsation levels.  
For optimal operation of the check valves, a back-pressure of at least  
25 psi is required. Operating at lower pressures can lead to improper  
seating of the valves and subsequently inaccurate flow rates.  
4.1.3 Pulse Damping  
The diaphragm-type pulse damper (inside the pump drawer)  
consists of a compressible fluid (isopropanol) held in an isolated  
cavity by an inert but flexible diaphragm. During the pumping phase  
of the pump cycle, the fluid pressure of the mobile phase displaces the  
diaphragm, compressing the fluid in the cavity and storing energy.  
During the pump refill phase the pressure on the diaphragm is reduced  
and the compressed fluid expands, releasing the energy it has stored.  
This helps to stabilize flow rate and pressure. The amount of mobile  
phase in contact with the pulse damper is small, only 0.25 mL at 2,500  
psi, and the geometry used insures that the flow path is completely  
swept, so solvent “memory effects” are virtually eliminated.  
To be effective, the pulse damper requires a backpressure of  
approximately 500 psi or greater.  
4-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4.2  
Electronic Control  
4.2.1 Microprocessor Control  
The pump is controlled by hybrid microprocessor circuitry which  
(1) provides control signals to the motor drive circuitry, (2) interfaces  
with the keyboard/display, (3) receives signals from the refill flag, and  
(4) provides external input/output (RS-232) interfacing. Firmware  
programming is stored in an EPROM.  
An eccentric cam provides refill in a fraction of the full cam cycle.  
The remaining revolution of the cam provides piston displacement for  
outward flow of the mobile phase. In addition to the rapid refill  
characteristics of the drive, the onset of refill is detected by an infrared  
optical sensor. The microprocessor changes the refill speed of the  
motor to an optimum for the set flow rate. At 1ml/min, the refill rate is  
more than five times faster than if the motor operated at constant  
speed. The optimum refill rate minimizes the resulting pulsation while  
avoiding cavitation in the pump head.  
The flow rate of any high pressure pump can vary depending on  
the operating pressure and the compressibility of the fluid being  
pumped. The Binary Solvent Delivery Module is calibrated at 1000 psi  
using an 80:20 mixture of water and isopropanol.  
4.2.2 DC Power Supply  
Power for the pump is provided by a switching power supply  
which accepts voltages from 90 – 240 VAC. Output is 24 VDC for the  
pumps and heater. A switching 5 VDC supply is also provided to  
power control and display circuits.  
4.2.3 Remote Interfacing  
RS-232C modular jacks are provided on the back panel. See  
Appendix A for information on operation via this connection.  
4.2.4 Motor Stall Detector  
The motor can stall and create a loud buzzing sound if the flow  
path connected to the pump's outlet becomes plugged, if the pressure  
exceeds the maximum pressure rating of the pump, or if the  
mechanism jams. If a motor stall occurs, the electrical current being  
supplied to the motor is turned off and the fault light is turned on.  
The Motor Stall Detector is enabled or disabled during power-up by  
pressing and holding the RUN/STOP and the PRIME buttons while  
the power is switched on. Release the buttons when the display  
displays "SFE". To enable the Motor Stall Detector press the UP-  
ARROW button and the display will display "On". To disable the  
4-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Motor Stall Detector press the DOWN-ARROW button and the  
display will display "OFF". To exit this mode and store the current  
setting in non-volatile memory, press the RUN/STOP button.  
The Motor Stall Detector uses a timer to determine if the camshaft has  
stopped turning or if the refill switch is defective. The timer begins  
timing after the pump accelerates or decelerates to its set point flow rate.  
If the Motor Stall Detector has been enabled, and the camshaft stops  
turning or the refill switch stops operating, the fault will be detected  
between the time it takes to complete 1 to 2 pump cycles. One  
revolution of the camshaft produces a delivery phase and a refill phase.  
The fault is canceled by using one of the following methods: (1) by  
pressing the RUN/STOP button on the front panel, (2) by sending a  
stop command "ST" via the serial communications port on the back  
panel, or (3) by connecting the PUMP-STOP input to COM on the  
back panel, or removing the connection between the PUMP-RUN  
input and COM if the PUMP-STOP input is permanently jumpered to  
COM on the back panel. Note: the PUMP-RUN, PUMP-STOP, and  
COM are an option and do not exist on the standard pump.  
4-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. MAINTENANCE  
Cleaning and minor repairs of the Binary pump can be performed  
as outlined below.  
NOTE: Lower than normal pressure, pressure variations, or  
leaks in the pumping system can all indicate possible  
problems with the piston seal, piston, or check valves.  
Piston seal replacement could be necessary after 1,000  
hours of running time. See Section 5.2.3.  
5.1  
Filter Replacement  
5.1.1 Inlet Filters  
Inlet filters should be checked periodically to ensure that they are  
clean and not restricting flow. A restriction could cause cavitation and  
flow loss in the pump. Two problems that can plug an inlet filter are  
microbial growth and impure solvents. To prevent microbial growth,  
use at least 10-20% organic solvent in the mobile phase or add a  
growth-inhibiting compound. If you pump 100% water or an aqueous  
solution without any inhibitors, microbes will grow in the inlet filter  
over time, even if you make fresh solution every day. Always use well  
filtered, HPLC grade solvents for your mobile phase.  
5.2  
Changing Pump Heads  
5.2.1 Removing a Pump Head  
CAUTION: The sapphire piston is fragile. Twisting the  
pump head during removal can cause the piston to break.  
Closely follow instructions during head removal and  
replacement to avoid breakage.  
As a guide to pump head assembly, the standard pump heads are  
shown in Figures 5-1 through 5-4. All of the Binary pump heads have  
a similar arrangement.  
1. Turn OFF the power to the Binary pump.  
2. Remove the inlet line and filter from the mobile phase reservoir. Be  
careful not to damage the inlet filter or crimp the Teflon™ tubing.  
7-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CAUTION: Check that the Allen nuts at the front of the  
pump head are secure before removing any tubing from the  
pump head.  
3. Remove the inlet line from the inlet check valve.  
4. Remove the outlet line from the outlet check valve.  
5. Remove inlet and outlet self-flush check valves.  
6. Momentarily turn ON the Binary pump and quickly turn OFF the  
power upon hearing the refill stroke. This reduces the extension of the  
piston and decreases the possibility of piston breakage.  
