Leica Digital Camera SP2600 User Manual

Leica SP2600  
Ultramilling attachment  
Supplementary instruction manual Leica SP2600  
To be used only in conjunction  
with instruction manual Leica SM2500 .  
V1.2 English – 07/2002  
Always keep this manual near the instrument.  
Read carefully prior to operating the instrument.  
Download from Www.Somanuals.com. All Manuals Search And Download.  
NOTE  
The information, numerical data, notes and value  
These are determined only by the contract provi-  
sions agreed between ourselves and our cus-  
tomers.  
judgments contained in this manual represent the  
current state of scientific knowledge and state-  
of-the-art technology as we understand it follow-  
ing thorough investigation in this field.  
Leica reserves the right to change technical  
specifications as well as manufacturing process-  
es without prior notice. Only in this way is it pos-  
sible to continuously improve the technology and  
manufacturing techniques used in our products.  
This document is protected under copyright laws.  
Any copyrights of this document are retained by  
Leica Microsystems Nussloch GmbH.  
Any reproduction of text and illustrations (or of  
any parts thereof) by means of print, photocopy,  
microfiche, web cam or other methods – includ-  
ing any electronic systems and media – requires  
express prior permission in writing by Leica Mi-  
crosystems Nussloch GmbH.  
We are under no obligation to update the present  
manual according to the latest technical devel-  
opments, nor to provide our customers with addi-  
tional copies, updates etc. of this manual.  
For erroneous statements, drawings, technical  
illustrations etc. contained in this manual we ex-  
clude liability as far as permissible according to  
the national legal system applicable in each indi-  
vidual case. In particular, no liability whatsoever  
is accepted for any financial loss or consequen-  
tial damage caused by or related to compliance  
with statements or other information in this man-  
ual.  
For the instrument serial number and year of  
manufacture, please refer to the name plate at  
the back of the instrument.  
Statements, drawings, illustrations and other in-  
formation as regards contents or technical de-  
tails of the present manual are not to be consid-  
ered as warranted characteristics of our  
products.  
© Leica Microsystems Nussloch GmbH  
Published by:  
Leica Microsystems Nussloch GmbH  
Heidelberger Str. 17 - 19  
D-69226 Nussloch  
Germany  
Telephone: +49 (0)6224 143-0  
Facsimile: +49 (0)6224 143-200  
Leica SP2600 – Ultramilling attachment  
3
Download from Www.Somanuals.com. All Manuals Search And Download.  
Table of contents  
1.  
Important information................................................................................................................ 6  
1.1 Symbols in the text and their meaning ................................................................................................ 6  
1.2 Designated use........................................................................................................................................ 6  
1.3 Information on this instruction manual ................................................................................................ 7  
2.  
Safety ............................................................................................................................................ 8  
2.1 Safety regulations ................................................................................................................................... 8  
2.2 Built-in safety devices ............................................................................................................................ 9  
2.3 Safety instructions ................................................................................................................................ 10  
3.  
4.  
Technical Data ......................................................................................................................... 13  
Installation................................................................................................................................. 14  
4.1 Overview - Leica SP2600 Ultramilling attachment / basic instrument Leica SM2500  
- instrument parts / functions .............................................................................................................. 14  
4.2 Leica SP2600 standard delivery .......................................................................................................... 15  
4.3 Site requirements.................................................................................................................................. 15  
4.4 Unpacking, transport, installation ...................................................................................................... 15  
4.5 Installing the Leica SP2600 ultramilling attachment........................................................................ 16  
4.6 Connecting the vacuum cleaner......................................................................................................... 17  
4.7 Installing additional accessories........................................................................................................ 18  
5. Operation ..................................................................................................................................... 19  
5.1 Overview of chapter 5 ‘Operation...................................................................................................... 19  
5.2 Adjusting the pre-millers in the calibrating device.......................................................................... 20  
5.3 Adjusting the finishing millers in the calibrating device ................................................................. 25  
5.4 Specimen holder systems - how to combine individual components .......................................... 29  
5.5 Clamping specimen .............................................................................................................................. 30  
5.5.1 Installing the base plates in the basic instrument ........................................................................... 30  
5.5.2 Clamping the specimen holders in the base plate with dove tail guide ....................................... 30  
5.5.3 Clamping the vacuum stages on/in the base plates ........................................................................ 31  
5.5.4 Connecting the vacuum pump / placing specimens onto the vacuum stages ............................ 32  
5.6 Preparatory steps prior to the milling procedure ............................................................................ 34  
5.7 Pre-milling .............................................................................................................................................. 36  
5.7.1 Perspective view of the pre-milling procedure................................................................................ 38  
5.8 Finishing milling ..................................................................................................................................... 40  
5.8.1 Perspective view of the finishing milling procedure ....................................................................... 42  
5.9 Switching the instrument off (switching off the control unit .......................................................... 43  
Supplementary instruction manual V 1.2 – 07/2002  
4
Download from Www.Somanuals.com. All Manuals Search And Download.  
Important information  
6.  
7.  
Troubleshooting ....................................................................................................................... 44  
Cleaning, disinfection and maintenance ............................................................................ 45  
7.1 Cleaning and disinfection .................................................................................................................... 45  
7.2 Maintenance.......................................................................................................................................... 46  
7.2.1 Resharpening the millers ..................................................................................................................... 46  
7.2.2 Replacing the light bulb in the calibration device............................................................................ 47  
8.  
Overview - accessories .......................................................................................................... 48  
8.1 Miller holders and millers .................................................................................................................... 48  
8.2 Base plates for specimen holders and vacuum stage adapters ................................................... 48  
8.3 Specimen holders ................................................................................................................................. 48  
8.4 Vacuum stage adapters ....................................................................................................................... 48  
8.5 Vacuum stages and matching lids / perforated specimen holders .............................................. 49  
8.6 Vacuum pumps ...................................................................................................................................... 49  
8.7 Additional accessories ........................................................................................................................ 50  
9. Warranty and service ................................................................................................................ 51  
Leica SP 2600 – Ultramilling attachment  
5
Download from Www.Somanuals.com. All Manuals Search And Download.  
1.  
Important information  
1.1 Symbols in the text and their  
meaning  
1.2 Designated use  
Warnings and cautions appear in a  
grey box and are marked by a warning  
triangle.  
Failure to comply with these warnings/  
cautions can lead to accidents, bodily  
injury and damage to instrument and/  
or accessories.  
Fig. 1.1  
The Leica SP2600 is designated to prepare milled  
surfaces and thin slices of specimen materials  
such as plastics and metals as well as of organic  
materials (such as bones, teeth) in medicine, biol-  
ogy and industry.  
Notes, i.e. important information for the  
user, appear in a grey box and are  
marked by an information sign.  
The Leica SP2600 ultramilling attachment is an  
accessory which cannot be operated as a stand-  
alone device but must be used in conjunction  
with the Leica SM2500 sliding microtome de-  
scribed in a separate instruction manual.  
Figures in brackets refer to item num-  
bers in illustrations or to the illustra-  
tions themselves.  
(5)  
Specimen surfaces are milled with a rotating  
spindle, equipped with either a pre- or a finishing  
miller.  
With each sledge stroke, the micrometer feed  
mechanism of the Leica SM2500 lowers the mill-  
er towards the specimen by the feed setting se-  
lected.  
Function keys that need to be pressed  
ENTER on the control unit are printed in bold  
and capital letters.  
The Leica SP2600 may be used only in conjunc-  
tion with accessories and supplementary instru-  
ments made by Leica.  
If the instrument is used for any other than  
the designated application, this will be  
considered an improper use of the product!  
Supplementary instruction manual V 1.2 – 07/2002  
6
Download from Www.Somanuals.com. All Manuals Search And Download.  
1. Important information  
1.3 Information on this instruction manual  
Instrument type  
This supplementary instruction manual must be used together with the Leica  
SM2500 sliding microtome instruction manual describing the basic instru-  
ment Leica SM2500, which is needed to operate the Leica SP2600 ultramill-  
ing attachment.  
All information provided in this supplementary instruction manual applies  
only to the instrument type indicated on the title page.  
A name plate indicating the instrument type and serial number is attached to  
the rear of the Leica SP2600 ultramilling device.  
Fig. 1.2  
Working with this instruction manual  
This supplementary instruction manual contains important instructions and  
information related to the operating safety and maintenance of the Leica  
SP2600 ultramilling attachment.  
