ResMed Portable Generator VPAP Adapt SV User Guide

VPAP™ Adapt SV  
User Guide  
English  
228234-Eng/3 08 10  
1
Download from Www.Somanuals.com. All Manuals Search And Download.  
Introduction  
VPAP™ Adapt SV  
The VPAP Adapt SV flow generator has been designed to be reliable, comfortable  
and easy to use. You can observe your treatment and make adjustments to  
settings using a keypad and LCD screen.  
CAUTION  
!
USA federal law restricts this device to sale by or on the order of a  
physician.  
User/Owner Responsibility  
The user or owner of this system shall have sole responsibility and liability for any  
injury to persons or damage to property resulting from:  
operation which is not in accordance with the operating instructions supplied  
maintenance or modifications carried out unless in accordance with  
authorized instructions and by authorized persons.  
Please read this manual carefully before use.  
Definitions  
WARNING  
!
!
Alerts you to possible injury.  
CAUTION  
Explains special measures for the safe and effective use of the device.  
Medical Information  
Intended Use  
The VPAP Adapt SV system is intended to provide non-invasive ventilatory  
support to treat adult patients with obstructive sleep apnea (OSA) and respiratory  
insufficiency caused by central and/or mixed apneas and periodic breathing. This  
device is intended for use in the hospital or home.  
Contraindications  
The VPAP Adapt SV is not a life support ventilator and may stop operating with  
power failure or in the unlikely event of certain fault conditions.  
Before using the VPAP Adapt SV tell your doctor if you have any of the following  
conditions:  
hypotension or significant intravascular volume depletion  
acute sinusitis or otitis media  
epistaxis (severe nosebleeds) causing a risk of pulmonary aspiration  
conditions predisposing to a risk of vomiting into mask  
Introduction  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
impaired ability to clear secretions  
pneumothorax or pneumomediastinum  
recent cranial trauma or surgery  
chronic hypoventilation.  
Below are general warnings and cautions. Further specific warnings,  
cautions and notes appear next to the relevant instructions in the manual.  
WARNINGS  
!
The VPAP Adapt SV is NOT a life support ventilator.  
The entire manual should be read before using the VPAP Adapt SV.  
Advice contained in this manual should not supersede instructions given by  
the prescribing physician.  
The VPAP Adapt SV should be used with masks and accessories  
recommended by ResMed or the prescribing physician. Use of incorrect  
masks and accessories may adversely affect the function of the  
VPAP Adapt SV.  
The VPAP Adapt SV is designed for use with masks that allow exhaled gases  
to be flushed out through vent holes. Exhaled gases will be rebreathed if the  
mask is worn with the machine turned off, or the vent holes are occluded. If  
this occurs over prolonged periods, suffocation may occur.  
In the event of power failure or machine malfunction, remove the mask.  
º
The air flow for breathing produced by this device can be as much as 11 F  
º
(6 C ) higher than the temperature of the room. Caution should be exercised  
º
º
if the room temperature is warmer than 90 F (32 C).  
The VPAP Adapt SV can be set to deliver pressures up to 25 cm H O. In the  
2
unlikely event of certain fault conditions, pressures of up to 40 cm H O for  
2
less than one second are possible.  
The VPAP Adapt SV is not suitable for use in the vicinity of flammable  
anesthetics.  
If oxygen is used with the VPAP Adapt SV, the oxygen flow should be  
stopped when the device is not operating. If oxygen flow continues when  
the device is not operating, oxygen may accumulate within the device and  
create a risk of fire.  
Do not use the VPAP Adapt SV if there are obvious external defects or  
unexplained changes in performance.  
Do not open the VPAP Adapt SV case. There are no user serviceable parts  
inside. Repairs and internal servicing should only be performed by an  
authorized service agent.  
CAUTIONS  
!
You should report unusual chest pain, severe headache or increased  
breathlessness to your physician. An acute upper respiratory tract infection may  
require temporary discontinuation of treatment.  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
The following side effects may arise during a course of therapy with the  
VPAP Adapt SV:  
drying of the nose, mouth, or throat  
bloating  
ear or sinus discomfort  
eye irritation  
skin rashes  
chest discomfort.  
Quick Setup Guide  
1
Attach the air tubing and pressure sensor tube to the VPAP Adapt SV and mask  
(page 9).  
Attach a humidifier if required (page 10).  
2
3
4
5
Switch on the VPAP Adapt SV. Make sure the mask is unblocked.  
Select the mask setting (page 21).  
Perform the Air Circuit Learn function (page 21) at first-time use, or if you are  
adding or removing a component of the system (eg, a humidifier or anti-bacterial  
filter).  
Fit the mask and commence breathing into it. The SmartStart™ function should  
trigger treatment to commence within the first couple of breaths. If not, press  
the Start/Stop key once to commence treatment (page 16).  
6
Note: You can observe treatment progress (eg, mask leak) via the Treatment  
screens (page 24).  
Quick Setup Guide  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
Remove your mask to stop treatment. If SmartStop is enabled, treatment will  
stop automatically. If the SmartStop function has been disabled, press the  
Start/Stop key once.  
7
VPAP Adapt SV System  
VPAP Adapt SV Components  
The VPAP Adapt SV flow generator is supplied with air tubing and a power cord.  
Masks are supplied separately, as you and your clinician have to decide which  
mask is best for you.  
Some people experience dryness of the nose, mouth and throat while using the  
VPAP Adapt SV system, especially during winter. Others may experience  
sneezing and/or a runny or blocked nose during the first few weeks of VPAP  
treatment. In many cases, these symptoms of nasal irritation can be resolved  
with a humidifier. See page 10 to connect a humidifier to the VPAP Adapt SV.  
LCD screen  
Control panel  
Carry handle  
Removable  
Pressure sensor  
front cover  
tube connector  
Air tubing connector  
Figure 1: VPAP Adapt SV flow generator  
The control panel allows you to adjust settings on the flow generator, and the  
LCD screen displays information about your treatment. See “Control Panel  
Functions” on page 17.  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
ResLink attachment point  
Air filter  
Power cord  
Figure 2: Rear of VPAP Adapt SV  
See page 14 to connect a ResLink™. See page 30 to replace the air filter.  
Proximal cuff  
Luer connector  
Tube clip  
Pressure sensor tube  
Connect this end to  
the flow generator  
Connect this end  
to the mask  
Air tubing  
Figure 3: Air tubing fully assembled  
All necessary air tubing components are packed with the VPAP Adapt SV:  
a 6’6" (2 m) length of air tubing  
a pressure sensor tube (to measure pressure at the mask) with Luer  
connectors  
tube clips (to hold the pressure sensor tube onto the air tubing)  
a proximal cuff (a special connector that attaches the pressure sensor tube  
into the main air tubing so you only need to make one attachment to the  
mask).  
The air tubing is supplied with the proximal cuff connected. Attach the tube clips  
evenly along the air tubing. Press the pressure sensor tube into the tube clips,  
and screw the Luer connectors together at the proximal cuff. Make sure the  
pressure sensor tube is not kinked, obstructed or twisted.  
