Safety regulations  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
4
 
Safety concept  
The individual modules of the Leica DM micros-  
copy series include an interactive CD-ROM with  
all relevant user manuals in several languages.  
Keep it in a safe place, and readily accessible  
to the user. User manuals and updates are also  
available for you to download and print from  
our web site at  
You can combine individual system articles with  
articles from external suppliers. Please read the  
user manual and the safety requirements of the  
supplier.  
Before installing, operating or using the instru-  
ments, read the user manuals listed above. In  
particular, please observe all safety instruc-  
tions.  
This operating manual describes the special  
functions of the individual modules of the Leica  
DM microscopy series and contains important  
instructions for their operational safety, main-  
tenance, and accessories.  
To maintain the unit in its original condition  
and to ensure safe operation, the user must  
follow the instructions and warnings contained  
in these user manuals.  
The "Safety concept" booklet contains addi-  
tional safety information regarding the service  
work, requirements and the handling of micro-  
scope, accessories and electrical accessories as  
well as general safety instructions.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
5
 
Symbols used in this operating manual  
Warning of a danger  
This symbol indicates especially impor-  
tant information that must be read and  
Danger due to hot surface.  
This symbol warns against touching  
accessible hot surfaces, e.g. those of  
light bulbs.  
complied with.  
Failure to comply can cause the following:  
Ϙ
Hazards to personnel  
Important information  
This symbol indicates additional infor-  
mation or explanations that intend to  
provide clarity.  
Ϙ
Functional disturbances or damaged instru-  
ments  
Warning of hazardous electrical voltage  
This symbol indicates especially impor-  
tant information that is mandatory to  
read and observe.  
Explanatory notes  
This symbol within the text points to addi-  
tional information and explanations.  
Failure to comply can cause the following:  
Ϙ
Hazards to personnel  
Ϙ
Functional disturbances or damaged  
instruments  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
6
 
Important Notes  
Description  
User Manual  
Accessories from third-party suppliers  
You can combine individual system articles with  
articles from external suppliers. Please read the  
user manual and the safety requirements of the  
supplier.  
The Leica DM750 P microscope meets today's  
state of the art of technology. Nevertheless,  
hazards may still arise during operation. The  
potential risks are described below.  
This user manual includes important instruc-  
tions related to operating safety, maintenance  
and accessories,  
Your Leica DM750 P microscope comes  
with an interactive CD-ROM with all  
relevant user manuals. Keep it in a safe place,  
and readily accessible to the user. User manu-  
als and updates are also available for you  
to download and print from our web site at  
Before installing, operating or using the  
instruments, it is required to read this  
user manual. In particular, please observe all  
safety instructions.  
Original condition  
To maintain the unit in its original condition  
and to ensure safe operation, the user must  
follow the instructions and warnings contained  
in these user manuals.  
Legal requirements  
Contact address  
Adhere to general and local regulations relat-  
ing to accident prevention and environmental  
protection.  
Leica Microsystems (Switzerland) Ltd.  
Industry Division  
Max Schmidheiny Strasse 201  
CH-9435 Heerbrugg (Switzerland)  
EC Declaration of Conformity  
Electricallyoperatedaccessoriesareconstructed  
based on the state of the art of technology  
and are provided with an EC Declaration of  
Conformity.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
7
 
Instructions on use  
Intended use  
Place of use  
Non-intended use  
Using the instrument in any way  
contrary to the specifications in the user  
manual can lead to bodily harm and damage to  
objects. Doing so can impair the functions of  
the protective measures provided.  
The Leica DM750 P microscope is an optical  
instrument which uses magnification and illu-  
mination to improve the visibility of objects. It  
is used for observing and documenting.  
Only use the instruments in closed, dust free  
rooms and between +10 °C and +40 °C. Protect  
the devices from oil, chemicals and extreme  
humidity. If using the devices outdoors, protect  
them from dust and moisture. Never use electri-  
cal devices outdoors. Install electrical devices at  
least 10 cm from the wall and away from flam-  
mable substances.  
The Leica DM750 P microscope may be used  
only in closed rooms and must be placed on a  
solid floor.  
Never install any other plug (NEMA  
5-15P) or unscrew any mechanical  
components unless expressly instructed to do  
so in the instructions.  
The Leica DM750 P microscope can be used in  
clean rooms without any problems.  
Avoid large temperature fluctuations, direct  
sunlight and vibrations. These conditions  
can distort measurements and micrographic  
images.  
The devices and accessories described in  
this operating manual have been tested  
for safety and potential hazards.  
Position the Leica DM750 P microscope  
so that it can be disconnected from the  
power supply at any time. The power cord must  
remain accessible at all times.  
In warm and warm-damp climatic zones,  
the individual components require  
special care in order to prevent the build-up of  
fungus.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
8
 