It is an important part of the Leica SP2600 ultramilling attachment.  
Prior to setting up and operating the Leica SP2600 together with the Leica  
SM2500 be sure to read carefully both this supplementary instruction  
manual as well as the Leica SM2500 instruction manual, with the exception  
of chapters 5.5.6 - 5.7 of the Leica SM2500 manual. These chapters describe  
installation of knife holders / knives and sectioning with the Leica SM2500  
and are not of importance for operating the Leica SP2600 ultramilling attach-  
ment.  
If additional requirements on accident prevention and environmen-  
tal protection exist in the country of operation, this supplementary  
instruction manual must be supplemented by additional appropri-  
ate instructions to ensure compliance with such requirements.  
Leica SP 2600 – Ultramilling attachment  
7
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.  
Safety  
2.1 Safety regulations  
This instrument has been built and tested in accordance with the same di-  
rectives and standards as the Leica SM2500.  
For a complete list of the directives and standards applied please refer to  
chapter 2.1, page 8 of the Leica SM2500 instruction manual.  
In order to maintain the SP2600 in a condition compliant with the above di-  
rectives and standards and to ensure safe operation, the operator must ob-  
serve the instructions and warnings contained in this supplementary in-  
struction manual as well as in the Leica SM2500 instruction manual.  
The Leica SP2600 ultramilling may be operated only in conjunction  
with the Leica SM2500 sliding microtome!  
Supplementary instruction manual V 1.2 – 07/2002  
8
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Safety  
2.2 Built-in safety devices  
The instrument is equipped with the following safety devices:  
1
Plexiglass protective cover  
The miller is equipped with a protective cover made of plexiglass (1).  
When the cover is moved to the upper limit position (1, Fig. 2.1) the miller  
is automatically blocked and cannot be started.  
For any manipulations in the milling area (such as inserting millers, fine  
adjustment of specimen height etc.) the protective cover (1, Fig. 2.1) must  
be in the upper limit position.  
Fig. 2.1  
1
To be able to start the miller, the protective cover must be lowered (1,  
Fig. 2.2).  
Never touch the milling spindle while the plexiglass cover is shut,  
no matter whether the spindle is rotating or not.  
Fig. 2.2  
Important note on the safety devices built in by the manufacturer:  
The safety devices provided by the manufacturer constitute only a basis for  
accident prevention.  
Major responsibility for accident prevention during the use of the instrument  
rests with the employer who has the operating authority for the instrument  
and, in addition, with the persons, designated by him, who operate, service  
and repair the instrument.  
To ensure safe and trouble-free operation of the instrument, be sure to  
comply with all instructions, warning and cautions in this supplementary  
instruction manual as well as in the Leica SM2500 instruction manual!  
Leica SP 2600 – Ultramilling attachment  
9
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.  
Safety  
Safety instructions - transport and installation  
2.3 Safety instructions  
Be sure to follow the unpacking in-  
structions (attached to the outside of  
the instrument shipping crate)!  
Be sure to read and comply with the  
safety instructions, warnings and cau-  
tions in this chapter as well as in  
chapter 2 of the Leica SM2500 instruc-  
tion manual, even if you are already fa-  
miliar with the operation and use of  
other Leica products.  
Never remove or modify any of the  
safety devices installed on the instru-  
ment and/or accessories!  
Before mounting the ultramilling at-  
tachment, make sure the bearing  
blocks of the microtome supporting the  
milling attachment and the bearing  
surfaces of the milling attachment are  
clean.  
Two people are required to place the  
milling attachment on the bearing  
blocks from above.  
Do not slide the milling attachment  
onto the bearing blocks to prevent the  
bearing surfaces of the microtome and  
milling attachment from getting  
scratched!  
The bearing surfaces must not become  
damaged under any circumstances!  
Safety instructions - general  
The Leica SP2600 ultramilling attach-  
ment may only be operated in conjunc-  
tion with the Leica SM2500 sliding mi-  
crotome.  
Prior to working with the Leica SP2600  
ultramilling attachment, be sure to  
read the following instruction manu-  
als:  
• The entire Leica SM2500 instruction  
manual except chapters 5.5.6  
through 5.7.  
• This supplementary instruction  
manual Leica SP2600.  
Caution:  
When releasing the lid of the milling  
attachment (can be lifted and used as  
a handle) be careful not to get your fin-  
gers caught between the lid and the  
body of the milling attachment!  
The Leica SP2600 ultramilling attach-  
ment may only be operated by skilled  
laboratory personnel. It may only be  
used for the designated application  
and operated in accordance with the  
instructions contained in this supple-  
mentary instruction manual and in the  
Leica SM2500 instruction manual!  
Supplementary instruction manual V 1.2 – 07/2002  
10  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2. Safety  
Safety instructions - transport and installation  
Safety instructions -  
working with the instrument  
Never remove or modify any of the  
safety devices installed on the instru-  
ment and/or accessories!  
Caution!  
Only the vacuum cleaner offered by  
Leica or other vacuum cleaner brands  
rated at 230 V/ 50 Hz and a maximum  
power draw of 4 A can be connected  
directly to the SM2500 control unit (see  
page 17), and this only in countries  
where the mains supply is 230 V / 50  
Hz!  
Otherwise, the vacuum cleaner must  
be connected directly to mains and op-  
erated via the on/off switch of the  
vacuum cleaner.  
Never touch the milling spindle while  
the plexiglass cover is shut, no matter  
whether the spindle is rotating or not.  
Always slide the plexiglass cover to  
the upper position prior to attaching  
the miller holders to the milling  
spindle!  
Wear protective gloves when working  
with potentially infectious material!  
Always slide the plexiglass cover to  
the upper position prior to checking  
the distance between milling edge  
and specimen surface! - Risk of in-  
jury!  
Safety instructions - inserting the millers into the  
holders / working with the calibrating device  
Twisted or deformed springs for the  
miller holders must be exchanged!  
Always check the distance between  
miller cutting edge and specimen sur-  
face thoroughly prior to milling!  
Under no circumstances may the  
miller cutting edge be positioned more  
than a few microns (= selected milling  
thickness) below the specimen sur-  
face when starting the milling proce-  
dure.  
When the millers are inserted into the  
holders the letter ‘V’ (pre-millers) or ‘F’  
(finishing miller) must be visible for the  
milling edge to point into the right di-  
rection.  
Caution:  
The gauge contact plate must never hit  
the milling edge.  
This could damage the surface of the  
contact plate, leading to incorrect  
measuring results.  
Be very careful when drawing the con-  
tact plate close to the milling edge.  
Always use the magnifier.  
When spraying alcohol to moisten the  
specimen surface, remove excess liq-  
uid as soon as it starts accumulating.  
Do not operate near open flames or  
near other sources of ignition!  
No liquids may enter the interior of the  
instrument during operation!  
Leica SP2600 – Ultramilling attachment  
11  
Download from Www.Somanuals.com. All Manuals Search And Download.  
2.  
Safety  
Safety instructions -  
Safety instructions -  
working with the instrument  
cleaning disinfection and maintenance  
Only authorized service technicians  
may access the internal components of  
the instrument for service and repair!  
Do not exceed a milling thickness set-  
ting of 5 µm with the following materi-  
als: brittle, very ductile materials such  
as titanium or fiberglass and com-  
pound samples containing such mate-  
rials (such as printed circuit boards).  
Cleaning detergents appropriate for  
the ultramilling attachment.  
Varnished surfaces:  
• mild laboratory detergents (slightly  
moistened cloth)!  
Non-varnished metal parts:  
• mild laboratory detergents,  
• xylene substitutes (slightly  
moistened cloth)  
During milling:  
• switch on the vacuum cleaner,  
• wear appropriate safety goggles,  
• wear appropriate mask (type of  
mask depends on the material being  
milled / size of the chips that form  
during milling).  
• alcohol (slightly moistened cloth).  
Goggles and mask prevent milling  
chips and microdust from entering  
your eyes and from being inhaled.  
For disinfection of the instrument and  
accessories use Leica Cryofect or  
other commercial disinfectants. - Use  
all disinfectants according to the  
manufacturer’s instructions.  
Caution - rotating miller:  
After switching off the milling spindle  
(button (37) - chapter. 5.4.11, p. 34 - in-  
struction manual Leica SM2500) the  
miller continues to rotate for a little  
while.  
When disinfecting, wear appropriate  
goggles and mask as well as protec-  
tive gloves.  