Note: ResMed recommends the following tubing products for use with the  
VPAP Adapt SV: replacement air hose (only) PN14986; replacement air delivery  
system (hose, sensor line, clips, proximal cuff) PN26923.  
Masks  
You will also need a ResMed mask system (supplied separately). For the latest  
set up your mask correctly for use with the VPAP Adapt SV. Your clinician can  
explain the features of the masks available and will discuss the mask that best  
suits your needs.  
VPAP Adapt SV System  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
Assembling the VPAP Adapt SV System  
Complete the following steps to assemble the VPAP Adapt SV system. Make  
sure the area around the unit is clean (dust free) and clear of bedding, clothes and  
any other potential blockages.  
To add a humidifier see page 10.  
WARNING  
!
Use only the air tubing system supplied with the VPAP Adapt SV system.  
Connect the power cord  
1
There are two sockets at the rear of the flow generator—the top one for an AC  
(standard mains electricity) and the lower one for a DC power cord (see “Using  
DC Power to Run the VPAP Adapt SV” on page 15). ResMed recommends using  
the AC power cord supplied with the unit.  
The customized ResMed power cord can be held in place with the locking clips.  
To insert the locking clip, pinch the free ends together and fit the pins into the  
holes on both sides of the socket. Insert the power cord into the socket. Push  
the locking clip down so that the groove holds the power cord in place. Plug one  
end of the power cord into the power socket at the rear of the unit, and the other  
end into a power outlet.  
Power switch  
AC locking clip  
AC power cord  
Plug the free end of the power cord into a power outlet.  
CAUTIONS  
!
!
Do not connect both AC and DC power cords to the VPAP Adapt SV at  
the same time.  
Be careful not to place the device where it can be bumped or where  
someone is likely to trip over the power cord.  
WARNING  
Make sure the power cord and plug are in good condition and the  
equipment is not damaged.  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
Connect air tubing  
2
Connect the air tube firmly onto the air outlet at the front of the unit.  
Air tubing connector  
Pressure sensor tube  
connector  
Figure 4: Front view of VPAP Adapt SV  
To attach the pressure sensor tube to the unit, screw the Luer connector onto  
the socket on the side of the unit.  
Figure 5: Twisting the Luer lock connectors together  
WARNING  
!
Use care when handling the air tubing. Be particularly careful when  
attaching the Luer connector to the VPAP Adapt SV. If you kink or  
excessively twist the sensor tube it may affect the protection offered by the  
mask pressure alarm system.  
Attach the mask to the air tubing.  
3
The mask system should already be assembled. If it is not, please refer to the  
user instructions supplied with the mask.  
Connect the mask system to the free end of the air tubing.  
Proximal cuff  
Figure 6: Assembling the VPAP Adapt SV system  
VPAP Adapt SV System  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
Humidifiers  
A humidifier may be required if you are experiencing dryness of the nose, throat,  
or mouth. The following ResMed humidifiers are compatible with the  
VPAP Adapt SV:  
HumidAire 2i  
heated humidifier  
HumidAire  
heated humidifier  
Humidifier accessory (HumidAire only)  
Medium air tubing  
21" (52 cm)  
Figure 7: Humidifiers compatible with the VPAP Adapt SV  
When using a humidifier, stop the VPAP Adapt SV by using the Start/Stop key or  
by removing the mask to trigger the SmartStop function. Do not switch the  
VPAP Adapt SV off at the power switch or the power source while it is running, as  
this could result in condensation in the motor.  
Please refer to the relevant humidifier user manual for detailed instructions and  
warnings.  
CAUTIONS  
!
Be very careful not to allow water to enter the VPAP Adapt SV  
as this could affect therapy and/or damage the device.  
Do not tip the humidifier while it is connected to the  
VPAP Adapt SV. If water enters the VPAP Adapt SV device, turn off the  
main power switch at the back of the unit and unplug it from the power  
outlet. Return the unit to your equipment supplier for checking.  
Connecting a Humidifier  
Note: After connecting or removing a humidifier, always perform the Air Circuit  
Learn function (see page 21).  
WARNING  
!
When using a humidifier, position it so that it is lower than your head will  
be during sleep, and at the same level or lower than the VPAP Adapt SV.  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
HumidAire 2i™  
The HumidAire 2i provides heated humidification for the VPAP Adapt SV.  
To attach the HumidAire 2i, remove the VPAP Adapt SV front cover and attach  
the humidifier docking station and water chamber. No other accessories are  
required for use. Please refer to the HumidAire 2i User’s Manual for details.  
Once you have performed the Air Circuit Learn function the VPAP Adapt SV is  
ready for use. See “Starting Treatment” on page 16.  
VPAP Adapt SV  
HumidAire 2i  
docking station  
HumidAire 2i  
water chamber  
Figure 8: HumidAire 2i attached to the VPAP Adapt SV  
Figure 9: HumidAire 2i attached to VPAP Adapt SV and mask  
WARNING  
!
Make sure that the water chamber is empty and thoroughly dried before  
transporting the HumidAire 2i.  
When using the carry bag, always separate the VPAP Adapt SV unit and  
the Humidaire 2i before placing in their respective compartments.  
VPAP Adapt SV System  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
HumidAire™  
You will need a medium size 21" (52 cm) air tube to connect the VPAP Adapt SV  
unit to the HumidAire™ (see Figure 7).  
1
Make sure both the HumidAire and the VPAP Adapt SV are turned off.  
Fill the HumidAire with water as described in the humidifier manual. Place the  
filled water chamber inside the HumidAire.  
Gently detach enough of the tube clips to release the end of the pressure sensor  
tube from the VPAP Adapt SV air tubing.  
2
Connect the VPAP Adapt SV air tubing to the left connector port on the humidifier,  
and the medium 21" (52 cm) air tubing to the right connector port. Close the  
HumidAire lid.  
Medium air tubing  
HumidAire  
Pressure sensor tube  
VPAP Adapt SV air tubing  
Place the VPAP Adapt SV on top of the HumidAire. Never place the  
VPAP Adapt SV unit underneath the humidifier. (This is to avoid water spilling into  
the unit.)  
3
Connect the free end of the medium air tubing to the VPAP Adapt SV.  
Connect the pressure sensor tube by twisting the Luer lock onto the  
VPAP Adapt SV (see page 9).  
4
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Connect the mask system to the free end of the long air tubing. The final  
assembly should look like this:  
5
Switch on the HumidAire and the VPAP Adapt SV.  
6
WARNING  
!
Make sure that the power cords and plugs are in good condition and the  
equipment is not damaged.  
Perform the Air Circuit Learn function (see page 21).  
For information about filling and maintaining your humidifier, refer to the  
7
HumidAire User’s Manual.  
The VPAP Adapt SV is now ready for use with the HumidAire. See “Starting  
Treatment” on page 16.  
CAUTION  
!
Set up the HumidAire and the VPAP Adapt SV unit so that the tubing to  
the mask runs higher than the machines. This will prevent condensation  
build-up.  
If condensation appears in the mask, turn the humidifier settings down.  