Instructions on use (continued)  
The responsible Leica affiliate must be  
consulted whenever the instrument is  
altered, modified or used in conjunction with  
non-Leica components that are outside of the  
scope of this manual.  
Transport  
Integration in third-party products  
When installing Leica products into third-party  
products, the manufacturer of the complete  
system or its dealer is responsible for follow-  
ing all applicable safety instructions, laws and  
guidelines.  
If at all possible, use the original packaging for  
shipping or transporting individual modules.  
In order to prevent damage from vibrations,  
disassemble all moving parts that (according to  
the user manual) can be assembled and disas-  
sembled by the customer and pack them sepa-  
rately.  
Unauthorized alterations to the device  
or noncompliant use shall void all rights  
to any warranty claims.  
Disposal  
Disposal must comply with locally applicable  
laws and regulations.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
9
Health risks and dangers of use  
Health risks  
The ergonomic design and construction of the  
Leica microscopy series are intended to reduce  
the exertion of the user to a minimum.  
Dangers during use  
Workplaces with microscopes facilitate  
and improve the viewing task, but they  
also impose high demands on the eyes and  
holding muscles of the user. Depending on the  
duration of uninterrupted work, asthenopia  
and musculoskeletal problems may occur. For  
this reason, appropriate measures for reduction  
of the workload must be taken:  
Ϙ
The Leica DM750 P microscope may only be  
connected to a grounded socket.  
Danger of infection  
Ϙ
The Leica DM750 P microscope may not be  
operated unless it is in proper functioning  
condition.  
Direct contact with eyepieces is a poten-  
tial transmission method for bacterial  
and viral infections of the eye.  
The risk can be kept to a minimum by  
using personal eyepieces for each indi-  
vidual or detachable eyecups.  
Ϙ
Ϙ
Ϙ
Optimum workplace layout  
Frequent changes of activity  
Thorough training of the personnel, giving  
consideration to ergonomic and organiza-  
tional aspects.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
10  
 
Information for the person responsible for the instrument  
Information for the person responsible for  
the instrument  
Ϙ
Ϙ
Do not use the Leica DM750 P microscope  
unless it is in perfect condition.  
Ϙ
Ϙ
Only original Leica replacement parts may  
be used in servicing the product.  
Ϙ
Ϙ
Ϙ
Ensure that the Leica DM750 P microscope  
is used only by qualified personnel.  
Inform your Leica representative or Leica  
Microsystems (Schweiz) AG, Industry  
Division, 9435 Heerbrugg, Switzerland,  
immediately of any product defect that  
could potentially cause injury or harm.  
After service work or technical modifica-  
tions, the unit must be reconfigured with  
observance to our technical requirements.  
Ensure that this user manual is always avail-  
able at the Leica DM750 P microscope.  
Ϙ
If the unit is modified or serviced by unau-  
thorized persons, is improperly maintained  
(as long as maintenance was not carried  
out by us), or is handled improperly, Leica  
will not accept any liability.  
Carry out regular inspections to make  
certain that the authorized users are adher-  
ing to safety requirements.  
Ϙ
If you use accessories made by third-party  
manufacturers with the Leica DM750 P  
microscope, be sure that each such manu-  
facturer confirms the safety-engineering,  
harmless usability of the product and  
observe the product's user manual.  
Ϙ
Ϙ
When instructing new users, do so thor-  
oughly and explain the meanings of the  
warning signs and messages.  
Ϙ
The electric installation in the building  
must conform to the national standard, e.g.  
current-operated ground leakage protec-  
tion (fault-current protection) is suggested.  
Assign individual responsibilities for start-  
ing, operating and servicing the instru-  
ment and monitor the observance of these  
responsibilities.  
Ϙ
Modifications and maintenance of the  
Leica DM750 P microscope may only be  
performed by professionals expressly  
authorized by Leica.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
11  
 