When switching off the mains switch  
on the control unit (see chapter 5.2, p.  
21 - Instruction manual Leica SM2500)  
the milling spindle will rotate briefly.  
Therefore, never switch off the control  
unit, while one of your hands is in the  
rotation zone of the miller.  
Dispose of potentially infectious  
specimen material as per the labora-  
tory regulations applicable in the  
country of operation.  
No liquids may enter the interior of the  
instrument during cleaning, disinfec-  
tion or maintenance!  
- Risk of injury!  
Never touch the milling spindle while  
it is rotating and/or while the plexi-  
glass cover is shut, no matter whether  
the spindle is rotating or not.  
Supplementary instruction manual V 1.2 – 07/2002  
12  
Download from Www.Somanuals.com. All Manuals Search And Download.  
3. Technical Data  
Ultramilling attachment  
Milling spindle (special design) ......................................................................................... type TSAV 60 x 160  
Motor, rotation speed selectable in 100-rpm steps ............................................................ 500 to 3.000 rpm  
Protective class ................................................................................................................................................... I  
Pollution degree .................................................................................................................................................. 2  
Dimensions and weight - ultramilling attachment  
Dimensions (H x W x D) ..................................................................................................... 300 x 315 x 240 mm  
Weight ........................................................................................................................................................... 18 kg  
Vacuum cleaner ‘Fakir S20’  
Nominal voltage ........................................................................................................................................... 230 V  
Nominal frequency ...................................................................................................................................... 50 Hz  
Power draw ............................................................................................................................................ max. 4 A  
These specifications are supplemented by the specifications con-  
tained in the Leica SM2500 instruction manual.  
Leica SP 2600 – Ultramilling attachment  
13  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4.  
Installation  
4.1 Overview - Leica SP2600 Ultramilling attachment / basic instrument Leica SM2500  
- instrument parts / functions  
1.1  
1.2  
2
2.1  
2.2  
1.3  
1.4  
1.5  
Fig. 4.1  
1
1. Leica SM2500 - basic instrument (sliding microtome)  
1.1 Leica SP2600 - ultramilling attachment  
1.2 Carrier for ultramilling attachment SP2600 (when not in use)  
1.3 Specimen holder system (not part of standard delivery)  
1.4 Carrying handle, black, rear (extendible - not visible in Fig. 4.1)  
1.5 Carrying handle, black, front (extendible)  
2. Leica SM2500 - control unit  
2.1 Control panel  
2.2 Emergency stop button  
3. Foot switch (not shown in Fig. 4.1 - for details on the foot switch please  
refer to p. 18, chapter 4.6 of this supplementary instruction manual and to  
p. 18, chapter 4.5.2 and p. 24, chapter 5.4.1 of the Leica SM2500 manual).  
Supplementary instruction manual V 1.2 – 07/2002  
14  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Installation  
4.2 Leica SP2600 standard delivery  
Millers and specimen holders are not  
included in the instrument standard  
delivery but have to be ordered sepa-  
rately, in accordance with each user’s  
specific application(s).  
Mounting instructions for all accesso-  
ries and/or references to the corre-  
sponding chapters of the instruction  
manual Leica SM2500 can be found in  
chapter 5.6 of this instruction manual.  
Standard delivery includes:  
1
1
3
Basic instrument Leica SP2600 (Ultramilling  
attachment)  
Calibrating device with dial gauge for miller  
adjustment  
Holders, each with two springs, for pre- and  
finishing millers, one holder delivered attached  
to the instrument.  
1
1
Carrier for Leica SP2600 Ultramilling device  
when not in use  
Tool set:  
4.3 Site requirements  
For site requirements, please refer to chapter 4.2  
of the Leica SM2500 instruction manual.  
1
1
1
1
1
Allen key, size 2.5, with handle  
Allen key, size 4.0  
Allen key, size 4.0 with handle  
Allen key, size 5.0 with handle  
Allen key, size, 8.0  
4.4 Unpacking, transport, installation  
Be sure to follow the unpacking in-  
structions (attached to the outside of  
the instrument shipping crate)!  
1
1
1
1
Power cord – type KN 2/5 7,5 V  
Box level  
Plexiglass block, 19 x 33 x 53 mm  
Balance weight, installed in one of the miller  
holders  
Flat belt  
Never remove or modify any of the  
safety devices installed on the instru-  
ment and/or accessories!  
The Leica SP2600 ultramilling attach-  
ment is supplied only in conjunction  
with the Leica SM 2500 microtome  
(Leica SM2500 microtomes already in-  
stalled at a customer’s site must be re-  
turned to the manufacturer for upgrad-  
ing with a Leica SP2600 ultramilling  
attachment).  
1
1
1
Dust cover  
Supplementary instruction manual for Leica  
SP2600, to be used in conjunction with the  
Leica SM2500 instruction manual  
- German  
- English  
For instructions on how to install / set up the Lei-  
ca SM2500 please refer to chapter 4 of the Leica  
SM2500 instruction manual.  
Leica SP2600 – Ultramilling attachment  
15  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Installation  
4.5 Installing the Leica SP2600 ultramilling attachment  
Place the carrier (1, Fig. 4.2) for the ultramilling attachment (3, Fig. 4.3) on  
the rear part of the Leica SM2500 microtome fitting the two pins of the  
carrier into the corresponding bores in the Leica SM2500 microtome.  
1
2
Fig. 4.2  
Lower the bearing blocks (2) of the microtome as far as they will go (use  
button (51), chapter 5.4.13, p. 36 - Leica SM2500 manual).  
Prior to placing the ultramilling attachment onto the microtome:  
Make sure the bearing blocks of the microtome supporting the  
milling attachment and the bearing surfaces of the milling attach-  
ment are clean.  
Two people are required to place the milling attachment on the  
bearing blocks from above.  
4
3
Do not slide the milling attachment onto the bearing blocks to pre-  
vent the bearing surfaces from getting scratched!  
The bearing surfaces must not become damaged under any cir-  
cumstances!  
Place ultramilling device (3) onto the microtome bearing blocks (two  
people!).  
Lift lid (4) of the milling attachment (serves as a handle - see Fig. 4.3).  
Caution:  
When releasing the lid of the milling attachment (can be lifted and  
used as a handle) be careful not to get your fingers caught be-  
tween the lid and the body of the milling attachment!  
Fig. 4.3  
5
Fasten the ultramilling attachment to the microtome with four screws (5)  
using a size 8 Allen key.  
Connect the connecting cable of the milling attachment to the control  
unit (see chapter 4.5.2, page 18 of the Leica SM2500 manual).  
Supplementary instruction manual V 1.2 – 07/2002  
16  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Installation  
4.6 Connecting the vacuum cleaner  
Milling dust is extracted by a vacuum cleaner, through a nozzle (1) located in  
the milling area.  
230 V / 50 Hz Vacuum cleaners with a power draw of maximum 4 A  
can be connected directly to the SM2500 control unit and thus be  
switched on/off via the vacuum cleaner button on the SM2500 con-  
trol unit.  
1
Fig. 4.4  
Installing vacuum cleaners rated at 230 V/ 50 Hz / max. power draw 4 A  
Insert the connection piece of the vacuum cleaner hose into nozzle (1) at  
the rear of the ultramilling attachment.  
Connect any one of the two adapter cables (both cables are identical)  
supplied as part of the Leica SM2500 standard delivery into the vacuum  
cleaner socket at the control unit (see chapter 4.5.2, p. 18 of the Leica  
SM2500 manual).  
Connect the vacuum cleaner power cord to the socket of the adapter  
cable.  
Switch on the on/off switch of the vacuum cleaner.  
The vacuum cleaner can now be switched on/off via the vacuum cleaner  
button on the control unit (button (35), chapter 5.4.11, p. 34 of the Leica  
SM2500 manual).  
Caution!  
Only the vacuum cleaner offered by Leica or other vacuum cleaner  
brands rated at 230 V/ 50 Hz and a maximum power draw of 4 A can  
be connected directly to the SM2500 control unit, and this only in  
countries where the mains supply is 230 V / 50 Hz!  
Otherwise, the vacuum cleaner must be connected directly to  
mains and operated via the on/off switch of the vacuum cleaner.  
Installing other vacuum cleaners:  
Insert the connection piece of the vacuum cleaner hose into nozzle (1) at  
the rear of the ultramilling attachment.  
Connect the power cord of the vacuum cleaner to mains.  
Operate the vacuum cleaner via the vacuum cleaner on/off switch.  