VPAP Adapt SV System  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
ResLink™  
The ResLink is a device that records data onto a SmartMedia™ card during your  
therapy. The card can be returned to your clinician so that they can observe your  
progress and adjust flow generator parameters if necessary. The ResLink  
connects to the back of VPAP Adapt SV as shown. For full details on using the  
ResLink, please refer to the ResLink User’s Manual.  
Anti-bacterial Filter  
Your clinician may recommend that you use an anti-bacterial filter. A filter should  
be placed in the air tubing between the mask and the flow generator. If using a  
humidifier, place the filter between it and the flow generator (see Figure 10  
below).  
Typically, the filter should be replaced daily. Check the filter instructions for full  
details.  
Whenever a filter is added, removed or replaced you must run the Air Circuit  
Learn function (see “Learn Circuit” on page 21).  
Note: Only hydrophobic anti-bacterial filters should be used with a  
HumidAire 2i.  
Anti-bacterial filters  
Figure 10: Inserting anti-bacterial filters  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using DC Power to Run the VPAP Adapt SV  
The VPAP Adapt SV can be run on DC in case of power failure. ResMed supplies  
separately a converter that allows a 12V or 24V battery to be connected to the  
DC input of the VPAP Adapt SV. When using a DC power supply, the power is  
turned on and off via the switch on the converter, and not the power switch at  
the back of the VPAP Adapt SV. The DC converter is a separate accessory that  
can be purchased through your ResMed distributor.  
The humidifier heater will be automatically disabled when the VPAP Adapt SV is  
operating from the DC supply.  
Power consumption of the VPAP Adapt SV varies with treatment settings.  
Adequate battery type and sizing is required before operation on DC, and it is  
recommended that you contact ResMed’s Technical Services division or see  
ResMed's website for information on your specific application.  
CAUTION  
!
The VPAP Adapt SV uses 30V DC power. Only connect the VPAP Adapt SV  
to DC power using the VPAP Adapt SV DC Converter (P/N 26928) specified  
for this application.  
Supplemental Oxygen  
The VPAP Adapt SV is designed to be compatible with the use of low flow  
oxygen therapy provided from an external source and supplied at a fixed rate of  
up to 15 L/min. At a fixed flow rate of supplemental oxygen flow, the inhaled  
oxygen concentration will vary, depending on the pressure settings, patient  
breathing pattern, mask selection, and the leak rate. This warning applies to most  
types of bilevel machines.  
WARNING  
!
If oxygen is used with this device, the oxygen flow must be turned off  
when the device is not operating.  
Explanation: When the device is not in operation, and the oxygen flow is  
left on, oxygen delivered into the ventilator tubing may accumulate  
within the device enclosure and create a risk of fire.  
Always begin VPAP Adapt SV therapy before the oxygen supply is  
turned on.  
Always turn the oxygen supply off before stopping VPAP Adapt SV  
therapy.  
Oxygen supports combustion. Oxygen should not be used while  
smoking or in the presence of an open flame.  
VPAP Adapt SV System  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Using the VPAP Adapt SV  
WARNING  
!
Before starting therapy with a new mask, select the correct mask type in the  
Settings menu (page 21). When adding or removing a new component,  
such as a mask, humidifier, or anti-bacterial filter, perform the Air Circuit  
Learn function (page 21).  
Starting Treatment  
When the system is fully assembled press the power switch at the back of the  
VPAP Adapt SV to ON. The Welcome screen will be displayed while the flow  
generator is warming up. Allow the unit to warm up for approximately 30  
seconds before placing the mask on your face.  
WELCOME  
Figure 11: Welcome screen  
Once the VPAP Adapt SV has warmed up, breathing into the mask should start  
therapy immediately. If not, press the Start/Stop key once to commence  
treatment.  
The VPAP Adapt SV restarts in the mode in which it stopped.  
Stopping Treatment  
If SmartStop is enabled, treatment will stop automatically when you remove your  
mask. If the SmartStop function has been disabled, press the Start/Stop key.  
Standby Mode  
VPAP ADAPT SV  
menu  
Figure 12: Standby screen  
Once you have stopped treatment, the VPAP Adapt SV will be in standby mode.  
In standby you can:  
press the power switch at the back to OFF to turn off the flow generator  
commence treatment again by breathing into the mask or pressing the  
Start/Stop key  
enter any of the menus by pressing the Left key.  
When in standby mode, a gentle (non-therapeutic) flow of air will still be present.  
Note: If the SMART DATA - AUTO APPEAR option is set to ON the screens  
available under the Results menu will be displayed automatically for  
approximately 20 minutes after your session is completed.  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Control Panel Functions  
The control panel of the VPAP Adapt SV has a keypad that allows you to:  
start or stop treatment  
adjust settings on the flow generator  
mute or stop alarms  
view treatment data.  
LCD display  
QuickView key  
Alarm LEDs  
Alarm mute key  
Start/Stop Left  
key key  
Up/Down Right  
key key  
Figure 13: The VPAP Adapt SV control panel  
LCD screen  
Displays information about flow generator settings, alarms and treatment. The  
LCD screen is backlit whenever any key is pressed. The lighting will turn off  
automatically when no button has been pressed for two minutes, or can be set  
to stay on (see “Options Menu” on page 23).  
An alarm message will overwrite any other message on the screen. The original  
screen will reappear when any key is pressed.  
Start/Stop key  
Press the key once to start or stop treatment.  
Up/Down key  
Moves backwards and forwards within a menu and between menus.  
Increases and decreases values of parameters—press once to adjust in  
single increments; hold down to move quickly through the values.  
Using the VPAP Adapt SV  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Left key (green)  
Enters a menu.  
Confirms and applies settings.  
Performs the function indicated by the text above it in the LCD (eg, ’enter,  
’change’).  
Right key (red)  
Performs the function indicated by the text above it in the LCD (eg, ’exit’).  
Cancels operations.  
Extended hold: exits to the top level of the menu.  
QuickView key  
Takes you immediately to the Treatment screens, with information on your  
current therapy. See “Treatment Screens” on page 24.  
Alarm LEDs and Alarm mute  
The Alarm LEDs are lights that indicate that an alarm has been triggered. Alarms  
can be muted by pressing the Alarm mute key once. The Alarm LEDs will  
remain lit for as long as the alarm is being triggered. See “Alarms” on page 25.  
Adjusting Settings  
By using the Left, Right and Up/Down keys you can adjust some aspects of  
your treatment. The settings can be seen in the LCD display.  
To change a setting, press the Up/Down key until you arrive at the menu you  
require.  
Press the Left key to enter the menu, then use the Up/Down key again to  
scroll through the options.  
Once you find the option you require press the Left key to select the function  
that is displayed in the bottom left of the screen eg, ’change, yes’.  
Use the Up/Down key to alter the parameters and press the Left key when  
you have the setting you require.  
Press the Right key to exit from the menu.  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Example  
You wish to change the setting for the mask you are using.  
