Leica Vacuum Cleaner ASP300 User Manual

Leica ASP300  
Advanced Smart Processor  
Automated vacuum  
tissue processor  
Instruction manual  
Leica ASP300  
V1.3 English - 12/2003  
Always keep this manual together with the instrument.  
Read this instruction manual carefully  
before working with the instrument.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NOTE  
The information, numerical data, notes and  
These are determined only by the contract  
provisions agreed between ourselves and our  
customers.  
value judgments contained in this manual rep-  
resent the current state of scientific knowl-  
edge and state-of-the-art technology.  
We are under no obligation to update the  
present manual according to the latest tech-  
nical developments, nor to provide our cus-  
tomers with additional copies, updates etc. of  
this manual.  
Leica reserves the right to change technical  
specifications as well as manufacturing pro-  
cesses without prior notice. Only in this way is  
it possible to continuously improve the tech-  
nology and manufacturing techniques used in  
our products.  
For erroneous statements, drawings, techni-  
cal illustrations etc. contained in this manual  
we exclude liability as far as permissible ac-  
cording to the national legal system applicable  
in each individual case. In particular, no liabili-  
ty whatsoever is accepted for any financial  
loss or consequential damage caused by or  
related to compliance with statements or other  
information in this manual.  
Statements, drawings, illustrations and other  
information as regards contents or technical  
details of the present manual are not to be  
considered as warranted characteristics of  
our products.  
This document is protected under copyright  
laws. Any copyrights of this document are re-  
tained by Leica Microsystems Nussloch  
GmbH.  
Any reproduction of text and illustrations (or of  
any parts thereof) by means of print, photo-  
copy, microfiche, web cam or other methods –  
including any electronic systems and media –  
requires express prior permission in writing by  
Leica Microsystems Nussloch GmbH.  
© Leica Microsystems Nussloch GmbH  
Published by:  
Leica Microsystems Nussloch GmbH  
Heidelberger Str. 17 - 19  
D-69226 Nussloch  
Germany  
Telephone: +49 (0) 62 24 143-0  
Fax: +49 (0) 62 24 143-200  
Leica ASP300  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of contents  
1. Important notes............................................................................................................................. 6  
1.1  
1.2  
Symbols used in this manual................................................................................................... 6  
Designated use .......................................................................................................................... 6  
2. Safety .............................................................................................................................................. 7  
2.1  
2.2  
2.3  
Safety instructions ..................................................................................................................... 7  
Warnings...................................................................................................................................... 8  
Integrated safety devices...................................................................................................... 10  
3. Instrument components and specifications .......................................................................... 11  
3.1  
3.2  
3.3  
3.4  
3.5  
Specific instrument options................................................................................................... 11  
Overview - instrument components .................................................................................... 12  
Standard delivery - packing list ............................................................................................ 14  
Technical data ......................................................................................................................... 15  
Compatible reagents ............................................................................................................... 18  
4. Setting up the instrument .......................................................................................................... 19  
4.1  
4.2  
4.3  
4.4  
4.5  
4.6  
4.7  
Site requirements..................................................................................................................... 19  
Electrical connection.............................................................................................................. 19  
Installing the accessories ..................................................................................................... 21  
Setting up the data connections.......................................................................................... 22  
Alarm functions ........................................................................................................................ 23  
Switching the instrument on ................................................................................................. 24  
Touch screen functions......................................................................................................... 26  
4.7.1 User interface - overview ...................................................................................................... 26  
4.7.2. Access control ......................................................................................................................... 27  
4.8  
Check list for first operation (at supervisor access level).............................................. 28  
5. Operation...................................................................................................................................... 29  
5.1  
Reagent handling..................................................................................................................... 29  
5.1.1 Fill/drain reagents (other than wax) .................................................................................... 29  
5.1.2 Changing the wax .................................................................................................................... 31  
5.2  
Setting up the instrument parameters ................................................................................ 34  
5.2.1 System setup............................................................................................................................. 34  
5.2.2 Editing the reagent list ............................................................................................................ 37  
5.2.3 Viewing the program list......................................................................................................... 40  
5.2.4 Adding and/or modifying programs .................................................................................... 41  
5.2.5 Favorites (maximum of 5 programs) .................................................................................... 44  
Instruction manual V 1.3 – 12/2003  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of contents  
5.2.6 Editing the stations .................................................................................................................. 46  
5.2.7 Reagent groups........................................................................................................................ 47  
5.3  
Programs.................................................................................................................................... 48  
5.3.1 Editing a selected program ................................................................................................... 49  
5.3.2 Starting a program ................................................................................................................... 51  
5.4  
5.5  
5.6  
5.7  
Reagent status ......................................................................................................................... 52  
System Monitor......................................................................................................................... 53  
On-line help ............................................................................................................................... 54  
Configuring a printer (optional)............................................................................................. 55  
6. Trouble shooting ........................................................................................................................ 56  
6.1  
6.2  
6.3  
6.4  
General ....................................................................................................................................... 56  
Power failure ............................................................................................................................. 56  
Trouble shooting ...................................................................................................................... 57  
Typical fill or drain problems ................................................................................................. 58  
7. Cleaning and maintenance ...................................................................................................... 59  
Clean programs ........................................................................................................................ 59  
7.1  
7.1.1 Retort clean programs ............................................................................................................ 59  
7.1.2 Wax clean programs .............................................................................................................. 61  
7.1.3 Smart clean ............................................................................................................................... 62  
7.2  
General cleaning steps .......................................................................................................... 63  
7.3.1 Daily cleaning and maintenance ......................................................................................... 65  
7.3.2 Periodic cleaning and maintenance................................................................................... 66  
7.4  
Preventative maintenance schedule ................................................................................. 67  
8. Warranty and service ................................................................................................................ 69  
Leica ASP300  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
1. Important notes  
1.1 Symbols used in this manual  
and their meaning  
1.2 Designated use  
The Leica ASP300 is an automated tissue pro-  
cessor for the following laboratory applications:  
Warnings  
appear in a grey box and are marked  
by a warning triangle  
fixation  
dehydration  
.
paraffin wax infiltration  
of histological tissue specimens.  
Notes,  
i.e. important user information appear  
in a grey box and are marked by an  
The Leica ASP300 must be operated exclu-  
sively with the reagents listed in  
information symbol  
.
Chapter 3.5 - “Compatible reagents“.  
Any other use of the instrument  
is considered improper!  
Flammable solvents and reagents are  
marked with this symbol.  
Instrument surfaces which become  
hot during operation are marked with  
this symbol.  
Figures in brackets refer to item nos.  
in drawings.  
(5)  
Function keys to be pressed on the  
instrument touch screen are written  
in bold-print capital letters.  
ENTER  
Instruction Manual V 1.3 – 12/2003  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Safety  
Make sure to comply with the safety instructions and warnings in this chapter.  
Make sure to read these instructions, even if you are already familiar with the operation and  
use of other Leica products.  
2.1 Safety instructions  
This instruction manual includes important in-  
formation related to the operating safety and  
maintenance of the instrument and it is an im-  
portant part of the product.  
This instrument has been built and tested in  
accordance with the following safety regula-  
tions on electrical measuring, control, regulat-  
ing and laboratory devices:  
DIN EN ISO 9001  
If additional requirements on accident  
prevention and environmental protec-  
tion exist in the country of operation,  
this instruction manual must be  
supplemented by appropriate instruc-  
tions to ensure compliance with such  
requirements.  
EN 61010-1  
EN 61010-2  
EN 61326-1 and EN 61326-2  
EN 61000-3-2 and EN 6100-3-3.  
Instrument type  
The protective devices on both in-  
strument and accessories may neither  
be removed nor modified.  
Only authorized and qualified service  
personnel may repair the instrument  
and access the instrument’s internal  
components.  
All information provided in this manual applies  
only to the instrument type indicated on the title  
page.  
A name plate indicating the instrument serial  
number is attached to the back of the instru-  
ment.  