Leica SP 2600 – Ultrafräse  
17  
Download from Www.Somanuals.com. All Manuals Search And Download.  
4. Installation  
4.7 Installing additional accessories  
Install further accessories as described in chapters 4.5.2 and 4.5.3, pages  
18 and 19 of the Leica SM2500 instruction manual.  
Connect the foot switch.  
The foot switch must be connected - otherwise the instrument will  
not be operational. --> For details on the foot switch, see chapter  
4.5.2, p. 18 and chapter 5.4.1, p. 24 of the Leica SM2500 instruction  
manual.  
If working with vacuum stages, also a vacuum pump must be in-  
stalled - --> see page 33, chapter 5.6.4 of this supplementary in-  
struction manual.  
Supplementary instruction manual V 1.2 – 07/2002  
18  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.1 Overview of chapter 5 ‘Operation’  
Prior to working with the Leica SP2600 ultramilling attachment, be  
sure to read the following instruction manuals:  
• The entire Leica SM2500 instruction manual except chapters  
5.5.6 through 5.7.  
• This supplementary instruction manual Leica SP2600.  
Chapter 5 of this supplementary instruction manual Leica SP2600 includes  
the following subchapters:  
5.1 Overview - chapter 5.  
5.2 Adjusting the pre-millers in the calibrating device  
5.3 Adjusting the finishing millers in the calibrating device  
5.4 Specimen holder systems - how to combine individual components  
5.5 Clamping specimens  
5.5.1 Installing the base plates in the basic instrument  
5.5.2 Clamping the specimen holders in the base plate with dovetail guide  
5.5.3 Clamping the vacuum stages on/in the base plates  
5.5.4 Connecting the vacuum pump/placing specimens onto the vacuum  
stages  
5.6 Preparatory steps prior to the milling procedure  
5.7 Pre-milling  
5.7.1 Perspective view of the pre-milling procedure  
5.8 Finishing milling  
5.8.1 Perspective view of the finishing milling procedure  
5.9 Switching the instrument off (switching off the control unit).  
Initial operation:  
The first time you operate the instrument, we recommend you work through  
the above subchapters, in numerical order as listed.  
Pay particular attention to the safety instructions..  
Leica SP 2600 – Ultramilling attachment  
19  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.2 Adjusting the pre-millers in the  
calibrating device  
Preparatory steps  
Connect mains power supply (20) to miller cali-  
bration device (18).  
Connect mains power supply (20) to mains: the  
light bulb is switched on.  
Place magnifier (16) onto calibration device.  
20  
Loosen regulating screws (6), (7) and (11) in  
miller holder (2) until they are flush with the  
inner surfaces of the miller holder.  
18  
If the screws protrude from the inner surfaces  
of the miller holder (2), spring (1) and miller (3)  
cannot be inserted into the correct position.  
16  
Twisted or deformed springs for the  
miller holders must be exchanged!  
1
Inserting the miller into the miller holder  
Insert spring (1) straight from above into miller  
holder (2) - the two upper edges of the spring  
must be flush with the upper surface of the  
miller holder.  
11  
6
Carefully insert miller (3) from the front - the  
letter ‘V’ must be visible.  
6
When the miller is inserted into the  
holder the letter ‘V’ must be visible. -  
That way, the milling edge is pointing  
into the right direction.  
4
Insert the miller (3) completely - it must fit  
against the rear and lower inner surface of the  
miller holder (2) and be flush with spring (1).  
Insert screw (4) and washer (5) and tighten  
slightly to fasten miller (3) .  
2
7
3
5
Fig. 5.2  
Supplementary instruction manual V 1.2 – 07/2002  
20  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Adjusting the miller  
Y
The miller (3) is adjusted in the holder (2) along  
three different axis (x, y and z - see Fig. 5.3, left).  
6
Adjustment along the x-axis is done outside the  
calibration device (18, Fig. 5.1).  
For adjustment along the z- and y-axis, the miller  
holder (2) with miller (3) has to be fastened in the  
calibration device.  
6
Z
X
Adjustment along the x-axis  
Adjust miller (3) as shown in Fig. (5.4a / 5.4b).  
Along the entire length of both sides of the  
miller, there must be a uniform distance be-  
tween miller (3) and miller holder (2) / spring  
(1) (see Fig. 5.4b).  
The miller is adjusted along the x-axis via the  
two regulating screws (6) (Allen key, size 2.5).  
Tighten (clockwise) and/or loosen (counter-  
clockwise) screws (6, Fig. 5.3) alternately un-  
til the miller is adjusted to a parallel position  
as shown in Fig. 5.4b.  
2
3
Fig. 5.3  
2
1
3
Do not tighten screws (6) too much. If  
screws (6) are too tight, screw (7, Fig.  
5.2) (for adjustment along the y-axis)  
can no longer be moved.  
Fig. 5.4a: WRONG  
Fig. 5.4b: CORRECT  
Leica SP2600 – Ultramilling attachment  
21  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Adjustment along the z-axis  
18  
1. Pull lever (12) at the left of calibration device  
(18) forward as far as it will go, thus moving  
contact plate (17) backwards to its rear limit  
position. This prevents the contact plate from  
hitting the milling edge during the adjustment  
procedure.  
12  
2. Once miller (3) has been correctly adjusted  
along the x-axis, place miller holder (2) with  
miller (3) next to block (8) into the calibration  
device.  
3. Fit the clamping device (9) (with the semicir-  
cular recess pointing in your direction) against  
the miller holder (2) and fasten with two  
knurled screws (10) (turn screws clockwise).  
16  
17  
10  
4. Rotate magnifier (16) forward to a position  
9
above the miller.  
8
2
3
Fig. 5.5  
5. With lever (12), bring contact plate (17) close  
to the miller (move lever backwards) until just  
a very small gap remains (see Fig. 5.6).  
(12)  
Caution:  
The gauge contact plate must never hit  
the milling edge.  
This could damage the surface of the  
contact plate, leading to incorrect  
measuring results.  
Be very careful when bringing the con-  
tact plate close to the milling edge.  
Always use the magnifier.  
17  
Fig. 5.6  
Supplementary instruction manual V 1.2 – 07/2002  
22  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
6. Insert the size-5 Allen key into regulating  
screw (4) insert size-2.5 Allen key into regu-  
lating screw (11) (Fig. 5.7).  
11  
3
4
The arrows in Fig. 5.7 indicate clock-  
wise sense of rotation.  
7
Fig. 5.7  
7. Tighten/loosen screws (11) and (4) alternately  
to adjust miller (3) along the z-axis. The miller  
edge is adjusted correctly when the outer  
edge of the miller is at right angles to the front  
surface of the contact plate (see Fig. 5.8a).  
The miller MUST NOT be positioned as shown  
in Fig. 5.8b and 5.8c!  
90°  
Fig. 5.8 a - CORRE
Fig. 5.8 b - WRON
Fig. 5.8 c - WRONG  
Tightening either regulating screw (4,  
11) (= clockwise rotation in the sense  
of the arrow) brings the corresponding  
side of the miller closer to the contact  
plate.  
8. If one of the two screws (4, 11) has been tight-  
ened as far as it will go, slightly release the  
opposite screw. This slackens the screw that  
had been completely tightened .  
Do not tighten screws (4, 11) too much.  
If screws (4, 11) are too tight, screw (7,  
Fig. 5.2) (for adjustment along the y-  
axis) can no longer be moved.  
Caution:  
Be very careful when bringing the  
contact plate close to the milling  
edge.  
Always use the magnifier.  
Make sure the milling edge does not  
hit the contact plate!  
Leica SP2600 – Ultramilling attachment  
23  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Adjustment along the y-axis  
12  
1. Select reference value ‘1’.  
The reference value will be needed  
14  
13  
later on to adjust the finishing miller to  
the same milling level as the pre-  
miller.  
Any reference value can be selected.  
Experience, however, shows that ‘1’ is  
the best value for new millers (i.e. mill-  
ers not yet resharpened).  
17  
3
2. To select the reference value, move lever (12),  
setting the dial of the inner (smaller) scale (13)  
to a position just below ‘1’ and the dial of the  
outer scale (14) to approximately ‘70’ (approx.  
10 o’clock) (Fig. 5.9a).  
7
Fig. 5.9 a  
3. Rotate regulating screw (7) clockwise, bring-  
ing the milling edge (3) close to the contact  
plate (17).  