Care instructions  
General instructions  
Cleaning coated parts and plastic parts  
Cleaning glass surfaces  
Ϙ
Protect the Leica DM750 P microscope  
against damp, vapors, acids, alkalis, and  
corrosive substances. Do not store chemi-  
cals in the vicinity.  
Ϙ
Ϙ
Ϙ
Ϙ
Ϙ
Dust and dirt particles should be removed  
with a soft brush or lint-free cotton cloth.  
Ϙ
Remove dust using a fine, dry and grease-  
free brush made from hair, by blowing with  
a squeeze blower, or by using a vacuum.  
Remove coarse debris with a moistened  
disposable cloth.  
Ϙ
Ϙ
Remove dirt using a clean cloth that has  
been moistened with distilled water.  
Ϙ
Protect the Leica DM750 P microscope from  
oil and grease. Never grease or oil mechani-  
cal parts or sliding surfaces.  
Clinging dirt can be removed using petro-  
leum ether or alcohol  
Remove tough dirt using alcohol, chloro-  
form or petroleum ether.  
Ϙ
Ϙ
Follow the instructions of the disinfectant  
manufacturer.  
Acetone, xylene or nitro-containing thin-  
ners must NOT be used.  
It is advisable to enter a service agreement  
with Leica Service.  
Never use chemicals to clean colored  
surfaces or accessories with rubberized  
parts. This could damage the surfaces,  
and specimens could be contaminated by  
abraded particles.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
12  
 
Accessories, maintenance and repair  
Accessories  
Maintenance  
Repairs and service work  
Only the following accessories may be used  
with the Leica DM750 P microscope:  
Ϙ
The Leica DM750 P microscope is basically  
maintenance-free. To ensure that it always  
operates safely and reliably, we recommend  
that you take the precaution of contacting  
the responsible service organization.  
Ϙ
Ϙ
Ϙ
Only original Leica Microsystems spare  
parts may be used.  
Ϙ
The Leica accessories described in this user  
manual.  
Before opening the instruments, switch off  
the power and unplug the power cable.  
Ϙ
Other accessories, provided that these have  
been expressly approved by Leica as being  
technically safe in this context.  
You can arrange for periodic inspections  
or, if appropriate, conclude a mainte-  
Avoid contact with powered electrical  
circuits, which can lead to injury.  
nance contract with them.  
Service address  
In case of problems, please contact us as  
follows:  
Ϙ
It is advisable to enter a service agreement  
with Leica Service.  
Ϙ
For maintenance and repair, only OEM spare  
parts may be used.  
Leica Microsystems (Switzerland) Ltd.  
Industry Division  
Max Schmidheiny Strasse 201  
CH-9435 Heerbrugg (Switzerland)  
Phone +41 71 726 33 28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
13  
 
Electrical data and ambient conditions  
Fuse Replacement  
Environment  
Unplug the instrument before changing any fuses. The  
Leica DM750 P has two fuses, which are located behind the  
power cord receptacle.  
Temperature for use  
+10 °C … +40 °C  
-20 °C … +52 °C  
+50 °F … +104 °F  
25 mm on 50 mm hard wood  
100 g / 6 ms  
Storage temperature  
Manipulation shock  
Use only fuse type: 5×20 mm, 1 Amp/250V, Quick acting  
Transport shock (unpacked)  
Transport shock (packed)  
Transport vibrations (unpacked)  
(# 13RFAG30003)  
800 mm free fall  
5–200 Hz / 1.5 g  
Electrical data  
Input: 100-240 V, 50/60 Hz, 5 W (3 W LED)  
Atmospheric pressure "in use" and for storage: 500–1060 mBar  
Humidity "in use" and for storage:  
Installation Category II (Overvoltage Category)  
Pollution degree 2  
20–90%  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Leica DM750 P  
Manual  
14  
 