VPAP Adapt SV  
menu  
Press down’ arrow on Up/Down key  
to move to next menu  
Press Left key to enter menu options  
SETTINGS  
enter  
exit  
(ASV MODE)  
MASK: FULL FACE  
1. Press Left key to select ’change.  
2. Press Up/Down key to reach setting.  
3. Press Left key to enter setting.  
4. Press Right key to exit.  
change exit  
LEARN CIRCUIT?  
yes exit  
SMARTSTOP: OFF  
change exit  
SOUND LEVEL: LOW  
change exit  
RESULTS  
enter  
exit  
exit  
exit  
OPTIONS  
enter  
SERVICING  
enter  
Figure 14: Using the keys to adjust settings  
Using the VPAP Adapt SV  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Menus  
The options available in the Settings menu will depend on the therapy you are  
receiving (either ASV or CPAP), as set by your clinician.  
VPAP ADAPT SV  
menu  
SETTINGS  
enter  
exit  
CPAP MODE  
exit  
ASV MODE  
MASK: FULL FACE  
change exit  
RAMP: 20min  
change  
MASK: FULL FACE  
change exit  
LEARN CIRCUIT?  
yes exit  
LEARN CIRCUIT?  
yes exit  
SMARTSTOP: OFF  
change exit  
SMARTSTOP: OFF  
change exit  
SOUND LEVEL: LOW  
change exit  
SOUND LEVEL: LOW  
change exit  
RESULTS  
enter  
exit  
- -  
MASK FIT: ***  
good  
exit  
These menus appear  
only if at least one  
Smart Data option  
has been enabled by  
the clinician.  
AVG PRESS: 10.4  
exit  
USAGE: 4.56hrs  
exit  
SERVICING  
enter  
exit  
SN: 1234567890123  
4567 exit  
PCB: 123456789012  
exit  
exit  
SW: SX432-0201  
Figure 15: Navigating the Patient menu.  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Settings Menu  
You will only see the options relevant to your therapy. In CPAP (Continuous  
Positive Airway Pressure) mode, the clinician sets a pressure that will be  
constant throughout treatment. In ASV mode, the VPAP Adapt SV continually  
measures your breathing and regulates the air pressure accordingly.  
Mask  
Select the type of mask that you will be using (you will only need to adjust this if  
you have more than one mask). The following table shows the setting that  
should be selected for each mask type. If your mask type is not mentioned  
below, please contact your clinician. Not all mask types can be used with the  
VPAP Adapt SV.  
Option in Settings menu  
Use setting with:  
Vista  
Vista™  
Ultra  
Ultra Mirage™  
Full Face  
Mirage™ Full Face Mask Series II  
Ultra Mirage™ Full Face Mask  
Mirage Quattro™ Full Face Mask  
Activa  
Activa™  
Selecting mask options in the Settings menu  
Learn Circuit  
The Air Circuit Learn function takes approximately 20 seconds to run. It allows  
the unit to ’learn’ what pressure it needs to build in to its calculations, based on  
the components in the system. Select LEARN CIRCUIT every time you add or  
remove a component (eg, mask, humidifier) or move to a significantly different  
altitude.  
To run this function:  
1. Set up the entire VPAP Adapt SV system (flow generator, mask, humidifier  
etc) as you wish to use it.  
2. Make sure that the mask is unobstructed so air can flow from the mask to  
the flow generator.  
3. Turn on the VPAP Adapt SV at the power switch.  
4. Move through the menus until you reach LEARN CIRCUIT and select ’yes’ by  
pressing the Left key.  
5. A message will remind you to check that the airflow from the mask is clear.  
Select ’start’ with the Left key.  
6. A progress screen is displayed while the Air Circuit Learn function is running.  
LEARNING CIRCUIT  
Using the VPAP Adapt SV  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. When the function is complete, a screen will appear saying that the circuit  
has been learned successfully. The next menu item is then displayed.  
Commence treatment as described in “Starting Treatment” on page 16.  
If there is a problem with the circuit a screen saying ’Invalid circuit’ will be  
displayed instead of the progress screen. Check that all the connections  
between the pieces of equipment are secure. You may need to return to the  
previous chapter of this manual to check that you have set up the system  
correctly.  
SmartStop  
Turn the SmartStop function on or off. When SmartStop is on, the  
VPAP Adapt SV will automatically stop therapy when the mask is removed.  
Note: The unit will automatically ’SmartStart’ whenever you start to breathe  
into the mask.  
Leak Alert  
If your clinician has enabled Leak Alert, you will hear an alert tone sound when  
you have a high leak from your mask. You can then adjust the mask on your face  
to remove the leak. If Leak Alert is turned off, an extra leak alert message will be  
displayed at the end of a session if high leak occurred. The message will be  
displayed until a key is pressed.  
Ramp (CPAP mode only)  
Ramp time is the length of time taken for the air pressure to build up to the  
maximum pressure set by your clinician. This setting allows you to increase or  
decrease the ramp time in 5-minute intervals.  
If you find that you are still awake when the air pressure starts to build you may  
wish to increase the Ramp time.  
Results Menu  
This menu has a series of screens showing information about your most recent  
treatment session. If enabled by your clinician the following data may be viewed:  
Mask-fit  
Average Pressure  
Usage.  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Mask-fit  
Gives a rating of how well the mask was fitting during your last session on the  
VPAP Adapt SV. The stars indicate how good the fit was (more stars = better fit).  
Star rating  
Definition  
*****  
****-  
***--  
Excellent  
Very good  
Good  
**---  
Adjust mask  
Adjust mask  
Adjust mask  
*----  
HIGH LEAK  
Average Pressure  
Provides information on the average air pressure during your most recent  
session on the VPAP Adapt SV.  
Usage  
Tells you how long the VPAP Adapt SV was in use during your most recent  
session.  
Options Menu  
The Options menu allows you to change:  
Smart Data (Auto Appear)  
Backlight  
Language.  
Smart Data  
If the SMART DATA - AUTO APPEAR option is set to ON, the screens available  
under the Results menu will be displayed automatically for approximately 20  
minutes after your session is completed. After this time the data will still be  
available in the Results menu.  
If you want to turn off AUTO APPEAR, select ’enter’ in the SMART DATA screen  
by pressing the Left key. The AUTO APPEAR screen will be displayed. Select  
’change’ by pressing the Left key and you will switch from ON to OFF (similarly,  
you can turn it from OFF to ON). Your clinician may also turn off AUTO APPEAR.  
Backlight  
On: Display a permanent backlight for the LCD.  
Auto: Backlight turns off if no button has been pressed for two minutes.  
Language  
Change the language of the LCD display. Languages available are: English,  
French, Spanish and Portuguese.  
Using the VPAP Adapt SV  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Servicing Menu  
This menu displays the following information:  
SN—the flow generators serial number  
PCB—the flow generators PCB number (PCB—printed circuit board)  
SW—the version of software currently installed on the flow generator.  
Treatment Screens  
There are three Treatment screens displayed in the VPAP Adapt SV LCD. The  
LCD will display the Treatment screens when treatment commences, and during  
treatment if no key is pressed for 20 minutes.  
Press the QuickView key at any point within a menu to display  
Treatment Screen 1. You can return to your original screen if you press  
the QuickView key again within 20 minutes.  