7. Unplug the power cord.  
8. Carefully remove the two Allen nuts at the front of the pump head.  
CAUTION: Use care when removing the pump head.  
Twisting the pump head can cause the piston to break.  
9. Carefully separate the pump head from the pump. Move the pump  
head straight out from the pump and remove it from the piston. Be  
careful not to break or damage the piston. Also remove the seal and  
seal backup washer from the piston if they did not stay in the pump  
head.  
10. Carefully separate the flush housing from the pump. Move the flush  
housing straight out from the pump and remove it from the piston. Be  
careful not to break or damage the piston. Also remove the self-flush  
seal from the piston if it did not stay in the flush housing.  
7-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OUTLET FLUSHING CHECK VALVE  
O-RING  
OUTLET CHECK VALVE  
PISTON  
SEAL BACKUP PLATE  
SELF-FLUSH SEAL  
PUMP HEAD  
SELF-FLUSH HOUSING  
INLET FLUSHING CHECK VALVE  
SEAL BACKUP WASHER  
SEAL  
INLET CHECK VALVE  
Figure 5-1. Stainless Steel Self-Flushing Pump Head Assembly  
HOLE PLUG  
OUTLET CHECK VALVE  
PISTON  
SEAL BACKUP PLATE  
NON-FLUSH GUIDE BUSHING  
PUMP HEAD  
SELF-FLUSH HOUSING  
O-RING  
SEAL BACKUP WASHER  
SEAL  
INLET CHECK VALVE  
Figure 5-2. Stainless Steel Non-Self-Flushing Pump Head Assembly  
7-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
OUTLET FLUSHING  
CHECK VALVE  
O-RING  
OUTLET CHECK VALVE  
PUMP HEAD  
PISTON  
SEAL BACKUP PLATE  
SELF-FLUSH SEAL  
SELF FLUSH HOUSING  
INLET FLUSHING CHECK VALVE  
SEAL BACKUP WASHER  
SEAL  
INLET CHECK VALVE  
Figure 5-3. Bioclean (PEEK™) Self-Flushing Pump Head Assembly  
HOLE PLUG  
OUTLET CHECK VALVE  
PISTON  
SEAL BACKUP PLATE  
NON-FLUSH GUIDE BUSHING  
PUMP HEAD  
SELF-FLUSH HOUSING  
O-RING  
SEAL BACKUP WASHER  
SEAL  
INLET CHECK VALVE  
Figure 5-4. Bioclean (PEEK™) Non-Self-Flushing Pump Head Assembly  
7-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.2.2 Cleaning the Pump Head Assembly  
NOTE: If you choose to remove the piston seal or self-flush  
seals, you should have a new set on hand to install after  
cleaning. It is not recommended that you reinstall used  
piston or self-flush seals since they are likely to be  
scratched and damaged during removal and would not  
provide a reliable seal if reused. If you decide to remove the  
seals, use only the flanged end of the plastic seal removal  
tool supplied with the seal replacement kit and avoid  
scratching the sealing surface in the pump head. See  
Section 5.2.3 for seal replacement instructions.  
1. Inspect the piston seal cavity in the pump head. Remove any foreign  
material using a cotton swab, or equivalent, and avoid scratching the  
sealing surfaces. Repeat for the self-flush housing. Be sure no fibers  
from the cleaning swab remain in the components.  
2. The pump head, check valves, and self-flush housing may be further  
cleaned using a laboratory grade detergent solution in an ultrasonic  
bath for at least 30 minutes, followed by rinsing for at least 10 minutes  
in distilled water. Be sure that all particles loosened by the above  
procedures have been removed from the components before re-  
assembly.  
CAUTION: When cleaning check valves, be sure that the  
ball is not against the seat in the ultrasonic bath. This may  
destroy the precision matched sealing surface and the valve  
will not check.  
CAUTION: If removing the check valves, keep them in the  
orientation shown below the entire time they are not  
installed in the pump head. The assemblies may fall apart,  
parts may be lost, and they may not operate properly when  
re-assembled.  
7-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. If the check valves have been removed, tighten the check valves on  
stainless steel pumps to 75 inch-pounds or enough to seal at maximum  
pressure. Do not exceed maximum torque. For Bioclean (PEEK™)  
pumps, tighten each check valve firmly by hand. Do not go ¼ turn past  
finger tight.  
NOTE: The inlet check valve has a larger opening (1/4"-28,  
flat-bottom seat) for the 1/8" inlet tubing; the outlet check  
valve has a smaller opening (#10-32, cone seat) for the  
1/16" outlet tubing. The inlet check valve must be connected  
at the larger opening in the pump head. See Figure 5-5.  
If the piston and flushing seals have been removed, insert new  
seals as described in Section 5.2.3, then continue with Section 5.2.5 to  
replace the pump head.  
7-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
RING  
SINGLE  
INLET HOLE  
INLET CHECK VALVE (SS)  
INLET CHECK VALVE (PEEK)  
MULTIPLE  
OUTLET HOLES  
RING  
OUTLET CHECK VALVE (PEEK)  
OUTLET CHECK VALVE (SS)  
OUTLET  
SELF-FLUSH  
CHECK VALVE  
INLET  
SELF-FLUSH  
CHECK VALVE  
TRANSPARENT  
WASHER  
CROSS BALL  
RETAINER  
Figure 5-5. Check Valves  
5.2.3 Replacing Piston Seals  
Lower than normal pressure, pressure variations, and leaks in the  
pumping system can all indicate possible problems with the piston  
seal. Depending on the fluid or mobile phase used, piston seal  
replacement is often necessary after 1000 hours of running time.  
Each replacement seal kit contains one seal, one backup washer,  
one self-flush seal, one non-flush guide bushing, two seal  
insertion/removal tools, and a pad to clean the piston when changing  
the seal.  
7-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.2.3.1 Removing the Seals  
1. Remove the pump head as described in Section 5.2.1. Use caution so  
as not to damage the sapphire piston.  
2. Insert the flanged end of the seal insertion/removal tool into the seal  
cavity on the pump head. Tilt it slightly so that flange is under the seal  
and pull out the seal.  
CAUTION: Using any other “tool” will scratch the finish.  