4. Slowly continue to turn screw (7) clockwise  
until the dial of scale (14) starts to move (i.e.  
until the milling edge touches the contact  
plate).  
14  
13  
5. Keep rotating screw (7) very carefully and  
slowly until the dial of scale (14) is in the up-  
per ‘0’ position and the dial of scale (13) is lo-  
cated exactly at ‘1’ (Fig. 5.9b).  
Caution:  
Be very careful when bringing the con-  
tact plate close to the milling edge.  
Always use the magnifier.  
Make sure the milling edge does not  
hit the contact plate!  
7
Fig. 5.9 b  
Supplementary instruction manual V 1.2 – 07/2002  
24  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.3 Adjusting the finishing millers in the  
calibrating device  
Carry out preparatory steps as per chapter 5.2,  
page 20 (‘Adjusting the pre-millers in the cali-  
brating device‘).  
Twisted or deformed springs for the  
miller holders must be exchanged!  
Insert spring and finishing miller as described  
in chapter 5.2, page 20 (‘Inserting the miller into  
the miller holder’).  
When the finishing miller is inserted  
into the holder, the letter ‘F’ must be  
visible. - That way, the milling edge is  
pointing into the right direction.  
Fig. 5.10  
Leica SP2600 – Ultramilling attachment  
25  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Adjusting the finishing miller  
Just like the pre-miller, the finishing miller is ad-  
justed in the miller holder along three different  
axis (see page 21).  
Adjustment along the x-axis  
As described for the pre-miller - see page 21.  
Adjustment along the z-axis  
Carry out steps 1 - 4 as described on page 22  
(Fig. 5.5) for the pre-miller.  
Next:  
5. With lever (12), bring contact plate (17) close  
to the miller (turn lever backwards) so that only  
a small gap remains (see Fig. 5.11).  
12  
Caution:  
The gauge contact plate must never hit  
the milling edge.  
This could damage the surface of the  
contact plate, leading to incorrect  
measuring results.  
Be very careful when bringing the con-  
tact plate close to the milling edge.  
Always use the magnifier.  
17  
Fig. 5.11  
Supplementary instruction manual V 1.2 – 07/2002  
26  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
6
7
Insert size-5 Allen key into regulating screw  
(4) insert size-2.5 Allen key into regulating  
screw (11) (Fig. 5.12) .  
11  
3
4
The arrows in Fig. 5.12 indicate clock-  
wise sense of rotation.  
7
Fig. 5.12  
Tighten/loosen screws (11) and (4) alternately  
to adjust miller (3) along the z-axis. The miller  
is adjusted correctly when angle ’a‘ equals  
angle ’b’, i.e. when the distance between each  
end of the milling edge and the contact plate  
is the same: in Fig. 5.13a the finishing miller  
correctly adjusted.  
The miller MUST NOT be positioned as shown  
in Fig. 5.13b and 5.13c!  
b
a
Tightening either regulating screw (4,  
11) (= clockwise rotation in the sense  
of the arrow) moves the corresponding  
side of the miller towards the contact  
plate.  
Fig. 5.13 a - CORRECT: angle a = angle b  
8. If one of the two screws has been tightened  
as far as it will go, slightly release the oppo-  
site screw . This slackens the screw that had  
been completely tightened.  
Fig. 5.13 b - WRONG  
Do not tighten screws (4, 11) too much.  
If screws (4, 11) are too tight, screw (7)  
(for adjustment along the y-axis) can  
no longer be moved.  
Caution:  
Be very careful when bringing the  
contact plate close to the milling edge.  
Always use the magnifier.  
Make sure the milling edge does not  
hit the contact plate!  
Fig. 5.13 c - WRONG  
Leica SP2600 – Ultramilling attachment  
27  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Adjustment along the y-axis  
12  
1
Select the same reference value as was set  
for the pre-miller.  
14  
13  
Pre-miller and finishing miller used for  
milling a specific sample must be ad-  
justed using the same reference value.  
This ensures that both milling edges  
hit the sample at exactly the same  
height.  
Minor differences in height can still be  
adjusted via the height adjustment but-  
ton on the control unit, once the miller  
holder has been attached to the milling  
spindle (see chapter 5.4.13, p. 36 the  
Leica SM2500 manual).  
17  
3
2
To select the reference value, move lever (12),  
setting the dial of the inner (smaller) scale (13)  
to a position just below ‘1’ and the dial of the  
outer scale (14) to approximately ‘70’ (approx.  
10 o’clock) (Fig. 5.14a).  
Fig. 5.14 a  
7
3
4
Rotate regulating screw (7) clockwise, bring-  
ing the milling edge (3) close to the contact  
plate (17).  
Slowly continue to turn screw (7) clockwise  
until the dial of scale (14) starts to move (i.e.  
until the milling edge touches the contact  
plate).  
Keep rotating screw (7) very carefully and  
slowly until the dial of scale (14) is in the up-  
per ‘0’ position and the dial of scale (13) is lo-  
cated exactly at ‘1’ (Fig. 5.14b).  
14  
13  
5
Caution:  
Be very careful when bringing the con-  
tact plate close to the milling edge.  
Always use the magnifier.  
Make sure the milling edge does not  
hit the contact plate!  
7
Fig. 5.14 b  
Supplementary instruction manual V 1.2 – 07/2002  
28  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.4 Specimen holder systems - how to combine individual components  
Base plate  
with vise  
0383 09267  
Vacuum stage  
adapter for base  
plate with vise  
0383 60407  
Vacuum stage  
55 x 100 mm  
0364 09011  
SPECIMEN  
Vacuum stage  
for lids  
30 - 50 mm  
0364 09001  
Lid, 30 mm  
0364 09005  
Lid, 40 mm  
0364 09007  
SPECIMEN  
Vacuum stage  
76 x 25 mm,  
for perforated  
specimen holders  
0364 18199  
Lid, 50 mm  
0364 09009  
Base plate with  
dovetail guide  
0383 09422  
Vacuum stage  
adapter for base  
plate with  
dovetail guide  
0383 20351  
Vacuum stage  
with variable  
surface:  
76 x 25, 76 x 65  
and 100 x 65 mm,  
for perforated  
specimen holders  
0364 18197  
Specimen holder,  
perforated,  
in 9 different sizes:  
5 x 5 perforations  
0364 26808  
5 x 10 perforations  
0364 26809  
10 x 10 perforations  
0364 26810  
10 x 15 perforations  
0364 26811  
15 x 15 perforations  
0364 26812  
20 x 10 perforations  
0364 26813  
20 x 15 perforations  
0364 26814  
Specimen clamp  
44 x 58 mm, for  
base plate with  
dovetail guide  
0383 09415  
SPECIMEN  
SPECIMEN  
SPECIMEN  
Round specimen  
holder,  
for base plate  
with dovetail  
guide  
25 x 15 perforations  
0364 26815  
30 x 15 perforations  
0364 26816  
0383 08992  
Configurations:  
1. Base plate with dovetail guide (0383 09422) and holder for round specimens (0383 08992)  
= order no. 0383 08993  
2. Base plate with dovetail guide (0383 09422) and specimen clamp 44 x 58 mm (0383 09415) with eccentric ring (0383 09425)  
= order no. 0383 20212  
3. Vacuum stage 76 x 25 mm with vacuum stage adapter (0383 60407) for base plate with vise (0383 09267)  
= order no. 0364 18200  
Leica SP 2600 – Ultramilling attachment  
29  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.  
Operation  
5.5 Clamping specimen  
For a detailed overview of the clamping systems available, please  
refer to chapter, 5.4, page 29.  
5.5.1 Installing the base plates in the basic instrument  
For detailed instructions on how to install the base plates in the  
basic instrument, please refer to chapters 5.5/5.5.1, pages 38/39 of  
the Leica SM2500 instruction manual.  
When working with the Leica SP2600 ultramilling attachment, al-  
ways insert the eccentric ring prior to installing the base plates.  
The eccentric ring blocks the specimen orientation mechanism  
thus providing enhanced stability during milling.  
For details on how to insert the eccentric ring, please refer to  
chapter 5.5/5.5.1, page 39 of the Leica SM2500 instruction manual.  
5.5.2 Clamping the specimen holders in the base plate with dove tail guide  
For detailed instructions on how to clamp the specimen clamp 44 x  
58 mm and/or the round specimen holder in the base plate with  
dovetail guide as well as on how to clamp specimens in these  
specimen holders, please refer to chapter 5.5.4 ‘Mounting speci-  
men holders onto the base plate with dovetail guide’, pages 42 - 44  
of the Leica SM2500 instruction manual.  