The Leica DM750 P  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
15  
 
Introduction  
Thank you for purchasing the Leica DM750 P  
Polarizing Microscope System from Leica  
Microsystems. The Leica DM750 P is a highly  
adaptable instrument with superior optics and  
precision engineering for educational and basic  
industrial polarized light microscopy applica-  
tions.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
16  
 
Unpacking  
Ϙ
Ϙ
Ϙ
Carefully remove the microscope and any  
components from the packing carton.  
Check the components against the planned  
configuration.  
Optional items such as contrast accessories,  
camera adapters, cameras, and carrying  
cases are not shipped as part of the stan-  
dard equipment. These items are delivered  
in separate packages.  
Ϙ
Please do not discard any of the packing  
materials. They should be used for safely  
storing and transporting the instrument  
should the need arise.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
17  
 
Get Ready!  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
18  
 
Attaching Analyzer Module or Analyzer/Bertrand Lens Module  
Tools used  
Allen key  
2. Align the shape of the module with the  
shape of the microscope stand.  
4. Tighten the set screw to lock the module  
into place.  
Ϙ
1. Slightly unscrew the set screw located on  
top of the stand with the Allen key tool  
provided.  
3. Insert the dovetail of the module into the  
stand. (Analyzer/Bertrand Lens Module  
shown).  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
19  
 
Viewing Tube  
1. Slightly unscrew the set screw located on  
the top of the Analyzer module.  
5. If you wear eyeglasses for microscope view-  
ing, keep the rubber eyeguards folded  
down. If you do not wear eyeglasses, you  
may find it useful to unfold the rubber  
eyeguards in order to block out ambient  
room light.  
Install the eyepiece with the crossline  
reticle into the right eyetube so that  
the raised feature on the eyepiece aligns and  
is seated with the orientation slot in the right  
eyetube.  
2. Align the shape of the polarizing viewing  
tube with the shape of the module.  
3. Insert the dovetail of the polarizing view-  
ing tube into the module and tighten the  
set screw firmly without forcing it. This  
precisely positions the body onto the opti-  
cal axis of the microscope.  
4. Insert the eyepieces into the eyetubes.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
20  
 
Installing Objectives  
If you have purchased a standard micro-  
scope configuration, you will notice the  
1. As you rotate the nosepiece clockwise,  
attach the objectives by screwing them  
into the nosepiece holes starting with the  
lowest magnification and advancing to the  
highest magnification.  
objectives are already installed on the nose-  
piece and the substage condenser is already  
installed on the stand. In this case, go to section  
"Operation". If you purchased your Leica DM750  
P by components and not by standard configu-  
ration, please continue in this section.  
Always rotate the nosepiece by using  
the knurled nosepiece ring.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Leica DM750 P  
Manual  
21  
 
Substage Condenser  
3. Swing the polarizer, which is connected to  
the bottom of the condenser mount, to the  
left.  
Substage Condenser  
2. Adjust the condenser mount to its lowest  
position using the condenser focus knob  
on the left-hand side of the substage.  
The Leica DM750 P has an open substage  
condenser mount. Therefore, the  
condenser needs to be installed.  
1. Raise the stage to its highest position using  
the coarse focus knob on the right side of  
the microscope.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
22  
 
Substage Condenser (cont.)  
4. Completely loosen the two thumbscrews  
on the condenser mount.  
6. Raise the condenser to its highest position  
using the condenser focus on the left-hand  
side of the substage  
7. Tighten the thumbscrews to roughly center  
the substage condenser so that the top lens  
of the condenser is centered beneath the  
objective in use. You will more accurately  
center and focus the condenser when you  
reach section Koehler Setup.  
5. Slide the substage condenser into the  
fork mount located underneath the stage  
by aligning the positioning pin on the  
condenser into the slot on the back of the  
fork.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
23  
Get Set!  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
24  
 