Press the Right key to move from a Treatment screen to the Standby screen.  
If an alarm is triggered while you are in the Treatment screens, the arrow will  
flash. Press the Down key until you reach the alarm message.  
Scroll through the screens in the usual way, using the Up/Down key.  
Treatment mode Pressure  
ASV  
5.0--15.0  
exit  
Figure 16: Treatment Screen 1  
Leak  
Respiratory rate  
LK:18L/min  
RR:12  
VT: 428  
Tidal volume  
Figure 17: Treatment Screen 2  
Minute ventilation  
MV:6.2  
Target ventilation  
TARG:5.8  
exit  
Figure 18: Treatment Screen 3  
Note: TARG is not a relevant parameter in CPAP mode.  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Alarms  
The VPAP Adapt SV unit is fitted with alarms to alert you to changes that will  
affect your treatment. The most common reason for an alarm to sound is  
because the system has not been properly assembled. Check that the air tubing  
and pressure sensor tube have been properly attached to the flow generator and  
mask (and humidifier if used).  
You can mute an alarm by pressing the Alarm Mute key once. Unless the alarm  
requires you to turn the power off, you can press any key again to remove the  
alarm message from the LCD screen and to access all menus. If the problem is  
still present, the alarm will sound and display again after one minute. The Alarm  
LED will remain lit for as long as the problem is present.  
We recommend that you test the audible alarm once a week. To do this, hold  
down the Mute key while the VPAP Adapt is starting up. If the alarm is  
functioning correctly it will beep twice.  
LCD message  
Alarm type  
Action  
LOW PRESSURE!  
Check circuit  
Low mask pressure  
alarm  
Check tube connections.  
LCD turns off  
Power fail alarm  
Turn off VPAP Adapt SV at power switch.  
Turn off VPAP Adapt SV at power switch.  
HIGH PRESSURE!  
Turn power off  
High pressure alarm  
HIGH AVG PRESS!  
High pressure alarm  
High mask leak alert  
Turn off VPAP Adapt SV at power switch.  
Adjust the mask.  
HIGH LEAK!  
Check circuit  
LOW PS!  
Check circuit  
Insufficient pressure  
Support alarm  
Check tubing.  
Run LEARN CIRCUIT.  
FLOW BLOCKED!  
Turn power off  
Flow blocked alarm  
Turn off VPAP Adapt SV. Check air  
circuit. Turn on VPAP Adapt SV.  
Low Mask Pressure Alarm  
Triggered by:  
pressure sensor line is blocked or disconnected  
air pressure at the mask has fallen below a set level  
mask is removed and SmartStop has been disabled.  
If a low pressure alarm sounds:  
1. Check that the air tube and sensor tube are connected properly.  
2. Turn the VPAP Adapt SV off and on again at the power switch. If the alarm  
persists, return the unit to ResMed for servicing.  
Alarms  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Will clear when:  
the low pressure condition is fixed  
treatment is stopped by pressing the Start/Stop key.  
Power Fail Alarm  
Triggered by:  
power failure  
machine is disconnected or switched off while delivering treatment.  
If a power fail alarm sounds:  
the flow generator stops delivering air pressure.  
Will stop:  
when the Alarm Mute key is pressed, or  
after two minutes, or  
when power is restored.  
CAUTION  
!
Remove the mask from your face if the power fails.  
High Pressure / High Average Pressure Alarms  
Triggered by:  
High Pressure Alarm: the mask pressure exceeds the set trigger level (30  
cm H O) for more than 700 milliseconds.  
2
High Average Pressure Alarm: the mask pressure exceeds the set trigger  
level (22 cm H O), averaged over approximately one minute.  
2
If a high pressure alarm sounds:  
1. The treatment will stop.  
2. Turn power off.  
3. Check that the air tube and sensor tube are connected properly.  
4. Turn power back on.  
5. Remove mask and perform Air Circuit Learn function.  
6. Try using the flow generator one more time.  
7. If the high pressure alarm activates repeatedly, discontinue use and return to  
ResMed for servicing. If the alarm does not recur, then continue to use as  
normal.  
Note: The alarm system is designed to ignore coughing. However, if a cough is  
especially intense and prolonged, it may trigger the high pressure alarm.  
Will stop when:  
the VPAP Adapt SV is turned off.  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
High Leak Alert  
Triggered by:  
high mask leak (greater than 30 L/min) for more than 20 seconds  
the mask being removed while SmartStop is disabled.  
If a high leak alert sounds:  
adjust the mask to minimize leak  
check that the air tube and sensor tube are connected properly.  
Will stop when:  
the mask leak has been rectified.  
Note: The High Leak alert can be turned off in the Leak Alert option in the  
Settings menu.  
Insufficient Pressure Support Alarm  
Triggered by:  
air pressure at the mask has failed to reach an expected level for 3 breaths  
the Air Circuit Learn function (page 21) has not been run after adding a new  
component (eg, new mask, humidifier) to the system.  
If an insufficient pressure support alarm sounds:  
check that the pressure sensor tubing is not kinked  
check that the pressure sensor tubing is properly connected  
run LEARN CIRCUIT.  
Will stop when:  
the pressure sensor tubing is unobstructed and properly connected to the  
flow generator and mask  
treatment is stopped (Start/Stop key or SmartStop)  
the VPAP Adapt SV is turned off.  
Flow Blocked Alarm  
Triggered by:  
blockage in air circuit.  
If a flow blocked alarm sounds:  
1. The treatment will stop.  
2. Turn power off.  
3. Check whether there is a blockage in the air circuit.  
4. Remove blockage.  
5. Turn power back on.  
6. If the flow blocked alarm activates repeatedly, discontinue use and return to  
ResMed for servicing. If the alarm does not recur, then continue to use as  
normal.  
Alarms  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Will stop when:  
blockage is removed.  
Cleaning and Maintenance  
You should regularly carry out the cleaning and maintenance described in this  
manual.  
CAUTION  
!
Do not wash the pressure sensor tube. If fluid enters the pressure sensor  
tube allow it to dry completely by hanging it in a clean place out of direct  
sunlight. If the pressure sensor tube cannot be dried completely it should be  
replaced.  
Daily  
1. Disconnect the air tubing and pressure sensor tube and hang them in a clean,  
dry place until next use. Do not hang the air tubing in direct sunlight as it may  
harden and crack over time.  
2. Clean the mask according to the mask user instructions.  
3. If you are using a humidifier, clean it according to the instructions in the  
manual.  
Weekly  
1. Remove the air tubing and pressure sensor tube from the VPAP Adapt unit  
and the mask.  
2. Remove the pressure sensor tube and proximal cuff from the air tubing.  
3. Wash the mask system according to the instructions supplied with it.  
4. Wash the air tubing and proximal cuff in warm water using mild liquid soap.  
Rinse thoroughly, hang, and allow to dry.  
5. Before next use, assemble the mask and headgear according to the mask  
user instructions.  
