3.2 How to Sit Correctly
3.4 Using the Memory Function
1. Know Your Device
2. Preparation
2.1 Battery Installation
1. Remove the battery cover.
The monitor automatically stores the results up to 60 sets for each user (1 and
2). It can also calculate an average value based on the last 3 measurement
values taken within 10 minutes.
To take a measurement, you need to be
relaxed and comfortably seated, under
comfortable room temperature. Avoid
bathing, drinking alcohol or caffeine,
smoking, exercising or eating 30 minutes
before taking a measurement.
Contents:
Monitor, arm cuff, instruction manual, storage case, battery set,
USB cable, blood pressure pass
Notes:
Automatic Blood
Pressure Monitor
• If there are only 2 measurement values in the memory for that period, the
average will be based on these 2 values.
Monitor:
2. Insert 4 “AA” batteries as
indicated in the battery
compartment.
A
I
• If there is 1 measurement value in the memory for that period, this is
displayed as the average.
• If the memory is full, the monitor will delete the oldest value.
• When viewing the measurement value taken without setting the date and
time, “-:--” is displayed instead of the date and time.
• Sit on a chair with your feet flat on the floor.
• Sit upright with your back straight.
• Sit with your back and arm being supported.
• The arm cuff should be placed on your arm at
the same level as your heart.
Model M3 IT
Instruction Manual
H
G
F
E
J
IM-HEM-7131U-E-EN-01-08/2013
2298868-3A
3. Replace the battery cover.
Notes:
L
K
B
C
To View the Measurement Values Stored in Memory
1. Select your User ID (1 or 2).
• When the low battery symbol (
off, then replace all batteries at the same time. Long life alkaline batteries
are recommended.
• The measurement values continue to be stored in memory even after the
batteries are replaced.
• The supplied batteries may have a shorter life.
) appears on the display, turn the monitor
Introduction
Thank you for purchasing the OMRON M3 IT Automatic Blood Pressure Monitor.
A. Display
B. Memory button
C. START/ STOP button
D. Date/ Time setting button
E. USER ID selection switch
F. Cuff wrap guide lamp
G. Up/ Down buttons
H. Blood pressure colour indicator
I. Battery compartment
J. AC adapter jack
(for optional AC adapter)
K. USB port
2. Press the
button.
The OMRON M3 IT is a compact, fully automatic blood pressure monitor, operating
on the oscillometric principle. It measures your blood pressure and pulse rate simply
and quickly. For comfortable controlled inflation without the need of pressure
pre-setting or re-inflation the device uses its advanced “IntelliSense” technology.
3.3 Taking a Measurement
The Memory number appears for a second
before the pulse rate is displayed.
Notes:
Alternating date/ time
display
• To cancel a measurement, press the START/STOP button to release the air
in the arm cuff.
The newest set is numbered “1”.
L. Air jack
Intended Use
Disposal of used batteries should be carried out in accordance with the
national/local regulations for the disposal of batteries.
• Remain still while taking a measurement.
This device is a digital monitor intended for use in measuring blood pressure and
pulse rate in adult patient population who can understand this instruction manual
with the arm circumference range printed on the arm cuff. The device detects the
appearance of irregular heartbeats during measurement and gives a warning signal
with the measurement result.
The monitor is designed to take measurements and store the measurement
values in the memory for 2 people using USER ID 1 and USER ID 2.
Note: The cuff wrap guide result appears on
the display with the measurement
values. The cuff wrap guide lamp will not
light.
2.2 Setting the Date and Time
Set the monitor to the correct date and time before taking a measurement for
the first time.
Arm cuff:
1. Select your User ID (1 or 2).
M
M. Arm cuff (Arm circumference 22 - 42 cm)
N. Air plug
O. Air tube
Please read this instruction manual thoroughly before using the device.
Please keep for future reference. For specific information about your own blood
pressure, CONSULT YOUR PHYSICIAN.
1. Press the
button.
: The arm cuff was wrapped snugly enough.
: The arm cuff was loose or not wrapped correctly.
2. Push
or
to change the year
N
O
2. Press the START/ STOP button.
Push
button to confirm the year and then the month flashes. Repeat the
Important Safety Information
The arm cuff will start to inflate automatically.
same steps to change the month, day, hour, and minutes.
3. Press the
or
button to view the values stored in memory.
Year
Month
Day
Warning: Indicates a potentially hazardous situation which, if not avoided,
: To the older values
: To the more recent values
START
INFLATE
DEFLATE
COMPLETE
could result in death or serious injury.
(General Usage)
Display:
Consult your physician before using the device in pregnancy including
pre-eclampsia, or if diagnosed with arrhythmia or arteriosclerosis.