3. Repeat the procedure for the low-pressure seal in the flush housing.  
4. Inspect, and if necessary, clean the pump head as described in Section  
5.2.2.  
5.2.3.2 Cleaning the Piston  
1. Once the pump head and self-flush housing are removed, gently  
remove the seal back-up plate by using either a toothpick or small  
screwdriver in the slot on top of the pump housing.  
2. Grasp the metal base of the piston assembly so that you avoid exerting  
any side load on the sapphire rod, and remove the piston from the slot  
in the carrier by sliding it up.  
3. Use the scouring pad included in the seal replacement kit to clean the  
piston. Gently squeeze the piston within a folded section of the pad  
and rub the pad along the length of the piston. Rotate the piston  
frequently to assure the entire surface is scrubbed. Do not exert  
pressure perpendicular to the length of the piston, as this may cause  
the piston to break. After scouring, use a lint-free cloth, dampened  
with alcohol, to wipe the piston clean.  
4. Grasp the metal base of the piston assembly, and insert it into the slot  
in the piston carrier until it bottoms in the slot.  
5.2.3.3 Replacing the Seals  
1. Place a high-pressure replacement seal on the rod-shaped end of the  
seal insertion/removal tool so that the spring is visible when the seal is  
fully seated on the tool. Insert the tool into the pump head so that the  
open side of the seal enters first, facing the high pressure cavity of the  
pump head. Be careful to line up the seal with the cavity while  
inserting. Then withdraw the tool, leaving the seal in the pump head.  
When you look into the pump head cavity, only the polymer portion of  
the seal should be visible.  
7-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Place a self-flush replacement seal on the seal insertion/removal tool  
so that the spring in the seal is visible when the seal is on the tool. As  
in the previous step, insert the tool and seal into the seal cavity on the  
flushing housing, taking care to line up the seal with the cavity, and  
then withdraw the tool. When the seal is fully inserted only the  
polymer part of the seal will be visible in the seal cavity.  
3. Place seal back-up washer over the high-pressure seal. Place seal  
back-up plate back into pump housing if it was removed. Orientation  
is not important in these cases.  
4. Attach the pump head as described in Section 5.2.5. Use caution so as  
not to damage the sapphire piston.  
5. Condition the new seal as described in Section 5.3.  
5.2.4 Changing the Piston  
1. Remove the pump head as described in Section 5.2.1. Use caution so  
as not to damage the sapphire piston.  
2. Grasp the metal base of the piston assembly so that you avoid exerting  
any side load on the sapphire rod, and remove the piston from the slot  
in the carrier by sliding it up.  
3. Grasp the metal base of the replacement piston assembly, and insert it  
into the slot in the piston carrier until it bottoms in the slot.  
4. Attach the pump head as described in Section 5.2.5. Use caution so as  
not to damage the sapphire piston.  
7-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.2.5 Replacing the Pump Head  
CAUTION: The sapphire piston is fragile. Twisting the  
pump head during replacement can cause the piston to  
break. Closely follow instructions during head removal and  
replacement to avoid breakage.  
1. Make sure that the inlet valve is on the bottom and the outlet valve is  
on the top. Carefully align the self-flush housing and gently slide it  
into place on the pump. If misalignment with the piston occurs, gently  
push up on the piston holder.  
2. Line up the pump head and carefully slide it into place. Be sure that  
the inlet valve is on the bottom and the outlet valve is on the top. Do  
not force the pump head into place.  
3. Finger tighten both knurled nuts into place. To tighten firmly,  
alternately turn nuts 1/4 turn while gently wiggling the pump head to  
center it.  
CAUTION: Use care when replacing the pump head.  
Twisting the pump head can cause the piston to break.  
4. Re-attach the inlet and outlet lines. Reconnect the self-flush lines and  
fittings to the self-flush check valves. Change the flushing solution.  
5.3  
Conditioning New Seals  
NOTE: Use only organic solvents to break-in new seals.  
Buffer solutions and salt solutions should never be used to  
break-in new seals.  
Using a restrictor coil or a suitable column, run the pump with a  
50:50 solution of isopropanol (or methanol) and water for 30 minutes  
at the back pressure and flow rate listed under PHASE 1 below and  
according to the pump head type. Then run the pump for 15 minutes at  
a back pressure and flow rate listed under PHASE 2 below.  
PHASE 1 (30 min.)  
PHASE 2 (15 min.)  
Pump Head Type  
Pressure  
Flow Rate  
Pressure  
Flow Rate  
10 mL  
1000 psi  
< 3 mL/min.  
3000-4000 psi  
3-4 mL/min.  
(SS or PEEK)  
5 mL  
1000 psi  
< 2.5 mL/min.  
1500 psi  
< 5 mL/min.  
(SS or PEEK)  
7-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.4  
Check Valve Cleaning and Replacement  
Many check valve problems are the result of small particles  
interfering with the operation of the check valve. As a result, most  
problems can be solved by pumping a strong solution of liquid,  
laboratory grade detergent through the check valves at a rate of  
1 mL/min for one hour. After washing with detergent, pump distilled  
water through the pump for fifteen minutes. Always direct the output  
directly to a waste beaker during cleaning. If this does not work, the  
check valve should be replaced.  
CAUTION: When removing the check valves, keep them in  
the orientation shown below the entire time they are not  
installed in the pump head. The assemblies may fall apart,  
parts may be lost, and they may not operate properly when  
re-assembled.  
5.5  
Pulse Damper Replacement  
5.5.1 Removing the Pulse Damper  
WARNING: There are potentially lethal voltages inside the  
pump case. Disconnect the line cord before removing the  
cover. Never bypass the power grounds.  
1. Make certain that the system has been depressurized. Unplug the power  
cord and remove the cover.  
2. Disconnect the tubing from the pulse damper.  
3. Disconnect the transducer from the circuit board.  
4. Remove the four screws that secure the pulse damper from the underside  
of the pump.  
5. Remove the pulse damper.  
5.5.2 Pulse Damper Refurbishing  
Refurbishing the pulse damper is a time-consuming  
procedure. You may want to return the pulse damper to have it  
7-11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
rebuilt. Do not attempt to refill or refurbish the pulse damper  
until you have a refurbishing kit. Instructions are furnished with  
the kit.  