6. Reconnect the pressure sensor tube to the air tubing, then connect both to  
the proximal cuff. Attach the proximal cuff to the mask.  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Air tubing  
Remove proximal cuff  
Remove the pressure sensor tube  
from air tubing  
Tubing clips  
Figure 19: Air tubing assembly removed for cleaning  
CAUTION  
!
Do not use bleach, chlorine-, alcohol-, or aromatic-based solutions  
(including all scented oils), moisturising or antibacterial soaps to clean  
the cushion, mask, air tubing, or the VPAP Adapt SV. These solutions  
may cause hardening and reduce the life of the product.  
º
º
Do not wash or dry the mask frame at a temperature above 176 F (80 C).  
Exposure to higher temperatures may reduce the life of the product.  
Do not hang the air tubing and pressure sensor tube in direct sunlight as  
the tubing may harden over time and eventually crack.  
Periodically  
1. The mask and air tubing are subject to normal wear and tear. Inspect them  
regularly for damage.  
2. Remove the power cord. Clean the exterior of the flow generator with a  
damp cloth and mild liquid soap.  
3. Inspect the air filter to check if it is blocked by dirt or contains holes. See full  
instructions below.  
WARNING  
!
!
Beware of electric shock. Do not immerse the flow generator or power cord  
in water. Always unplug the flow generator before cleaning and be sure that  
it is dry before reconnecting.  
CAUTION  
Do not attempt to open the VPAP Adapt SV. There are no user serviceable  
parts inside. Repairs and internal servicing should only be performed by an  
authorized service agent.  
Cleaning and Maintenance  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Air Filter  
Inspect the air filter every month and check if it is blocked by dirt or contains  
holes.  
Air filter cover  
The filter should be replaced every six months, or more often if the flow  
generator is operating in a dusty environment.  
WARNING  
!
Do not wash the air filter once it has become soiled. The air filter is not  
washable or reusable.  
Fitting the filter  
Remove the filter cover at the back of the flow generator.  
Remove and discard the old air filter.  
Insert a new filter, the blue tinted side facing towards you.  
Note: Match the cut corner of the filter with the shape of the casing.  
Replace the air filter cover. The blue tinted side of the filter should be facing  
out.  
Hypoallergenic Air Filter  
The ResMed hypo-allergenic filter is a standard filter to which an electrostatic  
filter has been bonded to capture very small particles. The filter is intended for  
those users of ResMed flow generators who will benefit from enhanced filtering  
of the air delivered during treatment.  
Inspect the air filter every month and check if it is blocked by dirt or contains  
holes.  
The filter should be replaced every three months, or more often if the flow  
generator is operating in a dusty environment.  
Fit as described above in “Air Filter”.  
Servicing  
This product (VPAP Adapt SV) should be inspected by an authorised ResMed  
Service Centre five years from the date of manufacture. Prior to this, the device  
is intended to provide safe and reliable operation provided that it is operated and  
maintained in accordance with the instructions provided by ResMed.  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Applicable ResMed warranty details are provided with the device at the time of  
original supply. Of course, as with all electrical devices, if any irregularity  
becomes apparent, you should exercise caution and have the device inspected  
by an authorized ResMed service center.  
If you feel the unit is not performing properly, see “Troubleshooting” on page 31.  
Servicing (Australia and New Zealand)  
This device is intended to provide safe and reliable operation when operated and  
maintained in accordance with the instructions provided by ResMed. To ensure  
that your device continues to provide optimum performance it is recommended  
that this product (VPAP Adapt SV) be inspected by an authorised ResMed  
Service Centre five years from the date of purchase.  
CAUTION  
!
Inspection and repair should only be performed by an authorized service  
agent. Under no circumstances should you attempt to service or repair the  
unit yourself.  
Troubleshooting  
If there is a problem, try the following suggestions. If the problem cannot be  
solved, contact your equipment supplier or ResMed. Do not attempt to open the  
flow generator.  
Problem  
Possible cause  
Solution  
Insufficient air delivered  
from flow generator  
Air filter is dirty.  
Replace air filter.  
Air tubing is kinked or  
punctured.  
Straighten or replace tubing.  
Flow generator operating  
erratically, ie, switching on  
and off, pressure varying at  
random  
Water may have entered the  
air tubing.  
Remove the water from the  
air tubing.  
Pressure rises  
inappropriately  
You are talking, coughing,  
voluntarily breath-holding  
(eg, while rolling over in bed),  
or intentionally breathing in  
an unusual manner.  
Avoid talking with the mask  
on, and breathe as normally  
as possible.  
Severe mask leak or mouth  
leak (greater than 30 L/min).  
Correct your leak condition.  
Alarm stays on continuously  
Blockage in air circuit.  
Ensure there is no barrier to  
the passage of air.  
Internal malfunction.  
Return the device to your  
equipment supplier for  
service.  
Troubleshooting  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Problem  
Possible cause  
Solution  
No LCD display  
Power not connected or  
switch at back is not on.  
Ensure the power cable is  
connected, and that the  
switch at the back of the unit  
is in the ’on’ position.  
FAULT message displayed  
on LCD  
Internal malfunction.  
Return the device to your  
equipment supplier for  
service.  
System Error 7 message  
Device power-on has been  
interrupted by breathing into  
the mask during warm up.  
Turn the flow generator off  
and on again. Do not place  
mask on face during warm  
up (30 seconds). Select  
LEARN CIRCUIT (page 21).  
If the error is not cleared,  
contact your equipment  
supplier for service.  
Internal malfunction.  
Return the device to your  
equipment supplier for  
service.  
Glossary  
Alarms  
When more than one alarm is activated at the one time, the highest priority  
alarm will be displayed.  
Backlight  
The backlight turns off if no button has been pressed for two minutes. It can  
also be kept permanently on by selecting On in the Backlight screen of the  
Options menu.  
bpm  
breaths per minute  
L/min  
liters per minute  
Mean  
The average value of the monitored parameter.  
Minute ventilation  
A measure of the respiratory rate x tidal volume.  
Proximal cuff  
A connector between the air tubing and the mask. The pressure sensor tub-  
ing attaches to the proximal cuff so that pressure at the mask can be  
detected and monitored.  
SmartStart  
A feature that starts the VPAP Adapt SV automatically when you breathe into  
the mask.  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
SmartStop  
A feature that stops the VPAP Adapt SV automatically when you remove the  
mask. You can turn SmartStop on or off in the Settings menu.  
Tidal volume  
The volume of air, in liters, breathed in or out in one breath.  
Technical Specifications  
Dimensions (H x W x D)  
5.6" x 9.8" x 11.5" (142 mm x 248 mm x 293 mm)  
Weight  
Flow generator with cover: 8.14 lbs (3.7 kg)  
Flow generator with HumidAire 2i (empty): 9.24 lbs (4.2 kg)  
Environmental conditions  
Operating temperature: 41 F to 95 F (+5 C to +35 C)  
Storage and transport temperature: -4 F to +140 F (-20 C to +60 C)  
º
º
º
º
º
º
º
º
Humidity—operating, storage and transportation: 10%–95% non-condensing  
Housing construction  
Injection moulded plastic  
Air tubing  
1 x 6’6” (2 m)  
Air filter  
Synthetic fiber  
Patient connection port  
22 mm taper, compatible with ISO 5356-1:2004 Anaesthetic & Respiratory  
Equipment—Conical Connectors  
Power supply  
AC Input range: 110–120V and 220–240V; 50–60Hz; 60VA  
DC Input range (if using DC Converter: 12–13.8V DC  
Performance  
Operating pressure range: 4–25 cm H O  
2
Sound pressure level  
<30 dBA (tested in accordance with the requirements of ISO 17510-1)  
Battery type  
Lithium  
Technical Specifications  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Alarm specifications  
High maximum mask pressure alarm  
Stops air delivery at 30 cm H O >700 msec.  