Leica ASP300  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Safety  
2.2 Warnings  
The safety devices installed in this instrument by the manufacturer only constitute the basis  
for accident prevention.  
Primarily responsible for accident-free operation is above all the institution which owns the  
instrument and, in addition, the designated personnel who operates, services or repairs the  
instrument.  
To ensure trouble-free operation of the instrument, make sure to comply with the following  
instructions and warnings.  
Transport and installation  
Warnings and instructions on transport and installation of the instrument can be found on a  
separate information sheet attached to the outside of the instrument’s shipping crate.  
Warnings - reagent handling  
To avoid damage to the instrument, only use the reagents listed in chapter 3.5!  
When handling reagents, follow the instructions of the manufacturer and make sure all  
laboratory regulations in force in your country are complied with.  
Dispose of waste solvents with care according to local regulations and the waste manage-  
ment policy of the company or institution.  
Use caution when handling hot, molten paraffin.  
Do not touch paraffin stations and retort walls – they can be very hot.  
Do not clean the reagent bottles in an automatic dishwasher.  
Material safety data sheets can be obtained from the supplier of the chemicals.  
They are also available on the Internet:  
Instruction manual V 1.3 – 12/2003  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Safety  
Warnings - cleaning and maintenance  
Prior to each maintenance switch the instrument off and disconnect mains power.  
Do not clean the instrument with solvents containing acetone or xylene. No liquid may be  
spilled into the internal components of the instrument.  
Empty the condensate bottle at regular intervals.  
Warnings - working with the instrument  
The Leica ASP300 may only be operated by trained laboratory personnel and as per the  
present instruction manual.  
In case of emergency switch off mains (item no. 5 in Fig. 3) at the rear of the instrument and  
unplug the power cord.  
The instrument must be installed so that the mains switch at the rear of the instrument can  
easily be accessed at any time.  
Always press PAUSE before opening the retort lid during processing to allow the retort to  
vent.  
Do not remove the wax drain hose or the remote fill/drain hose until the fill or drain has been  
entirely completed.  
After refilling / replacing the reagent bottles ensure that the lids are tight.  
The bottles must be properly pushed home into the connection manifolds at the rear inner  
wall of the reagent module.  
Fixatives containing mercuric salts, acetic or picric acid will corrode metallic components  
in the instrument.  
After each paraffin step a retort clean cycle must be run.  
Leica ASP300  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Safety  
2.3 Integrated safety devices  
The instrument is equipped with the following protection features:  
overpressure protection,  
protection against excessive vacuum,  
overcurrent protection,  
overheating protection.  
Overpressure protection  
The design of the pressure system ensures that no dangerous levels of  
pressure can build up.  
Protection against excessive vacuum  
The vacuum system is not capable of generating a dangerous vacuum  
condition.  
Overcurrent protection  
Overheating protection  
Overcurrent conditions are protected against by both the mains fuse  
and the separate heating power fuses.  
The design of the heating system ensures that no dangerous overheat-  
ing conditions can occur.  
Instruction manual V 1.3 – 12/2003  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Instrument components and specifications  
3.1 Specific instrument options  
Reagent Management System (RMS).  
The specimen cassette baskets may be used with or without cas-  
sette dividers and spring separators designed to separate cassettes  
and improve fluid penetration.  
In-process fluid recirculation (‘tidal agitation’) for continuous and ef-  
ficient mixing of fluids.  
Remote fill/drain system for reagents.  
Remote fill/drain system for paraffin.  
Automatic optical fluid level sensor.  
Active paraffin wax solvent removal feature.  
Magnetic stirrer.  
Programmable program start time delay.  
Programmable finish time for processing programs.  
3-step retort drain feature.  
Specimen processing under pressure, vacuum, alternating pressure  
/ vacuum cycles or under ambient pressure.  
User-programmable clean cycles.  
Leica ASP300  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Instrument components and specifications  
3.2 Overview - instrument components  
4
3
23  
22  
6
7
1
10  
9
5
21  
20  
13  
12 11 8  
19  
17  
18  
2
15  
14  
16  
Fig. 1  
Instruction manual V 1.3 – 12/2003  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Instrument components and specifications  
Overview - Instrument components  
Instrument components and accessories  
1
2
3
4
5
6
7
8
9
-
-
-
-
-
-
-
-
-
Basic instrument - processor module  
The processor module comprises three paraf-  
fin wax baths and the processing retort.  
Basic instrument - reagent module  
Wax bath lids  
Three baskets (21) contain the cassettes.  
Each basket can hold up to 300 cassettes.  
Retort lid  
All processing occurs in the stainless steel re-  
tort under the pressure, vacuum and tempera-  
ture conditions selected.  
Touch screen  
Cover flap of instrument console  
Instrument console comprising:  
Printer port  
The reagent bottles are stored in the reagent  
cabinet.  
Local alarm connection  
Remote alarm connection  
Floppy disk drive  
10 -  
11 -  
12 -  
Serial port  
13 - Activated carbon filter  
14 - Condensate bottle  
15 - Reagent bottle (13 units)  
16 - Drip tray  
17 - Wax drain connection  
18 - Remote drain connection  
19 - Retort  
20 - Wax baths  
21 - Specimen basket  
22 - Lid for specimen basket  
23 - Splash protector  
Leica ASP300  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Instrument components and specifications  
3.3 Standard delivery - packing list  
Leica ASP300 basic standard delivery:  
1
1
Leica ASP300 - Basic instrument  
Accessory kit consisting of:  
- 1 Basket hook  
0476 34100  
0476 35929  
0476 34713  
- 3 Specimen baskets, assy., including lid, dividers, springs and handle  
0476 34193  
(accessories for maximum and optimum usage of basket space)  
- 1 Reagent bottle, plastic  
0476 34274  
0476 34150  
- 2 Activated carbon filters, assy.  
- 1 Remote fill/drain hose, assy. with special connecting device  
for remote filling / draining of processing retort  
- 1 Paraffin drain hose  
0476 34716  
0476 34721  
0476 35923  
0336 35460  
0411 34604  
0476 35921  
0476 34202  
0476 35792  
0476 34770  
- 1 Paraffin wax scraper, plastic  
- 1 Lubricant ‘Molykote 111’, for valves and O-rings, 100 g  
- 1 Jumper cable - mains  
- 1 Maintenance kit (2 spare lids, 9 O-rings)  
- 1 Stirrer bar, assy.  
- 2 Sets of labels for reagent bottles, each set containing 15 labels  
- 1 Splash protector  
1
Set of power cords:  
- 1 Power cord D  
0411 13558  
0411 27822  
0476 34274  
0476 34278  
0476 35922  
0476 34145  
0476 34852  
0476 34162  
- 1 Power cord UK ST-BU F-5A  
13 Reagent bottles, plastic (included in the instrument)  
1
1
1
1
1
Condensate bottle, plastic (included in the instrument)  
Service kit for pump (included in the instrument)  
Drip tray (included in the instrument)  
Cover sheet for overflow (included in the instrument)  
Console cover (mounted to the instrument)  
1
1
1
3,5 Diskette, empty (taped to the back of console cover)  
Instruction manual Leica ASP300 multilingual  
Demo-Program, CD-Rom  
0704 37102  
Instruction manual V 1.3 – 12/2003  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Instrument components and specifications  
3.4 Technical data  
Nominal voltage:  
Two factory-preset voltage settings  
(not user-adjustable):  
100 to 120 V + 10% or  
230 to 240 V + 10%  
Nominal frequency:  
Main fuses:  
50 to 60 Hz  
2 fast-blowing melting fuses, 20 x 5 mm, UL-listed  
Order no.  