5.5.3 Pulse Damper Installation  
1. Position the pulse damper, aligning it with the four mounting holes in the  
bottom of the cabinet. The pressure transducer should be pointed toward the  
rear of the cabinet.  
2. From the underside of the pump cabinet, tighten the four screws to hold the  
pulse damper in place.  
3. Connect the tubing from the pump head to the port at the rear of the pulse  
damper (i.e., toward the rear of the cabinet). Connect the line from the  
prime/purge valve to the other port, toward the front panel.  
4. Connect the transducer’s wire harness connector to pressure board  
connector P3.  
5. Replace the cover on the pump.  
5.6  
Cleaning the Pump  
1. Disconnect the column inlet tube from the column.  
2. Direct the column inlet tube (the tube from the injector outlet) to a  
waste beaker.  
3. Set the flow rate to maximum.  
4. Turn the injector to the INJECT position.  
5. Pump 100% methanol or isopropanol through the pump and injector  
for 3 minutes.  
6. Pump 100% filtered, distilled water through the pump and injector for  
3 minutes.  
For stainless steel flow paths, proceed to Step 7; For PEEK ™ flow  
paths, skip to step 10.  
WARNING:  
Use standard laboratory procedures and  
extreme care when handling strong acids and bases.  
7. Pump a 20% nitric acid/water solution through the pump and injector  
for 3 minutes.  
8. Flush the pump and injector with 100% filtered, distilled water for at  
least 3 minutes.  
9. Pump 100% isopropanol through the pump and injector for 3 minutes.  
The pump is now prepared for any mobile phase or short- or long-term  
shutdown.  
7-12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
10. If storing the pump for more than 12 hours fill each pump head with  
100% isopropanol then seal the inlets and outlet with the supplied  
plugs.  
5.7  
Lubrication  
The Binary pump has modest lubrication requirements. The  
bearings in the pump housing and piston carrier are permanently  
lubricated and require no maintenance. A small dab of light grease  
such as Lubriplate 630-AA on the cam is the only recommended  
lubrication. Be sure not to get lubricant on the body of the piston  
carrier, as this can retard its movement and interfere with proper  
pumping.  
NOTE: Keeping the interior of the pump free of dirt and dust  
will extend the pump’s useful life.  
5.8  
Other Pump Maintenance  
The internal components of the pump can be accessed by  
removing the pump drawer. To do so, turn off power at front switch.  
Unplug power cord. Ensure there is no power to the system.  
Disconnect the inlet line and disconnect pump outlet at the “T”  
Connector Block above. Unscrew the two thumbscrews on the upper  
portion of the drawer until fully released (counterclockwise). Pull the  
drawer out as shown on next page.  
Note: Drawer can be removed completely.  
7-13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
The Pump mechanism, Control Boards, & Pulse Dampener (with Pressure  
Transducer) are accessible by removing pump drawer.  
7-14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.9  
Fuse Replacement  
Three fuses protect the Binary Solvent Delivery Module. Two of  
the fuses are located in the power entry module at the rear of the  
cabinet and are in series with the AC input line. The other fuse is  
located on the circuit board and is in series with the 24 VDC supply.  
Troubleshooting the fuses is straightforward. If the power cord is  
plugged in and the on/off power entry switch is on and the display  
does not light, check the two fuses in the power entry module. To gain  
access to these fuses, gently pry off the cover plate with a small flat-  
bladed screwdriver. Replace with fuses of the correct rating: 2 A slow-  
blo for 120 VAC systems.  
7-15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.10 Battery Replacement (If applicable) See attached photo for battery option.  
Depending on the version of drive board assembly installed, the board may not have a  
battery. If the printed circuit board does not have a battery, it is designed with circuitry that  
does not require a battery backup and you should disregard the following instructions.  
The battery provides power for the memory that holds the current  
pump configuration. If the pump is set at a flowrate and the power is  
turned off, when the power is turned back on the flowrate should  
appear as it was set. If this flowrate does not appear the battery will  
need replaced.  
CAUTION: Be sure to disconnect power cord before removing  
cover to insure there is no voltage present.  
CAUTION: Circuit boards can be damaged by Electro Static  
Discharge (ESD). Follow standard ESD procedures when  
handling circuit boards.  
1. Unplug the unit.  
2. Remove the cover.  
3. Turn the unit so that the pump heads are to the right. The battery can  
be seen in the lower right corner of the circuit board. The battery is  
circular and has a positive pole mark (+) on the top. Gently pull it  
from its socket.  
4. With the positive mark (+) up, gently slide the new battery into the  
battery socket. Be sure the battery is all the way into place. It must  
contact the base of the battery socket.  
5. Replace the cover to the unit.  
6. Plug the unit back in.  
PCA without Battery  
7-16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. PROBLEM SOLVING  
Quick Guide to Problem Solving  
You Notice  
This May Mean  
Possible Cause  
You Should  
1. Uneven pressure  
trace.  
2. Pressure drops.  
3. No flow out the  
outlet check valve.  
1. Bubble in  
check valve.  
2. Leaks in  
system.  
3. Dirty check  
valve.  
1. Solvent not properly  
degassed.  
2. Fittings are not tight.  
3. Mobile phase not properly  
filtered.  
4. Particles from worn piston  
seal caught in check valve.  
5. Plugged inlet filter.  
1. Check to be certain that mobile phase is  
properly degassed.  
2. Check connections for leaks by tightening  
fittings.  
3. Prime the system directly from the outlet  
check valve.  
4. Clean or replace the check valves. See  
Section 5.4.  
4. Bad check  
valve.  
5. Clean or replace inlet filter. See Section  
5.1.1.  
1. Uneven pressure  
trace.  
2. Pressure drops.  
3. Fluid between the  
pump head and the  
retainer.  
1. Leaks in  
system.  
2. The piston  
seal or  
1. Fittings not tight.  
1. Check all connections for leaks.  
2. Replace piston seal & diaphragm. See  
Sections 5.2 and 5.3.  
3. Check the piston for salt deposits. Clean as  
necessary. See Section 5.2.4.  
2. Long usage time since last  
seal / diaphragm change.  
3. Salt deposits on seal or  
diaphragm (especially if  
buffered aqueous mobile  
phases are used).  
diaphragm is  
worn.  