2
High mean mask pressure alarm  
Stops air delivery when mean mask pressure exceeds 22 cm H O (averaged  
2
over approx 1 minute).  
Low mask pressure alarm  
Mask pressure is less than 3 cm H O for longer than 5 seconds while the  
patient is using the VPAP Adapt SV.  
Insufficient pressure support alarm  
2
Is triggered if for 3 breaths in a row the maximum mask pressure achieved at  
inspiratory to expiratory transition is either <87.5% of the absolute mask  
pressure targeted at that point, or < absolute mask pressure targeted minus  
1 cm H O, whichever is the lower.  
2
High leak alarm  
Leak exceeds 30 L/min for more than 20 seconds.  
Flow blocked alarm  
Flow is less than 6 L/min for more than 20 seconds.  
IEC 601-1 classifications  
Class II, Type CF  
Class II medical electrical equipment provides protection against electric shock  
either by double insulation or reinforced insulation and does not require a  
protective earth.  
Inspiratory Trigger Characteristics  
The inspiratory trigger occurs at a flow greater than 0 LPM when tested with  
EPAP=5 cm H O, IPAP=15 cm H O, sinisoidal breath shape. The inspiratory  
2
2
trigger response time is <600 ms when tested with EPAP=5 cm H O,  
2
IPAP=15 cm H O, 0.67 L/sec inspiratory and expiratory flow rate. The time is  
2
measured from the beginning of inspiratory flow to the beginning of pressure  
rise to IPAP.  
Note: The ASV algorithm is designed to treat a patient with essentially normal  
lungs. Hence, trigger response is less sensitive and pressure ramp-up is more  
gentle than typical non-invasive ventilators.  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Displayed values  
Value  
Display  
Resolution  
Range  
Accuracy  
Proximal cuff  
Pressure  
-3 to 40 cm H2O  
0.5 cm H2O  
0.1 cm H2O  
Mass flow sensor  
Leak  
0–99 liters per minute  
8–30 breaths per minute  
0–4500 milliliters  
+/- 12 L/min  
+/- 10%  
1 liter per breath  
1 breath per minute  
1 milliliter  
Respiratory rate  
Tidal volume  
Minute ventilation  
Target MV  
+/- 10%  
0–40 liters per minute  
0–40 liters per minute  
+/- 10%  
0.1 liters per minute  
0.1 liters per minute  
+/- 10%  
Maximum low impedance flow at stated pressures  
Pressure (cm H O)  
Flow (L/min)  
2
4.2  
217  
218  
219  
8.6  
13.0  
Glossary of symbols  
Read the relevant operating manual before using  
Type CF equipment  
AC switch  
Class II  
Drip proof  
IPX1  
Technical Specifications  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Guidance and Manufacturer’s Declaration — Electromagnetic  
Emissions and Immunity  
Guidance and manufacturer’s declaration – electromagnetic emissions  
The VPAP Adapt SV is intended for use in the electromagnetic environment specified below. The  
customer or the user of the VPAP Adapt SV should assure that it is used in such an environment.  
Electromagnetic environment -  
guidance  
Emissions test  
Compliance  
RF emissions CISPR 11  
Group 1  
The VPAP Adapt SV uses RF energy  
only for its internal function.  
Therefore, its RF emissions are very  
low and are not likely to cause any  
interference in nearby electronic  
equipment.  
RF emissions CISPR 11  
Class B  
Class A  
The VPAP Adapt SV is suitable for use  
in all establishments, including  
domestic establishments and those  
directly connected to the public low-  
voltage network that supplies  
Harmonic Emissions  
IEC 61000-3-2  
Voltage Fluctuations/Flicker  
Emissions IEC 61000-3-3  
Complies  
buildings used for domestic purposes.  
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed  
and put into service according to EMC information provided in this document.  
Warnings: The VPAP Adapt SV should not be used adjacent to or stacked with other equipment.  
If adjacent or stacked use is necessary, the VPAP Adapt SV should be observed to verify normal  
operation in the configuration in which it will be used.  
The use of accessories (eg Humidifiers) other than those specified in this manual is not  
recommended. They may result in increased emissions or immunity of the VPAP Adapt SV.  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Guidance and manufacturer’s declaration – electromagnetic immunity  
The VPAP Adapt SV is intended for use in the electromagnetic environment specified below. The  
customer or the user of the VPAP Adapt SV should assure that it is used in such an environment.  
Electromagnetic  
environment –  
guidance  
IEC60601-1-2  
test level  
Compliance  
level  
Immunity test  
Electrostatic  
discharge (ESD)  
IEC 61000-4-2  
6 kV contact  
8 kV air  
6 kV contact  
8 kV air  
Floors should be wood,  
concrete or ceramic tile. If  
floors are covered with  
synthetic material, the  
relative humidity should be  
at least 30%.  
Electrical fast  
transient/burst  
IEC 61000-4-4  
2 kV for power  
supply lines  
2 kV  
Mains power quality should  
be that of a typical  
commercial or hospital  
environment.  
1 kV for input/  
output lines  
Not Applicable  
Surge  
IEC 61000-4-5  
1 kV differential  
mode  
1 kV differential  
mode  
Mains power quality should  
be that of a typical  
commercial or hospital  
environment.  
2 kV common  
mode  
2 kV common  
mode  
Voltage dips, short  
interruptions and  
voltage variations on  
power supply input  
lines.  
<5% Ut  
(>95% dip in Ut)  
for 0.5 cycle  
< 12V  
(>95% dip in 240V)  
for 0.5 cycle  
Mains power quality should  
be that of a typical  
commercial or hospital  
environment.  
40% Ut  
(60% dip in Ut)  
for 5 cycles  
96V  
If the user of the  
IEC 61000-4-11  
(60% dip in 240V)  
for 5 cycles  
VPAP Adapt SV requires  
continued operation during  
power mains interruptions,  
it is recommended that the  
VPAP Adapt SV be powered  
from an uninterruptible  
power source.  
70% Ut  
(30% dip in Ut)  
for 25 cycles  
168V  
(30% dip in 240V)  
for 25 cycles  
<5% Ut  
<12V  
(>95%dip in Ut)  
for 5 sec  
(>95%dip in 240V)  
for 5 sec  
Power frequency  
(50/60 Hz)  
magnetic field  
IEC 61000-4-8  
3 A/m  
3 A/m  
Power frequency magnetic  
fields should be at levels  
characteristic of a typical  
location in a typical  
commercial or hospital  
environment  
NOTE: Ut is the AC mains voltage prior to application of the test level.  