100 to 120 V; F 10 A 250 VAC  
230 to 240 V; F 5 A 250 VAC  
69411002  
69411000  
Nominal power:  
1,000 VA  
Dimensions (W x D x H), in mm:  
Dry weight, unpacked:  
Weight, packed:  
595 x 680 x 1,325  
approx. 140 kg  
220 kg  
Operating temperature range:  
Relative humidity:  
15 °C to 35 °C  
10% to 80% - non-condensing  
IEC 1010 classification:  
Protective Class 1  
Pollution Degree 2  
Overvoltage Installation Category II:  
800 V impulse (120 V systems)  
1500 V impulse (240 V systems)  
Maximum operating altitude:  
Local / remote alarm relays:  
maximum 2,500 m  
30 V DC, maximum 2 A  
2 connections:  
both single voltage-free, change-over contacts  
(can be operated as both normally-open and  
normally-closed connections)  
Paraffin wax baths  
Number of baths:  
Volume (liters):  
Melt time:  
3
4.3 l per bath  
approx. 10 hrs  
Temperature:  
40 to 65 °C  
+ 1 °C  
Temperature accuracy:  
Leica ASP300  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Instrument components and specifications  
Retort  
Capacity:  
max. 300 cassettes  
Reagent volume (in liters):  
Temperature (paraffin):  
Temperature (processing reagents):  
Temperature (cleaning reagents):  
Temperature accuracy:  
Filling time:  
4.3  
40 to 65 °C  
o
ambient or 35 to 55 C  
50 to 65 °C  
+ 1 K  
approx. 90 sec  
approx. 80, 120, 140 sec (adjustable)  
-70 kPa (g)  
Drain time :  
Impregnation vacuum:  
Impregnation pressure:  
Fill vacuum:  
35 kPa (g)  
-70 kPa (g)  
Drain pressure:  
35 kPa (g)  
General  
Reagent bottles  
Cleaning solution bottles:  
Maximum bottle volume:  
10  
3
4.3 l  
Printer:  
Optional accessory  
ON/OFF  
Pretest check:  
Fluid level sensor:  
ON/OFF  
Recirculation (pump in/out):  
(a) Time before 1st cycle:  
(b) Time between cycles:  
ON/OFF  
12 min  
20 min  
System setup  
Password status:  
Type of password:  
Supervisor / User  
alphanumeric, freely selectable  
Reagent Management Status:  
ON/OFF  
ON/OFF  
Software access control (‘Lock Mode’):  
Instruction manual V 1.3 – 12/2003  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Instrument components and specifications  
3.4 Technical Data (continued)  
Hardware and Software:  
Large color liquid crystal display (LCD) and touchscreen.  
User-friendly, intelligent software.  
3,5“ floppy drive and printer port.  
Alarm system with two remote alarm sockets.  
Password-protected instrument supervisor mode.  
Multiple specimen protection system.  
Capacities:  
15 programs of up to 10 reagent and 3 paraffin wax steps each.  
Time per program step: 0 to 99 hours , 59 minutes.  
Delay time: max. 7 days.  
Up to 300 cassettes can be processed simultaneously.  
Four user-programmable retort clean programs.  
Wax clean program.  
10 Reagent bottles.  
3 Wax baths.  
3 Cleaning solution bottles.  
1 Condensate bottle.  
Reagent temperature adjustable from 35 °C to 55 °C  
or ambient.  
Paraffin temperature adjustable from 40°C to 65 °C.  
Choice of three retort drain rates of 80, 120 and 140 seconds.  
Up to 100 reagent names in memory.  
Leica ASP300  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Instrument components and specifications  
3.5 Compatible reagents  
Fixatives  
1. Formalin, buffered or unbuffered  
2. Formalin saline  
Dehydration  
1. Ethanol  
2. Isopropanol  
3. Methanol  
4. Butanol  
5. Industrial Methylated Spirits  
Clearing  
1. Xylen (recommended)  
2. Toluol *  
3. Chloroform *  
*
Prior to using these reagents, contact  
your Leica representative or Leica di-  
rectly for information on necessary pro-  
tective measures. Using these reagents  
will require an increased number of main-  
tenance interventions by Leica Techni-  
cal Service.  
Paraffin wax  
1. Paraffin  
Caution:  
Reagents other than those listed here may damage some com-  
ponents of the instrument.  
Do not use acetone, benzene or trichlorethane in the instru-  
ment.  
Fixatives containing mercuric salts, acetic or picric acid will  
corrode metallic components in the instrument and shorten in-  
strument life.  
If you choose to work with such fixatives, it is essential to per-  
form a clean cycle which contains multiple water rinses each  
time after use, to minimize damage.  
In addition, we recommend frequent and regular preventive  
maintenance by the Leica Technical Service.  
Instruction manual V 1.3 – 12/2003  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Setting up the instrument  
4.1 Site requirements  
The instrument needs approx. 650 x 700 mm floor space.  
Practically vibration-free floor.  
Room temperature constantly between +15 °C and +35 °C.  
Relative humidity maximum 80%, non-condensing.  
Avoid vibrations, direct sunlight and heavy variation in temperature.  
The chemicals to be used in the Leica ASP300 are both flam-  
mable and noxious. Therefore, the instrument must be set up in  
a well ventilated area, free from any ignition sources.  
The room where the instrument is installed is not suitable for  
people to stay in for longer periods of time, unless it is  
equipped with a fume extraction system.  
Do not operate the instrument in rooms with explosion hazard.  
Moving the instrument  
The instrument must be installed such that the mains switch at  
the rear of the instrument (item no. 5 in Fig. 3) can easily be ac-  
cessed at any time.  
After the instrument has been unpacked (see unpacking instructions  
attached to outside of transport crate) it must be touched only at the  
points marked with dots ‘ ‘ for transport to its final site of installation.  
Once the instrument has been installed there, the brakes on the instru-  
ment rollers must be locked.  
4.2 Electrical connection  
Check the voltage label (Fig. 2) on the rear of the instrument to ensure  
that the instrument is set to the correct voltage range.  
Fig. 2  
The power supply voltage for the instrument is factory-preset  
and CANNOT be altered by the user.  
Leica ASP300  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Setting up the instrument  
Connecting the instrument to mains  
The instrument MUST be connected to an earthed mains power  
outlet socket.  
It is recommended that the ASP300 be plugged into a wall socket  
that has Ground Fault Circuit Interruption (GFCI) protection - as  
an additional electrical safeguard.  
Make sure to use the appropriate mains cable for the local volt-  
age supply (wall outlet).  
Connect mains cable (6) to mains power supply socket (1) (see Fig. 3).  
Jumper cable (3) connects the mains power supply output (2) to the  
electronics module input (4).  
Switch ON/OFF switch (5) ON.  
Switch ON/STOP switch at the right of the instrument ON.  
Electrical  
connections at the  
rear panel of the  
instrument  
1
6
5
4
2
3
Cables  
connected  
correctly  
Fig. 3  
Important:!  
Specification for connection (2) is as follows:  
100 - 120 V or 230 - 240 V, max. 200VA.  
Instruction manual V 1.3 – 12/2003  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Setting up the instrument  
4.3 Installing the accessories  
6
Unpack the activated carbon filter and in-  
sert it into the instrument (see Fig. 4). Make  
sure to insert the filter with the correct side  
facing upwards. See label (6) at the front of  
the filter - the arrow must point upwards.  
7
l
Unpack the magnetic stirrer and insert into  
the retort.  
Install the console flap (covers floppy drive  
and ports).  
Fig. 4  
Prior to filling the bottles with reagent for the  
first time, check the connections. If neces-  
sary, tighten manually.  
Remote fill / drain hose  
Connect the remote fill / drain hose as  
shown in Fig. 5.  
Important!  
8
When inserting the hose into the fill/drain  
connection (item 7 in Fig. 5) the connecting  
device (8) of the hose must lock with an au-  
dible click.  
7
Fig. 5  
Leica ASP300  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Setting up the instrument  
4.4 Setting up the data connections  
9
8
Printer (optional)  
Lift off the instrument console flap (12).  
Connect the printer to the printer port (7) via  
a standard Centronics cable (refer Fig. 6).  