Supplementary instruction manual V 1.2 – 07/2002  
30  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
5.5.3 Clamping the vacuum stages on/in  
the base plates  
Clamping the vacuum stages in the base plate  
with vise (Fig. 5.15)  
Fasten the vacuum stage on the vacuum stage  
adapter for the base plate with vise (with 4  
Allen screws (1) / Allen key size 4).  
The 4 size-four Allen screws are part of the  
‘Vacuum stage adapter’ standard delivery, the  
size-four Allen key comes as part of the Leica  
SP2600 standard delivery.  
Clamp the vacuum stage in the base plate with  
vise.  
For detailed instructions on how to do this,  
please refer to chapter 5.5.2, page 41 ‘Clamp-  
ing specimen stages’ of the Leica SM2500 in-  
struction manual: the vacuum stage is  
clamped the same way as the specimen  
stages described there.  
Fig. 5.15  
1
Clamping the vacuum stages in the base plate  
with dovetail guide (Fig. 5.16)  
Fasten the vacuum stage on the vacuum stage  
adapter for the base plate with dovetail guide  
(with 4 Allen screws (1) / Allen key size 3).  
The 4 size-three Allen screws are part of the  
‘Vacuum stage adapter’ standard delivery, the  
size-three Allen key comes as part of the Leica  
SM2500 standard delivery.  
Clamp the vacuum stage in the base plate with  
dovetail guide.  
For detailed instructions on how to do this,  
please refer to chapter 5.5.4, page 44 ‘Clamp-  
ing the round specimen holder’ of the Leica  
SM2500 instruction manual: the vacuum stage  
is clamped the same way as the round speci-  
men holder.  
Fig. 5.16  
Leica SP2600 – Ultramilling attachment  
31  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.  
Operation  
5.5.4 Connecting the vacuum pump /  
placing specimens onto the vacuum  
stages  
IMPORTANT:  
Prior to first use, every vacuum table  
surface must be milled once, just like  
a specimen surface. This ensures that  
the surface of the specimens milled on  
the table later on will be perfectly  
even.  
To mill a table surface, carry out one  
entire pre- and finishing milling cycle  
as per chapters 5.6 - 5.8, pages 34 - 43  
of this manual.  
Connecting the vacuum pump  
Fig. 5.17  
1 2  
Place vacuum pump (make sure the voltage  
rating of the vacuum pump is appropriate for  
the mains supply in your laboratory!) on the  
floor underneath the microtome bench.  
Connect vacuum pump to mains.  
Slide connecting hose (1) (part of standard  
delivery of Leica vacuum pumps) onto the  
nozzle on vacuum pump (2).  
Connect the other end of connecting hose (1)  
to the nozzle (3) of the vacuum stage.  
Place the specimen onto the vacuum stage.  
There are different types of vacuum  
stages for different specimen sizes -  
see next page.  
Fig. 5.18  
Supplementary instruction manual V 1.2 – 07/2002  
32  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Placing specimens onto vacuum stage for 30 -  
50-mm lids  
Fasten a lid suited to the size of the specimen  
to be milled onto the vacuum stage (slotted  
screws are part of standard delivery of  
vacuum stage lids - use slotted screwdriver).  
Place the specimen (1) onto the sintered sur-  
face (2) of the lid.  
Switch on the vacuum pump.  
Placing specimens onto vacuum stages for per-  
forated specimen holders  
Fig. 5.19  
1
2
Fasten a perforated specimen holder (3) suited  
to the size of the specimen on the vacuum  
stage 76 x 25 mm (4) or onto the vacuum stage  
with variable surface (not shown here).  
Place specimen on the perforated area of the  
vacuum stage  
Switch on the vacuum pump.  
Fig. 5.20  
3
The following is recommended for all  
vacuum stages, no matter of which  
type and size:  
If the specimen is smaller than the sin-  
tered or perforated stage surface, the  
entire stage surface not covered by the  
specimen should be covered with  
autoadhesive plastic film.  
Placing specimens onto vacuum stage 55 x 100  
mm  
Place the specimen directly on the sintered  
surface area of the vacuum stage.  
(Vacuum stage 55 x 100 mm not shown here).  
Fig. 5.21  
4
Leica SP2600 – Ultramilling attachment  
33  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.  
Operation  
5.6 Preparatory steps prior to the milling procedure  
For a detailed description of all control elements located on the  
control unit, please refer to chapters 5.3 (‘Overview - control  
panel’) and 5.4 (‘Description of each individual control element’) of  
the Leica SM2500 instruction manual.  
Select manual mode of operation MAN (button (5) - chapter 5.4.3, page  
26, Leica SM2500 instruction manual).  
In ‘MAN’ mode, carry out the following preparatory steps:  
Move the milling spindle to the upper limit position (button (48), chapter  
5.4.13, p. 36, Leica SM2500 manual) and open the plexiglass cover.  
Attached the miller holder (1) (containing the adjusted pre-miller) to the  
milling spindle (use Allen screws size 5 (2) / Allen key size 5 - part of  
SP2600 standard delivery).  
Always slide plexiglass cover to upper position prior to attaching  
the miller holders to the milling spindle!  
1
2
Fig. 5.22  
Supplementary instruction manual V 1.2 – 07/2002  
34  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Insert the specimen (see chapters 5.5.2 and 5.5.4, pages 30 and 33 of this  
manual) or - if you need to mill a vacuum table prior to first use - see  
chapter 5.5.3, page 31 of this manual).  
Wear protective gloves when working with potentially infectious  
specimen material!  
Move the specimen sledge underneath the miller (buttons (56) or (57),  
chapter 5.4.14, page 37, Leica SM2500 instruction manual).  
or  
Set the milling window (button (61) and others - see chapter 5.4.14, page  
37, Leica SM2500 instruction manual).  
Press button (51) (chapter 5.4.13, page 36, Leica SM2500 manual) to bring  
the cutting edge of the miller close to the specimen surface (coarse ad-  
justment - small gap must still remain).  
Always slide the plexiglass cover to the upper position prior to the  
distance check ‘milling edge / specimen surface’! - Risk of injury!  
Next, use buttons (50) and/or (49) (see chapter 5.4.13, page 36, Leica  
SM2500 instruction manual) for fine adjustment of the distance between  
milling edge and specimen surface.  
Check the remaining distance in three different zones of the specimen  
surface (beginning/center/end) moving the specimen sledge via buttons  
(56) - (59), see chapter 5.4.14, page 37, Leica SM2500 manual) - the cut-  
ting edge of the miller should just about touch the specimen surface.  
or  
Caution:  
Always check the distance between miller cutting edge and  
specimen surface thoroughly prior to milling!  
Under no circumstances may the miller cutting edge be positioned  
more than a few microns (= selected milling thickness) below the  
specimen surface when starting the milling procedure.  
If the surface height of the specimen is uneven, the cutting edge  
should just about touch the highest point of the specimen. - Then  
mill the specimen until the surface is completely level.  
Leica SP 2600 – Ultramilling attachment  
35  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.  
Operation  
5.7 Pre-milling  
Select automatic mode of operation ‘AUTO’ (button (5) - chapter 5.4.3, page  
26 Leica SM 2500 instruction manual).  
In ‘AUTO’ mode, set the milling parameters for the pre-milling procedure:  
Select STOP or PHOTO position (buttons (23) or (26) chapter 5.4.8, page  
31 - Leica SM2500 instruction manual).  
or  
Select milling thickness (button (10) / rotary selector switch (9) - see  
chapters 5.4.4 and 5.4.5, pages 27/28 - Leica SM2500 instruction manual).  
(--> see following page for recommended settings).  
Activate retraction (button (10) / rotary selector switch (9) - see chapters  
5.4.4 and 5.4.5, pages 27/28 - Leica SM2500 instruction manual).  
+
Select milling and return stroke speed.  
(button (19) / rotary selector switch (9) - see chapter 5.4.4, page 27 and  
chapter 5.4.7, page 30 - Leica SM2500 instruction manual).  
+
(see next page for recommended settings).  
Close plexiglass protective cover.  
Start miller rotation (button (37) - chapter 5.4.11, page 34 - Leica SM2500  
instruction manual).  