(Continued next page)  
Technical Specifications  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Guidance and manufacturer’s declaration – electromagnetic immunity (continued)  
The VPAP Adapt SV is intended for use in the electromagnetic environment specified below. The customer  
or the user of the VPAP Adapt SV should assure that it is used in such an environment.  
Immunity  
test  
IEC60601-1-2  
test level  
Complianc  
e level  
Electromagnetic environment –  
guidance  
Portable and mobile RF communications  
equipment should be used no closer to any  
part of the VPAP Adapt SV, including cables,  
than the recommended separation distance  
calculated from the equation applicable to the  
frequency of the transmitter.  
Recommended separation distance  
Conducted RF  
IEC 61000-4-6  
3 Vrms  
150 kHz to 80 MHz  
3 Vrms  
10 V/m  
d = 1.17 P  
Radiated RF  
IEC 61000-4-3  
10 V/m  
80 MHz to 2.5 GHz  
d = 0.35 P 80 MHz to 800 MHz  
d = 0.70 P 800 MHz to 2.5 GHz  
where P is the maximum output power rating  
of the transmitter in watts (W) according to the  
transmitter manufacturer and d is the  
recommended separation distance in meters  
(m).  
Field strengths from fixed RF transmitters, as  
determined by an electromagnetic site  
a
survey, should be less than the compliance  
b
level in each frequency range.  
Interference may occur in the vicinity of  
equipment marked with the following symbol:  
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.  
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by  
absorption and reflection from structures, objects and people.  
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and  
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted  
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an  
electromagnetic site survey should be considered. If the measured field strength in the location in which  
the VPAP Adapt SV is used exceeds the applicable RF compliance level above, the VPAP Adapt SV should  
be observed to verify normal operation. If abnormal performance is observed, additional measures may be  
necessary, such as reorienting or relocating the VPAP Adapt SV.  
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Recommended separation distances between portable and mobile RF communications  
equipment and the VPAP Adapt SV  
The VPAP Adapt SV is intended for use in an environment in which radiated RF disturbances are controlled.  
The customer or the user of the VPAP Adapt SV can help prevent electromagnetic interference by  
maintaining a minimum distance between portable and mobile RF communications equipment  
(transmitters) and the VPAP Adapt SV as recommended below, according to the maximum output power of  
the communications equipment.  
Separation distance according to frequency of transmitter  
m
Rated maximum output  
power of transmitter  
W
150 kHz to 80 MHz  
80 MHz to 800 MHz  
800 MHz to 2.5 GHz  
d = 1.17 P  
d = 0.35 P  
d = 0.7 P  
0.01  
0.1  
1
0.17  
0.37  
1.17  
0.04  
0.11  
0.35  
1.11  
0.07  
0.22  
0.70  
2.21  
7.00  
10  
3.69  
11.70  
100  
3.50  
For transmitters rated at a maximum output power not listed above, the recommended separation distance  
d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where  
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter  
manufacturer.  
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.  
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by  
absorption and reflection from structures, objects and people.  
Technical Specifications  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Limited Warranty  
ResMed warrants that your ResMed product shall be free from defects in  
material and workmanship for the period specified below from the date of  
purchase by the initial consumer. This warranty is not transferable.  
Product  
Warranty Period  
ResMed humidifiers, ResControl™, ResLink™, ResTraxx™  
ResMed flow generators  
1 Year  
2 Years  
90 Days  
Accessories, mask systems (including mask frame, cushion,  
headgear and tubing). Excludes single-use devices.  
Note: Some models are not available in all regions.  
If the product fails under conditions of normal use, ResMed will repair or replace,  
at its option, the defective product or any of its components. This Limited  
Warranty does not cover:  
a) any damage caused as a result of improper use, abuse, modification or  
alteration of the product;  
b) repairs carried out by any service organization that has not been expressly  
authorized by ResMed to perform such repairs;  
c) any damage or contamination due to cigarette, pipe, cigar or other smoke;  
d) any damage caused by water being spilled on or into a flow generator.  
Warranty is void on product sold, or resold, outside the region of original  
purchase. Warranty claims on defective product must be made by the initial  
consumer at the point of purchase.  
This warranty is in lieu of all other express or implied warranties, including any  
implied warranty of merchantability or fitness for a particular purpose. Some  
regions or states do not allow limitations on how long an implied warranty lasts,  
so the above limitation may not apply to you.  
ResMed shall not be responsible for any incidental or consequential damages  
claimed to have occurred as a result of the sale, installation or use of any ResMed  
product. Some regions or states do not allow the exclusion or limitation of  
incidental or consequential damages, so the above limitation may not apply to  
you. This warranty gives you specific legal rights, and you may also have other  
rights which vary from region to region.  
For further information on your warranty rights, contact your local ResMed dealer  
or ResMed office.  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Index  
right 18  
A
start/stop 17  
up/down 17  
air circuit learn function 21  
air filter 30  
hypoallergenic 30  
air pressure  
monitoring 23  
air tubing  
L
language  
change setting 23  
LCD display 31  
leak alert 22  
attaching to mask 9  
cleaning 28  
learn circuit 21  
Luer connector 9  
connecting 9  
air tubing system 6, 7, 9  
alarm  
M
mask  
flow blocked 27  
high leak 27  
connection 9  
indication of fit 23  
recommended 7  
select from menu 21  
medical information 3  
menus  
high pressure 26  
low mask pressure 25  
power fail 26  
alarm LED 18  
anti-bacterial filter 14  
how to use 19  
navigation 20  
B
backlight  
change setting 23  
O
Options menu 23  
oxygen  
C
cautions 4  
supplemental 15  
contraindications 3  
CPAP mode 21  
P
power cord  
D
connection 8  
pressure sensor tube  
connecting 9  
proximal cuff 7  
DC converter 15  
F
fault message 31  
Q
H
quick setup guide 5  
HumidAire  
connection 12  
HumidAire 2i 11  
humidifier  
connecting 10  
types 10  
R
Ramp 21, 22  
ResLink 14  
Results menu 22  
S
K
Settings menu 20  
Smart data 23  
SmartStop 22  
key  
alarm mute 18, 25  
left 18  
Index  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
standby mode 16  
starting treatment 16  
stopping treatment 16  
Supplemental oxygen 15  
System Error 7 message 32  
T
Treatment screens 24  
U
user/owner responsibility 3  
V
VPAP Adapt SV  
usage 23  
VPAP Adapt SV Components 6  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Quantum Mobility Aid Quantum 1402 User Guide
Radio Shack Headphones 43 216 User Guide
Radio Shack Universal Remote 15 2107 User Guide
Ricoh Printer L4130 User Guide
Samson Headphones SWC99BGT K; SWC99BGT I User Guide
Samsung Projection Television ST43L2HD User Guide
Sanyo Car Stereo System FXD 780RDS User Guide
Sanyo CRT Television DS13204, DS19204, DS25204 User Guide
Sealife Digital Camera DC300 DC310 User Guide
Sears Tent 30870002 User Guide