Basic setup  
As a safety precaution, the power cord  
2. Plug the microscope power cord into  
suitable grounded electrical outlet.  
4. Turn on the microscope using the switch on  
the bottom right-hand side of the micro-  
scope stand.  
has been grounded to the microscope  
base. Never use an adapter between the power  
cord and the power source; it will render the  
grounding feature ineffective.  
a
A grounded 3-wire cord is provided.  
3. Set the illumination control (knob is located  
on the bottom left-hand side of the micro-  
scope stand) to the lowest setting to start.  
The illumination control knob allows you  
to adjust the intensity of light produced by  
the illumination system.  
Work Surface  
Always use your microscope on a hard,  
stable surface.  
Power  
1. If the power cord is not already attached,  
attach it securely to the back of the micro-  
scope.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
25  
 
Using the Aperture Condenser  
The condenser is furnished with an iris  
diaphragm, which can be adjusted to  
match the effective numerical aperture of the  
objective.  
1. To open and close this diaphragm, simply  
rotate the knurled condenser ring to the  
right or to the left so that the line on the  
rotating ring aligns with the objective  
magnification in use.  
2. To start, fully open the iris diaphragm of the  
aperture condenser by rotating the ring to  
the extreme right.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
26  
 
Setting up Polarized Light  
The polarizer is located under the  
substage condenser in a swing in/out  
mount. Swing it into place and be sure it "clicks"  
into place.  
1. Loosen the polarizer locking thumbscrew,  
which is located on the right-hand side of  
the polarizer.  
2. Rotate the polarizer to the zero position.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
27  
 
Setting up Polarized Light (cont.)  
The Analyzer is located in the module  
between the viewing tube and the stand  
either in a slider module or in an Analyzer  
Bertrand Lens module. If you have a slider, slide  
the Analyzer into position.  
3. If you have an Analyzer/Bertrand Lens  
4. Increase the illumination intensity slightly.  
You will notice while looking into the  
microscope with the Polarizer and Analyzer  
in place and no specimen, that the field will  
look dark. If not, rotate the polarizer slightly  
until you have the darkest field of view.  
module, swing in the Analyzer by flipping  
the lower rocker of the module to the  
right (the "A" position/Analyzer in), while  
ensuring the upper rocker of the Analyzer/  
Bertrand Lens Module is flipped to the left  
(the "O" position/Bertrand Lens out).  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
28  
Focusing  
Focusing  
3. Raise the microscope stage using the  
coarse adjustment knob until you reach its  
positive stop or the specimen touches the  
objective.  
4. Look into the eyepieces and adjust the illu-  
mination intensity so that it is comfortable  
for viewing.  
1. Place a polarized light specimen on the  
stage.  
2. Rotate the 10× objective into working posi-  
tion.  
5. Using the fine focus adjustment knob, bring  
the specimen into sharp focus.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
29  
 
Viewing Tube Adjustment  
Adjust the eyetubes for your interpupillary  
distance.  
2. Focus the crosshair reticle by gripping the  
knurl and rotating the top of the eyepiece  
until the reticle is in clear focus.  
This time, focus the specimen by using  
the focusing capability in the focusing  
eyepiece. Do not raise or lower the stage to  
focus!  
1. Fold or unfold the eyetubes until you see  
one illuminated circle.  
3. While looking through only the crosshair  
eyepiece, bring the specimen into focus  
with the fine focus of the microscope.  
5. Grip the knurl on the focusing eyepiece  
with one hand and rotate the top of the  
eyepiece with the other hand until the  
specimen is in focus with this eye and this  
focusing eyepiece. This corrects for any  
vision differences between your right eye  
and left eye.  
The viewing tubes maintain a constant  
tube length for all interpupillary settings.  
This means that a change of interpupillary  
distance does not affect parfocality, magnifica-  
tion, or calibrations that depend on magnifica-  
tion.  
6. Now switch to a high magnification objec-  
tive (not an oil objective) and focus the  
microscope while viewing with both eyes.  
The higher magnifications have a shal-  
lower depth of field. Therefore, after  
focusing with a high magnification, you will  
find that when you change to a lower magni-  
fication, you only have to adjust the fine focus  
slightly, if at all.  
It may help to cover your left eye.  
To correctly set the eyepieces, be sure  
4. Now switch to look only through the  
other eyepiece (focusing eyepiece without  
crosshair) with the other eye.  
the crosshair eyepiece is seated securely  
in the slot of the right eyetube.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
30  
 