Do not use the device on the injured arm or the arm under medical treatment.
Do not apply the arm cuff on the arm while being on an intravenous drip or blood
transfusion.
Hour
Change Confirm
Y
Z
AA
P
Q
R
Minute
Change Confirm
To View the Average Value
1. Select your User ID (1 or 2).
Change Confirm
: Forward
Change Confirm
: Back
Consult your physician before using the device on the arm with an arterio-venous
(A-V) shunt.
S
T
AB
AC
Cuff Wrap Guide Lamp
Change Confirm
2. Press and hold the
button for
Do not use the device with other medical electrical (ME) equipment
simultaneously.
3. Press the START/STOP button to turn the monitor off.
more than 3 seconds.
Do not use the device in the area the HF surgical equipment, MRI, or CT scanner
exists, or in the oxygen rich environment.
Notes:
U
V
AD
• If the batteries have been replaced, the date and time setting will need to be
reset.
The air tube or the AC adapter cable may cause accidental strangulation in
infants.
AE
AF
• If the date and time are not set, “-:--” appears during or after measurement.
W
X
Notes:
Contained small parts that may cause a choking hazard if swallowed by infants.
If your systolic pressure is more than 210 mmHg
After the arm cuff starts to inflate, press and hold the START/ STOP
button until the monitor inflates 30 to 40 mmHg higher than your
expected systolic pressure.
• If the previous measurement was taken without setting the date and time,
the average value is not calculated.
(AC Adapter (optional) Usage)
P. OK symbol
Y. Memory symbol
Z. Average value symbol
AA. Date/Time display
AB. Movement error symbol
AC.Irregular heartbeat symbol
AD.Blood pressure level indicator
AE. Deflation symbol
Do not use the AC adapter if the device or the power cord is damaged. Turn off
the power and unplug the power cord immediately.
Plug the AC adapter into the appropriate voltage outlet. Do not use in a
multi-outlet plug.
3. Using the Device
• If there are no measurements results stored in the memory,
the screen to the right is displayed.
Q. Transfer indicator
R. USER ID symbol
S. DATA/FULL symbol
T. Systolic blood pressure
U. Diastolic blood pressure
V. Low battery symbol
W. Cuff wrap guide
3.1 Applying the Arm Cuff
Notes:
Remove tight-fitting clothing or tight rolled up sleeve from your left upper arm.
Do not place the arm cuff over thick clothes.
• The monitor will not inflate above 299 mmHg.
• Do not apply more pressure than necessary.
Never plug in or unplug the power cord from the electric outlet with wet hands.
Caution: Indicates a potentially hazardous situation which, if not avoided,
may result in minor or moderate injury to the user or patient or
damage to the equipment or other property.
1. Insert the air plug into the air jack
securely.
3. Remove the arm cuff.
4. Press the START/STOP button to turn the monitor off.
The monitor automatically stores the measurement result in its memory.
It will automatically turn off after 2 minutes.
AF. Pulse display / Memory number
To Transfer the Measurement Values
X. Heartbeat symbol
(General Usage)
(Flashes during measurement.)
To begin using the Medical Link software for the first time, please visit
Always consult your physician. Self-diagnosis of measurement results and
self-treatment are dangerous.
People with severe blood flow problems, or blood disorders, should consult a
physician before using the device, as the arm cuff inflation can cause bruising.
If there are any abnormalities during the measurement, remove the arm cuff.
Do not use this device on infants or persons who cannot express their intentions.
Do not inflate the arm cuff more than necessary.
1. Open the USB port cover.
Note: Wait 2-3 minutes before taking another measurement. Waiting between
measurements allows the arteries to return to the condition prior to
taking a measurement.
Display symbols:
Irregular Heartbeat Symbol (
)
Normal Heartbeat
When the monitor detects an irregular
rhythm two or more times during the
measurement, the irregular heartbeat
Pulse
2. Apply the arm cuff to your left
Do not use the device for any purpose other than measuring blood pressure.
Use only the approved arm cuff for this device. Use of other arm cuffs may result
in incorrect measurement results.
Do not use a mobile phone or other devices that emit electromagnetic fields,
near the device. This may result in incorrect operation of the device.
Do not disassemble the monitor or arm cuff.
Do not use in a location with moisture, or a location where water may splash on
the device. This may damage the device.
Blood pressure
upper arm.
Irregular Heartbeat
Using the Guest Mode
symbol ( ) will appear on the display
with the measurement values.
An irregular heartbeat rhythm is defined
as a rhythm that is 25% less or 25%
more than the average rhythm detected
2. Connect the USB cable to a PC.
Long
Short
The monitor stores measurement values for 2 users in the memory.
The guest mode can be used to take a single measurement for another user.