Pump makes a loud  
clanging or slapping  
noise (intermittent  
contact with cam).  
Piston carrier is  
catching in  
piston guide.  
1. Cap nut screws on the pump  
head are loose.  
2. Seal(s) are worn.  
1. Check cap nut screws on pump head.  
Tighten if necessary.  
2. Replace seals.  
3. Replace piston guide and seals. See  
Sections 5.2 and 5.3.  
3. Piston guide is worn  
No power when pump  
turned ON.  
Blown fuses in  
the power entry  
module.  
1. Power surge.  
2. Internal short.  
1. Replace only with the appropriate fuses (1A  
for 100/110 Vac or 1/2A for 220/240 Vac).  
2. Contact service technician if problem  
persists.  
Blue dye in mobile  
phase.  
Pulse damper  
diaphragm has  
burst.  
Sudden pressure drop when  
purging system.  
Replace pulse damper. See Section 5.5.  
Pump runs for 50  
pump strokes, then  
shuts down.  
Lower pressure  
limit is  
activating.  
1. Mobile phase is not properly  
filtered.  
2. Particles from worn seal  
trapped in the system (e.g.,  
tubing, filters, injection valve,  
column inlet).  
1. Check to be certain the low pressure limit is  
set to 0 psi.  
2. Only increase the low pressure limit after the  
pump attains operating pressure.  
3. Contact service technician.  
1. Pump shuts down  
after run is called  
even with no  
column connected.  
2. Pump runs to  
maximum pressure  
and shuts down.  
Clog in fluid  
system.  
1. Remove and clean both the inlet and  
bulkhead filters. See Section 5.2.  
2. If the problem persists, remove tubing from  
system one piece at a time until you find the  
clogged piece. Most clogs occur outside the  
pump itself.  
No power when pump  
turned ON. Fan does  
not run.  
Blown fuses in  
the power entry  
module.  
1. Power surge.  
2. Internal short.  
1. Replace only with the appropriate fuses (1A  
for 100-120 Vac or 1/2A for 220-240 Vac).  
2. Contact service technician if problem  
persists.  
Front panel appears  
OK but pump motor  
does not run.  
Blown fuse on  
the motor power  
circuit board.  
1. Power surge.  
2. Internal short.  
1. Replace only with the appropriate fuse .  
2. Contact service technician if problem  
persists.  
PEEK fittings or  
components leak.  
You cannot  
force PEEK  
parts with  
interference to  
seal by brute  
force tightening.  
1. Film of fluid between  
surfaces.  
2. Salt crystals between  
surfaces.  
1. Clean and dry mating surfaces.  
2. If scratched, replace defective part.  
3. Scratches in mating surfaces.  
7-17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. LIST OF REPLACEMENT PARTS  
BINARY, STAINLESS STEEL  
880201  
880202  
880203  
880204  
880407  
880721  
880651  
880353  
880354  
880414  
880511  
880971  
880806  
Seal Kit, Aqueous, 5mL  
Seal Kit, Organic, 5mL  
Seal Kit, Aqueous, 10mL  
Seal Kit, Organic, 10mL  
Check Valve Kit – Stainless Steel  
Replacement Inlet Filter Elements (Package of 2)  
Prime Purge Valve Rebuild Kit – Stainless Steel  
Series II-IV, Binary Piston, 5mL  
Series II-IV, Binary Piston, 10mL  
Self-Flush Check Valve Kit  
Binary Drive Assembly  
Front Panel Assembly  
Fuse, 2 Amp, 5x20 mm (10 pack)  
Specific to Left Hand Pump Drawer (with Pressure)  
880613  
880144  
880972  
880940  
880941  
Repl. Pulse Damper with Pressure Transducer - Stainless  
SMT Board Set with Pressure  
Binary Overlay with Pressure  
Left hand replacement drawer 5mL SS  
Left hand replacement drawer 10mL SS  
Specific to Right Hand Pump Drawer (without Pressure)  
880615  
880145  
880979  
880942  
880943  
Repl. Pulse Damper without Pressure Transducer - Stainless  
SMT Board Set without Pressure  
Binary Overlay without Pressure  
Right hand replacement drawer 5mL SS  
Right hand replacement drawer 10mL SS  
7-18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
BINARY, PEEK  
880201  
880202  
880203  
880204  
880408  
880721  
880652  
880353  
880354  
880414  
880511  
880124  
880806  
Seal Kit, Aqueous, 5mL  
Seal Kit, Organic, 5mL  
Seal Kit, Aqueous, 10mL  
Seal Kit, Organic, 10mL  
Check Valve Kit - PEEK  
Replacement Inlet Filter Elements (Package of 2)  
Prime Purge Valve Rebuild Kit - PEEK  
Series II-IV, Binary Piston, 5mL  
Series II-IV, Binary Piston, 10mL  
Self-Flush Check Valve Kit  
Binary Drive Assembly  
Front Panel Assembly  
Fuse, 2 Amp, 5x20 mm (10 pack)  
Specific to Left Hand Pump Drawer (with Pressure)  
880614  
880144  
880972  
880944  
880945  
Repl. Pulse Damper with Pressure Transducer - PEEK  
SMT Board Set with Pressure  
Binary Overlay with Pressure  
Left hand replacement drawer 5mL PEEK  
Left hand replacement drawer 10ml PEEK  
Specific to Right Hand Pump Drawer (without Pressure)  
880616  
880145  
880979  
880946  
880947  
Repl. Pulse Damper without Pressure Transducer - PEEK  
SMT Board Set without Pressure  
Binary Overlay without Pressure  
Right hand replacement drawer 5mL PEEK  
Right hand replacement drawer 10mL PEEK  
7-19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
APPENDIX A  
A.1 Rear Panel Serial Communications Port - Pump  
An RS-232C modular jack is provided on the back panel. A  
computer, with appropriate software, can be used as a remote  
controlling device for pump operation via this connection.  