7
Other connections  
10  
For additional data transfer and storage the  
instrument console is equipped with a stan-  
dard 1.44 MB floppy drive (10) for 3.5dis-  
kettes and with an RS 232 serial port (11) .  
11  
12  
Only instruments which satisfy the re-  
quirements for SELV circuits may be  
connected to interfaces (7), (8), (9)  
and (11).  
Fig. 6  
Instruction Manual V 1.3 – 12/2003  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Setting up the instrument  
4.5 Alarm functions  
The Leica ASP300 is equipped with two different alarm functions:  
Instrument alarm  
This alarm is generated from within the Leica ASP300. The instrument  
alarm is generated in all alarm situations.  
Remote alarm  
This alarm device is installed outside of the Leica ASP300.  
All alarms (all types of alarm) are forwarded to the remote dialer con-  
nected to the instrument, regardless of whether the local instrument  
alarm (type 1) is connected or disconnected.  
The optional remote alarm system works via a potential-free  
relay. If an alarm situation occurs, the circuit is closed.  
The remote alarm device connected to the instrument must be  
rated at less than 2 amp.  
A maximum voltage of 30 V DC may be present.  
Connecting a remote alarm device:  
A remote alarm device can be connected to the Leica ASP300 via a 6.3  
mm-diameter jack connector (optional accessory).  
For details on how to connect a remote alarm device to the Leica  
ASP300, please contact your local Leica office or the manufacturer  
Leica Microsystems Nussloch directly.  
Leica ASP300  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Setting up the instrument  
4.6 Switching the instrument on  
Once the instrument has been switched on, the mains switch (ON/OFF) (5) should always re-  
main in position ON.  
After switching it on the instrument will take  
a few minutes to initialize (see opposite -  
Fig. 7).  
After initialization, the FAVORITES screen will  
be displayed (Fig. 8).  
Screen saver  
Fig. 7  
A screen saver will operate turning the  
screen off when no key has been pressed  
for a (user-) programmable time.  
Press any part of the touchscreen to restore  
the screen.  
After having been restored, the screen will  
be inoperational for a few seconds to avoid  
accidental activation of any key symbols.  
.
Fig. 8  
Instruction Manual V 1.3 – 12/2003  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Setting up the instrument  
System diagnostics menu  
If, during initialization, ‘Touch here  
...’ is touched (refer Fig. 7) and sub-  
sequently the supervisor pass-  
word is entered, the SYSTEM  
DIAGNOSTICS menu (Fig. 9) opens  
up, giving access to standard in-  
strument settings.  
6
7
8
2
3
1
4
5
Attention!  
Serious instrument malfunctions  
may occur if any settings are  
modified wrongly.  
Fig. 9  
The following functions can be selected:  
1 - Access for service engineers only.  
2 - Access INSTALLATION menu.  
3 - Abort the current program.  
To exit this menu, the instrument has to be re-  
started - press: RESTART APPLICATION.  
Press YES to confirm ‘ARE YOU SURE YOU  
WANT TO ...’ (refer Fig. 10) .  
4 - Abort the current program and delete the  
current allocation of reagents to reagent  
bottles and retort.  
Initialization starts again from the screen  
shown in Fig. 7.  
5 - Delete all programs and reset instrument  
status (all lists will be empty).  
6 - Store actual instrument status on floppy  
disk.  
7 - Display run log.  
8 - Restart instrument.  
Fig. 10  
Leica ASP300  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Setting up the instrument  
4.7 Touch screen functions  
4.7.1 User interface - overview  
The Leica ASP300 is programmed and operated via a colored LCD  
touch screen.  
An on-line help feature provides detailed information on all instrument  
functions.  
HELP can be accessed from any screen by pressing the HELP button.  
Button symbols  
To activate any functions on the Leica ASP300 touchscreen, press the  
corresponding button symbols. Examples of buttons are:  
Enabled  
Disabled  
Buttons on the Leica ASP300 touch screen can either be enabled or dis-  
abled.  
A button that is disabled appears with a thinner border than a button  
that is enabled.  
Instruction manual V 1.3 – 12/2003  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Setting up the instrument  
4.7.2. Access control  
The Leica ASP300 may be configured to allow two levels of user ac-  
cess.  
‘Operator’ symbol  
Operator access level:  
Operators may run programs and view results. On this level, the OP-  
ERATOR symbol is displayed in the upper right corner of the touch  
screen.  
‘Supervisor’ symbol  
Supervisor access level:  
Supervisors may perform all Operator functions, and additionally cre-  
ate programs and perform the instrument sit-up functions.  
To access Supervisor level, press SUPERVISOR, enter the required  
password and confirm. The SUPERVISOR symbol will then be dis-  
played instead of the OPERATOR symbol and all enabled buttons will  
be surrounded by a black border instead of a blue one.  
Lock feature against unauthorized use  
The Leica ASP300 provides a feature called "Lock" to prevent ac-  
cess to any operations of the Leica ASP300 to unauthorised users.  
The ‘Lock’ mode and the corresponding ‘Lock’ password are config-  
ured from the SYSTEM SETUP menu.  
Leica ASP300  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Setting up the instrument  
4.8 Check list for first operation (at supervisor access level)  
Parameter settings  
Screen  
Press button  
Enter instrument name and select lan-  
guage.  
Activate printer (optional).  
Enter carbon filter age warning and pump  
warning threshold.  
Set all parameters as required, especially  
wax bath set temperature.  
Check whether date and time are correct.  
Add the desired reagents and edit the  
‘REPLACE REAGENTS’ warning thresholds.  
Allocate a reagent to each of station.  
Fill the reagent stations (bottles), either via  
the ASP300 remote drain/ fill function  
(SMART SCREEN) or manually (REAGENT  
STATUS).  
or  
Create the programs you need.  
Programs can be copied and modified.  
Allocate the most popular programs to the  
FAVORITES menu and, if desired, set a fin-  
ish time for these programs and allocate a  
symbol.  
Instruction manual V 1.3 – 12/2003  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.1 Reagent handling  
5.1.1 Fill/drain reagents (other than wax)  
1. Manually changing the reagents  
Remove the appropriate reagent bottle from the Leica ASP300 re-  
agent cabinet, remove the bottle screw cap.  
Drain/refill the reagent by pouring to/from a bulk container.  
On the REAGENT STATUS screen, verify correct allocation and set  
the reagent bottle as full.  
2. Use the ‘Remote Fill / Drain’ features of the Leica ASP300  
Connect the supplied remote fill/drain hose to  
the remote drain connection (refer Fig. 11).  
Important!  
Ensure that the hose is pushed all of  
the way onto the connection, until a  
clear click is heard  
Put the loose end of the remote fill/drain hose  
into the desired container for the drain/fill  
operation as desired.  
Fig. 11  
Leica ASP300  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Fill / drain reagents  
In FAVORITES, press MENU. In MENU FUNC-  
TIONS, press SMART SCREEN.  
The SMART SCREEN (Fig. 12) pops up.  
Fig. 12  
Select station  
Press key symbol  
Remote drain  
On the screen, select reagent bottle no. 3  
and the bulk container (Remote).  
The press STATION DRAIN.  
Check for leaks.  
Remote fill  
On the screen, select reagent bottle no. 3  
and the bulk container (Remote).  
The press STATION REFILL.  
The remote fill procedure takes less than 3  
minutes.  
Check for leaks.  
Instruction Manual V 1.3 – 12/2003  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.1.2 Changing the wax  
1
2
Fig. 13  
Draining the wax  
Connect the paraffin drain hose (2) as shown  
in Fig. 13 and insert the loose end of the hose  
into a remote collecting container.  
Make sure the hose end is securely kept in  
the remote container until both the drain has  
been completed and the hose has been  
purged with pressurized air.  
From the FAVORITES screen, press SMART  
SCREEN (see previous chapter).  
Select the wax bath to be drained and the  
remote container (‘Remote’).  
Then press STATION DRAIN.  