No measurement values are stored in the memory when the guest mode is
selected.
Pulse
Blood pressure
while the monitor is measuring the systolic and diastolic blood pressure.
1. Press and hold the START/STOP
Do not use the device in a moving vehicle (car, airplane).
Do not take measurements more than necessary. It may cause bruising due to
blood flow interference.
If the irregular heartbeat symbol ( ) displays with your measurement results,
we recommend you consult your physician. Follow the directions of your
physician.
button for more than 3 seconds.
The USER ID symbol and the Date/Time
display will disappear.
The bottom edge of the arm cuff
should be 1 to 2 cm above the elbow.
Air tube is on the inside of your arm
3. Connect the USB cable to the monitor.
Consult your physician before using the device if you had a mastectomy.
(AC Adapter (optional) Usage)
Movement Error Symbol (
)
and aligned with your middle finger.
Fully insert the power plug into the outlet.
1 - 2 cm
The movement error symbol is displayed if you move your body during the
measurement. Please remove the arm cuff, and wait 2 - 3 minutes.
Take another measurement, remain still during measurement.
When disconnecting the power plug from the outlet, do not pull the power cord.
Be sure to pull from the power plug safely.
2. Release the START/STOP button when the Date/Time display
When handling the power cord, take care not to do the following:
Do not damage.
Do not break it.
turns off.
Average Value Symbol (
)
Do not tamper with it. Do not forcibly bend or pull.
The arm cuff will start to inflate automatically.
The average value symbol is displayed when you press and hold the memory
button for more than 3 seconds. The most recent average value appears on
the display screen.
Do not twist.
Do not pinch.
Do not bundle during use.
Do not place under heavy objects.
To Delete All the Values Stored in Memory
1. Select your User ID (1 or 2).
3. Secure closed with the
Wipe the dust off from the power plug.
fabric fastener.
Always consult your physician. Self-diagnosis of measurement results and
self-treatment are dangerous.
Unplug monitor when not in use.
2. Press the Memory button, while the memory symbol (
)
Disconnect the power plug before cleaning.
Cuff Wrap Guide Lamp (
)
appears.
Use only the original adapter designed for this device. Use of unsupported
adapters may damage and/or may be hazardous to the device.
If the cuff was wrapped too loosely, it may cause unreliable values. If the
wrapping of cuff is too loose, the cuff wrap guide lamp does not light.
(Battery Usage)
Otherwise
lights in green. This is the function which is used as an aid in
Do not insert the batteries with their polarities incorrectly aligned.
Use only 4 “AA” alkaline or manganese batteries with this device. Do not use
other types of batteries. Do not use new and used batteries together.
Remove the batteries if the device will not be used for three months or more.
determining if the cuff is wrapped snugly enough.
3. While holding the
button
down, press the
Blood Pressure Colour Indicator
START/STOP button for
more than 3 seconds.
If your systolic or diastolic pressure is above the
standard range (135 mmHg for the systolic blood
pressure and/or 85 mmHg for the diastolic blood
pressure), the blood pressure colour indicator will light
in orange when the measurement result is displayed.
If they are within the standard range, no light will
appear.
Notes:
General Precautions
• When you take a measurement on the right
arm, the air tube will be at the side of your
• Do not forcibly crease the arm cuff or the air tube excessively.
• Do not press the air tube while taking a measurement.
• To unplug the air plug, pull on the air plug at the connection with the monitor, not
the tube itself.
elbow. Be careful not to rest your arm on the
air tube.
• The blood pressure can differ between the right arm and the left arm, and
therefore also the measured blood pressure values can be different.
OMRON recommends to always use the same arm for measurement. If the
values between both arms differ substantially, please check with your
physician which arm to use for your measurement.
Note: You cannot partially delete the values stored in the memory.
All values for the user you select will be deleted.
• Do not drop the monitor or subject device to strong shocks or vibrations.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not use the device outside the specified environment. It may cause an
inaccurate result.
• Read and follow the "Important information regarding Electro Magnetic
Compatibility (EMC)" in the "6. Specifications".
• Read and follow the "Correct Disposal of This Product" in "6. Specifications" when
disposing of the device and any used accessories or optional parts.
The JNC7* Guideline recommends the following guideline.
General Guidelines for Blood Pressure
Prehypertension at Office Hypertension at Home
Systolic Blood Pressure
Diastolic Blood Pressure
120 - 139 mmHg
80 - 89 mmHg
135 mmHg
85 mmHg
These are form statistical values for blood pressure.
* JNC7: The Seventh Report, 2003 Dec, of the Joint National Committee on
Prevention, Detection, Evaluation, and Treatment of Hi Blood
Pressure.
Download from Www.Somanuals.com. All Manuals Search And Download.
|