Koehler Set up  
Koehler Set up  
1. Close the Koehler field diaphragm on the base of the microscope so  
that its iris leaves are present within the field of view when looking  
into the eyepieces.  
3. Turn the condenser centering thumbscrews simultaneously to center  
the image of the field diaphragm in the field of view.  
2. Use the condenser focus knob on the left-hand side of the substage to  
bring the leaves of the field diaphragm into sharp focus.  
4. Open the field diaphragm until the iris leaves disappear just beyond  
the field of view.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
31  
 
Objective Centration  
Tools used  
2. Focus the polarizing sample.  
Ϙ
Centering keys  
3. Move a noticeable point of the sample to  
the center of the crosslines.  
1. Retrieve the two centering keys from the  
packaging and insert the centering tools  
into the centering holes located above the  
objective in use. Engage the ball end of the  
centering tools into the set screws, which  
are hidden in the centering holes.  
4. Loosen the stage brake and rotate the stage  
until the noticeable point of the sample is  
furthest away from the center of the cross-  
lines. This may even be outside of the field  
of view. This indicates the objective is not  
centered to the stage.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
32  
 
Objective Centration (cont.)  
5. Adjust the objective centering screws  
until the noticeable point of the sample is  
midway between the furthest position and  
the center of the crosslines. This moves the  
objective center closer to the stage center.  
6. Move the specimen (not the objective  
centering screws!) so that the noticeable  
point is at the center of the crosslines once  
again.  
7. Check to make sure that the noticeable  
point stays at the center of the crosslines  
when the stage is rotated. If not, repeat the  
centering process.  
Each objective must be centered sepa-  
rately. Also, there is a magnetic storage  
location for the objective centering tools in the  
back of the Analyzer or Analyzer/Bertrand Lens  
modules.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
33  
Bertrand Lens Operation  
If you have an Analyzer/Bertrand Lens  
You can now view the conoscopic image  
through the eyepieces.  
4. If you have an Analyzer/Bertrand Lens with  
a focusable Bertrand Lens, you can now  
turn the knob on the upper rocker to focus  
the Bertrand Lens image.  
module, you can view  
a
conoscopic  
image through the eyepieces.  
3. If necessary, you can fine tune the Bertrand  
Lens centration with the tools provided.  
1. Bring the portion of the specimen you  
would like to view in conoscopy (with the  
Bertrand Lens) to the center of the field  
2. Swing in the Bertrand Lens by flipping the  
upper rocker of the Analyzer/Bertrand Lens  
module to the "B" position. Be sure the  
Polarizer and Analyzer are also in position.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
34  
 
Time Delay Shutoff  
The Leica DM 750 P is equipped with  
a time delay shutoff capability, which  
automatically turns the illumination off after 2  
hours of no changes in the intensity control.  
Changing the status of the time delay  
shutoff  
Ϙ
Ϙ
The LED Illumination will flash to indicate  
the time delay shutoff status was changed.  
The LED Illumination will flash two times  
slowly then stay on when you disable the  
Time Delay Shutoff.  
1. Rotate the intensity control to the lowest  
setting.  
Ϙ
The time delay shutoff IS enabled as a  
default.  
2. Turn the power on.  
Ϙ
The LED Illumination will flash three times  
quickly then stay on when you enable the  
Time Delay Shutoff.  
3. Rotate the intensity control to full high  
intensity, then back to full low intensity all  
within 5 seconds.  
When you turn the power off and then  
turn the power on again, the system will  
be in the last Time Delay Shutoff status (Enabled  
or Disabled) and you will not see any blinking.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
35  
Go!  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
36  
 