+
Several queries will be displayed each of  
which has to be confirmed individually.  
Leica ASP300  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Filling with wax pellets  
Remove the splash protector.  
Fill the wax bath (Fig. 15) with wax pellets up  
to the top.  
Press BACK and REAGENT STATUS to ac-  
cess the reagent status table (Fig. 14). High-  
light the wax bath you just filled and press  
SET AS FULL.  
After about 90 minutes, add another 0.5 kg  
of wax pellets.  
The pellets will take a total of approx. 10  
hours to melt.  
Important!  
During the melting process, the paraf-  
fin level MAY NOT drop below the  
MIN level (2, Fig. 15).  
Once the paraffin is completely molten, re-  
insert the splash protector.  
Fig. 14  
Software control  
1
The software calculates when the wax will  
be ready for use and selects program start-  
ing times accordingly, i.e. the paraffin will  
definitely be molten when it is required for  
the first paraffin step.  
2
Fig. 15  
Instruction Manual V 1.3 – 12/2003  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
                                             
5. Operation  
Adding molten paraffin  
Molten wax may have a maximum temperature of 70°C.  
If filling with molten wax, do not fill above the MAX level (see Fig. 15).  
After a wax bath has been filled, it must be marked as ‘full’ in the  
REAGENT STATUS screen.  
Software control  
If the wax baths are filled with molten wax, it is possible to override  
the Leica ASP300 melting time calculation.  
For that purpose, in supervisor mode, access the  
SERVICE FUNCTIONS screen (Fig. 16).  
Press TIME UNTIL WAX MELTED.  
Enter the required delay time and press OK  
to confirm.  
Fig. 16  
If the instrument attempts a paraffin step with the paraffin only partly molten, severe instru-  
ment malfunctions may occur.  
Leica ASP300  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.2 Setting up the instrument parameters  
From the start screen, press MENU to go to the MENU FUNCTIONS  
screen, then press MORE.  
In the MORE MENU FUNCTIONS screen, press SYSTEM SETUP.  
5.2.1 System setup  
SYSTEM SETUP screen  
(Fig. 17).  
Fig. 17  
The DISPLAY/BUZZER field is practically self-explanatory:  
The number next to DISPLAY OFF indicates the remaining time (in min)  
until the screen saver (after the last user action) is activated.  
LCD CONTRAST can be set from 0 to 14.  
The BUZZER VOLUME can be set between 1 and 10.  
Instruction manual V 1.3 – 12/2003  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Program Options  
Selection of the options on how programs are carried out.  
STATION ALLOCATION:  
By age or Sequential  
By age  
-
RMS is activated, reagents will automatically be  
used in order of increasing cleanliness.  
Sequential - reagents will be used in sequential order of stations.  
WAX BATH ORDER: Auto or 1.; 2. ; 3.  
Wax bath order can be activated only when STATION ALLOCATION  
is set to Sequential.  
PROMPT FOR NUM BLOCKS :  
Enabled or Disabled  
Enabled-  
When starting a program, the actual number of  
blocks must be entered.  
This is recommended with the RMS being enabled.  
Number of blocks can be entered optionally (not  
mandatory).  
Disabled -  
Date / Time  
Be sure to verify that the date and time entered do actually correspond  
to the local time/date.  
Press the buttons to access the entry windows where time/date can be  
corrected, if necessary.  
Instrument  
Press WAX BATH SET TEMP (°C) to access the wax bath temperature  
entry window. Select the highest temperature compatible with the type  
of paraffin used as paraffin temperature will drop slightly once the paraf-  
fin is in the retort.  
Leica ASP300  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Security  
To determine the level of access to the instrument.  
SUPERVISOR PASSWORD:  
Enabled or Disabled  
Enabled -  
A password is required for access to the supervisor  
level.  
Disabled -  
All instrument and software functions are fully  
accessible to all staff.  
LOCK MODE: Enabled or disabled.  
Enabled -  
No screen functions can be accessed.  
To unlock, a password has to be entered.  
Disabled-  
Screen can be accessed at any time until the screen  
saver turns off the screen.  
Entry keyboard  
The keyboard is displayed any time text needs to be entered. The key-  
board headline (1) tells you what kind of text to enter. 30 characters can  
be entered into each entry field, though sometimes not all characters  
entered can be displayed.  
Important keys  
1
Shift:  
To shift to upper case symbols.  
AltGr :  
For access to a number of  
special characters.  
<-- :  
Deletes last character entered.  
Clear:  
Deletes entire line.  
Fig. 18  
Instruction manual V 1.3 – 12/2003  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.2.2 Editing the reagent list  
Adding new reagents  
1. Enter the reagent name and allocate the new reagent to the reagent  
group it belongs to.  
2. Define the stations (bottles) that will be filled with the new reagent.  
3. Fill the stations (bottles) according to the reagent list.  
Enter reagent names  
Enter reagent names via the REAGENTS screen.  
You must be logged on at supervisor access level to proceed.  
From the start screen, press MENU to go to the MENU FUNCTIONS  
screen, then press MORE to go to MORE MENU FUNCTIONS.  
Press REAGENTS.  
The SET UP REAGENTS AND WARNING THRESHOLDS screen opens.  
To add a reagent:  
Press INSERT.  
Enter the new reagent name.  
Press OK to confirm.  
You will then automatically be prompted to select the reagent  
group:  
Select the reagent group  
Allocate the new reagent to the desired  
group and press OK to confirm.  
Fig. 19  
Leica ASP300  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Allocation of reagents to the correct reagent group is the basis for  
compatibility monitoring. Allocation to the wrong reagent group  
can lead to reagent cross contamination.  
Changeable parameters  
Entering / modifying reagent thresholds  
Highlight the reagent to be modified.  
Press the headline key symbol of the parameter to be changed.  
Enter a new threshold or - if no warning is desired - press CLEAR to  
remove the threshold altogether.  
Press OK to confirm.  
The same threshold value applies for all reagent stations con-  
taining the same reagent.  
Changing reagent names or reagent groups  
Highlight the reagent.  
Press the corresponding headline key symbol.  
In the entry window (or via the keyboard), enter the new reagent  
group allocation.  
Press OK to confirm and save.  
Reedit all station and programs allocated to the reagent changed.  
Instruction manual V 1.3 – 12/2003  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Deleting reagents  
Highlight the reagent to be deleted in the SET UP REAGENTS AND  
WARNING THRESHOLDS screen.  
Press DELETE.  
Press OK in the screen prompt.  
Please remember that a reagent which is already used in a pro-  
gram cannot be deleted.  
Adding new reagents to stations  
Go to SMART FUNCTIONS to remote-fill the reagent bottle.  
If a bottle is filled manually, it must be defined as full afterwards.  
For that purpose:  
Press MENU to access the MENU FUNCTIONS screen.  
Press REAGENT STATUS.  
In the reagent list, highlight the station / reagent that has been filled  
manually.  
Press SET AS FULL to mark the station as ‘full’.  
Upon activation of the RMS all warning thresholds for the se-  
lected reagent are automatically reset to ‘0’.  
Leica ASP300  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.2.3 Viewing the program list  
This list (Fig. 20) displays all programs currently existing in the  
ASP300.  
You can create:  
up to 15 processing programs,  
3 retort cleaning programs,  
1 wax cleaning program.  
At supervisor access level:  
the processing program names  
can be edited,  
new processing programs can  
be created and existing ones  
can be deleted.  
Fig. 20  
Important!  
New processing programs are created by copying existing  
ones. Therefore, the list must contain at least one program at all  
times.  
Program duration cannot be edited directly.  
To alter the duration of a program, the duration of one or more indi-  
vidual program steps has to be modified.  
Retort and wax cleaning programs are predefined and cannot  
be altered.  
Instruction manual V 1.3 – 12/2003  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.2.4 Adding and/or modifying programs  
Creating a new program  
You must be logged on at supervisor level.  