A.1.1 Hardware Implementation  
The REMOTE INPUT serial communications port is configured  
for 9600 baud, 8 data bits, 1 stop bit, and no parity. The connector is a  
standard RJ-11 modular telephone type jack. When looking at the  
connector on the rear panel of the pump, pin 1 is at the top and pin 6 is  
at the bottom. The pin-out is:  
Pin  
1, 6  
2
3
4
Function  
Ground  
DSR (Handshaking input to pump)  
RXD (Serial data input to pump)  
TXD (Serial data output from pump)  
DTR (Handshaking output from pump)  
5
Special wiring considerations: Use the following chart for  
interfacing the pump's serial communications port to either a 25-pin or  
a 9-pin COM port on an IBM-PC type computer.  
Pump (RJ11)  
Signal  
Ground  
DSR  
RXD  
TXD  
IBM (DB25)a  
IBM (DB9)b  
1, 6  
2
3
4
5
7
20  
2
3
6
5
4
3
2
6
DTR  
a Jumper pins 4, 5, and 8 on DB25.  
b Jumper pins 1, 7, and 8 on DB9.  
Part Description  
Part Number  
Modular Cable  
Adapter RJ-11 to DB9  
Adapter RJ-11 to DB-25  
12-0677  
12-0672  
12-0671  
A.1.2 Hand-Shaking  
The pump uses hardware handshaking. The pump will not  
transmit on the TXD output if the DSR input is at a low logic level.  
And, the pump will not receive on the RXD input when the DTR  
A-1  
Download from Www.Somanuals.com. All Manuals Search And Download.  
output is at a low logic level. A low logic level is -3.0 to -15 volts and  
a high logic level is 3.0 to 15 volts.  
A.1.3 Command Interpreter  
The pump’s high-level command interpreter receives and responds  
to command packets. The pump will not send a message except when  
prompted, and it will send a response to every valid command as  
described below. The response to an invalid command is “Er/”.  
Each command is characterized by a unique two-letter command  
code, and only one command can be issued per line. Case is not  
important; that is, the command codes “PR” “Pr” “pR” and “pr” are all  
equivalent. Response strings sent by the pump are terminated by the  
“/” character.  
If the pump's response is "Er/", send a "#" to clear any characters  
which may be remaining in the command buffer. The pump will  
automatically clear all characters in the command buffer after one  
second elapses from the time at which the last character of an  
incomplete command was sent.  
The command packets are as follows:  
Command  
RU  
Response  
OK/  
Comments  
Sets the pump to the RUN state.  
ST  
OK/  
Sets the pump to the STOP state.  
FLxxx  
OK/  
Sets the flow rate to x.xx or xx.x mL/min where the range  
is fixed for the pump head size, i.e., for 0.01 to 9.99  
mL/min xxx = 001 to 999, for 0.1 to 39.9 mL/min xxx =  
001 to 399.  
FOxxxx  
FMxxxx  
OK/  
OK/  
Sets the flow rate to xx.xx or xxx.x mL/min where the  
range is fixed for the pump head size, i.e., for 0.01 to  
10.00 mL/min xxxx = 0001 to 1000, for 0.1 to 40.0 mL/min  
xxxx = 0001 to 0400.  
Sets the flowrate to x.xxx mL/min, i.e.,  
for 0.001 to 9.999mL/min xxxx = 0001 to 9999.  
for 10.00 to 12.00mL/min xxxx = 1000 to 1200.  
Reads the pump's current pressure, where:  
x, xx, xxx, or xxxx = Current pressure in PSI  
Reads the pump's current pressure and flowrate, where:  
x, xx, xxx, or xxxx = Current pressure in PSI  
y.yyy, y.yy, yy.yy, or yy.y = Flow rate in mL/min  
PR  
CC  
OK,x/  
(x, xx, xxx, or xxxx)  
OK,x,y.yy/  
(x, xx, xxx, or xxxx)  
(y.yyy, y.yy, yy.yy, or yy.y)  
The format is y.yy and yy.yy for a standard pump head,  
y.yyy for micro pump head or yy.y for a macro pump  
head.  
A-2  
Download from Www.Somanuals.com. All Manuals Search And Download.  
CS  
OK,x.xx,y,z,PSI,w,v,u/  
Reads the current pump setup, where:  
x.xxx, xx.xx, or xxx.x = Flow rate in mL/min  
y, yy, yyy, or yyyy = Upper pressure limit  
z, zz, zzz, or zzzz = Lower pressure limit  
PSI = Units (PSI, ATM, MPA, BAR, or KGC)  
w = Pump head size (0 = standard, 1 = macro)  
v = Run status (0 = stopped, 1 = running)  
u = Pressure Board present = 0; otherwise 1  
Identifies the pump type and EPROM revision x.xx  
Sets the upper pressure limit in PSI. The maximum value  
for xxxx is 5000 for the plastic head or 6000 for the steel  
head; the minimum value is the lower limit plus 100. The  
value must be expressed as four digits, i.e., for 900 PSI  
xxxx = 0900.  
(x.xxx, xx.xx, or xxx.x)  
(y, yy, yyy, or yyyy)  
(z, zz, zzz, or zzzz)  
ID  
UPxxxx  
OK,vx.xx SR3O firmware/  
OK/  
LPxxxx  
OK/  
Sets the lower pressure limit in PSI. The maximum value  
for xxxx is the current upper pressure limit setting minus  
100; the minimum value is 0. The value must be  
expressed as four digits, i.e., for 100 PSI xxxx = 0100.  
Puts the pump in fault mode. Turns on the FAULT LED  
and stops the pump immediately.  
SF  
RF  
OK/  
OK,x,y,z/  
Reads the fault status, where:  
x = Motor stall fault (0 = no, 1 = yes)  
y = Upper pressure limit fault (0 = no, 1 = yes)  
z = Lower pressure limit fault (0 = no, 1 = yes)  
Disables the keypad. (Default status at power-up is  
enabled.)  
KD  
OK/  
KE  
PCxx  
OK/  
OK/  
Enables the keypad.  
Sets the pressure compensation value, where xx = the  
operating pressure (in PSI divided by 100),  
i.e., for 0 PSI xx = 00, for 5000 PSI xx = 50.  
Reads the pressure compensation value in hundreds of  
PSI, i.e., for 0 PSI x = 0, for 5000 PSI xx = 50.  