Ready! Set! Go!  
Now all you need to do is change objectives, set  
the aperture condenser and field diaphragm  
appropriately for the objective magnification in  
use, and enjoy the view!  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
37  
 
Care of the Microscope  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Leica DM750 P  
Manual  
38  
 
General maintenance  
General  
Ϙ
The cord wrap allows you to wrap the cord  
in such a way that only the length you need  
is extended.  
Ϙ
Proper storage of accessories in the appro-  
priate positions is important to keep them  
safe from loss or damage. There are storage  
places for the objective centering tools and  
for two compensators in the rear of the  
Analyzer/Bertrand Lens module.  
Always carry the microscope using two  
hands. There is a handle on the back of  
the microscope and an undercut in the front for  
this purpose.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
39  
 
General maintenance (cont.)  
Ϙ
Ϙ
Ϙ
Keeping all optical components clean is  
important for maintaining good optical  
performance.  
Ϙ
It is very important to avoid the excessive  
use of solvent, so use them sparingly. The  
lint-free cloth, lens tissue or cotton swab  
should be moistened with solvent, but  
not be wet enough for the solvent to seep  
around the lens.  
concave front lens of fairly short radius or  
curvature. The surface of this front lens can  
be readily cleaned with a toothpick covered  
with a cotton tip, or with a small cotton  
swab. Moisten the cotton with metha-  
nol and squeeze it almost dry. Wipe the  
front lens lightly without applying undue  
force or scrubbing action. Make sure that  
the cotton tip contacts the concave lens  
surface. Check the objective with a magni-  
fier after cleaning.  
The microscope should always be covered  
with the plastic dust cover (provided with  
the instrument) when it is not in use.  
Ϙ
No part of the microscope is quite so vulner-  
able to collecting dirt, dust, and oil as the  
front lens of the objective. Whenever you  
encounter lack of contrast, cloudiness or  
poor definition, carefully check the condi-  
tion of the front lens with a magnifier.  
If any optical surface becomes coated with  
dust or dirt, clean the surface by blowing it  
off with a syringe or brushing it off with a  
camel hair brush before attempting to wipe  
the surface clean.  
Ϙ
If you need to remove the microscope’s  
viewing body, be careful not to accidentally  
touch the outer lens surface (located on  
the underside of the body). Fingerprints on  
this surface will reduce image clarity. This  
lens can be cleaned in the same manner as  
objectives and eyepieces.  
Ϙ
Optical surfaces should be cleaned with a  
lint-free cloth, lens tissue or a cotton swab  
moistened with methanol or commercially  
available glass cleaner. Do not use alcohol.  
Ϙ
The lower magnification objectives have  
fairly large front lenses and can be cleaned  
with a cloth or lens tissue wrapped around  
a finger and moistened with methanol. Do  
not use alcohol.  
Ϙ
Cleaning 40× and 100× objectives requires  
more care. Note: To achieve the high degree  
of flatness obtained with the higher magni-  
fication objectives, the objective has a small  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
40  
Mechanical Maintenance  
Mechanical Maintenance  
Ϙ
Use the dust cover. It is the most impor-  
tant way to keep the microscope in good  
mechanical and physical condition.  
Ϙ
The Leica DM750 P has a durable finish. All  
non-optical parts of the instrument can be  
cleaned with methanol, naptha, or soap  
and water. Avoid the use of all other organic  
solvents when cleaning the instrument.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
41  
 
Dimensions  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
42  
 
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
Manual  
Leica DM750 P  
45  

Land Pride Lawn Aerator CA2560 User Manual
Lennox Hearth Indoor Fireplace 875027M User Manual
LG Electronics Flat Panel Television PM9700 User Manual
Makita Grinder GD0600 User Manual
M Audio Stereo Amplifier Octane User Manual
Maytag Washer Dryer LAT5006 User Manual
Maytag Washer MGR8670W User Manual
Melissa Bread Maker 643 111 User Manual
Mellerware Slow Cooker 27565 User Manual
Midland Radio Two Way Radio GXT300 Series User Manual