In the VIEW/EDIT PROGRAMS screen (Fig. 20) highlight a program as  
similar as possible to the program you wish to create.  
Press COPY to copy the selected program.  
The new program will have the same name as the program copied,  
however, the figure “(2)“ will be added.  
Highlight the line containing the new program.  
Press PROGRAM NAME.  
Enter the new program name.  
Editing program steps  
Press EDIT (refer Fig. 20) to access the PROGRAM STEPS screen.  
In the headline you will find the PROGRAM NAME.  
The colors on the left border of the table indicate the reagent groups  
to which the reagent of the step belongs.  
The program steps (max. 13 steps per program) are displayed in the  
order in which they are carried out.  
The following characteristics can be edited:  
reagent name  
duration of step  
(with the exception of fill and drain steps)  
retort temperature  
(if “Ambient“ is selected, the retort temperature indication remains  
empty)  
type of pressure and/or vacuum cycle  
retort drain time  
delay step.  
Leica ASP300  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Editing program steps  
To edit a step, highlight the corresponding line and press the re-  
spective headline.  
In the entry windows that pop up enter / select the program step set-  
tings.  
Fig. 21  
Instruction manual V 1.3 – 12/2003  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Editing program steps (continued)  
Setting a delay step  
A so-called ‘delay step’ ensures that a program is finished at a specified  
time.  
Highlight the program step you wish to define as a delay step.  
Press DELAY.  
The delay symbol is moved to the selected step, thus defining the  
step as delay step.  
Copying steps  
Highlight the step to be copied.  
Press COPY.  
Use the MOVE UP / MOVE DOWN symbols to move program steps up  
or down within an existing program without having to recreate those  
steps.  
DELETING PROGRAM STEPS  
Highlight the step to be deleted.  
Press DELETE.  
Leica ASP300  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.2.5 Favorites (maximum of 5 programs)  
Favorites can be programmed to end either:  
“as soon as possible“ (ASAP) or  
at a predetermined finish time.  
Fig. 22  
Adding and/or modifying Favorites  
Highlight the program line to be altered.  
Press PROGRAMS.  
Select the desired program and press OK to confirm.  
Setting a finish time  
Press FINISH TIME.  
Enter the finish time (24-hour clock). Select ASAP, if the program finish  
time is not to be delayed.  
Press OK to confirm.  
Instruction manual V 1.3 – 12/2003  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
                          
5. Operation  
Assigning a finish day  
Press DAY.  
Select the desired week day in the list and press OK to confirm.  
Assigning an icon to a “Favorite“  
Press SYMBOL in the table header.  
Select the desired icon and press OK to confirm.  
The first icon on the list is empty - select this option, if you do  
not want to assign an icon to a specific program.  
Deleting a ‘Favorite’  
Highlight the Favorite to be deleted.  
Press DELETE.  
The ‘Favorite“ is deleted from the FAVORITES list only - the  
program itself is not deleted.  
Leica ASP300  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.2.6 Editing the stations  
Fig. 23 shows all stations / allocated reagents cur-  
rently defined in the instrument.  
Fig. 23  
Adding / changing reagents  
Allocating reagents  
Highlight the desired station.  
Only processing reagents (except paraffin)  
can be allocated to stations 1 – 10.  
Press REAGENT NAME.  
Only a cleaning reagent can be allocated to  
station 11.  
The SELECT REAGENT entry window pops up.  
Only those reagents that are compat-  
ible with the respective station will be  
available for selection.  
Only a cleaning alcohol can be allocated to  
station 12.  
Only cleaning water/detergent can be allo-  
cated to stations 13 and 13-ext.  
Highlight the desired reagent and press OK  
to confirm.  
Only paraffin wax can be allocated to the  
wax baths.  
Instruction Manual V 1.3 – 12/2003  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.2.7 Reagent groups  
In this screen (Fig. 24) you find the colors used for each  
reagent group in the graphic program display.  
Fig. 24  
Selecting a new color for a reagent group  
Highlight the line containing the reagent color you wish to change.  
Press COLOR in the table header: the entry window SELECT COLOR  
pops up.  
Select the desired color and press OK to confirm.  
If you want to close the entry field without changing the color, press  
CANCEL.  
Leica ASP300  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.3 Programs  
Programs can be started from two different screens:  
ALL PROGRAMS  
FAVORITES (max. 5 programs)  
Fig. 25  
Fig. 26  
A ”Favorite” is a processing program that is  
frequently used.  
In the Favorites screen, press  
ALL PROGRAMS.  
All settings are already programmed. Only the  
number of blocks has to be entered (if this  
function is enabled).  
Press the symbol corresponding to the pro-  
gram you wish to start.  
A graphic representation of the stations /  
retort will be displayed with the stations re-  
quired for the program you are about to start  
colored in the corresponding reagent group  
colors (Fig. 27).  
Starting a “Favorite“  
To start a Favorite, press the corresponding  
symbol on the Favorites screen - the program  
will be started immediately.  
Prior to starting it, modify the program as  
required.  
All processing step modifications made will  
only be applied to the current program.  
Instruction Manual V 1.3 – 12/2003  
48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.3.1 Editing a selected program  
Redefine the program finish time:  
Press FINISH TIME. In the entry window, set  
a new finish time.  
Fig. 27  
To modify a program prior to starting it, press  
EDIT to access the Edit Program screen.  
Fig. 28  
Enter the day when you want the program  
finished.  
Enter the finish time (24-hour clock format).  
Press OK to confirm.  
Fig. 29  
Leica ASP300  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Access the RUN OPTIONS function to display  
the options selected for the current program.  
LEVEL CHECK enabled:  
The lower level sensor in the retort is acti-  
vated.  
RECIRCULATION enabled:  
During each processing step, the reagent  
is pumped periodically from the bottle to the  
retort and back for continuous and efficient  
mixing of fluid. The first pump step is carried  
out after 16 minutes, all subsequent pump  
steps in 12-minute intervals.  
PRETEST enabled:  
After starting the program a test run is per-  
formed: the first reagent is pumped into the  
retort and back into the bottle.  
STIRRER enabled:  
Magnetic stirrer is on.  
PRINTER enabled:  
At the end of the program run, a full run log  
will be printed out.  
Fig. 30  
Important!  
The options selected will be applied to ALL programs!  
All run options can be enabled / disabled during the program run if necessary.  
The printer option will be displayed only if a printer has been configured for the instrument.  
Instruction Manual V 1.3 – 12/2003  
50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.3.2 Starting a program  
Fig. 31  
Press START to start the program. Press OK to confirm the query that  
follows. Enter the number of blocks, if the NUMBER OF BLOCKS op-  
tion has been enabled.  
The FINISH TIME button shows the time when the current program  
will be finished.  
Before a program in progress can be modified, the PAUSE symbol  
must be pressed.  
To modify or delete program steps, press EDIT and, in the entry win-  
dow, carry out any changes you wish to make. Press CONTINUE to re-  
start the program.  
The finish time of a program in progress can only be changed as  
long as the delay step has not yet been completed.  
Press TABLE to change from graphic to tabular program display,  
press GRAPHICAL to go back to graphic display.  
Leica ASP300  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.4 Reagent status  
In the reagent status screen (Fig. 32) you can display and update the  
actual reagent status of each station and you can set the stations  
as full or empty.  
Warning thresholds that trigger  
warning messages appear on a  
red background.  
To change a reagent status,  
highlight the corresponding line  
and press the corresponding  
key symbol .  
Fig. 32  
Reagent warnings  
Criteria for warning messages  
If the RMS is enabled, the instrument will give  
you warning messages when reagents have  
been used for too long.  
For normal reagents, the following warning cri-  
teria apply:  
number of blocks processed  
number of programs carried out  
number of day lapsed  
Warning messages for normal reagents are  
given based on up to 3 criteria; warning mes-  
sages for paraffin wax on up to 6 criteria.  
since the reagent has last been changed.  