Sets the pump head type, where:  
RC  
OK,x/  
(x or xx)  
OK/  
HTx  
x = 1 for a stainless steel 10 mL/min pump head  
x = 2 for a plastic 10 mL/min pump head  
x = 3 for a stainless steel 40 mL/min pump head  
x = 4 for a plastic 40 mL/min pump head  
x = 5 for a stainless steel 5 mL/min pump head  
x = 6 for a plastic 5 mL/min pump head  
The pump is stopped; and, the pressure compensation  
and pressure limits are initialized, when the head type is  
changed.  
RH  
OK,x/  
Reads the pump head type, where:  
x = 1 for a stainless steel 10 mL/min pump head  
x = 2 for a plastic 10 mL/min pump head  
x = 3 for a stainless steel 40 mL/min pump head  
x = 4 for a plastic 40 mL/min pump head  
x = 5 for a stainless steel 5mL/min pump head  
x = 6 for a plastic 5 mL/min pump head  
A-3  
Download from Www.Somanuals.com. All Manuals Search And Download.  
PI  
OK,a.aa,b,c,d,e,f,g,h,i,j,k,l,  
m,n,o,p,q/  
Reads the current pump setup, where:  
a.aaa, a.aa, aa.aa, or aa.a = Flow rate in mL/min  
b = Run status (0 = stopped, 1 = running)  
c or cc = Pressure compensation  
d = Pump head type (see RH command)  
e = Pressure Board present = 0; otherwise 1  
f = External control mode (0 = frequency, 1 =  
voltage)  
(a.aaa, a.aa, aa.aa, or  
aa.a)  
(c or cc)  
g = 1 if pump started and frequency controlled,  
else 0  
h = 1 if pump started and voltage controlled, else 0  
i = Upper pressure limit fault (0 = no, 1 = yes)  
j = Lower pressure limit fault (0 = no, 1 = yes)  
k = Priming (0 = no, 1 = yes)  
l = Keypad lockout (0 = no, 1 = yes)  
m = PUMP-RUN input (0 = inactive, 1 = active)  
n = PUMP-STOP input (0 = inactive, 1 = active)  
o = ENABLE IN input(0 = inactive, 1 =active)  
p = Always 0  
q = Motor stall fault (0 = no, 1 = yes)  
Resets the pump configuration to its default power-up  
state.  
RE  
OK/  
A.2 Pump Rear Panel 4-Pin and 10-Pin Terminal Board Connectors  
A 4-pin terminal board connector and a 10-pin terminal board  
connector are provided on the back panel. Any device capable of  
providing the proper run/stop logic level, flowrate control frequency,  
or flowrate control voltage can be used as a remote controlling device  
for pump operation via this connection. The terminal board  
connectors can be removed for ease of connecting wires, if desired, by  
pulling firmly rearward and should be reinserted firmly afterward.  
A.2.1 Pressure Fault and Motor Stall Fault Output  
The pump's output is on the 4-pin terminal board connector. The  
pinout is:  
Pin  
4
3
2
1
Function  
EVENT 1  
EVENT 2  
EVENT 3  
Ground  
This output is produced internally by a reed relay which has SPDT  
contacts with a 0.25 amp maximum, 50 VDC maximum, 0.2 ohm  
rating. The 4-pin connector allows wires to be connected to the  
EVENT 1 (Pole), EVENT 2 (NC), and EVENT 3 (NO) terminals.  
When the pump stops due to the sensed pressure exceeding the set  
pressure limits or if a motor stall fault occurs, the connection between  
A-4  
Download from Www.Somanuals.com. All Manuals Search And Download.  
the EVENT 1 terminal and the EVENT 2 and EVENT 3 terminals is  
affected. EVENT 2 is Normally Closed (connected to EVENT 1) until  
a fault occurs and then opens. EVENT 3 is Normally Open (not  
connected to EVENT 1) until a fault occurs and then closes.  
A.2.1.1 Upper and Lower Pressure Limit Range  
(pump A only)  
The pressure sensing transducer provides accurate, wide range  
pressure monitoring. Because of the sensitivity of the transducer,  
the zero reading may shift up to 0.1% of the full pressure scale  
over years of operational use. The user should also be aware that  
the resistance to flow of the fluid being pumped through the  
tubing and fittings may cause the pressure to variy with the flow  
rate and the viscosity of the mobile phase employed.  
If absolute accuracy is needed for the pressure safety limits:  
1. Disconnect the column from pumping system and operate the  
pump with the mobile phase and flow rate to be used in the  
analysis. Observe the resulting pressure displayed on the pump  
readout. The column will cause a pressure reading that adds to this  
basic reading due to system flow resistance.  
2. Set the upper limit shut-off to a pressure equal to the basic  
reading plus the safe operating pressure for the column to be used.  
For example, if the basic pressure reading (without the column) is  
7 PSI and the safe limit for the column is 25 PSI, set the maximum  
pressure limit to 32 PSI or less.  
3. If the mobile phase or flow rate is changed, reset the pressure  
limit as appropriate.  
4. Note that a lower pressure limit is available to prevent  
continued operation in the event of a leak. For proper operation,  
this must be set to a pressure higher than the basic pressure or it  
may not sense the reduced pressure.  
A.2.2 General Information on Inputs  
The pump's inputs are on the 10-pin terminal board connector. The  
pinout is:  
A-5  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Pin  
10  
9
Function  
VOLTAGE COM  
VOLTAGE IN  
FREQ IN  
8
7
6
5
4
3
2
1
ENABLE IN  
PUMP-RUN  
PUMP-STOP  
No connection  
No connection  
No connection  
COM  
A.2.3 General Information on Run, Stop, and Enable Inputs  
The PUMP-RUN, PUMP-STOP, and ENABLE IN inputs operate  
from an internal 5 VDC source and each one draws approximately  
0.008 amps when connected to COM. To activate either the PUMP-  
RUN, PUMP-STOP, or ENABLE IN input connect it to COM. Any  
device capable of switching 0.008 amps can be connected between the  
PUMP-RUN, PUMP-STOP, or ENABLE IN input and COM, such as:  
a switch contact, a relay contact, an open collector output, an open  
drain output, or any output with a high logic level output of 3.8 to 6.0  
volts and a low logic level output of 0.0 to 0.5 volts. A switch contact  
or a relay contact is preferred since this type of connection will  
provide isolation between the pump and the controlling device. The  
COM terminal is internally connected to the pump's chassis ground  
and should be connected to the controlling device's ground or zero volt  
terminal when the controlling device has an open collector output, an  
open drain output, or any output with logic level output.  