Select rotational speed of miller (buttons (39), (40) - chapter 5.4.11, page  
34 - Leica SM2500 instruction manual).  
or  
Select desired milling mode (button (15) - chapter 5.4.6. page 29 - Leica  
SM2500 instruction manual).  
Individual parameter settings depend on specimen material and  
size. Please refer to the following page for a list of recommended  
settings.  
Supplementary instruction manual V 1.2 – 07/2002  
36  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Recommended milling thickness settings for pre-milling  
10 µm - keep milling until surface is completely level.  
5 µm - three entire sledge strokes (milling and return stroke).  
3 µm - three entire sledge strokes (milling and return stroke).  
Do not exceed a milling thickness setting of 5 µm with the follow-  
ing materials: brittle, very ductile materials such as titanium or fi-  
berglass and compound samples containing such materials (such  
as printed circuit boards).  
Recommended sledge speed settings for pre-milling  
Milling stroke: 1.5 mm/s  
Return stroke: any speed  
Recommended rotational speed of miller for pre-milling  
2,000- 3,000 rpm (depending on specimen type)  
Recommended mode of operation for pre-milling  
Select ‘continuous’ mode of operation.  
Starting the milling procedure  
During milling:  
• switch on the vacuum cleaner,  
• wear appropriate safety goggles,  
• wear appropriate mask (type of mask depends on the material  
being milled / size of the chips that form during milling).  
Goggles and mask prevent milling chips and microdust from enter-  
ing your eyes and from being inhaled.  
Switch on the vacuum cleaner (via button (35) on the control unit - see  
chapter 5.4.11, page 34 of Leica SM2500 instruction manual) or via the  
on/off switch of the vacuum cleaner itself.  
Start sledge stroke (buttons (29), (30) - chapter 5.4.10, p. 33 - Leica  
SM2500 instruction manual).  
+
Carry out the milling steps as per above recommendation.  
--> See next page for a perspective view of the pre-milling procedure.  
Leica SP 2600 – Ultramilling attachment  
37  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.  
Operation  
5.7.1 Perspective view of the pre-milling procedure  
1
2
Fig. 5.23  
Pre-milling procedure  
The milling spindle carrying the pre-miller rotates clockwise (circular arrow  
(1)) while the specimen sledge passes underneath the miller in the direction  
of arrow (2).  
The pre-miller leaves a serrated surface texture which is then in a second  
milling procedure milled to a mirrorlike, even surface by the finishing miller  
(see Fig 5.24 p. 42).  
Supplementary instruction manual V 1.2 – 07/2002  
38  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Ending the pre-milling procedure  
Press RUN STOP or RUN ENABLE to stop the sledge (buttons (29), (30) -  
Chapter 5.4.10, page 33 - Leica SM2500 manual) - or, when working in  
PROG’ mode of operation, wait for the programmed number of strokes to  
be finished (see chapter 5.4.3, page 26, Leica SM2500 manual).  
or  
If the specimen sledge has not stopped in the front limit position, activate  
manual mode of operation MAN (button (5) - chapter 5.4.3, page 26,  
Leica SM2500 manual) and move the specimen to the front limit position  
(via buttons (58) or (59) - chapter 5.4.14, page 37- SM2500 manual) .  
Stop miller rotation (button (37) - chapter 5.4.11, page 34 - Leica SM2500).  
After switching off the milling spindle (button (37) - chapter. 5.4.11,  
p. 34 - instruction manual Leica SM2500) the miller continues to  
rotate for a little while!  
Risk of injury!  
Slide plexiglass protective cover to upper position.  
Never touch the milling spindle while it is rotating and/or while  
the plexiglass cover is shut, no matter whether the spindle is ro-  
tating or not.  
Remove the miller holder containing the pre-miller.  
Leica SP 2600 – Ultramilling attachment  
39  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.  
Operation  
5.8 Finishing milling  
Attach the miller holder containing the finishing miller to the milling  
spindle (Allen screws / Allen key size 5) - see Fig. 5.21, page 34).  
Always slide the plexiglass cover to the upper limit position prior  
to attaching the miller holders to the milling spindle!  
Bring the cutting edge of the miller close to the specimen surface and  
thoroughly check the distance between miller cutting edge and speci-  
men surface as described for the pre-miller on page 35.  
If necessary, adjust the milling window (button (61) and others - see  
chapter 5.4.14, page 37 of Leica SM2500 instruction manual).  
Setting the milling parameters for the finishing milling procedure  
Once again activate automatic mode of operation ‘AUTO’ and set the  
milling parameters using the same control unit buttons as described for  
the pre-miller on page 36.  
Individual parameter settings depend on specimen material and  
size. Please refer to the recommendations below and on the fol-  
lowing page.  
Recommended milling thickness settings for finishing milling  
5 µm - keep milling until surface is completely level.  
3 µm - three entire sledge strokes (milling and return stroke).  
2 µm - three entire sledge strokes (milling and return stroke).  
1 µm - three entire sledge strokes (milling and return stroke).  
Do not exceed a milling thickness setting of 5 µm with the follow-  
ing materials: brittle, very ductile materials such as titanium or fi-  
berglass and compound samples containing such materials (such  
as printed circuit boards).  
Supplementary instruction manual V 1.2 – 07/2002  
40  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Recommended sledge speed settings for finishing milling  
Milling stroke: 0.5 mm/s to 1.0 mm/s  
Return stroke: any speed  
Recommended rotational speed of miller for finishing milling  
< 2.000 rpm  
Recommended mode of operation for finishing milling  
Select ‘continuous’ mode of operation.  
Starting the milling procedure  
During milling:  
• switch on the vacuum cleaner,  
• wear appropriate safety goggles,  
• wear appropriate mask (type of mask depends on the material  
being milled / size of the chips that form during milling).  
Goggles and mask prevent milling chips and microdust from enter-  
ing your eyes and from being inhaled.  
Switch on the vacuum cleaner (via button (35) on the control unit - see  
chapter 5.4.11, page 34 of Leica SM2500 instruction manual) or via the  
on/off switch of the vacuum cleaner itself.  
Start sledge stroke (buttons (29), (30) - chapter 5.4.10, p. 33 - Leica  
SM2500 instruction manual).  
Carry out the milling steps as per above recommendation.  
+
Important note:  
If you need to mill several specimens with identical results (e.g.  
for examining a specimen layer located at a certain level), com-  
pletely identical parameters have to be selected for each speci-  
men (for pre-milling as well as for the finishing milling procedure).  
--> See following page for a perspective view of the finishing milling proce-  
dure.  
Leica SP 2600 – Ultramilling attachment  
41  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5.  
Operation  
5.8.1 Perspective view of the finishing milling procedure  
1
2
Fig. 5.24  
Finishing milling procedure  
The milling spindle carrying the finishing miller rotates clockwise (circular  
arrow (1)) while the specimen sledge passes underneath the miller in the di-  
rection of arrow (2).  
The serrated surface structure created by the pre-miller (see Fig 5.23 p. 38)  
is converted by the finishing miller into a mirrorlike, completely even sur-  
face.  
Supplementary instruction manual V 1.2 – 07/2002  
42  
Download from Www.Somanuals.com. All Manuals Search And Download.  
5. Operation  
Ending the finishing milling procedure  
Press RUN STOP or RUN ENABLE to stop the sledge (buttons (29), (30) -  
chapter 5.4.10, page 33 - Leica SM2500 manual) - or, when working in  
PROG’ mode of operation, wait for the programmed number of strokes to  
be finished (see chapter 5.4.3, page 26, Leica SM2500 manual)  
If the specimen sledge has not stopped in the front limit position, activate  
manual mode of operation MAN (button (5) - chapter 5.4.3, page 26, Leica  
SM2500 manual) and move the specimen to the front limit position (via  
buttons (58) or (59) - chapter 5.4.14, page 37- SM2500 manual) .  
or  
Stop miller rotation (button (37) - chapter 5.4.11, page 34 - Leica SM2500).  
After switching off the milling spindle (button (37) - chapter. 5.4.11,  
p. 34 - instruction manual Leica SM2500) the miller continues to  
rotate for a little while!  
Risk of injury!  
Remove the specimen for subsequent microscopic examination.  
Slide plexiglass protective cover to upper position.  
Never touch the milling spindle while it is rotating and/or while  
the plexiglass cover is shut, no matter whether the spindle is rotat-  
ing or not.  
Remove the finishing miller.  
5.9 Switching the instrument off (switching off the control unit)  
Once all milling procedures have been completed or at the end of your  
workday, switch off the mains switch at the rear of the control unit.  