The warning messages are displayed at the  
end of a cleaning cycle. From there, you can  
go directly to the REAGENT STATUS screen.  
For paraffin wax, three additional criteria can  
be applied:  
number of blocks processed  
number of programs carried out  
number of days lapsed  
In addition, warning messages (on a yellow  
background) are displayed in the program  
start screens.  
since the wax has last been cleaned.  
Instruction Manual V 1.3 – 12/2003  
52  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.5 System Monitor  
The SYSTEM MONITOR screen supplies important infor-  
mation on the Leica ASP300 system status.  
PUMP AGE:  
Indication of hours of service of the pressure/  
vacuum pump. After approx. 1,000 hours of  
use, have Leica Technical Service carry out a  
preventive maintenance of the pump.  
CARBON FILTER AGE:  
Indication of hours of service of the activated  
carbon filter. The carbon filter must be ex-  
changed after approx. 4 months of use.  
Fig. 33  
After replacing the filter press RESET.  
ROTARY VALVE RETRIES:  
If a figure higher than 50 is indicated, a preventive maintenance is  
recommended.  
FILL/DRAIN:  
This screen displays the average fill / drain times of all stations  
(in seconds).  
Short term = average duration of last 5 fills / drains.  
Long term = average duration of last 20 fills / drains.  
If one of the figures displayed exceeds 250, call Leica Technical  
Service.  
Leica ASP300  
53  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.6 On-line help  
The Leica ASP300 features an extensive on-line Help system.  
For information on basic problems refer to the Help Index (to access  
press INDEX on the Help screen).  
Use the UP/ DOWN arrows to  
scroll.  
Press PREVIOUS TOPIC (1) to  
return to the index.  
Press BACK (2) to go back to the screen  
from where you accessed Help.  
CUSTOM HELP (3) is only displayed, if user-  
specific help files have been installed.  
2
1
3
Fig. 34  
Detailed information on how to create and install customer-specific help  
files can be found in the “What is Custom Help?“ help text.  
Accessing ‘Help’ in the event of error messages  
Press DETAILS, or  
Press the HELP button on the error message.  
Instruction Manual V 1.3 – 12/2003  
54  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.7 Configuring a printer (optional)  
To select and configure a printer, proceed as follows:  
With a Centronics standard cable, connect the printer to  
the printer port (refer chapter 4.4, Fig. 6).  
Printing run logs  
Printing is possible when the PRINT symbol is displayed.  
For the PRINT symbol to be displayed, a printer must have been  
configured for the instrument.  
The following lists and protocols can be printed:  
reagent list,  
stations list,  
reagent status list,  
all programs,  
run log,  
error log.  
Leica ASP300  
55  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Trouble shooting  
6.1 General  
The Leica ASP300 is equipped with a trouble shooting system.  
If an error is detected, detailed error messages and corresponding user  
instructions on what to do are displayed on the touch screen.  
For all correctable error conditions the instrument will take  
corrective actions to protect the specimens.  
Details of all events are recorded in “Run Log”.  
Run log can be viewed by pressing MENU and then RUN LOG.  
6.2 Power failure  
Power failure during a processing program  
In the event of a power failure, the instrument automatically compen-  
sates any impact the power failure might have on the specimens.  
Power failure during a paraffin or retort clean cycle  
If a power failure occurs during a clean cycle, the total duration of  
the cycle will be extended by the duration of the power failure.  
Instruction manual V 1.3 – 12/2003  
56  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Trouble shooting  
6.3 Trouble shooting  
Check for error messages indicating the nature of the problem.  
Press the Help button on the message for any information relating to  
the cause.  
Check the Run Log for any information on the nature of the failure.  
Use the SMART SCREEN functions to individually test opera-  
tions such as fill, drain, pressurizing and evacuation operations.  
Power failure  
Verify that the mains lead is plugged in to the wall socket, and that, if  
applicable, the wall socket is switched ON.  
Verify that the jumper lead or UPS lead are fully inserted into their re-  
spective socket.  
Verify that the power switches on the rear (adjacent to the power  
lead) and on the side of the instrument are both switched ON.  
If, by carrying out the above listed steps, you cannot identify the  
problem, call Leica Technical Service.  
Leica ASP300  
57  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6. Trouble shooting  
6.4 Typical fill or drain problems  
Fill and drain problems may be due to several causes:  
1. There is insufficient reagent  
Verify that both the reagent bottles and the wax stations have  
sufficient reagent.  
2. Insufficient pressure or vacuum  
The Leica ASP300 is unable to create sufficient vacuum (for a fill) or  
pressure (for a drain) in the retort.  
Verify that the reagent bottle in question is correctly inserted in  
the reagent cabinet.  
Verify that all reagent bottles are fully pushed home into their  
sockets.  
Check whether there is some debris under the retort lid seal  
causing a leak.  
3. There is a blockage in the air or reagent lines  
If there is reagent in the retort that cannot be drained, try to remove  
the blockage as follows:  
Heat the retort to the maximum possible temperature (consistent  
with the reagent currently in the retort, if any) and leave it at this  
temperature for at least 15 minutes.  
Whilst the retort is at maximum temperature, attempt to fill or  
drain.  
If the retort is empty:  
Run the extended retort clean program. Use the maximum set-  
tings for number of clean cycles and temperature (consistent  
with the type of reagents/solvents used in the clean program).  
If you cannot eliminate the problem by carrying out the above  
listed steps, call Leica Technical Service  
Instruction manual V 1.3 – 12/2003  
58  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Cleaning and maintenance  
7.1 Clean programs  
You find the clean programs at the end of the programs  
list (Fig. 35). To display a clean program, highlight the re-  
spective line and press EDIT.  
The ASP300 offers three retort clean programs and one paraffin  
clean program. Clean programs cannot be copied or deleted.  
7.1.1 Retort clean programs  
A clean program comprises a  
minimum of three and a maxi-  
mum of four steps:  
1. Cleaning Solvent  
1 step from station 11  
2. Cleaning Alcohol  
1 step from station 12  
3. Cleaning Water  
1 step from station 13  
Optional step between  
steps 2 and 3:  
Cleaning Water / Detergent  
1 step from station 13-ext.  
Fig. 35  
The steps of retort clean programs are listed in the order in which  
they are carried out.  
The clean programs skip any steps not required to complete the re-  
tort clean.  
Leica ASP300  
59  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Cleaning and maintenance  
Editing steps  
To edit a program step highlight the corresponding line and press  
the respective table header.  
To disable a step, set the number of cycles to “0“. Only cleaning wa-  
ter steps can be disabled.  
The following characteristics of each step can be edited:  
Reagent name - only those reagents that are compatible with the re-  
agent station being edited will be displayed.  
Number of cycles - the duration displayed is calculated based on  
the number of cycles.  
Retort temperature - temperatures from 50 to 65 °C can be set.  
Station 13-ext  
The term “13-ext“ stands for an additional station which can be used  
instead of station 13.  
Station 13 always contains cleaning water.  
If a program contains a “13-ext“ step, during the clean cycle the op-  
erator is prompted to insert bottle “13-ext“ when needed, and to re-  
move the bottle as soon as the step has been completed.  
Only reagents belonging to the cleaning water group can be al-  
located to station 13-ext.  
Station 13-ext may be used for clean programs only.  
Instruction manual V 1.3 – 12/2003  
60  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Cleaning and maintenance  
7.1.2 Wax clean programs  
The Leica ASP300 contains a wax clean fea-  
ture that removes solvents (xylene) from used  
wax.  
Follow the instructions on the left (Fig. 36) to run  
a wax clean program.  
Fig. 36  
The duration of the wax clean program is  
user-programmable.  
Follow the instructions in Fig. 37.  
Fig. 37  
Leica ASP300  
61  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Cleaning and maintenance  
7.1.3 Smart clean  
The Smart Clean prepares the ASP300 for transport.  
Prior to starting a Smart Clean, a complete retort clean program must be  
run.  