A.2.4 Run and Stop Inputs  
The pump's motor can be commanded to run or stop from the back  
panel inputs when the pump’s flowrate is controlled from the front  
panel or when the pump’s flowrate is controlled by the voltage or  
frequency input. There two modes of operation for the run and stop  
inputs which are described below:  
Dual Signal Pulse: In this mode of operation both the PUMP-RUN  
and PUMP-STOP inputs are normally at a high logic level. To start  
the pump, pulse the PUMP-RUN input to a low logic level for a  
minimum of 500 mS. To stop the pump, pulse the PUMP-STOP input  
to a low logic level for a minimum of 500 mS.  
Single Signal Level: To enable this mode of operation the PUMP-  
STOP input must be permanently connected to COM with a jumper  
wire. To start the pump, put a low logic level on the PUMP-RUN  
input. To stop the pump, put a high logic level on the PUMP-RUN  
input.  
A-6  
Download from Www.Somanuals.com. All Manuals Search And Download.  
A.2.4 Enable Input  
When activated (ENABLE IN is at a low logic level), the  
ENABLE IN input disables flowrate control on the front panel and  
enables flowrate control on the back panel.  
A.2.5 General Information on Voltage and Frequency Inputs  
Special programming and circuitry allows this pump to be operated  
remotely with the flowrate controlled by voltage or frequency inputs.  
To select the remote mode of operation:  
a.) With the pump plugged in and the rear panel power switch  
OFF, press in and hold the "DOWN ARROW" button while  
turning the power switch ON.  
b.) Release the "DOWN ARROW" button and either a U (closest  
approximation to V for voltage) or an F (for frequency) will be  
displayed.  
c.) Select the desired remote operating mode by pressing the  
"DOWN ARROW" button to toggle between the voltage and  
frequency mode.  
d.) Press the "RUN/STOP" button to place the pump in normal  
operating mode.  
e.) To enable the currently selected remote mode (voltage or  
frequency), connect the rear panel ENABLE IN connection to  
the COM connection.  
f.) When in the remote mode (ENABLE IN at a low logic level)  
all front panel buttons remain active except the flow setting  
increase/decrease capability.  
A-7  
Download from Www.Somanuals.com. All Manuals Search And Download.  
APPENDIX B  
EZ GRIP FITTING GUIDE  
A-8  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Scientific Systems Inc.  
Warranty Statement  
Scientific Systems, Inc. (SSI) warrants that instruments or equipment manufactured by it for a period  
thirty-six (36) months from date of shipment to customer, against defects in materials and  
workmanship under normal installation, use and maintenance. Expendable items and physical  
damage caused by improper handling or damage caused by spillage or exposure to any corrosive  
environment are excluded from this warranty. The warranty shall be void for Polyetheretherketone  
(PEEK) components exposed to concentrated Nitric or Sulfuric acids which attack PEEK, or  
methylene chloride, DMSO or THF which adversely affect UHMWPE seals and PEEK tubing. Any  
defects covered by this warranty shall be corrected by replacing or repairing, at SSI’s option, parts  
determined by SSI to be defective.  
Spare or replacement parts and accessories shall be warranted for a period of 12 months from date of  
installation at customer against defects in materials and workmanship under normal installation, use  
and maintenance. Defective Product will be accepted for return only if customer returns them to SSI  
within thirty (30) days from the time of discovery of the alleged defect, and prior to return, obtains a  
Return Goods Authorization (RGA) number from SSI, and provides SSI with the serial number of  
each instrument to be returned. Freight costs for the return of defective Product is the responsibility  
of SSI. SSI shall specify the freight carrier for returns.  
The warranty shall not apply to any Product that has been repaired or altered except by SSI or those  
specifically authorized by SSI, to the extent that such repair or alteration caused the failure, or to  
Product that has been subjected to misuse, negligence, accident, excessive wear, or other causes not  
arising out of a defect in material or workmanship.  
The warranty shall not apply to wear items, specifically:  
Check Valves  
Piston and Wash Seals  
Pulse-Damper Diaphragms  
Filter Elements  
Pistons  
Inlet Lines  
The following is the exclusive procedure by which to make claims under this warranty. Customer  
shall obtain SSI’s oral or written authorization to return the product and receive a Return Goods  
Authorization (RGA) number. The Product must be returned with the RGA number plainly visible  
on the outside of the shipping container to SSI. It must be securely packed in a rigid container with  
ample cushioning material, preferably the original packaging. All claimed defects must be specified  
in writing, including the RGA number, with the written claim accompanying the Product. Product  
shall be shipped to SSI at customer’s expense. SSI shall bear the expense of return shipment.  
If it appears to SSI that any Product has been subjected to misuse, negligence, accident or excessive  
wear, or is beyond the warranty period, customer shall be notified promptly. SSI shall notify  
customer of its finding and provide an estimate to repair such Product at the then current rates for  
parts and service. SSI shall either repair the product per customer’s authorization or shall return  
such Product not repaired to customer at customer’s expense. SSI may invoice customer for the  
freight costs of any Product shipped back to customer by SSI which is not covered under the  
warranty.  
Limitations of Warranty. THE FOREGOING WARRANTIES AND LIMITATIONS ARE CUSTOMER’S  
EXCLUSIVE REMEDIES AND ARE IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR  
A-9  
Download from Www.Somanuals.com. All Manuals Search And Download.  
IMPLIED, INCLUDING WITHOUT LIMITATION ANY WARRANTY OF MERCHANTABILITY OR  
FITNESS FOR A PARTICULAR PURPOSE.  
A-10  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Sharp Answering Machine FO 475 User Manual
Smooth Fitness Treadmill 935HR User Manual
Sony Portable CD Player D EJ106CK User Manual
Sony Portable DVD Player DVP FX950 User Manual
Sony Webcam VGN NS User Manual
Stanley Black Decker Chainsaw CS05 User Manual
Star Micronics Printer FR 10 User Manual
Swann DVR DVR8 Net Plus User Manual
Swann TV Receiver SW112 DVB User Manual
Tannoy Speaker Tweeter User Manual