When switching off the mains switch on the control unit (see  
chapter 5.2, p. 21 - Instruction manual Leica SM2500) the milling  
spindle will rotate briefly.  
Therefore, never switch off the control unit, while one of your  
hands is in the rotation zone of the miller.  
- Risk of injury!.  
Leica SP 2600 – Ultramilling attachment  
43  
Download from Www.Somanuals.com. All Manuals Search And Download.  
6.  
Troubleshooting  
6.  
Troubleshooting  
For all instructions on error detection and trouble shooting, please refer to  
chapter 6, page 64 of the Leica SM2500 instruction manual .  
Supplementary instruction manual V 1.2 – 07/2002  
44  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Cleaning, disinfection and maintenance  
7.1 Cleaning and disinfection  
Only authorized service technicians may access the internal com-  
ponents of the instrument for service and repair!  
Cleaning detergents appropriate for the ultramilling attachment -  
varnished surfaces:  
• mild laboratory detergents (slightly moistened cloth)!  
non-varnished metal parts:  
• mild laboratory detergents,  
• xylene substitutes (slightly moistened cloth)  
• alcohol (slightly moistened cloth).  
No liquids may enter the interior of the instrument during cleaning!  
Cleaning  
Clean the basic instrument Leica SM2500 as per the instructions in chap-  
ter 7.1, pages 65 - 67 of the Leica SM2500 instruction manual.  
Clean the Leica SP2600 with a soft, slightly moistened cloth using one of  
the detergents listed above.  
Clean diamond millers with a soft cloth moistened with alcohol.  
For disinfection of the instrument and accessories use Leica Cryo-  
fect or other commercial disinfectants. - Use all disinfectants ac-  
cording to the manufacturer’s instructions!  
No liquids may enter the interior of the instrument during disinfec-  
tion!  
When disinfecting, wear appropriate goggles and mask as well as  
protective gloves!  
Dispose of potentially infectious specimen material as per the  
laboratory regulations applicable in the country of operation!  
Disinfection  
Remove potentially infectious milling waste / remaining samples and dis-  
pose of as per the regulations applicable.  
Disinfect instrument and accessories as per the instructions of the disin-  
fectant manufacturer.  
Leica SP 2600 – Ultramilling attachment  
45  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Cleaning, disinfection and maintenance  
7.2 Maintenance  
Only authorized service technicians may access the internal com-  
ponents of the instrument for service and repair!  
No liquids may enter the interior of the instrument during mainte-  
nance!  
Carry out the maintenance steps to be done by the user as per the in-  
structions in chapter 7.2, pages 67 - 70 of the Leica SM2500 instruction  
manual.  
In addition to the maintenance steps listed in the Leica SM2500 manual,  
the following maintenance work is required for the Leica SP2600:  
chapter 7.2.1 - Resharpening the millers.  
chapter 7.2.2 - Replacing the light bulb in the calibration device.  
7.2.1 Resharpening the millers  
The millers need to be resharpened as soon as the milled specimen sur-  
faces no longer have a smooth, mirrorlike appearance and/or when mill-  
ing residues / residues of a grimy appearance start to accumulate on the  
specimen surface.  
Supplementary instruction manual V 1.2 – 07/2002  
46  
Download from Www.Somanuals.com. All Manuals Search And Download.  
7. Cleaning, disinfection and maintenance  
7.2.2 Replacing the light bulb in the calibration device  
Remove base cover (2) of calibration device (1) (unscrew 4 slotted  
screws (3)).  
Loosen light bulb holder (5) (metal bridge with lamp socket) (unscrew 2  
slotted screws (4)).  
Take replacement light bulb (8) out of sponge rubber cube (7).  
Remove broken light bulb (6) from lamp socket and replace with new  
light bulb (8).  
Place another replacement bulb (8) into the sponge rubber cube (7).  
Put light bulb holder (5) back in place.  
Place ring with cable (9) onto bore (10) in the light bulb holder.  
Fasten light bulb holder with two screws (4).  
Put sponge rubber cube (7) with new replacement light bulb (8) back into  
the calibration device.  
Put base cover (2) back in place and fasten with 4 screws (3).  
7
1
8
5
6
3
10  
4
9
2
Fig. 7.1  
Leica SP 2600 – Ultramilling attachment  
47  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8. Overview - accessories  
8.1 Miller holders and millers  
Miller holder  
Holder for pre- and finishing miller  
Pre-miller  
Pre-miller D - with diamond edge, cutting angle -1°, clearance angle +3°  
Finishing miller  
Finishing miller D - with diamond edge, cutting angle +3°, clearance  
angle +4°, for milling plastics  
Finishing miller D - with diamond edge, cutting angle -8°, clearance angle  
+1.5°, for milling precious and nonferrous heavy metals  
Finishing miller D - with diamond edge, cutting angle -1°, clearance angle  
+1.5°, for milling light metals  
Finishing miller D - with diamond edge, cutting angle -5°, clearance angle  
+1.5°, for milling light metals  
Counterbalance insert - for Leica SP2600 - to be used instead of a pre- or  
finishing miller  
8.2 Base plates for specimen holders and vacuum stage adapters  
Base plate with vise  
Base plate with dovetail guide  
8.3 Specimen holders  
Specimen clamp 44 x 58 mm,  
for base plate with dovetail guide  
Round specimen holder, for base plate with dovetail guide  
8.4 Vacuum stage adapters  
Vacuum stage adapter for base plate with vise  
Vacuum stage adapter for base plate with dovetail guide  
Supplementary instruction manual V 1.2 – 07/2002  
48  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8. Overview - accessories  
8.5 Vacuum stages and matching lids / perforated specimen holders  
Vacuum stage for Leica SP2600 - basic body for lids 30 - 50 mm  
Lid, 30 mm, for Leica SP2600 vacuum stage  
Lid, 40 mm, for Leica SP2600 vacuum stage  
Lid, 50 mm, for Leica SP2600 vacuum stage  
Vacuum stage for Leica SP2600 - 55 x 100 mm  
Vacuum stage, 76 x 25 mm  
Available either configured (with vacuum stage adapter for base plate  
with vise) or without vacuum stage adapter.  
To be used with several different perforated specimen holders:  
specimen holder, 5 x 5 perforations  
specimen holder, 5 x 10 perforations  
specimen holder, 10 x 10 perforations  
specimen holder, 10 x 15 perforations  
specimen holder, 15 x 15 perforations  
specimen holder, 20 x 10 perforations  
specimen holder, 20 x 15 perforations  
specimen holder, 25 x 15 perforations  
specimen holder, 30 x 15 perforations  
Vacuum stage, with variable surface: 76 x 25 mm, 76 x 56 mm  
and 100 x 65 mm.  
To be used with the same perforated specimen holders as vacuum stage  
76 x 25 mm.  
8.6 Vacuum pumps  
Vacuum pump, 230 V, 50 Hz  
Vacuum pump, 120 V, 60 Hz  
Leica SP 2600 – Ultramilling attachment  
49  
Download from Www.Somanuals.com. All Manuals Search And Download.  
8
Overview - accessories  
8.7 Additional accessories  
Vacuum cleaner and vacuum cleaner accessories  
Vacuum cleaner ‘Fakir S20’, with connection hose  
230 V / 50 Hz only  
Set of vacuum cleaner bags for Fakir S 20  
Illumination  
Halogen spotlight illumination  
230 V/50 Hz only  
Supplementary instruction manual V 1.2 – 07/2002  
50  
Download from Www.Somanuals.com. All Manuals Search And Download.  
9. Warranty and service  
9.  
Warranty and service  
For all information on warranty and service, please refer to chapter 9, pages  
74 - 76 of the Leica SM2500 instruction manual.  
Leica SP 2600 – Ultramilling attachment  
51  
Download from Www.Somanuals.com. All Manuals Search And Download.  

Lathem GPS Receiver ATX GPS User Manual
Lincoln Welder IM985 User Manual
Macurco Carbon Monoxide Alarm DVP 120 User Manual
Magnavox TV VCR Combo MDV560VR 17 User Manual
Makita Impact Driver 6910 User Manual
Makita Impact Driver LXDT06 User Manual
Manitowoc Ice Griddle C0836 72 User Manual
McAfee Automobile Accessories ASH300 User Manual
Melissa Microwave Oven 753 123 User Manual
Micro Innovations Mouse KB575M User Manual