To start a Smart Clean, press SMART CLEAN on the MORE MENU FUNC-  
TIONS screen and press OK to confirm.  
The Smart Clean takes approx. 15 minutes and consists of 3 phases:  
Phase 1:  
Make sure all reagent containers are inserted correctly.  
All internal tubing is cleaned with pressurized air.  
Phase 2:  
Before phase 2 starts, the operator is prompted to remove all reagent  
bottles from the instrument.  
All internal tubing is again cleaned with pressurized air.  
Phase 3  
Before phase 3 starts, the operator is prompted to insert an empty bottle  
into station 11.  
The tubing is cleaned with pressurized air for the third time - remaining  
rests of reagents, if any, are forced back into station 11.  
Empty and clean all reagent bottles thoroughly before reinsert-  
ing them into the instrument to make sure there will be no re-  
agent spills from the bottles that could cause damage during  
transport.  
The SMART clean is ideal for removing residual liquid from the  
internal pipes of the instrument, as it thoroughly purges them  
with pressurized air. A SMART clean should be carried out  
once every month.  
Instruction manual V 1.3 – 12/2003  
62  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Cleaning and maintenance  
7.2 General cleaning steps  
Paraffin wax baths  
Warning! Be careful as the walls of the wax baths are very hot  
and may cause burns!  
Wipe the wax baths and lids clean.  
The lid can be removed for cleaning purposes.  
If they are dirty, remove the wax strainers from the wax baths. Clean,  
dry and reinsert them.  
Ensure that the air vent hole at the top rear right hand corner is un-  
obstructed.  
Instrument exterior  
Clean the instrument exterior as necessary. Wipe with a damp cloth  
moistened with mild detergent and dry.  
Do not use solvents on painted surfaces and/or the touch  
screen!  
Checking the drip tray  
Regularly check the drip tray for spilled  
reagents (refer Fig. 38).  
Fig. 38  
Leica ASP300  
63  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Cleaning and maintenance  
Draining the condensate bottle  
Remove and empty the condensate bottle  
and reinsert (Fig. 39).  
Fig.39  
Exchanging the activated carbon filter  
Exchange the filter as shown in Fig. 40.  
Replace the filter at least every four months  
– make sure to insert correctly –  
(see arrow on label)!  
Perform a Smart Clean prior to changing the  
filter.  
Fig. 40  
Filterlabel  
Instruction Manual V 1.3 – 12/2003  
64  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Cleaning and maintenance  
7.3.1 Daily cleaning and maintenance  
Cleaning the retort lid  
Remove wax from the inside of the retort lid with the plastic scraper  
provided. Make sure to remove all wax from around the retort lid  
seal.  
For convenience, the lid may be removed during cleaning by lifting it  
vertically and sliding it towards the console.  
Cleaning the retort  
The retort may be wiped clean using a cloth moistened with solvent  
(xylene or alcohol) or mild detergent. Ensure that the air holes at the  
top front of the retort are open - remove all dirt!  
Cleaning the retort strainer  
Use alcohol or xylene to clean the strainer located at the bottom of  
the retort. For convenience, the strainer may be taken out to remove  
all solid dirt.  
Cleaning the touch screen  
On the MENU FUNCTIONS screen press CLEAN TOUCH SCREEN.  
Clean the touchscreen.  
Press ENABLE EXIT and then BACK to return to MENU FUNCTIONS.  
Cleaning the processor module top surface  
Remove both lids for cleaning.  
First scrape the stainless steel surface with the plastic scraper and  
then wipe down.  
Reagent and condensate bottle seal lubrication  
To ensure easy removal of the reagent and condensate bottles, lu-  
bricate the O-ring seals on the plug-in nozzles with the O-ring lubri-  
cant supplied.  
Check regularly whether the reagent bottles are still properly con-  
nected to the connection manifolds in the instrument. Retighten, if  
necessary.  
Leica ASP300  
65  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Cleaning and maintenance  
This procedure is particularly important with bottles containing  
chloroform. Chloroform tends to swell the bottle seals, making  
them difficult to remove. This problem can be prevented by  
regular lubrication.  
Checking the retort and wax bath lid seals  
Regularly check the retort and wax bath lid seals for damage. If a  
seal is damaged, it must replaced without delay.  
7.3.2 Periodic cleaning and maintenance  
Cleaning the reagent bottles  
Remove the reagent bottles from the reagent cabinet.  
Empty and clean them using a bottle brush and a laboratory deter-  
gent in warm water.  
Never clean the reagent bottles in an automatic dishwasher.  
Wipe the stainless steel inner walls of the reagent cabinet with a  
damp cloth moistened with a mild detergent.  
Refill and reconnect the bottles once cleaned.  
Make sure the bottle lids are tight.  
Make sure the bottles are properly pushed home into the connection  
manifolds at the rear inner wall of the reagent module.  
Instruction manual V 1.3 – 12/2003  
66  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Cleaning and maintenance  
7.4 Preventative maintenance schedule  
Maintenance activity  
daily  
weekly  
monthly  
every  
4 months  
Lubricate reagent bottle O-rings  
and check for damage.  
√ ∗  
Wipe dry retort and lid  
after completing retort clean cycle.  
Clean exterior of instrument  
with soft cloth and minimal  
amount of xylene.  
Check retort strainer  
for any tissue or wax debris.  
Remove residual paraffin  
from inner surface of wax bath lid.  
Inspect and, if necessary,  
clean retort lid seal.  
Inspect and, if necessary,  
clean wax bath lid seal.  
Inspect and, if necessary,  
adjust paraffin levels.  
Inspect reagent container fluid levels.  
* if the automatic fill and drain function is actually being used  
Leica ASP300  
67  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Cleaning and maintenance  
Maintenance activity  
daily  
weekly  
monthly  
every  
4 months  
Inspect and, if necessary,  
clean wax bath strainers .  
Inspect wax bath air vent and,  
if necessary, clear.  
Inspect and empty condensate bottle.  
Clear inlet holes.  
Examine activated carbon filter.  
Replace carbon filter.  
Remove fluid level sensor (float),  
inspect and clean.  
Inspect electrical connectors  
at the rear panel of the instrument.  
Inspect and, if necessary,  
clean inside reagent bottles.  
Check that the bottle caps and seals  
(O-rings) of the reagent bottles are in  
place and tight.  
Check that the reagent bottles are fully  
inserted in the connection manifolds.  
Instruction manual V 1.3 – 12/2003  
68  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8. Warranty and service  
Warranty  
Leica Microsystems Nussloch GmbH guarantees that the contractual  
product delivered has been subjected to a comprehensive quality con-  
trol procedure based on the Leica in-house testing standards, and that  
the product is faultless and complies with all technical specifications  
and/or agreed characteristics warranted.  
The scope of the warranty is based on the content of the concluded  
agreement. The warranty terms of your Leica sales organization or the  
organization from which you have purchased the contractual product  
shall apply exclusively.  
Technical service information  
If you require technical service or replacement parts, please contact  
your Leica sales representative or dealer who sold the product.  
Please provide the following information:  
Model name and serial number of the instrument.  
Location of the instrument and name of the person to contact.  
Reason for the service call.  
Date of delivery.  
Decommissioning and disposal  
The instrument or parts of the instrument must be disposed of in compli-  
ance with the local laws.  
Leica ASP300  
69  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Notice  
Instruction manual V 1.3 – 12/2003  
70  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Leica Marine RADAR MX535 User Manual
Liebherr Beverage Dispenser WS 40 User Manual
Life Fitness Stepper Machine 90SW User Manual
Life Fitness Treadmill M051 00K26 A261 User Manual
Lochinvar Water Heater SSB i s 01 User Manual
Lynx Indoor Furnishings L LS User Manual
McCulloch Trimmer 115376827 User Manual
Memorex CRT Television MT2024 User Manual
Memorex DVD Player MVD 2037 O MQXD User Manual
Metra Electronics Power Screwdriver 95 2004 User Manual